WATO EX-65

Anesthesia System

Physical Specifications
Dimensions and Weight
Height 1370 mm
Width 780 mm (not including breathing system)
945 mm (including breathing system)
Depth 690 mm
Weight <145 kg
(without vaporizers and cylinders)
Top Shelf
Weight limit 30 kg
Width 305 mm
Length 545 mm Tslope 0.0 - 2.0 s
Work Surface Apnea I: E 4:1~1:8
Height 850 mm ∆Papnea 3 - 60 cmH2O
Area 1635 cm2 Positive End Expiratory Pressure (PEEP)
Drawer (3Xdrawers, Internal Dimension) Type Integrated, electronic controlled
Height 130 mm Range OFF, 3~30 cmH2O
Width 415 mm Ventilator Performance
Depth 320 mm Driving pressure 280 kPa to 600 kPa
Bag Arm Peak gas flow 120 L/min + Fresh Gas Flow
Height 1150 mm Monitoring Parameters
Length 312 mm Minute volume 0 ~ 100 L/min
Connection ISO 22mm OD, 15mm ID Tidal volume 0~3000 ml
Casters Inspired oxygen (FiO2) 18% ~ 100%
Diameter 125 mm Airway pressure -20 ~ 120 cmH2O
Brakes Center brake system with Lock / Unlock icons I:E 50:1 ~ 1:50
Rate 0 ~120 bpm
Ventilator Specifications PEEP 0 ~ 70 cmH2O
Modes of Ventilation Resistance (R) 0 ~ 600 cmH2O/(L/s)
Manual/Spontaneous Ventilation/Bypass Compliance (C) 0 ~ 300 ml/cmH2O
Volume Control Ventilation (VCV) with PLV function Elasticity (E) 0.003 to 10 hPa/mL(cmH2O/mL)
Pressure Control Ventilation (PCV) with/without volume guarantee (VG) Control Accuracy
Synchronized Intermittent Mandatory Ventilation Volume delivery 5 mL to 60 mL: ±10 mL
(SIMV-Volume Controlled and SIMV-Pressure Controlled) 60 mL to 210 mL: ±15 mL
Pressure Support Ventilation (PS) with apnea backup 210 mL to 1500 mL: ± 7% of the set value
Synchronized Intermittent Mandatory Ventilation Volume Guarantee Pressure Pinsp, Plimit, ∆Psupp, ∆Papnea
(SIMV-VG) ± 2.5 cmH2O or ± 7% of the set value,
Continuous Positive Airway Pressure/Pressure Support Ventilation whichever is greater
(CPAP/PS) PEEP OFF: ± 3.0 cmH2O
Compensation 3 to 30 cmH2O: ± 2.0 cmH2O, or ± 8% of the
Circuit gas leakage compensation and automatic compliance set value, whichever is greater
compensation Rate ± 1bpm or ± 10% of the set value, whichever
Ventilation Parameters Range is greater
Patient Size Adult, Pediatric, Neonate I:E 2:1 to 1:4: ± 10% of the set value
Tidal volume 10~1500 mL (Volume Mode) Other range: ± 25% of the set value
5~1500 mL (Pressure Mode) Tip:Ti ± 8%
Pinsp 5~80 cmH2O Tinsp ± 0.2s
Plimit 10~100 cmH2O Trigger Window ± 10%
∆Psupp 3~60 cmH2O Flow Trigger ± 1L/min
0, 3~60 cmH2O (CPAP/PS) Pressure Trigger ± 2cmH2O
Rate 2~100 bpm Exp% ± 10%
I:E 4:1 - 1:8 Monitoring Accuracy
Inspiratory pause (Tip:Ti) OFF, 5% - 60% Volume monitoring 0 to 60 mL: ± 10 mL
Inspiratory time (Tinsp) 0.2 - 10.0 s 60 to 210 mL: ± 15 mL
Trigger window 5% - 90% 210 to 3000 mL: ± 7% of the real reading
Flow trigger 0.2 ~ 15 L/min Pressure monitoring ± 2.0 cmH2O or ± 4% of the real reading,
Pressure trigger -20~ -1 cmH2O whichever is greater
Expiration termination level 5% - 80% Rate ± 1bpm or ± 5% of the real reading,
Minimum Rate 2 - 60 bpm whichever is greater
I:E 2:1 to 1:4: ± 10% of the reading Vaporizers
Other range: no defined. Vaporizer Mindray V60 Anesthetic Vaporizer or Penlon
MV ± 0.1L/min or ± 8% of the real reading, Sigma Delta Anesthetic Vaporizer
whichever is greater Support agents Halothane, Enflurane, Isoflurane,
O2 concentration ± (2.5% of volume percentage + 2.5% of gas Sevoflurane
concentration) Position MAX.2
Trend Graph Mounting mode Selectatec®, with interlocking function
Continuous trend information with time discrete events for the latest Plug-in®, with interlocking function
48 hours
Trend Table Modules
Continuous trend information together with time discrete events for Anesthesia Gas (AG) Module
the latest 48 hours Monitor gases CO2, N2O, Halothane, Enflurane, Isoflurane,
Alarm Log Book Sevoflurane, Desflurane, MAC, Paramagnetic
500 events storage, first in first out O2 (optional)
Alarm setting Warm-up time 45 s (ISO accuracy mode)
Tidal volume Low: 0 ~ 1595 ml 10min (full accuracy mode)
High: 5 ~ 1600 ml Pump rate Adu/Ped: 150, 180, 200 ml/min
Minute volume Low: 0 ~ 99 L/min Neo: 100, 110, 120 ml/min
High: 0.2 ~ 100 L/min Range CΟ2: 0% ~ 10%
Inspired oxygen Low: 18% ~ 98% Des: 0% ~ 18 %
High: OFF, 20% ~ 100% Sev: 0% ~ 8%
Apnea alarm VTe < 10ml measured in 20s Enf, Iso, Hal: 0% ~ 5%
Paw < (PEEP + 3) cmH2O in 20s O2/N2O: 0% ~ 100%
Airway pressure low 0 ~98 cmH2O Carbon Dioxide (CO2) Modules
Airway pressure high 2 ~100 cmH2O Method Infrared absorption
Sustained airway pressure alarm: 15s Module type Mindray side-stream
Subatmospheric pressure alarm: Paw < -10 cmH2O Capnostat mainstream
Alarm silence countdown timer: 120 to 0 seconds Oridion micro-stream
Lung Recruitment Tool (optional)
Maneuver Multi-Step and One-Step Recruitment Work mode Standby or measurement
One-Step Recruitment Pressure Hold: 20 to 60 cmH2O Displayed numerics EtCO2, FiCO2
Hold Time: 10 to 40s Waveform Capnography
PEEP on Exit: Off, 3 to 30 cmH2O Side-Stream Carbon Dioxide (CO2) Module
Multi-Step Recruitment Increasing PEEP progressively Measurement range 0 ~ 152 mmHg
(with a maximum of 7 stages) Accuracy ±2 mmHg (0 ~ 40 mmHg)
± 5% of the real reading (41 ~ 76 mmHg)
Ventilator Components ± 10% of the real reading (77~152 mmHg)
Flow Sensor Resolution 1 mmHg
Type Variable orifice flow sensor Pump rate Neonatal: 100 mL/min or 120 mL/min
Location Inspiratory and expiratory port Adult/children: 120 mL/min or 150 mL/min
Oxygen Sensor Warming-up time < 1 min, enter the ISO accuracy mode
Type Galvanic fuel cell After 1 min, enters the full accuracy mode
FiO2 displayed 18% to 100% Response time <5 s@100 mL/min
Accuracy ± (volume fraction of 2.5 % +2.5 % gas level) <5 s@120 mL/min
Response Time ≤20 seconds Measured by using neonatal watertrap and
Ventilator Screen 2.5 m neonatal sampling line
Display type Color capacitive touch screen <6.5 s@120 mL/min
Display size 15 inch <6 s@150 mL/min
Pixel format 1024 x 768 Measured by using adult watertrap and
Brightness Adjustable 2.5 m adult sampling line
Screen display configurable Mainstream CO2 Module
Display parameters All setting and alarm parameters (including Measurement range 0 ~ 150 mmHg
Breath rate, I/E ratio, Tidal volume, Minute Accuracy ± 2 mmHg (0 ~ 40 mmHg)
volume, PEEP, MEAN, PEAK, PLAT, and O2 ± 5% of the reading (41 ~ 70 mmHg)
concentration, EtCO2, N2O, Aesthesia gas ± 8% of the reading (71 ~ 100 mmHg)
concentration, BIS) ± 10% of the reading (101 ~ 150 mmHg)
Display waveforms P-T, F-T, V-T, CO2, BIS, O2, Anesthetic gas, N2O Resolution 1 mmHg
Spirometry loops P-V, F-V and F-P Response time <2 s
Timer On screen timer Alarm limit EtCO2 High: OFF, 2 ~ 150 mmHg
Communication Ports EtCO2 Low: OFF, 0 ~ 148 mmHg
One RS-232C connector and one DB9 connector FiCO2 High: OFF, 1 ~ 150 mmHg
Ethernet (RJ-45) Micro-stream CO2 Module
USB Measurement range 0 ~ 99 mmHg
VGA Accuracy 0 ~ 38 mmHg: ± 2 mmHg
39 ~ 99 mmHg: ± (5 % of the reading + 0.08 %
of (the reading minus 38 mmHg))
Sampling rate 50 ml/min O2 Controls
Sampling accuracy -7.5 ml/min ~ + 15 ml/min Method N2O shut off with loss of O2 pressure
Initialization time 30s Supply failure alarm ≤ 220.6 kPa ± 34.2kPa
Response time ≤ 2.9s O2 Flush 25 ~ 75 L/min
Rising time ≤ 190 ms O2-N2O Link system
Alarm range EtCO2 High: OFF, 2 ~ 99 mmHg Type Mechanical
EtCO2 Low: OFF, 0 ~ 97 mmHg Range O2 concentration not lower than 25%
FiCO2 High: OFF, 1 ~ 99 mmHg Auxiliary O2 Flowmeter
BIS Module Range 0 ~ 15 L/min
Measured parameters EEG Indicator Flow tube
BIS/BIS L, BIS R 0 ~ 100 Electronic Flow Meters
Sweep speed 6.25 mm/s,12.5 mm/s, 25 mm/s or 50 mm/s O2 flow range 0 to 15 L/min
Alarm limit BIS high: 2 ~ 100 Air flow range 0 to 15 L/min
BIS low: 0 ~ 98 N2O flow range 0 to 10 L/min
Calculated parameters SQI/SQI L, SQI R; EMG/EMG L, EMG R; SR/SR L, Accuracy between -10% and +10% of the indicated
SR R; SEF/SEF L, SEF R; TP/TP L, TP R; BC/BC L, value (under 20°C and 101.3 kPa, for flow
BC R; sBIS L, sBIS R; sEMG L, sEMG R; ASYM between 10% and 100% of full scale)
Agent Consumption Calculation Optimizer
Calculation range 0 to 3000 ml Only available when AG or CO2 Module is loaded
Accuracy ± 2 mL, or ± 25% of the real reading,
whichever is larger Environmental Specifications
Operating
Electrical Specifications Temperature 10 ~ 40°C
Current Leakage Relative humidity 15% ~ 95% (noncondensing)
100 ~ 240V < 500 μA Barometric (Kpa) 70 ~ 106 kPa
Power and Battery Backup Storage
Power input 220-240 V, 50/60 Hz, 6A Temperature -20 ~ 60°C for main unit,
100-120 V, 50/60 Hz, 7A -20 ~ 50°C for O2 sensor
100-240 V, 50/60 Hz, 7A Relative humidity 10% ~ 95% (noncondensing)
Auxiliary electrical outlets Barometric 50 ~ 106 kPa
Up to 4 outlets (3A for each, total 5A) Electromagnetic Compatibility
Battery backup 90 minutes in case of one battery or Immunity Complies with all requirements of IEC 60601-
240 minutes in case of two batteries (powered 1-2
by new fully-charged batteries with 25°C Emissions Complies with all requirements of IEC 60601-
ambient temperature) 1-2
Battery type Build-in Li-ion battery, 10.95 VDC, 4500 mAh
Safety feature In case of electricity and battery failure, Breathing System Specification
manual ventilation, gas delivery and agent Breathing system volume (Pre-pak)
delivery are possible Automatic ventilation 2850 ml
Manual ventilation 1800 ml
Pneumatic Specifications Breathing system volume (Non Pre-pak)
ACGO (Auxiliary Common Gas Outlet, Integrated) Automatic ventilation 2600 ml
Connector ISO 22 mm OD and 15 mm ID Manual ventilation 1800 ml
Pipeline Supply System Components
Gas type O2, N2O and Air Carbon dioxide absorbent canister
Pipeline input range 280 to 600 kPa Absorbent capacity: 1500 mL
Pipeline connections DISS or NIST Integrated expiratory limb water trap
Pipeline Supply Pressure Gauges Capacity: 6 mL
Display type Mechanical Breathing Circuit Parameters
Ranges 0 to 1000kPa System leakage ≤ 60 mL/min at 3 kPa
Accuracy ± (4% of the full scale reading + 8% of the Compliance ≤4 mL/100Pa (Manual mode)
actual reading) Automatically compensates for compression
Cylinder Supply losses within the breathing circuit in
Cylinder Supply E Cylinder (American style or UK style) mechanical mode
O2 Input Range 6.9 to 20 MPa Expiration resistance < 6.0 cm H2O @60 L/min
N2O Input Range 4.2 to 6 MPa Inspiration resistance < 6.0 cm H2O @60 L/min
Air Input Range 6.9 to 20 MPa System Pressure Gauge
Cylinder Connections Pin-Index Safety System (PISS) Range -20 ~ 100 cmH2O
Yoke Configuration O2, N2O, Air Accuracy ± (2% of the full scale reading + 4% of the
Cylinder Supply Pressure Gauges actual reading)
Display type Mechanical Ports and Connectors
Air Range 0 to 25 MPa Exhalation 22 mm OD / 15 mm ID conical
O2 Range 0 to 25 MPa Inhalation 22 mm OD /15 mm ID conical
N2O Range 0 to 10 MPa Manual bag port 22 mm OD /15 mm ID conical
Accuracy ± (4% of the full scale reading+8% of the Bag-to-Ventilator Switch
actual reading) Type Bi-stable
Control Switch between manual and mechanical Materials
ventilation All materials in contact with exhaled patient gases are autoclavable,
Integrated Adjustable Pressure Limiting (APL) Valve except flow sensors (being not capable of being autoclaved), O2 sensor,
Range SP, 5 ~ 70 cmH2O and mechanical pressure gauge.
Tactile knob indication at above 30 cmH2O All materials in contact with patient gas are latex free.
Accuracy ± 3 cmH2O or ± 15% of the setting value,
which is greater, but is not more than + 10 Suction Device
cmH2O Venturi Suction Regulator
Anesthetic Gas Scavenging System (AGSS) Gas source Air, from system gas source
Size (H x W x D) 430 x 132 x 114 mm Minimum flow 20 L/min
Type of disposal system Maximum vacuum ≥72 kPa at supply gas pressure of 280 kPa;
Active: High-flow or Low-flow ≥73 kPa at supply gas pressure of 600 kPa
Passive Continuous Suction Regulator
Applicable standard ISO 80601-2-13 Supply Negative Pressure Suction
Pump rate 75 ~ 105 L/min (High-flow) Maximum vacuum 517.5 mmHg to 540 mmHg (69 kPa to 72 kPa)
25 ~ 50 L/min (Low-flow) with external vacuum applied of 540 mmHg
and 40 L/min free flow
Pressure relief device: Pressure compensation opening to the air Maximum flow 39 L/min to 40 L/min with external vacuum
State indication of the disposal system: The float falls below the "MIN" applied of 540mmHg and 40 L/min free flow
mark on the sight glass when the disposal system does not work or the Minimum flow 20 L/min
pump rate is lower than 25 L/min (Low-flow) or 75 L/min (high-flow). --------------------------------------------------------------
Connector of the disposal system: ISO 9170-2 Please contact your local Mindray sales representative for the most
current information.

www.mindray.com
P/N: ENG-WATO EX-65 datasheet-210285X4P-20210330