Professional Documents
Culture Documents
An Improved Automated Immunoassay For C-Reactive Protein On The Dimension Clinical Chemistry System
An Improved Automated Immunoassay For C-Reactive Protein On The Dimension Clinical Chemistry System
125–131
C-reactive protein (CRP) is an acute phase reactant that This paper describes a new improved particle-enhance d
has the ability to activate complement after binding to turbidimetric CRP assay on the Dimension1 system that
antigen, and in combination with macrophages to kill o ers the high analytical sensitivity and extended
bacteria and tumour cells [1± 3]. The concentration of upper assay limit required for a broad range of clinical
applications. In addition, the new assay also has the
ability to signal antigen excess up to 2000 mg/l CRP.
The assay principle and performance are detailed in this
* e-mail: weitq@dadebehring.com article.
J ournal of Automated M ethods & M anag ement in Chemistry ISSN 1463± 9246 print/ISSN 1464± 5068 online # 2000 Taylor & Francis Ltd 125
http ://www.tandf.co.uk /journals
T. Q. Wei et al. Improved automated immunoassay for C-reactive protein
126
T. Q. Wei et al. Improved automated immunoassay for C-reactive protein
1000 R3 T a bl e 1. P e r f or ma nce of t he i m pr ov e d CR P as s ay on
D imension1 system–interference from physiological substances.
R1
R2
3
Absorbance Difference x 10
127
T. Q. Wei et al. Improved automated immunoassay for C-reactive protein
T able 2. Comparison of the CRP results of myeloma patients measured with the Binding Site
radial-immunodif fusion assay and the improved CRP method on D imension error 1 system.
M yeloma sample Paraprotein ( g/l) RID result ( mg/l) Dimension1 result ( mg/l)
Each Dimension1 result was measured in duplicate while each RID result was obtained by a
single determination.
IgM values of the two patients are 26.1 and 23.7 g/l,
Like other direct agglutination assays, for a given amount
respectively. The corresponding CRP values reported
of antibody particles, the particle± antigen complex for-
with the RID method were 0.0 and 59.4 mg/l, respect-
mation increases with the amount of CRP to a point
ively, as compared to 7 0.1 and 59.9 mg/l measured by beyond which there is less complex formed. This
the improved CRP assay (table 2) . This observation phenomenon of less complex formation with increasing
indicates the myeloma sera containing elevated IgM amounts of antigen indicates antigen excess and is called
did not interfere with the improved Dimension1 CRP the `hook e ect’ or `prozone e ect’. This assay started to
assay. In addition, sera from myeloma patients contain- show antigen excess between 360 and 400 mg/l (® gure 5) .
ing elevated IgG or IgA were also tested by the improved However, a software routine was incorporated into the
assay and the RID method, no signi® cant interference method parameters to identify and signal the antigen
was found in these studies (table 2) . Icteric, lipemic and excess situation. Samples with CRP concentrations either
haemolysed sera from patients were also tested using both above the assay range (250.0 mg/l) or in antigen excess
methods, no signi® cant di erence in the measured CRP situations trigger an error message (either `assay range’ or
values was detected (data not shown). `antigen excess’, respectively). This allows operators to
retest the sample by dilution. Figure 5 shows the signal
Both within-run and total precisions were excellent
over analyte concentrations spanning the range from 0.0
across the assay range, as summarized in table 3. The
to 2000. 0 mg/l CRP. Any sample between the upper
data were obtained using a Dimension1 system (model
assay limit (250.0 mg/l) and 2000 mg/l was ¯ agged.
X L), but are representatives of the precision observed for
Theoretically, even with samples above 2000 mg/l, the
all of the three instrument models used in this testing method should be able to ¯ ag antigen excess, but it was
(AR, X L and RxL). While precision on individual not tested for all the reagent lots manufactured.
instruments provides important information about the
assay, it does not indicate the total method variability The results of a split-sample method comparison study,
such as might be observed in a multi-site pro® ciency shown in ® gure 3 for the subject assay in comparison to
survey. We thus performed an inter-laborator y compar- the Behring Nephelometric analyser, demonstrated very
ison (ILC) study as described above. The results are good correlation. The patient samples used in this study
reported in table 4. The overall standard deviations, include serum, plasma and disease state specimens, e.g.
which may be taken as a realistic predictor of the varia- icteric, haemolysed and lipemic samples as well as mye-
bility expected in multi-site surveys, indicate very good loma specimens with elevated IgG, IgA and IgM. For
multi-laborator y performance with one reagent and one maximum robustness of the comparisons investigated, we
calibrator lot. used a number of reagent lots and multiple instruments
128
T. Q. Wei et al. Improved automated immunoassay for C-reactive protein
T able 4. Inter-laboratory comparison study for the improved CRP method on the D imension1 system.{
M easured mean Overall SD Overall
Samples ( mg/l) ( mg/l) CV ( %)
{ All samples were measured in ® ve replicates per day for 5 days using 12 Dimension1 instruments.
50 The Bland± Altman form of the di erence plot [27, 28] is
also provided in ® gure 4 to show the measure of agree-
ment between the two methods. It is apparent that there
0 is no obvious relationship between the di erences and
0 50 100 150 200 250 300
measured concentrations. The mean di erence and the
Result by Behring Nephelomet (mg/L) standard error of the mean di erences (SEM) were
Figure 3. Comparison of CRP results as reported on the calculated to be 7 0.8 and 0.265 mg/l, respectively.
D imension1 system with results from the Behring N ephelometer The 95% con® dence interval for the mean di erence
( N Latex CRP M ono) assay. T he data represent duplicate (estimated as the mean § 2 £ SEM) was 7 0.3 and
determinations for 622 clinical specimens. Each set of duplicates 7 1.36 mg/l. Although this con® dence interval does not
was performed on one of three D imension1 systems used for the include 0.0 mg/l, it is at a level comparable to the
complete study ( one in the laboratories of the authors and two in sensitivity of the assay (0.2± 0.5 mg/l), and therefore in-
separate clinical hospital laboratories) . T hree Behring N ephel- dicates that bias between the Behring Nephelometric
ometer analysers, two BN II and one BN A model, were employed method and the improved assay is negligible.
during the study. T wo reag ent lots and two calibrator lots were
used for the Dimension 1 system, and one standard lot with four A direct comparison of serum with plasma results was
reag ent lots were used on the Behring N ephelometer analysers. performed on specimens to which CRP had been added.
Values for the regression line are as follows: r ˆ 0:992; This approach was used to demonstrate the equivalence
n ˆ 622; Sy=x ˆ 7:04 mg/l; slope ˆ 0:993 § 0:004; of the two sample matrices because of the lack of avail-
intercept ˆ ¡0:374 § 0:402 mg /l. able matched draws from patients with adequate CRP
concentrations to span the assay range. This study, which
included the anticoagulant sodium EDTA and lithium
and calibrators in the study, as detailed in the caption of heparin showed the equivalence of the two specimen
® gure 3. The entire study occurred during a 6-month types and serum. The linear regression statistics obtained
period. were sodium EDTA result ˆ 1.01 £ serum result ‡ 0.7
(mg/l, n ˆ 45), and lithium heparin result ˆ 0.98 £ serum
We also tested the same set of data with Passing and result 7 0.3 (mg/l, n ˆ 53). Actual patient plasma speci-
Bablok regression [25, 26]. The advantage of using mens containing CRP, when compared using the
Passing and Bablok regression is the elimination of the Dimension 1 system and the Behring Nephelometer
e ect of extreme points that could be over weighted by analyser gave correlation slopes not statistically di erent
standard linear regression. The regression statistics of from the correlation with serum specimens.
129
T. Q. Wei et al. Improved automated immunoassay for C-reactive protein
50
shows a dilution study performed with a serum sample
diluted to 0.2 mg/l CRP with phosphate-bu ered saline.
+3SD
+2SD The results indicate linearity su cient for the measure-
ment of CRP below the consensus cut-o level of 3± 5 mg/
(mg/L)
0
-2SD l [4, 14].
-3SD
Shelf life and calibration intervals are also important
-50 performance criteria for a clinical assay. FlexTM reagent
cartridges, calibrated and periodically measured over 90
days, showed a maximum rate of change of 5% over a 60-
day period of testing when tested with the upper four
-100
levels of calibrator. The zero-concentration calibrator
0 50 100 150 200 250 300
showed no drift outside the limit of detection (0.5 mg/l).
Average of CRP Results Based on this, a 60-day calibration interval was assigned.
on Dimension system an d Behring Nephelometer In continuing studies extended over 1 year using a 60-
(mg/L) day calibration interval, the overall drifts for all calibra-
Figure 4. D iå erence plot of the data used in gure 3 with the tor levels were less than 5% at each calibrator level, thus
mean diå erence ( bold line) and SD of the mean diå erence ( thin a shelf life of at least 12 months was determined for this
line) . Both two SD and three SD lines are shown in the g raph. method.
SEM represents the standard error of the mean diå erence.
A detailed comparison of the performance characteristics
of the improved assay and the current CRP method on
The accuracy of the method was further evaluated by the Dimension1 system is shown in table 5. Compared to
recovery of the international standard, CRM 470. By the the current commercial CRP assay, the improved
addition technique we found within 100 § 7% recovery method is ® ve± 10 times more sensitive and has ¹2.5
for three FlexTM reagent and calibrator lots. times the assay range. It is also equipped with extra
The limit of detection was determined to be between 0.2 features, e.g. standardization with the IFCC reference
and 0.5 mg/l when de® ned as the concentration corre- CRM 470 [30, 31], antigen excess detection and faster
sponding to two standard deviations above the 0.0 mg/l throughput on certain instrument models (table 5) .
level … n ˆ 20†. This range was determined using four
In conclusion, this new assay o ers more sensitive, precise
FlexTM reagent lots on six Dimension1 RxL instruments
and accurate CRP measurements than most other com-
conducted at three external clinical sites and the author’s
laboratory. Reproducibility studies performed with the mercially available assays can deliver. The advantage s
0.0 mg/l level using NCCLS protocol EP5-T2 showed a make this improved assay suitable for a wide variety of
total SD of less than 0.2 mg/l with multiple instruments clinical applications on this clinical chemistry system.
and reagent lots, and thus supported the results for the The extended upper assay limit also decreases the need
limit of detection. for re-testing of post-diluted samples and provides a faster
turn-around time and lower cost per reportable result
Linearity was assessed by ® tting the data to a quadratic that is coupled with the improved throughput. We
model and by testing signi® cance of the coe cient of the believe the addition of this improved CRP assay en-
second-degree term [29]. This analysis indicated that the hances the utility of the Dimension1 system in laboratory
assay’s linearity extended beyond 260 mg/l. Although settings where workstation consolidation is advantage ous.
T able 5. Performance characteristics of the improved CRP assay as compared to those of the
current CRP method on the D imension1 system.{
Characteristics T he improved CRP assay ( RCRP) CRP assay ( CRP)
{ CRP values shown here are the values anchored to the IFCC standard CRM 470 [27] .
{ Performed using linear regression. Passing and Bablok regression was also tested with the same
data and gave a slope of 0.98 and an intercept of 0.07 mg/l.
} The current commercial Dimension1 CRP method demonstrated resistance to antigen excess
up to 800 mg/l.
130
T. Q. Wei et al. Improved automated immunoassay for C-reactive protein
300
Dimension 1 is a registered trademark of Dade Behring
250
Inc.
Behring Nephelometer1 is a registered trademark of
200 Dade Behring Inc.
150
References
100
1. GAMBINO, R., Clin Chem. 43 (1997), 2017.
2. SHRIVE, A. K., C HEETHAM, G. M. T., HOL DEN, D, M YL ES, D. A. A.,
50
TURNELL, W. G., VOL ANAKIS, J. E., P EP YS, M. B., BL OOMER, A. C.
and G REENHOUGH, T. J., N ature Structural Biology, 3 (1996), 346.
0 3. K IND, C. R. H. and P EP YS, M. B., Int. M ed., 5 (1984), 112.
0 500 1000 1500 2000 4. TIETZ, N. W., T extbook of Clinical Chemistry (Philadelphia, PA : W. B.
Saunders) (1999), 42 ± 72 (specimen collection and processing), 481
mg/L CRP Added (Interim Consensus Reference Intervals for 14 Plasma Proteins in
Figure 5. Representative change of reaction rate with increasing Human serum Referenced to CRM 470/RPPHS), 713.
5. GAMBINO, R., C-reactive protein ± an underutilized test. Lab
CRP concentration demonstrating the antigen excess phenomenon Report for Physicians (1989), 11 :41.
of the improved CRP method. T he antigen excess ag was 6. R IDER, P. M., R IFE, N., P F EIFFER, M. A., SACKS, F. M., M OUE , L.
validated using two reag ent lots on 12 instruments. Samples with A., GOL DMA N, S., FL A KIER, G. C., et al., Circulation , 98 (1998), 839.
CRP values between the upper assay limit and 2000 mg/l triggered 7. L IUZZO, G., BUFFON, A., BIASUCCI, L. M., CA L L IMORE, J. R.,
CA L IGIURI, G., VITELLI, A., A LTA MURA, S., et al., Circulation , 98
either the ‘assay range’ or ‘antigen excess’ ag. T his routine (1998), 2370.
assures that no falsely low CRP values are reported due to antig en 8. R IDKER, P. M., GLYNN, R. J. and H ENNEKENS, C. H., Circulation , 97
excess and allows the operator to dilute and retest the sample. (1998), 2007.
9. R IDKER, P. M., CUSHMA N, M., S TA MPF ER, M. J., TRACY, R. P. and
H ENNEKENS, C. H., Circulation , 97 (1998), 425.
10. M ORROW, D., R IFAI, N., A NTMA N, E. M., WEINER, D. L., M CCA BE,
9 C. H., CA NNON, C. P. and BR AUNWAL D, E., J . Am. Coll. Cardiol., 31
by the Improved Meth od on Dimension
(1998), 1460.
8
11. FEL D, R. D., J . Clin. Lig and Assay, 20 (1997), 313.
7 12. P IETIL A, K. O., H ARMOINEN, A. P., J OKINIITTY, J. and PA STERNACK,
CRP Measured in mg/L
131