Dear user,
We have made every effort to ensure the accuracy of the contents of this manual.
Should any errors be detected, we would greatly appreciate to receive suggestions
to improve the quality of the contents of this manual.
The above not withstanding, we can assume no responsibility for any errors in this
manual or their eventual consequences.
About this section:
Due to our policy of continuous development, this manual
- xxx xxmay notxx
x xx include
xxx features
x xxx
that have been added to later firmware versions. We reserve rights to make
- xx xx
modifications to this manual without any further notice.
x x xxx x xxxxx x
- xxxx xxx xx xxxx xxx xx x
For more detailed technical data about the MC5 Multifunction Calibrator, please
contact the manufacturer.
- xxx xx x xx xx xxx x xxx
- xx xx x x xxx x xxxxx x
- xxxx xxx xx xxxx xxx xx x
© Copyright 2002
- xxx xx x xx xx xxx x xxx
OY BEAMEX AB
P.O. Box 5 - xx xx x x xxx x xxxxx x
68601 Pietarsaari - xxxx xxx xx xxxx xxx xx x
FINLAND
Tel +358 - 6 - 7840111
Fax +358 - 6 - 7840404
E-mail: sales@beamex.com
Internet: http://www.beamex.com
(8822000) / UEMC5_HART-Appendix / 002182 (003082)
Contents
Contents
General 1
MC5 and HART ............................................................................................... 1
HART Instruments and Beamexs Calibration Database Software ................. 2
Contents
136
General
General
HART (Highway Addressable Remote Transducer) is a digital com-
munication protocol that uses sine waves superimposed on the stan-
dard 4 - 20 mA current. Because the HART signal is small, and its
average value is zero it does not affect the accuracy of the analog
current signal.
HART is a master-slave communication protocol, which means that
during normal operation, each slave (field device) communication
is initiated by a master communication device. Two masters (a pri-
mary master and a secondary master) can connect to a HART loop.
The HART communication protocol is an open standard owned by
the member companies in the HART Communication Foundation
(HCF). For additional information contact:
HART Communication Foundation, 9390 Research Boulevard,
Suite I-350, Austin, Texas, 8759 USA.
Internet: http://www.hartcomm.org
137
Appendix 1, User Guide for MC5 HART Option
External Supply
MC5 measures the 250 ohm
+
externally supplied X
Input signal Supply
output current of the
I etc.
transmitter. The
HART ® terminal is
used for digital com-
munication. Remem-
T/C INT. RJ 2-w xmtr
T/C OR EXT
138
Connecting MC5 and a HART Instrument
Low V
60 VDC/30 VAC
+ +
X + +
Input signal Input signal X
I - I -
T/C OR EXT
T/C OR EXT
WIRES ONLY
WIRES ONLY
3 & 4-w meas V, , V, ,
3 & 4-w meas
HART® HART ®
SENSOR MEASURE & SIMULATE OUTPUT MEASURE
SENSOR MEASURE & SIMULATE OUTPUT MEASURE
ET E
ET E
Max input:
Low V Max input:
Low V
60 VDC/30 VAC
60 VDC/30 VAC
139
Appendix 1, User Guide for MC5 HART Option
HART
Open the settings window by press- Communication
ing settings
D/Menu, 2 Current
C/Others and E: Current Measurement
140
HART and MC5’s Basic Mode
page 138 for information on how to Searching for HART Device at: 12
connect MC5 and a HART instru-
ment. Devices found:
Note.
Measuring the Analog Output (AO) of a HART instrument does not require
any special actions. Just connect the Analog Output to MC5’s current mea-
surement terminals in the E module.
141
Appendix 1, User Guide for MC5 HART Option
Device Information
The adjacent picture displays the
06.10.2000 16:01
data of the Device Info window. HART DEVICE INFORMATION
The following fields are editable: Tag 3051C
Descriptor BEAMEX
Close
142
HART and MC5’s Basic Mode
Device Settings
The adjacent picture displays the 06.10.2000 16:09
data of the Device Settings window. DEVICE SETTINGS
Tag
The following fields are editable: 3051C
Polling Address 0
Tag, Output
Polling Address, PV/Range Unit mbar
PV/Range Unit, Lower Range 0.00
Upper Range 1000.00
Lower Range, Damping (s) 0.40
Upper Range, Transfer Function Linear
Damping, Burst Mode Off
Close
143
Appendix 1, User Guide for MC5 HART Option
For the Digital Output (PV), MC5 opens the Instrument Input Win-
dow for checking/editing the settings. Use B/Next Page Function
Key to check/edit the Instrument Output Window data.
06.10.2000 16:28 19.04.2000 15:29
INSTRUMENT INPUT INSTRUMENT OUTPUT
Pressure Current
Unit mA
Pressure type Gauge
Range 0% 4.000000
Unit mbar 100 % 20.000000
Range 0% 0.000000
100 % 1000.0000
29.05.2002 15:42
Warning! Input Pressure [P1:INT20C]
Trim Lower
gauge
0% Range
Do not alter any settings unless 0.0433 kPa
Trim Upper
you have the knowledge and are Output
100% Range
Current [E: Meas]
authorized to trim the HART in- 4.3362 Check Any
mA
Value
strument. Analog Output (AO) Select
Output
Error Digital Output (PV)
1.00 -0.04 [AO]
% of span
+
Device
Note. Information
0 Device
Some of the menu options may be dis- Settings
abled when trimming certain HART in- End
struments. Refer to the HART - HART
0% 50% Adjustment
100%
instrument’s manual for device depen-
dent information concerning the trim-
ming procedure.
144
HART and MC5’s Basic Mode
146
Calibrating a HART Instrument
calibration procedure.
-
0% 50% 100%
Accept
Stop
Readings
4.3374 mA
View
Calibration
Error
1.00 -0.06 Results
% of span
+
Zero
- Pressure
0% 50% Module 100%
Close
MENU
147
Appendix 1, User Guide for MC5 HART Option
Unit mA
Pressure type Gauge
Range 0% 4.000000
Unit mbar 100 % 20.000000
Range 0% 0.000000
100 % 1000.0000
148
Calibrating a HART Instrument
Note.
Some of the menu options may be disabled when trimming certain HART
instruments. Refer to the HART instrument’s manual for device dependent
information concerning the trimming procedure.
149