Manual

Arioso Power(TOC)
System
Manual

Arioso Power(TOC)
System
Inspection Certificate

•Item : Water Purification System

•Model :

•Serial No. :

•Date :

•Origin : Republic of Korea

We hereby certify that goods has been inspected before


shipment and found in good order.

Authorized Signature
Index
Manual

1 General Description 5

2 System Structure 8

3 TOC meter Structure 10

4 Installation Condition 11

5 Installation Method Ⅰ 12

6 Installation Method Ⅱ (Diagram) 14

7 Display 16

8 Operating 22

9 Volume Dispensing ∙ Dispensing Time 24

10 Filter Exchange and Display / Default Value 26

11 Error Message / Default Value 33

12 TOC Meter Error & UV Lamp 35

13 Adapter 36

14 Using Method of Level Sensor 37

15 Using Method of Water Softner 38

16 Pre-Treatment system 40

17 Accessories 42

18 Parts for a instrument 43

19 Maintenance 44
1 General Description

1. Arioso Power (TOC) System


The Power System is the unique water purification system combining
2 systems-RO System & UP System. It can produce simultaneously from the city/ground water.
And, it is a perfect instrument in space budget and efficiency.

1) Feed Water : Tap Water / Ground Water (Water Softner required)

2) Included filters
- Pretreatment : Pre filter + 10 ” A/C Filter
- RO System : RO Pack
- UP System : ① UP Pack
② 0.2 m Capsule filter
- TOC System (option)

3) STD Accessories : Manual, Power Cord, Level Sensor, Hand Lever

4) Optional Accessories
① 254/185 UV Lamp
② MW 5000 UF filter
③ Printer
④ Foot-pedal s/w
⑤ Pressure regulater
⑥ Wall-mounting bracket
⑦ Water Tank (20L / 30L / 40L / 60L)
⑧ Frame

5) Applications
① General laboratory tests
② Preparations of stock solution
③ Washing & rinsing for glasswares
④ Analytical Instruments
⑤ Microorganism analysis
⑥ Cell Culture
⑦ DNA/RNA Tests

Arioso Power (TOC) System 5


6) Flow Schematic


① ② ③ ④ ⑤ ⑦ ⑧ Ⓐ Ⓒ Ⓓ
Ts




Tank

Drain RO Product UP Product

∙ RO System ① Pretreatment system ② Temp sensor ③⑧ Conductivity sensor


④⑥ Pressure sensor ⑦ RO pack ⑨ Water sensor

∙ UP System Ⓐ UP pack Ⓑ 254/186 UV lamp Ⓒ MW5000 UF filter


Ⓓ Resistivity sensor Ⓔ 0.2 m final filter (PF final filter)

2. Features
∙ Certified CE, ISO 9001, ISO 14001, NRTL
∙ Awarded Korean excellent products
∙ Auto calibraion for water sensor
∙ Extensible point of use
∙ Frost alarm message
∙ 3, 5 steps water level sensor
∙ Filter exchange indicator : Beep & charactor display
∙ Languages : English, Korean, etc
∙ Self-diagnosis functions

6 Human Corporation
3. Specifications

Arioso Power Arioso Power Arioso Power Arioso Power Arioso Power Arioso Power Arioso Power
Ⅰ Ⅰ+70 Ⅱ Ⅱ+70 Ⅲ Ⅱ(TOC) Ⅲ(TOC)
Product flow rate Max. RO : 15 L/Hr Max. RO : 25 L/Hr
(RO, UP) Max. 15 L/Hr Max. 25 L/Hr Max. 35 L/Hr Max. 25 L/Hr Max. 35 L/Hr
UP : 70 L/Hr UP : 70 L/Hr

Display & Controller Large color 70 x 52 touch screen ← ←


Ambient temp : -10 to 100 ← ←
Feed water temp : 0 to 100 ℃ ← ←
Feed water : 0 to 999 ㎲/㎝ ← ←
Water quality display
Pure water (RO) : 0.2 to 250.0㎲/㎝ ← ←

Ultra Pure (UP) : Up to 18.3 ㏁·㎝ ← ←


- TOC : 0 to 200 ppb

Digital Pressure Sensor (0~10kg/㎠)


Pressure display
① Feed water pressure ② RO membrane pressure

· Various valves · Auto sensor calibration(RO, UP)


Self-check function
· System operation status · Self-diagnosis

Cleaning function · Auto regular flushing · Auto regular cleaning · Auto cleaning lower than setting value

· Filter replacement alarm & display (A/C, RO/UP pack, UF, UV lamp, 0.2㎛CF, etc)
System function · Volumetric time setting (RO/UP) · Language(English, Korean, etc)
· Auto/Manual Pressure S/W · Extensible point of use

Scholar Scholar-UV
Type _ _ Bio Integrate
PF PF
0.2㎛ final filter STD STD STD STD

Pyrogen free final filter STD STD

254/185 UV lamp STD STD STD

MW 5000 UF filter STD STD

Water Quality

-TOC (ppb) 5~10 5~10 0~5 0~5 5~10 0~5

-Endotoxin (Eu/㎖) NA < 0.125 NA < 0.125 < 0.001 < 0.001

-Bacteria (cfu/㎖) <1 <1 <1 <1 <1 <1

-Particles ( > 0.22㎛/㎖) <1 <1 <1 <1 <1 <1

-RNases (ng/㎖) - < 0.01 - < 0.01 < 0.01 < 0.01

-DNases (pg/㎖) - - - - <4 <4

-DNases (pg/ ) - <4 - <4 - -

Dimensions (WxDxH) 320 x 490 x 475 mm

Power 230V / 110V, 50/60Hz, 60W

Standard Accessories Pretreatment System, ROmembrane, UP Pack, Level Sensor, Manual, Power Cord
· Printer · Foot-Pedal switch · Pretreatment system · Low pressure valve, Wall-mounting bracket
Optional Accessories · Water tank(20L, 30L, 40L, 60L) · Vent filter for PE water tank · Frame

∙ Feed water requirement ; TOC < 50 ppb

Arioso Power (TOC) System 7


2 System Structure

1. Pretreatment System (HMC-PT-3)


Air Vent Button
(Red Button)

Output
[From this port to main instrument Input
using Blue 6mm(OD) tubing] [From tap water to this port
using Blue 10mm(OD) tubing]

10 ” A/C Filter 10 ” Pre Filter

2. Main System / Front


Display

UP/RESET switch

0.2 m Capsule filter

Locking system

8 Human Corporation
3. Main System / Front-inside

RO Pack

254/185 nm UV Lamp
MW 5000 UF Filter

UP Pack

Product / Drain valve


Ⓐ valve Control Water volume with Ⓐ valve ,
Ⓑ valve and then fix Ⓐ valve by turning Ⓑ valve clockwise.
(※Do not fasten too strongly.)
‘Auto
( drain valve’is installed in Arioso Power I.)

4. Main System / Back Side


① ② ③ ④ ⑤ ⑥⑦ ⑧
① Input Port [6mm(OD)]
Connect to output of Pre-treatment system
using Blue 6mm(OD) tubing

② Pure Water Product Port


Connect to Water tank using White 6mm(OD) tubing

③ Drain Port
Connect to Sink using Blue 6mm(OD) tubing
(TOC Version) ④ TOC Input (Option)
Connect to Input of TOC Meter using transparent 4mm(OD) tubing
① ② ③ ⑤ ⑦ ⑧
⑤ Printer Port (Option) - 4 pin
Connect to Printer

⑥ TOC Meter Power Port (Option) - 5 pin


Connect to Power of TOC meter

⑦ Level Sensor - 2, 3, 6 pin


After connect level sensor, put it Water Tank

⑧ Power switch
(General version without TOC Meter) Connect to Power

Arioso Power (TOC) System 9


3 TOC Meter Structure (Option)

1. Front

① Water Flow Controller


Control Water Flow of TOC Meter (15-20㎖/min : approx.
30-35drops / 10 sec) Set the ball in‘② Flow Meter’
between RED lines.

② Flow Meter for TOC

Ball RED Line

※ If there is air inside Flow Meter, Ball could go down.

Hole
(Fit for the back-left rubber feet of main system.)

2. Back

∙ TOC Meter Power Port


Connect to Main system ⑥ TOC Meter Power Port
(Refer to page 09)

∙ TOC Meter Input


Connect to Main system ④ TOC Input using transparent
4mm(OD) tubing

∙ TOC Meter Output


Connect to Sink using transparent 4mm(OD) tubing
※ Make sure that tubing is not twisted and set upside.

10 H u m a n C o r p o r a t i o n
4 Installation Condition

1. Environmental Conditions
1) Indoor installation

2) Altitude should be up to 2000m


3) Ambient temperature : 5℃ ~ 40℃
(If the temperature is lower the 5℃, the system can be frozen up)
4) Maximum relative humidity : 80%

5) Power : 230 ± 10%V, 50/60㎐ ± 1㎐, 60W

6) Pollution degree 2

7) Installation Categories Ⅱ

CAUTION!! - Refer to accompanying documents

CAUTION!! - ELECTRIC SHOCK

“ To avoid electrical shock, do not open the cover.


Refer servicing to qualified personnel only.”

8) The instrument should be placed on steady workbench that is free


from a strong vibration source.
9) The location room should be free from a strong electro magnetic
interference source or harmful or corrosive gases.

Arioso Power (TOC) System 11


5 Installation Method Ⅰ

1. Pretreatment System (Refer to Page 8)


1) Connect the Input of Pretreatment System into the Tap Water. [Blue 10mm(OD) tubing]
※ 2) Before connect the main system, wash 10” A/C filter by opening Tap Water.
[Drain minimum 60 sec.]
3) Connect the Output of Pretreatment into the Input Port in back side of main system.
[Blue 6mm(OD) tubing]

2. Main System (Refer to Page 9)


① Input Port [6mm(OD)]
Connect to output of Pre-treatment system
using Blue 6mm(OD) tubing

② Pure Water Product Port


Connect to Water tank using White 6mm(OD) tubing

③ Drain Port
Connect to Sink using Blue 6mm(OD) tubing

④ TOC Input (Option)


Connect to Input of TOC Meter using transparent 4mm(OD) tubing

⑤ Printer Port (Option) - 4 pin


Connect to Printer

⑥ TOC Meter Power Port (Option) - 5 pin


Connect to Power of TOC meter

⑦ Level Sensor - 2, 3, 6 pin


After connect level sensor, put it Water Tank

⑧ Power switch
Connect to Power

3. Main System (Refer to Page 8)


1) Unscrew the locking system at under of Main body and Open it.
2) Lock the door again when you are using the instrument.

12 H u m a n C o r p o r a t i o n
4. Connect Front Pack (Refer to Page 9)
1) Connect RO pack Line
2) Connect UP Pack line
3) Connect 0.2 m Capsule Filter
4) Connect 254/185 UV Lamp (Option)
5) Connect MW 5000UF filter (Option)

5. TOC Meter Installation Method


① Put TOC Meter back-right of Main system.

② Insert the back-left rubber feet


of main system into the hole of TOC meter.

Arioso Power (TOC) System 13


6 Installation Method Ⅱ (Diagram)

1. Pre-treatment System Diagram

Air Vent RED button

Blue 10mm(OD)
tubing

10 ” A/C Filter 10 ” Pre Filter

① Open the tap after connecting lines.


② Drain water for approx.
60sec to wash A/C filter.
③ Connect to input port of Main system
after close the tap. (Page 9)

14 H u m a n C o r p o r a t i o n
2. Back side of Main System

① ② ③ ④ ⑤ ⑥ ⑦ ⑧

Printer

Pretreatment
Output

TOC Meter

TOC drain

sink

① Blue 6mm(OD) input tubing


② White 6mm(OD) (RO product) tubing
•Level Sensor Type ③ Blue 6mm(OD) (RO Drain) tubing
⑴ 2 step level sensor ④ Transparent 4mm(OD) (TOC meter Input) tubing
⑵ 3 step level sensor ⑤ Printer cable (4pin)
⑶ 5 step level sensor
⑥ TOC meter Power (5pin)
⑦ Level Sensor
⑧ Power Cable port & Switch

Arioso Power (TOC) System 15


7 Display

1. Main Display

UP/RESET

MODE : Setting Operating Condition

R : Produce / Stop RO water Key

U : Produce / Stop UP water Key

SET : Checking Instrument

: ① Press this, Produce / Stop UP water


UP/RESET ② Press for 3sec, back to Main display

2. Operation Display

1) A 1.2 bar/F or M 1.2 bar/F

① A (Auto) : Only works when the Tap water pressure is in 1~5bar(kg/㎠).


It doesn’t work when the Tap water pressure is over 5bar(kg/㎠) or
blow than1bar(kg/㎠). (Default Value : 1~5bar, Modifiable at MODE )
② M (Manual) : Only works regardless of Tap water pressure
(Remarks : It works even though there is no supplied water!)
③ 1.2 bar/F : Present Feed Water pressure

④ Setting Method : SET → NEXT → Auto or Manual

16 H u m a n C o r p o r a t i o n
2) 5.2 bar/RO or - bar/RO (without Pressure sensor)
① Pressure of RO Membrane (Default Value : 3 to 7 bar)
Regulate Pressure of RO Membrane, the ratio of RO product / Drain volume
by controlling Product / Drain Valve
생산비율
Model Auto Drain Valve Auto Drain Valve ;
Product Drain
It automatically adjust
Arioso Power Ⅰ 4 6 Built-in the ratio of product
Arioso Power Ⅱ 4 6 and drain water
No volume.
Arioso Power Ⅲ 5 5

② Analog Pressure guage is equipped when it does not have Pressure sensor of RO membrane.

3) feed 148
RO
0.2 s /cm
① RO Product : 0.2 s /cm
② Feed water Conductivity : 148 s /cm (Default Value : < 350 s /cm)

4) TOC 2 ppb
UP
18.3 MΩ∙cm

① UP Product : 18.3MΩ∙cm at 25℃


② TOC : 2 ppb (Option)

5) 15.0 ℃/F or 15.0 ℃/F

① 15.0℃ : Temperature of Feed Water


② : Beep ON
③ : Beep OFF
④ Setting method : SET → Beep ON or Beep OFF

6) 21.0 ℃ Amb

① 21.0℃ Amb : The present Ambient Temperature (Default Value : > 5℃)
② Caution: The system can be frozen up if the temperature is lower than 5℃!

7)
∙ Water Tank
( 2step ,3step ,5step )

Arioso Power (TOC) System 17


3. Mode Display

※ For changing Default value Touch display → change the value with or key.

1) RO / UP : Setting replacement period of RO / UP Pack


Value Setting ① 1st Value : 1st Alert message RO/UP pack replacement
② 2nd Value : 2nd Alarm message RO/UP pack replacement

2) AC / UV
Setting replacement period of AC Filter, UV Lamp (Unit ; hour)
Time Setting

3) RO / UP RO / UP
or
Volume Dispensing Dispensing Time
(Option)

∙ Setting dispensing Volume ∙ Setting dispensing time


(Vloume : 0.01 to 99.0 L)

※ SET → Set the Volume Dispensing or Dispensing Time first.

4) Replace period for 0.2 m Capsule Filter (or PF Filter) and MW5000 UF Filter
0.2CF / MW5000
※ In the case of Scholar - PF, Scholar - UV - PF Type
Time Setting
Replace 0.2CF →PF Filter!

5) Feed
Setting Max value of Feed water Conductivity. (Unit ; s /cm)
Water µs/cm

6) Pressure ① FD : Setting Feed Water Pressure (Unit ; kg/㎠, bar)


Value Setting ② RO : Setting RO membrane Pressure (Unit ; kg/㎠, bar)

7) Running time Display total accumulated running time of


RO / UP / AC / UV LAMP RO Pack, UP Pack, AC Filter, UV Lamp 0.2 m CF, UF Filter. (Unit ; hour)

18 H u m a n C o r p o r a t i o n
8) TOC
Running time of UV Lamp for TOC Meter (Unit ; hour)
UV Time

9) Temp ① Amb : Setting Minimum value of Ambient temperature to prevent freezing (Unit ; ℃)
Amb / Feed water ② Feed Water : Setting Maximum value of Feed Water temperature.

10) TOC
Display Error No. for TOC Meter.
Error No

11) Print
Printing Function

12) Language
Language selecting function

13) Setting
On Off
Value

① On : Operating with set value in Mode function


② Off : Operating regardless default values.
(※Pressure Function is maintained.)

Arioso Power (TOC) System 19


4. SET Display

1) Parts Check ① SOL . 1 : No1. valve Operating check


② SOL . 2 : No2. valve Operating check
③ SOL . 3 : No3. valve Operating check
④ SOL . 4 : No4. valve Operating check
⑤ SOL . 5 : No5. valve Operating check

2) Functional Check ① RO Product : Check RO Product

② Auto Cleaning : Check Auto Cleaning

③ TOC : Check TOC Function

④ UP Product : Check UP Product

⑤ UV Lamp : Check UV Lamp

3) Calibration Calibration in case RO sensor (Conductivity Sensor),


Sensor UP sensor(Resistivity sensor) has Error
※ Please Inquire about Calibration to Head Office.

4) Volume Dispensing or Dispensing Time

Choose one between above 2 functions.


① Volume Dispensing (Option) : Dispensing RO and UP after putting dispensing volume you want.
(Volume range : 0.01 to 99.0 L)
② Dispensing Time : Setting time for Dispensing RO and UP

20 H u m a n C o r p o r a t i o n
5) Using Time Reset ∙ Reset running time of each filters. (Please Inquire it to Head Office)

6) BEEP ON or BEEP OFF

∙ Setting Alarm or touch sound.

7) AUTO or Manual

① AUTO : Only works when Feed Water pressure is in 1~5 bar (kg/㎠)

② Manual : Works regardless of feed water pressure


(Even works without supplied water)
Caution ; The instrument can be defective because
it woks even though there is no supplied water.
8) Default Value

① Yes : Going back to factory setting values.


② No : Remaining the current setting values.

9) Auto Flushing

① Yes : Auto cleaning UP system for 120sec. after Cleaning RO system for 120sec.
every 1 hour in standby display.
② No : Cancel upper function.

10) Volume Calibration

Calibration to get high accuracy Once equipped the Flow meter

Range : 0 ~ 200
Range : 0 ~ 200

Ex) Calibrate after serveral tests. Set 100ml as dispensing volume

produce more than 100ml produce less than 100ml

Set the above valve as lower than 100 Set the above valve as higher than 100

Test the above procedure again till it is calibrated.

Arioso Power (TOC) System 21


8 Operation

1) Power S/W ON

Main Display

2) R Touch : RO water Product Key


RO
① Self Test : Auto checking Feed water pressure, storage tank etc.,
② Auto Cleaning : Auto Cleaning inside of instrument.
※ Extension : When RO conductivity isn’ t good enough,
it automatically extends cleaning for max. 600 sec.
③ RO product : Touch RO to stop producing RO water.
Pressure : 3 to 7 bar
(Default Valve)

Touch Part

U Touch: UP product key


UP

3) To produce UP,
touch UP or right side s/w ON
UP/RESET
To stop producing UP, re-touch them.

Operation Display

22 H u m a n C o r p o r a t i o n
4) Waiting Display

① MODE : Switch over MODE display

② S/B : ∙ Switch over Standby display.


∙ To release function, re-touch the key
- Function -
※ During Standby, Auto cleaning circulated.
Once UP water (inside of instrument) value is lower than 10 MΩ∙ cm
① RO System : Auto cleaning for 120 sec. and Extend max 600sec.
② UP System : Auto cleaning for 60 sec. and Extend max 600sec.
※Standby function could shorten the filter replacement period!

③ TOC (Option) : Measuring TOC value, when TOC Meter is equipped.

Arioso Power (TOC) System 23


9 Volume Dispensing, Dispensing Time
※ While you use this function,“Filter Exchange message”is not Displayed

1) Volume Dispensing setting method.

ex) set RO product 13.25L

RO key touch (change the color)

After touching set 10

After touching set 3

select
After touching set 0.2

After touching set 0.05

Operating

※ Calibrate volume, when there is difference between


product volume and set volume.

SET → NEXT → Volume Calibration → put value
24 H u m a n C o r p o r a t i o n ( refer to page 21 )
2) Dispensing Time setting method.
- Calculate as much as set dispensing volume.
- Setting method

Main Display → SET → Dispensing Time → Main →

RO / UP
Main Display → MODE → → Set time → Main →
Dispensing Time

R or U key

※ Product volume results from Setting time


(Product volume can vary depending on water temp, filter condition)
∙ Water temperature : 7℃
∙ Input Conductivity : 185 s /㎝
∙ Model : Arioso Power Ⅰ (Scholar Type)
∙ Dispensing Time Product volume Dispensing Time Product volume
15 sec. 50 ㎖ 3 min. 26 sec. 700 ㎖
30 sec. 100 ㎖ 5 min. 5 sec. 1L
47 sec. 150 ㎖ 6 min. 7 sec. 1.2 L
63 sec. 200 ㎖ 7 min. 15 sec. 1.4 L
80 sec. 250 ㎖ 8 min. 10 sec. 1.6 L
90 sec. 300 ㎖ 9 min. 15 sec. 1.8 L
120 sec. 400 ㎖ 10 min. 15 sec. 2.0 L
2 min. 26 sec. 500 ㎖

Arioso Power (TOC) System 25


10 Filter replacement and Display

1. Pretreatment System
1) Prefilter
1) Replacement Period
① Approx. 20 ~ 40days (depends on using time feed water condition.)
② When the filter becomes to brown color (Refer to sticker posted on the housing.)
2) Replacement Method
① Make the system stop
② Close the Tap Water.
③ Remove Air by pressing the Air Removing Button (Red one) in the housing.
④ Replace PreFilter

2) A/C filter
1) Replacement Period

Default Value : 400hr

2) Display

After replace it,


touch screen
for 5sec until
you hear the
“beep”

∙ Alarm turned on
∙ Message blinking (Message is steadily displayed when it is used over 400 hours.)

3) Replacement Method
① Same as Prefilter replacement Method
② After replace a filter, touch screen for 5sec and you can hear“beep”.

※ Replace A/C Filter can disappear by touch screen,


but it appears when you turn on the system again without replacing filter.

26 H u m a n C o r p o r a t i o n
3) RO Filter (Pack)
1) Replacement Period

Default Value : ① 1st value : 20 s /㎝


② 2nd value : 30 s /㎝

2) Display

•1st value •2nd value

∙ Alarm turned on
∙ System stop
∙ Alarm turned on
∙ Message blinking

3) Replacement Method
① Make the system stop
② Close the Tap Water
③ Depressurize the system by touching‘RO’to dispense
water for 20 sec. and stop it.
④ Shut power OFF to the unit and disconnect the electrical power to the unit.
⑤ Open the front door and replace RO pack.
(Connect the Inlet, drain and product tubings.)
⑥ Power it on and open the tap water to operate it again.

Arioso Power (TOC) System 27


4) UP Pack
1) Replacement Period

Default Value : ① 1st value : 13.0MΩ∙ ㎝


② 2nd value : 10.0MΩ∙ ㎝

2) Display
•1st value •2nd value

∙ Alarm turned on
∙ Stop producing UP water and be recycled.
∙ Alarm turned on
∙ Message blinking

3) Replacement Method
① Make the system stop
② Close the Tap Water
③ Depressurize the system by touching‘UP’to dispense
water for 20 sec. and stop it.
④ Shut power OFF to the unit and disconnect the electrical power to the unit.
⑤ Open the front door and replace UP pack.
⑥ Power it on and open the tap water to operate it again.

4) Structure of UP Pack

④③ ② ①
① Inlet Line
② Inlet Line •2 Line : Scholar, Bio
③ Outlet Line •4 Line : ∙ Scholar - UV
④ Product Line ∙ Integrate

28 H u m a n C o r p o r a t i o n
5) 0.2㎛ Capsule Filter or PF (Pyrogen Free) Filter
1) Replacement Period

Default Value : ① 1st value : 280hr


② 2nd value : 300hr
※ Replace‘PF Filter’in case of Scholar-PF, Scholar-UV-PF Type.

2) Display
•1st value •2nd value

∙ Alarm turned on ∙ Alarm turned on


∙ Message blinking ∙ massage is steadily displayed

3) Replacement Method
∙ Replace 0.2 m CF (or PF filter)

touch screen
for 5sec and
you can
hear“beep”

※ Replace 0.2 CF message can disappear by touching it.


But it appears when you turn on the system again without replacing filter.

Arioso Power (TOC) System 29


6) MW5000 UF Filter
1) Replacement Period

Default Value : ① 1st value : 280hr


② 2nd value : 300hr

2) Display
•1st value •2nd value

∙ Alaram turned on ∙ Alaram turned on


∙ Message blinking ∙ Message is steadily displayed

3) Replacement Method
① Make the system stop
② Close the Tap Water ( Remove the Pressure inside)
③ Replace MW 5000 filter

touch screen
for 5sec and
you can
hear“beep”

※ Replace UF Filter message can disappear by touching it.


But it appears when you turn on the system again without replacing filter.

30 H u m a n C o r p o r a t i o n
4) Structure and Caution
Applicable for Bio / Integrate Type

※ After replacing MW5000 UF Filter or UP Pack,


press the‘UP’button and then open
the White Cap of MW 5000 UF filter for removing
White Cap air inside. If the water starts to come out from the
vent, close the White Cap. You can use the
instrument now.

Blue Cap ※ This air removal is also needed for the first use
after installation of Bio/Integrate type system.

MW 5000 UF Filter
Remove the air in the MW 5000 UF Filter!!
Or the product volume and purity can be decreased.

7) UV Lamp
1) Replacement Period

Default Value : ① 1st value : 1800hr


② 2nd value : 2000hr

2) Display
•1st value •2nd value

∙ Alarm turned on ∙ Alarm turned on


∙ Message blinking ∙ Message is steadily displayed

Arioso Power (TOC) System 31


3) Replacement Method

① Close the Tap Water


② Turn the instrument off and disconnect power cord from unit.
③ Replace UV lamp (Remove the Pressure inside)
④ Make sure the UV lamp is again connected to the UV lamp adapter.

touch screen
for 5sec and
you can
hear“beep”

※ Replace UV Lamp message can disappear by touch screen,


but it appears when you turn on the system again without replacing filter.

4) UV lamp structure

① ① Outlet Line : Yellow tubing


② Inlet Line : White tubing
③ Adapter

※ Warning!
This unit is equipped with an ultraviolet lamp.
Ultraviolet radiation is harmful to the eyes and skin.
Do not attempt to observe the lamp directly.
Some water will drain from the UV lamp when it is disconnected.

32 H u m a n C o r p o r a t i o n
11 Error Message

1)
Feed Pressure High ∙ Alarm turned on

•Solution

Default Value : 5kg/㎠

① Adjust the feed water pressure to less than 5kg/㎠


② Install pressure regulator (Ask for it to Head-Office).

2)
Feed Pressure Low ∙ Alarm turned on

•Solution

Default Value : 1kg/㎠

① Use booster pump


② Change feed water pressure mode to Manual mode

SET → next → Auto or Manual

※ At manual mode, Instrument can be operated even without Water supply.

3)
Feed Temp Exceeding ∙ Alarm turned on

•Solution

Default Value : 35.0℃

Decrease the temperature of feed water. Otherwise, filter can be damaged.

Arioso Power (TOC) System 33


4)
Feed Purity Exceeding ∙ Alarm turned on

•Solution

Default Value : 350 s /㎝

① Install the Pretreatment system.


② If the feed water is ground water, Water softner should be installed.

5)
Check drain valve ∙ Alarm turned on

•Solution

Default Value : RO membrane pressure - 3~7㎏/㎠

① When Pressure of RO membrane is out of Default value,


Purity and Product volume can be influenced by it.
② Adjust Default Value by turning Drain Valve (Page 09)
③ Ratio of Product and Drain volume

생산비율
Model Auto Drain Valve Auto Drain Valve ;
Product Drain
It automatically adjust
Arioso Power Ⅰ 4 6 Built-in the ratio of product
Arioso Power Ⅱ 4 6 and drain water
No volume.
Arioso Power Ⅲ 5 5

6) Freezing Caution

or

•Solution

Default Value : 5℃

Increase the ambient temperature to prevent Freezing.

34 H u m a n C o r p o r a t i o n
12 TOC Meter

1) Error message
No Error Message Error Code Error Reason Remark
1 Change UV Lamp 01 Replace UV Lamp (After using 4000hr)
2 Error 02 02 Quality of Sample water is out of the Specification
(Below than 1 MΩ∙ ㎝)
3 Error 03 03 Temp. Sensor 1 (Error on Thermistor)
4 Error 06 06 Temp. of Sample water is over the Specification
(higher than 35℃)
5 Error 07 07 Temp. of sample water is out of the specification (lower than 5℃)
6 Error 08 08 Temp. Sensor 2 (Error on Thermistor)
7 Error 10 10 Value of TOC is out of the range of Specification
(Over 200ppb)
8 Error 18 18 UV Lamp doesn’t illuminate. (Disconnection of UV lamp)
9 System Check Failed 19 Message error, measurement is not possible.

2) TOC meter UV lamp replacement method.

TOC Drain Tube UV Lamp.


TOC UV Lamp Slowly life
it up.

TOC Inlet Tube

Arioso Power (TOC) System 35


13 Adapter

1. Power Supply

2. UV Lamp Adapter

36 H u m a n C o r p o r a t i o n
14 How to use Level Sensor

When you open the Tank cap,


make sure that level sensor doesn’
t
get twisted.

You should Install Level


sensor vertically.

•Level Sensor Type


① 2 step level sensor
② 3 step level sensor
③ 5 step level sensor

Arioso Power (TOC) System 37


15 Water Softner for Pure Water System

1. Name of No.
① Cap
② Main Body
③ Output Line or Drain in Regeneration
④ Input Line (Tap Line)
⑤ Input Valve

38 H u m a n C o r p o r a t i o n
Installation ①

Output

Sink Table Input


Water Softner Pure Water System

2. Installation (Refer to Page 38)


1) Connect ④ Input Line to Tap Water (10mm(OD) Blue tubing)
2) Connect ③ Output to Pure Water System or pretreatment system
3) Open the Tap water.

3. Regeneration method
1) Close the Tap water.
2) Close the ⑤ valve.
3) Disassemble the ③ Output line and put it in the sink.
4) Open the ① cap.
5) Insert 500~600g salt and close the ① cap.
6) Remain it for 30 minutes and open the ⑤ valve.
And, open the tap water slightly as the water drips drop by drop for 10 minutes.
7) Open the tap water more for 10 minutes to remove the salt all in it.
(flow rate : 1L/min. approx.)
8) Close the tap water and connect ③ Output line to the main instrument.
Now, it is ready to use again!

4. Regeneration & Replacement Period


① Available volume of Water : Approx. 2000L (In case of the Hardness of feed water is 1000ppm)
② Regeneration Period : 1 week
③ Replacement period of Cation Resin in Water Softner : Approx. 6 ~ 10 months
Volume of Cation Resin : Approx. 3L

Arioso Power (TOC) System 39


16 Pretreatment System

HMC - PT - 1 HMC - PT - 2 HMC - PT - 3

1 2 3
Specifications
Picture No. Part No. Part No. Remark
1) 10” Prefilter : 1ea
In case of plenty of particles and
1 2) 10” Housing : 1ea HMC - PT - 1
rust in the feed water
3) Stand
1) 10” Prefilter : 2ea
In case of plenty of particles and
2 2) 10” Housing : 2ea HMC - PT - 2
rust in the feed water
3) Stand
1) 10” Prefilter : 1ea
2) 10” A/C filter : 1ea In case of plenty of particles, rust
3 3) 10” Housing : 2ea
HMC - PT - 3
and organics in the feed water
4) Stand

※ Installation Method
Input (In) Product (Out)

Sink Table Pure Water System


IN : Connect to Tap Water (Input)
OUT : Connect to Pure Water System (Product)

40 H u m a n C o r p o r a t i o n
HMC - WS 10 HMC - WS 10 - 1 HMC - WS 10 - 2

4 5 6
Specifications
Picture No. Part No. Part No. Remark

1) 10” Prefilter : 1ea


4 2) Stand
HMC - WS 10

① When the feed water is the Ground water.


1) Water Softner : 1ea
② In case of plenty of particles and
2) 10” Prefilter : 1ea
5 3) 10” Housing : 1ea
HMC - WS 10 - 1 organics and rust in the feed water
4) Stand ③ In case of plenty of Hardness
((Ca++, Mg++) in the feed water

④ When the conductivity of feed water


is more than 300~400 g/㎝
1) Water Softner : 1ea
2) 10” Prefilter : 1ea
6 3) 10” Housing : 1ea
HMC - WS 10 - 2
4) Stand

10” Prefilter HMC - 10 PE Replacement Period : Approx. 30 days /


Filters When the filter becomes to brown color

10” A/C filter HMC - 10 AC Replacement Period : Approx. 180 days

Arioso Power (TOC) System 41


17 Accessories

•Wall-mounting bracket •Frame


A B
A-type
430(W) x 630(D) x 1350(H) mm
Model : RO, UP Series

B-type
520(W) x 650(D) x 1350(H) mm
Model : Power Series

•Automatic Regeneration Softner

Tank
Model Volume In/Outlet Flow Rate
Flow
ResinRate
(cm) Salt (cm)
HMC-AR6 6L 20mm 1000L/Hr 15 x 45 24 x 42 x 72

HMC-AR20 10L 20mm 1500L/Hr 20 x 45 24 x 42 x 72

•Water Tank (Pedestal option)

20L 30L 40L 60L

42 H u m a n C o r p o r a t i o n
18 Parts

③ ② ①

① UP Sensor

② RO Sensor

③ Feed Sensor

④ Sol Valve 1

⑤ Temperature Sensor

⑧ ⑪ ⑥ Sol Valve 3

⑦ Sol Valve 4

⑧ Sol Valve 5

⑥ ⑫ ⑨ Pressure sensor No.1 (Feed Pressure)
⑩ ⑭
⑩ Pressure sensor No.2

⑪ Main Board
⑮ ⑫ SMPS
⑬ ⑬ TOC Board (Option)

⑭ UV Lamp Adapter

⑮ Pump Adapter
④ ⑯
⑯ Auto Drain Valve
(Installed in Arioso Power Ⅰ only)

Arioso Power (TOC) System 43


19 Maintenance

1) Sol valve Check method

SET → Parts Check →

•Check method
Sound Pump Symptom Note
SOL .1 O O Produce RO water, Drain
SOL .2 X X -
SOL .3 O X -
SOL .4 O X UV Lamp on Option
SOL .5 O X -
SOL .1 + SOL .2 O O Produce RO water, Drain
SOL .1 + SOL .2 + SOL .3 O O Increase Drain water
SOL .1 + SOL .2 + SOL .4 O O Drain, UV Lamp on Option

SOL .1 + SOL .2 + SOL .4


O O Drain
+ SOL .5

SOL .1 + SOL .3 O O Drain


SOL .1 + SOL .4 O O Drain
SOL .1 + SOL .5 O O Produce RO water, Drain

2) Functional check method

SET → Functional Check →

•Check method
working valve Pump Symptom Note
RO Product SOL .1 O Produce RO water, Drain
UP Product SOL .1, 4, 5 O Produce UP water, Drain
Auto Cleaning SOL .1, 3 O Drain
UV Lamp SOL .4 - UV lamp on Option
TOC SOL .1, 4, 5 O Produce UP water, TOC lamp on Option

44 H u m a n C o r p o r a t i o n
3) Trouble Shooting Details
Trouble Parts Problem / Possible Causes Solution

•When the system works, unknown noise occurs.


Sol 1 •Unable to produce RO Water • Replace Sol 1

•No drain comes out

Sol 2

•When RO Water droplets slowly without stopping


Sol 3 • Replace Sol 3
the production during‘Auto Drain’
.
•Even after pressing UP → unable to produce UP water
Sol 4 •Over the process of UP Water production, • Replace Sol 4

RO water drips.

•Even after pressing UP → unable to produce UP water


Sol 5 •Over the process of UP Recycling, UP water drips. • Replace Sol 5

•Unable to execute Recycling

Pump Unable to operate the system at all. Replace Pump

Product/Drain Valve Unable to adjust the volume of RO Product / Drain Replace Product / Drain Valve

Even the system works properly but suddenly


Pressure S/W Replace Pressure S/W
the system reboots/resets and initializes to Self Test

① Check the Power Cord


② Check the Fuse
Unable to display
③ Check the Power Connector
④ Check the Power Supply

① Check the UV Lamp Adapter


② Check the Power Supply
Error on UV lamp lighting
③ Check the Slave Board
④ Check the Relay Board

① Even there is no water inside water tank ① Change the location of the
but Tank Full message displays
Tank Full Floater of Level Sensor
② When the water Tank is full,
② Check the Slave Board
unable to display Tank Full

When the system is on following condition ;


- Low Pressure The Valve doesn’t work.
- Tank Full (It is normal.)
- Replace RO Pack

Arioso Power (TOC) System 45


4) Unable to produce RO Water
1 ) The feed water is not supplied properly from the tap water. - Check
2) Due to defective pump or error, unable to supply the feed water sufficiently to the filters.
- Check the operating status and replace the pump
3) Due to exhausted RO Pack, the filter membrane has been clogged - Replace the Filter
4) When you replace the RO Pack, the cutting parts are not removed.
- Check and remove the Cutting Parts
5) In case the Solenoid Valve No.1 (2-way) is defective. - Check & Replace it
6) When the Procuct/Drain valve has not been adjusted properly or is defective.
- Check & adjust the pressure gauge at front side.
If you unable to adjust, please replace to new one.

5) Unable to produce UP Water


1 ) In case unable to produce RO Water properly.
2) In case the Solenoid Valve No.4 (3-way) & Solenoid Valve No.5 (3-way) are defective.
- Check Solenoid Valve No.4 (3-way) first and replace and test again
3) When the UP Pack has been clogged - Replace it
- Due to exhausted filter, the filter has been clogged by particles or dust - Replace the Filter.
- If the cutting parts are not removed properly - Check the removal of Cutting Parts

6) LCD doesn’t work even after main power is switched on


1 ) Main Power Connector Error - Check and Replace
2) Fuse has been disconnected - Replace to spare Fuse
3) Due to power supply error, unable to supply electricity to the controller - Check & Replace

7) When‘Low Pressure’Message appears


1 ) In case the feed water (tap water) has not been supplied sufficiently - Check
2) In case the pressure S/W is defective - When the feed water pressure is low, check the water
quality/status If necessary / the pretreatment system
should be added.

8) When the system goes back to initialization repeatedly during running the system.
(Repeatedly execute self test, auto drain / Reset system)
1 ) Lack of feed water

2) Auto Pressure S/W - Manual Position SET → NEXT → Auto or Manual
3) Pressure S/W error - Pressure S/W error can be occurred by lack of feed water pressure
4) When the Level Sensor is defective or Sensor has been wrongly placed.

9) When the pressure of RO membrane cannot be adjusted as 3 to 7 bar


automatically in Arioso Power Ⅰ
1 ) Adjust it using‘Product/Drain valve’
.
2) Check the‘Auto drain valve’and replace it.

46 H u m a n C o r p o r a t i o n
M E M O
Human Corporation
11, Dogok-ro 62-gil, Songpa-Ku, Seoul, Korea
(304-19, Jamsil Bon-Dong)
TEL. 82-2-418-4182 FAX. 82-2-423-6430
Homepage : www.humancorp.co.kr