User Manual
en
Table of Contents
Warning! MACOSEAL twin must be used in compliance with all specifications and operational procedures
listed in this manual.
Warning! MACOSEAL twin must be used under the continuous control of trained persons.
Warning! Please read the User Manual carefully before use.
Warning! Follow the operating instructions while operating MACOSEAL twin.
Warning! Cables and accessories others than those specified, may result in increased emission or
decreased immunity of the equipment or system.
Warning! The equipment or system should not be used adjacent to or stacked with other equipment. If
adjacent or stacked use is necessary, the equipment or system should be observed to verify
normal operation in the configuration in which it will be used.
Warning! If any of the components of MACOSEAL twin are exposed to blood, they must be cleaned with
an appropriate disinfectant solution (see chapter five).
Warning! Disconnect MACOSEAL twin from power source before performing any maintenance and
cleaning procedure.
Caution! Portable and mobile RF communication equipment can affect MEDICAL ELECTRICAL
EQUIPMENT, See table 2, 4 and 6, page 4 - 6, for guidance.
Caution! Inspect all parts of the instrument for the appearance of defects before use.
Caution! To guarantee the safe use, only accessories delivered with MACOSEAL twin can be used.
Caution! If multiple seals are made along a length of tubing they should be at least 1/2 inch (1 cm) apart.
Multiple seals made closer than the recommended distance may result in rupture of the sealed
tubing segment
Caution! In case of sealer malfunction (intermittent operation, poor quality seals, sealing time seems too
long or too short) call the technical service for assistance.
Caution! Periodically check the pattern of the sealing visually (see section 4.2)
Note! Contact MacoPharma Technical Services for the disposal of the MACOSEAL twin (and
MACOSEAL_UME_04
1.1 Introduction
MACOSEAL twin is a fully automatic system for sealing PVC tubes, especially tubes in blood pack systems. After
the sealing procedure, the tube is easily pulled apart due to the distinct sealing pattern, with no damage to the
blood inside the tubes.
The base unit is connected by the cable to the hand unit, where the sealing takes place when the trigger is
pressed. Different types and sizes of tubes can be used, and the necessary sealing time is self-adaptive to fit the
tubes that are being used.
MACOSEAL twin uses radio frequency (RF) energy to generate heat for sealing. Users are requested to be
cautious of potential electrical shocks or hazards while handling this sealer. When the hand unit is connected to
the base unit, always keep your fingers away from the electrodes in the slot. Never place any object other than
the PVC tube between the electrodes.
Table 2:
< 5% Uτ < 5% Uτ
(> 95% dip in Uτ) (> 95% dip in Uτ)
for 5 sec for 5 sec
Power frequency 3 A/m 3 A/m Power frequency magnetic fields
(50/60 Hz) magnetic should be at levels characteristic of a
field typical location in a typical
commercial or hospital environment
MACOSEAL_UME_04
IEC 61000-4-8
NOTE Uτ is the a.c. mains voltage prior to application of the test level.
Recommended separation
distance
Conducted RF 3 Vrms 3V d = 1.17 √P
IEC 61000-4-6 150 kHz to 80 MHz
NOTE 1 At 80MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
the 36200-00 is used exceeds the applicable RF compliance level above, it should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the 36200-00.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m
MACOSEAL_UME_04
NOTE 1 At 80MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
MACOSEAL_UME_04
Handle connectors
Power Switch
Power Inlet
Type Label
Release lock.
(One on each side)
List of all cable and maximum lengths of cable, transducers and other accessories with which the manufacturer
claims compliance (cables, accessories used other than those listed may impact emission/immunity)
Item Specification
9 362 98100 Europe (supplied as Three-wire (10A), 2.5 m long (8ft), female plug
Power cable
standard)
Hand unit - Multicolour LED, indicates Green at normal power, Orange during seal, Red during
error indication.
- Indicator lights red (fast blink) during one second in case of short circuit, sealing
function is blocked.
- Indicator lights red (slow blink) once a second in case of over heating, sealing function
is blocked.
- Type and serial number.
MACOSEAL_UME_04
3. The above list is subject to change, refer to the packing list for accurate description of contents.
If parts are missing or if parts are damaged, report it immediately.
4. Please keep all shipping and packaging materials, as they may be required for later transportation, at least
during the warranty period.
1. Connect the handle cable, blue at end of cable, to one of the connectors on the base unit front.
Make sure connector snaps in secured position.
2. Insert the power cable into the power entry module on the back panel of the base unit.
3. Plug the power cable into an alternating current outlet.
4. Perform a test seal on an empty or water filled tube to ensure proper operation.
Warning! The instrument must always be connected to an outlet with protected ground and from appropriate
alternating current mains source, 100 to 230 V ~.
Caution! Ensure that all cables are connected before turning the power on.
1. Connect the handle cable, blue at end of cable, to the available connector on the base unit front.
Make sure connector snaps in secured position.
2. Perform a test seal on an empty or water filled tube to ensure proper operation.
Warning! Ensure that the power switch is turned off before installing additional handle.
MACOSEAL_UME_04
During the whole sealing process, a light on the front of the hand unit lights orange. When this light goes to
green, the blue trigger may be released, and the sealing is complete. The intelligent sense control in the unit
detects controls and adjusts the sealing process to give the best sealing quality for the type of tubes that are
being used.
See description in section 1.4 for the different light indications on the units in case of problem.
The Hand unit consists of ergonomic handgrip with trigger and electrode (which can be removed for cleaning).
The handle has integrated cable for connection to base unit.
The Base unit consists of the power supply module and RF-generator module with control circuits.
MACOSEAL_UME_04
Caution! Inspect all parts of the instrument for the appearance of defects before use.
Warning! This equipment emits a low level of electromagnetic (non-ionizing) radiation while sealing.
Misuse or direct contact between tissue and RF electrode can result in severe RF burns.
It should not be used near high frequency sensitive electronic equipment.
1. Put the tube to be sealed down to the bottom of the slot of the hand unit.
2. Check that the tube is placed between the electrodes in the slot.
3. Press the trigger to bring the two electrodes closer together until the orange light on front of the handle
comes on.
The sealing time is normally 0.4 to 1.5 seconds; after maximum of 6 seconds the RF is turned off.
Note! Do not pull the tube and keep the trigger fully pressed down until the orange light changes to
green.
4. When the orange light switches to green, sealing is finished, then release the trigger and remove the tubing.
5. The centre of the sealed pattern is very thin and pulling both sides will divide the tube into two pieces.
Note! If the user releases the seal button at any time, the sealing process stops.
Caution! If multiple seals are made along a length of tubing they should be at least 1cm (½ inch) apart.
Multiple seals made closer than the recommended distance may result in rupture of the sealed tubing
segment.
Caution! In case of sealer malfunction (intermittent operation, poor seals quality) call the technical service for
assistance.
Caution! If the sealing time seems too long or too short, call the technical service for assistance.
Caution! Periodically check the pattern of the sealing visually (see picture below).
Over heated (slow red blink). Let the unit cool down until the green light comes on.
If the sealer still doesn’t start, see chapter 6, for further information.
Warning! For you own safety always turn off the power switch and disconnect the Power cable.
It is advisable to wear medical gloves for cleaning drops of blood.
Caution! Blood and blood products must be treated as potentially infectious at all times. In the event of blood
spillages, appropriate protective clothing should be worn during clean-up procedures. After removing
residual biological material, surfaces which have been in contact with blood or components must be
disinfected using a chemical agent considered as a "sterilant" (e.g. isopropanol 70%). Alternatively, a
freshly prepared solution of diluted sodium hypochlorite (household bleach) may be used to disinfect
surfaces which will not be harmed by the solution. Diluted solutions from 1 part bleach to 10 parts
water may be used. Regardless of the “sterilant” or disinfection solution used, remember to remove
any residue to ensure that the surfaces of the equipment are not subject to corrosion or discoloration.
Discard all materials in contact with blood according to institutional policies regarding the disposal of
biohazardous materials.
Caution! Do not disinfect or sterilize any part of the sealer through autoclave or with ethylene oxide
gas, to do so will render the sealer unusable and invalidate the warranty.
Do not use chemical or abrasive cleaners such as acetone, ammonia or similar.
Do not use sharp edged tools for cleaning, which could damage the finish of the units.
Caution! Disconnect hand unit from base unit before performing any cleaning procedure.
Caution! Do not allow liquid to flow in the electronic part of the equipment.
Figure 1
Lock
Lock
Figure 2 Figure 3
Figure 4
Clean the hand unit and both electrodes with a soft lintfree cloth moistened with mild detergent.
The removed trigger with electrode can be washed under tap water.
Dry carefully and ensure that the electrodes are completely dry to prevent sparks.
MACOSEAL_UME_04
After cleaning, inspect the electrodes for any mechanical damage or wear out.
Do not use damaged part.
Figure 5 Figure 6
Locking Locks in
point (one place (one on
on each each side)
Figure 7 Figure 8
1. Place the trigger with electrode onto the
handle, ensure that the trigger is close to the
handle, see Fig 5
2. Make sure that the electrode is positioned so
as to be parallel with the fixed electrode and
aligned with the corresponding cut-in portion
on the handle, see fig 6.
3. Pull the trigger with electrode back on to the
handle until the locks on each side “click” fig 7
and fig 8
4. Press the trigger twice and ensure that the
electrode moves smoothly, is in position and
not loose.
LED doesn't light green No Voltage. Check Power Supply and cable.
Power switch off Check power switch.
Internal Fuse blown. Seek MacoPharma technical service for
assistance.
Other cause. Seek MacoPharma technical service for
assistance.
LED does not light or does not base unit inoperative. Seek MacoPharma technical service for
light at the same time as the assistance.
LED on the Hand Unit
Bad sealing. Active electrode doesn’t move Seek MacoPharma technical service for
smoothly on handle. assistance.
Rim on active electrode Change hand unit or seek MacoPharma
damaged. technical service for assistance.
Other cause. Seek MacoPharma technical service for
assistance.
Hard to divide tube after seal Cable broken. Seek MacoPharma technical service for
assistance.
Other cause. Seek MacoPharma technical service for
assistance.
Intermittent sealing. Hand unit defective. Seek MacoPharma technical service for
assistance.
LED lights orange after the trigger Hand unit defective. Seek MacoPharma technical service for
has been released. assistance.
LED doesn’t light. Cable broken. Seek MacoPharma technical service for
assistance.
Other cause. Seek MacoPharma technical service for
assistance.
MACOSEAL_UME_04
7.1 Warranty
MacoPharma warrants that the equipment shall be free from defects in material and workmanship when delivered to the
original purchaser. MacoPharma’s sole obligation shall be limited to the repair or replacement, at MacoPharma’s option and
expense, of the defective part or unit for a period of one (1) year following the date of initial delivery.
The warranty extends only to the original purchaser and is not assignable or transferable, and shall not apply to auxiliary
equipment, or disposable accessories.
MacoPharma warrants that the equipment is fit for the purposes and indications described in the labeling when used in
accordance with the directions for use. Unless the equipment is used in accordance with these instructions, this warranty is
void and of no effect. No other expressed or implied warranty exists, including any warranty of merchantability or fitness for a
particular purpose. MacoPharma’s sole obligation and purchaser’s exclusive remedy for breach of warranty shall be limited
to the repair or replacement at MacoPharma’s option. MacoPharma shall not be liable for proximate, incidental or
consequential damages. Modifications, alterations, recalibrations or abuse, and service by other than a MacoPharma
authorized representative will void the warranty.
7.2 Service
Service under warranty period
While under MacoPharma warranty, the instrument must not be opened by unauthorized personnel.
Contact MacoPharmaTechnical Service for service and repair information for all MACOSEAL twin instruments.
Shipping costs for all units returned to MacoPharma shall be paid by the customer. The unit must be packed in its original
box or in another MacoPharma approved box that will provide adequate protection during shipment. To ensure prompt
return, a MacoPharmaTechnical Service representative must be notified before shipping any unit for repair. When calling
MacoPharma Technical Service, please be prepared to provide part number and serial number of the unit. A service request
number will be issued and should accompany all communications. A brief written description of the problem should be
attached to the instrument when it is returned for service. MacoPharma will not be responsible for unauthorized returns or for
units damaged in shipment due to improper packing.
Service after warranty period
After MacoPharma warranty period, the MACOSEAL twin will continue to be serviced by MacoPharmaTechnical Service.
If the instrument has been purchased by thecustomer, he/she will have the choice, either to have the service ensured by
MacoPharma Technical Service, or to perform it themselves. Please contact MacoPharma Technical Service to dispose of
the instrument.
If customers want to perform the service themselves, MacoPharma Technical Service will make available on request the
service manual including non confidential information (as block diagrams, component part lists, descriptions, calibration
instructions), periodic preventive maintenance guide, and any other non confidential information which will assist the
customer’s appropriately qualified technical personnel to repair those parts of equipment which are designated by the
manufacturer as repairable.
Product Disposal
For MACOSEAL twin disposal, please ensure the following:
• Do not dispose MACOSEAL twin as unsorted municipal waste.
• Collect the MACOSEAL twin separately.
• Use the collection and return systems available to you.
For more information on return, recovery or recycling of MACOSEAL twin, please contact your local MacoPharma
representative.
Preventive Inspection and Maintenance
At least every second year, the MACOSEAL twin must be fully inspected by:
- either the qualified service organization of the customer,
- or by MacoPharma Technical Service.
Interchangeable Parts Replacement (User level maintenance)
User level maintenance is limited to changing base unit or hand unit.
MACOSEAL_UME_04
MACOSEAL_UME_04