288 Full

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 5

Original article

J Clin Pathol: first published as 10.1136/jclinpath-2014-202693 on 19 January 2015. Downloaded from http://jcp.bmj.com/ on June 7, 2023 at Gerstein Science Information Centre Serials
Short-term and long-term storage stability
of heparin plasma ammonia
Lora Dukic, Ana-Maria Simundic

▸ Additional material is ABSTRACT time may sometimes be carried out. However,


published online only. To view Aims Ammonia is an extremely unstable analyte and existing original reports about short-term and long-
please visit the journal online
(http://dx.doi.org/10.1136/
requires special attention during sampling, transport and term ammonia stability are inconsistent. According
jclinpath-2014-202693). storage. The aim of this study was to evaluate the to the manufacturer of our assay (Randox
stability of ammonia in lithium-heparin plasma during Laboratories Ltd, Crumlin, UK), if plasma is sepa-
University Department of
Chemistry, Medical School short-term (at +4°C) and long-term (at −20°C) storage. rated within 30 min, ammonia is stable for 2 h at
University Hospital Sestre Methods Twenty plasma samples were used for short- +2 to +8°C.5 Some authors even recommend
milosrdnice, Zagreb, Croatia term stability assessment. Each sample was divided into long-term plasma storage at −70°C.6
five aliquots and stored in stoppered tubes at +4°C, for The aim of our study was to verify the manufac-
Correspondence to
Lora Dukic, University 1, 2, 3, 4 and 24 h from initial testing. Fifteen plasma turer’s declaration by assessing the stability of
Department of Chemistry, samples were used for long-term stability assessment. ammonia in lithium-heparin plasma. Stability was
Medical School University Each sample was divided into eight aliquots and stored assessed during standard storage conditions in our
Hospital Sestre milosrdnice, in stoppered tubes at −20°C for 3, 24, 48 h and 1, 2, laboratory for short-term (at +4°C) and long-term
Vinogradska c 29, Zagreb
4, 8 and 12 weeks from initial testing. Ammonia (at −20°C) sample storage.
10 000, Croatia;
lora.dukic@gmail.com concentration was determined on a Beckman Coulter
AU2700 chemistry analyser using Randox ammonia
Received 25 September 2014 enzymatic UV method. Bias was calculated from initial
Revised 3 December 2014 MATERIALS AND METHODS
value for each time point and compared with quality From December 2013 until May 2014, routine
Accepted 15 December 2014
specifications defined by Royal College of Pathologists of

Section. Protected by copyright.


Published Online First patient samples for which ammonia testing was
19 January 2015 Australasia. requested were collected. Sampling was performed in
Results The average bias exceeded the total allowable 4.5 mL Lithium Heparin plasma tubes (Greiner
error after storage of samples for 1 h at +4°C and 3 h Bio-One GmbH, Kremsmünster, Austria) in a single
at −20°C. venipuncture without stasis. On sampling, whole
Conclusion Ammonia is not stable during storage at blood was transported on ice immediately to the bio-
+4°C and −20°C in lithium-heparinised plasma and chemistry laboratory. On delivery to the laboratory,
should therefore be analysed immediately. the specimen was centrifuged for 10 min at 1800g in
a benchtop centrifuge (Rotofix 32, Hettich,
Tuttlingen, Germany). Specimens with visible haem-
INTRODUCTION olysis, icteria and lipaemia were not included in the
Ammonia determination is clinically significant in study. Plasma was immediately separated from cells
various pathological states such as in patients with and 300 μL aliquot was processed for ammonia ana-
symptoms of neuromuscular and cerebral distur- lysis on a Beckman Coulter AU2700 chemistry ana-
bances caused by hepatopathy, oncology patients lyser (Beckman Coulter, Tokyo, Japan) using Randox
receiving aggressive chemotherapy and patients ammonia enzymatic UV method (Randox
taking the antiepileptic drug valproic acid.1 2 In neo- Laboratories Ltd, Crumlin, UK). Daily quality
nates and children, ammonia is often requested when control was performed using Ammonia/Ethanol
metabolic disturbance is suspected.2 Guidelines for Control Level 1 and 2 (Randox Laboratories Ltd,
detection and management of hyperammonaemia Crumlin, UK). Within-run precision coefficients of
suggest that levels of ammonia up to 200 μmol/L are variation declared by the manufacturer were 3.6%,
associated with acquired conditions such as sepsis, 2.5% and 2.8% for the ammonia concentrations
chemotherapy or liver dysfunction, while ammonia 57.2, 167.0 and 305.0 μmol/L, respectively. Our own
levels higher than 200 μmol/L indicate metabolic laboratory data for precision on patient samples are:
disorders.3 within-run precision 14.4% and 3.7% at concentra-
Ammonia is an extremely unstable analyte and tions 24.3 and 102.5 μmol/L, respectively.
requires special attention during sampling, trans- Following initial testing, remaining plasma was
port and storage. It has been recommended that immediately divided in several portions of 300 μL,
heparinised or EDTA plasma samples be kept in an according to the protocol described below.
ice bath after sampling, centrifuged immediately on Twenty plasma samples were used for short-term
arrival at the laboratory, aliquoted and analysed stability assessment. Each sample was divided in
within 15–30 min, for accurate analysis of five aliquots and stored in stoppered tubes at +4°C,
ammonia.4 Guidelines for the investigation of for 1, 2, 3, 4 and 24 h from initial testing.
hyperammonaemia even recommend repeated Fifteen plasma samples were used for long-term
ammonia sampling to exclude possible artefactual stability assessment. Each sample was divided in
To cite: Dukic L, cause of high ammonia values.3 Unfortunately, eight aliquots and stored in stoppered tubes at
Simundic A-M. J Clin Pathol ammonia testing is not universally available and −20°C for 3, 24, 48 h and 1, 2, 4, 8 and 12 weeks
2015;68:288–291. delayed analysis due to prolonged sample transport from initial testing. After thawing, plasma
288 Dukic L, et al. J Clin Pathol 2015;68:288–291. doi:10.1136/jclinpath-2014-202693
Original article

J Clin Pathol: first published as 10.1136/jclinpath-2014-202693 on 19 January 2015. Downloaded from http://jcp.bmj.com/ on June 7, 2023 at Gerstein Science Information Centre Serials
36.3 (IQR 28.5–67.8) μmol/L. For plasma samples used for long-
term stability assessment (at −20°C; N=15), the median of initial
concentration was 18.7 (IQR 16.9–25.4) μmol/L.
Figure 1 and table 1 show the median ammonia concentra-
tions and respective bias (absolute and relative) in samples
stored at +4°C during different time points. After 1 h of plasma
storage at +4°C, observed bias had already exceeded RCPA cri-
teria for total allowable error (CI for estimated bias exceeded
the limit of acceptance). There was a gradual increase in
ammonia concentration during the first 4 h of storage, followed
by almost twofold increase of initial concentration of ammonia
after 24 h of storage. Median concentrations between serial
measurements in samples stored at +4°C differed significantly
(p<0.001).
Figure 2 and table 2 contain data for samples stored at −20°C
followed over 12 weeks, at different time points. There was a
Figure 1 The increase of ammonia concentrations during short-term statistically significant difference between median concentrations
storage of lithium-heparin plasma aliquots in stoppered tubes at +4°C of ammonia across different time intervals during long-term
(N=20). storage at −20°C ( p<0.001). The median bias had already
exceeded criteria for total allowable error after 3 h of storage at
−20°C after initial testing.
specimens were visually inspected and centrifuged for 10 min
on 1800g, if necessary. DISCUSSION
All analyses in this study were carried out in triplicate. For Our study shows that measurement of ammonia in aliquoted
this study, only residual patient samples were used. The study and stoppered plasma samples stored for only 1 h in a refriger-
was approved by the Hospital Ethical Committee. ator at +4°C and stored for 3 h at −20°C results in unacceptable

Section. Protected by copyright.


bias. Therefore, neither short-term nor long-term storage of
Statistical analysis heparin plasma at +4°C and −20°C, respectively, is possible for
For statistical analysis, median of triplicate measurements was reliable ammonia determination.
used, as recommended by CLSI.7 For each measurement, bias Stability under various storage conditions has been previously
from initial value was calculated according to the following studied by several authors. Howanitz et al6 have studied the effect
formula: of storage of sodium heparin plasma at +4°C, +22°C, −20°C and
Bias ¼ median at specified time point  median at initial value −70°C. The observed increases of ammonia were much smaller
than in our study. Ammonia increased only 5.6% from the initial
Bias was expressed in absolute values and in percentages. value in plasma stored for 1 h at 4°C (in our study the increase was
Calculations were made with Microsoft Excel 2003 (Microsoft 18%). After 24 h of storage at −20°C the increase observed by
Corporation, Washington, USA). Measurements were expressed Howanitz et al was 18%, whereas we already observed an increase
as median and IQR. Median differences between serial measure- of 39% after 3 h of storage at −20°C. However, their study was
ments were compared by non-parametric Friedman test for carried out on healthy volunteers who have generally lower con-
serial measurements. p Value <0.05 was used as a significance centrations of ammonia, whereas we studied short-term ammonia
level. Statistical analysis was carried out using the MedCalc stat- stability on hospital inpatients with various diseases, whose initial
istical software V.13.1.2.0 (MedCalc Software, Mariakerke, concentrations were substantially higher. Moreover, Howanitz
Belgium). et al studied samples stored in glass tubes and we used plastic
Average bias values were compared with specifications for tubes. It has been reported that differences in blood collection
allowable error from RCPA (Royal College of Pathologists of tube material and tube components may affect sample quality.9
Australasia).8 For ammonia values ≤50 μmol/L, defined allowable De Fonseca-Wollheim studied the rate of increase of ammonia
error was considered ±5 μmol/L and for values ≥50 μmol/L concentration in dipotassium EDTA plasma stored at 0°C, +20°
allowable bias was ±20%. C and +37°C at six different time points between 0 and
90 min. The mean ammonia increase in that study was 1.3-fold
RESULTS at +20°C and 6.5-fold at +37°C in relation to ammonia con-
The median initial concentrations of ammonia of plasma samples centration measured in a sample stored at 0°C, as reference.
used in short-term (at +4°C) stability assessment (N=20) was This study also assessed the stability of ammonia at −38°C.

Table 1 Short-term stability of ammonia in lithium-heparin plasma aliquots during storage for 1, 2, 3, 4 and 24 h at +4°C in stoppered tubes
(N=20)
Initial value 1h 2h 3h 4h 24 h

Ammonia concentration (μmol/L) 36.3 (28.5 to 67.8) 42.5 (37.3 to 72.4) 43.0 (35.4 to 74.4) 49.2 (36.3 to 77.3) 46.8 (33.5 to 78.2) 71.7 (46.4 to 104.8)
median and IQR
Bias (μmol/L) median and 95% CI / 4.6*(3.8 to 7.1) 8.2*(4.1 to 8.9) 8.4*(6.3 to 10.2) 9.3*(7.1 to 13.3) 24.6*(21.1 to 43.6)
Bias (%) / 18 20 25 26 78
*All data points that did not meet RCPA (Royal College of Pathologists of Australasia) criteria.

Dukic L, et al. J Clin Pathol 2015;68:288–291. doi:10.1136/jclinpath-2014-202693 289


Original article

J Clin Pathol: first published as 10.1136/jclinpath-2014-202693 on 19 January 2015. Downloaded from http://jcp.bmj.com/ on June 7, 2023 at Gerstein Science Information Centre Serials
36.2* (22.1 to 41.5)
51.2 (42.3 to 57.5)
12 weeks

220
25.0* (19.7 to 31.8)
39.2 (35.5 to 56.9)
Long-term stability of ammonia in lithium-heparin plasma aliquots during storage for 3, 24 and 48 h and 1, 2, 4, 8 and 12 weeks at −20°C in stoppered tubes (N=15)
8 weeks

171
21.1* (12.1 to 25.6)
39.1 (31.4 to 46.1)
Figure 2 The increase of ammonia concentrations during long-term
storage of lithium-heparin plasma aliquots in stoppered tubes at −20°C

4 weeks
(N=15).

136
14.0* (12.3 to 21.8)
37.3 (27.7 to 44.1)
Interestingly, the authors did not observe a significant increase
of ammonia after 24 h of storage at −38°C.10 The stability at
−20°C was, unfortunately, not studied in their work.

2 weeks
Stability of ammonia was also studied in equine EDTA

104
plasma. The increase of ammonia was significant after 24 h of

Section. Protected by copyright.


storage at +4°C, whereas ammonia was stable for 21 days if
stored at −20°C.11 It has to be pointed out that equine and

8.8* (5.7 to 13.2)


25.4 (17.6 to 25.5)
human matrices have differences, which may explain observed
differences in stability of ammonia in the specimens. Moreover,

1 week
the study assessed the stability of ammonia in EDTA plasma,
whereas in our study heparin plasma was used. This might at

59
least partially explain the observed differences.

4.4* (2.6 to 11.3)


22.3 (19.8 to 29.7)
There are several reasons for such high in vitro instability of
ammonia in human plasma. Ammonia concentration in hepari-
nised plasma is pH-dependent.12 Substantial changes in pCO2
caused by inappropriate sampling and storage conditions cause
48 h

changes in the rate of ammonia formation in heparinised blood


46
specimens.13 When laboratory glassware was used to test
19.5 (15.4 to 29.1)

ammonia, great difficulty was encountered owing to contamin-


4.2*(2.6 to 6.4)

*All data points that did not meet RCPA (Royal College of Pathologists of Australasia) criteria.
ation by ammonia-containing detergents. Lots of issues related
to specimen collection and determination of ammonia were
resolved by introduction of more sophisticated specimen collec-
24 h

25

tion devices and disposables, however, special precautionary


measures should still be respected. Exposure to cigarette smoke
23.4 (20.6 to 27.2)

5.6* (2.3 to 7.0)

pollution causes false elevated ammonia values, because


ammonia-forming compounds are added routinely to tobacco in
order to enhance flavour and maximise the quantity of free
nicotine in the smoke.14 Moreover, ammonia is continuously
3h

39

generated in the dipotassium EDTA plasma sample as a result of


the production of ammonia caused by γ-glutamyltransferase
18.7 (16.9 to 25.4)

hydrolysis of glutamine.15
Information on adequate short-term and long-term storage
Initial value

conditions for accurate ammonia measurement is important for


routine everyday work as well as in research settings. The prea-
/

nalytical phase is recognised as the most vulnerable part of the


total testing process and it is therefore extremely important to
Bias (μmol/L) median and
(μmol/L) median and IQR
Ammonia concentration

recognise and properly manage all potential sources of variabil-


ity of laboratory test results.16 17
As already pointed out, accurate information on short-term
storage stability is necessary for laboratories that receive plasma
Table 2

Bias (%)

samples from distant laboratories where ammonia testing is not


95% CI

available. Moreover, according to ISO standard 15 189, each


laboratory is obliged to define storage protocols for different
290 Dukic L, et al. J Clin Pathol 2015;68:288–291. doi:10.1136/jclinpath-2014-202693
Original article

J Clin Pathol: first published as 10.1136/jclinpath-2014-202693 on 19 January 2015. Downloaded from http://jcp.bmj.com/ on June 7, 2023 at Gerstein Science Information Centre Serials
analytes.18 Long-term ammonia sample stability is also import- Handling editor Tahir S Pillay
ant in the research setting, where samples need to be stored for Contributors LD as first author and A-MS as coauthor have contributed in data
longer periods of time. collection and analysis, design and writing of the manuscript.
Hyperammonaemia is a severe clinical condition owing to Funding This study was supported by the Ministry of Science, Education and
central nervous system abnormalities such as seizures, drowsi- Sports, Republic of Croatia, project number: 134-1340227-0200.
ness and apnoea. In children and neonates, hyperammonaemia Competing interests None.
(ammonia concentration >100 and 150 mmol/L, respectively) is Ethics approval Medical School University Hospital Sestre milosrdnice Ethical
highly suggestive of potential inherited deficiencies of urea Committee.
cycle enzymes.19 Differential diagnosis and timely management Provenance and peer review Not commissioned; externally peer reviewed.
of patients presenting with hyperammonaemia are critical for
patient outcome.20 A false increase of ammonia may mask
other potentially dangerous pathological processes in the REFERENCES
central nervous system and lead to misdiagnosis, or, alterna- 1 Newman DJ, Price CP. Nonprotein nitrogen metabolites. In: Burtis CA, Ashwood ER,
tively, lead to additional diagnostic workup to investigate the eds. Tietz fundamentals of clinical chemistry. 5th edn. Philadelphia, PA: W.B.
cause of the hyperammonaemia and cause delayed diagnosis. Saunders Company, 2001:417–19.
The frequency of false positive findings of ammonia due to 2 Thomas L, ed. Ammonia. In: Clinical laboratory diagnostics: use and assessment of
clinical laboratory results. 1st edn. Frankfurt: TH-Books Verlagsgesellschaft,
preanalytical errors has been estimated to account for almost 1998:186–92.
48% of all cases of hyperammonaemia in children.21 It is 3 UK National Metabolic Biochemistry Network. Guidelines for the investigation of
therefore absolutely necessary not to delay the analysis of hyperammonemia. http://www.metbio.net/docs/MetBio-Guideline_
ammonia for longer than 15–30 min. All deviations from the AMUP100834-21-07-2010.pdf
4 Nikolac N, Omazic J, Simundic AM. The evidence based practice for optimal sample
recommended procedure might lead to false increases of quality for ammonia measurement. Clin Biochem 2014;47:991–5.
ammonia concentration, cause diagnostic errors and jeopardise 5 Randox Ammonia (NH3) Enzymatic UV-Method. Package insert. Randox
patient safety. Laboratories Ltd. REV 001
Experimental data on short-term and long-term storage of 6 Howanitz JH, Howanitz PJ, Skrodzki CA, et al. Influences of specimen
plasma ammonia differ substantially in their design and method- processing and storage conditions on results for plasma ammonia. Clin Chem
1984;30:906–8.
ology. Various authors have studied different time points at dif- 7 CLSI. EP07-A2—Interference Testing in Clinical Chemistry; approved guideline. 2nd

Section. Protected by copyright.


ferent temperatures. We have explored stability of ammonia in edn. Wayne, PA: Clinical Laboratory Standards Institute, 2005.
heparin plasma stored at conditions most commonly used in 8 Royal College of Pathologists of Australasia Allowable Limits of Performance.
routine laboratories (+4°C and −20°C). One possible limitation http://www.westgard.com/rcpa-australasian-quality-requirements.htm#alcohol (accessed
1 Jun 2014).
of our study is that ammonia concentrations in wide ranges 9 Bowen RA, Remaley AT. Interferences from blood collection tube components on
were not included. We believe that our results may help labora- clinical chemistry assays. Biochem Med (Zagreb) 2014;24:31–44.
tory professionals and researchers to define their optimal 10 da Fonseca-Wollheim F. Preanalytical increase of ammonia in blood specimens from
storage protocols during sample transport as well as provide healthy subjects. Clin Chem 1990;36:1483–7.
evidence-based acceptance criteria. 11 Lindner A, Bauer S. Effect of temperature, duration of storage and sampling
procedure on ammonia concentration in equine blood plasma. Eur J Clin Chem Clin
In conclusion, although manufacturers may contend that Biochem 1993;31:473–6.
ammonia is stable for 2 h at +2 to +8°C if plasma is separated 12 da Fonseca-Wollheim F. The influence of pH and various anions on the distribution
within 30 min, we were not able to confirm this in our study. of NH+4 in human blood. Eur J Clin Chem Clin Biochem 1995;33:289–94.
Therefore, manufacturers’ claims on heparinised plasma storage 13 da Fonseca-Wollheim F, Heinze KG. The influence of pCO2 on the rate of ammonia
formation in blood. Eur J Clin Chem Clin Biochem 1992;30:867–9.
stability need revision. Our results indicate that ammonia 14 Seeman JI, Carchman RA. The possible role of ammonia toxicity on the exposure,
cannot be reliably measured if stored in lithium-heparinised deposition, retention, and the bioavailability of nicotine during smoking. Food Chem
plasma tubes for 1 h at +4°C and for 3 h at −20°C. Toxicol 2008;46:1863–81.
15 da Fonseca-Wollheim F. Deamidation of glutamine by increased plasma
γ-glutamyltransferase is a source of rapid ammonia formation in blood and plasma
specimens. Clin Chem 1990;36:1479–82.
Take home messages 16 Simundic AM. Preanalytical phase—an updated review of the current evidence.
Biochem Med (Zagreb) 2014;24:6.
17 Lippi G, Becan-McBride K, Behúlová D, et al. Preanalytical quality improvement:
▸ Ammonia is an extremely unstable analyte and requires in quality we trust. Clin Chem Lab Med 2013;51:229–41.
special attention during sampling, transport and storage. 18 Fuentes-Arderiu X. Clinical sample stability and measurement uncertainty. Clin
Chem Lab Med 2014;52:e37–8.
▸ We have assessed the stability of ammonia in
19 National organization for rare disorders (NORD). The physician guide to urea cycle
lithium-heparin plasma during short-term (at +4°C) and disorders. Danbury, 2012.
long-term (at −20°C) storage. 20 Auron A, Brophy PD. Hyperammonemia in review: pathophysiology, diagnosis, and
▸ Lithium-heparin plasma ammonia is not stable when stored treatment. Pediatr Nephrol 2012;27:207–22.
for 1 h at +4°C and for 3 h at −20°C. 21 Maranda B, Cousineau J, Allard P, et al. False positives in plasma ammonia measurement
and their clinical impact in a pediatric population. Clin Biochem 2007;40:531–5.

Dukic L, et al. J Clin Pathol 2015;68:288–291. doi:10.1136/jclinpath-2014-202693 291


Sažetak

Cilj: Amonijak je izrazito nestabilan analit koji zahtjeva posebnu pažnju tijekom uzorkovanja,
transporta i pohrane. Cilj ovog istraživanja bio je utvrditi stabilnost amonijaka u uzorku litij-
heparinizirane plazme tijekom kratkoročne (+4°C) i dugoročne pohrane (-20°C).
Metode: Za procjenu kratkoročne stabilnosti koristili smo plazmu od dvadeset ispitanika.
Svaki uzorak je razdijeljen u 5 alikvota i pohranjen u začepljene epruvete na +4°C na 1, 2, 3, 4
i 24 sata od početne analize. Za procjenu dugoročne stabilnosti korišteno je petnaest
uzoraka plazme. Svaki uzorak je razdijeljen u 8 alikvota koji su pohranjeni u začepljene
epruvete na -20°C na 3, 24 i 48 sati i 2, 4, 8 i 12 tjedana od početne analize. Koncentraciju
amonijaka smo odredili na biokemijskom analizatoru Beckman Coulter AU 2700 koristeći
Randox Ammonia enzimatsku UV metodu. Odstupanje od inicijalne koncentracije u svakoj
vremenskoj točki je uspoređeno s unaprijed definiranim kriterijima prihvatljivosti prema
Royal College of Pathologists of Australasia.
Rezultati: Prosječno odstupanje bilo je veće od kriterija prihvatljivosti prilikom pohrane
uzoraka tijekom sat vremena na +4°C i 3 sata na -20°C.
Zaključci: Amonijak nije stabilan tijekom pohrane na +4°C i -20°C u litij-hepariniziranoj plazmi
i stoga ga treba odmah analizirati.

Disclaimer: This abstract has been translated and adapted from the original English-language
content. Translated content is provided on an "as is" basis. Translation accuracy or reliability
is not guaranteed or implied. BMJ is not responsible for any errors and omissions arising
from translation to the fullest extent permitted by law, BMJ shall not incur any liability,
including without limitation, liability for damages, arising from the translated text.

You might also like