PC Based Electroencephalogram System
M.E.Chandrasiri, R.M.T.M.Dhanapala, W.G.K.G.Kumari, and R.Ranaweera, Member IEEE

Abstract-In this study a prototype of high performance PC which can be recorded on the surface of scalp as EEG [3].
based Electroencephalogram (EEG) system was developed The frequency content of EEG signals is highly dependent
which integrates a hardware circuit design for initial signal on the activity in cerebral cortex and according to the
conditioning and the digital signal processing design for frequency bands of EEG signals they can be classified in
further signal conditioning and analysis. Since EEG signals
to wave groups as alpha, beta, theta and delta [1].
are very weak and is from a source with high output
impedance, in addition to amplification of the signal and EEG is widely used in different application areas such
filtering, design of a differential pre-amplifier with high as health/clinical research, brain mapping and human
common-mode rejection ratio and high signal-to-noise ratio computer interfacing. It is used as a diagnostic tool in
was very important. In order to minimize the requirements clinical applications for neurological diseases such as
of the notch filters to remove power line interference, Driven epilepsy, Parkinson's disease etc [3]. Another application
Right Leg (DRL) technique was integrated in the analog of EEG is brain computer interfacing (BCI), which creates
design for reducing common mode voltage of the patient direct communication between brain and computer or
body. The instrumentation amplifier with DRL was followed peripherals. This can be used as a method to control
by a high pass filter stage, two amplifier stages and a low pass
prosthesis or devices for patients who have lost the ability
filter stage. The Digital signal processing techniques were
implemented in LabVIEW (National Instruments) with to move or speak [3]. Apart from these applications, EEG
further filtering for power line noise, ability to identify the is also used for research applications such as cognitive
most significant EEG band present in acquired signal, and studies and sleep research [2].
capability for further analysis of the signal. EEG signals are very weak signals as they have very
A high performance single channel EEG system with low voltage levels in micro volt range (approximately
high input impedance, high signal to noise ratio, and high 100µV) [1]. Because of that high gain amplification and
amplification range was achieved for the use under non high sensitivity is required to get a clean signal. Also, EEG
shielded conditions. The performance of individual blocks as signals have different frequency bands in the range of 0.5-
well as the integrated system was verified. Testing results
100Hz [1]. Original source of the EEG has high
showed that the performance of all parts of the EEG system
satisfies the design specifications. The effectiveness of the impedance. Due to the high source impedance, the front
design was verified by comparing to results from Great end of the design should have high input impedance, at
Lakes Neurotechnologies BioRadio 150 system as well as least ten times that of the source impedance [2].
using EEG signals obtained during visual and motor tasks. Due to the very low amplitude of EEG signals, they
are more susceptible to noise interference. Interference can
Index Terms—DRL, EEG, ERP. be present because of the capacitive coupling between the
subject, the power lines and the earth and also due to the
I. I NTR ODUC T ION capacitive coupling between the measurement cables and
the power lines [3][1]. There can be magnetically induced
An electroencephalogram (EEG) is a interference also due to external equipment, which are
neurophysiologic measurement of electrical activity in the picked up by the cable loops and other exposed areas [2].
brain as recorded by electrodes placed on the scalp. These Since the EEG signals are in the low frequency range they
resulting traces are known as EEG signals [3][7]. Neurons can be corrupted by both high and low frequency
in the brain communicate with each other by exchanging interference. Due to differences in polarization between
small electric pulses between them. This is called as firing the electrodes and the skin, there can also be a DC offset
of neurons. When thousands of neurons fire voltage between the input terminals [7].
simultaneously in synchrony, it creates an electric field
II. M E THO DO LO GY
M.E.Chandrasiri is with the Department of Electrical and Electronic
Engineering, Faculty of Engineering, University of Peradeniya, Sri Lanka
(e-mail: malithi1989@gmail.com). 1. System design
R.M.T.M.Dhanapala is with the Department of Electrical and Electronic The acquired weak EEG signal was initially
Engineering, Faculty of Engineering, University of Peradeniya, Sri Lanka conditioned using an analog circuit, where the signal was
(e-mail:thilinimdk@gmail.com).
W.G.K.G.Kumari is with the Department of Electrical and Electronic
amplified and filtered. The conditioned signal was further
Engineering, Faculty of Engineering, University of Peradeniya, Sri Lanka processed and analyzed using a digital system. The signal
(e-mail: wgkalani@gmail.com). is first amplified by a high quality instrumentation
R.Ranaweera is with the Department of Electrical and Electronic amplifier with high common mode rejection ratio
Engineering, Faculty of Engineering, University of Peradeniya, Sri Lanka
(phone: 94-81-239-3401; fax: 94-81-238-5772; e-mail:
(CMRR), which measures the voltage difference between
rdbranaweera@gmail.com). two locations on the scalp. Driven Right Leg circuit was