Designation: F640 – 07

Standard Test Methods for


Determining Radiopacity for Medical Use1
This standard is issued under the fixed designation F640; the number immediately following the designation indicates the year of original
adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A superscript
epsilon (´) indicates an editorial change since the last revision or reapproval.
This standard has been approved for use by agencies of the Department of Defense.

1. Scope* priate safety and health practices and determine the applica-
1.1 These test methods cover the determination of the bility of regulatory limitations prior to use.
radiopacity of materials and products utilizing X-ray based 2. Referenced Documents
techniques, including fluoroscopy, angiography, CT (computed
tomography) and DEXA, also known as DXA, (dual energy 2.1 ASTM Standards:2
X-ray absorptiometry). The results of these measurements are B209 Specification for Aluminum and Aluminum-Alloy
an indication of the likelihood of locating the product within Sheet and Plate
the human body. D3182 Practice for Rubber—Materials, Equipment, and
1.2 Types of Tests—There are three methods of tests de- Procedures for Mixing Standard Compounds and Prepar-
scribed, differing in the method of determining radiopacity. ing Standard Vulcanized Sheets
1.2.1 Method A—Radiopacity is (1) qualitatively deter- E94 Guide for Radiographic Examination
mined by viewing image(s) of a test sample and the image E1316 Terminology for Nondestructive Examinations
background, with or without the use of a body mimic, or (2) F647 Practice for Evaluating and Specifying Implantable
quantitatively determined as a specific difference in optical Shunt Assemblies for Neurosurgical Application
density or pixel intensity between the image of a test sample 3. Terminology
and the image background, with or without the use of a body
mimic. 3.1 Definitions—For definitions of terms relating to X-ray
1.2.2 Method B—Radiopacity is determined by (1) qualita- procedures, refer to Terminology E1316.
tively comparing image(s) of a test sample and a user-defined 3.2 Descriptions of Terms:
standard without the use of a body mimic, or (2) quantitatively 3.2.1 body mimic, n—a piece of material, a phantom, a
determining the specific difference in optical density or pixel cadaver, or an animal utilized to mimic the appropriate X-ray
intensity between the image of a test sample and the image of attenuation through a particular part of the human body.
a user-defined standard without the use of a body mimic. 3.2.2 digital resolution, n—the number of pixels per inch in
1.2.3 Method C—Radiopacity is determined by (1) qualita- a digital image.
tively comparing image(s) of a test sample and a user-defined 3.2.2.1 Discussion—This may be different in the x and y
standard with the use of body mimic or (2) quantitatively directions
determining the specific difference in optical density or pixel 3.2.3 grayscale range, n—the number of levels in pixel
intensity between the image of a test sample and the image of intensity resolved in the digital image.
a user-defined standard with the use of a body mimic. 3.2.3.1 Discussion—This is normally 256 levels in an 8-bit
1.3 The values stated in SI units are to be regarded as the grayscale image
standard. 3.2.4 optical density, n—the range of values of optical
1.4 This standard does not purport to address all of the density as measured by a densitometer; in this test method the
safety concerns, if any, associated with its use. It is the expected range is 0.50 to 1.50.
responsibility of the user of this standard to establish appro- 3.2.5 optical density difference, n—the difference in optical
density units between two regions or objects in an image,
reported to at least two digits to the right of the decimal point.
1
These test methods are under the jurisdiction of ASTM Committee F04 on
Medical and Surgical Materials and Devices and are the direct responsibility of
2
Subcommittee F04.15 on Material Test Methods. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Current edition approved March 1, 2007. Published March 2007. Originally contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
approved in 1979. Last previous edition approved in 2000 as F640 – 79 (2000). Standards volume information, refer to the standard’s Document Summary page on
DOI: 10.1520/F0640-07. the ASTM website.

*A Summary of Changes section appears at the end of this standard.


Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.

1
F640 – 07
3.2.6 pixel intensity, n—the grayscale level of a pixel 6.4 Rubber Blankets—Blankets incorporating X-ray absorb-
between 0 and 255, as determined by the digital analysis ers may be used to mask areas outside that covered by the body
program. mimic material (this prevents undercutting). Lead sheets may
3.2.7 pixel intensity difference, n—the difference in gray- also be used for masking.
scale level between two regions or objects in an image, 6.5 Back-Scatter Protection, as described in Guide E94, or
reported to within the significance capability of the digital as appropriate with the specific X-ray imaging system.
analysis program. 6.6 Densitometer—The densitometer shall be capable of
3.2.8 user-defined standard, n—a comparison standard se- measuring the optical density over the range of 0.0 to 3.0
lected by the user. optical density units, minimum. It shall have a measuring
3.2.8.1 Discussion—This standard may be an existing medi- accuracy of 60.02 optical density units or better. The densi-
cal product or a material in a particular form, it may be a tometer shall have been calibrated within six months previ-
commercially available standard, or it may be one developed ously by a method and calibration standard traceable to the
by the user. National Institute of Standards and Technology. This is not
required if digital analysis is used.
4. Summary of Test Methods 6.7 Step Tablet, 3 for calibrating densitometers. This is not
4.1 The test specimen is placed so it sits in the middle of the required if digital analysis is used.
X-ray image area in the X-ray imaging system. X-ray images
are made at specified voltages, times, and currents that are 7. Test Specimens
typical of those used in the X-ray diagnosis of humans. 7.1 Material—The material may be in any form. For com-
Preferred settings are those appropriate for the product and for paring results between materials, best results will be obtained
the particular area of body interest (for example, leg, heart, and by utilizing the same form and dimensions for each material.
so forth). The radiopacity of the test specimen is evaluated in 7.2 Product—The product or specific part or section of the
terms of the criteria described for the test method selected. product may be utilized in any desired configuration.
5. Significance and Use NOTE 1—For plastics, a 2.0-mm thick sheet of material is often molded
especially for testing. For example, see the description of sample in
5.1 These methods are intended to determine whether a Practice D3182.
material, product, or part of a product has the degree of
radiopacity desired for its application as a medical device in the 8. Imaging Conditions
human body. 8.1 The test shall be performed at appropriate conditions for
5.2 These methods allow for both qualitative and quantita- the imaging system, the product or material, and the area of the
tive evaluation in different comparative situations. body within which the product is intended for use.
6. Apparatus 8.2 Imaging conditions shall be described in the test report.
6.1 X-Ray Imaging System. 9. Procedure—Method A
6.2 X-Ray Film or Digital Image Acquisition System—The 9.1 Test Specimen Placement—Place the test specimen(s) in
film or digital imaging system must be appropriate for the the middle of the X-ray imaging area. If a body mimic is used,
imaging conditions used. A grid may be used. place the test specimen(s) as appropriate in, on, or under the
6.3 Body Mimic for Methods A and C (not all-inclusive): body mimic and ensure that the test specimen(s) are in the
6.3.1 Aluminum Sheet—A 10.0 6 0.15-mm thick aluminum middle of the X-ray imaging area. If digital analysis will be
sheet may be used. If so specified in the standard for the used, place the calibration standard within the image so that it
medical device, a 15.0 6 0.15-mm thick sheet may be used. does not interfere with the desired imaging of the test speci-
The aluminum sheet shall be $99 % in purity, or type 1100 or men(s) and body mimic.
purer, in accordance with Specification B209. 9.2 X-Ray Exposure—As necessary for the imaging system,
6.3.2 Animal—An appropriate animal, or portion of appro- the product or material, and the area of the body within which
priate animal, with which to perform the tests may be used. the product is intended for use.
6.3.3 Cadaver—A human body, or portion of human body, 9.2.1 If using film and a body mimic, the exposure shall be
with which to perform the tests may be used. of such duration that an optical density of 0.8 to 1.2 is obtained
6.3.4 Calibration Standard—For digital analysis, a standard for the background or through the body mimic, if one is used.
that creates a completely clear area and a completely opaque 9.3 Film Development (If Used)—Develop the X-ray film in
area in the image to allow setting of the full range of 256 accordance with the manufacturer’s instructions. If a digital
grayscale levels. analysis method will be used, convert the developed film
6.3.5 Metal, Plastic, or Composite—A metal, plastic, or
composite material of appropriate dimensions may be used.
6.3.6 Phantom—A device that mimics a portion of the body 3
SRM 1001 has been found suitable for this purpose. The sole source of supply
may be used; this device may be as complex as a manufactured of the apparatus known to the committee at this time is National Institute of
torso with appropriate densities representing all portions of the Standards and Technology (NIST), 100 Bureau Dr., Stop 1070, Gaithersburg, MD
20899-1070, https://srmors.nist.gov. If you are aware of alternative suppliers, please
anatomy within the torso, or may be as simple as a tub of water. provide this information to ASTM International Headquarters. Your comments will
6.3.7 Step Wedge—A step wedge may be used if it has the receive careful consideration at a meeting of the responsible technical committee,1
requisite thickness steps. which you may attend.

2
F640 – 07
image(s) to digital format using an appropriate digital scanning within the image so that it does not interfere with the desired
or photographic method. imaging of the test specimen(s) and body mimic.
9.4 Qualitative Analysis—Visually compare the image(s) of 11.2 X-Ray Exposure—As necessary for the imaging sys-
the test specimen to the background on the film or in the digital tem, the product or material, and the area of the body within
image (whether original or converted from film). which the product is intended for use.
9.5 Quantitative Analysis: 11.3 Film Development—If film is used, develop the X-ray
9.5.1 Measurement of Optical Density (Film): film in accordance with the manufacturer’s instructions. If a
9.5.1.1 Background—Measure the background or body digital analysis method will be used, convert the developed
mimic optical density with a densitometer to determine film image(s) to digital format using an appropriate digital
whether it is within the specified range of 0.8 to 1.2. scanning or photographic method.
9.5.1.2 Test Specimen Image—Measure the optical density 11.4 Qualitative Analysis—Visually compare the image(s)
of the image of the test specimen with a densitometer. of the test specimen(s) to the user defined standard(s) on the
9.5.1.3 Optical Density Difference—Subtract the optical film or in the digital image (whether original or converted from
density of the test specimen from the optical density of the film).
background. If applicable, compare this difference to the 11.5 Quantitative Analysis—Compare the image(s) of the
required contrast specified in the standard for the medical test specimen(s) with the user defined standard(s) in accor-
device. dance with 9.5.
9.5.2 Measurement of Pixel Density (Digital):
9.5.2.1 Set Grayscale Range—Set the fully clear portion of 12. Report
the imaged calibration standard to level 0 and the fully opaque 12.1 The report shall include the following:
portion of the imaged calibration standard to 255. 12.1.1 All test equipment, including source type, filter type,
9.5.2.2 Background—Measure the background pixel inten- detector type, machine geometry, machine type and model
sity. numbers, and film type and resolution (if film is used) or
9.5.2.3 Test Specimen—Measure the pixel intensity of the imaging system resolution (if digital analysis is used).
test specimen image. Pixel intensity may be measured at 12.1.2 All test conditions, including the specific values of
multiple points and averaged or used to determine radiopacity kVp and mA·s, the source-to-detector distance, the object-to-
at different points in the test specimen. detector distance, and, if film is used, the focus-film distance
9.5.2.4 Pixel Intensity Difference—Subtract the pixel inten- and film exposure settings.
sity of the test specimen from the pixel intensity of the 12.1.3 If applicable, specification requirements, as listed in
background. the requirements of the standard specification for the medical
device and the number and title of the medical device specifi-
10. Procedure—Method B cation.
12.1.4 Test sample description, including the manufacturer
10.1 Test Specimen Placement—Place the test specimen(s)
name; the type of device or part; the dimensions, including
in the middle of the X-ray imaging area. If digital analysis will
diameter, wall thickness, and so forth; and the specific mate-
be used, place the calibration standard within the image so that
rial(s), including elements or chemical formula(e), and the type
it does not interfere with the desired imaging of the test
of radiopaque additive (if any) and how it occurs in the part,
specimen(s) and body mimic.
such as a uniform dispersion, line of specific cross section, and
10.2 X-Ray Exposure—As necessary for the imaging sys-
so forth.
tem, the product or material, and the area of the body within
12.1.5 Body mimic description, (1) for a phantom to include
which the product is intended for use.
the manufacturer name and model number, and the type of
10.3 Film Development—If film is used, develop the X-ray
device, for example, “torso” or “knee,” (2) for a cadaver to
film in accordance with the manufacturer’s instructions. If a
include the location and dimensions, and (3) for an animal to
digital analysis method will be used, convert the developed
include the location and dimensions.
film image(s) to digital format using an appropriate digital
12.1.6 User-defined standard description, including materi-
scanning or photographic method.
als and dimensions.
10.4 Qualitative Analysis—Visually compare the image(s) 12.1.7 Test sample placement in comparison to standard
of the test specimen to the user-defined standard on the film or and/or body mimic (that is, on top of, within, underneath, near,
in the digital image (whether original or converted from film). and so forth).
10.5 Quantitative Analysis—Compare the image(s) of the 12.1.8 Optical and X-ray image(s) of the test specimen(s)
test specimen(s) with the user-defined standard(s) in accor- and, if using digital analysis, the calibration standard.
dance with 9.5. 12.1.9 Optical density or pixel intensity readings for all
measurements as described in 9.5, 10.5, and 11.5.
11. Procedure—Method C
11.1 Test Specimen Placement—Place the test specimen(s) 13. Precision and Bias
and the user-defined standard as appropriate in, on, or under the 13.1 The precision and bias of these test methods have not
body mimic and ensure that the test specimen(s) and user- yet been established.
defined standard(s) are in the middle of the X-ray imaging area. 13.2 It is intended to develop data for the overall method
If digital analysis will be used, place the calibration standard and the measurement of density from a round-robin test.

3
F640 – 07
14. Keywords
14.1 implants; radiopacity; surgical applications; surgical
devices; surgical implants; test methods; X-ray

APPENDIX

(Nonmandatory Information)

X1. RATIONALE

X1.1 General—The initial development of this standard any requirements for minimum levels of radiopacity, nor is
was based on the need to determine the location in the body of there a preferred test method (of the three described). The
plastic parts of small diameter. This led to a proposed require- standard specification for the medical devices may specify
ment for an optical density contrast measured on the medical materials and values for some parameters of the test method,
device. However, radiopacity is a property of many types of such as for the body mimic and the X-ray test voltage
medical devices and its required numerical level is influenced
by many variables, some of the principal being: type of X1.3 Background Density—It has been proposed that, for
material; size, thickness, and configuration; part of body or many applications, an optical density contrast of 0.10 is
circulatory system; and X-ray energy applied in the procedure. satisfactory. For larger pieces and thin sections of the human
X1.2 Specification Values—This standard does not have body, a contrast of 0.05 optical density units may be adequate.

SUMMARY OF CHANGES

Committee F04 has identified the location of selected changes to this standard since the last issue (F640 – 79
(2000)) that may impact the use of this standard. (Approved March 1, 2007.)

(1) Changed the title from Radiopacity of Plastics for Medical an X-Ray Imaging System and removed the specific beam
Use to Determining Radiopacity for Medical Use. filtration requirement.
(2) Included all materials and products in any form and (13) Removed Sampling.
included all X-ray based imaging. (14) Added to list of possible body mimics.
(3) Changed “medical device” to “product” throughout. (15) Removed the Comparison Standards section.
(4) Altered Methods A, B and C to allow both qualitative and (16) Changed Conditioning to Imaging Conditions and gener-
quantitative evaluation. alized to allow for different machines.
(5) Added digital imaging. (17) Added conversion of film images to digital format.
(6) Removed intrinsic radiopacity. (18) Moved the Calculation to within the Procedure section for
(7) Changed penetrameter to body mimic throughout. each Method.
(8) Added a user-defined standard. (19) Updated Report requirements.
(9) Removed the definition of optical density. (20) Updated Keywords.
(10) Removed or generalized specific phrases related only to (21) Updated X1 Rationale, changed title from Variables in
plastics. Radiopacity of Plastics in Medical Devices, and moved infor-
(11) Removed the recommendation for a listing of radiopacity mation from X2 Comparison of The Three Test Methods.
density difference in product specifications. (22) Removed X2 Comparison of the Three Test Methods.
(12) Under Apparatus, generalized the X-Ray Machine to be (23) Added Summary of Changes.

4
F640 – 07
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.

This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org).

You might also like