Download as pdf or txt
Download as pdf or txt
You are on page 1of 268

ZEISS

S7 / OPMI® PROergo®

Instructions for Use


G-30-1435-en
Version 11.0
2018-11-30
S7 / OPMI PROergo

About This Manual These Instructions for Use are part of the delivery scope.
• Please read them carefully before using the device.
• Keep them at the site of use of the device.
• Store them for the entire service life of the device.
• Pass them on to any subsequent owner or user of the device.

Help with orientation – At the beginning of these Instructions for Use there is a summary of the
different sections; this provides an overview of all the topics.
– The contents of each chapter are specified in detail at the beginning of
each chapter.
– The index facilitates searching via keywords.

Scope The following Instructions for Use apply to OPMI PROergo with the following
suspension systems and identifications:
– S7 floor stand, reference number: 000000-1087-851
– S7 ceiling mount, rigid, reference number: 000000-1103-022
– S7 ceiling mount with lifting column, reference number: 000000-1226-405
– S7 wall mount with short arm, reference number: 000000-1193-109
– S7 wall mount with long arm, reference number: 000000-1254-579
– S7 Centro suspension system, reference number: 000000-1254-580

Trademarks OPMI und PROergo is a trademark/are trademarks or registered trademark(s)


of Carl Zeiss Meditec AG or other companies of the ZEISS Group in Germany
and/or other countries.

Manufacturer's details Carl Zeiss Meditec AG


Goeschwitzer Strasse 51-52 Fax: + 49 (0) 7364 - 20 4823
07745 Jena Email: info.meditec@zeiss.com
Germany Internet: www.zeiss.com/med

Subject to changes in design and scope of delivery, as well as ongoing


technical development. Printed in Germany.

Copyright © Carl Zeiss Meditec AG 2018


All rights reserved. Reprints and reproductions, in whole or in part, are
prohibited. However, these Instructions for Use may be digitized and archived
for internal use. Access to this data by third parties is not permissible and must

Version 11.0
Page 2 G-30-1435-en
S7 / OPMI PROergo

be prevented.

Version 11.0
G-30-1435-en Page 3
S7 / OPMI PROergo

Version 11.0
Page 4 G-30-1435-en
S7 / OPMI PROergo

Summary of the section

Section: Safety measures 7

Section: Device overview 57

Section: Preparations for use 139

Section: Operation 179

Section: What to do in the event of malfunctions 197

Section: Care and maintenance 213

Section: Device data 227

Section: Keyword index 259

Version 11.0
G-30-1435-en Page 5
S7 / OPMI PROergo

Version 11.0
Page 6 G-30-1435-en
S7 / OPMI PROergo Safety measures

Safety measures

Key to symbols ................................................................................. 9


Hazard symbols ..........................................................................................9
Information symbols ...................................................................................9

Target group ..................................................................................10

Field of use ..................................................................................... 10


Intended use ............................................................................................10
Normal use ...............................................................................................10
Reasonably foreseeable misuse .................................................................11

Liability and warranty ..................................................................... 11

Notes for the operator .................................................................... 12

Requirements for operation ............................................................15


Prior to the very first use ...........................................................................15
Before every use .......................................................................................16
During use ................................................................................................18
After every use .........................................................................................19
Risk of burn injuries caused by high illumination intensity .........................20
Stand safety equipment ............................................................................23

Symbols and labels on the device ................................................... 28


Labels on the microscope .........................................................................28
Labels on the connection panel ................................................................31
Labels on S7 floor stand ...........................................................................32
Additional labels on S7 floor stand (instrument tray option) ......................38
Labels on S7 ceiling mount .......................................................................40
Additional label on S7 ceiling mount with lifting column ..........................44

Version 11.0
G-30-1435-en Page 7
Safety measures S7 / OPMI PROergo

Labels on S7 wall mount and on S7 Centro suspension system ................ 46


Additional labels on S7 wall mount .......................................................... 51
Labels on the light sources of the suspension systems .............................. 52
Labels on the wall power outlet (optional) ............................................... 55

Version 11.0
Page 8 G-30-1435-en
S7 / OPMI PROergo Safety measures

Key to symbols

We would like to inform you of the safety aspects that must be observed
when handling this device. This chapter contains a summary of the most
important information concerning matters relevant to instrument safety.

Hazard symbols

The following safety information has been incorporated into the Instructions
for Use. Please note this information and be particularly careful in these cases.

WARNING Indicates a hazard which may lead to fatal injuries or severe injuries if
precautions are not followed.
CAUTION Indicates a hazard which may lead to moderately severe injuries if
precautions are not followed.
NOTE Indicates a hazard which may lead to minor injuries or property damage
if precautions are not followed.

Information symbols

The following information symbols are used in these Instructions for Use:
– Lists
 Requirements for an action
• Prompt for action
 Result of an action
Additional information and hints

Version 11.0
G-30-1435-en Page 9
Safety measures S7 / OPMI PROergo

Target group

These Instructions for Use are intended for physicians, nurses and other
medical and technical staff who prepare, operate or maintain the device after
appropriate training. It is the duty of the device owner/operator to train and
brief all operating personnel.
Additional service activities are not part of these Instructions for Use. They will
be performed by staff specially trained for this purpose by ZEISS.

Field of use

Intended use

OPMI PROergo is a surgical microscope intended for the illumination and


magnification of the treatment and surgical area and for the support of
visualization in treatment and surgical procedures.

Normal use

OPMI PROergo is a microscope designed for dental treatment and for surgical
procedures in the field of dentistry.
Through its selection of illumination systems (halogen, xenon) and its
mechanical construction, it is adapted to the requirements of daily tasks
involved in treatment and surgery applications.

Version 11.0
Page 10 G-30-1435-en
S7 / OPMI PROergo Safety measures

Reasonably foreseeable misuse

CAUTION Injury to the patient's eye!


• The S7 / OPMI PROergo must not be used for ophthalmic procedures!
• Make sure that no light enters the patient's eyes.

CAUTION Live video images and recorded video sequences, video clips (cut sequences)
and single images shall not be used for diagnosis purposes. Visualized images
may contain variations in shape, contrast and color.

Liability and warranty

The warranty and liability depend on the contractually specified conditions.

Loss of warranty!
The manufacturer is not liable for damage caused by unauthorized
modifications to the device. Furthermore, such unauthorized modifications
will render the warranty and any claims under warranty invalid.

Version 11.0
G-30-1435-en Page 11
Safety measures S7 / OPMI PROergo

Notes for the operator

WARNING The device must not be modified without any approval!


Do not modify the device without approval from the manufacturer. If the
system is modified after consultation with the manufacturer, suitable
inspections and testing must be completed to ensure subsequent safe use.
The manufacturer is not liable for damages caused by an unauthorized use
of the device.
The correct use of the device is absolutely vital for safe operation. Therefore,
please thoroughly familiarize yourself with the content of these Instructions
for Use before starting up the system. Please comply with the Instructions for
Use of the other system components as well. You can obtain additional
information from ZEISS Service.
• Only operate the device within the scope of its intended use.
• This device must be used by properly instructed and trained personnel
only. It is the duty of the customer or institution operating the equipment
to train and instruct all staff using the equipment.
• Keep the Instructions for Use accessible to the operating personnel at all
times.
• Note the symbols and labels attached to the device! (See Page 28.)
• Only use original accessories or consumables approved by ZEISS.

Legal regulations • Check regularly that the legal regulations applicable in your country with
regard to accident prevention and work safety are being complied with.
• This device must not be modified without the manufacturer’s approval. If
the system is modified after consultation with the manufacturer, suitable
inspections and testing must be completed to ensure subsequent safe use.
The manufacturer is not liable for damage caused by an unauthorized
modifications to the device.
• If required by the applicable national laws and regulations, connect the
device to the "special substitute power supply".

Ambient conditions • Do not use the power-operated devices included in the delivery package

Version 11.0
Page 12 G-30-1435-en
S7 / OPMI PROergo Safety measures

– in areas where there is a risk of explosion,


– or at a distance of less than 25 cm from flammable anesthetics or
volatile solvents such as alcohol, benzine or similar substances.
• Do not use or store the device in a damp place. Do not expose the system
to dripping water, running water, sprayed water or splashes.
• Ensure that fluids cannot enter the device.
• Ensure that the installation conditions and the use of the device meet
surgical requirements:
– Low vibration
– Clean environment
– Avoidance of extreme mechanical stress

Electrical safety • Always switch off the device before you plug it into or unplug it from the
electricity network, if you are not going to use it for a prolonged period of
time, or if you want to clean its surface.
• Switch off the system at the power switch if you notice any smoke, sparks
or unusual noise. In this case, do not use the device again before it has
been repaired by our Service department.
A potential equalization connector is provided on the instrument's connection
panel (see IEC60601-1). This connector can be used to connect other active
devices to the same ground potential or for redundant connection to the
protective ground.
• Do not use multiple sockets!
• Never open the device! The device contains freely accessible live
components. If you remove the housing, you run the risk of electric shock.
• Do not touch the device if your body is electrostatically charged and the
device is not grounded.
• Please note the information on EMC (electromagnetic compatibility ) in
the chapter "System data" on Page 254.

Transport & Service • The system may only be transported over long distances (e.g. during
relocation, return for repair, etc.) in its original packaging or in special
return packaging. Please contact your dealer or ZEISS Service for this
purpose.
• This instrument is a high-grade technological product. To ensure optimum
performance and safe working order, we recommend having it checked
by our service representative as part of regular scheduled maintenance.

Version 11.0
G-30-1435-en Page 13
Safety measures S7 / OPMI PROergo

• In order to prevent any impairment of the device’s safety due to age,


wear, etc., the user must ensure that the device is subjected to the
necessary safety checks (see section "Care and maintenance").
• Modifications and repairs of this device or any equipment operated
together with this device may only be performed by ZEISS SERVICE or
other specialists authorized by Carl Zeiss.

Optics and light • Never look at the sun through the binocular tube, the objective lens or an
eyepiece.
• Avoid looking directly into the light source, e.g. into the microscope
objective lens or into the light guide, in order to prevent damage to
the eye!
• Start with the lowest brightness setting and gradually increase brightness
up to the necessary and still admissible level.
• With increasing age of the light source, the actual illumination intensity
delivered at the respective setting decreases (this is a normal property of
the system).

Version 11.0
Page 14 G-30-1435-en
S7 / OPMI PROergo Safety measures

Requirements for operation

Prior to the very first use

The system is installed by ZEISS Service or by a specialist designated by ZEISS.


Make sure that the following requirements continue to be met for subsequent
operation:
 The connecting components have been properly connected. The screw
connections have been firmly tightened.
 All cables and plugs are in perfect condition.
 The voltage set on the system corresponds to the rated line voltage on the
site of installation.
 The device is connected with the power cord supplied.
 When connecting the device to a network, make sure that the network
does not carry a hazardous voltage.
Specific information about the floor stand:
 The power cord may only be plugged into a power outlet that
is properly grounded.
More specific information about the ceiling mount:
 The ceiling anchor has been correctly installed by the competent
construction personnel.
 This ceiling anchor complies with the specifications contained in our
printed "Planning Manual".
 The device may only be connected to a power outlet that
is properly grounded.

Version 11.0
G-30-1435-en Page 15
Safety measures S7 / OPMI PROergo

Before every use

• When mounting accessory equipment, please make sure that the


admissible total weight of the surgical microscope is not exceeded.
(see "maximum load" label or the section "Technical data").
• Before every use, make certain to compensate for any added weight; this
will enable the surgical microscope to maintain its balance in all positions
of the working range.
• Before every use, limit the stroke of the suspension arm to prevent any
contact with the patient if the surgical microscope is lowered accidentally.
• Check the surgical microscope for sufficient freedom of movement. The
device itself or accessories may be damaged if conflicting with each other.
• To make sure that accessories do not fall down, check before each use
that they are firmly seated and the securing screws are firmly tightened.
• When connecting accessories, make sure that there is sufficient clearance
and that the surgical microscope does not touch the patient.
• Before every use, check the user settings in the software to avoid
unexpected behavior of the system.
• Always route cables so that they do not hinder the user.
• Never connect any unknown or defective devices to the electrical
interfaces (e.g. for data, video or foot control panel).
• Connect components only to properly grounded outlets.
• When setting up your system, make sure to comply with the requirements
of IEC 60601-1, chapter 16 (in some countries IEC 60601-1-1 may still be
valid).
• Reattach any covers or caps that have been removed. Close any existing
openings with the corresponding caps.
Before starting the system, remove the protective cover used for
transportation from the objective lens.
• If the xenon lamp is used beyond its maximum service life of 500 hours,
a sudden failure may occur. Change to the backup lamp (see Page 204)
or replace the lamp module in a timely manner (see Page 222). Reset the
service hour counter of the xenon illumination.

Version 11.0
Page 16 G-30-1435-en
S7 / OPMI PROergo Safety measures

• If the halogen lamp is used beyond its maximum service life of 50 hours,
sudden failure may occur and interrupt operation. Replace the halogen
lamp in a timely manner (see chapter "Service and maintenance" on
Page 220).
• Keep a replacement lamp (only for halogen light source) ready at hand.
• Never cover any ventilation openings. This may cause the light source of
the device to overheat and fail.
• Never use force when trying to connect any electrical connectors (plugs,
sockets). If a connection cannot be made, check to make sure the plug fits
the socket. If any of the connectors are damaged, have the ZEISS service
team or authorized service representative repair them.
• Position the suspension system in such a way that you can slide or swing
it away from the patient at any time.
• Prior to using the wireless foot control panel (FCP WL), make sure that its
batteries are fully charged. Inadequate power supply to the wireless foot
control panel may cause the device to malfunction.
• Go through the checklist in the chapter "Operation".
• Depending on how the surgical microscope is being used, take
appropriate precautions to ensure that, if necessary, the surgical
procedure or treatment can be completed without the full functionality
of the microscope (e.g. without the lighting).
• If you operate the device in a sterile area, make sure that you use the
corresponding sterile accessories for the device.
The single-use sterile drapes can be used to cover the device.
• When fitting the sterile drape, ensure there is sufficient room for
movement of the microscope carrier and surgical microscope.
• Observe the instructions for use for the sterile drapes and make sure
that the drape is attached in such a way that your field of view is not
obstructed.

Version 11.0
G-30-1435-en Page 17
Safety measures S7 / OPMI PROergo

More specific information about the ceiling mount:


• Move the lift arm or the lifting column down to its lower end position.
• Do not activate the lifting column or the lift arm during surgery. Do not
use the lifting column or the lift arm for focusing.
Specific information about the floor stand:
• Use the grab handle on the floor stand for moving.
• Mind the maximum vertical clearance when passing through doorways.
• Avoid any jolting.
• Risk of toppling – only move the device over steps or raised edges if two
people are available to hold and lift it!
• Be extremely careful when moving over slopes.Be extremely careful when
moving over slopes.
• Do not park the device on sloped surfaces.Do not park the device on
sloped surfaces.
• Secure the device in position by pressing down the immobilizers on the
stand base to prevent the stand from rolling away by itself.
• Do not exceed the instrument tray’s load capacity of 13 kg.

During use

• Never look directly into the light source, e.g. a microscope objective lens
or a light guide, as this might cause phototoxic injury to the eye.
• Never leave a device unattended when its light source is switched on.
• Defective or unidentified accessories may lead to increased leakage
current on the device and injure the patient. Never connect any defective
or unidentified accessories and never touch the power outlet or video
interfaces while in contact with the patient.
• If you replace the lamp shortly after it has failed, the lamp will still be very
hot. Use heat-resistant protective gloves when replacing the lamp to
prevent burns.
• If there is a failure and you cannot get any help from the chapter "What
to do in the event of malfunctions", attach a sign "Out of order" to the
system and contact ZEISS service representative.
• Do not pull at the light guide, power cord or any other connecting cables.

Version 11.0
Page 18 G-30-1435-en
S7 / OPMI PROergo Safety measures

• Constantly monitor the system during use.


• Risk of crushing - mind your fingers!
Fingers may be crushed in the areas marked with the "Risk of crushing"
sign.
Do not touch these areas while the device is being moved.

After every use

• Always use the master switch to turn off the device if it is not in use.
• Inadequate, flawed or incorrect cleaning or disinfection that does not
comply with these Instructions for Use can expose the patient or medical
staff to a considerable risk of infection.
• Step on at least three pedals to secure the floor stand in position.
Make sure that the floor stand is stable and cannot roll away.
• Move ceiling mount to its standby position.

Version 11.0
G-30-1435-en Page 19
Safety measures S7 / OPMI PROergo

Risk of burn injuries caused by high illumination intensity

General
The xenon illumination is a light source with high intensity. If used improperly,
excessive illumination intensities may lead to third-degree burns. Keep the
illuminated tissue moist and make sure it is sufficiently rinsed. Carefully
monitor the effect of the illumination on the tissue, particularly under the
following circumstances:
– During longer interventions on skin and tissue using objective lenses with
short focal depth (short working distance),
– During interventions on poorly perfused tissue,
– When the brightness control for the xenon lamp is set to a high
illumination intensity.
Please bear in mind that the values displayed for the light intensity may
not be used as a measuring function. They are intended only as an aid to
orientation.
• Only change the xenon lamp module after switching off the system.
The igniter generates a high voltage when the lamp is on.
• Any type of radiation has a damaging effect on biological tissue. This is
also true for the light illumination the surgical field. You should therefore
reduce brightness, the illuminated field size and the light exposure time to
the absolute minimum required.
If the brightness can no longer be reduced in the case of malfunctions, you
must take this into account in the light exposure time.

Various factors influence the risk of burn injuries:


System-related factors
– The wavelength range is limited by filters to the visible range between
400 nm and 700 nm in order to reduce the risk of burn injuries.
– The illumination intensity decreases with increasing number of operating
hours of the light source. In case of a lamp exchange, illumination
increases to the originally high value.
Surgery-related factors
– The selected intensity of the light source is a major factor for the risk of
injury. It should always be set to the minimum required for the surgical
procedure to be performed.

Version 11.0
Page 20 G-30-1435-en
S7 / OPMI PROergo Safety measures

– The size of the luminous field influences the risk of injury in two respects:
– With a large illuminated-field diameter, areas of the skin are
illuminated which are not monitored as strictly by the surgeon and
which are not moistened sufficiently. These areas are particularly
exposed to the risk of injury. Such injuries can be prevented by
adjusting the luminous field size to the smallest size required for that
particular operation.
– If the luminous field is reduced in size, the intensity increases because
the light is more strongly focused. So, if possible, the intensity should
be lowered as soon as the size of the luminous field is reduced.
– A long surgical procedure increases the risk of injury, in particular if
a standard procedure takes considerably longer than usual.
– Injuries in the peripheral area can be prevented by covering this area with
wet, sterile gauze. The risk is increased if dry drapes are used to cover such
areas.
– The interaction of heat and antimicrobial substances in incision foils may
lead to an increase in the patient's reaction to these substances.
– It should also be considered that some areas of the body may be more
sensitive than others.
– Certain preparations of the surgical field, local vasoconstrictive
medications and incision drapes may also result in a higher risk of injury.
Patient-related factors
– The general condition of a patient's health may contribute to the risk of
injury.
– The skin type may also play a major role for the risk of injury.
– Certain medications also affect the sensitivity to light.

Recommendations
Due to the large number of different factors involved and the lack of scientific
publications on this topic, ZEISS cannot provide guidance on acceptable
illumination intensities and exposure durations. However, the surgical
microscope has several features that can help the user to reduce the risk of
burn injuries:

Version 11.0
G-30-1435-en Page 21
Safety measures S7 / OPMI PROergo

– Using the buttons on the handgrip or foot control panel, the surgeon can
then set the illumination intensity to the value required for the procedure.
Please note that the intensity increases with decreasing luminous field size
if the Spot function is used. For this reason, the intensity should only be
set only after the size of the luminous field has been changed.
– Never leave a device unattended when its light source is switched on.
– Switch off the light when the microscope is not in use, and make sure that
it is not aimed at unprotected bare skin.
Please note that most burns affected the skin around the incision. The most
important measures to prevent burns are reduction of the size of the
illuminated field using the spot function and the coverage of peripheral areas
with sterile, moist gauze. The area of the incision should be constantly
moistened.

Concluding remarks
ZEISS recommends that the illumination of the surgical field be kept to the
necessary minimum, for the patient's safety and for a good microscopic
image.
Please refer to the relevant warnings and precautions.

Version 11.0
Page 22 G-30-1435-en
S7 / OPMI PROergo Safety measures

Stand safety equipment

Halogen lighting system


1 Manual switching to the backup lamp
The lamp housing contains a backup lamp which automatically swings
into position when the first lamp fails. If this automatic function fails,
you can switch on the backup lamp by pressing this button.
2 Retina protection filter GG 475
The filter selector knob has four positions:

0 No filter
1 GG475 filter: to protect the patient's eyes against unnecessary (blue)
radiation during treatment (retinal injury).
2 KK 40 filter: to increase color temperature
3 No filter

Version 11.0
G-30-1435-en Page 23
Safety measures S7 / OPMI PROergo

Xenon lighting system

CAUTION The xenon lamp has a limited service life of 500 h!


If used beyond its maximum service life, the xenon lamp may explode.
• Please replace the xenon lamp in due time.
• Reset the operating time counter.

CAUTION Risk of injury caused by lamp rupture!


Lamp rupture (audible as a loud bang) may lead to jamming of the lamp
module and/or failure of the electronics modules.
• Before opening the lamp housing, ensure that the device is moved to
a position in which possibly falling particles cannot put the patient or user
at risk.
• Do not continue using the system if the lamp module is jammed or the
illumination is no longer operational due to defective electronic modules.
Contact our Service Department.
1 Switching to the backup lamp
The lamp module contains two xenon lamps. The second lamp is used as
a backup lamp which has to be swung into the illumination beam path
when the first lamp fails.
• If the xenon lamp fails, open the lamp module as follows:
Press the button (3)
 The lamp module will eject partially.
• Pull out the lamp module as far as it will go.
• Turn knob (1) through 180° until it snaps in place. This moves the backup
lamp into the illumination beam path.
• Push the lamp module all the way back into the lamp housing.
2 Indicator: backup lamp is in use
When the segment in knob (1) lights up, the backup lamp is in use.
If the first lamp has failed and the backup lamp is in use, make sure to have
a backup lamp module ready at hand as a precaution.

Version 11.0
Page 24 G-30-1435-en
S7 / OPMI PROergo Safety measures

3 2
2

Version 11.0
G-30-1435-en Page 25
Safety measures S7 / OPMI PROergo

Ceiling mount with lifting column

Lifting column

1 2 Selector switch
for setting the optimum viewing height of the surgical microscope or for
downward movement into the working position.
As long as the selector switch remains in the turned position, the lifting
column moves upward (2) or downward (1), depending on the switch
position. When you release the selector switch, the lifting column stops
immediately.
The switching technology causes a delay of approx. 2 seconds at the upper
and lower end positions of the lifting column. After this time elapses, you can
move the column in the opposite direction again by activating the selector
switch again.
After being in operation for a maximum of one minute, the lifting column
must be turned off for at least nine minutes to prevent the motor from
overheating.

– Before raising or lowering the mount, make sure that there is sufficient
clearance to avoid collision with other objects.

CAUTION Risk of injury caused by lifting column


Activating the lifting column may cause injuries.
• Do not activate the lifting column during surgery.
• Do not use the lifting column for focusing.
• Make sure that the patient is not put at risk by the motorized adjustment
of the lifting column.

Version 11.0
Page 26 G-30-1435-en
S7 / OPMI PROergo Safety measures

Version 11.0
G-30-1435-en Page 27
Safety measures S7 / OPMI PROergo

Symbols and labels on the device

NOTE Please comply with the warning labels and notices!


• If you notice that any label is missing on your device or has become
illegible, please contact us or one of our authorized representatives.
We will provide you with replacements.

Labels on the microscope

1 "Balance setting of lateral tilt motion" label


Label on the knob for balance setting of the lateral tilt motion.

2 "Friction adjustment of vertical axis" label


Label on the knob for setting the required friction of the vertical axis.

3 "Balance setting of front-to-back tilt axis" label


Label on the knob for tilting the microscope upwards.

4 "Balance setting of front-to-back tilt axis" label


Label on the knob for tilting the microscope downwards.

5 "OPMI PROergo" device label


The device label contains the following information:

– Manufacturer (company name) – Order no.


– Device name – Serial number

Zero position 6 "Marking for zero position"


Indication of the position in which the microscope body is decoupled from
its suspension.
7 "Friction adjustment of front-to-back tilt axis" label
Label on the knob for setting the required friction of the front-to-back tilt
axis.
8 "Friction adjustment of lateral tilt axis" label
Label on the knob for setting the required friction of the lateral tilt axis.

Version 11.0
Page 28 G-30-1435-en
S7 / OPMI PROergo Safety measures

2
8

3
6

2
1 4

Version 11.0
G-30-1435-en Page 29
Safety measures S7 / OPMI PROergo

9 "Filter selector" marking


The illustration on the left shows the marks on the filter selector knob:

10 Labeling of zoom and focusing switches


Labeling of the zoom and focusing switches on the handgrips.

11 "Focusing knob" marking


The illustration on the left shows the marks on the focusing knob:

12 "Manual focusing" symbol


Identifies the manual focusing knob for manual adjustment of the image
sharpness.

OPMI PROergo 13 Device name


This imprint identifies the device.

10

13 9

12
10

11

Version 11.0
Page 30 G-30-1435-en
S7 / OPMI PROergo Safety measures

Labels on the connection panel

1 "Potential equalization" label (floor stand only)


Identifies the potential equalization connection for other active devices or
for redundant connection to the protective ground.
2 "CAN" connector label
CAN Identifies the CAN bus connector.
3 Warning sign "Power output socket" (floor stand only)
Only connect devices with the correct electrical ratings.

4 Label "Power output socket" (floor stand only)


with permissible electrical values for connected devices.
5 "Video out" connector label
video out Identifies the Y/C video connection port.
6 "Connector of wired foot control panel" label

remote 7 Remote connector


Identifies a connector to which devices with a maximum breaking
capacity of 24V/0.5A may be connected.

2
7 3
4

Version 11.0
G-30-1435-en Page 31
Safety measures S7 / OPMI PROergo

Labels on S7 floor stand

1 "Risk of crushing" warning label


Fingers may be pinched between the carrier and suspension arms.
Do not touch this area while moving the surgical microscope.

2 "Maximum load" warning label


The maximum load on the suspension arm must not exceed 14 kg.
Max.14 kg

3 "Friction of suspension arm's upward/downward movement" label


Label on the knob for adjusting the friction of the suspension arm's
upward/downward movement.

- Balance + 4 "Balance setting" label


Balance setting must be performed after all accessories have been
mounted on the microscope.
5 "Friction of suspension arm's swivel movement" label
Label on the knob for adjusting the friction of the suspension arm's swivel
movement.

6 "Friction of carrier arm's upward/downward movement" label


Label on the knob for adjusting the friction of the carrier arm's upward/
downward movement.

Version 11.0
Page 32 G-30-1435-en
S7 / OPMI PROergo Safety measures

1 1

2 1

5 4 1 3

Version 11.0
G-30-1435-en Page 33
Safety measures S7 / OPMI PROergo

7 "Observe the Instructions for Use" label


Observe the Instructions for Use.

8 "Date of manufacture" plate


This plate indicates the date of manufacture for the device (year-month).

9 Rating plate
The rating label contains the following information:
– Manufacturer's symbol
– Manufacturer (company name) Carl Zeiss Meditec AG
– Manufacturer's address Goeschwitzer Strasse 51-52
07745 Jena, Germany
– Serial number
– Device name S7 / OPMI PROergo
– Reference number
– Rated Voltage (115V): 100V - 125V
(230V): 220V - 240V
– Connected load (115V): max 850VA
(230V): max 1250VA
– Line frequency range 50Hz - 60Hz
– Maximum total weight (with Max. 190 kg
installed safe workload)
– CSA approval
– CE marking

10 "Observe disposal regulations" indicating label


Do not dispose of electrical or electronic devices along with normal
domestic waste. For more information on the disposal of electrical and
electronic devices, please see the chapter "Maintenance and care".

Version 11.0
Page 34 G-30-1435-en
S7 / OPMI PROergo Safety measures

8 9 10

Version 11.0
G-30-1435-en Page 35
Safety measures S7 / OPMI PROergo

11 "Observe transport position" warning label


Indicates the transport position of the system. Before transporting the
system always ensure that it is in this position to avoid any damage.

12 SIP label
The SIP label contains the following information:
– Manufacturer (company name)
– Manufacturer's contact details,
i.e. phone number, fax number and email address of the local contact
for the national ZEISS sales organization.
XXXX XXXXXX – SIP number
A unique identification number assigned to your device.

13 "Observe the Instructions for Use" label


Observe the Instructions for Use.

Version 11.0
Page 36 G-30-1435-en
S7 / OPMI PROergo Safety measures

11

12

13

Version 11.0
G-30-1435-en Page 37
Safety measures S7 / OPMI PROergo

Additional labels on S7 floor stand (instrument tray option)

1 "Maximum load on instrument tray" indicating label


The maximum load of accessories on the instrument tray must not exceed
13 kg.

Version 11.0
Page 38 G-30-1435-en
S7 / OPMI PROergo Safety measures

Version 11.0
G-30-1435-en Page 39
Safety measures S7 / OPMI PROergo

Labels on S7 ceiling mount

1 "Risk of crushing" warning label


Fingers may be pinched between the carrier and suspension arms. Do not
touch this area while moving the surgical microscope.

2 "Maximum load" warning label


The maximum load on the suspension arm must not exceed 14 kg.
Max.14 kg

3 "Friction of suspension arm's swivel movement" label


Label on the knob for adjusting the friction of the suspension arm's swivel
movement.

4 "Friction of carrier arm's upward/downward movement" label


Label on the knob for adjusting the friction of the carrier arm's upward/
downward movement.

5 "Observe the Instructions for Use" label


Observe the Instructions for Use.

6 "Date of manufacture" plate


This plate indicates the date of manufacture for the device (year-month).

7 "Friction of suspension arm's upward/downward movement" label


Label on the knob for adjusting the friction of the suspension arm's
upward/downward movement.

8 "Observe disposal regulations" indicating label


Do not dispose of electrical or electronic devices along with normal
domestic waste. For more information on the disposal of electrical and
electronic devices, please see the chapter "Maintenance and care".

- Balance + 9 "Balance setting" label


Balance setting must be performed after all accessories have been
mounted on the microscope.

Version 11.0
Page 40 G-30-1435-en
S7 / OPMI PROergo Safety measures

1 2 1 3

5
1 6 5

9 8 7

Version 11.0
G-30-1435-en Page 41
Safety measures S7 / OPMI PROergo

10 SIP label
The SIP label contains the following information:
– Manufacturer (company name)
– Manufacturer's contact details,
i.e. phone number, fax number and email address of the local contact
for the national ZEISS sales organization.
XXXX XXXXXX – SIP number
A unique identification number assigned to your device.

11 Rating label (for S7 ceiling mount with rigid column)


The rating label contains the following information:
– Manufacturer's symbol
– Manufacturer (company name) Carl Zeiss Meditec AG
– Manufacturer's address Goeschwitzer Strasse 51-52
07745 Jena, Germany
– Serial number
– Device name S7 / OPMI PROergo
– Reference number
– Rated Voltage (115V): 100V - 125V
(230V): 220V - 240V
– Connected load (115V): max 450VA
(230V): max 450VA
– Line frequency range 50Hz - 60Hz
– CSA approval
– CE marking

Version 11.0
Page 42 G-30-1435-en
S7 / OPMI PROergo Safety measures

12 Rating label (for S7 ceiling mount with lifting column)


The rating label contains the following information:
– Manufacturer's symbol
– Manufacturer (company name) Carl Zeiss Meditec AG
– Manufacturer's address Goeschwitzer Strasse 51-52
07745 Jena, Germany
– Serial number
– Device name S7 / OPMI PROergo
– Reference number
– Rated Voltage (115V): 100V - 125V
(230V): 220V - 240V
– Connected load (115V): max 450VA
(230V): max 450VA
– Line frequency range 50Hz - 60Hz
– CSA approval
– CE marking

10 11/12

Version 11.0
G-30-1435-en Page 43
Safety measures S7 / OPMI PROergo

Additional label on S7 ceiling mount with lifting column

1 "Lifting column operation" indicating sign


Switch for moving the vertical position upward or downward.

2 Lifting column activation time


This label represents the maximum permissible operating time as well as
the subsequent mandatory rest phase for the lifting column and indicates
the ratio between the operating time and the rest phase.
After being in operation for a maximum period of one minute, the lift
column must be turned off for at least nine minutes to prevent the motor
from overheating.

Version 11.0
Page 44 G-30-1435-en
S7 / OPMI PROergo Safety measures

2 1

Version 11.0
G-30-1435-en Page 45
Safety measures S7 / OPMI PROergo

Labels on S7 wall mount and on S7 Centro suspension system

1 "Risk of crushing" warning label


Fingers may be pinched between the carrier and suspension arms. Do not
touch this area while moving the surgical microscope.

2 "Maximum load" warning label


The maximum load on the suspension arm must not exceed 14 kg.
Max.14 kg

3 "Friction of suspension arm's swivel movement" label


Label on the knob for adjusting the friction of the suspension arm's swivel
movement.

4 "Friction of carrier arm's upward/downward movement" label


Label on the knob for adjusting the friction of the carrier arm's upward/
downward movement.

5 "Observe the Instructions for Use" label


Observe the Instructions for Use.

6 "Date of manufacture" plate


This plate indicates the date of manufacture for the device (year-month).

7 "Friction of suspension arm's upward/downward movement" label


Label on the knob for adjusting the friction of the suspension arm's
upward/downward movement.

8 "Observe disposal regulations" indicating label


Do not dispose of electrical or electronic devices along with normal
domestic waste. For more information on the disposal of electrical and
electronic devices, please see the chapter "Maintenance and care".

- Balance + 9 "Balance setting" label


Balance setting must be performed after all accessories have been
mounted on the microscope.

Version 11.0
Page 46 G-30-1435-en
S7 / OPMI PROergo Safety measures

1 2 1 3

1 6 5

9 8 7

Version 11.0
G-30-1435-en Page 47
Safety measures S7 / OPMI PROergo

10 SIP label
The SIP label contains the following information:
– Manufacturer (company name)
– Manufacturer's contact details,
i.e. phone number, fax number and email address of the local contact
for the national ZEISS sales organization.
XXXX XXXXXX – SIP number
A unique identification number assigned to your device.

11 Rating label on wall mount with short arm


The rating label contains the following information:
– Manufacturer's symbol
– Manufacturer (company name) Carl Zeiss Meditec AG
– Manufacturer's address Goeschwitzer Strasse 51-52
07745 Jena, Germany
– Serial number
– Device name S7 / OPMI PROergo
– Reference number
– Rated Voltage (115V): 100V - 125V
(230V): 220V - 240V
– Connected load (115V): max 450VA
(230V): max 450VA
– Line frequency range 50Hz - 60Hz
– CSA approval
– CE marking

Version 11.0
Page 48 G-30-1435-en
S7 / OPMI PROergo Safety measures

12 Rating label on wall mount with long arm


The rating label contains the following information:
– Manufacturer's symbol
– Manufacturer (company name) Carl Zeiss Meditec AG
– Manufacturer's address Goeschwitzer Strasse 51-52
07745 Jena, Germany
– Serial number
– Device name S7 / OPMI PROergo
– Reference number
– Rated Voltage (115V): 100V - 125V
(230V): 220V - 240V
– Connected load (115V): max 450VA
(230V): max 450VA
– Line frequency range 50Hz - 60Hz
– CSA approval
– CE marking

10 11/12

Version 11.0
G-30-1435-en Page 49
Safety measures S7 / OPMI PROergo

13 Rating label on S7 Centro suspension system


The rating label contains the following information:
– Manufacturer's symbol
– Manufacturer (company name) Carl Zeiss Meditec AG
– Manufacturer's address Goeschwitzer Strasse 51-52
07745 Jena, Germany
– Serial number
– Device name S7 / OPMI PROergo
– Reference number
– Rated Voltage (115V): 100V - 125V
(230V): 220V - 240V
– Connected load (115V): max 450VA
(230V): max 450VA
– Line frequency range 50Hz - 60Hz
– CSA approval
– CE marking

13

Version 11.0
Page 50 G-30-1435-en
S7 / OPMI PROergo Safety measures

Additional labels on S7 wall mount

14 "Risk of crushing" indicating label


Fingers may be pinched between the carrier and suspension arms. Do not
touch this area while moving the surgical microscope.

14

14

Version 11.0
G-30-1435-en Page 51
Safety measures S7 / OPMI PROergo

Labels on the light sources of the suspension systems

xenon light source


1 Warning sign "Not for Ophthalmology"
The xenon light source may not be used for ophthalmological surgeries.
Make sure that no xenon light enters the patient's eyes.

2 "CAUTION" warning label


The lamp housing contains an integrated xenon light source.
Take note of the description on Page 222 "Changing the xenon lamp
module".
3 Lamp change
The plate graphically shows the three steps involved in changing the lamp.
A - Press the button
B - Pull out the lamp insert module
C - Change bulbs

1 4 "Filter selector for light source" indicating sign


0 This label indicates the position of the filter wheel.

1 2 5 "Manual switching to backup lamp" indicating label


These numerical symbols refer to switching from the standard lamp to the
backup lamp.
6 "Instructions for Use" sign
Observe the Instructions for Use or further applicable documents.

Version 11.0
Page 52 G-30-1435-en
S7 / OPMI PROergo Safety measures

1 2 3 4 5 6

Version 11.0
G-30-1435-en Page 53
Safety measures S7 / OPMI PROergo

Halogen light source


1 "Filter selector for halogen light source"
This label indicates the position of the filter wheel.

2 "Instructions for Use" sign


Observe the Instructions for Use or further applicable documents.

Version 11.0
Page 54 G-30-1435-en
S7 / OPMI PROergo Safety measures

Labels on the wall power outlet (optional)

1 Label - Cable connection for the foot control panel

2 Remote connector
Identifies a connector to which devices with a maximum breaking
capacity of 24V/0.5A may be connected.

Version 11.0
G-30-1435-en Page 55
Safety measures S7 / OPMI PROergo

Version 11.0
Page 56 G-30-1435-en
S7 / OPMI PROergo Device overview

Device overview

PROergo surgical microscope on S7 suspension system .................60


Properties .................................................................................................60
System components .................................................................................62

OPMI PROergo surgical microscope ...............................................64


Set-up ......................................................................................................64
Control elements, Displays, Connections ...................................................66
Binocular tubes and eyepieces ..................................................................72

Floor stand S7 ................................................................................ 76


Set-up ......................................................................................................76
Suspension arm ........................................................................................78
Carrier arm, stand column, stand base ......................................................80
Connector panel of S7 floor stand ............................................................82
Instrument tray (option) ............................................................................84

S7 ceiling mount .............................................................................86


Set-up ......................................................................................................86
Support arm and suspension arm .............................................................88
Connector panel of S7 ceiling mount ........................................................90

S7 ceiling mount with lifting column ..............................................92


Set-up ......................................................................................................92
Ceiling flange with lifting column .............................................................94
Support arm and suspension arm .............................................................96
Connector panel of S7 ceiling mount with lifting column ..........................98

S7 wall mount and S7 Centro suspension system ......................... 100


Set-up ....................................................................................................100
Support arm and suspension arm ...........................................................102

Version 11.0
G-30-1435-en Page 57
Device overview S7 / OPMI PROergo

Connector panel of S7 ceiling mount ..................................................... 104


Connectors for the optional wall outlet (ceiling mount) .......................... 106

Illumination system .......................................................................108


Halogen lighting system ......................................................................... 110
Xenon lighting system ........................................................................... 112

Control panel and menu overview ................................................116


Structure of the menus .......................................................................... 118
Optional foot control panel .................................................................... 134
Foot control panel for zoom and focus .................................................. 135
14-function foot control panel, wired (FCP) and wireless (FCP WL) ......... 136

1Chip HD camera (option) .............................................................138

Version 11.0
Page 58 G-30-1435-en
S7 / OPMI PROergo Device overview

Version 11.0
G-30-1435-en Page 59
Device overview S7 / OPMI PROergo

PROergo surgical microscope on S7 suspension system

Properties

The apochromatic optics of the microscope provide superb optical quality.


The microscope image displays optimum contrast and excellent detail
recognition along with outstanding depth of field and maximum 3D
perception. The 1:6 ratio zoom system allows the magnification of the overall
system to be set as required by the surgical procedure.

Suspension systems
The S7 floor stand is a suspension system for ZEISS surgical microscopes.
It comprises a suspension arm, a carrier arm, a stand column and a stand
base.
There is a handle mounted on the stand column. It is used to move the stand
and to suspend the foot control panel. The stand column is equipped on both
sides with cable supports for winding up cables before the unit is moved to
another location.
The stand base is extremely easy to move on its four casters. It has been
designed in such a way that high stability is ensured even with unfavorable
loading of the stand. The locking tabs permit you to reliably secure the floor
stand in position at the site of use.
Because the stand moves so smoothly, it is easy to underestimate how heavy
it is. Therefore, move the stand slowly and carefully!
The S7 ceiling mount with rigid column is a suspension system for ZEISS
surgical microscopes. It comprises a suspension arm, a carrier arm, a ceiling
column and a ceiling flange.
The S7 ceiling mount with lifting column is a suspension system for ZEISS
surgical microscopes. It comprises a suspension arm, a carrier arm, a ceiling
flange and a lifting column. This motorized lifting column permits the viewing
height to be perfectly matched to the surgeon's requirements.
The S7 wall mount and S7 Centro suspension system are carrier systems
for ZEISS surgical microscopes. The wall mount comprises a suspension arm,
carrier arm and wall anchor plate. The Centro suspension system comprises
a suspension arm and carrier arm.
The carrier arm of all suspension systems contains the control unit with all
electrical supply systems required for the control of a motorized surgical
microscope. You can control the motorized functions using the handgrips,

Version 11.0
Page 60 G-30-1435-en
S7 / OPMI PROergo Device overview

or via a foot control panel (option). A light guide (fiber optic illumination)
directs the light from the lamp housing in the suspension system to the
surgical microscope.
The suspension arm of all types of suspension system permits almost effortless
positioning of the surgical microscope. The spring force of the suspension arm
can be adjusted within a range from 6 to 14 kg. permitting reliable balancing
of the microscope even with accessory equipment attached. The downward
movement of the suspension arm can be limited using an adjustable
positioning aid.
A control panel allows the adjustment of user-defined settings such as the
zoom, focus, brightness and video settings, and the programming of the
configurable handgrip buttons.

Options for your surgical microscope


The SpeedFokus autofocus option permits you to select features (ROI, region
of interest) in the image of the surgical field and to rapidly focus on them
automatically or at the press of a button.
The "1Chip HD camera" option provides high-resolution video equipment for
your surgical microscope.
ZEISS Service can also retrofit existing surgical microscopes with the
"1Chip HD camera" option on request.
The "TRIO 610 HD video camera" option with the TRIO 600 CCU camera
control unit is a non-medical accessory for surgical microscopes. The video
camera system can be connected via the C-mount interface to a surgical
microscope from ZEISS.
ZEISS Service can also retrofit existing surgical microscopes with the
"TRIO 610 HD video camera" option on request.

Version 11.0
G-30-1435-en Page 61
Device overview S7 / OPMI PROergo

System components

1 S7 ceiling mount
2 S7 ceiling mount with lifting column
3 S7 wall mount
4 Floor stand S7
5 S7 central stand
6 OPMI PROergo surgical microscope
with 180° tiltable tube and magnetic eyepieces.

The overall system comprises a suspension system and a microscope.


All options described in these Instructions for Use can be ordered separately.

Version 11.0
Page 62 G-30-1435-en
S7 / OPMI PROergo Device overview

1 2

6 4

Version 11.0
G-30-1435-en Page 63
Device overview S7 / OPMI PROergo

OPMI PROergo surgical microscope

Set-up

The OPMI PROergo surgical microscope comprises the following modules:


1 Microscope body
2 Balancing system including magnetic brakes
This system allows balancing of the surgical microscope. When the
magnetic brakes are unlocked, the surgical microscope can be positioned
almost effortlessly. When the button on the back of the handgrip is
released, the microscope is reliably fixed in position above the surgical
field.
3 Axis of microscope rotation
4 coupling
for mounting the surgical microscope on the suspension system.
5 Support arm for the surgical microscope
6 Handgrips
for moving the surgical microscope. Using the buttons on the handgrips,
you can control important functions (e.g. focusing, zooming, brightness,
still/video camera release).
Using the buttons on the back of the handgrips you can release the
magnetic brakes.
NOTE Injury to the patient
If the system is in an extremely unbalanced state, the unit can move
uncontrollably out of position.
• The surgical microscope and the suspension system used must be
balanced correctly to permit almost effortless guiding of the surgical
microscope.
• For this reason, hold the surgical microscope tightly at its handgrips
before loosening the friction adjustment screw (see Page 78) on the
suspension arm.

Version 11.0
Page 64 G-30-1435-en
S7 / OPMI PROergo Device overview

1 2 3 4 5

6 2

Version 11.0
G-30-1435-en Page 65
Device overview S7 / OPMI PROergo

Control elements, Displays, Connections

1 Friction adjustment of vertical axis


Use this knob to adjust the friction of the vertical axis as required.
2 Balance setting of lateral tilt motion
Use this knob to adjust the balance setting of the lateral tilt motion.
3 Friction of the lateral tilt axis
Use this knob to adjust the friction of the lateral tilt axis as required.
4 Balancing the front-to-back tilt axis (motorized)
2 When you press this button, the microscope slowly tilts upward.
5 Balancing the front-to-back tilt axis (motorized)
1 When you press this button, the microscope slowly tilts downward.
6 Friction of the front-to-back tilt axis
Use the knob to adjust the friction of the front-to-back tilt axis as required.

Zero position 7 Marking for zero position


In the zero position, the microscope body is decoupled from the
suspension (see "Balancing the front-to-back tilt motion", Page 170).

Version 11.0
Page 66 G-30-1435-en
S7 / OPMI PROergo Device overview

4
2
1
5

6
2

Version 11.0
G-30-1435-en Page 67
Device overview S7 / OPMI PROergo

8 Rotating the video image (option)


This knob allows you to rotate the video image, i.e. the video image
can be rotated to provide better orientation for the assistant or for
documentation purposes. The zero position of the knob is indicated by
a clickstop.
9 Dust cover
10 Handgrips (removable)
for moving the surgical microscope and triggering the motorized OPMI
functions.
11 Clamping screw
for optimum positioning of the handgrips.
12 Focusing knob
for the manual setting of the image definition (focus, working distance)
13 Filter selector
The filter selector knob has the following positions:
H Green filter: enhances the contrast of tissue with a strong blood
supply and makes even the finest structures visible.
O Orange filter: used especially for working with plastic fillings to
prevent the material from premature hardening.

No filter

Medium-sized pinhole diaphragm: provides a medium-sized


illuminated field.
Small pinhole diaphragm: provides a small illuminated field.

Version 11.0
Page 68 G-30-1435-en
S7 / OPMI PROergo Device overview

10 11 12 13

Version 11.0
G-30-1435-en Page 69
Device overview S7 / OPMI PROergo

14 Release buttons, freely programmable


In combination with an S7 suspension system, specific functions of the
suspension system can be assigned to these buttons (e.g.: focus, zoom,
brightness control, still and video camera release, etc.) Configuring these
buttons is described on Page 124 (Handgrip Settings menu).
15 Zoom release button
T For setting the magnification factor in a range from 0.4x to 2.4x.
16 Focus release key
F For infinitely variable focusing within the working distance of 200 mm to
415 mm.
17 Release buttons, freely programmable
In combination with an S7 suspension system, specific functions of the
suspension system can be assigned to these buttons (e.g.: focus, zoom,
brightness control, still and video camera release, etc.) Configuring these
buttons is described in chapter "Description/Options menu".
18 Release button for magnetic brakes
The magnetic brakes of the surgical microscope are released for as long
as you press this button.
19 Locking the handgrips in position
Using this locking device, you can lock each handgrip in different
positions.
20 Knurled nut
for removing the handgrips.
21 Knobs instead of handgrips
After removing the handgrips, mount the knobs supplied to protect the
connector contacts and to be able to guide the surgical microscope.
These knobs can be provided with sterile caps.
22 Labeling fields
Labeling fields are provided for the freely programmable handgrip
buttons. On delivery, you receive a labeling set (small stickers) for each
handgrip, permitting you to label the handgrips in accordance with your
specific button assignment.
In addition to the abbreviations of the available functions, the labeling set
includes the letters A, B, C and D.
If several users with different button configurations are working on the
system, operating errors can be avoided by labeling the freely
programmable buttons with A, B, C and D instead of specific functions as
in the handgrips menu (see Page 124).

Version 11.0
Page 70 G-30-1435-en
S7 / OPMI PROergo Device overview

14

15

16

F
T
17

18

19
20

21
F

22
T

Version 11.0
G-30-1435-en Page 71
Device overview S7 / OPMI PROergo

Binocular tubes and eyepieces

CAUTION Damage to the eyes due to high light radiation!


Looking into a light source through a tube, an objective lens or an eyepiece
may cause eye injury.
• Never look into a light source or the sun with a tube, an objective lens or
an eyepiece.
Optionally you can equip the surgical microscope with the 180° swiveling
tube or the folding tube 170/f260.
The optional stereo observer is equipped with either a straight tube or
a swiveling tube.
1 Eyepiece mounting
2 Eye distance adjustment wheel (interpupillary distance)
The correct position has been reached when the two eyepiece images
merge into one. You can read off the PD setting on the adjustment knob.
3 Eye distance scale (interpupillary distance)
4 "normal/larger" symbol for magnification changer
5 Adjustment wheel for integrated magnification changer
6 (released/locked) symbol for manual tube rotation

7 Button for manual tube rotation

Version 11.0
Page 72 G-30-1435-en
S7 / OPMI PROergo Device overview

Straight tube

45° inclined tube

180° tiltable tube


2

Foldable Tube f170 f260

1 2 3 4 5 6 7

Version 11.0
G-30-1435-en Page 73
Device overview S7 / OPMI PROergo

Widefield eyepieces with magnetic coupling

NOTE Eyepieces have a magnetic field!


Please keep in mind the usual rules for the handling of magnets with regard
to eyepieces that have been removed from the tube:
• Do not place the eyepiece near instruments which may be magnetizable.
• Do not place the eyepiece on sensitive electronic devices such as infusion
pumps, pacemakers, measuring devices or magnetic data media such as
floppy disks, audio and video tapes or credit cards.
• Always store the unmounted eyepiece in its original packaging.
1 Eyecup
Always adjust the eyecups in such a way that the entire field of view can
be seen.

– Observation with Screw-in the eyecups all the way.


eyeglasses:
– Observation without Rotate the eyecups out in order to adjust
eyeglasses: them to the field of view of the observer.

2 Diopter setting ring


The eyepieces provide ametropia compensation between -8 D and +5 D.
Eyeglass wearers who wear eyeglasses during their work set the diopter
setting ring to the zero position.
Then turn the ring until the optimum setting has been achieved.
An integrated brake holds the setting ring in the position set.
3 Diopter scale
For reading the preset refractive value.

Version 11.0
Page 74 G-30-1435-en
S7 / OPMI PROergo Device overview

Version 11.0
G-30-1435-en Page 75
Device overview S7 / OPMI PROergo

Floor stand S7

Set-up

1 Suspension arm
2 Lamp housing (either xenon or halogen illumination)
3 Support arm
4 Control panel
5 Connector panel
6 Handle
7 Cable supports (2x, on the right and left of the stand column)
8 Stand column
9 Stand base

Version 11.0
Page 76 G-30-1435-en
S7 / OPMI PROergo Device overview

7
8

Version 11.0
G-30-1435-en Page 77
Device overview S7 / OPMI PROergo

Suspension arm

1 Lamp housing with halogen or xenon illumination


The stand is equipped with an illumination system for light guides.
Each lamp housing contains a backup lamp. When a halogen illumination
system is used, the backup lamp will be automatically swung in if the first
lamp fails. When a xenon illumination system is used, you must manually
swing the backup lamp into the beam path if the first lamp fails.
2 Friction
Knob for adjusting the friction of the suspension arm's upward/downward
movement.
3 Clamping screw for limiting downward movement
The movement of the suspension arm can be limited in the downward
direction.
Adjustment range: from the horizontal position of the suspension arm to
its lower stop.
4 Balancing
- Balance + Knob for setting the spring force for balancing. After all the accessories
have been fitted on the surgical microscope, this adjustment screw can be
used to set the balance of the suspension arm. The balance setting
procedure is described in the chapter "Preparations"/"Adjusting the
system".

Version 11.0
Page 78 G-30-1435-en
S7 / OPMI PROergo Device overview

Version 11.0
G-30-1435-en Page 79
Device overview S7 / OPMI PROergo

Carrier arm, stand column, stand base

1 Control and display panel


The control and display panel permits you to control all electrical functions
of the S7 floor stand and the surgical microscope.
2 Friction
Knob for adjusting the friction of the carrier arm's swivel movement.
3 Handle
for moving the floor stand. Before relocating the stand, you can hang the
foot control panel and wind up the cables on the lower end of the handle.
4 Connector panel
for connecting the system to line power. Further connectors are provided
for powering peripheral equipment. A foot control panel can be
connected to a multi-point connector.
5 Friction
Knob for adjusting the friction of the suspension arm's swivel movement.
6 Cable clip
for winding up the power cord and the cable of the foot control panel.
7 Bracket for foot control panel
8 Locking tab
For steerable casters (4 pcs).
To secure the stand in position, press down at least two locking tabs.
Press down to lock the caster.
Press up to release the caster.
9 Steerable caster
The four casters roll very smoothly, permitting the unit to be easily
wheeled to the site of use.

Version 11.0
Page 80 G-30-1435-en
S7 / OPMI PROergo Device overview

Version 11.0
G-30-1435-en Page 81
Device overview S7 / OPMI PROergo

Connector panel of S7 floor stand

CAUTION Risk of electric shock caused by unapproved or defective accessories!


Unapproved or defective accessories can cause increased leakage current!
• When setting up your system, be sure to comply with the requirements
of IEC 60601-1, section 16 (in some countries IEC 60601-1-1 may still be
valid).
• Comply with the Medical Device Operator Ordinance or other national
laws.
1 Opening for system cable of external video camera
2 Remote connection
For controlling external devices (e.g. MEDIALINK 100) with a maximum
switching capacity of 24 V / 0.5 A.
3 Potential equalization pin
This connector can be used to connect other active devices to the same
ground potential or for redundant connection to the protective ground.
4 connection strip for switching element (option)
Connection option for a foot control panel
5 Power output socket
For medical devices with power consumption of 115 V~ with a maximum
of 400 VA and for 230 V~ with a maximum of 800 VA.
The current of this power output socket can not be switched off with the
power switch.
6 Power input socket
7 Strain relief
The strain relief prevents inadvertent disconnection of the following
electrical connections:
– power cord,
– video connection cable,
– connection cable for the foot control panel (optional).
8 BNC video port (option)
For VBS video signal output.
9 Y/C video port (option)
For S-VHS video signal output.

Version 11.0
Page 82 G-30-1435-en
S7 / OPMI PROergo Device overview

10 Display window for rated voltage


The rated voltage displayed here must match the rated voltage of the
power grid available at the installation site. You can adjust the sliding
switch using a suitable tool.
11 CAN bus connector (not supported)

1 2 3 4 5 6 7 8 9 10 11

Version 11.0
G-30-1435-en Page 83
Device overview S7 / OPMI PROergo

Instrument tray (option)

The S7 floor stand can be equipped or retrofitted with an instrument tray (1).
In the case of retrofitting, our Service staff or an authorized person will install
the instrument tray on your suspension system.
The instrument tray (1) can carry a maximum of 13 kg. It is designed to
securely carry items such as a video recorder. Further accessories can be
installed on the instrument tray using the strap provided.

NOTE Risk of injury caused by non-observance of the following information!


• Do not exceed the instrument tray’s (1) load capacity of 13 kg.
• Make sure that the accessory equipment is positioned as securely as
possible on the instrument tray.
• Secure the accessory equipment using the two stud bolts on the
instrument tray.
• If required, secure further accessory equipment on the tray using the
strap provided.
• Remember there is a risk of collision and crushing when suspension arm
(2) is folded to its moving position. A "Risk of crushing" warning label is
therefore attached on the left and right of suspension arm (2).
• Please read the relevant Instructions for Use before starting up the
accessory equipment.
• Never pull or push at the accessory equipment (3) in order to move the
floor stand. Always use only the grab handle (4) to move the floor stand.

Version 11.0
Page 84 G-30-1435-en
S7 / OPMI PROergo Device overview

Version 11.0
G-30-1435-en Page 85
Device overview S7 / OPMI PROergo

S7 ceiling mount

Set-up

1 Ceiling flange
2 Stand column
3 Control panel
4 Connector panel
5 Lamp housing (either xenon or halogen illumination)
6 Suspension arm
7 Support arm

Version 11.0
Page 86 G-30-1435-en
S7 / OPMI PROergo Device overview

7
3

6
4

Version 11.0
G-30-1435-en Page 87
Device overview S7 / OPMI PROergo

Support arm and suspension arm

1 Lamp housing with halogen or xenon illumination


The stand is equipped with an illumination system for light guides.
Each lamp housing contains a backup lamp. When a halogen illumination
system is used, the backup lamp will be automatically swung in if the first
lamp fails.
When a xenon illumination system is used, you need to swing the backup
lamp into position manually if the first lamp fails.
2 Friction
Knob for adjusting the friction of the suspension arm's upward/downward
movement.
3 Clamping screw for limiting downward movement
The movement of the suspension arm can be limited in the downward
direction.
Adjustment range: from the horizontal position of the suspension arm to
its lower stop.
4 Balancing
- Balance + Knob for setting the spring force for balancing. After all the accessories
have been fitted on the surgical microscope, this adjustment screw can be
used to set the balance of the suspension arm. The procedure for setting
the balance is described in the section "Preparations for use".
5 Friction
Knob for adjusting the friction of the suspension arm's swivel movement.
6 Friction
Knob for adjusting the friction of the carrier arm's swivel movement.
7 Control and display panel
The control and display panel permits you to control all electrical functions
of the S7 suspension system and the surgical microscope.
8 Connector panel
Further connectors are provided for powering peripheral equipment.
A foot control panel (option) can be connected to a multi-point
connector.

Version 11.0
Page 88 G-30-1435-en
S7 / OPMI PROergo Device overview

Version 11.0
G-30-1435-en Page 89
Device overview S7 / OPMI PROergo

Connector panel of S7 ceiling mount

CAUTION Risk of electric shock caused by unapproved or defective accessories!


Unapproved or defective accessories can cause increased leakage current!
• When setting up your system, be sure to comply with the requirements
of IEC 60601-1, section 16 (in some countries IEC 60601-1-1 may still be
valid).
• Comply with the Medical Device Operator Ordinance or other national
laws.
1 Opening for system cable of external video camera
2 Remote connection
For controlling external devices (e.g. MEDIALINK 100) with a maximum
switching capacity of 24 V / 0.5 A.
3 Potential equalization pin
This connector can be used to connect other active devices to the same
ground potential or for redundant connection to the protective ground.
4 connection strip for switching element (option)
Connection option for a foot control panel
5 Strain relief
The strain relief prevents inadvertent disconnection of the following
electrical connections:
– video connection cable,
– connection cable for the foot control panel (optional).
6 BNC video port (option)
For VBS video signal output.
7 Y/C video port (option)
For S-VHS video signal output.
8 Display window for rated voltage
The rated voltage displayed here must match the rated voltage of the
power grid available at the installation site. You can adjust the sliding
switch using a suitable tool.
9 CAN bus connector (not supported)

Version 11.0
Page 90 G-30-1435-en
S7 / OPMI PROergo Device overview

1 2 3 4 5 6 7 8 9

Version 11.0
G-30-1435-en Page 91
Device overview S7 / OPMI PROergo

S7 ceiling mount with lifting column

Set-up

1 Ceiling flange
2 Lifting column
3 Control panel
4 Connector panel
5 Lamp housing (either xenon or halogen illumination)
6 Suspension arm
7 Support arm

Version 11.0
Page 92 G-30-1435-en
S7 / OPMI PROergo Device overview

7
3

6
4

Version 11.0
G-30-1435-en Page 93
Device overview S7 / OPMI PROergo

Ceiling flange with lifting column

1 Lifting column
2 Selector switch
for setting the optimum viewing height of the surgical microscope or for
3 4 downward movement into the working position.
As long as you hold the selector switch in the turned position, the lifting
column moves downward (3) or upward (4), depending on the switch
position. When you release the selector switch, the lifting column stops
immediately.
The switching technology causes a delay of approx. 2 seconds at the
upper and lower end positions of the lifting column. After this time
elapses, you can move the column in the opposite direction again by
activating the selector switch again.
The lifting column is used to move the microscope into position for
surgery prior to the surgical procedure.
Do not constantly move the lifting column up and down, because this will
trigger the thermal cut-out, which automatically deactivates the drive
motor. If this occurs, the lifting column cannot be moved until the motor
has cooled down.
Note: After being in operation for a maximum period of one minute, the
lifting column must be turned off for at least nine minutes to prevent the
motor from overheating.

CAUTION Risk of injury caused by lifting column


Activating the lifting column may cause injuries.
• Do not activate the lifting column during surgery.
• Do not use the lifting column for focusing.
• Make sure that the patient is not put at risk by the motorized adjustment
of the lifting column.

Version 11.0
Page 94 G-30-1435-en
S7 / OPMI PROergo Device overview

Version 11.0
G-30-1435-en Page 95
Device overview S7 / OPMI PROergo

Support arm and suspension arm

1 Lamp housing with halogen or xenon illumination


The stand is equipped with an illumination system for light guides.
Each lamp housing contains a backup lamp. When a halogen illumination
system is used, the backup lamp will be automatically swung in if the first
lamp fails.
When a xenon illumination system is used, you need to swing the backup
lamp into position manually if the first lamp fails.
2 Friction
Knob for adjusting the friction of the suspension arm's upward/downward
movement.
3 Clamping screw for limiting downward movement
The movement of the suspension arm can be limited in the downward
direction.
Adjustment range: from the horizontal position of the suspension arm to
its lower stop.
4 Balancing
- Balance + Knob for setting the spring force for balancing. After all the accessories
have been fitted on the surgical microscope, this adjustment screw can be
used to set the balance of the suspension arm. The procedure for setting
the balance is described in the section "Preparations for use".
5 Friction
Knob for adjusting the friction of the suspension arm's swivel movement.
6 Friction
Knob for adjusting the friction of the carrier arm's swivel movement.
7 Control and display panel
The control and display panel permits you to control all electrical functions
of the S7 suspension system and the surgical microscope.
8 Connector panel
Further connectors are provided for powering peripheral equipment.
A foot control panel (option) can be connected to a multi-point
connector.

Version 11.0
Page 96 G-30-1435-en
S7 / OPMI PROergo Device overview

Version 11.0
G-30-1435-en Page 97
Device overview S7 / OPMI PROergo

Connector panel of S7 ceiling mount with lifting column

CAUTION Risk of electric shock caused by unapproved or defective accessories!


Unapproved or defective accessories can cause increased leakage current!
• When setting up your system, be sure to comply with the requirements
of IEC 60601-1, section 16 (in some countries IEC 60601-1-1 may still be
valid).
• Comply with the Medical Device Operator Ordinance or other national
laws.
1 Opening for system cable of external video camera
2 Remote connection
For controlling external devices (e.g. MEDIALINK 100) with a maximum
switching capacity of 24 V / 0.5 A.
3 Selector switch
2 1 For setting the optimum viewing height of the surgical microscope or for
moving it upwards into the standby position.
While the selector switch is held in the turned position, the lifting column
moves upward (1) or downward (2), depending on the switch position.
When you release the selector switch, the lifting column stops
immediately.
The switching technology causes a delay of approx. 2 seconds at the
upper and lower end positions of the lifting column. After this time
elapses, you can move the column in the opposite direction again by
activating the selector switch again.
After being in operation for a maximum of one minute, the lifting column
must be turned off for at least nine minutes to prevent the motor from
overheating.
4 Potential equalization pin
This connector can be used to connect other active devices to the same
ground potential or for redundant connection to the protective ground.
5 connection strip for switching element (option)
Connection option for a foot control panel
6 Strain relief
The strain relief prevents inadvertent disconnection of the following
electrical connections:
– video connection cable,
– connection cable for the foot control panel (optional).
7 BNC video port (option)
For VBS video signal output.

Version 11.0
Page 98 G-30-1435-en
S7 / OPMI PROergo Device overview

8 Y/C video port (option)


For S-VHS video signal output.
9 Display window for rated voltage
The rated voltage displayed here must match the rated voltage of the
power grid available at the installation site. You can adjust the sliding
switch using a suitable tool.
10 CAN bus connector (not supported)

1 2 3 4 5 6 7 8 9 10

Version 11.0
G-30-1435-en Page 99
Device overview S7 / OPMI PROergo

S7 wall mount and S7 Centro suspension system

The functions of the S7 Centro suspension system are identical to those of the
S7 wall mount.

Set-up

1 Wall anchor plate (for S7 wall mount only)


2 Control panel
3 Connector panel
4 Lamp housing (either xenon or halogen illumination)
5 Suspension arm
6 Support arm

Version 11.0
Page 100 G-30-1435-en
S7 / OPMI PROergo Device overview

6 5

Version 11.0
G-30-1435-en Page 101
Device overview S7 / OPMI PROergo

Support arm and suspension arm

1 Lamp housing with halogen or xenon illumination


The stand is equipped with an illumination system for light guides.
Each lamp housing contains a backup lamp. When a halogen illumination
system is used, the backup lamp will be automatically swung in if the first
lamp fails.
When a xenon illumination system is used, you need to swing the backup
lamp into position manually if the first lamp fails.
2 Friction
Knob for adjusting the friction of the suspension arm's upward/downward
movement.
3 Clamping screw for limiting downward movement
The movement of the suspension arm can be limited in the downward
direction.
Adjustment range: from the horizontal position of the suspension arm to
its lower stop.
4 Balancing
- Balance + Knob for setting the spring force for balancing. After all the accessories
have been fitted on the surgical microscope, this adjustment screw can be
used to set the balance of the suspension arm. The procedure for setting
the balance is described in the section "Preparations for use".
5 Friction
Knob for adjusting the friction of the suspension arm's swivel movement.
6 Friction
Knob for adjusting the friction of the carrier arm's swivel movement.
7 Control and display panel
The control and display panel permits you to control all electrical functions
of the S7 suspension system and the surgical microscope.
8 Connector panel
Further connectors are provided for powering peripheral equipment.
A foot control panel (option) can be connected to a multi-point
connector.

Version 11.0
Page 102 G-30-1435-en
S7 / OPMI PROergo Device overview

Version 11.0
G-30-1435-en Page 103
Device overview S7 / OPMI PROergo

Connector panel of S7 ceiling mount

CAUTION Risk of electric shock caused by unapproved or defective accessories!


Unapproved or defective accessories can cause increased leakage current!
• When setting up your system, be sure to comply with the requirements
of IEC 60601-1, section 16 (in some countries IEC 60601-1-1 may still be
valid).
• Comply with the Medical Device Operator Ordinance or other national
laws.
1 Opening for system cable of external video camera
2 Remote connection
For controlling external devices (e.g. MEDIALINK 100) with a maximum
switching capacity of 24 V / 0.5 A.
3 Potential equalization pin
This connector can be used to connect other active devices to the same
ground potential or for redundant connection to the protective ground.
4 connection strip for switching element (option)
Connection option for a foot control panel
5 Strain relief
The strain relief prevents inadvertent disconnection of the following
electrical connections:
– video connection cable,
– connection cable for the foot control panel (optional).
6 BNC video port (option)
For VBS video signal output.
7 Y/C video port (option)
For S-VHS video signal output.
8 Display window for rated voltage
The rated voltage displayed here must match the rated voltage of the
power grid available at the installation site. You can adjust the sliding
switch using a suitable tool.
9 CAN bus connector (not supported)

Version 11.0
Page 104 G-30-1435-en
S7 / OPMI PROergo Device overview

1 2 3 4 5 6 7 8 9

Version 11.0
G-30-1435-en Page 105
Device overview S7 / OPMI PROergo

Connectors for the optional wall outlet (ceiling mount)

CAUTION Risk of electric shock caused by unapproved or defective accessories!


Unapproved or defective accessories can cause increased leakage current!
• When setting up your system, be sure to comply with the requirements
of IEC 60601-1, section 16 (in some countries IEC 60601-1-1 may still be
valid).
• Comply with the Medical Device Operator Ordinance or other national
laws.

1 Remote connection
For controlling external devices (e.g. MEDIALINK 100) with a maximum
switching capacity of 24 V / 0.5 A.
2 connection strip for switching element (option)
Connection option for a foot control panel

Version 11.0
Page 106 G-30-1435-en
S7 / OPMI PROergo Device overview

Version 11.0
G-30-1435-en Page 107
Device overview S7 / OPMI PROergo

Illumination system

Two different types of illumination are available for the suspension system.
1 Halogen illumination
The halogen light source is equipped with an illumination system for
fiber illumination. The lamp housing contains a backup lamp which
automatically swings into position when the first lamp fails.
2 XENON illumination
The xenon light source is equipped with an illumination system for fiber
illumination. The xenon lamp generates light whose spectrum resembles
that of natural daylight. Regardless of the brightness setting, the color
temperature of the light always remains the same. Normal daylight film
without any additional conversion filters can therefore be used for
photographic documentation. The lamp module contains two xenon
lamps. The second lamp is used as a backup lamp which must be manually
swung into the illumination beam path when the first lamp fails. You have
to pull out the lamp module all the way before being able to swing in the
backup lamp.

CAUTION Risk of injury caused by xenon light in the facial area


Surgery using xenon light in the facial area may seriously harm the patient's
eyes.
• The patient's eyes must not be exposed to xenon light.
• This xenon light source must not be used in ophthalmology.

Version 11.0
Page 108 G-30-1435-en
S7 / OPMI PROergo Device overview

Version 11.0
G-30-1435-en Page 109
Device overview S7 / OPMI PROergo

Halogen lighting system

The illumination system has been designed for fiber illumination. The lamp
housing contains a backup lamp which automatically swings into position
when the first lamp fails.
1 Lamp module
2 Ventilation grid
NOTE Overheating of the lamp module!
If ventilation grids are covered e.g. by drapes, this may lead to overheating
of the ventilation modules and deactivation of the lamp.
• Never cover the ventilation grids.
• Lamps switched off due to overheating will be automatically switched on
again when they have cooled down.
3 Flap
The flap is used as the mechanical indicator for the operating status of the
halogen lamps.
– When the flap is closed, the main lamp is operative.
– When the flap is open, the main lamp has failed. The backup lamp
is on.
4 Manual switching to the backup lamp
If the automatic activation function fails, press this button to switch on the
backup lamp.
5 Opening the lamp module
When you press this button, the slide-in lamp module is partially ejected.
Pull out the lamp module all the way to change the lamp.
6 Filter selector
The filter selector knob has four positions:

0 No filter
1 GG475 filter: to protect the patient's eyes against unnecessary (blue)
radiation during surgery (retinal injury).
2 KK 40 filter: to increase color temperature
3 No filter

Version 11.0
Page 110 G-30-1435-en
S7 / OPMI PROergo Device overview

1
2

Version 11.0
G-30-1435-en Page 111
Device overview S7 / OPMI PROergo

Xenon lighting system

CAUTION Risk of injury caused by xenon light in the facial area


Surgery using xenon light in the facial area may seriously harm the patient's
eyes.
• The patient's eyes must not be exposed to xenon light.
• This xenon light source must not be used in ophthalmology.
The xenon illumination system has been designed for fiber illumination.
The xenon lamp generates light whose spectrum resembles that of natural
daylight. Regardless of the brightness setting, the color temperature of the
light always remains the same. Normal daylight film without any additional
conversion filters can therefore be used for photographic documentation.
The lamp housing contains two xenon lamps. The second lamp is used as
a backup lamp which has to be swung into the illumination beam path when
the first lamp fails.

NOTE Overheating of the lamp module!


If ventilation grids are covered e.g. by drapes, this may lead to overheating
of the ventilation modules and deactivation of the lamp.
• Never cover the ventilation grids!
• Lamps switched off due to overheating will be automatically switched on
again when they have cooled down.
1 Lamp module
2 Manual switching to the backup lamp
• If the xenon lamp fails, open the lamp module as follows: Press the
button (7). The lamp module will eject partially.
• Pull out the lamp module as far as it will go.
• Turn knob (2) through 180° until it snaps in place. This moves the backup
lamp into the illumination beam path.
• Push the lamp module all the way back into the lamp housing.
• Reset service hour counter (5) to "0". Use a pointed object and press it into
the recess of reset button (6).

Version 11.0
Page 112 G-30-1435-en
S7 / OPMI PROergo Device overview

Version 11.0
G-30-1435-en Page 113
Device overview S7 / OPMI PROergo

When inserting a new lamp module, make sure that the knob (2) is set to "1".
If the first lamp fails, switch to the second lamp in logical sequence.
3 Indicator: backup lamp is in use
When the segment in knob (2) lights up, the backup lamp is in use.
4 Filter selector
The filter selector knob has two positions:
0 No filter
1 Filter moved into position
(there is no integrated filter in the standard configuration)

5 Counter
The counter records the service hours of the xenon light source.
• Change the xenon lamps after a maximum operating time of 500 hours to
avoid any rupture of the xenon lamps. Then reset the counter to "0" by
pressing the "Reset" button (6).
6 Reset button
The reset button resets the service hour counter to "0".
7 Opening the lamp module
When you press this button, the slide-in lamp module is partially ejected.
• To change the lamp, pull out the lamp module as far as it will go. Turn the
knob (2) through 180° until it snaps in place. This moves the backup lamp
into the illumination beam path.

CAUTION The xenon lamp has a limited service life of 500 h!


If used beyond its maximum service life, the xenon lamp may explode.
• Please replace the xenon lamp in due time.
• Reset the service hour counter to "0" after replacing the lamp.

Version 11.0
Page 114 G-30-1435-en
S7 / OPMI PROergo Device overview

3
4

Version 11.0
G-30-1435-en Page 115
Device overview S7 / OPMI PROergo

Control panel and menu overview

The control panel consists of a graphic display (1) with control keys (3).
All functions can be interactively controlled via menus. The display shows
the selected functions and settings. Using control buttons (3) you can move
through the menus and activate a selection field. The field activated is then
highlighted by a dark background. Press Enter button (4) to confirm your
selection or entry.
For entering, briefly press any of control buttons (3). In some settings,
the control buttons have a repeat function, i.e. constant pressing of a button
causes the command to be repeated.
The menus are clearly laid out and user-friendly. The name of the menu is
given in bar (2) at the bottom. At the right-hand end of the bar, a small
pictogram is shown for quick orientation. At the left-hand end of the bar,
the activation field (OK) is located. If the background of this field is dark,
press the "Enter" button to confirm your entry.
The following diagram depicts how the menus are linked.
Power switch (5)
For switching the system on and off. When the system has been switched on,
the green lamp in the switch is lit.
The power switch contains an automatic circuit breaker. If a short circuit
occurs, the automatic circuit breaker will respond and switch the system off.

2
4
3

Version 11.0
Page 116 G-30-1435-en
S7 / OPMI PROergo Device overview

Version ......

Please wait ...

USER 1
USER 2 NONE

USER 3 NONE

50 % 1.4 Off 50 % 300 mm Off 50 % 50 %

Off Off

4.9 x 300 mm 20 %

Standard

Close

55 mm + 0.00 + 0.00 12.5 x

12.5 x

Off

Version 11.0
G-30-1435-en Page 117
Device overview S7 / OPMI PROergo

Structure of the menus

– If no entry is made for about 30 seconds, the preceding menu will


automatically reappear.
– The data you have entered or changed will not be saved until you press
the OK button or the Enter button, or when the preceding menu appears
after approx. 30 seconds because no further entries have been made.

Firmware version 1.1 Start of menu


After you have switched on the system, the software is loaded. The growing
size of the ZEISS logo with the words "Please wait..." below it indicates that
this test is still in progress. The release number is displayed in the top left
Please wait ... corner. The system then performs a self-test.
After successful completion of the test, the "User Selection" menu is
automatically displayed.

User menu
The User menu comprises one standard profile and three other user profiles.
USER 1
When a user profile is selected, the system imports the settings stored under
USER 2
this profile, e.g. focusing speed, zoom link setting, zoom speed and handgrip
USER 3
settings. The stored start values ("Start at") for the magnification factor,
the focus position and the brightness are automatically set. The three user
profiles are used to adjust these settings for different users.
When the system is started up for the first time, the Standard User is
displayed; at every subsequent start-up, the user active last is displayed.
– The Standard User profile allows the adjustment of settings, but these
settings cannot be saved (fixed factory settings).
– In the Standard User profile, all "Start at" fields are deactivated. No bar is
displayed; it is not possible to adjust any settings.

Version 11.0
Page 118 G-30-1435-en
S7 / OPMI PROergo Device overview

Edit User Menu


1
Three different users or applications (e.g. Endo, Paro, Cosmetic) and their
profiles can be stored in the User menu. The Edit User menu allows you to
Dr. L. Weber enter your personal user name. When Edit icon (1) after the user name is
activated, the input keyboard is displayed. This keyboard allows you to enter
a 12-digit user name or application name. Then activate the "OK" key (2)
using the arrow keys and press the "Enter" button to save the name you
entered.
Dr. L. Weber
A user or application name is displayed in the input field. If no name has been
entered, only User 1, 2 or 3 will be displayed.
After selecting a user or an application, press the "Enter" button to return to
the "Surgery" menu.

Functions of buttons:
– When the input keyboard has been activated with the Edit button (1),
the input function is active.

2
– Select the individual characters using the arrow keys and confirm the
characters by pressing the "Enter" button.

Delete one or more characters from right to left by pressing the


"Enter" button.
Space Space bar; press the "Enter" button to insert a blank space.
Shift: switches to capital letters and back again when the "Enter"
button is pressed.
Cursor control, for insertion or deletion of individual characters in
<> the text.

You must activate the OK button (2) using the arrow buttons and then
press the Enter button to save the name entered under the selected user.
After selecting a user or an application, press the "Enter" button to return to
the "Surgery" menu.

Version 11.0
G-30-1435-en Page 119
Device overview S7 / OPMI PROergo

Surgery menu
The Surgery menu is constantly displayed during operation. The menu
4.9 x 300 mm 20 % includes the icons (from left to right) for the following menus: Zoom, Focus,
Illumination. Below: User, Options, Video and Balance. The displays below the
Standard zoom, focus, illumination and user icons show the current settings for total
magnification, working distance, brightness level and selected User.

– The display shows the name of the user or application shortened to


10 characters.
– The displayed values are rounded. which are only provided for
information, not for measurement purposes.
Use the control buttons to activate the appropriate selection field and press
the enter button to go to the desired menu.

Zoom menu
You can adjust the settings for the following functions in the Zoom menu:

50 % 1.4
Speed:
The zoom speed from minimum to maximum speed from 5% to 100% in
increments of 5%.
Start at:
the zoom start value. Enter the magnification factor at which the zoom system
should start when the system is switched on or is switched to a different user.
The magnification factor can be set and saved within a range between
γ = 0.4 x and γ = 2.4 x in increments of 0.1.
γ is only the magnification factor of the zoom system. The total magnification
provided by the system is also determined by the eyepieces and the working
distance used.
• Activate the function desired (Speed or Start at) and move the indicator
bar using one of the arrow buttons ( , ), until the reading required is
displayed in the field at the top.
• Press the Enter button to save the value set and to return to the Surgery
menu. If you do not press any key, the display will automatically jump back
to the Surgery menu after approx. 30 seconds.
In the Standard User profile, all "Start at" fields are deactivated. No bar is
displayed; it is not possible to adjust any settings.

Version 11.0
Page 120 G-30-1435-en
S7 / OPMI PROergo Device overview

Focus menu
You can adjust the settings for the following functions in the Focus menu:
Zoom Link:
On 50 % 300 mm
The Zoom Link function lets you couple the focusing speed to the current
On
magnification. This means that the focusing speed is automatically reduced
when a high magnification is used. Whether the Zoom Link function has been
activated or deactivated is stored in the respective user profile.

Speed:
Focus
The focusing speed from minimum to maximum speed from 5% to 100% in
increments of 5%.
OnOn

Start at:
focus start value. Adjust the working distance at which the focusing system
should start when you call your user profile. The setting can be adjusted and
saved within a range of 200 to 415 mm in increments of 1 mm.
The value shown in the graphic display is not a measuring function. It is only
intended as an aid for orientation and may not be used as a measurement
value.
• Activate the function desired and move the indicator bar using one of the
arrow buttons ( , ) until the reading required is displayed in the field
at the top.
• Press the Enter button to save the value set and to return to the Surgery
menu. If you do not press any button, the display will automatically
jump back to the Surgery menu after approx. 30 seconds and save the
values set.

Version 11.0
G-30-1435-en Page 121
Device overview S7 / OPMI PROergo

Illumination menu
You can adjust the settings for the following functions in the Illumination
menu:
Zoom link
On 50 % 50 % Intensity:
On the illumination intensity from minimum to maximum brightness from
5% to 100% in increments of 5%. This setting can be adjusted for the xenon
illumination and the halogen illumination.
Start at:
the brightness start value. Use one of the arrow buttons ( , ) to adjust the
illumination intensity in percent. The xenon or halogen lamp will then go on
at this brightness level after activation.
Illumination
The "Start at" field will be active only if "Zoom Link" is in the "Off" position.
On If Zoom Link is On, the "Start at" field will not be active. No bar is displayed;
On
it is not possible to adjust any settings.
Zoom Link:
the automatic light readjustment. When this function is active (ON),
the illumination is adjusted in such a way that the user is provided with
a subjectively constant brightness level in the eyepiece, regardless of the
zoom factor set. Whether the Zoom Link function has been activated or
deactivated is stored in the respective user profile.
• Press the Enter button to save the value set and to return to the "Surgery"
menu. If you do not press any button, the display will automatically jump
back to the Surgery menu after approx. 30 seconds and save the values
set.

Version 11.0
Page 122 G-30-1435-en
S7 / OPMI PROergo Device overview

Options menu "Options"


Activate the "Options" icon and press the enter button to access the
"Options" menu. The Options menu comprises the following submenus
(from the top):
– "Handgrip Settings" menu
– "Foot Control" menu
– "Binocular Settings" menu
– "Eyepiece Settings" menu
– "Magnetic Brakes" menu
– "Autofocus" menu (option)

Version 11.0
G-30-1435-en Page 123
Device overview S7 / OPMI PROergo

Handgrip Settings menu


This menu allows you to assign one of the listed functions to each of the four
1 2
buttons (A, B, C, D) for every user or application:
– Light on/off: switching the illumination on and off.
– Light +/-: adjusting the brightness level of the illumination.
– "Video Capture": triggers the device connected to the "Remote" socket
Freeze/Live
(on the connector panel of the suspension system).
– Photo: triggering an electrical contact in the suspension system. Here,
a camera can be connected (option).
– Autofocus: triggering the SpeedFokus autofocus option.
– Video Gain +/-: adjusting the brightness level of the integrated video
system (option).
– OSD on/off: activating/deactivating the on-screen display of the
integrated video system (option).
– "Freeze/Live": the current live video image is frozen and displayed as a full-
screen freeze frame on the monitor.
The "Video Gain +/-" and the "OSD on/off" functions are offered for selection
only if your system is equipped with an integrated video camera (option).
Without an integrated video camera (option), these fields remain blank.
In addition, you can either assign the same functions to both handgrips by
pressing button (1) or assign different functions to the left and right handgrip
by pressing button (2).
Function assignment:
Use the arrow buttons ( , ) to select the required function from the table.
Use the arrow buttons ( , ) to select the handgrip button required (A-D).
After you have pressed the "Enter" button, a check mark ( ) indicates that
the selected function has been assigned to the selected button.
When you press the "Enter" button again, your selection is saved and you
return to the "Surgery" menu.

Foot Control menu


Your suspension system permits the connection of a foot control panel
(option). This menu allows you to assign one of the listed functions to each of
the buttons C and D on the foot control panel for every user:
– "Video Capture": triggers the device connected to the "Remote" socket
(on the connector panel of the suspension system).
– Autofocus: triggering the SpeedFokus autofocus option.

Version 11.0
Page 124 G-30-1435-en
S7 / OPMI PROergo Device overview

Assignment of functions:
Use the ( , ) arrow buttons to select the required function from the table.
Use the ( , ) arrow buttons to select the button required (C or D).
After you have pressed the "Enter" button, a check mark ( ) indicates
that the selected function has been assigned to the selected button.
When you press the "Enter" button again, your selection is saved and you
return to the "Surgery" menu.

Binocular Settings memory function


55 mm + 0.00 + 0.00
This menu only has a memory function. It provides sliders which enable each
user to set his/her interpupillary distance and prescription using one of the
arrow buttons ( , ) and to save these values. No motorized setting is made
on the eyepieces or tube.
This function is intended to allow OR staff to adjust the surgical microscope
to the surgeon's settings before the procedure and, if necessary, to retrieve or
change the stored values.
– You can set the interpupillary distance in 1 mm increments from 55 mm
to 80 mm.
– When using widefield eyepieces with a magnetic coupling, you can set
your prescription between –5 D and 8D.
Eyepiece Settings menu
12.5 x
In this menu, you can mark the type of eyepiece (10x or 12.5x) used on the
12.5 x
surgical microscope. This is required for the correct display of the total
10 x magnification provided by the system. Use the arrow buttons ( , ) to
select the eyepiece type. Press the "Enter" key to save the eyepiece type used.
Press the Enter button to save the value set and to return to the Surgery menu.
If you do not press any key, the display will automatically jump back to the
Surgery menu after approx. 30 seconds.

On Magnetic Brakes menu


On This menu permits you to activate or deactivate the magnetic brakes.
In the "On" mode, the magnetic brakes are active.
In the "Off" mode, the magnetic brakes are inactive.
In this mode, you can adjust the friction of the microscope's axes using the
relevant adjustment knobs.
If "Off" has been activated by pressing the Enter button, the magnetic brakes
will be unlocked and remain unlocked when you press release button(1) the
next time.
1
To reactivate the magnetic brakes, select "On" and press the Enter button.
The magnetic brakes are now active again, i.e. when you press release button

Version 11.0
G-30-1435-en Page 125
Device overview S7 / OPMI PROergo

(1), the magnetic brakes are unlocked and you can position the balanced
surgical microscope. When you let go of release button (1), the magnetic
brakes are locked and the surgical microscope is fixed in position again.

CAUTION Injury to the patient


If the system is in an extremely unbalanced state, the unit can move
uncontrollably out of position.
• Balance the completely equipped surgical microscope before surgery
without the patient. to permit almost effortless guidance of the surgical
microscope.
• Hold the handgrips of the surgical microscope firmly before releasing the
magnetic brakes.

Version 11.0
Page 126 G-30-1435-en
S7 / OPMI PROergo Device overview

Option "SpeedFocus (Autofocus)"


The SpeedFokus autofocus option permits you to select and automatically
focus on a feature in the ROI (region of interest, measuring field) in the image
of the surgical field.
The SpeedFokus (autofocus) function is based on the analysis of the video
image contrast: during focusing of the microscope's Varioskop optics,
SpeedFokus continuously determines the video image contrast in the selected
ROI. In this way, SpeedFokus controls the microscope's Varioscope optics and
adjusts it to the point of maximum contrast. This corresponds at the same
time to the point of greatest image sharpness.
Autofocus menu
The Autofocus menu offers you the following functions for selection:
– ROI, on/off: activating/deactivating the ROI display.
– ROI Size, small/medium/large: selecting the ROI size.
– ROI Position: branching to the AF - ROI Position submenu to define the
ROI position.
– "Brake Automatic, on/off", activating/deactivating the automatic start of
the focusing process
– Camera Position, left/right: entering the orientation of the external video
camera.
– Advanced Settings: branching to the Advanced Settings submenu to enter
the technical data of the external video camera.
– The "Camera Position left/right" and "Advanced Settings" functions are
only offered for selection if your system is equipped with an external video
camera. If you have an integrated video camera, these fields remain blank.
– As long as you are entering any settings in the Autofocus menu or its
submenus, the ROI is displayed as a square frame in the video image on
the monitor. A few seconds after your last action, the frame disappears
from the screen.
Menu item "ROI, on/off"
The position and size of the ROI are displayed on the video monitor in the
form of a square frame. Use the "ROI on/off" function to activate or
deactivate this display on the video monitor.
Menu item "ROI Size, small/medium/large"
Permits you to select the ROI size.
Three different ROI sizes are available for selection: small (20x20 pixels),
medium (36x36 pixels) and large (50x50 pixels).

Version 11.0
G-30-1435-en Page 127
Device overview S7 / OPMI PROergo

The default setting for Standard User is the large ROI. You can change the size
of the ROI, but the change cannot be saved for the Standard User.
For specific users, the ROI size can be set and saved for each user as required.
Submenu "AF - ROI Position"
This submenu permits you to define the ROI position within the microscope's
field of view.
When you start your system, the ROI is always positioned at the center of the
field of view. The ROI is visualized as a square in the graphic display. Use the
arrow buttons to shift the ROI.
Press the OK button to return to the Autofocus menu. If you do not press any
of the buttons, the display will automatically jump back to the Surgery menu
after approx. 20 seconds.
Menu item "Brake Automatic, on/off"
This function permits you to activate or deactivate the automatic start of the
focusing process.
For repositioning the OPMI PROergo, you have to unlock its magnetic brakes.
Press the release button for the magnetic brakes on one of the handgrips. The
magnetic brakes of the microscope are unlocked for as long as you press one
of these release buttons.
If you have selected "Brake Automatic, on", focusing by the SpeedFokus
(autofocus) option will start automatically as soon as the magnetic brakes
have been locked, i.e. as soon as you let go of the release buttons.
If you have selected "Brake Automatic, off", focusing by the SpeedFokus
(autofocus) option will not start automatically when the magnetic brakes
are locked, i.e. when you let go of the release buttons. You can trigger the
focusing process using the handgrips or the foot control panel. This is only
possible if the Autofocus option has been selected in the handgrip menu
or foot control menu (for foot control panels with 8 or 14 functions).
You can stop focusing by the SpeedFokus (autofocus) option by pressing one
of the release buttons on the handgrips or on the foot control panel (with 8
or 14 functions) during the focusing process.
Menu item "Camera Position"
If you are using an external video camera, the orientation of the video camera
must be known to the video autofocus.
This function is only available in systems equipped with an external video
camera.
The video image must have the same orientation as the image in the main
eyepieces of the surgical microscope.
In addition, the video autofocus needs the information whether the video
camera is connected on the left or right, i.e. whether the beam splitter

Version 11.0
Page 128 G-30-1435-en
S7 / OPMI PROergo Device overview

projects into the main beam path from the left or right, to enable the video
focus to perform corrections in the right direction during the focusing
process.
Submenu "Advanced Settings"
– This submenu is only available in systems equipped with an external video
camera.
– This submenu contains camera-specific settings. These settings only need
to be changed when a different video camera is installed.
– The SpeedFokus (autofocus) option does not function properly if
parameters have not been correctly set.
The "Advanced Settings" submenu contains the following camera-specific
settings:
– Video Focal Length
– Camera Chip Size (size of the CCD array)
Video Focal Length and Camera Chip Size (size of the CCD array):
These settings must conform to the technical specifications of the video
camera used.
If you are using the "1Chip HD Camera" option on a surgical microscope
featuring SpeedFokus, you have to configure the following software settings
on the surgical microscope to ensure correct visualization of the image
format:
• set the focal length of the video optics to f = 50 mm and the size of the
CCD array to 1/3".

Version 11.0
G-30-1435-en Page 129
Device overview S7 / OPMI PROergo

Balance menu
The Balance menu has been designed as a user prompt and consists of six
images explaining the balancing procedure. You can scroll through the
Balance menu using the arrow buttons ( , ). The OK button has no
function.
Press the enter button or the arrow buttons ( , ) several times to return
to the "Surgery" menu.

The balancing procedure is described in the "Preparations" chapter under


"Adjusting the system"

Version 11.0
Page 130 G-30-1435-en
S7 / OPMI PROergo Device overview

Laser menu
Since this surgical microscope model does not allow the attachment of
a laser, the Laser menu is not used.
You should therefore check that the default setting "NO NE" - No Laser is
displayed in all menus (1) to ensure problem-free operation.
1 The User Selection menu permits you to access the laser selection menu
via the respective Edit button using the arrow button ( ). After you have
User 1 NoNONE
Laser activated Laser menu button (1), a pull-down menu is automatically opened
User 2 No Laser
TwinER in the display.
User 3 No Laser
TwinER
Note: The pull-down menu will be automatically opened if you approach from
the left with the arrow button ( ). If you select the field using the ( , )
arrow buttons, the pull-down menu can be opened by pressing the ENTER
key.
Use the arrow keys ( , ) to switch to None- No Laser.
By activating the setting "None" and pressing the Enter button, you can
transfer the selection to the Laser menu (1).
If you do not press any key, the display will automatically jump back to the
Surgery menu after approx. 30 seconds.

Version 11.0
G-30-1435-en Page 131
Device overview S7 / OPMI PROergo

Malfunctions (Error message menu)


This window appears on your display when a malfunction or incorrect
operation has occurred. At the same time, this is also indicated by three
1 beeps.
The type of malfunction present and a code number are displayed in field (1).
HR:-9005
The letters preceding the code number indicate the system component in
HANDGRIP RIGHT which the malfunction has occurred:
handgrip button pressed S - suspension system, HR - right handgrip, HL - left handgrip, F - foot control
Cancel
panel, V - video, O - surgical microscope, C - (core) software.
This code number permits you to identify the malfunction that has occurred,
E.g.:
HR:-9005
RIGHT HANDGRIP
handgrip button pressed

You can delete the error message by pressing the "Enter" button. Internally,
the system keeps a record of the malfunctions occurred.
If you cannot eliminate the malfunction, you may perhaps be able to continue
working by switching off the suspension system or separating the plug-in
connection to the surgical microscope (see "Failure of main functions",
Page 199).

Acoustic signals

Three successive beeps – Error message in case of a malfunction


or incorrect user operation.
One short beep – Confirms that the contact in the Remote
socket is activated.
– After power-on of the suspension
system.

Version 11.0
Page 132 G-30-1435-en
S7 / OPMI PROergo Device overview

Version 11.0
G-30-1435-en Page 133
Device overview S7 / OPMI PROergo

Optional foot control panel

The foot control panel permits you to control various functions of


a suspension system or a surgical microscope. The assignment of the
functions to the buttons on the foot control panel is described in the following
pages. You can only control functions which are implemented in the system
configuration used (stand, surgical microscope).

Version 11.0
Page 134 G-30-1435-en
S7 / OPMI PROergo Device overview

Foot control panel for zoom and focus

1 FOCUS : reducing the working distance


2 FOCUS : increasing the working distance
3 ZOOM : increasing the field of view, reducing magnification
4 ZOOM : reducing the field of view, increasing magnification

2 4

3
1

Version 11.0
G-30-1435-en Page 135
Device overview S7 / OPMI PROergo

14-function foot control panel, wired (FCP) and wireless (FCP WL)

A retrofit kit for the 14 function foot control panel is available as an option.
The prerequisite for this is that your device is equipped with the foot control
panel connector socket.
Please observe the G-30-1707 Instructions for Use which are contained in the
delivery scope for the 14 function foot control panel.

1 FCP WL (wireless 14 function foot control panel) - (view from above)


2 FCP (wired 14 function foot control panel) - (rear aspect with plugged in
connection lead)
3 Upgrade Kit

Version 11.0
Page 136 G-30-1435-en
S7 / OPMI PROergo Device overview

1 2 3

Version 11.0
G-30-1435-en Page 137
Device overview S7 / OPMI PROergo

1Chip HD camera (option)

1 1Chip HD camera (option)


flanged to a beam splitter
2 CCU (Camera Control Unit)
Control unit for the 1Chip HD camera

2 1

Version 11.0
Page 138 G-30-1435-en
S7 / OPMI PROergo Preparations for use

Preparations for use

Attaching the equipment ..............................................................142


Attaching the tube and eyepieces ...........................................................142

Connections, establishing ............................................................. 144


Connecting the strain relief .....................................................................144
Connecting the suspension system - S7 floor stand .................................146
Connecting the suspension system - S7 ceiling mount with
rigid column ...........................................................................................148
Connecting the suspension system - S7 ceiling mount with
lifting column .........................................................................................150
Connecting the suspension system - S7 wall mount ................................152
Connecting the video monitor and the MEDIALINK 100 (option) ............154

1Chip HD camera option ...............................................................156


Installing the 1Chip HD camera ..............................................................156
1Chip HD camera with iris diaphragm for video lenses ............................157
Cable connections for the 1Chip HD camera ...........................................158

Adjusting the system - S7 floor stand ........................................... 160


Setting the suspension arm weight compensation ..................................160
Limiting the suspension arm's downward movement .............................160

Adjusting the system - Ceiling mount, wall mount and S7


Centro suspension system ............................................................ 162
Setting the suspension arm weight compensation ..................................162
Limiting the suspension arm's downward movement .............................162

Adjusting the system - S7 ceiling mount with lifting column ........164


Set ergonomic working height ................................................................164
Setting the suspension arm weight compensation ..................................166
Limiting the suspension arm's downward movement .............................166

Version 11.0
G-30-1435-en Page 139
Preparations for use S7 / OPMI PROergo

Balancing the surgical microscope ................................................168

Adjusting the surgical microscope ................................................173

Preparing the device for sterile use ..............................................175


Attaching asepsis caps ........................................................................... 175
Attaching the drapes ............................................................................. 175

Relocating the S7 floor stand .......................................................176

Version 11.0
Page 140 G-30-1435-en
S7 / OPMI PROergo Preparations for use

WARNING Risk of serious injury due to unauthorized modification!


You must not modify this device without the permission of the manufacturer.
The manufacturer is not liable for damages caused by an unauthorized use
of the device.
Unauthorized modifications of the device may result in severe, irreversible
injuries to users and/or patients.
• If you modify the device after consultation with the manufacturer,
you must carry out the appropriate inspections and tests to ensure that
it can still be used safely.

System kit Any additional equipment you connect to medical electrical devices must
demonstrably comply with the applicable standards and directives (e.g.
IEC 60950-1 for data processing equipment). In addition, all configurations
must meet the requirements set out in the standards for medical electrical
systems (see IEC 60601-1, chapter 16).
Anyone connecting additional equipment to medical electrical devices is
a system configurer and, as such, is responsible for compliance of the system
with the standard requirements for medical electrical systems.
Please note that local legislation takes priority over the requirements of the
above-mentioned standards. If you have any questions, contact your local
dealer or Carl Zeiss Service.

NOTE This equipment must not be modified without the manufacturer’s


approval!

R017 Your devices may not be modified without the manufacturer’s approval.
If the system is modified after consultation with the manufacturer, suitable
inspections and testing must be completed to ensure subsequent safe use.
The manufacturer is not liable for damages caused by an unauthorized use
of the device. Furthermore, such unauthorized modifications will render the
warranty and any claims under warranty invalid.

Version 11.0
G-30-1435-en Page 141
Preparations for use S7 / OPMI PROergo

Attaching the equipment

Attaching the tube and eyepieces

CAUTION Risk of injury from the surgical microscope or its accessories


falling down!
• Never change modules and accessories during a surgical procedure or
above the patient.
• Attach the accessory modules to the microscope and firmly tighten the
respective securing screw.
• Before every use and after re-equipping the instrument, make sure that
the modules are securely locked in position.
• Make sure all the securing screws are firmly tightened!
Mounting and removing modules and accessories may lead to unintended
lowering of the suspension arm.
• Ensure that the maximum permissible load is never exceeded.
• Always re-balance the surgical microscope after re-equipping the system.
• Bring the suspension arm in a position convenient for you and firmly
tighten friction adjustment knob (1).
• Loosen securing screw (4) by a few turns.
• Remove cover (3) and store it in a safe place.
• Place the binocular tube (5) on the microscope body and firmly tighten the
securing screw (4).
• Further accessories may be installed between the binocular tube and the
microscope body. Lock these modules in position in the same way using
securing screw (4).
• Insert the widefield eyepieces (7) as far as they will go into their respective
mounts (6). The magnetic coupling reliably secures them in position.
If you wish to use documentation equipment, we can supply an eyepiece with
a reticle to aid focusing. The retrofitting of a reticle to an eyepiece can only
be performed in the factory or by our Service staff. Always install the eyepiece
with the reticle on the same side of the binocular tube where the
documentation equipment is located.

Version 11.0
Page 142 G-30-1435-en
S7 / OPMI PROergo Preparations for use

1 2 3 4 5 6 7

Version 11.0
G-30-1435-en Page 143
Preparations for use S7 / OPMI PROergo

Connections, establishing

Connecting the strain relief

You can secure the power plug and the multipoint connector of the switching
component against inadvertent loosening by installing the two cables in strain
relief device (1).
After you have mounted strain relief device (1), the cable must have the
following length:
– 320 mm from the strain relief device up to and including power
connector (9).
– 320 mm from the strain relief device up to connector (10) of the foot
control panel.
• Form a loop with the cable (3).
• Open flap (4).
• Feed the cable through the opening (5).
• Close the flap (6).
• Tighten the cable until it encloses the flap (7).
• Check the cable length.

Version 11.0
Page 144 G-30-1435-en
S7 / OPMI PROergo Preparations for use

2 3 4

5 6 7 8

320 mm

10

Version 11.0
G-30-1435-en Page 145
Preparations for use S7 / OPMI PROergo

Connecting the suspension system - S7 floor stand

WARNING Danger! Electrical voltage!


• To avoid the risk of electrical shock, this device may only be connected to
a power supply network which is provided with a protective ground
conductor.

CAUTION Risk of tripping over cables lying on the floor!


If cables are installed inappropriately on the floor, this presents an increased
R057
tripping hazard and interferes with the flow of work. By tripping and falling
down, users may suffer tissue trauma.
• If possible, cables should only be laid outside of the area of movement.

NOTE Observe the rated voltage


Prior to leaving the factory, the device is adjusted to the rated voltage of the
R017 respective country of destination. The voltage setting is displayed in a small
window (1) on the support arm. If the set voltage does not comply with the
line voltage available at the site of installation, this can cause damage to the
device.
• Check the voltage setting. The set voltage must comply with the line
voltage available at the site of installation. If this is not the case, you must
not operate the system.
• If the rated voltage is not correctly set, adjust the sliding switch (5) using
a suitable tool.

• Connect the plug of the foot control panel to the connection strip (2) of
the stand if you wish to use a foot control panel.
• Connect the stand using the power cord provided (4).
• If required, use the potential equalization pin (1) to connect the stand to
the equipotential bonding strip in the OR.

Version 11.0
Page 146 G-30-1435-en
S7 / OPMI PROergo Preparations for use

1 2 3 4 5

Version 11.0
G-30-1435-en Page 147
Preparations for use S7 / OPMI PROergo

Connecting the suspension system - S7 ceiling mount with rigid column

• Check the voltage indicator for (2).


The stand is factory preset to the rated voltage of the country where it will
be used. The rated voltage indicated at window (2) must correspond to
the rated voltage available on the site of installation.
• If the rated voltage is not correctly set, adjust the sliding switch (2) using
a suitable tool.
• Plug the connector of the foot control panel (option) into connector (1) of
the suspension system.

Version 11.0
Page 148 G-30-1435-en
S7 / OPMI PROergo Preparations for use

1 2

Version 11.0
G-30-1435-en Page 149
Preparations for use S7 / OPMI PROergo

Connecting the suspension system - S7 ceiling mount with lifting column

• Check the voltage indicator for (2).


The stand is factory preset to the rated voltage of the country where it will
be used. The rated voltage indicated at window (2) must correspond to
the rated voltage available on the site of installation.
• If the rated voltage is not correctly set, adjust the sliding switch (2) using
a suitable tool.
• Plug the connector of the foot control panel (option) into connector (1) of
the suspension system.

Version 11.0
Page 150 G-30-1435-en
S7 / OPMI PROergo Preparations for use

1 2

Version 11.0
G-30-1435-en Page 151
Preparations for use S7 / OPMI PROergo

Connecting the suspension system - S7 wall mount

• Check the voltage indicator for (2).


The stand is factory preset to the rated voltage of the country where it will
be used. The rated voltage indicated at window (2) must correspond to
the rated voltage available on the site of installation.
• If the rated voltage is not correctly set, adjust the sliding switch (2) using
a suitable tool.
• Plug the connector of the foot control panel (option) into connector (1) of
the suspension system.

Version 11.0
Page 152 G-30-1435-en
S7 / OPMI PROergo Preparations for use

1 2

Version 11.0
G-30-1435-en Page 153
Preparations for use S7 / OPMI PROergo

Connecting the video monitor and the MEDIALINK 100 (option)

NOTE Use of video signal isolation


When video signals are led out of the OR, e.g. for a conference, this may
cause excessive leakage and contact currents.
• Use galvanic isolation for isolating the video signals.

The image quality obtained with Y/C connections and cables shorter than
10 m is always better than that obtained with BNC connections of the same
length. If cable lengths exceed 10 m, the image quality provided by a Y/C
connection will decrease markedly; in these cases, only BNC connections
should be used.
If possible, you should position the individual units in such a way that you can
use Y/C connections. The Y/C connection provides the best image quality,
in particular for images recorded with MEDIALINK 100.
If this is not possible, position the individual units in such a way that you can
at least use the Y/C cable for the video input port of the MEDIALINK 100.

Remote control of MEDIALINK 100 using freely configurable buttons


The freely configurable handgrip buttons of the surgical microscope and the
freely configurable buttons (C and D) on the foot control panel allow you to
control external units.
The signal generated by the configurable buttons is present at the Remote
socket on the connector panel of the suspension system.
• To enable the remote control function, connect the Remote socket on the
connector panel of the suspension system to the Remote socket on
MEDIALINK 100.

Set-up
1 Standard video monitor
2 Standard video recorder, video printer, video projector
3 MEDIALINK 100
4 CCU connector panel (on back of the system) for the 1Chip HD camera
5 Connector panel of the suspension system

Version 11.0
Page 154 G-30-1435-en
S7 / OPMI PROergo Preparations for use

Devices with SpeedFocus S7 / OPMI Sensera with 1Chip HD camera option,


(Autofocus) option for external MEDIALINK 100 option and video monitor /
camera and 1Chip HD video recorder
camera option

Video monitor 1
VIDEO IN
Video recorder 2

CCU of the 1Chip


HD camera option Y/C
BNC
(FBAS)

MEDIALINK 100 3

Y/C

SD - Y/C Remote

CCU for the 1Chip HD camera

Version 11.0
G-30-1435-en Page 155
Preparations for use S7 / OPMI PROergo

1Chip HD camera option

Installing the 1Chip HD camera

Please observe the "1Chip HD camera" Instructions for Use (G-30-1946).


These Instructions for Use are included automatically in systems with the
"1Chip HD camera" option. They contain all the additional explanations for
the "1Chip HD camera" option.
The 1Chip HD camera is flanged permanently to a beam splitter or to an iris
diaphragm for video lenses. Normally the beam splitter with the 1Chip HD
camera remains installed on the surgical microscope.
However, the beam splitter with the 1Chip HD camera can be removed from
the surgical microscope if necessary.

CAUTION Risk of infection due to dangling camera cable!


If the beam splitter and the permanently connected 1Chip HD camera are
R508 removed from the surgical microscope and the remaining, non-sterile camera
cable is hanging down from the surgical microscope, the camera cable could
contaminate the sterile area with pathogens under certain circumstances.
This may cause infections.
• Make sure that the remaining camera cable is not hanging down from
the surgical microscope.

Version 11.0
Page 156 G-30-1435-en
S7 / OPMI PROergo Preparations for use

1Chip HD camera with iris diaphragm for video lenses

Normally the 1Chip HD camera is located on the right side as viewed from the
main eyepieces. The iris diaphragm for video lenses on the 1Chip HD camera
allows you to re-equip your surgical microscope quickly. The 1Chip HD camera
can be attached to and removed from the beam splitter, and therefore the
output ports of the beam splitter (left/right sides) can also be changed.
After changing sides, you will have to adjust the position of the image.
For further details please refer to the Instructions for Use of the "1Chip HD
camera" (G-30-1946), in the chapters "Using the CCU to configure the 1Chip
HD camera / Configuring the picture settings". These Instructions for Use are
included automatically in systems with the "1Chip HD camera" option.

1 Iris diaphragm for video lenses (301675-0000.000)


The iris diaphragm for video lenses is installed permanently on the 1Chip
HD camera. Use the aperture setting to adjust the brightness of the video
image.
2 Adjusting ring for the aperture
3 1Chip HD camera
4 Connector socket for the camera cable
5 Bayonet mount

1 2 3

4
5

Version 11.0
G-30-1435-en Page 157
Preparations for use S7 / OPMI PROergo

Cable connections for the 1Chip HD camera

CAUTION Risk of electric shock caused by unapproved or defective accessories!


Unapproved or defective accessories can cause increased leakage current!
• When setting up your system, be sure to comply with the requirements
of IEC 60601-1, section 16 (in some countries IEC 60601-1-1 may still be
valid).
• Comply with the Medical Device Operator Ordinance or other national
laws.

CAUTION Risk of tripping over cables lying on the floor!


If cables are installed inappropriately on the floor, this presents an increased
tripping hazard and interferes with the flow of work. By tripping and falling
R507
down, users may suffer tissue trauma.
• If possible, cables should only be laid outside of the area of movement.
Consult the Instructions for Use for the "1Chip HD Camera" (G-30-1946) for
details of the required cable connections.
You can find a list of the video accessories available for use with your surgical
microscope in the section "video accessories for the 1Chip HD camera option"
on Page 247.

1Chip HD camera on the beam splitter, Splitter 20

Camera cable
DC 12V - 15V
from the permanently installed
suspension in suspension system
system

CCU

S-Video DVI HD SDI HD SDI


(max. 5m)

video recorder video recorder video recorder video recorder


video monitor video monitor video monitor video monitor

Version 11.0
Page 158 G-30-1435-en
S7 / OPMI PROergo Preparations for use

Version 11.0
G-30-1435-en Page 159
Preparations for use S7 / OPMI PROergo

Adjusting the system - S7 floor stand

Setting the suspension arm weight compensation

CAUTION Risk of crushing - mind your fingers!


Fingers may be crushed in the areas marked with the "Risk of crushing" label.
• Do not touch these areas while the system is being moved or brought in
its working / transport position.
Do not perform the balance setting until the surgical microscope and all
accessories have been mounted!
• Move the suspension arm into its horizontal position and tightly hold it
with one hand.
• Loosen knob (3) for adjusting the friction of the suspension arm's upward/
downward movement.
• Move the suspension arm slightly up and down. At the same time, turn
adjustment screw (1) until you think that the spring force is sufficient to
compensate for the weight of the surgical microscope and accessories.
Continue to perform this adjustment until the surgical microscope moves
neither upward nor downward. The surgical microscope is fully balanced
when it remains stationary in all positions within the work area.
Rotating clockwise increases the spring force. turning the screw counter-
clockwise reduces the spring force.

Limiting the suspension arm's downward movement

CAUTION Risk of injury caused by lowering of the surgical microscope!


• Limit the stroke of the suspension arm to prevent any contact with the
patient should the surgical microscope be lowered accidentally.
• Loosen locking lever (2) for the upward/downward movement.

Version 11.0
Page 160 G-30-1435-en
S7 / OPMI PROergo Preparations for use

• Lower the surgical microscope until it can be focused on the surgical field
(according to the focal length of the lens). At the same time, ensure that
there is a sufficiently safe distance between the microscope and the
surgical field.
• Firmly retighten locking lever (2) for the upward/downward movement.
• Lower the surgical microscope to its lowest position again, and check the
distance between the microscope and the surgical field.
– The limit of downward travel is only effective in the lower part of the
suspension arm's vertical movement range (starting from the arm's
horizontal position).
– If you wish to use the suspension arm's full range of vertical movement,
lower the suspension arm as far as it will go and then firmly retighten
locking lever (2).

1 2 3

Version 11.0
G-30-1435-en Page 161
Preparations for use S7 / OPMI PROergo

Adjusting the system - Ceiling mount, wall mount and S7


Centro suspension system

Setting the suspension arm weight compensation

CAUTION Risk of crushing - mind your fingers!


Fingers may be crushed in the areas marked with the "Risk of crushing" label.
• Do not touch these areas while the system is being moved or brought in
its working / transport position.
Do not perform the balance setting until the surgical microscope and all
accessories have been mounted!
• Move the suspension arm into its horizontal position and tightly hold it
with one hand.
• Loosen knob (3) for adjusting the friction of the suspension arm's upward/
downward movement.
• Move the suspension arm slightly up and down. At the same time, turn
adjustment screw (1) until you think that the spring force is sufficient to
compensate for the weight of the surgical microscope and accessories.
Continue to perform this adjustment until the surgical microscope moves
neither upward nor downward. The surgical microscope is fully balanced
when it remains stationary in all positions within the work area.
Rotating clockwise increases the spring force. turning the screw counter-
clockwise reduces the spring force.

Limiting the suspension arm's downward movement

CAUTION Risk of injury caused by lowering of the surgical microscope!


• Limit the stroke of the suspension arm to prevent any contact with the
patient should the surgical microscope be lowered accidentally.
• Loosen locking lever (2) for the upward/downward movement.

Version 11.0
Page 162 G-30-1435-en
S7 / OPMI PROergo Preparations for use

• Lower the surgical microscope until it can be focused on the surgical field
(according to the focal length of the lens). At the same time, ensure that
there is a sufficiently safe distance between the microscope and the
surgical field.
• Firmly retighten locking lever (2) for the upward/downward movement.
• Lower the surgical microscope to its lowest position again, and check the
distance between the microscope and the surgical field.
– The limit of downward travel is only effective in the lower part of the
suspension arm's vertical movement range (starting from the arm's
horizontal position).
– If you wish to use the suspension arm's full range of vertical movement,
lower the suspension arm as far as it will go and then firmly retighten
locking lever (2).

1 2 3

Version 11.0
G-30-1435-en Page 163
Preparations for use S7 / OPMI PROergo

Adjusting the system - S7 ceiling mount with lifting column

Set ergonomic working height

CAUTION Risk of injury caused by lifting column


Activating the lifting column may cause injuries.
• Do not activate the lifting column during surgery.
• Do not use the lifting column for focusing.
• Make sure that the patient is not put at risk by the motorized adjustment
of the lifting column.

Bring the surgical microscope into a position convenient for you before each
surgical procedure, and set the optimum ergonomic working height of the
suspension system via the lifting column (without patient!).
As long as you keep selector switch (2) in the turned position, lifting column
(1) moves upward or downward, depending on the switch position. When
you release the selector switch, the lifting column stops immediately.
After being in operation for a maximum of one minute, the lifting column
must be turned off for at least nine minutes to prevent the motor from
overheating.
• Before raising or lowering the stand or mount, make sure that there is
sufficient clearance so as to avoid any collision with other objects.
The lifting column is used to move the microscope into position for surgery
prior to the surgical procedure.

Do not constantly move the lifting column up and down, because this will
trigger the thermal cut-out, which automatically deactivates the drive motor.
If this occurs, the lifting column cannot be moved until the motor has cooled
down.

NOTE Risk of injury to the head when using ceiling mounts!


Your head may collide with the device while the device is lowered.
• Move ceiling mount to its standby position when it is not in use.

Version 11.0
Page 164 G-30-1435-en
S7 / OPMI PROergo Preparations for use

Version 11.0
G-30-1435-en Page 165
Preparations for use S7 / OPMI PROergo

Setting the suspension arm weight compensation

CAUTION Risk of crushing - mind your fingers!


Fingers may be crushed between the carrier and suspension arms.
• Never touch the area between the suspension and carrier arms while you
are performing the balancing procedure or moving the suspension arm.
Do not perform the balance setting until the surgical microscope and all
accessories have been mounted!
• Move the suspension arm into its horizontal position and tightly hold it
with one hand.
• Loosen knob (3) for adjusting the friction of the suspension arm's upward/
downward movement.
• Move the suspension arm slightly up and down. At the same time, turn
adjustment screw (1) until you think that the spring force is sufficient to
compensate for the weight of the surgical microscope and accessories.
Continue to perform this adjustment until the surgical microscope moves
neither upward nor downward. The surgical microscope is fully balanced
when it remains stationary in all positions within the work area.
Rotating clockwise increases the spring force. turning the screw counter-
clockwise reduces the spring force.

Limiting the suspension arm's downward movement

CAUTION Risk of injury caused by lowering of the surgical microscope!


• Limit the stroke of the suspension arm to prevent any contact with the
patient should the surgical microscope be lowered accidentally.
• Loosen locking lever (2) for the upward/downward movement.
• Lower the surgical microscope until it can be focused on the surgical field
(according to the focal length of the lens). At the same time, ensure that
there is a sufficiently safe distance between the microscope and the
surgical field.
• Firmly retighten locking lever (2) for the upward/downward movement.

Version 11.0
Page 166 G-30-1435-en
S7 / OPMI PROergo Preparations for use

• Lower the surgical microscope to its lowest position again, and check the
distance between the microscope and the surgical field.
– The limit of downward travel is only effective in the lower part of the
suspension arm's vertical movement range (starting from the arm's
horizontal position).
– If you wish to use the suspension arm's full range of vertical movement,
lower the suspension arm as far as it will go and then firmly retighten
locking lever (2).

1 2 3

Version 11.0
G-30-1435-en Page 167
Preparations for use S7 / OPMI PROergo

Balancing the surgical microscope

NOTE Risk of injury to the patient!


With an incorrectly balanced system, brake release may lead to uncontrolled
movements of the suspension arm.
• Use the surgical microscope in perfectly balanced condition only!
• Balance the surgical microscope before surgery, without the patient.
• The subsequent test must not be carried out when the system is
positioned over the patient and has to be done at a safe distance from
other persons and devices.
• To check correct balancing of the system, unlock the brakes. For this step,
hold the surgical microscope by both handgrips. If the system has been
correctly balanced, the surgical microscope can be moved almost
effortlessly.
To permit almost effortless guidance of the surgical microscope, the
microscope and the suspension system used must be correctly balanced.
If the system is in an extremely unbalanced state, the unit can move
uncontrollably out of position. For this reason, hold the surgical microscope
tightly at its handgrips before loosening the friction adjustment knob on the
suspension arm.

Adjusting the friction of the microscope's axis of rotation


• Use knob (1) to adjust the friction of the microscope's axis of rotation.

Lateral tilt motion, balancing


• Loosen friction adjustment knob (3).
• Use knob (2) to adjust the spring force in such a way that the surgical
microscope is balanced about its lateral tilt axis and remains stationary in
the viewing direction required.
– If the surgical microscope tends to move clockwise, turn knob (2)
counterclockwise.
– If the surgical microscope tends to move counterclockwise, turn knob
(2) clockwise.

Version 11.0
Page 168 G-30-1435-en
S7 / OPMI PROergo Preparations for use

Adjusting the friction of the lateral tilt axis


• Use knob (3) to adjust the friction for the lateral tilt axis as required.

Version 11.0
G-30-1435-en Page 169
Preparations for use S7 / OPMI PROergo

Balancing the front-to-back tilt motion

NOTE Unintentional movement of the system


With an incorrectly balanced system uncontrolled movements of the
suspension system may occur.
• If uncontrolled movements occur, move the surgical microscope,
by overcoming the braking effect, into a position in which the patient is
not put at risk.
For initial balancing or after a change of the accessories, proceed as follows:
• Deactivate the magnetic brakes by activating the Off button in the
Magnetic Brakes menu and then pressing the Enter button. The magnetic
brakes are now off.
• Press the release button (5) to open the magnetic brake.
• Loosen friction knob (4).
Zero position
• Press the button (1) until the motor of the balancing system stops.
This decouples the microscope body from its mount.
Check zero position: The triangle and the line must coincide. If they don't,
readjust manually.
The microscope has been decoupled from its suspension and has
adopted its center-of-gravity position (zero position) if the white ring
with marking (3) does not move when you turn the microscope.
• Tap the key (2).
Allow the microscope to hang in its center-of-gravity position while you
press the button (2).
After approx. 15 seconds, the microscope body has again been coupled on
its suspension. You can check this: when you move the microscope body,
the white ring (3) also moves.
• Continue pressing button (2) and check whether the microscope body
remains balanced in every tilt position. Depending on the accessories
attached, this can take some time (approx. 1 minute or longer). To check
this, slightly lift the microscope body and tilt it into different positions to
see whether the microscope body has been correctly balanced.
• If you have pressed button (2) for too long, you can correct the balance
by pressing button (1).

Version 11.0
Page 170 G-30-1435-en
S7 / OPMI PROergo Preparations for use

2
1

4
5

Version 11.0
G-30-1435-en Page 171
Preparations for use S7 / OPMI PROergo

Correcting the balance of the front-to-back tilt motion


After a slight change in the position of accessories (e.g. the angle of the
coobservation tube) it may become necessary to correct the balance setting.
– If the microscope body tends to tilt forward (in your direction), press
button (2).
– If the microscope body tends to tilt backward, press button (1).

Adjusting the friction of the front-to-back tilt axis


Use knob (4) to adjust the friction of the front-to-back tilt axis as required.

2
1

Version 11.0
Page 172 G-30-1435-en
S7 / OPMI PROergo Preparations for use

Adjusting the surgical microscope

Notes on adjustment
– To permit fast setup of the microscope, it is advisable to store the pupillary
distances and prescriptions of the different users so that they can be easily
preset during system preparation.
– Do not wear any bifocal or progressive lenses for the adjustment
procedure and subsequent work with the microscope.
Lenses of this kind make correct diopter setting impossible and result in
unsatisfactory imaging.
– Optimum image clarity in the eyepieces and at the optical port for
documentation equipment (photo, video) can only be achieved through
careful focusing and precisely adjusting the binocular tube.

Adjusting the binocular tube

Adjusting the • Position the microscope vertically above a flat object, e.g. a piece of paper
pupillary distance with writing on.
• Bring the eyepieces on the binocular tube to eye distance, so that the two
eyepiece images (the object and the periphery of the field of view) merge
to produce a single image.

Adjusting the eyepieces The following procedure must be performed separately for each eyepiece in
the order described.
• Set to the smallest magnification on the microscope.
• Set the diopter adjustment ring at the eyepiece to zero position.
• Look through the eyepiece and move the microscope to focus the image.
• Set the microscope to the highest magnification and correct the fine
focusing until the image is defined sharply.
• Now set to the smallest magnification once again without changing the
working distance.
• Set the diopter adjustment ring at the eyepiece to its maximum value (plus
direction).
• Look through the eyepiece and slowly turn the diopter adjustment ring to
the minus direction until the image is sharply defined.

Version 11.0
G-30-1435-en Page 173
Preparations for use S7 / OPMI PROergo

• Repeat the entire procedure for the second eyepiece.


 The microscope is now adjusted so that a constantly sharp image is
produced throughout the entire magnification range without your having
to refocus after changing the magnification value.
If this is not the case, repeat the procedure.

Adjusting the eyecups • Adjust the eyecups so that you can survey the entire field of view.
– Viewing with eyeglasses: screw in the eyecups.
– Viewing without eyeglasses: remove the eyecups.

Adjusting the working distance and magnification


• Position the microscope above the surgical field so as to ensure
convenient work.
• Focus roughly on the object by changing the distance between
microscope and object.
• Set the microscope to the highest magnification and correct the fine
focusing until the image is defined sharply.
• Set the microscope to the required magnification. The focal plane is
retained.

CAUTION Risk of injury to the patient caused by parts falling down!


Due to collisions or incorrect installation, parts may fall down and injure the
patient.
• Before using the device, check that all parts are properly and securely
mounted.
• Check that the floor stand is secured in a steady position and can be
pivoted without causing collisions.

Version 11.0
Page 174 G-30-1435-en
S7 / OPMI PROergo Preparations for use

Preparing the device for sterile use

CAUTION Risk of infection due to soiled aseptic accessories and drapes!


• Only use sterile asepsis accessories and drapes suitable for the system!
• The products contained in the asepsis sets must be cleaned, disinfected
and sterilized before each use. This also applies to the first use after
delivery.
• Make sure that the operating personnel have been instructed in the
sterile operation of the device.

Attaching asepsis caps

The device can be equipped with resterilizable products for sterile use.
The asepsis sets available from ZEISS contain caps and handgrips which can
be sterilized in autoclaves. For more detailed information about sterilization,
please refer to the enclosed "Preparation of resterilizable products" manual
for each Asepsis Set.

Attaching the drapes

You can also use disposable sterile drapes to cover the system. When draping
the system, make sure there is enough slack in the drapes to allow for
movement of the microscope carrier and surgical microscope.

Version 11.0
G-30-1435-en Page 175
Preparations for use S7 / OPMI PROergo

Relocating the S7 floor stand

CAUTION Risk of crushing - mind your fingers!


Fingers may be crushed in the areas marked with the "Risk of crushing" label.
• Do not touch these areas while the system is being moved or brought in
its working / transport position.

NOTE Mechanical force effect when the device is being transported!


If the following precautions are not observed, the device may topple over
and injure persons:
• Use the grab handle for transporting the device.
• Mind the maximum vertical clearance when passing through doorways.
• Avoid any jolting.
• Be extremely careful when moving over slopes.
• Do not park the device on sloped surfaces.
Please observe the following points when relocating the stand:
• Turn off the device at the power switch.
1 • Remove the power plug from the power outlet.
2 • Move the suspension arm into its transport position (see illustration on
the right).
3
• Wind up the foot control panel cable on one of the cable holders (2),
and hang up the foot control panel on the holder (3).
• Wind up the power cord on the other cable holder (2).
• Release all the locking tabs (4) or the casters.

4 • Hold the device by the grab handle (1) and carefully push it into its new
position. Ensure the device is standing on even ground.
• Press at least three of the locking tabs (4) and make sure that the stand is
no longer able to move away by itself.
Because the stand moves so smoothly, it is easy to underestimate how heavy
it is. Therefore, move the stand slowly and carefully!

Version 11.0
Page 176 G-30-1435-en
S7 / OPMI PROergo Preparations for use

NOTE Transport over long distances!


The system may be transported over long distances (e.g. during relocation,
return for repair, etc.) only in its original packaging or in special return
packaging.
• Please contact your dealer or ZEISS Service for this purpose.

Version 11.0
G-30-1435-en Page 177
Preparations for use S7 / OPMI PROergo

Version 11.0
Page 178 G-30-1435-en
S7 / OPMI PROergo Operation

Operation

Checklist for S7 floor stand ..........................................................180

Checklist for wall mount, ceiling mount and S7 Centro


suspension system ........................................................................ 184

Checklist for S7 ceiling mount with lifting column ....................... 189

Switching the device on and off ...................................................194

Operating procedure ....................................................................194


Working with the SpeedFokus (autofocus) option ...................................196

Version 11.0
G-30-1435-en Page 179
Operation S7 / OPMI PROergo

Checklist for S7 floor stand

WARNING Precautions in case of device failure


• We recommend you to take adequate precautions, depending on the
application, to enable the surgical procedure or treatment to be finished
without using this microscope, if required.

WARNING Function deterioration of the light source


With increasing service life of the light source, the actual light intensity
decreases (this is a normal system property).
• Before every use, ensure that the available light quantity is sufficient.
• If necessary, change the lamp.
After changing the lamp, check the brightness setting.
• Keep a replacement lamp (e.g. OR lamp) ready at hand.

CAUTION Risk of injury!


Incorrect configuration settings may lead to unexpected behavior of the
device and may cause injury to the patient.
• Always check the settings of the selected user profile before using the
system.

CAUTION Function failure recognized too late!


• To avoid the occurrence of any defects during surgery, perform
a function test before every surgical procedure (without patient)!
For safety reasons, the device must not be used if a malfunction is noticed
before the start of a procedure.
• Correct the malfunction (see "What to do in the event of malfunctions")
or contact ZEISS Service.
Always check the following points before surgery (without the patient!):
• Make sure that there is sufficient clearance between the instrument and
the accessories installed and that the protective cover (1) is removed from
the objective lens.
• Turn on the device at the power switch on the stand or mount.
XENON illumination
 Check that the xenon illumination is intact.

Version 11.0
Page 180 G-30-1435-en
S7 / OPMI PROergo Operation

 Check that the xenon illumination is on and the green indicator lamp is lit.
If the first lamp has failed and the backup lamp is in use (segment in the
switching knob lights up), make sure to have a backup lamp module ready
to hand as a precaution.

CAUTION The xenon lamp has a limited service life of 500 h!


If used beyond its maximum service life, the xenon lamp may explode.
• Please replace the xenon lamp in due time.
• Reset the service hour counter to "0" after replacing the lamp.
• For the lamp change procedure and how to reset the service hour
counter, see Page 112.
Please replace the xenon lamp in due time.
Please note the following guidelines – example:
– If operated for 4 hours/day on 5 days/week, the lamp needs to be
exchanged after 25 weeks maximum.
– If operated for 8 hours/day on 5 days/week, the lamp needs to be
exchanged after 12 weeks maximum.
Checking the illumination system
 Check that the lamp and the backup lamp are intact.
Checking the function of the handgrips
 Check that the handgrips are located in a position convenient for you,
or one or both handgrips have been removed and replaced by the knobs
supplied.
Check the zoom
• Press the appropriate release button(s) on the handgrips.
Check the focus
• Press the appropriate release button(s) on the handgrips.

Version 11.0
G-30-1435-en Page 181
Operation S7 / OPMI PROergo

Checking the functions of the magnetic brakes


• Press release button (1).
The magnetic brakes are active if Brakes are set to "Off" in the Surgery/
Options/Magnetic Brakes menu.
Checking the functions of the knobs
• Check the proper functioning of focusing knob (4) and filter selector
knob (5).
Eyepieces / binocular tube
 Check that binocular tube (3) has been securely mounted.
 Check that securing screw (2) has been firmly tightened.
 Check that the surgical microscope and the tube are in a position
convenient for you.
 Check that the correct interocular distance has been set.
 Check that the eyecups have been adjusted in such a way that you can see
the full field of view.
 Check that the correct prescription has been set on the diopter scale.
 Check that image quality is correct throughout the entire magnification
range.
Balance
 Check that the surgical microscope has been correctly balanced.

1 2 3 4 5

Version 11.0
Page 182 G-30-1435-en
S7 / OPMI PROergo Operation

Balancing
 The suspension arm is properly balanced. If the suspension arm is properly
balanced, the surgical microscope should remain immobile in all positions
of the working range and should not swivel up or down.
Stroke limit
 Check that the minimum working distance (height) from the surgical field
has been adjusted.
Stand base
 At least two of the immobilizers have been pressed down to prevent the
stand from rolling away by itself.
Accessories
 Proper functioning of all other equipment (surgical microscope,
coobservation equipment, video system, etc.) has been checked using the
relevant user's manuals.

Optional foot control panel


 Check that the plug of the foot control panel has been connected.
 Check that the power switch of suspension system has been switched on.
• Check that all functions assigned to the respective buttons on the foot
control panel are working properly.

SpeedFokus (Autofocus) option


Settings and matching when using an external video camera
 Check that the video camera position, the focal length of the video optics
and the size of the CCD array have been correctly entered in the Autofocus
menu.
 Check that the video system and the surgical microscope have been
matched to each other, i.e. when the image is in focus in the microscope
eyepieces, the video image on the monitor must also be in focus (also
applies to the integrated video camera).
 Check that the image on the monitor has the same orientation as the
image in the main eyepieces.
Focusing process
Check that focusing can be activated according to the programming of the
buttons:
 Using buttons A, B, C and/or D of the handgrips.
 Using button C and/or D of the foot control panel, if a foot control panel
with 8 or 14 functions (option) is provided.

Version 11.0
G-30-1435-en Page 183
Operation S7 / OPMI PROergo

Checklist for wall mount, ceiling mount and S7 Centro


suspension system

WARNING Precautions in case of device failure


• We recommend you to take adequate precautions, depending on the
application, to enable the surgical procedure or treatment to be finished
without using this microscope, if required.

WARNING Function deterioration of the light source


With increasing service life of the light source, the actual light intensity
decreases (this is a normal system property).
• Before every use, ensure that the available light quantity is sufficient.
• If necessary, change the lamp.
After changing the lamp, check the brightness setting.
• Keep a replacement lamp (e.g. OR lamp) ready at hand.

CAUTION Risk of injury!


Incorrect configuration settings may lead to unexpected behavior of the
device and may cause injury to the patient.
• Always check the settings of the selected user profile before using the
system.

CAUTION Function failure recognized too late!


• To avoid the occurrence of any defects during surgery, perform
a function test before every surgical procedure (without patient)!
For safety reasons, the device must not be used if a malfunction is noticed
before the start of a procedure.
• Correct the malfunction (see "What to do in the event of malfunctions")
or contact ZEISS Service.
Always check the following points before surgery (without the patient!):
• Make sure that there is sufficient clearance between the instrument and
the accessories installed and that the protective cover (1) is removed from
the objective lens.
• Turn on the device at the power switch on the stand or mount.
XENON illumination
 Check that the xenon illumination is intact.

Version 11.0
Page 184 G-30-1435-en
S7 / OPMI PROergo Operation

 Check that the xenon illumination is on and the green indicator lamp is lit.
If the first lamp has failed and the backup lamp is in use (segment in the
switching knob lights up), make sure to have a backup lamp module ready
to hand as a precaution.

CAUTION The xenon lamp has a limited service life of 500 h!


If used beyond its maximum service life, the xenon lamp may explode.
• Please replace the xenon lamp in due time.
• Reset the service hour counter to "0" after replacing the lamp.
• For the lamp change procedure and how to reset the service hour
counter, see Page 112.
Please replace the xenon lamp in due time.
Please note the following guidelines – example:
– If operated for 4 hours/day on 5 days/week, the lamp needs to be
exchanged after 25 weeks maximum.
– If operated for 8 hours/day on 5 days/week, the lamp needs to be
exchanged after 12 weeks maximum.
Checking the illumination system
 Check that the lamp and the backup lamp are intact.

Version 11.0
G-30-1435-en Page 185
Operation S7 / OPMI PROergo

Checking the function of the handgrips


 Check that the handgrips are located in a position convenient for you,
or one or both handgrips have been removed and replaced by the knobs
supplied.
Check the zoom
• Press the appropriate release button(s) on the handgrips.
Check the focus
• Press the appropriate release button(s) on the handgrips.

Version 11.0
Page 186 G-30-1435-en
S7 / OPMI PROergo Operation

Checking the functions of the magnetic brakes


• Press release button (1).
The magnetic brakes are active if Brakes are set to "Off" in the Surgery/
Options/Magnetic Brakes menu.
Checking the functions of the knobs
• Check the proper functioning of focusing knob (4) and filter selector
knob (5).
Eyepieces / binocular tube
 Check that binocular tube (3) has been securely mounted.
 Check that securing screw (2) has been firmly tightened.
 Check that the surgical microscope and the tube are in a position
convenient for you.
 Check that the correct interocular distance has been set.
 Check that the eyecups have been adjusted in such a way that you can see
the full field of view.
 Check that the correct prescription has been set on the diopter scale.
 Check that image quality is correct throughout the entire magnification
range.
Balance
 Check that the surgical microscope has been correctly balanced.

1 2 3 4 5

Version 11.0
G-30-1435-en Page 187
Operation S7 / OPMI PROergo

Balancing
 The suspension arm is properly balanced. If the suspension arm is properly
balanced, the surgical microscope should remain immobile in all positions
within the working range and should not swivel up or down.
Stroke limit
 The minimum working distance (height) from the surgical field has been
adjusted.

Optional foot control panel


 Check that the plug of the foot control panel has been connected.
 Check that the power switch of suspension system has been switched on.
• Check that all functions assigned to the respective buttons on the foot
control panel are working properly.

SpeedFokus (Autofocus) option


Settings and matching when using an external video camera
 Check that the video camera position, the focal length of the video optics
and the size of the CCD array have been correctly entered in the Autofocus
menu.
 Check that the video system and the surgical microscope have been
matched to each other, i.e. when the image is in focus in the microscope
eyepieces, the video image on the monitor must also be in focus (also
applies to the integrated video camera).
 Check that the image on the monitor has the same orientation as the
image in the main eyepieces.
Focusing process
Check that focusing can be activated according to the programming of the
buttons:
 using buttons A, B, C and/or D of the handgrips.
 using button C and/or D of the foot control panel, if a foot control panel
with 8 or 14 functions (option) is provided.

Version 11.0
Page 188 G-30-1435-en
S7 / OPMI PROergo Operation

Checklist for S7 ceiling mount with lifting column

WARNING Precautions in case of device failure


• We recommend you to take adequate precautions, depending on the
application, to enable the surgical procedure or treatment to be finished
without using this microscope, if required.

WARNING Function deterioration of the light source


With increasing service life of the light source, the actual light intensity
decreases (this is a normal system property).
• Before every use, ensure that the available light quantity is sufficient.
• If necessary, change the lamp.
After changing the lamp, check the brightness setting.
• Keep a replacement lamp (e.g. OR lamp) ready at hand.

CAUTION Risk of injury!


Incorrect configuration settings may lead to unexpected behavior of the
device and may cause injury to the patient.
• Always check the settings of the selected user profile before using the
system.

CAUTION Function failure recognized too late!


• To avoid the occurrence of any defects during surgery, perform
a function test before every surgical procedure (without patient)!
For safety reasons, the device must not be used if a malfunction is noticed
before the start of a procedure.
• Correct the malfunction (see "What to do in the event of malfunctions")
or contact ZEISS Service.
Always check the following points before surgery (without the patient!):
• Make sure that there is sufficient clearance between the instrument and
the accessories installed and that the protective cover (1) is removed from
the objective lens.
• Turn on the device at the power switch on the stand or mount.
XENON illumination
 Check that the xenon illumination is intact.

Version 11.0
G-30-1435-en Page 189
Operation S7 / OPMI PROergo

 Check that the xenon illumination is on and the green indicator lamp is lit.
If the first lamp has failed and the backup lamp is in use (segment in the
switching knob lights up), make sure to have a backup lamp module ready
to hand as a precaution.

CAUTION The xenon lamp has a limited service life of 500 h!


If used beyond its maximum service life, the xenon lamp may explode.
• Please replace the xenon lamp in due time.
• Reset the service hour counter to "0" after replacing the lamp.
• For the lamp change procedure and how to reset the service hour
counter, see Page 112.
Please replace the xenon lamp in due time.
Please note the following guidelines – example:
– If operated for 4 hours/day on 5 days/week, the lamp needs to be
exchanged after 25 weeks maximum.
– If operated for 8 hours/day on 5 days/week, the lamp needs to be
exchanged after 12 weeks maximum.
Checking the illumination system
 Check that the lamp and the backup lamp are intact.

Version 11.0
Page 190 G-30-1435-en
S7 / OPMI PROergo Operation

Checking the function of the handgrips


 Check that the handgrips are located in a position convenient for you,
or one or both handgrips have been removed and replaced by the knobs
supplied.
Check the zoom
• Press the appropriate release button(s) on the handgrips.
Check the focus
• Press the appropriate release button(s) on the handgrips.

Version 11.0
G-30-1435-en Page 191
Operation S7 / OPMI PROergo

Checking the functions of the magnetic brakes


• Press release button (1).
The magnetic brakes are active if Brakes are set to "Off" in the Surgery/
Options/Magnetic Brakes menu.
Checking the functions of the knobs
• Check the proper functioning of focusing knob (4) and filter selector
knob (5).
Eyepieces / binocular tube
 Check that binocular tube (3) has been securely mounted.
 Check that securing screw (2) has been firmly tightened.
 Check that the surgical microscope and the tube are in a position
convenient for you.
 Check that the correct interocular distance has been set.
 Check that the eyecups have been adjusted in such a way that you can see
the full field of view.
 Check that the correct prescription has been set on the diopter scale.
 Check that image quality is correct throughout the entire magnification
range.
Balance
 Check that the surgical microscope has been correctly balanced.

1 2 3 4 5

Version 11.0
Page 192 G-30-1435-en
S7 / OPMI PROergo Operation

Checking the lifting column


• Check the function of the lifting column by operating the selector switch.
Balancing
 The suspension arm is properly balanced. If the suspension arm is properly
balanced, the surgical microscope should remain immobile in all positions
of the working range and should not swivel up or down.
Stroke limit
 The minimum working distance (height) from the surgical field has been
adjusted.
Checking accessories
• Using their respective Instructions for Use, check all other pieces of the
device equipment (surgical microscope, co-observation equipment, video
system, etc.) to make sure they are functioning properly.

Optional foot control panel


 Check that the plug of the foot control panel has been connected.
 Check that the power switch of suspension system has been switched on.
• Check that all functions assigned to the respective buttons on the foot
control panel are working properly.

SpeedFokus (Autofocus) option


Settings and matching when using an external video camera
 Check that the video camera position, the focal length of the video optics
and the size of the CCD array have been correctly entered in the Autofocus
menu.
 Check that the video system and the surgical microscope have been
matched to each other, i.e. when the image is in focus in the microscope
eyepieces, the video image on the monitor must also be in focus (also
applies to the integrated video camera).
 Check that the image on the monitor has the same orientation as the
image in the main eyepieces.
Focusing process
Check that focusing can be activated according to the programming of the
buttons:
 using buttons A, B, C and/or D of the handgrips.
 using button C and/or D of the foot control panel, if a foot control panel
with 8 or 14 functions (option) is provided.

Version 11.0
G-30-1435-en Page 193
Operation S7 / OPMI PROergo

Switching the device on and off

 The included power cord (floor stand only) is connected.


• Observe the checklists starting on Page 180.
This is how to switch the device on or off:
• To switch it on, press the power switch (1) on the stand. When the device
is switched on, the indicator light in the power switch is illuminated.
1
• To deactivate it, press the power switch (1) on the stand.
The current for the power outlet socket cannot be switched off with the
power switch.

CAUTION Never leave the device unattended when it is switched on!


If a patient or an object is exposed to the light source for a longer period of
time, this could cause burns or fire damage.
• Never leave a device unattended when its light source is switched on.

Operating procedure

CAUTION Risk of injury caused by light in the facial area


Surgery using light in the facial area may seriously harm the patient's eyes.
• The patient's eyes must not be exposed to light.
• This light source must not be used in ophthalmology.

Version 11.0
Page 194 G-30-1435-en
S7 / OPMI PROergo Operation

CAUTION Risk of burns caused by light!


Changing the magnification on the foldable tube also changes the light
intensity. Light can cause burn injuries.
• After adjusting the magnification always check the light intensity and the
illuminated-field diameter.
• Reduce the light intensity to the absolute minimum required.

CAUTION Risk of injury caused by light action!


Light action can cause injuries in the peripheral area. The risk is increased if
dry drapes are used to cover such areas.
• To avoid this, use wet, sterile gauze to cover the relevant areas.
• Turn on the device at the power switch on the stand or mount.
• Set the required brightness of the surgical field illumination on the
suspension system.
• Check the system using the checklist.
• Swing the surgical microscope over the surgical field into an ergonomic
position within the working distance.
• Select the lowest magnification (zoom).
• Look through the eyepieces and lower the surgical microscope until the
surgical field comes into focus. This will allow you to achieve a rough
focus.
• Select the highest magnification (zoom).
• Look through the eyepieces and activate the focusing function until the
image of the surgical field appears sharply focused.
• Select the magnification required (zoom). Look through the eyepieces of
the binocular tube. Adjust the eyepieces so that both the edge of the field
of view and the microscope image are sharply focused.
• Bring the handgrips into a position convenient for the procedure planned
and tighten locking screws (1).
1 • Switch off the system when not in use.

CAUTION Live video images and recorded video sequences, video clips (cut sequences)
and single images shall not be used for diagnosis purposes. Visualized images
may contain variations in shape, contrast and color.

Version 11.0
G-30-1435-en Page 195
Operation S7 / OPMI PROergo

Working with the SpeedFokus (autofocus) option

The SpeedFokus autofocus option permits you to select an area of special


interest to you in the surgical field. SpeedFokus focuses on this selected area
at the press of a button. Focusing is triggered on the buttons of the handgrip
or foot control panel as configured in the Options menu.
• Look through the eyepieces and activate the focusing function until the
image of the surgical field appears sharply focused. The image on the
video monitor must also be sharply focused.
• Watch the image on the S-video monitor (which is connected to the
S-video connector) and position the measuring field (ROI), i.e. the square
frame, in the area in which you are particularly interested.
• Activate focusing by the SpeedFokus autofocus function.
If the ROI in the surgical field is marked by major height differences with
steeply sloping edges, it may happen that SpeedFokus focuses on the wrong
feature in the surgical field. You can correct this as follows:
• Slightly shift the ROI and/or reduce its size.
• Re-activate focusing by the SpeedFokus autofocus function.

HD SDI or DVI HD video monitor


on HD SDI or DVI
connector
CCU of the 1Chip
HD camera option

SD-video monitor
S-Video SpeedFokus
on S-video
(optional)
connector

Version 11.0
Page 196 G-30-1435-en
S7 / OPMI PROergo What to do in the event of malfunctions

What to do in the event of malfunctions

Failure of main functions ..............................................................199


Failure of magnetic brakes ......................................................................199
Failure of the SpeedFokus (Autofocus) option .........................................199
Failure of the motorized focusing function ..............................................200
Failure of the halogen lamps ...................................................................202
Lamp failure of the xenon illumination ....................................................204

Causes of malfunctions and remedies ........................................... 207

Version 11.0
G-30-1435-en Page 197
What to do in the event of malfunctions S7 / OPMI PROergo

Version 11.0
Page 198 G-30-1435-en
S7 / OPMI PROergo What to do in the event of malfunctions

Failure of main functions

WARNING Risk of injury caused by failure of main functions!


If one of the main functions (focusing, zoom) is unavailable or is not
functioning correctly, this may result in serious injury to the patient.
• Deactivate the electronic components and proceed in manual mode.

Failure of magnetic brakes

In the event of magnetic brake failure (magnetic brakes are closed), you can
manually position the microscope by overriding the locking effect of the
brakes.

Failure of the SpeedFokus (Autofocus) option

Failure of the SpeedFokus (autofocus) option while the rest of the OPMI
PROergo system is working correctly on the S7 system:
You can continue using your OPMI PROergo / S7 system, but you can no
longer benefit from the SpeedFokus autofocus option until it has been
repaired by the service staff.
• Continue surgery using the motorized focusing system of the OPMI
PROergo.
• Focus the image manually on the OPMI PROergo.

Version 11.0
G-30-1435-en Page 199
What to do in the event of malfunctions S7 / OPMI PROergo

Failure of the motorized focusing function

Motor running constantly


• Bring the suspension arm into its horizontal position and tighten locking
knob (4) firmly.
• Loosen and remove knurled nut (1).
• Fold up cover (2) and pull out cable connection (3) for the surgical
microscope (large connector) at the side. Carefully lower the cover and
allow it to rest lightly on the cable.
• Unplug cable connection (3) for the surgical microscope (large
connectors).
• Manually focus by turning focusing knob (5)
or
• focus by moving the suspension arm up or down.

Motor not working


• Manually focus by turning focusing knob (5)
or
• focus by moving the suspension arm up or down.

Version 11.0
Page 200 G-30-1435-en
S7 / OPMI PROergo What to do in the event of malfunctions

1
2

3
4

Version 11.0
G-30-1435-en Page 201
What to do in the event of malfunctions S7 / OPMI PROergo

Failure of the halogen lamps

NOTE Overheating of the lamp module!


If ventilation grids are covered e.g. by drapes, this may lead to overheating
of the lamp housing and deactivation of the lamp.
• Never cover the ventilation grids.
• Lamps switched off due to overheating will be automatically switched on
again when they have cooled down.
The lamp housing contains a backup lamp which automatically swings into
position when the first lamp fails. The open flap (2) indicates that the backup
lamp is in operation.

Switching manually to the backup lamp


Press button (3) to manually activate the backup lamp.

If the backup lamp fails:

NOTE Risk of burns caused by hot lamp!


If you replace the lamp shortly after it has failed, the lamp will still be
very hot.
• Allow the lamp to cool down before changing it.
• Use heat-resistant protective gloves when replacing the lamp.
• Turn on the stand at its power switch.
• Press button (4) to eject lamp module (1) partially. Remove the lamp
module and replace the lamp or insert a backup lamp module.
• Switch the suspension system back on. Adjust the brightness of the
illumination system as required.

Version 11.0
Page 202 G-30-1435-en
S7 / OPMI PROergo What to do in the event of malfunctions

Version 11.0
G-30-1435-en Page 203
What to do in the event of malfunctions S7 / OPMI PROergo

Lamp failure of the xenon illumination

CAUTION Risk of injury caused by lamp rupture!


Lamp rupture (audible as a loud bang) may lead to jamming of the lamp
module and/or failure of the electronics modules.
• Before opening the lamp housing, ensure that the device is moved to
a position in which possibly falling particles cannot put the patient or user
at risk.
• Do not continue using the system if the lamp module is jammed or the
illumination is no longer operational due to defective electronic modules.
Contact our Service Department.

NOTE Overheating of the lamp module!


If ventilation grids are covered e.g. by drapes, this may lead to overheating
of the lamp housing and deactivation of the lamp.
• Never cover the ventilation grids.
• Lamps switched off due to overheating will be automatically switched on
again when they have cooled down.

NOTE Risk of burns from hot lamp!


If you replace the lamp shortly after it has failed, the lamp will still be very
hot.
• Allow the lamp to cool down before removing it.
• Always use heat-resistant protective gloves when replacing the lamp!

Switching to the backup lamp


• Turn off the suspension system at the power switch before switching to
the backup lamp.
The lamp module contains two xenon lamps. The second lamp serves as
a backup lamp which can be swung into position if the first lamp fails.
If the first xenon lamp fails, open the lamp module (2) as follows:
• Press the button ( 3). The lamp module will eject partially.
• Pull out the lamp module as far as it will go.

Version 11.0
Page 204 G-30-1435-en
S7 / OPMI PROergo What to do in the event of malfunctions

• Turn knob (1) through 180° until it snaps in. This swings the second xenon
lamp (backup lamp) into the beam path.
• Push the lamp module all the way back into the lamp housing.

Version 11.0
G-30-1435-en Page 205
What to do in the event of malfunctions S7 / OPMI PROergo

• Reset the operating hours counter to "0". Use a pointed object and press
it into the recess of reset button (1).
• Switch the suspension system back on again at the power switch.
If the first lamp has failed and the backup lamp is in use (segment in the knob
is lit), make sure to have a backup lamp module ready to hand as
a precaution.

Version 11.0
Page 206 G-30-1435-en
S7 / OPMI PROergo What to do in the event of malfunctions

Causes of malfunctions and remedies

For your safety


This instrument is a high-grade technological product. To ensure optimum
performance and safe working order, we recommend having it checked
by our service representative as part of regular scheduled maintenance.
If a failure occurs which cannot be remedied by referring to the section
"What to do in the event of malfunctions", attach a sign to the system
stating that it is out of order and contact our service representative.
microscope

Fault Possible reason Corrective measure Cross


reference
Handgrip buttons without Power failure. Contact in-house electrician.
function.
Suspension system not Switch on the stand at the
switched on. power switch.
Automatic circuit breaker in Press power switch again.
power switch of suspension
If the system does not work,
system has been activated.
operate the surgical
microscope manually.
Zoom or focus without Defective motor. Adjust zoom or focus
function. manually.
Notify Service.
Zoom or focus system always Open the cover on the
moves to upper or lower end suspension arm and
position. disconnect the cable to
the surgical microscope
(large connector).
Notify Service.
Magnetic brakes inoperative. Magnet brakes have been Switch menu to "Off":
deactivated ("On") in the Brakes are active.
menu: Surgery/Options/
Magnetic Brakes.
Defective magnetic brakes. Manually position
microscope by overcoming
the braking effect.
Notify Service.

Version 11.0
G-30-1435-en Page 207
What to do in the event of malfunctions S7 / OPMI PROergo

Fault Possible reason Corrective measure Cross


reference
Surgical microscope tilts Surgical microscope not Balance surgical
forward / backward or balanced or incorrectly microscope.
sideward. balanced.
Motion of surgical Friction adjustment knobs Loosen friction adjustment
microscope too stiff. tightened too firmly. knobs and adjust friction
correctly.

S7 suspension system

Fault Possible reason Corrective measure Cross


reference
No function at all. Power failure. Contact in-house
electrician.
Suspension system not Switch on the stand at Page 116
switched on. the power switch.
Automatic circuit breaker in Press power switch again.
power switch of suspension If there continues to be
system has been activated. no function, contact our
service representative.
Surgical field illumination on Defective main and backup Change lamp or insert Page 220
microscope not working. lamps. backup lamp module.
Thermal cut-out in lamp Remove the cause of
housing has been activated. overheating. For example,
drapes could be covering
the ventilation grid of the
lamp housing. When the
lamp module has cooled
down, the illumination
switches on again.
Stand electronics failure. Illuminate surgical area
with an OR illuminator.
Notify Service.

Version 11.0
Page 208 G-30-1435-en
S7 / OPMI PROergo What to do in the event of malfunctions

Fault Possible reason Corrective measure Cross


reference
Insufficient surgical field Brightness level set too low. Adjust brightness on the Page 122
illumination. suspension system's display
panel or using the foot
control panel.
Halogen lamp not properly Properly push halogen Page 220
plugged into lamp mount. lamp into lamp mount.
Defective light guide Notify Service.
(illumination not uniform). The light guide probably
needs to be replaced.
Motorized focusing and / Stand electronics failure. Manually adjust focusing
or zoom function of surgical and / or zoom on surgical
microscope is inoperative. microscope.
Motion of suspension arm Friction adjustment knob Slightly loosen friction Page 78
too stiff. tightened too firmly. adjustment knob. Page 78
Floor stand wobbles. Floor not level. Stand base Adjust the position of the
not appropriately positioned. stand base.
No video image. Connection cable not Check all connections.
properly connected.
No video image, only color Camera head not connected. Connect the camera head. Page 154
bars.
Image with no image. Y/C cable between CCU Plug connector in properly.
and monitor not properly
plugged in.
Defective Y/C cable between Use different Y/C cable or
CCU and monitor. BNC cable.
Video indicator lamp is lit, Monitor not switched on Switch on monitor or set Please
but there is no image. or not correctly set. it to correct channel. refer to
monitor's
Set monitor to default status,
Instructions
or increase brightness and
for Use.
contrast.
Connecting cable between Plug connector in properly.
CCU and monitor or
between CCU and surgical
microscope not or not
properly plugged in.

Version 11.0
G-30-1435-en Page 209
What to do in the event of malfunctions S7 / OPMI PROergo

Fault Possible reason Corrective measure Cross


reference
Defective connecting cable Replace the cable.
between video unit and
monitor.
If a video recorder is Connect CCU and monitor
connected between CCU directly.
and monitor: defective
video recorder.
Defective CCU or camera Notify Service.
head in surgical microscope,
or defective connecting
cable between CCU and
surgical microscope.
Picture too dark / Monitor not correctly set. Set monitor to default status, Please
snowing on the monitor or increase brightness and refer to
contrast. monitor's
Instructions
for Use.
Brightness of video unit Increase brightness of
set too low. video unit.
Surgical field illumination Increase brightness of Page 122
too dark. illumination.
Picture too bright / Monitor not correctly set. Set monitor to default Please
image on monitor status, or reduce refer to
appears milky. brightness and/or monitor's
contrast. Instructions
for Use.
Brightness of video unit Reduce brightness of
set too high. video unit.
Surgical field illumination Reduce brightness of Page 122
too bright. surgical field illumination.

Version 11.0
Page 210 G-30-1435-en
S7 / OPMI PROergo What to do in the event of malfunctions

SpeedFokus video autofocus

Fault Possible reason Corrective measure Cross


reference
SpeedFokus video Camera control unit (CCU) • Switch on MediLive
autofocus is inoperative. switched off. controller (CCU).
Monitor switched off. • Switch on the monitor.
Failure of the SpeedFokus • Notify Service.
video autofocus system.
Poor focusing result Focusing of the video • Focus the microscope
or none at all. camera not matched to image, then the video
focusing of the surgical image.
microscope.
Measuring field (ROI) • Correct the ROI position Page 128
incorrectly positioned. in submenu "AF - ROI
Position".
Suspension system moved • Do not move suspension
during focusing. system during focusing.
No object in focusing • Reposition surgical
range. microscope.
No object with sufficient • Reposition surgical
contrast in focusing microscope or focus
range. manually.
In external video camera: Incorrect entry for camera • Call up Autofocus Page 127
Poor focusing result orientation (left/right). menu and change
or none at all. Camera Position.
Incorrect entry for Video • Call up Advanced Page 129
Focal Length. Settings submenu
and enter correct
focal length.
Video camera not correctly • Turn video camera
oriented. so that the video image
corresponds to main
observer's viewing
direction.
Focused video image on Setting of diopter rings on • Set diopter rings to zero Page 74
monitor, but unfocused eyepieces has been changed. or to the correct value.
image in surgical
microscope

Version 11.0
G-30-1435-en Page 211
What to do in the event of malfunctions S7 / OPMI PROergo

Fault Possible reason Corrective measure Cross


reference
Focused image in Focusing of video camera • Focus the microscope
surgical microscope, has not been matched to image, then the video
but unfocused video focusing of surgical image.
image on monitor microscope.

Focusing cannot Different function assigned • Call up the Handgrip Page 124
be activated to handgrip button Settings menu and
(A, B, C or D). assign the Autofocus
function to one of the
buttons.
Different function assigned • Call up the Foot Control Page 124
to foot control panel menu and assign the
button (C or D). Autofocus function to
one of the buttons.

Version 11.0
Page 212 G-30-1435-en
S7 / OPMI PROergo Care and maintenance

Care and maintenance

Care of the device ......................................................................... 214


Cleaning .................................................................................................214
Sterilization ............................................................................................215
Disinfection ............................................................................................216

Device maintenance ......................................................................217


Maintenance intervals .............................................................................217
Safety check ...........................................................................................218
Changing the halogen lamp ...................................................................220
Xenon lamp module, changing ...............................................................222
Spare parts .............................................................................................224
Disposal ..................................................................................................225

Version 11.0
G-30-1435-en Page 213
Care and maintenance S7 / OPMI PROergo

Care of the device

NOTE Risk of injury to the head when using ceiling mounts!


Your head may collide with the device while the device is lowered.
• Move ceiling mount to its standby position when it is not in use.

Cleaning

Cleaning the optical surfaces


The multilayer T* super anti-reflection coating of the optical components
(e.g. eyepieces, objective lenses) ensures optimum image quality.
Even very slight soiling or finger marks can reduce the quality of the image.
To protect the device's internal optics against dust, the device should never
be left unused without the objective lens, binocular tube and eyepieces
attached. After use, cover the device to protect it from dust. Always store
additional objective lenses, eyepieces and accessories in dust-free cases when
they are not being used.
Clean the outer surfaces of optical components (eyepieces, objective lenses)
only as required:
• Do not use aggressive or abrasive cleaning agents.
• Remove dust from the optical surfaces using a squeeze blower or a clean,
grease-free brush.
For regular cleaning of the surgical microscope’s objective lenses and
eyepieces, we recommend using the optics cleaning kit available from ZEISS.
For the catalog number, please see the section "Device data".

Fogging of optical surfaces


To keep the objective lenses from fogging, we recommend using an anti-
fogging agent. Anti-fogging agents provided by eyecare professionals for
use with eyeglass lenses are also suitable for ZEISS objective lenses.
• Observe the Instructions for Use pertaining to the anti-fogging agent
concerned.
An anti-fogging agent does not ensure fog-free eyepiece optics. It cleans
eyepiece optics and protects them against dirt, grease, dust, lint and
fingerprints.

Version 11.0
Page 214 G-30-1435-en
S7 / OPMI PROergo Care and maintenance

Cleaning the mechanical surfaces


All of the system's mechanical surfaces can be wiped clean with a damp cloth.
Do not use any aggressive or abrasive cleaning agents.
Remove any residue using a mixture of 50% ethyl alcohol and 50% distilled
water plus a dash of household dish-washing liquid.

Sterilization

CAUTION Risk of infection due to soiled aseptic accessories and drapes!


• Only use sterile asepsis accessories and drapes suitable for the system!
• The products contained in the asepsis sets must be cleaned, disinfected
and sterilized before each use. This also applies to the first use after
delivery.
• Make sure that the operating personnel have been instructed in the
sterile operation of the device.

Asepsis sets
The asepsis sets available from ZEISS contain rubber caps and hand grips
that can be sterilized in autoclaves. For more detailed information about
sterilization, please refer to the enclosed "Preparation of resterilizable
products" manual for each Asepsis Set.

Sterile drapes (drapes)


Sterile single-use drapes are available to cover the system.
The drapes recommended by ZEISS are listed Page 243 in the section
"Asepsis ordering data".

• Loosely attach the sterile drapes so that there is sufficient space for the
microscope supporting stand and the surgical microscope to move.

Version 11.0
G-30-1435-en Page 215
Care and maintenance S7 / OPMI PROergo

NOTE Do not cover ventilation slots


Covering the ventilation slots can cause the device to overheat, resulting in
damage.
• Do not cover the ventilation slots in order to ensure that the lamps are
able to cool sufficiently and to avoid any loss in lamp functionality.

• Loosely fasten the drape using the adhesive strips.

Disinfection

CAUTION This could damage the paint on the device.


The use of unapproved disinfectants can cause damage to the paintwork on
the device. Flakes of paint could come loose and fall into the surgical field.
• Use a disinfectant based on aldehyde or alcohol. The addition of
quaternary compounds is acceptable.
• To avoid damaging the surfaces, do not use any disinfecting components
other than those specified below.
The maximum application concentrations are:
– For alcohol (tested with 2 propanol): 60%
– For aldehyde (tested with glutaraldehyde): 2%
– For quaternary compounds (tested with DDAC): 0.2%

Version 11.0
Page 216 G-30-1435-en
S7 / OPMI PROergo Care and maintenance

Device maintenance

Electromagnetic No regular inspections or maintenance procedures are required in order to


compatibility (EMC) maintain electromagnetic compatibility (EMC).

Maintenance intervals

Regular preventive maintenance is required in order to ensure safe and correct


operation of the device as well as a long service life.
The following table lists the relevant maintenance activities and shows at
which maximum intervals they must be performed.
The maintenance activities that you are able to perform yourself are described
below.
Any other maintenance activities that is not described here in detail requires
device-specific knowledge.
Please contact your local ZEISS Service or authorized partner to arrange for
the timely performance of these maintenance activities.

Interval Maintenance tasks


Every 2 years Visual inspection and function test of the following modules:
– Suspension system with carrier arm
– Illumination and power supply unit
– Surgical microscope with binocular tube
– Light guide
– Brakes
– Power cord
– Function test of limit of downward travel
– Connected accessories
– Function test of video system (option)
In accordance with Safety check
country-specific
requirements

Version 11.0
G-30-1435-en Page 217
Care and maintenance S7 / OPMI PROergo

Safety check

WARNING Improper maintenance!


Cables and plug connections may come loose and parts of the housing may
become live. The user and patient may suffer an electric shock.
• Make sure that the regular technical safety checks required for this
system in accordance with the applicable national regulations are
performed on schedule and to the stipulated extent.
The safety checks must only be performed by the provider of your overall
system.
The scope of the system safety checks should include at least the following
points:
– Availability of the Instructions for Use
– Visual inspection of the device and accessories for damage, as well as
legibility of the labels
– Leakage current test
– Test of protective ground conductor
– Function and wear test of the steerable casters and locking tabs
– Function test of all switches, buttons, sockets and indicator lamps of the
device

Version 11.0
Page 218 G-30-1435-en
S7 / OPMI PROergo Care and maintenance

Version 11.0
G-30-1435-en Page 219
Care and maintenance S7 / OPMI PROergo

Changing the halogen lamp

NOTE Risk of burns caused by hot lamp!


If you replace the lamp shortly after it has failed, the lamp will still be
very hot.
• Allow the lamp to cool down before changing it.
• Use heat-resistant protective gloves when replacing the lamp.
The following steps are required for changing the lamp:
• Turn off the device at the power switch.
• Press button (2), the lamp module is partially ejected. Pull out lamp
module (1).
• Remove defective halogen lamp (3, 4) from spring-loaded mount.
• Pull the ceramic base (7) from the contact pins of the halogen lamp.
• Plug the ceramic base (7) onto the contact pins of the new halogen lamp.
• Insert the new halogen lamp (3, 4). Make sure you do not touch the lamp
bulb (6) or the interior of the reflector (5).
• Press the halogen lamp into the spring-loaded mount.
• Push the lamp module, including the new halogen lamp, back into the
device.
• Turn on the system at its power switch.
Only use 12 V, 100 W halogen lamps available under the specified Cat. No.
For the catalog number, please see the section "Device data".
To use the service life of your halogen lamps as economically as possible, we
recommend that you proceed as follows: If main lamp (3) has failed, remove
it and replace it by backup lamp (4). Install the new halogen lamp instead of
the backup lamp.

Version 11.0
Page 220 G-30-1435-en
S7 / OPMI PROergo Care and maintenance

5 6 7

Version 11.0
G-30-1435-en Page 221
Care and maintenance S7 / OPMI PROergo

Xenon lamp module, changing

CAUTION Risk of injury caused by lamp rupture!


Lamp rupture (audible as a loud bang) may lead to jamming of the lamp
module and/or failure of the electronics modules.
• Before opening the lamp housing, ensure that the device is moved to
a position in which possibly falling particles cannot put the patient or user
at risk.
• Do not continue using the system if the lamp module is jammed or the
illumination is no longer operational due to defective electronic modules.
Contact our Service Department.

NOTE Risk of burns from hot lamp!


If you replace the lamp shortly after it has failed, the lamp will still be very
hot.
• Allow the lamp to cool down before removing it.
• Always use heat-resistant protective gloves when replacing the lamp!

CAUTION Injury caused by improper handling!


Improper handling of the xenon lamp may lead to damage or injury.
• The lamp module may be changed only by appropriately trained
personnel.
• Please observe the following points:
– First switch off the suspension system at the power switch.
– Only change the lamp module after it has cooled down completely!
In case of malfunction, the high pressure inside the hot lamp may
cause the lamp to burst. The hot surface of the xenon lamp may also
cause burns.
– Tap the key (1). The lamp module is slightly ejected.
– Pull out the lamp module as far as it will go.
– Slide the original transport case (2)over the module, Bolt (3) must
engage in hole (4). This unlocks the stop.
– Remove the old module and install the new lamp module by
proceeding in the reverse order.

Version 11.0
Page 222 G-30-1435-en
S7 / OPMI PROergo Care and maintenance

– Check the function of the xenon lamp and backup xenon lamp.
– Pack the old lamp module (5) in the transport packaging of the new
lamp module. Fill in the enclosed return card and send the old lamp
module to the nearest ZEISS service representative.
– Only use original the transport case (2), as it also provides explosion
protection should the xenon lamps be faulty.

Version 11.0
G-30-1435-en Page 223
Care and maintenance S7 / OPMI PROergo

Spare parts

Asepsis sets

Description Order number


Asepsis caps 27 mm (pack of 6) 305810-9008-000
Asepsis caps 22 mm (pack of 6) 305810-9001-000
asepsis caps for 180° tiltable tube (pack of 6) 305810-9003-000

Halogen lighting system

Description Order no.


Halogen lamp, 12 V, 100 W 380079-9040-000

Xenon lighting system

Description Order number


Complete replacement lamp module with 2 xenon 304977-9036-700
lamps in transport container and with return card;
in exchange for a returned module with defective
xenon lamps

Complete xenon lamp module with 2 xenon lamps 304977-9036-000


(new component)

Version 11.0
Page 224 G-30-1435-en
S7 / OPMI PROergo Care and maintenance

Disposal

User information on the disposal of electrical and electronic devices


This symbol means that the product must not be disposed of as normal
domestic waste.
The correct disposal of electrical or electronic devices helps to protect the
environment and to prevent potential hazards to the environment and/or
human health which may occur as a result of improper handling of the devices
concerned.
For detailed information on the disposal of the product, please contact your
local dealer or the device manufacturer or its legal successor. Please also
consult the manufacturer's current information provided on the Internet.
In the event of resale of the product or its components, the seller is required
to inform the buyer that the product must be disposed of in accordance with
the applicable national regulations currently in force.
For customers in the European Union
Please contact your dealer or supplier if you wish to dispose of electrical or
electronic devices.
Information on disposal in countries outside the European Union
This symbol is only applicable in the European Union. For the disposal of
electrical and electronic devices, please observe the relevant national
legislation and other regulations applicable in your country.
Disposal of outer packaging material
• Dispose of the outer packaging material according to your local directives
and laws.

Version 11.0
G-30-1435-en Page 225
Care and maintenance S7 / OPMI PROergo

Version 11.0
Page 226 G-30-1435-en
S7 / OPMI PROergo Device data

Device data

Technical data .............................................................................. 228


Essential performance features S7 / OPMI PROergo ................................228
Surgical microscope ................................................................................228
Stands/mounts .......................................................................................230
Technical data on video equipment ........................................................238
Ordering data .........................................................................................239

Compliance ................................................................................... 250

Environmental conditions ............................................................. 251

Guidelines and manufacturer's declaration concerning


electromagnetic compatibility ...................................................... 252

EMC – Electromagnetic compatibility IEC 60601-1-2: 2014


(4th edition) ..................................................................................254
Electromagnetic interference ..................................................................254
Electromagnetic immunity for all ME equipment and systems .................255
Electromagnetic immunity for ME equipment and systems
that are not life-sustaining ......................................................................256

System enhancements .................................................................. 257

Version 11.0
G-30-1435-en Page 227
Device data S7 / OPMI PROergo

Technical data

Essential performance features S7 / OPMI PROergo

S7 / OPMI PROergo does not have any essential performance features as


defined in IEC 60601-1.
In lieu of testing the device's essential performance features, certain device
functions defined by the manufacturer were tested.
The following device functions have been defined for the immunity test:
– Visualization of the object must be maintained (minor image distortions
are permitted).
– The lighting for object illumination must be continuously maintained
(interruptions of less than one minute are permitted).
– The zoom must be adjustable.
– Focusing must be possible.
– Movement of all stand axes in all directions must be maintained.
– Operation via the foot control panel must be maintained.
– Operation via the control panel must be maintained.
– The video image must be available.

Surgical microscope

Component Properties
Magnification Motorized zoom system, zoom ratio 1:6,
magnification factor γ = 0.4x - 2.4x.
Total magnification 2.5x – 14.9x at a working distance of 200 mm
(10x eyepiece) 1.5x – 8.9x at a working distance of 415 mm
Total magnification 3.1x – 18.6x at a working distance of 200 mm
(12.5x eyepiece) 1.8x – 11.1x at a working distance of 415 mm
Field of view 14 – 83 mm at a working distance of 200 mm
(10x eyepiece) 23 – 140 mm at a working distance of 415 mm

Version 11.0
Page 228 G-30-1435-en
S7 / OPMI PROergo Device data

Component Properties
Fields of view 12–71 mm at a working distance of 200 mm
(12.5x eyepiece) 20–119 mm at a working distance of 415 mm
Focus Continuous motorized focusing via Varioskop
focusing range 200 - 415 mm
Note: The values shown on the display are different
from the real values. They are rounded values and
serve for information only.
Focal lengths f At a working distance of 200 mm: f = 278 mm
At a working distance of 300 mm: f = 366 mm
At a working distance of 415 mm: f = 466 mm
Light field diameter At a working distance of 200 mm:
min. 11 mm – max. 85 mm
At a working distance of 415 mm:
min. 19 mm – max. 142 mm
Tubes / Eyepieces 180° tiltable binocular tube, f = 170 mm.
10x or 12.5x widefield eyepieces with magnetic
coupling.
Weight Approx. 9.2 kg (without tube and eyepieces)

Version 11.0
G-30-1435-en Page 229
Device data S7 / OPMI PROergo

Stands/mounts

Electrical data (applies to all S7 stands/mounts)

Component Properties
Power supply Only connect the floor stand to wall outlets which
are provided with a properly connected protective
ground conductor.
Rated Voltage (115 V): 100–125 V
(230 V): 220–240 V
Power consumption Floor stand:
(115 V): max. 850 VA
Remote connection socket (230 V): max.1250 VA
View of connector side Ceiling mount with rigid column or lifting column:
(115 V): max. 450 VA
2 (230 V): max. 450 VA
Rated Frequency 50–60 Hz
3 1 Fuses Automatic circuit breaker
Electrical – Power outlet (floor stand only)
outputs (115 V): max. 400 VA
1 (230 V): max. 800 VA
2 – Surgical microscope
– Remote socket for an external signal with max.
3
24 V / 0.5 A (floor stand only)
The device is designed for continuous operation.

Halogen lighting system

Fiber-optic
illumination 2 halogen reflector lamps 12 V 100 W in quick-
change module for light guide, with GG 475 and KK
40 filters
Fully automated switching between lamps.

Version 11.0
Page 230 G-30-1435-en
S7 / OPMI PROergo Device data

Xenon lighting system


Fiber optic Xenon short-arc reflector lamp
illumination Color temperature: approx. 5000 K
Rated output: 180 W, rated voltage approx. 14.5 V,
lamp current: 12 A, ignition voltage: 20 kV
Backup lamp in lamp housing, manually selectable.

Version 11.0
G-30-1435-en Page 231
Device data S7 / OPMI PROergo

Floor stand S7
Mechanics
Component Properties
Support arm Length: 450 mm
Swivel angle: 360°
Suspension arm Length: 865 mm
Swivel angle: 305°
Lifting range: -320 mm / +400 mm
coupling Swivel angle: 350°
Stand height 1730 mm
Base dimensions 640 x 640 mm
Max. weight load on 14 kg (complete microscope equipment,
suspension arm including accessories)
Weight Approx. 177 kg (including microscope)
Maximum total weight 190 kg (with installed safe workload)

305°

Version 11.0
Page 232 G-30-1435-en
S7 / OPMI PROergo Device data

S7 ceiling mount
Mechanics
Component Properties
Support arm Length: 450 mm
Swivel angle: 296°
Suspension arm Length: 865 mm
Swivel angle: 296°
Lifting range: -320 mm / +400 mm
coupling Swivel angle: 350°
Admissible 14 kg (complete microscope equipment,
max. load on including accessories)
suspension arm
Weight (including Approx. 80 kg
microscope)

Version 11.0
G-30-1435-en Page 233
Device data S7 / OPMI PROergo

S7 ceiling mount with lifting column


Mechanics
Component Properties
Support arm Length: 450 mm
Swivel angle: 296°
Suspension arm Length: 865 mm
Swivel angle: 296°
Lifting range: -320 mm / +400 mm
coupling Swivel angle: 350°
Max. weight load 14 kg (complete microscope equipment,
on suspension arm including accessories)
Weight (including Approx. 133 kg
microscope)

Version 11.0
Page 234 G-30-1435-en
S7 / OPMI PROergo Device data

Technical data of lifting column


Rated Voltage 24 V DC
Current Consumption 5A
Vertical lift, total 350 mm
Lifting force 2000 N
Lifting speed 5.5 mm/sec
Interval (operation / pause) 1 min / 9 min

Version 11.0
G-30-1435-en Page 235
Device data S7 / OPMI PROergo

S7 wall mount
Mechanics
Component Properties
Support arm Length: 450 mm
Swivel angle: 190°
Suspension arm Length: 865 mm Length: 1065 mm
Swivel angle: 360° Swivel angle: 360°
Lifting range: +400 mm Lifting range: +540 mm
-320 m -390 mm
coupling Swivel angle: 350°
Admissible 14 kg (complete microscope equipment,
max. load on including accessories)
suspension arm
Weight (including Approx. 68 kg Approx. 69 kg
microscope)

Version 11.0
Page 236 G-30-1435-en
S7 / OPMI PROergo Device data

S7 central stand
Mechanics
Component Properties
Support arm Length: 450 mm
Swivel angle: 190°
Suspension arm Length: 1065 mm
Swivel angle: 360°
Lifting range: +540 mm
-390 mm
coupling Swivel angle: 350°
Max. weight load 14 kg (complete microscope equipment,
on suspension arm including accessories)
Weight (including Approx. 53 kg
microscope)

540
390

Version 11.0
G-30-1435-en Page 237
Device data S7 / OPMI PROergo

Technical data on video equipment

1Chip HD camera (option)


• See Instructions for Use G-30-1946 "1Chip HD Camera"

TRIO 610 (Option)


• See Instructions for Use G-30-1725 "TRIO 610 with CCU TRIO 600"

Version 11.0
Page 238 G-30-1435-en
S7 / OPMI PROergo Device data

Ordering data

Only operate the device using the accessories recommended by ZEISS and
included in the delivery package. To find the contact person responsible for
orders in your country, visit:
www.zeiss.com/med
For authorized ZEISS partners, the contractual product configurations,
accessories and spare parts apply.
Please contact your authorized ZEISS partner for more information.
The given catalog number refers to the basic component. Further components
for the respective function are evident from the descriptive text. For this
reason, you should also cite the descriptive text in any orders.

Ordering data for OPMI PROergo surgical microscope

Description Order number


PROergo surgical microscope 302932-9901-000
OPMI mount adapter for OPMI Sensera and PROergo 000000-1117-216
Dust cover, blue, with ZEISS logo 000000-1055-278
S light guide, 2.0 m 303481-9020-000

Ordering data for S7 suspension systems

Description Order number


S7 floor stand for OPMI® PROergo
– S7 floor stand, basic equipment 000000-1087-851
– S7/1 floor stand covering 000000-1120-147
S7 ceiling mount with rigid column, full without ceiling
flange 000000-1103-022
S7 ceiling mount with lifting column without
pre-installation set 000000-1226-405
S7 wall mount, with short suspension arm 000000-1193-109
S7 wall mount, with long suspension arm 000000-1254-579
S7 central stand 000000-1254-580

Version 11.0
G-30-1435-en Page 239
Device data S7 / OPMI PROergo

Ordering data for country-specific cables for the S7 floor stand

Description Order no.


Europe 000000-0301-997
USA 000000-0147-000
UK 000000-0400-264
Switzerland 309850-9011-000
Argentina 000000-0434-527
China 000000-0475-507
Brazil 000000-0527-730

Ordering data for illumination

Description Order number


Halogen light source for S7 suspension systems 000000-1119-657
Illumination, xenon
– Xenon Superlux 180 light source for S 7
suspension systems 000000-1517-743
– UV protective glasses, tinted 000000-0480-028

Ordering data for tubes

Description Order no.


Tiltable binocular tube, f = 170 mm, 180° 303791-0000-000
Foldable tube f170/f260 incl. Rotate and PROMAG 303771-9020-000
(not to be used with a mouth switch)

Ordering data for eyepieces

Description Order number


12.5x push-in widefield eyepiece, asph. (2 pcs required) 305543-9901-000
10x push-in widefield eyepiece (2x required) 305542-0000-000
12.5x push-in widefield eyepieces
(1x with "format" reticle)
– 12.5x push-in widefield eyepiece, asph. 305543-9901-000
– 12.5x push-in widefield eyepiece, asph.
with "format" reticle 000000-1023-188

Version 11.0
Page 240 G-30-1435-en
S7 / OPMI PROergo Device data

Description Order number


10x push-in widefield eyepieces
(1x with "format" reticle)
– 10x push-in widefield eyepiece 305542-0000-000
– 10x push-in widefield eyepiece with "format"
reticle 000000-1023-184
12.5x widefield push-in eyepieces
(1x with "focusing aid" reticle)
– 12.5x widefield push-in eyepiece 305543-9901-000
– 12.5x widefield push-in eyepieces
(1x with "focusing aid" reticle) 000000-1023-186
10x widefield push-in eyepieces
(1x with "focusing aid" reticle)
– Widefield push-in eyepiece 10x 305542-0000-000
– 10x widefield push-in eyepieces
(1x with "focusing aid" reticle) 000000-1023-183

Ordering data for coobservation

Description Order number


Stereo co-observation tube, complete, with straight
tube and 10x eyepieces
– Stereo co-observation module with 2 pivot joints 000000-1063-869
– Straight binocular tube, f=170 mm 303765-0000-000
– Sleeve for tubes with screw thread (2 pieces) 305542-0107-000
– 10x push-in widefield eyepiece (2x required) 305542-0000-000
Stereo co-observation tube, complete, with 180° tube
and 10x eyepieces
– Stereo co-observation module with 2 pivot joints 000000-1063-869
– Tiltable binocular tube, f = 170 mm, 180°
(2x required) 303791-0000-000
– 10x push-in widefield eyepiece (2x required) 305542-0000-000

Version 11.0
G-30-1435-en Page 241
Device data S7 / OPMI PROergo

Ordering data for angled lens

Description Order number


Angled lens without lateral output for documentation
accessories 000000-1099-230
Angled lens without documentation port with tube
dovetail
– 1 angled lens without lateral output for
documentation accessories 000000-1099-230
– 1 rotary dovetail for binocular tubes 301007-0000-000
Angled lens with lateral output port for documentation 000000-1079-239
accessories
Angled lens with documentation port with tube dovetail
– Angled lens with lateral output port for
documentation accessories 000000-1079-239
– Rotary dovetail for binocular tubes 301007-0000-000

Ordering data for accessories

Description Order number


Double iris diaphragm for increasing the depth of field.
– Adapter with double iris diaphragm 303354-0000-000
Wall socket for S7 ceiling mount 000000-1351-565
Foot control panel for focus and zoom operation
– Socket for foot control panel for S7 suspension
systems for OPMI Sensera / OPMI PROergo
(only required for operation of existing foot
control panels) 000000-1191-636
– Foot control panel for zoom and focus
– S7 foot control panel for zoom and focus 305946-9200-000
– Socket for foot control panel for S7 suspension
systems for OPMI Sensera / OPMI PROergo 000000-1191-636
Video Instrument tray for S7 floor stand
– Instrument tray for S7 floor stand 000000-1352-922
Set for attaching a Heraeus light at S7 ceiling mount
with lifting column 000000-1217-044

Version 11.0
Page 242 G-30-1435-en
S7 / OPMI PROergo Device data

Asepsis ordering data

Description Order number


Asepsis caps
(re-sterilizable rubber caps for the OPMI PROergo)
– Asepsis caps 27 mm (6 caps per pack) 305810-9008-000
– Asepsis caps 22 mm (6 caps per pack) 305810-9001-000
– Asepsis caps, for PD adjustment, 180° tiltable tube
(rubber caps for the interpupillar distance setting
on the 180° tiltable tube, 6 caps per pack) 305810-9003-000
Sterile disposable drapes
– OPMI Drapes sterile
(Type 70, ZEISS VisionGuard drapes for the OPMI
with 2 observation tubes and camera. Dimensions:
122 cm x 209 cm) pack with 5 drapes 306070-0000-000
– DRAPES sterile
(type 88, ZEISS Drapes for handgrips)
pack of 20 326088-0000-000
– VisionGuard Replacement Lenses
20 per box 306001-0000-000
– VisionGuard Lens Protectors
10 per box 306002-0000-000

Cleaning agents

Description Order no.


Optics cleaning kit 000000-1216-071
Microfiber cleaning cloth 000000-1254-655

Ordering data for spare parts

Description Order number


Halogen lamp 12 V, 100 W (2 pcs) 380079-9040-000
N-xenon lamp module incl. 2 xenon lamps in
transport case for Superlux 180 304977-9036-000

Version 11.0
G-30-1435-en Page 243
Device data S7 / OPMI PROergo

Ordering data for SpeedFokus with external HD video camera

Description Order number


SpeedFokus for external camera with standard cable
– Integrated MediLive video control unit/SpeedFokus 308203-9240-000
– Standard cable set, 5m external video rack 000000-1294-229
– Electrical components kit for S7 suspension systems
for OPMI Sensera / ProErgo 308203-9210-000
SpeedFokus for external camera with MediLive
mounting kit (floor stand only)
– Integrated MediLive video control unit/SpeedFokus 308203-9240-000
– Instrument tray for S7 floor stand 000000-1352-922
– 5m S-VHS connection cable 2x 4 pin Mini DIN 000000-0300-281
– Cover for OPMI without video 000000-1184-640
– Electrical components kit for S7 suspension systems
for OPMI Sensera / ProErgo 308203-9210-000
SpeedFokus for external camera with Lemo cable
– Integrated MediLive video control unit/SpeedFokus 308203-9240-000
– Lemo 2x Y/C connecting cable, 10m 000000-0418-950
– SpeedFokus connecting cable for Lemo 000000-1293-810
– Cover for OPMI without video 000000-1184-640
– Electrical components kit for S7 suspension systems
for OPMI Sensera / ProErgo 308203-9210-000

Version 11.0
Page 244 G-30-1435-en
S7 / OPMI PROergo Device data

Ordering data for 1Chip HD camera option for S7 / OPMI PROergo

Description Order no.


1Chip HD camera:
– 1Chip HD camera for the OPMI PROergo/Sensera 301900-8200-000
– System cable DVI-D, 5 m 302681-8767-000
Beam splitter for the 1Chip HD camera:
– Splitter 20 (beam splitter 20) 301503-9901-000
– Splitter 50 (beam splitter 50) 301513-9902-000
– Angled lens with lateral output port for
documentation accessories 000000-1079-239
– Angled lens with lateral output port for
documentation accessories (with rotatable adapter):
– Angled lens with lateral output port for
documentation accessories 000000-1079-239
– Rotatable dovetail for binocular tubes 301007-0000-000
Iris diaphragm for video lenses 301675-0000.000
SpeedFokus for the 1Chip HD camera
– Integrated MediLive video control unit/SpeedFokus 308203-9240-000
With the "1Chip HD Camera" option for OPMI PROergo you automatically
receive the "1Chip HD Camera" Instructions for Use (G-30-1946).

Ordering data for UC Dental PROergo

Description Order number


Upgrade for OPMI PROergo with 1Chip HD camera:
– UC 1Chip HD camera - OPMI PROergo/Sensera 301900-8200-500
– System cable DVI-D, 5 m 302681-8767-000
– Beam splitters for the 1Chip HD camera (if customer
does not have):
– Splitter 20 (beam splitter 20) 301503-9901-000
– Splitter 50 (beam splitter 50) 301513-9902-000
– Angled lens with lateral output port for
documentation accessories 000000-1079-239

Version 11.0
G-30-1435-en Page 245
Device data S7 / OPMI PROergo

Description Order number


With the "1Chip HD Camera" option for OPMI PROergo you automatically
receive the "1Chip HD Camera" Instructions for Use (G-30-1946).
SpeedFokus upgrade for the 1Chip HD camera
– UC SpeedFokus for 1Chip HD camera on OPMI
Sensera and OPMI PROergo 308203-9240-500
SpeedFokus for other external cameras
– UC SpeedFokus for external cameras on the
OPMI Sensera and standard cable set
– UC SpeedFokus for external camera on
OPMI PROergo 308203-9240-500
– Standard cable set for SpeedFokus, 5m 000000-1294-229
– UC SpeedFokus for external cameras on
OPMI PROergo and Lemo cable set
– UC SpeedFokus for external camera on
OPMI Sensera and OPMI ProErgo 308203-9240-500
– Y/C cable, external 000000-0418-950
– SpeedFokus connecting cable for LEMO 000000-1293-810
UC foot control panel for zoom and focus
– S7 foot control panel for zoom and focus 305946-9200-000
– UC foot control panel connector 000000-1237-540
UC Superlux xenon illumination for the OPMI PROergo
– UC Superlux 180 for S7 000000-1295-527
– Tinted UV protective eyewear 000000-0480-028

Ordering data for the 3-CCD camera (option)

Description Order number


TRIO 610 high definition camera system
– Trio 610 HDTV camera system with CCU TRIO 600 308203-2300-000
– System cable TRIO 610, straight connector, 4m 308203-2350-000
TRIO 610 extension cable 4m 308203-2400-000
With the "TRIO 610 HD camera" option you automatically receive the
"TRIO 610 with TRIO 600 CCU" Instructions for Use (G-30-1725).

Version 11.0
Page 246 G-30-1435-en
S7 / OPMI PROergo Device data

Video accessories for the 1Chip HD camera option

Description Order no.


External video recording
Sony HD recorder 308203-9435-000
HD recorder – Sony HVO-500 MD HD with cables
– Sony HVO-500MD HD recorder 000000-0575-031
– Adapter cable 000000-0576-774
– Connection cable – video recorder to AUX interface 000000-1052-814
– DVI monitor cable, 2m 308203-3020-000
– Cable tie set (20 pcs) 301687-9106-000
Keyboard USB EN 308203-2220-000
Foot switch, 10m 000000-1179-860
10 m-network cable, green, 2x RJ45 305946-8660-000
Monitors - external
Sony 21" full HD monitor 308203-9430-000
NDS 24" HDTV monitor with 5m HD-SDI video cable
and stand base:
– NDS 24" HDTV monitor 308203-2021-000
– Base for 23"/24" HDTV monitor 000000-0498-394
– HD-SDI video cable, 75 Ohm 2xBNC pin 5m 305989-8763-000
NDS 26" HDTV monitor with 5 m-HD-SDI video cable
and stand base (not for KSK 150 FC):
– NDS 26'' HDTV monitor 308203-2100-000
– Base for 23"/24" HDTV monitor 000000-0505-511
– HD-SDI video cable, 75 Ohm 2xBNC pin 5m 305989-8763-000
NDS 42" HDTV monitor with 5 m-HD-SDI video cable
and stand base: (not for KSK 150 FC)
– NDS 42" HDTV monitor 308203-2030-000
– HD-SDI video cable, 75 Ohm 2xBNC pin 5m 305989-8763-000

Version 11.0
G-30-1435-en Page 247
Device data S7 / OPMI PROergo

Description Order no.


Video cables
– System cable DVI-D, 5m 302681-8767-000
– DVI monitor cable, 2m 308203-3020-000
– 5m S-VHS connection cable 2x 4 pin Mini DIN 000000-0300-281
– 10m S-VHS connection cable 2x 4 pin Mini DIN,
sturdy and gold-plated contacts 308203-3070-000
– HD-SDI video cable, 75 Ohm 2xBNC pin 5m 305989-8763-000
– HD-SDI video cable, 75 Ohm 2x BNC pin 10 m 305989-8764-000
– HD-SDI video cable, 75 Ohm 2x BNC pin 15m 305989-8766-000
– Component HD YPbPr cable, 10m 308203-3080-000
Country-specific power cables for video recorders,
monitors and MEDIALINK 100:
– Europe 000000-0149-592
– United States 397902-9058-000
– United Kingdom 309850-9003-000
– Switzerland 000000-0228-202
– Argentina 000000-0434-527
– China 000000-0475-506
– Brazil 000000-0526-671

Ordering data for external video recording

Description Order no.


MEDIALINK 100 308203-2200-000
Network cable 10m, green 2xRJ45 305946-8660-000
Foot switch 10m 000000-1179-860
Keyboard USB EN 308203-2220-000
Remote control 308203-2210-000

Video accessories
For external video accessories for this surgical microscope, please refer to the
separate product overview G-30-1888.

Version 11.0
Page 248 G-30-1435-en
S7 / OPMI PROergo Device data

Foot control panel

Description Order number


Foot control panel for zoom and focus 000000-1267-950
Upgrade kit for 14-function foot control panel, 304970-9055-000
wired (FCP)
Upgrade kit for 14-function foot control panel, 304970-9060-000
wireless (FCP WL)
3 m connecting cable for 14-function foot control 304970-8735-000
panel (FCP & FCP WL)
(cable with angle plug)

Version 11.0
G-30-1435-en Page 249
Device data S7 / OPMI PROergo

Compliance

Directives and standards with which the following devices are


compliant:
– The device complies with Medical Device Directive 93/42/EEC: Class I
It is labeled with:
– This device satisfies the US American and Canadian safety regulations for
medical electrical devices/systems.
It is labeled with:
– The device complies with Directive 2011/65/EU on the restriction of
the use of certain hazardous substances in electrical and electronic
equipment.

Classification of the device according to IEC 60601-1


The device is classified as follows:
– Level of protection against electric shock: Class1
– Level of protection against entry of water: IPX0
– Electromagnetic compatibility (EMC): Fulfills IEC 60601-1-2, Class B
(as per CISPR 11)
– Operating mode: continuous operation

Version 11.0
Page 250 G-30-1435-en
S7 / OPMI PROergo Device data

Environmental conditions

For operation

Feature Admissible range


Temperature + 10°C ... + 40°C
Relative humidity 30 % ... 75 %
Air pressure 700 hPa ... 1060 hPa

For transport and storage

Feature Admissible range


Temperature Without 1Chip HD camera: - 40°C ... + 70°C
With 1Chip HD camera: - 20°C ... + 70°C
Relative humidity 10 % ... 90 %
(without condensa-
tion)
Air pressure 500 hPa ... 1060 hPa

Version 11.0
G-30-1435-en Page 251
Device data S7 / OPMI PROergo

Guidelines and manufacturer's declaration concerning


electromagnetic compatibility

The device is subject to specific precautions with regard to electromagnetic


compatibility (EMC) in Professional Healthcare Facility Environments. In order
to avoid the occurrence of EMC interference, the device may only be installed,
operated and maintained in the manner indicated in these Instructions for
Use, and only using components supplied by ZEISS.

WARNING Functional deterioration!


Do not install or operate the S7 / OPMI PROergo in close proximity to other
devices (with the exception of the combination of the devices described in
these Instructions for Use), as this can impair the functioning of the S7 / OPMI
PROergo.
• If it cannot be avoided that the S7 / OPMI PROergo is operated in
proximity to other devices, the proper function of the S7 / OPMI PROergo
must be monitored.

WARNING Functional deterioration!


Electrical devices can influence each other as a result of their electromagnetic
radiation. The use of non-approved components can cause increased
emissions or reduce the immunity of the S7 / OPMI PROergo.
• Only use accessories, transformers, cables and spare parts which are
specified in these Instructions for Use or which are approved by ZEISS for
the S7 / OPMI PROergo.

Version 11.0
Page 252 G-30-1435-en
S7 / OPMI PROergo Device data

WARNING Deterioration of performance!


• Do not use any portable or mobile RF communication equipment or
transmitters (including peripheral devices such as antenna cables or
external antennas) in the proximity of the S7 / OPMI PROergo (minimum
distance 30 cm). Otherwise it is possible that the functioning of the S7 /
OPMI PROergo may be affected or its performance impaired.
• Do not use cell phones near the S7 / OPMI PROergo. They represent
a potential risk for the proper functioning of medical equipment.
Malfunctions may be caused by a variety of local factors. Such
malfunctions cannot be predicted or assessed. in advance.
• Please follow the EMC guidelines on the following pages.

NOTE Danger due to electromagnetic radiation!


The S7 / OPMI PROergo may be disturbed by other devices, even when these
other devices comply with the applicable emission requirements according to
CISPR.
• Do not use the S7 / OPMI PROergo when it is located next to or stacked
on top of other devices.
• If operation next to or stacked on top of other devices is unavoidable, the
S7 / OPMI PROergo must be monitored to ensure that it is operating
normally in this configuration.

Version 11.0
G-30-1435-en Page 253
Device data S7 / OPMI PROergo

EMC – Electromagnetic compatibility


IEC 60601-1-2: 2014 (4th edition)

Electromagnetic interference

The S7 / OPMI PROergo is intended for operation in an electromagnetic


environment as specified below. The customer or the user of the S7 / OPMI
PROergo is responsible for ensuring that the device is operated in such an
environment.

Emitted interference measurements Compliance


HF emissions as per Group 1
CISPR 11
HF emissions as per Class B
CISPR 11
Harmonic emissions Class A
as per IEC 61000-3-2
Emission of voltage fluctuations/ Compliant
flicker as per IEC 61000-3-3

Version 11.0
Page 254 G-30-1435-en
S7 / OPMI PROergo Device data

Electromagnetic immunity for all ME equipment and systems

The S7 / OPMI PROergo is intended for operation in the electromagnetic


environment specified below. The customer or the user of the device S7 /
OPMI PROergo is responsible for ensuring that it is operated in such an
environment.

Electromagnetic IEC 60601 test level Compliance level


immunity tests
Electrostatic discharge ±8 kV ±8 kV
(ESD) as per IEC 61000-4-2 Contact discharge Contact discharge
±2 kV, ±4 kV, ±8 kV, ±15 kV ±2 kV, ±4 kV, ±8 kV, ±15 kV
Contact discharge Contact discharge
Fast transient/burst immunity ±2 kV for ±2 kV for
as per IEC 61000-4-4 power supply lines power supply lines
±1 kV for ±1 kV for
input/output lines input/output lines
Surges as per ±0.5 kV, ±1 kV ±0.5 kV, ±1 kV
IEC 61000-4-5 Phase neutral voltage Phase neutral voltage
±0.5 kV, ±1 kV, ±2 kV ±0.5 kV, ±1 kV, ±2 kV
Phase/neutral to ground voltage Phase/neutral to ground voltage
Magnetic field for supply 30 A/m 30 A/m
frequency (50/60 Hz) as per 50 Hz or 60 Hz 50 Hz or 60 Hz
IEC 61000-4-8
Voltage dips, short interruptions 0 % UT for 1/2 cycle 0 % UT for 1/2 cycle
and voltage variations as per at 0°, 45°, 90°, 135°, 180°, 225°, at 0°, 45°, 90°, 135°, 180°, 225°,
IEC 61000-4-11 270° and 315° 270° and 315°
0 % UT for 1 cycle 0 % UT for 1 cycle
and and
70 % UT for 25 cycles 70 % UT for 25 cycles
Single-phase: at 0 degrees Single-phase: at 0 degrees
0 % UT for 250 cycles 0 % UT for 250 cycles

Version 11.0
G-30-1435-en Page 255
Device data S7 / OPMI PROergo

Electromagnetic immunity for ME equipment and


systems that are not life-sustaining

The S7 / OPMI PROergo is intended for operation in the electromagnetic


environment specified below. The customer or the user of the S7 / OPMI
PROergo is responsible for ensuring that the device is operated in such an
environment.

Electromagnetic IEC 60601 test level Compliance level


immunity tests
Conducted HF disturbances as 3V 3V
per IEC 61000-4-6 150 kHz to 80 MHz
6V 6V
ISM-bands from 150 kHz to
80 MHz
SIP/SOPs with length <3 m tested
Radiated HF disturbances 3 V/m 3 V/m
as per IEC 61000-4-3 80–2700 MHz
Radiated HF disturbances 27 V/m 27 V/m
from near fields of wireless 380–390 MHz
communication devices
28 V/m 28 V/m
as per IEC 61000-4-3
430–470 MHz,
800–960 MHz,
1700–1990 MHz,
2400–2570 MHz
9 V/m 9 V/m
704–787 MHz,
5100–5800 MHz

Version 11.0
Page 256 G-30-1435-en
S7 / OPMI PROergo Device data

System enhancements

CAUTION As per IEC 60601-1-1 and IEC 60601-1, Section 16.2, this system's power
outlet is a multiple electrical outlet intended for use with an ME system.
Connecting electrical devices not approved by ZEISS can lead to a reduced
level of safety on the ME system.
• When configuring your ME system (using the power output socket),
make sure to comply with the requirements of IEC 60601-1, Section 16,
as well as the following measures:
– Do not place multiple electrical outlets on the floor.
– Do not connect any additional multiple electrical outlets.
– Only connect compatible components to the system.
– Do not exceed the max permissible electrical load of multiple
electrical outlets.
– Use multiple electrical outlets only for components of the system.

NOTE Connect accessories


Any additional equipment connected to medical electrical devices must
demonstrably comply with the applicable standards and directives
(e.g. IEC 60950 for data processing equipment).
Furthermore, all configurations must comply with the standards for medical
systems (see IEC 60601-1, section 16; in some countries IEC 60601-1-1 may
still be valid).
Anyone who connects auxiliary equipment to medical electrical devices is
a system configurer, and as such, is responsible for ensuring that the system
complies with the applicable system standards.
Please note that local legislation takes priority over the requirements set out
in the above-mentioned standards. If you have any questions, please contact
ZEISS Service or your local dealer.
The device S7 / OPMI PROergo can be extended with the accessories offered
by ZEISS (see ordering data/accessories) to form an already tested system. This
device combination has been approved by ZEISS.

Version 11.0
G-30-1435-en Page 257
Device data S7 / OPMI PROergo

Version 11.0
Page 258 G-30-1435-en
S7 / OPMI PROergo Keyword index

Keyword index

Numerics

1Chip HD camera option ........................................................................156


1Chip HD Camera with iris diaphragm ....................................................157

Accessories for 1Chip HD camera option ................................................247


Acoustic signals ......................................................................................132
Adjusting the eyepieces ..........................................................................173
Adjusting the pupillary distance ..............................................................173
Adjusting the surgical microscope ..........................................................173
Adjusting the system ......................................................................160, 164
Adjusting the system - Ceiling mount, wall mount and S7
Centro suspension system ......................................................................162
Advanced Settings, submenu ..................................................................129
AF - ROI Position submenu .....................................................................128
After every use .........................................................................................19
Ambient conditions ..................................................................................12
Area of application ...................................................................................10
Asepsis sets ............................................................................................175
Attaching the equipment ........................................................................142
Attaching the tube and eyepieces ...........................................................142
Automatic circuit breaker .......................................................................116
Available rated voltage ...........................................................................146

Backup lamp in use ..........................................................................24, 114


Balance menu .........................................................................................130
Balancing .............................................................................78, 88, 96, 102
Balancing the front-to-back tilt motion ...................................................170
Balancing the surgical microscope ..........................................................168
BNC video port .....................................................................82, 90, 98, 104
Bracket for foot control panel ...................................................................80

Cable connections ..................................................................................158


Cable connections for the 1Chip HD camera ...........................................158
Cable holder .............................................................................................80
Camera chip size (size of CCD array) .......................................................129

Version 11.0
G-30-1435-en Page 259
Keyword index S7 / OPMI PROergo

Care ....................................................................................................... 213


CCD array, size of .................................................................................. 129
CCD position .......................................................................................... 128
Ceiling anchor .......................................................................................... 15
Ceiling flange ..................................................................................... 86, 92
Ceiling mount with lifting column ............................................................ 26
Changing the halogen lamp ................................................................... 220
Changing the xenon lamp module ......................................................... 222
Checklist for S7 ceiling mount with lifting column .................................. 189
Checklist for S7 ceiling mount with rigid column .................................... 184
Checklist for S7 Centro suspension system ............................................. 184
Checklist for S7 floor stand .................................................................... 180
Checklist for S7 wall mount ................................................................... 184
Clamping screw for limiting downward movement .............. 78, 88, 96, 102
Cleaning ................................................................................................ 214
Cleaning the optical surfaces ................................................................. 214
Connecting a video monitor ................................................................... 154
Connecting the MEDIALINK 100 ............................................................ 154
Connecting the strain relief device ......................................................... 144
Connector for foot control panel, etc. .......................... 82, 90, 98, 104, 106
Connector panel .................................................................. 76, 86, 92, 100
Connector panel for S7 ceiling mount with lifting column ........................ 98
Control elements, displays, connections ................................................... 66
Control panel ............................................. 76, 80, 86, 88, 92, 96, 100, 102
Control panel and menu overview ......................................................... 116
Country of destination ........................................................................... 146
Crushing and collision, risk of .................................................................. 84

Device overview ....................................................................................... 57


Diopter scale ............................................................................................ 74
Diopter setting ring .................................................................................. 74
Disinfection ............................................................................................ 216
Display ................................................................................. 80, 88, 96, 102
Disposal ................................................................................................. 225
Disposal symbol ........................................................................... 34, 40, 46
Drapes ................................................................................................... 175

Edit User Menu ...................................................................................... 119


Electrical safety ........................................................................................ 13
EMC - Electromagnetic compatibility ...................................................... 254
Emergency backup power supply ............................................................. 12
Enter application name .......................................................................... 119
Enter user name ..................................................................................... 119
Environmental conditions ....................................................................... 251
Error message menu .............................................................................. 132

Version 11.0
Page 260 G-30-1435-en
S7 / OPMI PROergo Keyword index

Establishing connections .........................................................................144


Eyecups ....................................................................................................74
Eyepieces, technical data ........................................................................229

Failure of magnetic brakes ......................................................................199


Failure of main functions ........................................................................199
Fault .......................................................................................................211
Filter GG475 .....................................................................................23, 110
Filter KK 40 ......................................................................................23, 110
Filter selector ..................................................................................110, 114
Focal length of objective lens ..................................................................229
Focal lengths ..........................................................................................229
Focus menu ............................................................................................121
Focus start value .....................................................................................121
Focusing .................................................................................................229
Focusing function, failure .......................................................................200
Focusing process ....................................................................................128
Focusing range .......................................................................................229
Focusing speed .......................................................................................121
Foot Control menu .................................................................................124
Foot control panel ..................................................................................134
Friction ...........................................................................78, 80, 88, 96, 102
Function of buttons ................................................................................119
Functions of S7 Centro suspension system ..............................................100

Graphic display .......................................................................................128

Halogen illumination system .............................................................23, 110


Handle ...............................................................................................76, 80
Hazard symbols ..........................................................................................9

Illuminated field diameter .......................................................................229


Illumination intensity ..............................................................................122
Illumination menu ..................................................................................122
Illumination system .................................................................................108
Immobilizers .............................................................................................80
Information symbols ...................................................................................9
Instrument tray (option) ............................................................................84
Instrument tray strap ................................................................................84
Instrument tray, supplied strap .................................................................84
Intended use ............................................................................................10

Version 11.0
G-30-1435-en Page 261
Keyword index S7 / OPMI PROergo

Intensity start value ................................................................................ 122


Iris diaphragm for video lenses ............................................................... 157

Key to symbols .......................................................................................... 9

Labels on S7 wall mount and on S7 Centro suspension system ................ 46


Lamp failure ................................................................................... 202, 204
Lamp housing ...................................................................... 76, 86, 92, 100
Lamp housing, halogen or xenon ......................................... 78, 88, 96, 102
Lamp module ................................................................................. 110, 112
Laser menu ............................................................................................ 131
Legal regulations ..................................................................................... 12
Liability and warranty .............................................................................. 11
Lifting column .............................................................................. 26, 92, 94
Light ± ................................................................................................... 124
Lights on/off .......................................................................................... 124
Limit to downward travel ..................................................... 78, 88, 96, 102
Limiting the downward travel of the suspension arm ............. 160, 162, 166

Magnetic brakes, failure ........................................................................ 199


Magnification ........................................................................................ 228
Magnification factor .............................................................................. 228
Main beam path .................................................................................... 128
Malfunctions .......................................................................................... 132
Manual switching to the backup lamp ........................................... 110, 112
Manufacturer’s address ............................................................................. 2
Manufacturer’s information ....................................................................... 2
Menus, relationship ............................................................................... 116
Mounting the 1Chip HD Camera ............................................................ 156
Moving the device to a different location ............................................... 176

Normal use ........................................................................................ 10, 11

Observe the rated voltage ...................................................................... 146


Opening for system cable .......................................................... 90, 98, 104
Opening the lamp module ............................................................. 110, 114
Operating procedure ............................................................................. 194
Operation .............................................................................................. 179
Option, video autofocus ........................................................................ 127

Version 11.0
Page 262 G-30-1435-en
S7 / OPMI PROergo Keyword index

Options button .......................................................................................123


Options menu ................................................................123, 125, 131, 132
Orange filter .............................................................................................68
Ordering data .........................................................................................239
OSD on/off .............................................................................................124

Photo .....................................................................................................124
Pinhole diaphragm, medium-sized ............................................................68
Pinhole diaphragm, small ..........................................................................68
Potential equalization pin .....................................................82, 90, 98, 104
Power output socket ................................................................................82
Power switch ..........................................................................................116
Preparations for use ...............................................................................139
Prior to the very first use ...........................................................................15
Protective ground conductor ....................................................................15

Rated voltage display window ..............................82, 90, 98, 104, 106, 158
Rating label ................................................................34, 42, 43, 48, 49, 50
Red-free ...................................................................................................68
Regulatory information ...........................................................................250
Remote connection ......................................................82, 90, 98, 104, 106
Requirements for operation ......................................................................16
Risk of burn injuries caused by high illumination intensity .........................20
Risk of collision and crushing ....................................................................84
ROI (region of interest, measuring field) ..................................................127
ROI, display of ........................................................................................127
ROI, position of ..............................................................................127, 128
ROI, shifting of .......................................................................................128
ROI, size of .............................................................................................127

S7 ceiling mount connector panel .....................................................90, 104


S7 central stand ......................................................................................100
S7 floor stand connector panel .................................................................82
Safety checks ..........................................................................................218
Safety measures .........................................................................................7
Selector switch ...................................................................................26, 94
Setting an ergonomic working height .....................................................164
Setting the suspension arm balance ........................................160, 162, 166
SIP label ........................................................................................36, 42, 48
Spare parts .............................................................................................224
Specifications .........................................................................................228
Speed .....................................................................................................121
Speed setting .........................................................................................120

Version 11.0
G-30-1435-en Page 263
Keyword index S7 / OPMI PROergo

Stand base ......................................................................................... 76, 80


Stand column .............................................................................. 76, 80, 86
Standard profile ..................................................................................... 118
Stands/mounts ....................................................................................... 230
Start at .......................................................................................... 120, 121
Start of menu ........................................................................................ 118
Start value, illumination intensity ........................................................... 122
Steerable caster ....................................................................................... 80
Sterilization ............................................................................................ 215
Strain relief device ............................................................... 82, 90, 98, 104
Structure of the menus .......................................................................... 118
Submenu Advanced Settings .................................................................. 129
Support arm .............................................. 76, 80, 86, 88, 92, 96, 100, 102
Surgery menu ........................................................................................ 120
Suspension arm ......................................... 76, 78, 86, 88, 92, 96, 100, 102
Suspension arm, limiting downward travel ............................. 160, 162, 166
Switching to the backup lamp ......................................................... 24, 204
Switching to the spare lamp .................................................................... 23
Symbols and labels on the device ............................................................. 28
System enhancements ........................................................................... 257

Target audience ....................................................................................... 10


Three beeps ........................................................................................... 132
Transport ................................................................................................. 13
Tube and eyepieces for main microscope ................................................. 72
Tubes, technical data ............................................................................. 229

User menu ............................................................................................. 118


User profile ............................................................................................ 118

Ventilation grid ...................................................................................... 110


Video accessories for the 1Chip HD camera option ................................ 247
Video autofocus .................................................................................... 127
Video autofocus, working with .............................................................. 196
Video camera (CCD position), orientation of .......................................... 128
Video camera, external .......................................................... 127, 128, 129
Video camera, integrated ....................................................................... 124
Video capture ........................................................................................ 124
Video Focal Length ................................................................................ 129
Video Gain± .......................................................................................... 124
Video image contrast ............................................................................. 127
Voltage setting ...................................................................................... 146

Version 11.0
Page 264 G-30-1435-en
S7 / OPMI PROergo Keyword index

Wall anchor plate ...................................................................................100


Weight ...................................................................................................229
What to do in the event of malfunctions ................................................197
Widefield eyepieces with magnetic coupling .............................................74

Xenon illumination system ................................................................24, 112

Y/C video port ......................................................................82, 90, 99, 104

Zero position ..............................................................................28, 66, 170


Zoom link .......................................................................................121, 122
Zoom Link function ................................................................................121
Zoom menu ............................................................................................120
Zoom speed ...........................................................................................120
Zoom start value .....................................................................................120

Version 11.0
G-30-1435-en Page 265
Keyword index S7 / OPMI PROergo

Version 11.0
Page 266 G-30-1435-en
S7 / OPMI PROergo

(Blank page for your notes . . .)

Version 11.0
G-30-1435-en Page 267
Carl Zeiss Meditec AG
Goeschwitzer Strasse 51-52 Fax: + 49 (0)7364 - 20 4823
07745 Jena Email: info.meditec@zeiss.com
Germany Internet: www.zeiss.com/med

You might also like