Dräger VentStar® Helix heated (N) plus

Consumables and Accessories

The Dräger VentStar® Helix heated (N) plus breathing circuits have been
specially designed to suit the high requirements in neonatal ventilation.
They provide high performance of ventilating neonates in every working
mode.
02 | Dräger VentStar® Helix heated (N) plus

Benefits

Attuned components with high system performance

To meet this requirement the VentStar Helix heated (N) plus breathing circuits are tested for compatibility with
the Babylog family and ventilators with Neo-option, as well as with BabyFlow plus and other Dräger accessories
for neonates. With the high quality typical of Dräger, these disposable products support to reduce the risk of
cross contamination and offer excellent durability for daily use. Additionally, all gas-carrying components in the
circuits are free of PVC, DEHP and Bisphenol A.

Constant heating performance

The integrated new heating technology with embedded double helix wire supports homogeneous heat
distribution to provide optimum air quality for the patient and was also designed to reduce condensation to
a minimum. Additionally, the smoothbore material and the clear lumen of the circuits help minimise the flow
resistance, support a constant flow and therefore optimise the ventilation performance.

Differentiation to serve individual user needs

Dräger provides two different variants of the new heated circuits for neonates. The VentStar Helix dual heated
(N) plus with heating in both limbs improves the routine work flow because there is no need for a water trap.
For customers who are used to and prefer water trap circuit systems, the VentStar Helix heated (N) plus is the
right choice. With a water trap in the expiratory limb in the customary Dräger design, the user can retain the
usual handling attitude and processes. The design also makes it easy to remove the condensate.

Specialised for High-Frequency ventilation

As an alternative treatment for premature babies and neonatal patients with critical lung diseases, High
Frequency ventilation is being applied more and more frequently. Both heated breathing circuits have been
specially designed with this application in mind. The products both offer an optimised High Frequency
performance, in part due to the low compliance. Through the combined performance together with the
Babylog VN500, Dräger provides a system solution for a powerful and consistent High Frequency Oscillation
delivery. The VentStar Helix heated (N) plus breathing circuits can be used for common ventilation modes and
make an additional circuit change unneccesary as well as saving time through improved workflow.

Solution for every patient

With the two new circuits, Dräger has extended its heated portfolio and can now provide specialised products
for the whole range of patient categories. For neonates with a tidal volume up to 100 mL the VentStar Helix
dual/heated (N) can be used. Additionally, the VentStar Helix heated is the best choice for pediatric and adult
patients with a tidal volume of at least 100 mL.
Dräger VentStar® Helix heated (N) plus | 03

System Components

Soft Connector N

For connecting the VentStar Helix heated (N) breathing circuits to the
Dräger Neo-Nebulizer 8411030.
D-42601-2015

Double connector 11ID/15OD

For connecting the MP02650 and MP02608 to all Dräger ventilators


used with this medical accessory, with the exception of Babylog
devices.
D-7481-2016

Neonatal Flow Sensors

Neonatal flow sensors are based on proven hot-wire anemometry


techniques. Due to extremely fast and accurate measurement results, as
well as extremely low inspiratory and expiratory resistances, it is clearly
superior to other methods.
D-9121-2009
04 | Dräger VentStar® Helix heated (N) plus

Related Products

VentStar® Helix – Heated Breathing Circuits

Breathing gas humidification with heated hose systems in mechanically


ventilated patients is an essential component in today’s intensive care.
Choosing the right hose system can support the therapy success.
D-35855-2011

BlueSet heated (N) – Reusable Breathing Sets

The BlueSets heated were developed for the connection to the


Fisher & Paykel MR 850 humidifier. Thanks to the inspiratory heated
breathing circuit and the expiratory water trap, humidified breathing gas
is conveyed to the patient. On top of that all parts included within the
BlueSets can be autoclaved and reused.
D-101205-2013

Dräger BabyFlow® plus

The superior non-invasive respiratory support system, designed to


maximize performance and patient comfort.
D-48842-2015
Dräger VentStar® Helix heated (N) plus | 05

Technical Data

Dimensions
Connectors Device side, Inner 11 mm
Patient side, Inner 15 mm
Length 1.70 m including 40 cm incubator extension

Material
Breathing hoses EVA, PE, TPE
Connectors EVA, PE
Y-piece PP
Humidifier chamber PP, SBC, silicone, aluminum
Water trap (MP02608) PP, PE, TPE, POM, SAN, steel
All gas-conducting components are PVC-free

Performance characteristics MP02650


Resistance at 2.5/5/15/30 L/min inspiratory <0,1/0,1/0,6/2,0 mbar (or hPa or cmH2O)
Resistance at 2.5/5/15/30 L/min expiratory <0,1/0,2/0,6/2,0 mbar (or hPa or cmH2O)
Compliance at 60 mbar (breathing circuit: inspiration and <0,8 mL/hPa (or mL/mbar)
expiration)
Leakage at 60 mbar/40 mbar <30 mL/min
Leakage at 20 mbar <15 mL/min

Performance characteristics MP02608


Resistance at 2.5/5/15/30 L/min inspiratory <0,1/0,1/0,6/2,0 mbar (or hPa or cmH2O)
Resistance at 2.5/5/15/30 L/min expiratory <0,1/0,2/1,0/3,3 mbar (or hPa or cmH2O)
Compliance at 60 mbar (breathing circuit: inspiration and <1.0 mL/hPa (or mL/mbar)
expiration)
Leakage at 60 mbar/40 mbar <30 mL/min
Leakage at 20 mbar <15 mL/min

Warm-up time 30 min

Humidifier output
Invasive ventilation at 4 to 60 L/min >33 mg/L
Non-invasive ventilation at 4 to 60 L/min >10 mg/L

Ordering Information
Order number
VentStar® Helix dual heated (N) plus MP02650
VentStar® Helix heated (N) plus MP02608

Optional system components


Soft Connector N MP03826
Double connector 11ID/15OD MP02655
Neonatal flow sensors 8410185 / 8411130

Related Products
VentStar® Helix heated MP02606 / MP02607
BabyFlow® Plus MP03700
BlueSet heated (N) 8414987 / 8411041 / 8411153
06 | Dräger VentStar® Helix heated (N) plus

Ordering Information

Further system components and related products can be found in the accessory catalogue
Not all articles are available worldwide
Dräger VentStar® Helix heated (N) plus | 07

Notes
08 | Dräger VentStar® Helix heated (N) plus

Notes

91 024 33| 17.07-1 | HQ | PP | Änderungen vorbehalten | © 2016 Drägerwerk AG & Co. KGaA
Not all products, features, or services are for sale in all countries.
Mentioned Trademarks are only registered in certain countries and not necessarily in the country
in which this material is released. Go to www.draeger.com/trademarks to find the current status.

CORPORATE HEADQUARTERS REGION DACH REGION MIDDLE EAST, AFRICA REGION CENTRAL
Drägerwerk AG & Co. KGaA Drägerwerk AG & Co. KGaA Drägerwerk AG & Co. KGaA AND SOUTH AMERICA
Moislinger Allee 53–55 Moislinger Allee 53–55 Branch Office Dräger Panama S. de R.L.
23558 Lübeck, Germany 23558 Lübeck, Germany P.O. Box 505108 Complejo Business Park,
www.draeger.com Tel +49 451 882 0 Dubai, United Arab Emirates V tower, 10th floor
Fax +49 451 882 2080 Tel +971 4 4294 600 Panama City
info@draeger.com Fax +971 4 4294 699 Tel +507 377-9100
. contactuae@draeger.com Fax +507 377-9130
Manufacturer: REGION EUROPE . contactcsa@draeger.com
Drägerwerk AG & Co. KGaA Drägerwerk AG & Co. KGaA REGION ASIA PACIFIC .
Moislinger Allee 53-55 Moislinger Allee 53–55 Draeger Singapore Pte. Ltd.
23542 Lübeck, Germany 23558 Lübeck, Germany 25 International Business Park
Tel +49 451 882 0 #04-20/21 German Centre
Fax +49 451 882 2080 Singapore 609916
info@draeger.com Tel +65 6308 9400
. Fax +65 6308 9401
asia.pacific@draeger.com
.