HP 331A-332A-333A-334A - Technical Data
HP 331A-332A-333A-334A - Technical Data
HP 331A-332A-333A-334A - Technical Data
|
HEWLETT hp,
(hp, PACKARD S31A/332A
ANALYZERS 333-4/334a
. LA rate
See
USES
Measure total broadband distortion
AUTOMATIC
Measure voltage and amplifier gain
Measure noise and hum level
Measure envelope distortion of AM RF carriers
% # ” om Ce eect
HIGH-PASS FILTER
DESCRIPTION In order to reduce the effect of hum components, a high
pass filter is provided which attenuates frequencies below
Distortion Analyzers have gone solid state, offering ex- 400 Hz. The filter may be activated by a front panel
tended tuning range, greater set-level sensitivity, improved switch when measuring distortion of signals greater than
selectivity and greater overall accuracy The Model 331A, 1 kHz in frequency.
332A, 333A, 334A Distortion Analyzers measure total dis-
tortion down to 0.1% full scale at any frequency between
5 Hz to 600 kHz; harmonics are indicated up to 3 MHz.
These instruments measure noise as low as 50 microvolts,
and measure voltages over a wide range of level and fre-
quency. All four models may be used as sensitive wide
range transistorized voltmeters for general purpose voltage
and gain measurements.
NZ
DISTORTION ANALYZER
The Distortion Analyzer consists of a high impedance
broadband amplifier, a tunable frequency rejection circuit,
All
and a metering circuit. The solid state rejection circuit
utilizes a capacitively tuned Wien bridge network which
provides greater than 80 dB of fundamental rejection.
Automatic bridge balancing (fundamental nulling) is
available in the ¢) Models 333A and 334A. Maximum in-
put sensitivity at 0.1% distortion setting corresponds to
300 »V rms for measuring low-level residuals. Input impe-
dance is one megohm for both voltmeter and distortion Switchable High-Pass Filter attenuates all frequencies be-
operation with a single input terminal being used for both low 400 Hz eliminating hum and noise components illus-
modes of operation. trated in the typical curve above.
Model 331A/332A/333A/334A Page 3
Elimination Characteristics:
Fundamental Rejection >80 dB
Second Harmonic Accuracy for a fundamental of:
5 to 20 Hz: better than +1 dB
20 Hz to 20 kHz: better than +0.6 dB
20 kHz to 100 kHz: better than —1 dB
100 kHz to 300 kHz: better than —2 dB
300 kHz to 600 kHz: better than —3 dB
The switchable RF Detector at the input of the instrument
has a frequency range of 550 kHz to 65 MHz. Input con- Distortion Introduced by Instrument:
nector is located on the rear panel of the instrument. <0.03% from 5 Hz to 200 kHz
<0.06% from 200 kHz to 600 kHz
HIGH IMPEDANCE VOLTMETER Meter indication is proportional to the average value of
The transistorized metering circuit of the @ 331A a sine wave.
through 334A, employs feedback to insure stability and Frequency Calibration Accuracy:
a flat frequency response from 5 Hz to 3 MHz. The volt- Better than +2% from 10 Hz to 200 kHz
meter mode offers 13 ranges in 10 dB steps. Range is from Better than —3% from 5 to 10 Hz
300 nV to 300V rms full scale. The bandwidth is 5 Hz to Better than +8% from 200 to 600 kHz
3 MHz for the 1 mV to 30 V ranges, 5 Hz to 500 kHz
for the 300 V ranges, and 20 Hz to 500 kHz for the 300 Input Impedance: Distortion Mode: 1 Megohm shunted
pV range. The average responding meter is calibrated to by less than 60 (*80) pF (10 megohms shunted by
the rms value of a sine wave. <10 pF with HP 10001A divider probe).
Voltmeter Mode: 1 Megohm shunted by 30 (*50) pF
VU OPTION AVAILABLE 1 to 300 V rms; 1 Megohm shunted by 60 (*80) pF,
OPTION: 01 provides an indicating meter having VU 300 nV to 0.3V rms.
ballistic characteristics. *With rear input modification.
Input Level For Distortion Measurements: 0.3V rms _ for
100% set level or 0.245 V for 0 dB Set Level. (up to
300 V may be attenuated to set level reference).
DC Isolation: Signal ground may be +400 Vdc from ex-
ternal chassis.
R.M.S. VOLTS Voltmeter Range: 300 »V to 300 V rms full scale (13
ranges) 10 dB per range.
VU Voltmeter Accuracy: (Using front panel input terminals)
HEWLETT Gp) PACKARD
RANGE +5Y,
300 wv 20 Hz - 500 kHz
VU Meter Scale
| mV -30V 5 Hz -3 MHz
1O0V - 300V 5 Hz - 500 kHz
REAR TERMINAL AVAILABLE
Available on special order — rear terminals in parallel Noise Measurements: Voltmeter residual noise on the
with the front input terminals for direct connection to the 300 pV range: <25 pV rms, when terminated in 600
analyzers at the rear when rack mounted and used in a ohms, <30 pV rms terminated with a shielded 100 k
system. ohm resistor.
Page 4 Model 331A/332A/333A/334A
Output: Approximately 0.1 V rms output for full scale MODEL 333A
meter deflection.
Same as Model 331A except as indicated below:
Output Impedance: 2 kQ.
Power Supply 115 or 230 Volts +10%, 50 to 1000 Hz, Automatic Nulling Mode:
approximately 4 watts. Terminals are provided for exter- Set Level: At least 0.2 V rms.
nal battery supply. Poistive and negative voltages be- Frequency Ranges:
tween 30V and 50V are required. Curren drain from X1, manual null tuned to less than 3% of set level;
each supply is 40 mA. total frequency hold-in +0.5% about true manual null.
X10 through X10 k, manual null tuned to less than
Outline Drawing: 10% of set level; total frequency hold-in +1% about
true manual null.
331A/332A/333A/334A
Automatic Null Accuracy:
| | 5 Hz to 100 Hz: Meter reading within 0 to +3 dB of
NOTE
DIMENSIONS IN INCHES AND (MILLIMETERS)
manual null.
@ E!A RACK HEIGHT(INCLUDING FILLER STRIP)
FOR CABINET HEIGHT (INCLUDING FEET) ADD 34 100 Hz to 600 kHz: Meter reading within 0 to +1.5
2 (7,9) TO EIA RACK HEIGHT jt 33ee dB of manual null.
® REAR APRON RECESS pace
(10,3)
1 kHz.
Frequency Calibration Accuracy:
|Q(4u2,6) —————__y+ mo
Better than +3% from5 Hz to 200 kHz
Better than +8% from 200 kHz to 600 kHz
SIDE R
(132,6) Power Supply: Same as Model 331A except current drain
] = i a from each supply is 80 mA.