Download as pdf or txt
Download as pdf or txt
You are on page 1of 3

Agilent ChemStation Software Modules

G2070BA/G2071BA/G2072BA/G2073BA/G2170BA/G2171BA/G2171BA/G2180BA/G2090BA/
G2710BA/G1601BA/G2201BA

B.02.01 Upgrade Site Prep Checklist

Purpose of Procedure Procedure Checklist: ‰ Available LAN interface with


TCP/IP protocol installed and a
To assure that the upgrade of LAN card (J4100A JetDirect card,
Software Requirements
Agilent instruments and systems can G1369A LAN card or equivalent)
be completed successfully by careful ‰ Windows 2000 Professional SP4 or
for the instrument, if needed.
preparation and evaluation of the Windows XP Professional SP2 *
installation site and by ensuring the ‰ Administrator logon required to the
‰ A printer supported by the current
availability of appropriate utilities, ChemStation PC
version of the ChemStation. For
consumables and supplies. ‰ TEMP variable points to an existing
further information please contact
directory (e.g. TEMP=C:\TEMP)
your local service representative or
‰ Check if a printer driver is installed
Customer Responsibilities the ChemStation installation manual
on the system
of your appropriate technique.
Customers should ensure that all ‰ Verify that the regional settings are
necessary operating supplies, set to English-US
‰ The voltage setting of the computer
consumables and usage dependent
system and the power cables has to
items such as columns, vials, PC Hardware Requirements be correct.
syringes and solvents required for
the successful installation of ‰ All PC hardware needs to be listed
instruments and systems are respectively in the Windows 2000 or
available. Installation sites should Windows XP Professional Hardware *Detailed information on how to set up
be prepared in accordance with the Compatibility List Windows XP Professional for optimal
following specifications. (see www.microsoft.com ) ChemStation usage is located on the
ChemStation CD-ROM under the
Important Information ‰ Microsoft Windows compatible manuals\installation\Configure and
pointing device Maintain your Agilent PC directory.
If you have problems in providing
any of the following, please contact *Please verify if your instruments
‰ Pentium IV processor operating at
your local Agilent office for support GPIB connection.
1.5 MHz or higher for LC, LC/MS
assistance. Assistance with user
and GC systems
specific applications may be
provided but should be contracted
separately. Users of the instrument ‰ At least 512 MB of RAM for
should be present throughout the - 2D instrument configuration
installation and familiarization - 3D instrument configuration
otherwise important operational, - LC/MS instrument configuration
maintenance and safety information - instrument configuration with
may be missed. ChemStore C/S

‰ CD-ROM / DVD-ROM drive


Additional Information
For detailed steps to perform an ‰ Hard-disk with at least 40 GB free
upgrade installation refer to the capacity
“Upgrade Preparation Guide for
Rev. B.02.01 ChemStation” (P/N ‰ 1280x1024 recommended resolution
G2170-90226). or better graphics adapter and
monitor, True Colors

‰ 82350A/B GP-IB card available


with one free IRQ on the PC and
one free PCI slot **.

Issued: May 2006– Rev 1.0 Copyright © 2006 Agilent Technologies PHS/CAS/IBS Product Support Page 1 of 3
Agilent ChemStation Software Modules
G2070BA/G2071BA/G2072BA/G2073BA/G2170BA/G2171BA/G2171BA/G2180BA/G2090BA/
G2710BA/G1601BA/G2201BA

B.02.01 Upgrade Site Prep Checklist

Instrument Hardware ‰ It is recommended to upgrade to the


Requirements LC, GC and minimum firmware listed in the
LC/MS For CE systems only: table using the GC Firmware Update
Utility:
‰ Check if Firmware revisions of If the firmware is not current, ask the
Instrument Minimum Firmware
instruments and modules are customer to upgrade to the minimum
Revision
appropriate for ChemStation firmware:
B.02.01. 6890N GC Firmware >= N.05.05
Mainframe: 2.3
System LAN assembly
DAD: 1.2
04.7B3
For LC systems only:
6890Plus, Firmware >= A.03.08
6890A
‰ If Firmware is not current, ask For GC systems only:
customer to upgrade to minimum 6850 GC Firmware >= A.05.04
GC Column Catalog: A GC Column serial# >= LAN assembly
Firmware A.06.02/B.01.02 required
Database Utility is available to transfer US10243001 04.7B3
with Rev. B.02.01 ChemStation.
user-defined GC columns from 6850 GC Firmware >= A.03.03
Please use the record outlined in the
G2070AA to G2070BA ChemStation. serial# <=
“1100/1200 IQ Attachment Form”.
This document is available from the US00003200
User-defined column entries in the GC Agilent 6850 Firmware >= A.05.04
EPI Warehouse @
Column catalog are not transferred Handheld
http://whadmin.cos.agilent.com/sear
automatically during an upgrade from Controller
ch/AdvPubForm.asp by searching
G2070AA to G2070BA ChemStation. (G2629A)
for “1100/1200 PM Checklist”. The
The GC Column Database Utility can be 35900E Firmware >= E.01.02
1100/1200 IQ Attachment Form
used after the upgrade to transfer the 5890 Series II Firmware >= A.03.02
AND the 1100/1200 PM Checklist
user-defined GC Columns. 4890D Firmware >= A.06.00
will be retrieved. The 1100/1200 IQ
Attachment Form is to be used only ‰ The GC Firmware Update Utility is
The Utility is available on the
once. provided on the ChemStation
ChemStation CD-ROM under the
Support directory. B.02.0x CD-ROM under the
‰ The FW upgrade process is Support directory. Please check the
described in the manual coming Agilent Technologies website for
with the FW upgrade tool, placed on ‰ If the firmware is not current, ask the latest firmware:
the ChemStation CD-ROM. the customer to upgrade to the http://www.chem.agilent.com/script
minimum firmware. For previously s/cag_firmware.asp
For LC/MS systems only: qualified systems, please use the GC
IQ Attachment form to record the
‰ Update the LC/MSD firmware by change. This document is available
running the executable from the EPI Warehouse @
c:\Chem32\MS\FIRMWARE\ http://whadmin.cos.agilent.com/sear
msupdate.exe while connected to the ch/AdvPubForm.asp by searching
LC/MSD. for “GC IQ Attachment”.

‰ Before upgrading the LC/MSD


ChemStation to revision B.02.01,
perform a dual polarity Autotune to
verify instrument tune performance
and to generate a record of current
tune parameters.

Issued: May 2006– Rev 1.0 Copyright © 2006 Agilent Technologies PHS/CAS/IBS Product Support Page 2 of 3
Agilent ChemStation Software Modules
G2070BA/G2071BA/G2072BA/G2073BA/G2170BA/G2171BA/G2171BA/G2180BA/G2090BA/
G2710BA/G1601BA/G2201BA

B.02.01 Upgrade Site Prep Checklist

ChemStation Solutions
For GC systems only: Additional Familiarization Upgrades
Check for installed Add-On Solutions.
from Rev. A to Rev. B ChemStation
‰ Companion
Before installing the Add-On solution,
‰ Explain the modified ChemStation
verify the correct revision of the Add-On ‰ Retention Time Locking
Solution Products for ChemStation structure (e.g. ChemStation.ini)
B.02.01 using the using the link to the ‰ HeadSpace
LSM Marketing Homepage for the ‰ Outline the new integrator features
compatibility matrix/Product Support ‰ Other: _________________
Plan: ‰ Outline the integrator changes and
http://lsbu.marketing.agilent.com/article/ inform about the available
ashow.asp?article_id=710 documentation *
ChemStation System Customization ‰ Outline the change for the “old”
Rev. A/Rev. B. Upgrades: Check for customized solutions:
All Add-On Products need to be spectra tool and inform about the
uninstalled BEFORE upgrading to the available documentation*.
‰ Check if “User.mac” is present in
B.02.01 ChemStation: xx:/hpchem/core directory or *Detailed information regarding the
xx:/chem32/core directory integration/spectra update, macro
General:
changes etc. can be found in the
‰ Ask customer if any customization “Upgrade Preparation Guide for
‰ ChemStore has been done and back up all Rev. B.02.01 ChemStation, Pub
customized macro files. number G2170-90226”.
‰ ChemAccess
‰ For Rev. A Upgrades only:
‰ ChemStation Plus Security Pack Regarding impact on customized
macros due to the upgraded
‰ Other: _________________ ChemStation structure point
customer to the available
documentation *.
For LC, LC/MS systems only:
‰ Purify Software

‰ Active Splitter

‰ G1979A Multi-Signal Output


Accessory (Side Box)

‰ Analyst

‰ Easy-Access

‰ Data Browser

‰ Other: _________________

Issued: May 2006– Rev 1.0 Copyright © 2006 Agilent Technologies PHS/CAS/IBS Product Support Page 3 of 3