Huang 2007

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

Biosensors and Bioelectronics 22 (2007) 1694–1699

A portable potentiostat for the bilirubin-specific


sensor prepared from molecular imprinting
Chun-Yueh Huang a , Mei-Jywan Syu b,∗ , Yong-Shuen Chang b ,
Chih-Hung Chang c , Tse-Chuan Chou b , Bin-Da Liu d
a Institute of Communication Engineering, National University of Tainan, Taiwan, ROC
b Department of Chemical Engineering, National Cheng Kung University, Tainan, Taiwan 70101, ROC
c Department of Electronic Engineering, Kun Shan University of Technology, Taiwan, ROC
d Department of Electrical Engineering, National Cheng Kung University, Taiwan, ROC

Received 25 May 2006; accepted 28 July 2006


Available online 8 September 2006

Abstract
A portable amperometric potentiostat was designed and implemented in this work. It was developed to acquisit the current signals produced from
bilirubin by an electrochemical sensor. Based on an SOC-based chip, this potentiostat has the merits of moderate accuracy, small size, low cost, and
high portability. The bilirubin electrode was prepared by synthesizing a thin layer of bilirubin imprinted poly(methacrylic acid-co-ethylene glycol
dimethacrylate) onto the Au layer. With the molecularly imprinted polymer (MIP) film, specific detection of bilirubin was successfully achieved.
The cyclic voltammogram of the electrode was measured from this assembled potentiostat. The performance from a commercial potentiostat was
considered rather stable and was used as a reference to examine and evaluate the performance of the assembled potentiostat. The detected current
signals by the bilirubin sensing were obtained. Linear calibration with a sensitivity of 1.344 ± 0.38 ␮A/mg dl was achieved. Our experimental
results showed that the proposed potentiostat’s performance could achieve sufficient performance. The evaluation was also made from the aspects
such as reset time and steady-response time. The self-assembled potentiostat thus demonstrated its ability in precise detection of bilirubin from an
electrode layered with the imprinted polymer film.
© 2006 Elsevier B.V. All rights reserved.

Keywords: Potentiostat; Amperometric; SOC-based chip; Bilirubin; Imprinted poly(methacrylic acid-co-ethylene glycol methacrylate)

1. Introduction al., 1997). A review on the application of molecularly imprinted


polymer to the biomimetic sensors was published (Haupt and
Bilirubin is an end product from hemoglobin metabolism Mosbach, 2000).
(Wang et al., 2006). Total and direct bilirubins are usually mea- Based on molecularly imprinted polymers (MIP), the elec-
sured to screen or to monitor liver and gall bladder disfunctions. trochemical sensors for a variety of applications can be made
The bilirubin test can be used to judge a liver disease or a (Sergeyeva et al., 1999; Piletsky and Turner, 2002). Theo-
blocked bile duct. Owing to the physiological importance of phylline was used as the template to form an MIP film on
bilirubin, it was chosen as the target molecule to be investigated indium tin oxide (ITO) and silicon platforms. The cyclic voltam-
in this work. The electrochemical sensor for bilirubin was pre- mogram of theophylline measured by the MIP–ITO electrode
pared from molecular imprinting in this work. The molecular was compared with those obtained from different electrode and
imprinting technique is an important tool for the measurements blank solution (Kindschy and Alocijia, 2005). Gate effect of
particularly for biosensors (Blanco-López et al., 2004) because thin layer of MIP, imprinted poly(methacrylic acid-co-ethylene
it has the advantages of high selectivity, sensitivity, and low cost. glycol methacrylate), was investigated (Yoshimi et al., 2001).
In 1997, the MIP for sensor technology was discussed (Kriz et The diffusive permeability of the thin MIP layer was sensitive
to the target template molecule. As a result, by the presence of
the template, the electrode performance was enhanced.
∗ Corresponding author. Tel.: +886 6 2757575x62631; fax: +886 6 2344496. Electrochemical biosensors, which can show the detected
E-mail address: syumj@mail.ncku.edu.tw (M.-J. Syu). data by means of potential or current, are commonly applied in

0956-5663/$ – see front matter © 2006 Elsevier B.V. All rights reserved.
doi:10.1016/j.bios.2006.07.036
C.-Y. Huang et al. / Biosensors and Bioelectronics 22 (2007) 1694–1699 1695

DNA identification, protein classification, neural recording, glu- dimethacrylate) thin film. The performance from the proposed
cose determination, or pH variation detection. In 2000, a review potentiostat was investigated from the cyclic voltammogram,
paper describing and discussing the interaction and functional- current signals, calibration sensitivity, reset time and steady-
ization of the electrode surface with the immobilized enzyme response time.
(or protein) layer was proposed (Willner and Katz, 2000). By
immobilizing pyruvate oxidase on a screen-printed electrode, 2. Material and methods
the amperometric biosensor for the analysis of human sali-
vary phosphate was constructed (Kwan et al., 2005). In the 2.1. A SOC-based portable potentiostat
signal processing, each electrochemical sensor needs a poten-
tiostat to maintain the electrochemical stability for the sensor Basically, a potentiostat is an electronic device that controls
and to convert the sensor’s output into an analog signal. So, the the voltage difference between a working electrode and a ref-
potentiostat is an indispensable device for the electrochemical erence electrode. The two electrodes are basic components of
sensors. an electrochemical sensor. By injecting currents into the sen-
Many researchers devoted themselves to develop the single- sor through a counter electrode, the potentiostat controls the
chip potentiostat so that the chip size and cost can be both electrochemical reaction. In other words, the potentiostat has
reduced. Fidler (Fidler et al., 1992) designed a potentiostat based two tasks. Firstly, it measures the potential difference between a
on a voltage-controlled current source for the amperometric working electrode and a reference electrode without polarizing
gas sensors. The circuit maintained electrochemical stability the reference electrode, and compares the potential difference
in the sensor as well as buffered the current output. There with a preset voltage. Secondly, it injects a current flowing from
are many other research reports, for instance, Turner (Turner a counter electrode to a working electrode in order to counter-
et al., 1987) presented a basic CMOS integrated potentiostat, act the difference between the preset voltage and the existing
Kakerow (Kakerow et al., 1995) used a monolithic potentiostat, working electrode potential. After completing the two tasks, the
Bandyopadhyay (Bandyopadhyay et al., 2002) proposed a multi- outputs of the potentiostat are the currents flowing from a counter
channel potentiostat, and Frey (Frey et al., 2003) reported an electrode to a working electrode. The controlled variable in the
integrated potentiostat for biosensor chips. Murari (Murari et al., potentiostat is the sensor’s preset voltage. Such an approach is
2005) also developed an integrated potentiostat for neurotrans- usually used to measure the amperometric sensor’s signals.
mitter sensing. The emphasis from these researchers was merely Fig. 1 shows the circuit diagram of a portable potentio-
focused on the integration of potentiostats and sensors, there is stat. The proposed potentiostat is mainly constructed by a
no further effort on the remote transmission of potentiostats. C8051F020 mixed-signal microprocessor, a current-to-voltage
Nevertheless, in 2005, a miniature implantable in vivo teleme- converter, three inverter amplifier, two low-pass filters, and an
try monitoring system for biosensing was proposed (Beach et RS232 serial data transfer interface. The C8051F020 chip is
al., 2005). In authors’ previous work (Huang et al., 2004a,b), an 8051 based CPU with 12-bit digital-to-analog converters,
a potentiostat designed with portability would be described and 12-bit analog-to-digital converters, and digital peripherals. A
implemented in this work. Additionally, an SOC-based chip was personal computer (PC) is used to control the experimental
further used for the portable potentiostat to improve the system process and collect the experimental data of potentiostat. The
performance as well as to reduce the cost. As a result, the mea- proposed potentiostat design specifications are described as fol-
surement of the proposed potentiostat can be carried out in daily lows: the programmable counter voltage ranges between −3 and
life. 0 V under the resolution 1 mV, the measured reference voltage
Consequently, a potentiostat was designed and assembled ranges between 0 and 1 V under the resolution 1 mV, and the
for the electrochemical detection of bilirubin by the electrode measured current ranges between 0 and 100 ␮A under the reso-
coated with imprinted poly(methacrylic acid-co-ethylene glycol lution 0.1 ␮A.

Fig. 1. The circuit diagram of the portable potentiostat.


1696 C.-Y. Huang et al. / Biosensors and Bioelectronics 22 (2007) 1694–1699

In this portable potentiostat, the control component is real-


ized by a 12-bit ADC (ch2), a 12-bit ADC (ch7), and a 12-bit
DAC in the C8051F020 (SOC-chip). An 8051 microprocessor
is used for signal generation, data acquisition, and experimen-
tal management. The ch2-ADC measures the voltage difference
between a reference electrode and a working electrode and mon-
itors whether the potential of the reference electrode reached
the preset voltage or not. According to the output of ch2-ADC,
the microprocessor generates a corresponding control signal to
DAC. Then, DAC produces a bias voltage to control the injected
current into the counter electrode. This feedback control pro-
cedure continues until the voltage measured by the reference
electrode is close to the preset polarization voltage. The control
algorithm of the amperometric potentiostat is shown as below,
Algorithm (Amperometric potentiostat). Given: the set poten-
tial (or voltage difference) between the working electrode and
the reference electrode (Vset )
Fig. 2. The circuit diagrams of (a) the current-to-voltage converter; (b) the
Step 1:
inverter amplifier.
Initialize System Clock, Port I/O and PCA
Calibrate ADC for Measurement
Start ADC Sampling at 47.85 kHz dropping across the resistor represents a corresponding cur-
Establish the initial counter electrode voltage (VC) rent of the sensor. The circuit diagram of a current-to-voltage
Step 2: While (1) converter is shown in Fig. 2(a). The conversion ratio of the
{
current-to-voltage converter (Franco, 1988) can be expressed
Set up ADC2’s and ADC7’s AMUX, Throughput, Gain
Monitor the potential between the working electrode and the reference as
  
electrode: VWR R3 R2
Monitor the current of the counter electrode: ICW Vo = − 1 + 2 Rf Ii (1)
Verror = Vset − VWR RG R1
If (|Verror | < 1 mV)
Subsequently, an inverter amplifier as shown in Fig. 2(b)
VC (t + 1) = VC (t)
Else is required to optimize the output voltage level of current-to-
sign = sign (Verror ) voltage converter with respect to the maximum resolution of the
case of |Verror | { ADC. Then, the output voltage of the inverter amplifier, propor-
(1 mV < |Verror | ≤3 mV): tional to the electrochemical sensor current, is measured by the
VC (t + 1) = VC (t)–sign * 1 mV, break;
ch7-ADC. A low-pass filter is used to cancel the electronic noise
(3 mV <|Verror | ≤9 mV):
VC (t + 1) = VC (t)–sign * 2 mV, break; from the measured signal.
(9 mV < |Verror | ≤33 mV):
VC (t + 1) = VC (t)–sign * 3 mV, break; 2.2. Chemicals
(33 mV < |Verror |):
VC (t + 1) = VC (t)–sign * 9 mV, break;
}
The Al2 O3 plates were purchased from Laser Tek,
} Inc., Taiwan. They were cleaned with H2 O2 /ammonia,
H2 O2 /concentrated sulfuric acid, and H2 O2 /concentrated HCl
sequentially. The ammonia solution was purchased from Showa,
In the adjustment strategy of the counter voltage, a fuzzy-like
Inc., Japan. Sulfuric acid and HCl were from Union Chemi-
modification method is used to speed up the time for the adjust-
cal Works Ltd, Inc., Taiwan. Methacrylic acid (MAA) used as
ment. The adjustment algorithm can be expressed as follows:
the monomer was obtained from Acros, Inc. Ethylene glycol
dimethacrylate (EGDMA) was from Fluka, Inc. ␣-Bilirubin and
IF the error of VWR is tiny, THEN the adjustment is zero. benzophenone were obtained from Sigma, Inc. and Lancaster,
IF the error of VWR is small, THEN the adjustment is small. Inc., respectively. All chemicals were used as received.
IF the error of VWR is medium, THEN the adjustment is
medium. 2.3. Fabrication of MIP-modified electrode
IF the error of VWR is large, THEN the adjustment is large.
A square surface area of either 0.5 cm × 0.5 cm or
In the above feedback control procedure, a current-to-voltage 1.0 cm × 1.0 cm or 2.0 cm × 2.0 cm sputtered with platinum (Pt)
converter measures the sensor current. The sensor current was formed on the Al2 O3 plate. The Pt electrode was soaked in
between a counter electrode and a working electrode pass 10 mM allyl mercaptan solution for 24 h. After drying, 0.5 ␮l
through a current measurement resistor (10 k). The voltage of 0.15 M benzophenone was dropped onto the allyl mercaptan-
C.-Y. Huang et al. / Biosensors and Bioelectronics 22 (2007) 1694–1699 1697

Fig. 3. Schematic illustration of the fabrication of the MIP film onto the electrode.

coated Pt electrode. The treated electrode was then spinned at confirmed the feasibility of the electrochemical detection on
1000 rpm and dried at 40 ◦ C. Then, 0.5 ␮l of pre-polymerization bilirubin by the MIP modified electrode. The scan rate was set to
mixture comprised of methacrylic acid (MAA), ethylene glycol be 7.5 mV/s. In fact, even without the MIP film, bilirubin can be
methacrylate (EGDMA) and bilirubin in solvent was spin-coated oxidized into other compound. However, the binding specificity
onto the electrode at 1000 rpm. The bezophenone solution and of bilirubin toward this MIP film can be further helpful to reduce
the pre-polymerization solution were both purged with nitrogen interference problem.
flow before the polymerization. The UV light was used for the
irradiation polymerization of the above solution onto the elec- 3.2. The current signals of bilirubin from the MIP modified
trode. The template, bilirubin, was removed from the polymer electrode
layer by placing the electrode in 10 mM NaOH with intensive
stirring and it was repeated 10 times. The fabrication of the MIP The measured results from the proposed potentiostat show
onto the substrate is illustrated in Fig. 3. that the current responses increase with the bilirubin concen-
tration. The detected signals are given in Fig. 5. Very good
2.4. Cyclic voltammetry and amperometric measurements current signals corresponding to the injection of bilirubin solu-
tions were obtained from the potentiostat in A-D mode. The
The cyclic voltammetry measurements were performed in current signal profiles were then used to produce the calibration
10 mg/ml bilirubin at a scan rate of 7.5 mV/s and the scanned curves as shown in Fig. 6(a)–(c). Fig. 6(a) is the calibration curve
voltage was between −0.4 and +0.8 V. The electrodes for cyclic obtained from the MIP modified electrodes with surface areas
voltammetry measurements all had the same surface areas of of 1.0 cm × 1.0 cm. The detection sensitivity from the proposed
0.5 cm × 0.5 cm or 2.0 cm × 2.0 cm. Amperometric measure- potentiostat was 1.344 ± 0.383 ␮A/mg dl with a relative stan-
ments were performed at the potential of 0.55 V with the injec- dard deviation (R.S.D.) of 28.48% while the one obtained from
tion of bilirubin solution at an interval of 5 min. The reference the commercial potentiostat was 1.600 ± 0.194 ␮A/mg dl with a
electrode is the standard calomel electrode (SCE) and the counter R.S.D. of 11.68%. It should be noticed that the deviation might
electrode is the Pt wire electrode. be a combined result from either the potentiostat or the MIP
modified electrode. Evaluation of the detection performance on
3. Results and discussions the assembled potentiostat could not actually be distinguished
from the deviation caused by the irreproducibility of the MIP
With the assembled potentiostat, the calibration of current electrode.
signal with respect to bilirubin concentration was carried out.
The gain of the IC was also tuned to achieve optimal cur-
rent signals. The preliminary test was accomplished before the
experiments. Then, the detection experiments were carried out
using this potentiostat. Basically, there were two modes designed
for this potentiostat, one was in AD (analog-to-digital) mode
and the other was in scan-mode. The potentiostat in AD-mode
was used to measure the current signal of the bilirubin con-
centration and the one in scan-mode was used for the cyclic
voltammetry measurement of the solution. The working elec-
trode used for the current detection was the MIP film coated Pt
electrode.

3.1. Cyclic voltammetry measurement from the scan-mode


potentiostat

The cyclic voltammetry profile of the MIP electrode obtained


from the bilirubin solution by the proposed scan-mode potentio-
stat is shown in Fig. 4. The profile shows a significant oxidation Fig. 4. Cyclic voltammograms measured from the proposed potentiostat in scan-
current around the potential of 0.4 V. Hence, the profile already mode.
1698 C.-Y. Huang et al. / Biosensors and Bioelectronics 22 (2007) 1694–1699

Fig. 5. The detected current signal profiles of the bilirubin concentration from
the proposed potentiostat.

Further examination of the current response from single injec-


tion of bilirubin solution was made. The response signals by the
AD-mode potentiostat were evaluated from the aspects such as
the reset time and steady-response time. Gathered the data from
several batches (not shown here), the times required to reach a
steady state before the measurement were 1,983 ± 126 s (6.34%)
from the proposed potentiostat. Although the proposed potentio-
stat was slightly slower than the commercial potentiostat, they
could reach similar performance. For the steady-response time, it
was 41.33 ± 2.15 s (5.21%) from the proposed potentiostat. Sim-
ilarly, the deviation on the reproducibility of the MIP electrode
also contributed to the standard deviation of the comparison
made between the potentiostats.

3.3. Calibration of bilirubin concentration against current


signal

The calibration curves were further compared to those


obtained from a 2.0 cm × 2.0 cm MIP electrode, the coincidence
among the data was rather high. In Fig. 6(a), the detection was
performed by an MIP electrode of 1.0 cm × 1.0 cm. The detec-
tion area of 1.0 cm × 1.0 cm achieved a detection sensitivity of
1.344 ␮A/mg dl while 5.685 ␮A/mg dl (with a standard devia-
tion (S.D.) of 0.393 and a relative standard deviation (R.S.D.) of
6.91%) for a corresponding area of 2.0 cm × 2.0 cm as shown in
Fig. 6(b). The detection sensitivity was 4.23-fold with respect
to a detection area increment of four-fold. Therefore, within the
experimental range of detection, the sensitivity is almost lin- Fig. 6. Calibration curves of the bilirubin concentrations from the proposed
early proportional to the electrode surface area. In conclusion, potentiostat. (a) With a detection area of 1.0 cm × 1.0 cm; (b) with a detection
the detection precision performance was rather good. area of 2.0 cm × 2.0 cm; (c) with the bilirubin concentration in the range of
To further achieve the limit of detection (LOD), the current 0.1–1.0 mg/dl and a detection area of 1.0 cm × 1.0 cm.
signals were measured at a lower bilirubin concentration range
within 1.0 mg/dl. Fig. 6(c) shows the calibration result. The sur- confirmed. With this LOD, the detection of bilirubin concentra-
face area of the MIP modified electrode was 1.0 cm × 1.0 cm. A tion from the serum of a normal adult can be expected.
calibration parameter of 0.950 ± 0.041 ␮A/mg dl with an R.S.D. In addition, the noises of the detected current signals were
of 4.26% was obtained. Therefore, the LOD at this range was calculated. Therefore, the signal-to-noise ratio can be obtained
C.-Y. Huang et al. / Biosensors and Bioelectronics 22 (2007) 1694–1699 1699

Table 1 Acknowledgments
The signal-to-noise-ratio data calculated from the detection current signals
Bilirubin (mg/dl) N (␮A) S (␮A) S/N The grant supported from ROC Ministry of Education Ex-
1 0.096 3.680 38.140
91-E-FA09-5-4 on this work is appreciated. The discussion of
2 0.151 5.240 34.702 the work with the team members of this project is helpful and
3 0.372 6.186 16.638 highly appreciated.
4 0.328 7.276 22.194
5 0.797 9.040 11.337 References
6 0.141 10.942 77.515
7 0.159 12.287 77.509
Bandyopadhyay, A., Mulliken, G., Cauwenberghs, G., Thakor, N., 2002. Pro-
Average 0.278 7.807 39.719
ceedings of the 2002 IEEE International Symposium on Circuits and Systems
ISCAS 2(2002) May 26–29, pp. 740–743.
Beach, R.D., Conlan, R.W., Godwin, M.C., Moussy, F., 2005. IEEE Trans.
as shown in Table 1. Obviously, judging from the S/N ratio in Instrum. Meas. 54 (1), 61–72.
the table, the detected current signal was significant enough for Blanco-López, M.C., Lobo-Castaňón, M.J., Miranda-Ordieres, A.J., Tuňón-
measuring the bilirubin concentration. Blanco, P., 2004. Trends Anal. Chem. 23 (1), 36–48.
Fidler, J.C., Penrose, W.R., Bobis, J.P., 1992. IEEE Trans. Instrum. Meas. 41
(2), 308–310.
4. Conclusions Franco, S., 1988. Design with Operational Amplifiers and Analog Integrated
Circuits. McGraw Hill.
In this paper, we propose a portable potentiostat and success- Frey, A., Jenkner, M., Schienle, M., Paulus, C., Holzapfl, B., Schindler-Bauer,
fully verify its performance by using the MIP film fabricated Pt P., Hofmann, F., Kuhlmeier, D., Krause, J., Albers, J., Gumbrecht, W.,
Schmitt-Landsiedel, D., Thewes, R., 2003. Proceedings of the International
sensor for the detection of bilirubin. According to the experimen- Symposium on Circuits and Systems, ISCAS ’03, vol. 5, May 25–28, pp.
tal results, the proposed potentiostat’s performance can achieve 9–12.
almost the same accuracy as the commercial potentiostat’s. Haupt, K., Mosbach, K., 2000. Chem. Rev. 100, 2495–2504.
Nice current response corresponding to the change of bilirubin Huang, C.Y., Lin, H.Y., Wang, Y.C., Liao, W.Y., Chou, T.C., 2004a. Proceedings
concentration obtained from the proposed potentiostat in AD- of 2004 Asia-Pacific Conference on Circuits and Systems, Tainan, Taiwan,
December 6–9, pp. 633–636.
mode was achieved. Accordingly, a linear calibration curve was Huang, C.Y., Wang, Y.C., Chen, H.C., Ho, K.C., 2004b. Proceedings of The 2004
obtained. The performance of the proposed potentiostat for the International Conference on Intelligent Sensors, Sensor Networks and Infor-
electrochemical sensing of bilirubin was made from the aspects mation processing, Melbourne, Australia, December 14–17, pp. 331–336.
such as the reset time and steady-response time. The calibration Kakerow, R.G., Kappert, H., Spiegel, E., Manoli, Y., 1995. Transducers 95,
of bilirubin concentration from 0.1 to 1.0 mg/dl is also excellent, Eurosensors IX, vol. 1, pp. 142–145.
Kindschy, L.M., Alocijia, E.C., 2005. Biosens. Bioelectron. 20, 2163–2167.
which is then feasible for the detection of the bilirubin con- Kriz, D., Ramstrom, O., Mosbach, K., 1997. Anal. Chem. 69, 345A–349A.
centration for a normal adult. Although the performance from Kwan, R.C.H., Leung, H.F., Hon, P.Y.T., Cheung, H.C.F., Hirota, K., Renneberg,
the assembled potentiostat may not be as equally good as the R., 2005. Anal. Biochem. 343, 263–267.
commercial one, the ability to achieve precisely amperometric Murari, K., Stanacevic, M., Cauwenberghs, G., Thakor, N.V., 2005. IEEE Eng.
detection of bilirubin by the MIP modified electrode was con- Med. Biol. Mag. 24 (6), 23–29.
Piletsky, S.A., Turner, A.P.F., 2002. Electroanalysis 14 (5), 317–323.
firmed. The cyclic voltammetry measurement from the proposed Sergeyeva, T.A., Piletsky, S.A., Brovko, A.A., Slinchenko, E.A., Sergeeva, L.M.,
scan-mode poentiostat was available as well. Besides, being Panasyuk, T.L., El’skaya, A.V., 1999. Analyst 124, 331–334.
different from the commercial potentiostat, the proposed poten- Turner, R.B.F., Harrison, D.J., Baltes, H.P., 1987. IEEE J. Solid-State Circuits,
tiostat is even suitable for processing the electrochemical signals SC-22, 473–478.
of the amperometric sensor outdoors. Furthermore, the porta- Wang, X., Chowdhury, J.R., Chowdhury, N.R., 2006. Current Paediatr. 16,
70–74.
bility and convenience of the proposed potentiostat are quite Willner, I., Katz, E., 2000. Angrew Chem. Int. Ed. 39, 1180–1218.
obvious. In conclusion, the proposed potentiostat has the merits Yoshimi, Y., Ohdaira, R., Iiyamai, C., Sakai, K., 2001. Sensor Actuator B 73,
of moderate accuracy, small size, low cost, and high portability. 49–53.

You might also like