Distillatore Manuale
Distillatore Manuale
Arioso Power(TOC)
System
Manual
Arioso Power(TOC)
System
Inspection Certificate
•Model :
•Serial No. :
•Date :
Authorized Signature
Index
Manual
1 General Description 5
2 System Structure 8
4 Installation Condition 11
5 Installation Method Ⅰ 12
7 Display 16
8 Operating 22
13 Adapter 36
16 Pre-Treatment system 40
17 Accessories 42
19 Maintenance 44
1 General Description
2) Included filters
- Pretreatment : Pre filter + 10 ” A/C Filter
- RO System : RO Pack
- UP System : ① UP Pack
② 0.2 m Capsule filter
- TOC System (option)
4) Optional Accessories
① 254/185 UV Lamp
② MW 5000 UF filter
③ Printer
④ Foot-pedal s/w
⑤ Pressure regulater
⑥ Wall-mounting bracket
⑦ Water Tank (20L / 30L / 40L / 60L)
⑧ Frame
5) Applications
① General laboratory tests
② Preparations of stock solution
③ Washing & rinsing for glasswares
④ Analytical Instruments
⑤ Microorganism analysis
⑥ Cell Culture
⑦ DNA/RNA Tests
⑥
① ② ③ ④ ⑤ ⑦ ⑧ Ⓐ Ⓒ Ⓓ
Ts
⑨
Ⓑ
Ⓔ
Tank
2. Features
∙ Certified CE, ISO 9001, ISO 14001, NRTL
∙ Awarded Korean excellent products
∙ Auto calibraion for water sensor
∙ Extensible point of use
∙ Frost alarm message
∙ 3, 5 steps water level sensor
∙ Filter exchange indicator : Beep & charactor display
∙ Languages : English, Korean, etc
∙ Self-diagnosis functions
6 Human Corporation
3. Specifications
Arioso Power Arioso Power Arioso Power Arioso Power Arioso Power Arioso Power Arioso Power
Ⅰ Ⅰ+70 Ⅱ Ⅱ+70 Ⅲ Ⅱ(TOC) Ⅲ(TOC)
Product flow rate Max. RO : 15 L/Hr Max. RO : 25 L/Hr
(RO, UP) Max. 15 L/Hr Max. 25 L/Hr Max. 35 L/Hr Max. 25 L/Hr Max. 35 L/Hr
UP : 70 L/Hr UP : 70 L/Hr
Cleaning function · Auto regular flushing · Auto regular cleaning · Auto cleaning lower than setting value
· Filter replacement alarm & display (A/C, RO/UP pack, UF, UV lamp, 0.2㎛CF, etc)
System function · Volumetric time setting (RO/UP) · Language(English, Korean, etc)
· Auto/Manual Pressure S/W · Extensible point of use
Scholar Scholar-UV
Type _ _ Bio Integrate
PF PF
0.2㎛ final filter STD STD STD STD
Water Quality
-Endotoxin (Eu/㎖) NA < 0.125 NA < 0.125 < 0.001 < 0.001
-RNases (ng/㎖) - < 0.01 - < 0.01 < 0.01 < 0.01
Standard Accessories Pretreatment System, ROmembrane, UP Pack, Level Sensor, Manual, Power Cord
· Printer · Foot-Pedal switch · Pretreatment system · Low pressure valve, Wall-mounting bracket
Optional Accessories · Water tank(20L, 30L, 40L, 60L) · Vent filter for PE water tank · Frame
Output
[From this port to main instrument Input
using Blue 6mm(OD) tubing] [From tap water to this port
using Blue 10mm(OD) tubing]
UP/RESET switch
Locking system
8 Human Corporation
3. Main System / Front-inside
RO Pack
254/185 nm UV Lamp
MW 5000 UF Filter
UP Pack
③ Drain Port
Connect to Sink using Blue 6mm(OD) tubing
(TOC Version) ④ TOC Input (Option)
Connect to Input of TOC Meter using transparent 4mm(OD) tubing
① ② ③ ⑤ ⑦ ⑧
⑤ Printer Port (Option) - 4 pin
Connect to Printer
⑧ Power switch
(General version without TOC Meter) Connect to Power
1. Front
Hole
(Fit for the back-left rubber feet of main system.)
2. Back
10 H u m a n C o r p o r a t i o n
4 Installation Condition
1. Environmental Conditions
1) Indoor installation
6) Pollution degree 2
7) Installation Categories Ⅱ
③ Drain Port
Connect to Sink using Blue 6mm(OD) tubing
⑧ Power switch
Connect to Power
12 H u m a n C o r p o r a t i o n
4. Connect Front Pack (Refer to Page 9)
1) Connect RO pack Line
2) Connect UP Pack line
3) Connect 0.2 m Capsule Filter
4) Connect 254/185 UV Lamp (Option)
5) Connect MW 5000UF filter (Option)
Blue 10mm(OD)
tubing
14 H u m a n C o r p o r a t i o n
2. Back side of Main System
① ② ③ ④ ⑤ ⑥ ⑦ ⑧
Printer
Pretreatment
Output
TOC Meter
TOC drain
sink
1. Main Display
UP/RESET
2. Operation Display
16 H u m a n C o r p o r a t i o n
2) 5.2 bar/RO or - bar/RO (without Pressure sensor)
① Pressure of RO Membrane (Default Value : 3 to 7 bar)
Regulate Pressure of RO Membrane, the ratio of RO product / Drain volume
by controlling Product / Drain Valve
생산비율
Model Auto Drain Valve Auto Drain Valve ;
Product Drain
It automatically adjust
Arioso Power Ⅰ 4 6 Built-in the ratio of product
Arioso Power Ⅱ 4 6 and drain water
No volume.
Arioso Power Ⅲ 5 5
② Analog Pressure guage is equipped when it does not have Pressure sensor of RO membrane.
3) feed 148
RO
0.2 s /cm
① RO Product : 0.2 s /cm
② Feed water Conductivity : 148 s /cm (Default Value : < 350 s /cm)
4) TOC 2 ppb
UP
18.3 MΩ∙cm
6) 21.0 ℃ Amb
① 21.0℃ Amb : The present Ambient Temperature (Default Value : > 5℃)
② Caution: The system can be frozen up if the temperature is lower than 5℃!
7)
∙ Water Tank
( 2step ,3step ,5step )
※ For changing Default value Touch display → change the value with or key.
2) AC / UV
Setting replacement period of AC Filter, UV Lamp (Unit ; hour)
Time Setting
3) RO / UP RO / UP
or
Volume Dispensing Dispensing Time
(Option)
4) Replace period for 0.2 m Capsule Filter (or PF Filter) and MW5000 UF Filter
0.2CF / MW5000
※ In the case of Scholar - PF, Scholar - UV - PF Type
Time Setting
Replace 0.2CF →PF Filter!
5) Feed
Setting Max value of Feed water Conductivity. (Unit ; s /cm)
Water µs/cm
18 H u m a n C o r p o r a t i o n
8) TOC
Running time of UV Lamp for TOC Meter (Unit ; hour)
UV Time
9) Temp ① Amb : Setting Minimum value of Ambient temperature to prevent freezing (Unit ; ℃)
Amb / Feed water ② Feed Water : Setting Maximum value of Feed Water temperature.
10) TOC
Display Error No. for TOC Meter.
Error No
11) Print
Printing Function
12) Language
Language selecting function
13) Setting
On Off
Value
20 H u m a n C o r p o r a t i o n
5) Using Time Reset ∙ Reset running time of each filters. (Please Inquire it to Head Office)
7) AUTO or Manual
① AUTO : Only works when Feed Water pressure is in 1~5 bar (kg/㎠)
9) Auto Flushing
① Yes : Auto cleaning UP system for 120sec. after Cleaning RO system for 120sec.
every 1 hour in standby display.
② No : Cancel upper function.
Range : 0 ~ 200
Range : 0 ~ 200
Set the above valve as lower than 100 Set the above valve as higher than 100
1) Power S/W ON
Main Display
Touch Part
3) To produce UP,
touch UP or right side s/w ON
UP/RESET
To stop producing UP, re-touch them.
Operation Display
22 H u m a n C o r p o r a t i o n
4) Waiting Display
select
After touching set 0.2
Operating
RO / UP
Main Display → MODE → → Set time → Main →
Dispensing Time
R or U key
1. Pretreatment System
1) Prefilter
1) Replacement Period
① Approx. 20 ~ 40days (depends on using time feed water condition.)
② When the filter becomes to brown color (Refer to sticker posted on the housing.)
2) Replacement Method
① Make the system stop
② Close the Tap Water.
③ Remove Air by pressing the Air Removing Button (Red one) in the housing.
④ Replace PreFilter
2) A/C filter
1) Replacement Period
2) Display
∙ Alarm turned on
∙ Message blinking (Message is steadily displayed when it is used over 400 hours.)
3) Replacement Method
① Same as Prefilter replacement Method
② After replace a filter, touch screen for 5sec and you can hear“beep”.
26 H u m a n C o r p o r a t i o n
3) RO Filter (Pack)
1) Replacement Period
2) Display
∙ Alarm turned on
∙ System stop
∙ Alarm turned on
∙ Message blinking
3) Replacement Method
① Make the system stop
② Close the Tap Water
③ Depressurize the system by touching‘RO’to dispense
water for 20 sec. and stop it.
④ Shut power OFF to the unit and disconnect the electrical power to the unit.
⑤ Open the front door and replace RO pack.
(Connect the Inlet, drain and product tubings.)
⑥ Power it on and open the tap water to operate it again.
2) Display
•1st value •2nd value
∙ Alarm turned on
∙ Stop producing UP water and be recycled.
∙ Alarm turned on
∙ Message blinking
3) Replacement Method
① Make the system stop
② Close the Tap Water
③ Depressurize the system by touching‘UP’to dispense
water for 20 sec. and stop it.
④ Shut power OFF to the unit and disconnect the electrical power to the unit.
⑤ Open the front door and replace UP pack.
⑥ Power it on and open the tap water to operate it again.
4) Structure of UP Pack
④③ ② ①
① Inlet Line
② Inlet Line •2 Line : Scholar, Bio
③ Outlet Line •4 Line : ∙ Scholar - UV
④ Product Line ∙ Integrate
28 H u m a n C o r p o r a t i o n
5) 0.2㎛ Capsule Filter or PF (Pyrogen Free) Filter
1) Replacement Period
2) Display
•1st value •2nd value
3) Replacement Method
∙ Replace 0.2 m CF (or PF filter)
touch screen
for 5sec and
you can
hear“beep”
2) Display
•1st value •2nd value
3) Replacement Method
① Make the system stop
② Close the Tap Water ( Remove the Pressure inside)
③ Replace MW 5000 filter
④
touch screen
for 5sec and
you can
hear“beep”
30 H u m a n C o r p o r a t i o n
4) Structure and Caution
Applicable for Bio / Integrate Type
Blue Cap ※ This air removal is also needed for the first use
after installation of Bio/Integrate type system.
MW 5000 UF Filter
Remove the air in the MW 5000 UF Filter!!
Or the product volume and purity can be decreased.
7) UV Lamp
1) Replacement Period
2) Display
•1st value •2nd value
touch screen
for 5sec and
you can
hear“beep”
4) UV lamp structure
※ Warning!
This unit is equipped with an ultraviolet lamp.
Ultraviolet radiation is harmful to the eyes and skin.
Do not attempt to observe the lamp directly.
Some water will drain from the UV lamp when it is disconnected.
②
32 H u m a n C o r p o r a t i o n
11 Error Message
1)
Feed Pressure High ∙ Alarm turned on
•Solution
2)
Feed Pressure Low ∙ Alarm turned on
•Solution
3)
Feed Temp Exceeding ∙ Alarm turned on
•Solution
•Solution
5)
Check drain valve ∙ Alarm turned on
•Solution
생산비율
Model Auto Drain Valve Auto Drain Valve ;
Product Drain
It automatically adjust
Arioso Power Ⅰ 4 6 Built-in the ratio of product
Arioso Power Ⅱ 4 6 and drain water
No volume.
Arioso Power Ⅲ 5 5
6) Freezing Caution
or
•Solution
Default Value : 5℃
34 H u m a n C o r p o r a t i o n
12 TOC Meter
1) Error message
No Error Message Error Code Error Reason Remark
1 Change UV Lamp 01 Replace UV Lamp (After using 4000hr)
2 Error 02 02 Quality of Sample water is out of the Specification
(Below than 1 MΩ∙ ㎝)
3 Error 03 03 Temp. Sensor 1 (Error on Thermistor)
4 Error 06 06 Temp. of Sample water is over the Specification
(higher than 35℃)
5 Error 07 07 Temp. of sample water is out of the specification (lower than 5℃)
6 Error 08 08 Temp. Sensor 2 (Error on Thermistor)
7 Error 10 10 Value of TOC is out of the range of Specification
(Over 200ppb)
8 Error 18 18 UV Lamp doesn’t illuminate. (Disconnection of UV lamp)
9 System Check Failed 19 Message error, measurement is not possible.
1. Power Supply
2. UV Lamp Adapter
36 H u m a n C o r p o r a t i o n
14 How to use Level Sensor
1. Name of No.
① Cap
② Main Body
③ Output Line or Drain in Regeneration
④ Input Line (Tap Line)
⑤ Input Valve
38 H u m a n C o r p o r a t i o n
Installation ①
Output
④
3. Regeneration method
1) Close the Tap water.
2) Close the ⑤ valve.
3) Disassemble the ③ Output line and put it in the sink.
4) Open the ① cap.
5) Insert 500~600g salt and close the ① cap.
6) Remain it for 30 minutes and open the ⑤ valve.
And, open the tap water slightly as the water drips drop by drop for 10 minutes.
7) Open the tap water more for 10 minutes to remove the salt all in it.
(flow rate : 1L/min. approx.)
8) Close the tap water and connect ③ Output line to the main instrument.
Now, it is ready to use again!
1 2 3
Specifications
Picture No. Part No. Part No. Remark
1) 10” Prefilter : 1ea
In case of plenty of particles and
1 2) 10” Housing : 1ea HMC - PT - 1
rust in the feed water
3) Stand
1) 10” Prefilter : 2ea
In case of plenty of particles and
2 2) 10” Housing : 2ea HMC - PT - 2
rust in the feed water
3) Stand
1) 10” Prefilter : 1ea
2) 10” A/C filter : 1ea In case of plenty of particles, rust
3 3) 10” Housing : 2ea
HMC - PT - 3
and organics in the feed water
4) Stand
※ Installation Method
Input (In) Product (Out)
→
→
40 H u m a n C o r p o r a t i o n
HMC - WS 10 HMC - WS 10 - 1 HMC - WS 10 - 2
4 5 6
Specifications
Picture No. Part No. Part No. Remark
B-type
520(W) x 650(D) x 1350(H) mm
Model : Power Series
Tank
Model Volume In/Outlet Flow Rate
Flow
ResinRate
(cm) Salt (cm)
HMC-AR6 6L 20mm 1000L/Hr 15 x 45 24 x 42 x 72
42 H u m a n C o r p o r a t i o n
18 Parts
③ ② ①
① UP Sensor
② RO Sensor
③ Feed Sensor
④ Sol Valve 1
⑨
⑤ Temperature Sensor
⑧ ⑪ ⑥ Sol Valve 3
⑦ Sol Valve 4
⑧ Sol Valve 5
⑦
⑥ ⑫ ⑨ Pressure sensor No.1 (Feed Pressure)
⑩ ⑭
⑩ Pressure sensor No.2
⑤
⑪ Main Board
⑮ ⑫ SMPS
⑬ ⑬ TOC Board (Option)
⑭ UV Lamp Adapter
⑮ Pump Adapter
④ ⑯
⑯ Auto Drain Valve
(Installed in Arioso Power Ⅰ only)
•Check method
Sound Pump Symptom Note
SOL .1 O O Produce RO water, Drain
SOL .2 X X -
SOL .3 O X -
SOL .4 O X UV Lamp on Option
SOL .5 O X -
SOL .1 + SOL .2 O O Produce RO water, Drain
SOL .1 + SOL .2 + SOL .3 O O Increase Drain water
SOL .1 + SOL .2 + SOL .4 O O Drain, UV Lamp on Option
•Check method
working valve Pump Symptom Note
RO Product SOL .1 O Produce RO water, Drain
UP Product SOL .1, 4, 5 O Produce UP water, Drain
Auto Cleaning SOL .1, 3 O Drain
UV Lamp SOL .4 - UV lamp on Option
TOC SOL .1, 4, 5 O Produce UP water, TOC lamp on Option
44 H u m a n C o r p o r a t i o n
3) Trouble Shooting Details
Trouble Parts Problem / Possible Causes Solution
Sol 2
RO water drips.
Product/Drain Valve Unable to adjust the volume of RO Product / Drain Replace Product / Drain Valve
① Even there is no water inside water tank ① Change the location of the
but Tank Full message displays
Tank Full Floater of Level Sensor
② When the water Tank is full,
② Check the Slave Board
unable to display Tank Full
8) When the system goes back to initialization repeatedly during running the system.
(Repeatedly execute self test, auto drain / Reset system)
1 ) Lack of feed water
⇨
2) Auto Pressure S/W - Manual Position SET → NEXT → Auto or Manual
3) Pressure S/W error - Pressure S/W error can be occurred by lack of feed water pressure
4) When the Level Sensor is defective or Sensor has been wrongly placed.
46 H u m a n C o r p o r a t i o n