Maglumi Components Maintenance

01 Washer
02 Chamber
03 Pipettor
04 Other maintennance

FOCUS MAKES PROFESSION 2


01
PART ONE
Washer

FOCUS MAKES PROFESSION P3


01
PART ONE Washer lift

Clean the floccule in this


area consisting of serum
and microbeads
Clean with WD-40 to
smooth the movement of
the washer lift

Clean the waste needle


with syringer and Clorox.

FOCUS MAKES PROFESSION P4


01
PART ONE Washer Channel

Check the tightness of the


white gear

Clean the crystal in case of stuck cuvette in the channel

FOCUS MAKES PROFESSION P5


01
PART ONE Wash Pump
Check the wahser pump
volumn is sufficient or not.
If not adjust the pump
volumn according to the
following steps.

Steps 2: Rotate the black cylinder to


adjust the spring length. Lengthen the
Step 1: Loose the
spring to increase the pump volumn,
screw inside.
shorten the spring to decrease.
Note: Tighten the
screw after the
adjustment.

FOCUS MAKES PROFESSION P6


01
PART ONE Wash Pump

Check the tightness of value tubing screws.


Crystal on the surface is the signal of loose connection.

FOCUS MAKES PROFESSION P7


01
PART ONE Peristaltic Pump

Reasons of washer peristaltic soaked abnormally:


1.Broken tubes
2.Crystals inside the black cylinder
3.Tightness in this area needs to be adjusted
4.The signal cables of washer needles are loose
5.The waste needle is blocked.

FOCUS MAKES PROFESSION P8


01
PART ONE Peristaltic Pump

The peristaltic tubing should not


touch the wall black cylinder or it may
damage the tubing.

FOCUS MAKES PROFESSION P9


02
PART TWO
Chamber

FOCUS MAKES PROFESSION P10


02
PART TWO
Chamber Cleaning
 Clean the crystals inside the waste needle and waste tube with syringer and
clorox

 Check the connection of waste pump sensor

Clean the lift window

Check the anti-flow tube is blocked or not

FOCUS MAKES PROFESSION P11


02
PART TWO Chamber Cleaning
Clean the crystal inside the chamber

FOCUS MAKES PROFESSION P12


02
PART TWO Chamber Cleaning

Fig Make sure the starter tubes are 1-2 mm out of the starter
injector and stable during the movement of chamber lift. In
case of adjustment needed, adjust the screws marked 1 or/and 2

FOCUS MAKES PROFESSION P13


02
PART TWO Chamber Transport

Make sure every single cuvette hole is in the middle of the two
metal plates. If not, adjust the parameter in service software.

FOCUS MAKES PROFESSION P14


02
PART TWO Chamber Waste Pump

Soak with system tubing cleaning solution

FOCUS MAKES PROFESSION P15


02
PART TWO Chamber Waste Pump

• Check the tightness of PMT signal cable(Both PMT


side and back panel side)

• Check if the signal cable gets too close to the


power cable of washer/chamber/pipetor. Or it may
affect the result.

FOCUS MAKES PROFESSION P16


02
PART TWO Starter pump

Loose this screw to disassemble the while runner

Loose this screw to adjust the starter pump volumn

FOCUS MAKES PROFESSION P17


02
PART TWO Starter pump

Take out the black cap and put some hot water(about 40℃) for
washing

Clean the crystal in the rubber ring

FOCUS MAKES PROFESSION P18


03
PART THREE
Pipettor

FOCUS MAKES PROFESSION P19


03
PART THREE Pipeting Needle

Reconnect all the cables to do with the pipeting arm including adaptor signal
cable, flat cable of Y axis and X axis, make sure all the cables are not loose.

FOCUS MAKES PROFESSION P20


03
PART THREE
Needle washing

Clean the needle inside with syringer and medicinal alcohol

Needle tip should be soaked with medicinal alcohol for 15 min

FOCUS MAKES PROFESSION P21


03
PART THREE
Inject pump

Solutions to inject error:


1. Reconnect the relative cable
2.Clean the inject pump motor and add lube or WD40.

FOCUS MAKES PROFESSION P22


04
PART FOUR
Other Maintenance

FOCUS MAKES PROFESSION P23


04
PART FOUR
Waste Tubing Cleaning

Clean the T-joint waste tubing in this area and the waste pump tubing.

FOCUS MAKES PROFESSION P24


04
PART FOUR Loader

Solutions to loader movement error:


1.reconnect the relative cables
2.clean the sensor and sensor holes of the loader
3.change the guide pulley

FOCUS MAKES PROFESSION P25


04
PART FOUR PCB Cleaning

Use blower to blow away the dust on


PCB.
DO NOT USE Banister Brush!

FOCUS MAKES PROFESSION P26


04
PART FOUR Tank

Wash the waste tank and wash tank with clorox first and
then wash with distilled water in case of bacteria

FOCUS MAKES PROFESSION P27


04
PART FOUR BGW After Maintenance

TEST RESULT CV
BGW1
BGW2
BGW3
BGW4
BGW5
L-le
L-Ri

FOCUS MAKES PROFESSION P28


Thanks