Download as pdf or txt
Download as pdf or txt
You are on page 1of 26

See discussions, stats, and author profiles for this publication at: https://www.researchgate.

net/publication/326400881

Dental Implants: An Overview

Article in Dental Update · July 2018

CITATIONS READS

0 29,078

1 author:

Abdulhadi Warreth
Ajman University
31 PUBLICATIONS 321 CITATIONS

SEE PROFILE

All content following this page was uploaded by Abdulhadi Warreth on 14 July 2018.

The user has requested enhancement of the downloaded file.


ImplantDentistry

Abdulhadi Warreth
Najia Ibieyou, Ronan Bernard O'Leary, Matteo Cremonese and Mohammed Abdulrahim

Dental Implants: An Overview


Abstract: Dental implants are widely used and are considered to be one of several treatment options that can be used to replace missing
teeth. A number of implant-supported treatment options have been used successfully to replace a single tooth and multiple teeth, as well
as a completely edentulous jaw. However, as the number of patients who have dental implants is increasing, dental personnel are more
likely to see patients with implant-supported restorations or prostheses. Nevertheless, dental implants may fail as a result of mechanical
complications, such as screw loosening or due to biological causes like peri-implant diseases. As result, dental personnel should be able
to recognize these complications and the factors that have negative effects on the success of such implant-supported restorations or
prostheses. Therefore, a basic knowledge of dental implants is necessary for every dental student, hygienist and dentist.
CPD/Clinical Relevance: Maintenance of implant-supported restorations and prostheses requires long-term follow-ups. It is the
responsibility of the patient to maintain good oral hygiene and also of the dental personnel who look after the patient to ensure a durable
restoration and prosthesis.
Dent Update 2017; 44: 596-620

Dental implants (also known as oral or integration is influenced by several factors, implants is known as peri-implant tissue
endosseous implants) have been used to such as implant material, bone quality and is comprised of soft (mucosa) and
replace missing teeth for more than half and quantity, and the implant loading hard (bone) tissues. The peri-implant
a century. They are considered to be an condition.2,3 soft tissue has similar features to the soft
important contribution to dentistry as As the use of dental implants tissue that surrounds teeth.7-10 It consists
they have revolutionized the way by which has become much more common, dental of a junctional epithelium and connective
missing teeth are replaced with a high personnel are more likely to see patients tissue. The junctional epithelium is
success rate.1-3 This success depends on the who have implant supported/retained attached to the implant and/or abutment
ability of the implant material to integrate restorations. Nevertheless, dental implants surface through a hemi-desmosomal
with the surrounding tissue. However, this are affected by diseases in a similar manner attachment. Connective tissue is present
to teeth and may also fail after several apical to the junctional epithelium and
months or years in service.4-6 Therefore, coronal to the crest of alveolar bone.10
Abdulhadi Warreth, BDentSc, it is not unreasonable to suggest that Connective tissue fibres are found to be
MDentSc(TCD), PhD(TCD), MFD RCSI, the implant and the peri-implant tissue positioned close to the implant surface
Division of Restorative Dentistry and should be examined on a routine basis in but not attached to it, and predominantly
Periodontology, Dublin Dental University a similar manner to that which is carried arranged in a circular manner. Connective
Hospital, Trinity College, Dublin, Najia out for periodontal examination.7 So, when tissue fibres also arise from the crest of
Ibieyou, BDentSc, MDentSc(TCD), a deviation from the norm is found, the alveolar bone and from the periosteum
PhD(TCD), Postgraduate student, treatment may be carried out in practice or and are oriented parallel to the implant/
Institute of Molecular Medicine, Trinity by a specialist, depending on the severity abutment surface and extend towards
College, Dublin, Ronan Bernard of the condition. Accordingly, the dentist the oral epithelium. Thus, the junctional
O’Leary, Fifth Year Dental Science, should be equipped with basic knowledge epithelium and connective tissue form
Matteo Cremonese, Third Year Dental of dental implants. Hence, it is the aim of a protective seal between the oral
Science, Dublin Dental University environment and the peri-implant bone
this article to provide this basic information
Hospital, Trinity College, Dublin and which plays a vital role in the success
which is needed by every dental student and
Mohammed Abdulrahim, BDentSc of the implant treatment outcome. The
dentist alike.
MDentSc(TCD), PhD(TCD), Oral Medicine junctional epithelium and the connective
Department, Faculty of Dentistry, tissue are collectively known as the
Benghazi University, Benghazi, Libya.
Implant-soft tissue interface biologic width, which is comparable to
The tissue that surrounds that found around teeth.11
596 DentalUpdate July/August 2017
ImplantDentistry

Implant-bone interface and categorized into four classes, as described


osseointegration in Figure 2 and Table 1. Some factors which
For dental implants to succeed, affect osseointegration are discussed below
intimate contact between the peri-implant and summarized in Table 2.
bone and the implant surface should
be achieved and maintained. Therefore, Implant placement methods
an integration between the implant Surgical implant placement may
surface and the bone is required for be carried out in one- or two-stage methods
the success of any implant system. This (Figure 3). The one-stage method is also
integration is known as osseointegration, known as the non-submerged method.
and is defined as a direct structural Using this technique, the bone is prepared to
and functional connection between receive the implant. The implant is fitted into
ordered living bone and the surface of the prepared bone (osteotomy). However,
a load-carrying implant.12 Under light the coronal part of the implant is kept
microscopy, successful osseointegration above the bone crest, protruding through
shows direct apposition of bone on the soft tissue, and is exposed to the oral
implant surface (Figure 1). However, when environment during the healing stage.22 The
the bone-implant interface is examined restoration can be attached immediately
using electron microscopy, the implant after the implant placement surgery or may
surface is found to be separated from also be delayed.
the surrounding bone by an amorphous The advantages of the one-stage
layer, a granular electron-dense layer, or a method include:16
layer of uncalcified collagen fibrils13,14 with  The avoidance of a second surgical
a thickness that ranges from 100 nm to procedure;
400 nm.13 Nevertheless, this layer appears  The lack of a micro-gap between the
not to have a negative impact on the implant and the abutment at the alveolar
success of the osseointegration. Inversely, bone crest level, resulting in a less crestal
when the connection between implant bone resorption; Figure 1. A histological image of bone-implant
surface and bone is mediated by a layer of interface. Bone formation around the implant
 The prosthetic procedure is simplified and
labelled with different chelating agents (fluoro-
connective tissue, osseointegration fails to less chair time per patient is required; and
chromes). The implant is the large black area.
occur.5,15,16  A non-loaded, immediate, or delay-
It is important to mention that,
as a result of the absence of periodontal
ligaments between the implant and its
surrounding bone, when the implants are
loaded, they move within the bone due to
bone elastic deformation.6 Furthermore,
osseointegrated implants cannot be
moved orthodontically.
Several factors are
reported to play a role in obtaining
osseointegration.17,18 As an example,
poor bone quality was found to be
associated with a high implant failure
rate when compared with bone of a high
quality.19 Clinical studies have reported
that dental implants in the maxillary Figure 2. The classification of bone according to its quality: Class I (A), Class II (B), Class III (C) and Class
arch (especially for the posterior maxilla) IV (C).
have lower survival rates than those in
the mandibular arch.19 This is usually
attributed to the differences in bone  Type I: almost the entire bone is composed of homogeneous compact bone;
quality between the two arches.20 Bone  Type II: a thick layer of compact bone surrounds a core of dense trabecular bone;
quality, as classified by Lekholm and Zarb21  Type III: a thin layer of cortical bone surrounds a core of dense trabecular bone; and
is based on radiographic assessment as  Type IV: a thin layer of cortical bone surrounding a core of low density trabecular bone.
well as resistance during the implant
Table 1. Classification of bone according to its quality.21
drilling procedure. Accordingly, bone is
July/August 2017 DentalUpdate 597
ImplantDentistry

loaded protocol can be implemented. undue trauma which can negatively affect the implant platform below the bone crest
One of the drawbacks that the healing. However, bone with optimum may be implemented. The countersink
may be associated with this surgical quality and quantity is a prerequisite for allows the placement of the cover-screw
protocol is that the implant is exposed this method to be used. Nevertheless, level with the bone crest. The raised flap is
to the oral environment, which may lead the method can be clinically successful. then repositioned and sutured to conceal
to contamination of the surgical site. Examples of the implants that can be placed the cover-screw and the implant (Figure
Furthermore, the implant may be exposed to using the one-stage technique include 3). After a few months, the second stage
the Solid-Screw Implant® (Straumann surgery is carried out. In this stage, the
UK, Crawley, W Sussex), AdVent Implant® implant site is re-opened, the cover-screw is
(Zimmer, FLA, USA) and Single-stage Implant accessed and then replaced with a healing
 Bone quality and quantity System® (BioHorizons, AL, USA). abutment, which is also known as a sulcus
 Implant shape In contrast, the two-stage former or transmucosal abutment (Figure 4).
 Implant surface macro-structure Afterwards, the healing abutment is replaced
method is also known as the submerged
 Implant micro-structure (roughness) with a provisional or final restoration. This
technique (Figure 3). In this method, two
 Material biocompatibility surgical protocol is suitable for use when the
surgical procedures are carried out. The
 Surgical techniques quality of bone is not optimum and when
first surgery involves installing the implant
 Heat generation during the implant bone graft materials are used in conjunction
into the bone, and a cover-screw (also
placement surgery with the implant. Examples of an implant
known as a sealing-screw) is attached to
 Implant primary (initial) stability
the implant platform. A countersink bone system used for the two-stage procedure
 Implant loading
preparation that allows for placement of include the Fixture MK III® (Nobel Biocare,
Uxbridge, UK), MAX 2.5® Implant (Bicon Inc,
Table 2. Some factors affecting osseointegration.
Boston, MA, USA) and OSSEOTITE® 2 Certain
Implant (Biomet 3i, Maidenhead, UK).
It is important to mention that
the cover-screw is used to prevent tissue
growth into the implant or over its platform.
It is attached to the implant using a screw-
driver with a light finger force. It is essential
to confirm that the cover-screw is fully
seated and no gap is left between the cover-
screw and the implant platform. The cover-
screw has a low profile which facilitates
the suturing procedure and allows the two
edges of the cut mucosa to be brought
close together without undue tension. If
there is too much tension, it may deteriorate
and preclude the healing.23 Conversely, the
healing abutment has a high profile and
protrudes through the peri-implant mucosa
Figure 3. A schematic presentation of an implant placed according to the one-stage (left) and two- to the oral cavity. Therefore, the healing
stage (right) implant placement methods. Note the transmucosal part (the neck) penetrating the peri- abutment is available in different lengths,
implant mucosa in the one-stage method. depending on the distance between the
implant platform and the surface of the
peri-implant mucosa. It is also available in
a variety of diameters, which is selected
according to the implant diameter. The
cover-screw and the healing abutment are
shown in Figure 4.

Implant stability
Implant stability (lack of mobility)
is divided into primary and secondary. The
primary, also known as initial stability, is
achieved during implant placement surgery.
It is believed that primary stability plays a
Figure 4. An image of a cover screw (left) and healing abutments (middle and right)
vital role in reaching osseointegration, upon
598 DentalUpdate July/August 2017
ImplantDentistry

July/August 2017 DentalUpdate 599


ImplantDentistry

600 DentalUpdate July/August 2017


ImplantDentistry

which secondary stability depends. are American Dental Association (ADA) used to overcome angle mismatching
Implant stability is produced approved. The most commonly used implant problems.
by close contact between the implant and systems include Nobel Biocare, Straumann, Implants are also available as
the host bone. The factors that may affect AstraTech, Bicon, BioHorizon, Biomet 3i, hollow and solid. Hollow implants allow
primary stability may be categorized into Intralock, and Zimmer. All are constructed more contact with bone but are weaker
three factors; those related to surgical site on the same basic concepts but there are than solid implants, which makes them
(local) or related to implant or surgical differences in the patented technology and more susceptible to mechanical failure and
method used in placement of the implants. materials. fracture. An example of a hollow implant
Local factors, such as bone quality and In general, dental implants may is the Hollow Cylinder Implants® made by
volume, may affect the degree of bone-to- be classified as a one- or two-piece implant. Straumann and ITI (Basel, Switzerland).
implant contact and consequently affect Irrespective of the implant type
primary stability. As an example, larger bone- The one-piece implant and for descriptive purposes, the implant
to-implant contact fractions were observed In the first type, the implant and usually consists of an implant body and
in bone sites of higher density. The implant the abutment are formed as a single solid neck. The implant body is the part of the
factors include shape, length, diameter unit. In this case, there is no screw-joint implant that is buried in the osteotomy.
and surface texture. For instance, tapered between the implant and the abutment. The coronal part of the implant is denoted
implants lead to higher insertion torque The lack of a screw-joint is considered an as the neck, through which the abutment/
values than cylindrical implants, which advantage as there is no screw-loosening, attachment is connected to the implant.
was considered to be due to the greater dangerous fracturing or micro-motions The coronal part may be smooth (one- and
frictional surface of the tapered implants between the abutment and the implant. two-piece) and placed above the crest of the
and associated with high primary stability The one-piece implants may be used when bone, or roughened (two-piece), in which
(see below). A surgical technique, such as narrow implants are indicated, such as in the the platform is usually placed below or level
that which leads to bone condensation replacement of the maxillary lateral incisors with the crestal bone. When the coronal part
during implant placement surgery or a and lower incisors, or when bone volume is smooth and placed above the crest of the
mismatch between the osteotomy and is limited and the use of standard implants bone and penetrates peri-implant mucosa,
implant diameter (with the implant diameter is not suitable. These types of implants are it is known as the transmucosal part. The
being slightly greater than the osteotomy) installed only with the one-stage implant surface of the transmucosal part is usually
results in satisfactory primary stability.24 placement method. Examples of a one- highly polished and is available in different
Also, the use of implants with self-taping piece implant are the one-piece 3.0 Dental lengths. It may also have a straight or a bevel
blades results in a lower primary stability in Implant® (BioHorizons) and Y-TZP Ceramic profile and may be augmented with micro-
medium-density bone when compared with Implant® (Nobel Biocare). grooves in order to optimize healing around
those without such blades.25 However, this the implants.
issue is contradictory. Placing the smooth (machined)
The two-piece implant
Secondary stability represents part of the implant below the bone crest
The two-piece implant type may lead to its resorption.29 However,
integration of the implant as a result of new
consists of an implant to which an abutment less crestal bone changes were observed
bone formation through its remodelling.27,28
or a restoration/attachment is connected, when the smooth part was located above
Therefore, this stability depends on bone
usually with a screw. It is more commonly the crestal bone level, irrespective of the
activities and factors that influence such
used than the one-piece implant type. With implant type; one- or two-piece implants.29
activities throughout the patient’s life.27
this implant type, both the one- and the Accordingly, it has been recommended that
The general consensus is that peri-implant
two-stage implant surgery protocol can be the smooth-rough border should coincide
bone is in a continuous active remodelling
implemented. with the alveolar bone crest.29
state which maintains osseointegration and
Angled implants in which their
provides secondary stability.27-29
coronal part is angled in relation to the
It is important to mention that,
main implant body are also available. These
Features to consider when
when the implant is inserted into the host
angled implants are useful in the anterior
choosing an implant system
bone, spaces may exist in the bone-implant Five features can be used to
region when placing non-angled implants
interface. These spaces are initially filled describe the dental implant body: shape,
in their optimum position is not possible. An
with blood that comes from injured blood surface macro- and micro-structure, length
example of angled implants is the Co-axisä
vessels, forming a fibrin network which is and diameter. These features are important
implant (the Southern Implants, UK) in
the important step towards the formation of when an implant system is chosen.
which the neck is at an angle to the long
osseointegration.
axis of the implant body. It is useful to use
when the long axis of a prospective implant 1. Shape (geometry)
Dental implant types is not along the long axis of the potential Implant shape may generally
In the worldwide market, restoration. An angled abutment, such as be tapered or parallel (straight-walled).
there is a wide range of dental implant Regular Neck synOcta® angled abutment The tapered type in general has more
systems available, but only a few brands (Straumann) is also available and can be primary stability than the parallel type.30
July/August 2017 DentalUpdate 601
ImplantDentistry

The use of tapered implants results in lateral edentulous space and in softer bone, such
compression of bone and increased stiffness as type IV (Figure 2), where primary stability
of the interfacial bone, which is reported is not always easy to achieve.32 They may
to increase the implant primary stability.13 also be used immediately or early after
Tapered implants were found to require a tooth extraction.32,33 The use of a tapered
higher insertion torque and less insertion implant with a wide platform achieves
time than parallel implants. A higher a satisfactory emergence profile of the
insertion torque gives a better implant restoration.
Figure 5. A representation of the most commonly primary stability.31 Tapered implants are also
used implant threads: V-shaped thread (left); used to avoid damaging the converging
square thread (middle) and a reverse buttress 2. Surface macro-structure (threads)
roots of adjacent teeth that bind the The implant macro-structure
(right).
is represented as threaded or non-thread
(thread-less). The threaded type is the
most commonly used implant design.
The threads are usually incorporated into
the implant design to improve the initial
stability and dissipate interfacial stress in
a more favourable way. As the threaded
implants provide better mechanical and
biological outcomes, non-thread implants,
such as cylinder (press-fit) implants, are
less likely to be used and are replaced by
the threaded type. Thread features such as
thread depth, thread thickness, face angle,
pitch and helix angle are considered to
be factors that determine the functional
thread surface and affect the biomechanical
load distribution of the implant.
Figure 6. Bone resorption at alveolar crest occurs after tooth extraction which may preclude the use of There are three thread shapes
a long implant as the crestal bone has to be trimmed down to maintain at least one millimetre of bone which are most regularly used when a
buccally and lingually at the bone crest region. dental implant is described (Figure 5). These
are V-shaped, square-shaped or reverse
buttress.34,35 An animal study conducted
Implant length: a long implant should be considered whenever the condition permits.
by Steigenga and colleagues36 revealed
Implant diameter: ideally, the implant should be approximately the same diameter as the effects of thread type on peri-implant
the root of the tooth it is replacing. bone formation. The study showed that
a. Wide implant: implants with a square thread design had
i. Poor quality bone; significantly more bone-implant contact
ii. Limited ridge height with adequate mesio-distal and bucco-lingual width; and and greater reverse-torque measurements
iii. Immediate implant placement (after tooth extraction). than observed when the V-shaped and
b. Narrow implant: reverse buttress thread designs were tested.
i. Used to replace maxillary lateral incisors or mandibular incisors; A threaded implant may also
ii. Limited edentulous space; be classified as a self-taping or pre-taping
iii. Limited ridge width (to avoid ridge augmentation surgery); implant.37 A self-taping implant is an
iv. When it is not possible to achieve good emergence profile with a wide implant body; implant which is designed to make its
and own threads as it is being placed into the
v. Converging adjacent tooth roots. prepared osteotomy. On the other hand, in
pre-taped implants, threads are prepared
Tapered implant: on the surface of the osteotomy using a
i. In type IV bone, where primary stability is difficult to achieve; tap drill (taper). The produced threads will
ii. Narrow or concave bone; accommodate the threads of the implant.
iii. Converging adjacent roots; and The pre-taping method is sometimes
iv. Iimmediate and early implant placement. recommended, such as in the case of dense
bone (type I and II) (Figure 2). However, pre-
Table 3. Some implant features that should be considered when an implant is selected, and their
taping implants achieved lower primary
indications.
stability than the self-taping implants.38
602 DentalUpdate July/August 2017
ImplantDentistry

3. Surface texture (micro-structure) function of contact between the implant may fail at an earlier stage than standard
Implant surface texture describes surface and bone, the longer the implant, implants,20 as peak failure rates of short
the roughness of the implant surface. the greater the surface contact and primary dental implants were 4−6 years, and 6−8
Therefore, the implant surface is either stability. However, the increase in implant years for the standard implants.52
smooth (machined) or can be of a variety stability does not occur linearly to the It is important to note that bone
of roughness. A rough-surfaced implant increase of the implant length. For instance, resorption following tooth extraction may
has a larger surface area than that of its a 10 mm implant has about 30% more result in the thinning of the alveolar bone
counterpart smooth implant. It is found surface area than a 7 mm implant, while a 13 crest, which may preclude placement of
to be associated with positive healing of mm implant has 20% more surface area than an implant with an adequate length and
peri-implant tissue and encourages the a 10 mm implant.46 diameter, as shown schematically in Figure
formation of osseointegration.39 The increase The bone of the edentulous 6. Therefore, bone mapping and a CT-scan or
in surface area distributes forces to which ridge may not be sufficient for placing an Cone-Beam Computed Tomography may be
the implant is exposed in a more favourable implant with the optimum length. Therefore, required.
manner. It also provides better primary several techniques have been suggested to
stability than that attained when the implant compensate for the deficiency in the residual 5. Implant diameter
surface is smooth.40 Histomorphometric and ridge, either before or simultaneously with The implant diameter is
removal torque studies with roughened implant placement. Among these methods measured from the crest of the widest
implant surfaces have revealed greater are guided bone regeneration, block grafts, thread to the same point on the opposite
bone apposition41 and higher removal sinus lifting procedures, inferior alveolar side of the implant.53 According to the
torque values than implants with smoother nerve repositioning methods, and bone diameter, implants may be classified as mini
surfaces.42 distraction.47 These surgical methods are when diameter is ≤2.7 mm; narrow when the
In general, two methods for the successful and can be used to increase bone diameter is >2.7 mm but ≤3.75 mm; regular
alteration of implant surface texture have height.47 However, they are not without risks when it ranges from 3.75−5 mm; and wide
been described in the literature: subtractive and may lead to several complications and when the diameter is >5 mm.
and additive methods. In the subtractive undesirable treatment outcomes.5,47 This may The implant diameter plays
method, the implant surface is roughened encourage the dentist and patient to avoid an important role in the success of oral
by removal of its surface materials usually such surgical methods and to use short implants and has a major impact on the
by blasting and/or acid etching.42-44 In the implants, therefore the implant is installed implant’s ability to withstand occlusal load.54
additive method, a biocompatible material, with less invasive surgical procedure and Selecting an implant of a suitable diameter
such as titanium or hydroxyapatite, is added the cost is reduced. Nevertheless, when a is governed by the dimensions of the
to the surface42 (see below). Some examples short implant is used, factors that affect the edentulous space (bucco-lingual and mesio-
of rough surface implants include: grit osseointegration, such as implant shapes, distal) (Figure 7), as well as the bone quality.
blasting with titanium oxide produced by surface texture, and thread designs, should Moreover, it is also affected by the type of
Astra Tech (Mannheim, Germany); Sand- be carefully selected to achieve a satisfactory tooth being replaced.
blasted Large-grit Acid-etched (SLA®) long-term outcome.47,48 However, earlier An increase in the diameter of
implants from Straumann; Acid-etched studies have reported that shorter implants an implant is associated with an increase
Implants® from Biomet 3i; and Plasma- are unpredictable and fail more frequently in its surface area. For instances, increasing
sprayed® (molten titanium sprayed on the than longer implants.46,49 In addition, the diameter in a 3 mm implant by 1 mm
implant surface) produced by Straumann longer implants had statistically higher increases the surface area by 35% over
and Frialit (Dentsply?, Weybridge, UK). survival rates when compared with shorter the same length.55 Also, a 3.75 x 10 mm
It is important to note that, if implants.50 For instance, it has been reported implant has 61% less surface area than a 6
the rough implant surface is exposed to the that survival rates after two years were mm diameter implant of the same length.33
oral environment, it may encourage plaque 93.1% for 5 mm implants and 98.6% for 9.5 Furthermore, an increase in the diameter
accumulation and interfere with its removal, mm implants.51 Furthermore, short implants and a change in the threads may lead to an
and subsequently may induce peri-implant
disease (see below).6,45

4. Implant length
Implant length is determined by
the distance between the top surface of the
implant platform and the apex. In general,
the length of the standard implant ranges
from 7−18 mm.33 Selection of an implant
of the required length is governed by the
Figure 7. The implant should be placed in the site that was previously occupied by the tooth being
available vertical bone height, width and
replaced, and surrounded by an adequate amount of bone (right). Two implants may be used to
quality which will accommodate the implant
replace a molar tooth, which results in the dissipation of the occlusal forces in a satisfactory manner.
(Figure 6). As implant primary stability is a
July/August 2017 DentalUpdate 603
ImplantDentistry

increase in the implant surface area of more implant is placed deeply below the crest of and mandibular incisors. They are also
than 300%. This increase in the surface area bone, the crown height is increased, which suitable when bone quantity is insufficient,
may lessen stresses to the crestal bone areas may lead to mechanical failure of implant or when the roots of adjacent teeth are
and reduce both crestal bone loss and early components and compromise aesthetic converging. They may also be used with a
loading implant failure.55 treatment outcomes. When the implant is removable implant-supported overdenture.
It is important to mention that, placed more superficially, restoration may be However, the use of an implant with a small
when the implant is installed, it should be in deemed impossible and aesthetic treatment diameter is not without disadvantages,
close contact with the surrounding bone of outcome is also compromised.33 such as mechanical failure of the implant
not less than 1 mm thickness on its buccal When a molar tooth is replaced, component. Furthermore, obtaining a
and lingual surface, and preferably 1.5 mm the use of two implants may be an option, as good emergence profile of the restoration
or more between the implant surface and its dissipation of occlusal loads are favourable. may also be a problem. Hence, a detailed
adjacent tooth (Figure 7). For instance, when However, placement of implants close to examination of each patient’s condition
an implant of 4 mm is selected, the bucco- each other is associated with difficulty in should be taken before a specific implant is
lingual and mesio-distal dimensions of the obtaining an optimal emergence profile, selected, and alternative treatment options,
edentulous space should be a minimum of interferes with oral hygiene and leads to such as a fixed (conventional or resin-
6.0 and 7.0 mm, respectively. However, it has chronic inflammation and bone resorption. bonded) prosthesis, may be considered.
been suggested that, in the aesthetic zone, Short and wide implants may It is important to distinguish
maintaining a minimum of 3 mm of bone be used to compensate for the decrease in between the implant diameter and platform
between adjacent implants is beneficial, the vertical bone height of the edentulous diameter as they may not be equal. The
as bone height as well as the inter-dental space when surgery cannot be considered. implant platform represents the part of the
papilla are more likely to be maintained.11 They may also be used when the quality of implant that is connected to the prosthetic
Consequently, implants with a smaller the bone bed is not optimal.56 Wide implants (abutment) counterpart. Table 3 displays
diameter at the implant-abutment interface can be used to increase implant stability,57 examples of implant features that should be
may be used when multiple implants are to thus improving stress distribution within considered when an implant is selected.
be placed.11 the surrounding bone.47 Furthermore, the
The diameter of the roots use of a wide diameter implant may reduce
is usually estimated at 2 mm apical to the stress on the retained screws. Wide
Implant materials
the cemento-enamel junction. With this implants are also used for the replacement of The most commonly used
measurement, an implant with a diameter posterior teeth and immediately after tooth materials in dental implants are either bio-
that matches, or is slightly smaller than, the extraction (Table 1).58 inert, such as commercially pure titanium (Cp
tooth being replaced is selected. In order Several situations do not allow Titanium) and titanium alloy, or bio-active
to obtain a restoration with an optimal the use of wide diameter implants59 and ceramics such as hydroxyapatite, tri- and
emergence profile, the implant platform is narrow implants are an alternative. For tetra-calcium phosphate and bio-glass.18
usually placed at about 2 mm apical to the example, narrow implants are suitable For more than five decades,
cemento-enamel of the adjacent teeth. If an for replacing maxillary lateral incisors titanium was the most commonly used
material in dental implants due to its bio-
compatibility, as well as its mechanical and
physical properties, such as resistance to
corrosion, high strength and low weight60
Depending on its oxygen content, Cp
titanium may be categorized into four
grades; grade I contains the least oxygen
while grade IV contains the most (0.18%
versus 0.4%).18 Titanium alloy consists of 90%
titanium, 6% vanadium, and 4% aluminium
and is classified as grade V.34
Titanium is a non-noble metal
which has the ability to form a very adherent
self-repairing and protective surface oxide
layer, which prevents further titanium
corrosion. This layer forms immediately
when the titanium is exposed to oxygen.
The formed oxide layer on Cp titanium is
similar to that which is formed on titanium
alloys.34 Titanium dioxide (TiO2) forms the
Figure 8. A schematic representation of the screw-joint connections: the external connection and the
main constituent of this oxide layer, however,
butt joint (left) and the internal connection and the slip joint (right).
other oxides, such as Titanium oxide (TiO)
604 DentalUpdate July/August 2017
ImplantDentistry

July/August 2017 DentalUpdate 605


ImplantDentistry

and Titanium pentoxide (Ti2O5) may also found that hydroxyapatite mechanical failure accommodates the prospective restoration.
exist. Incorporation of other chemical occurs primarily at the interface between
elements, such as carbon, traces of nitrogen the metal substrate and hydroxyapatite coat Abutment-implant connections
or chlorine, into the oxide layer have been (adhesive failure), irrespective of the implant (interfaces)
reported.61 design. This may have a negative effect on
When an implant is put to
The release of metallic ions implant osseointegration.71 Nevertheless,
function, it is connected with the restorative/
from the titanium implant surface may the risk for hydroxyapatite-coat degradation
prosthetic components. The connection
occur and increase as the implant surface and loosening (delamination) are still a
type can be classified as internal or external.
area increases.62 It has been suggested remaining concern.
In the internal connection systems, the
that ionic release may interfere with the With improvement in technology,
apical part of the abutment is inserted into
normal peri-implant bone mineralization ceramic materials are extended for use
an access hole in the implant platform. In
and remodelling, which could lead to as implant substrates. This is because
the external systems, a protrusion located
the failure of the implant.63 Furthermore, ceramics such the yttrium-stabilized
titanium release may induce hypersensitivity tetragonal zirconia polycrystalline has above the implant platform is inserted into
in susceptible patients, which may have an improved mechanical properties, superior a recess in the apical part of the abutment
undesirable impact on implant success.64 wear and corrosion resistance, with a high (Figure 8). The connection is also classified as
However, this issue is still debatable flexural strength. These characteristics a slip joint; when there is a space between
and more clinical and further laboratory may make them a potential alternative opposing mating surfaces, and a friction
investigations are required.24,64 Nevertheless, to conventional titanium implants for fit when such space does not exist. The
available literature indicates that Cp titanium supporting overdentures.72 Three types of connection may be further categorized as a
has a long-term successful performance. zirconia-containing ceramic systems are bevel (conical) joint or a butt joint (Figure 8).
In addition, the surface of the titanium most commonly used in dentistry; yttrium- The connection may have
implant, which was previously contaminated stabilized tetragonal zirconia poly-crystals, an anti-rotational component, such as
in the peri-implantitis case, was found to alumina-toughened zirconia and zirconia- hexagonal, octagonal, cone hex, cylinder
reintegrate with bone which was treated to toughened alumina. However, these non- hex, cam tube and pin/slot or be without
remove the contaminant.65-67 metallic materials are expected to replace Cp an anti-rotational device, such as a cone
Cp titanium and titanium alloys titanium and its alloys.73 Nevertheless, based (Morse taper). The function of the anti-
can make up the entire implant or can be on their systematic review of literature, rotational component is to stabilize and
used as a substrate to which a coating of Andreiotelli and colleagues74 concluded prevent abutment rotation.77,78 Likewise,
bio-active material, such as hydroxyapatite, that ceramic, in particular zirconia, implants the connection usually has a screw but is
is attached. are not yet suitable as an alternative to sometimes screw-less and relies entirely on
To speed up the healing titanium implants. Nevertheless, they the friction fit for its stability, such as Bicon®
process and osseointegration, implant potentially could be a successful material dental (Bicon Inc, Boston, MA, USA).
surfaces are coated with ceramics.68 The for use in implants, but this has not yet The first implant connection
ceramics may be bio-active, such as calcium been supported by clinical investigations.75 type used with a dental implant was
phosphates, or inert, such as aluminium However, ceramics such as zirconia are used described by P-I Brånemark.12 It was an
oxide and zirconium oxide. Examples of nowadays as abutments and crowns as they external hex, therefore consisting of six
calcium phosphate coating materials are have good clinical outcomes.73 sides, each two adjacent sides make a
hydroxyapatite and fluorapatite.68 The It is not unreasonable to 60-degree angle and had a height of 0.7
bio-active ceramics are reported to act as conclude that the prospective implant mm. The hex was originally used to carry
osseoinductive materials which encourage should be selected carefully and a restorative and insert the implant into the prepared
and accelerate bone apposition around the driven approach should be implemented to host bone (osteotomy). The hex was not
implants. Furthermore, coatings that have avoid an unwanted result.76 Thus, thorough aimed for use as an anti-rotational device,
similar properties to that of the extra-cellular investigation should be carried out to as the implants were mainly used to restore
matrix provide a favourable environment for guarantee the best possible outcome. The completely edentulous dental arches with
osteoblasts, osteoclasts and their progenitor edentulous area should be viewed in three implant-supported overdentures with
cells, that are responsible for the healing dimensions: mesio-distal, bucco-lingual and multiple implants. Consequently, rotational
of bone.69 Therefore, an early and strong corono-apical. The mesio-distal dimension displacement of the overdenture was not
implant stability is achieved and the risk of of the edentulous space should also be an issue. However, as the use of dental
implant failure is reduced.69 thought of as two interrelated spaces implants progressed and extended for use
Ceramics are initially used in the (inter-radicular and restorative). The inter- in replacing single and multiple missing
additive methods in which ceramic coatings radicular space holds the implant and can teeth, the use of a guiding index and an
are added to the metal implant. However, be found between the roots of the two anti-rotational device is needed. To fulfil this
high bond strength between the coating adjacent. Hence, a precise radiograph image requirement the original external hexagonal
material and the substrate is required to of the area is important. The restorative connections were modified and are now
withstand functional stresses and to avoid space should be carefully investigated as it available in different heights including
fragmentation of the coating materials.70 It is extends between the two adjacent teeth and 0.9, 1.0 and 1.2 mm and with various sizes.
606 DentalUpdate July/August 2017
ImplantDentistry

Furthermore, several types of internal connection is used with a narrow implant,


connections were also introduced and are the connection is exposed to vertical or
widely used nowadays. oblique loads. Although the screw itself may
In general, when the connection be protected from loading, the implant neck
is an internal type, the occlusal load is may not be able to resist such a load and will
usually dissipated through the implant mechanically fail80.81 as most of the occlusal
body and the screw is more likely to forces are transferred to the implant walls.81
Figure 9. Measurement of rotational freedom. A be protected from the imposed load.
passive fit of the abutment (blue) into a recess Loose screws were reported to occur less Screw-joint
(hexagonal) in the implant platform (a dotted frequently with internal connections than
When the implants and the
circle). The space between the two components is with external ones.79 However, the implant
restoration/prosthesis are connected
represented by the red area. The rotational freedom neck should be strong enough to resist
degree during abutment rotation is indicated by the together by a screw, the connection is
such loads. Nevertheless, when the internal
letter ‘A’. known as a screw-joint.16,77,82 For example,
when the single restoration (crown) is screw-
a b retained, one screw-joint is usually found to
connect the restoration to the implant. When
the restoration is cement-retained, there
is also one screw-joint, but it is between
the abutment and the implant (see below).
The screw-joint is also found with the fixed
implant-supported prosthesis in a similar
way as that described for the cement- and
screw retained single implant-supported
restoration. In the fixed implant-supported
overdentures (FISOs), there is a screw-
joint between the frame-work and the
implants, whereas in the removable implant-
supported overdentures (RISOs), there is a
screw-joint between the attachment system
and the implant.6,16 The attachment systems
Figure 10. An intra-oral radiograph showing a single implant-supported crown replacing the right are discussed later in the article. In some
second molar (a). The cuspal inclinations are lowered and flattened, but the occlusal table is widened situations when a screw-retained restoration
which creates a cantilevering effect and exposes the restoration, the screw and the implant to high
is used, there may be two screw-joints: one
tipping forces that may lead to their mechanical failure. A diagram of an implant-supported restoration;
between the implant and the abutment,
the implant is oriented so occlusal loading is directed along its long axis (b).
and one between the abutment and the
restoration/prosthesis.
1. Implant-abutment interface design/type. When the screw is tightened,
2. Rotational freedom (misfit). there are two opposing forces that act on
3. Manufacturing allowances (tolerance). the implant platform and the abutment or
4. The settling (embedment). restoration/attachment that form the joint.
5. Repeated opening and closing of the screw. One of these forces tries to hold the joint
6. The applied torque value: over and under torqueing the screw. together and is known as the clamping
7. Loading of restoration. force. The other force is called the separating
8. Prefabricated metal- and costume-made cylinders. force as it tries to disengage the screw-joint
9. The casting process: components away from each other. Hence,
a. Casting alloy;
the two forces are acting against each other.
b. Investment; and
As a tightening torque is applied to the
c. The finishing/polishing method.
screw, a tension (pre-loaded) is generated
10. Screw design and materials:
in the screw. Consequently, the screw shank
a. Shank or shank-less screws (a shank-less screw is usually less resilient than that with
and threads are tense and an elastic recovery
a shank);
is generated, thus creating the clamping
b. Shape and diameter of screw’s head;
force between the mating surfaces.16,77,82
c. Materials from which a screw is made of such as gold, titanium and gold-coated
To obtain an effective clamping
screws.
force, the tension created in the screw
material should be less than that of the
Table 4. Factors that affect screw-joint stability.
material’s elastic limit (Young’s modulus)
July/August 2017 DentalUpdate 607
ImplantDentistry

so no permanent plastic deformation or the conventional gold alloy and titanium occurs, which consequently leads to screw
screw fracture occurs. Maximum screw-joint alloy screws.86 Likewise, higher pre-loads loosening.94 Inversely, too little torque or a
stability can be achieved with a maximum were associated with gold-coated screws lower torque value which cannot produce
pre-load when the proportional limit of the when compared with that obtained from the required screw pre-loading needed to
screw is approached. Thus, to obtain this, screws made of uncoated gold or titanium hold the mating surfaces together exhibits
the applied torque should be 75% of the alloy for all insertion torques, as well as when greater micro-motion at the screw-joint,95
torque required to cause screw permanent the screws were re-tightened.87 which consequently causes screw loosening
deformation. In order to hold the implant Manufacturing tolerance is and may lead to its fatigue and fracture.
components together, a maximum clamping another factor that affects the screw- Therefore, it is vital to use the manufacturer’s
force and a minimal separating force are joint stability. It is defined as unplanned recommended tightening torque, which
required. Therefore, the clamping force deviations from the theoretical dimension should be within the elastic range of the
overcomes the separating force. of the shaft and its mating recess as some screw’s materials, as mentioned earlier.96,97
deviations from a perfect fit are expected, It is also essential to ensure consistent
Factors affecting screw-joint but not planned. Hence, this indicates an tightening torque values are applied.
stability insignificant value of misfit between the Therefore, torque gauges (control) should
matting surfaces. This misfit allows for what be used and manual torqueing should be
Lack of screw-joint stability is is known as rotational freedom (play) to avoided.91 It is also important to calibrate
reflected in loosening of the screw. It is occur. The rotational freedom is calculated the torqueing devices to obtain consistent
considered as one of the most common by the formed angle between the clockwise torqueing.98
problems associated with the use of implant- and anti-clockwise rotation of the anti- Torqueing the screw should be
supported restorations.83 One of many rotational components of the screw-joints carried out carefully and a counter-torque
factors that play a role in the stability of (Figure 9). The rotational freedom may vary device should be used to avoid disturbing
the screw-joint is the friction coefficient of from 1.6 to 5.3 degrees.88 The most stable the osseointegration. Hence, the use of a
the materials used in the fabrication of the and predictable screw-joint may be expected counter-torque device is recommended as
implant components, such as the abutment, when the rotational freedom is lower it reduces transmission of the tightening
implant and screw. The friction coefficient than two degrees.89 Hence, the produced torque to the implant-bone interface. On
has an effect on the generated pre-loading. rotational freedom affects the stability of the average, about 90% of the recommended
Tightening torque and consequently the screw-joint. pre-load tightening torque is transmitted
developed pre-load is inversely affected by Furthermore, the presence of to the implant-bone interface when the
the friction between the mating surfaces.84 a micro-roughness on the implant and counter-torque device is not used. This value
In general, during screw torqueing, friction abutment mating surface, which is worn is reduced to only 10% when the counter-
occurs between the implant surface and the away as a result of screw torqueing, leads torque device is used.94
opposing abutment surface, between the to what is called settling (embedment Overloading of the restoration
head screw and the abutment surface and relaxation). Consequently, part of the may lead to screw loosening and failure.
between the screw threads (male) and the clamping force is lost and the screw Therefore, the occlusion should be adjusted
implant threads (female). As such, when a becomes loose. The mean loss of pre-load and occlusal forces should be directed along
screw is tightened, only 10% of the torque may be up to 40% of the original pre-load the long axis of the implant, whenever
is converted into screw pre-load, while the value 15 hours after screw torqueing.90 possible (Figure 10). This can be achieved
other 90% of the tightening torque is lost as To reduce the settling effect, it has been by construction of a restoration in which
friction.84,85 In order to maximize pre-loading, suggested that the implant screws should its occlusal morphology is constructed
the friction between mating surfaces should be retightened ten minutes after the initial according to the mechanical principals
be reduced. This can be achieved by coating torque application as a routine clinical that favour this concept. For instance, the
the mating surfaces with other materials, procedure.91.92 All screw types were reported cuspal inclination should be flattened and
such as carbon film or the screw with to display some decline in pre-load with the incisal guidance made shallow to avoid
tungsten carbide. This process is known as repeated tightening. This decline occurs bending moments caused by the lateral
dry lubrication and the coating material is irrespective of the insertion torque and component of the occlusal forces.99 The
denoted as a dry lubricant. Both carbon and abutment type.87 As screws lose pre-load occlusal table of the prospective restoration
tungsten carbide coatings were reported to following placement, their re-tightening may be reduced by 30−40% of the tooth
reduce the friction coefficient and improve is required from time to time during the being replaced (Figure 10) and cantilevering
pre-loading.84 Torq-Tite® abutment screws restoration’s life. the restoration should be avoided. Use
(Nobel Biocare, Uxbridge, UK) are made The screw pre-load should of an occlusal splint is recommended for
of titanium alloys and are coated with a be high enough to maintain the joint patients with parafunctional habits such as
carbon layer and Gold-Tite® abutment screws integrity and reduce the possibility of the bruxism. The implant should be placed in
(BIOMET 3i) are titanium screws with a screw loosening and fracturing.93 However, the site that was previously occupied by the
gold-plated surface. Both screw types were when excessive torque is applied, slippage tooth being replaced, and surrounded by
found to be associated with lower friction between the screw threads (male) and an adequate amount of bone (Figure 7). It
coefficients and greater pre-load values than the implant internal threads (female) should also be oriented along the long axis
608 DentalUpdate July/August 2017
ImplantDentistry

July/August 2017 DentalUpdate 609


ImplantDentistry

of the tooth being replaced and within implant interface, which may also help in maxilla. Therefore, when a patient whose
the occlusal table. However, when a molar maintaining the crestal bone level. missing teeth were replaced with an implant-
tooth is replaced, the use of two implants A recent meta-analysis,102 supported restoration attends the dental
may be considered in order to dissipate the including 13 human randomized clinical clinic, one of the following restoration/
occlusal loads satisfactorily, as mentioned trials (RCTs), has shown a significantly less prosthesis is usually present:
earlier (Figure 7). mean crestal change at platform-switching  An implant-supported single restoration
Some of the other factors that implants, compared with when the implant (crown) (Figure 11);
may affect the screw-joint stability are platform dimensions matches the abutment  A fixed implant-supported prosthesis;
displayed in Table 4. (0.49 mm versus 1.01 mm). However, the  A removable implant-supported partial
use of platform-switch did not preserve denture (Figure 12); and
Platform switching concept the crestal bone better than when the  A fixed or removable implant-supported
This concept was based on switching concept was not used, when prosthesis (overdenture) (Figure 13).
clinical observations where the implant thin mucosal tissues on crestal bone were Treatment options for
platform diameter was wider than the present.103 Furthermore, the stress within the replacement of missing teeth with dental
abutment.100 It is assumed that, when screw-joint was found to increase when the implants are shown in Figure 14.
this principle is used, the crestal bone platform-switching concept is implemented.
loss after implant placement is less than This may lead to failure of the screw-joint
when the implant platform and the connection.104,105 Therefore, this concept
abutment pose a similar diameter.100 This should be used with substantial care.
concept is theoretically explained on the
bases of moving the micro-gap between Types of restorations/
the platform and the abutment inward prostheses for missing teeth
from the outer edge and consequently Implant-supported restorations
away from the bone.101 It also results (prostheses) may be used to replace a
in an increase in horizontal soft tissue single or multiple missing teeth, as well
dimension, which may protect the bone as completely edentulous mandible and
crest and limits its resorption.102 It also
shifts the stress between the implant and
abutment away from the cervical bone- a

Figure 11. A clinical image of a missing upper


right centre incisor (1.1) replaced with a single Figure 13. Clinical pictures of an upper
cement-retained, implant-supported crown. The Figure 12. (a, b) Clinical images of multipale edentulous maxilla restored with a RISO. (a) Four
abutment (a) and the restoration (b) is made missing maxillary teeth restored with a partial dental implants placed in the anterior region.
of porcelain fused to metal. The papilla failed denture which gains its support/retention from (b) The implants are connected with a CAD/
to fill the inter-dental space on the mesial and the teeth, alveolar ridge as well as from an CAM designed and fabricated bar. Four locator
distal aspect of the restoration. This may have a implant placed in the right canine region. The attachments (matices) are attached to the bar.
negative effect on the aesthetic outcome if the fitting surface of the denture showing the patrix (c) The fitting surface of the RISO showing the
patient has a high lip-line. of a locator attachment. patrices of the attachment.

610 DentalUpdate July/August 2017


ImplantDentistry

1. An implant-supported single restoration 3. A removable implant-supported prosthesis reduced. However, this type of prosthesis is
(crown) In certain clinical situations, more expensive than removable ones. It also
When a single tooth is replaced, multiple missing teeth cannot be restored requires more implants to support and retain
the restoration is usually either cemented with a fixed implant-supported restoration. the prosthesis.
to the abutment or screwed to the implant Instead, they are restored with a removable FISOs are of two basic types:
(Figure 11). This is known as a cement- prosthesis which is fundamentally similar to hybrid and porcelain fused to metal.
retained restoration and a screw-retained that which is used in replacing a completely The hybrid prosthesis is made of a metal
restoration, respectively. As mentioned edentulous jaw with a removable implant- substructure, acrylic and denture teeth. The
earlier, in the cement-retained restoration, supported overdenture (RISO) (Figure 12). In porcelain fused to metal prosthesis is made
the abutment is attached to the implant this case, in addition to the available teeth, of a metal substructure and porcelain in a
body through a screw-joint and the one or more implants with attachment similar way to that used in the fabrication of
restoration is cemented to the abutment systems are usually used. The attachment the conventional porcelain-fused-to-metal
in a similar fashion to that which is used systems are discussed later in the article. restoration. It is more expensive than the
in the conventional crown. Therefore, the hybrid and is difficult to make, but it is the
abutment is used to connect the crown 4. Implant-supported overdenture for better option when the vertical restorative
to the implant. In the screw-retained completely edentulous jaws space is limited.
implant restorations, the restoration and When the jaw is completely Conversely, the RISOs are
the abutment are a single unit which edentulous, there are two treatment options removable prostheses that can be removed
is attached to the implant directly by a for its restoration; namely a fixed or a and replaced by the patients. They are used
screw.16,106,107 removable implant-supported overdenture in combination with attachment systems
(FISO or RISO). A FISO is when the prosthesis (see below).
2. A fixed implant-supported prosthesis (fixed is permanently fixed to the implants through The number of implants
bridge) screw-joints between the prosthesis and the used with the RISOs may be reduced.
This is when multiple teeth are implants.108 This is so it cannot be removed For instance, in the case of edentulous
missing and replaced with a prosthesis by the patient. The prosthesis is supported mandible, the number may be reduced to
that cannot be removed by the patient. In by several implants (usually four or more). two implants, which are usually placed in
principle, this type of restoration resembles When such prostheses are indicated, it is a the anterior region of the mandible. The
that described for a single-implant favourable option for many patients. The two-implant supported overdenture option
supported crown: cement- or screw- volume of the prosthesis, and consequently is recommended as the first-choice standard
retained restorations. the tissue coverage by the prosthesis, are of care for an edentulous mandible.108-110

Missing tooth Missing multiple Missing multiple teeth Completely edentulous jaw
adjacent teeth but not adjacent

Single Fixed Removable implant-


implant- implant- supported denture
supported supported
crown bridge

Fixed implant-supported Removable implant-supported


overdenture (FISO) overdenture (RISO)

Hybrid

Screw-retained Cement-retained
Porcelain fused to metal

Figure 14. Treatment options for replacement of missing teeth with dental implants.

July/August 2017 DentalUpdate 611


ImplantDentistry

When two-implant supported overdentures of crowns or fixed prostheses (bridges), abutment types can be found in Table 5.
are used, the attachments permit movement and that are supported by implants, may
of the overdenture during function and be divided into two types, depending on Screw-retained restorations
allow the mucosa of the residual ridge to how they are connected to the implants; In this case, the retention of
be involved in dissipating the imposed cement-retained and screw-retained. As the restoration relies on the retaining
force. Therefore, it is important to note mentioned earlier, in the cement-retained screw. Nevertheless, the restoration can be
that, in order to obtain good support restoration the abutment is required to removed and/or replaced when required,
from the residual ridges, the RISOs should connect the restoration to the implant, without damage or need for a new
extend to cover the supporting tissues in while in the screw-retained restoration restoration. The adaptation between the
a similar fashion as that covered when the the abutment and the restoration form restoration and the underlying implant is
conventional complete denture is used. one unit. In addition, there are five types usually better than that in the case of its
of abutments which are available for use cement-retained counterpart. It can be used
The abutments in single and fixed implant-supported when the vertical restorative space is limited
The restorations that consist restorations.16,35 A summary of these as the retention depends on the screw, but

Custom-made abutments
 They are made of a plastic/wax pattern with/without a metal-machined interface ring;
 The pattern is made (wax) or adjusted (plastic) to the required form, shape and angle;
 The pattern is then used to create a metal abutment in a similar procedure to the conventional lost-wax technique;
 An abutment plastic/wax pattern is attached to the implant analogue, which is submerged in a working cast;
 The restoration is then made to fit the abutment also in the conventional method;
 UCLA plastic patterns are an example of these types of abutments;
 They require an impression of the implant platform.

Pre-machined (prefabricated/ready-made) modifiable metal abutments


 They are prefabricated abutments;
 They are adjustable and modifiable intra- and extra-orally;
 They cannot be used when the implant is placed in an improper position or with improper angulation;
 An impression of the abutment, not the implant, is taken using a manufactured impression coping;
 The conventional crown and bridge procedures are used when provisional or final restorations are made.

Pre-machined (pre-fabricated/ready-made), non-modifiable metal abutments


 They are pre-fabricated abutments that cannot be modified or altered;
 The abutment that is suitable for the specific clinical condition is selected;
 The abutment is attached to the implant body;
 An impression of the abutment, not the implant, is taken using a manufactured impression coping;
 The conventional crown and bridge procedures are used when provisional or final restorations are made.

All-ceramic abutments
 They are made entirely of ceramic;
 They are available in ready-made or customizable forms;
 They are indicated for use in cases when aesthetics are essential, and when thin biotype gingiva exists so that metal show through is
avoided.

CAD/CAM milled abutments


 They are made from a block of titanium or ceramic;
 An implant platform level impression may be required depending on the manufacturers;
 A working cast is fabricated then scanned optically to generate exact 3D images of the region;
 The information is sent to the milling machine to produce the abutment;
 It eliminates certain negative factors that may be associated with the conventional method of abutment fabrication, such as an
improper fit and incorporation of porosity;
 This type of abutment is more expensive than the other abutment types.

Table 5. Different abutment types.

612 DentalUpdate July/August 2017


ImplantDentistry

is contra-indicated when mouth opening order to remove them. The removal of excess of attachment systems used for RISOs are
is limited, as the use of the different tools cement may be not possible, which may displayed in Table 6.
required for screwing and torqueing the result in soft tissue problems and to peri-
screws may not be possible. However, the implant disease (see below).16,106,107 Therefore, Peri-implant tissue response
use of a screw-retained restoration may be its use should be avoided when the implant- to bacterial insult and peri-
considered when the implant platform is abutment connection is deeply embedded implant diseases
situated deep sub-mucosally, as complete sub-mucosally, which may preclude its
Despite their high success
removal of cement is not always possible removal. Furthermore, removing the
rate, implant failures are also reported to
when a cement-retained restoration is used. cement is not a predictable procedure and
occur. Several factors that have already
The screw type is not indicated when the may cause the abutment/restoration to be
been mentioned earlier which influence
screw hole is pointed at the labial surface scratched,111 leading to plaque accumulation.
such success should be considered when
as this compromises the aesthetics. Hence, Marginal adaptation between the abutment
treatment is planned.7 The implant may fail
the implant should be placed in its optimal and the restoration may also be inferior
before it is put to function as a result of its
position and angulation to avoid negative to that obtained when the screw-retained
failure to integrate with the peri-implant
effects on aesthetics, otherwise an angled restoration is used. It is also not suitable
tissue during the healing stage. This type of
abutment may provide an acceptable when the vertical restorative space is limited,
failure is categorized as an early failure. The
alternative. In the posterior region, the as retention may be compromised.
implant may also lose its integration and
occlusal morphology of the restoration may
fail at a later stage, months or even years
be difficult to obtain as the hole through The attachment systems after implant placement. This is known as
which the screw is tightened occupies
An attachment is defined as a late failure.113 The criteria for dental implant
a major part of the occlusal table of the
mechanical device used for the fixation, success are displayed in Table 7.
restoration. Furthermore, the access hole
retention and stabilization of a dental One of the complications that
may weaken the porcelain and lead to its
prosthesis.112 It is used with implant- is reported to affect the peri-implant tissue
fracturing. It is important to mention that,
supported removable partial dentures is caused by the inflammatory response of
if screw loosening of one restoration occurs
and overdentures. The attachment usually this tissue to bacteria that forms a biofilm on
in a fixed-implant supported restoration, a
consists of two parts. One part is attached to the implant surface.114,115 It occurs when the
cantilevering effect can arise and put the
the implant, while the other part is attached balance between the host’s defence and the
other abutment, implants, screw and the
to the prosthesis. Five types of attachment bacterial load shifts in favour of the bacteria.
peri-implant bone at risk as they are exposed
systems are available and compatible with This tissue response may be limited to the
to tremendous forces. Also, the screw
the main implant systems. The attachment peri-implant soft tissues (mucosa) or may
loosening is not an unreal problem with the
systems that are commonly used with RISOs also extend to and affect the peri-implant
screw-retained restoration. However, the
include: bar/clip, balls, locators, magnet bone and lead to its resorption.
ability to retrieve the restoration/prosthesis
and telescopic crown.108,112 The use of a Both tissue responses to
easily to allow its cleaning (and of the peri-
bar system allows splinting of two or more bacterial insult are collectively known as
implant tissues) is a significant advantage of
implants together. The other attachment peri-implant diseases, and are classified as
screw-retained restorations.
types may be used individually and also peri-implant mucositis or peri-implantitis.116
in combination with the bar system. The In peri-implant mucositis, the inflammatory
Cement-retained restorations attachments are attached to the implant by response is not essentially different from
The cement-retained restoration screws, resulting in a screw-joint. Features that which occurs in gingiva when it is
is indicated when mouth opening is
restricted, and when the implant angulation
is not optimal without a major negative 1. The different designs of the attachment systems are used to gain retention, support
effect on the aesthetic outcome of the and stability of the overdenture.
restoration.106-108 The occlusal morphology 2. They consist of a matrix (female) and a patrix (male):
can be easily constructed in the normal way,  The matrix accommodates the patrix; and
as in conventional restorations. The materials  The patrix frictionally fits and engages the matrix.
and techniques used for the fabrication 3. The joint that is made between the patrix and the matrix may be rigid (when
of the cement-retained restoration are no movements exist between the patrix and matrix) or resilient (when there are
similar to those used in the fabrication of movements).
conventional restorations. The trial stage and 4. The involved dental implants are either splinted or non-splinted.
the final cementation procedure are almost 5. A bar is usually used to connect the implants (splinted).
identical to those used in conventional 6. Bars may be custom-made, pre-fabricated (ready-made) or CAD/CAM milled.
restorations. However, it may not be possible 7. An individual attachment system is usually used in a non-splinted manner or combined
to remove the cement-retained restorations with a bar system.
if permanent cementing media is used.
Table 6. Features of attachment systems used for RISOs.
Therefore, restorations have to be cut in
July/August 2017 DentalUpdate 613
ImplantDentistry

1. That an individual, unattached implant is immobile when tested clinically.


2. That a radiograph does not demonstrate any evidence of peri-implant radiolucency.
3. That vertical bone loss is less than 0.2 mm annually following the implant’s first year of service.
4. That individual implant performance is characterized by an absence of signs and symptoms, such as pain, infections, neuropathies,
paresthesia or violation of the mandibular canal.
Table 7. Criteria for dental implant success114

1. Bone remodelling after implant placement loss is still debatable. Nevertheless, the
2. Reformation of a ‘biologic width’ current literature presents several factors
3. Presence of rough/smooth interface which may contribute to this loss, such as
4. Presence of a micro-gap at implant-abutment/restoration interface surgical trauma, reformation of a ‘biologic
5. Surgical trauma width’ and presence of a rough/smooth
6. Occlusal overloading interface. However, the factors that are
7. A ‘stress shielding’ phenomenon most commonly cited to cause such bone
8. Incomplete removal of luting cement resorption are displayed in Table 8.
9. Peri-implant disease

Table 8. Factors that may contribute to or cause crestal bone loss.6


Role of the patient and the
dental professionals
Each dental implant and
restoration/prosthesis should be evaluated
exposed to pathogenic bacteria and leads as its progression rate, may differ from
clinically and radiographically in a similar
to gingivitis.117 Therefore, in principle, peri- that which is commonly seen in chronic
manner to the treatment of periodontal
implant mucositis resembles gingivitis. periodontitis.6,7.117 For instance, the
disease. Oral hygiene should be observed
The onset and progression of protective connective tissue capsule, which
and regular check-ups should be
mucositis may be affected by a decrease was found to separate the periodontal
scheduled. Therefore, after a physiologic
in the vascularity and an increase in lesion from the alveolar bone around teeth
tissue remodelling period and at the time
collagen to fibroblast ratio in the peri- in the case of chronic periodontitis, does
of prosthesis installation, clinical and
implant connective tissue, and by the way not exist around implants.7 Therefore, the
radiographic examinations of the peri-
they are arranged around the implant self-limiting process is not present around
implant tissue should be carried out and
surface.117 Clinically, peri-implant mucositis implants, which may provide an explanation
is characterized with bleeding on gentle for the fast development and progression of used as a baseline to monitor any change
probing. It is a treatable disease and the the peri-implant disease. in the tissue and to intervene if required.
damage is reversible. However, it may It should be mentioned that When any deviation from the norm is
progress into peri-implantitis if untreated.6,7 dental implants may fail as a result of these found, intervention is then considered and
There are no major differences in the diseases if they are not treated as they carried out. In general, oral hygiene should
bacteria that were found to be associated lead to bone resorption, and eventually be monitored and different oral hygiene
with mucositis and peri-implantitis. This may to mobility and failure of the affected aids should be demonstrated and the
indicate that mucositis is the origin of peri- implant.8-10 patient encouraged to use them as often as
implantitis.117 It is important to remember required.
On the other hand, peri- that resorption of peri-implant crestal In general, care for dental
implantitis occurs when both the peri- bone occurs within the first year of implants has two phases: patient self-care
implant mucosa and bone are affected. It implant placement and continues to occur and professional clinical maintenance
resembles chronic periodontitis in natural to a lesser degree afterwards. It occurs aspects.122 It is the responsibility of the
teeth. However, some differences do exist. irrespective of the implant placement patient to maintain good oral hygiene.
For instance, the crestal bone loss occurs in a method (sub-merged or non-submerged). Patient self-care consists of a daily oral
circumferential fashion around the affected Based on a 15-year retrospective study, hygiene procedure in which toothbrush
implant, unlike bone resorption seen in Adell and colleagues119 reported that crestal (manual/powered and single tufted ones),
chronic periodontitis. The circumferential bone loss during the healing period and the auxiliary aids such as inter-proximal
shape of the peri-implantitis lesions may first year after connecting the prosthesis, brushes, dental floss/tape and mouthrinses
be attributed to the lack of periodontal was about 1.5 mm. Thereafter, there was may be used. A combination of these
ligament, and to the surface topographies only 0.1 mm bone loss annually. In another aids, whenever it is necessary, should be
of the involved implants which facilitate study, an average of 0.9 mm crestal bone considered and demonstrated. For instance,
the spread of infection apically as well as was lost during the first year and no more powered toothbrushes, which have different
laterally.6,117 The extent and the composition than 0.07 mm annually in the following interchangeable bristle heads (flattened,
of cells in the peri-implantitis, as well years.120,121 The exact cause of this bone rubber cup-like, short- and long-pointed in
614 DentalUpdate July/August 2017
ImplantDentistry

July/August 2017 DentalUpdate 615


ImplantDentistry

shape) that suit different clinical situations management of peri-implant diseases is not leads to screw loosening or even to its
may be used. When they are used properly, within this article’s scope. fracture. In the former situation, the screw
the result is a healthy environment around may be replaced, but in the latter situation
the implant. However, it is important Complications associated with the removal of the screw may not be
to mention that limiting the number of implant-supported restorations possible and the treatment is complicated,
auxiliary aids, their simplicity and the time and prostheses which is beyond the scope of this article.
required for their use are important for To minimize the occurrence of screw-
Several biological and
patients’ compliance as they play a vital role mechanical complications are reported loosening or fracture, the recommended
in this aspect.123 with the use of dental implants to support/ torque should be implemented using a
As already mentioned, dental retain restorations and prostheses. For torque driver that ensures that the right
implants are affected by and may fail as a instance, screws used to connect different amount of torque is achieved.33
result of the peri-implant disease which combinations of the implant-supported Mechanical superstructure
can be detected only by regular clinical restorations/prostheses may become loose failure may also occur when the material’s
and radiographic examinations. Therefore, and need to be retightened or replaced. mechanical properties and/or thickness
when an implant is affected by the peri- Screw loosening may be due to it not being is not optimum or when the occlusal
implant disease, the patient should be made adequately torqued or over-torqued or due design is not correctly designed. The
aware of the situation and a treatment plan to micro-movements that occur as a result of superstructure failure may also occur as
should be implemented and regular follow- the manufacturing tolerance.33,35 An under- a result of lack of passivity when several
ups arranged. However, there is a lack of torqued screw fails to deliver the tension implants are connected together. The lack
consensus on how peri-implant disease that is required to produce the optimum of passivity may overload the implants
is treated. Nevertheless, the Cumulative clamping force between the screw-joint and place the superstructure under
Interceptive Supportive Therapy (CIST) components. Re-tightening is there for tremendous pressure, that may lead to its
protocol that was presented by Lang required. Screw re-tightening can be easily failure. To check for passivity a test called a
and colleagues123 may be followed when achieved when the restoration is a screw- ‘Sheffield test’ or a ‘one-screw test’ is usually
peri-implant disease is found. The CIST is retained type. However, when the restoration carried out. However, the passivity problem
a systemic comprehensive protocol. This is cement-retained, cutting the restoration may be avoided by the use of computer-
protocol is based on clinical parameters to gain access to the screw may be the only aided design/computer-aided manufacture
such as peri-implant pocket depth (PIPD), solution, especially when permanent cement (CAD/CAM) technology.
bleeding on probing (BoP) and peri-implant is used. When a provisional cement is used, Acrylic or porcelain veneer may
bone loss on which clinical diagnosis the use of crown removal may be tried.35 also fail when the bulk of these materials
is made. Accordingly, a treatment plan When the screw is over-torqued are inadequate. For instance, when a
and continuous follow-up strategy are to a degree which places the screw material limited vertical restorative space does not
constructed. A summary of this protocol in tensile stress that exceeds its elastic limit, allow the use of the optimum thickness of
is presented in Table 9. However, the the screw may be plastically elongated. This the material. Depending on the degree of

Clinical parameters Clinical Diagnosis Treatment Protocols

*PIPD (shallow), Healthy peri-implant tissues No treatment is needed, just regular check-ups and
No plaque enhancement of oral hygiene
No **BoP
*PIPD (shallow) Mucositis A. Mechanical debridement and polishing using a
Plaque is present rubber cup and non-abrasive paste and regular check-
**BoP is present ups and enhancement of oral hygiene

*PIPD ≤5 mm Mucositis B. Treatment includes treatment A with antiseptic


cleaning

*PIPD >5 mm associated with bone loss of up to Peri-implantitis C. Same as treatment B in addition to the use of local
2 mm or systemic antibiotic
*PIPD >5 mm associated bone loss >2 mm Severe peri-implantitis D. Same treatment C combined with surgery (access
flap, resective method or regenerative technique)

*Peri-Implant Pocket Depth; **Bleeding on Probing


Table 9.The clinical parameters, diagnosis and a summary of the CIST protocol for treatment of peri-implant diseases.123

616 DentalUpdate July/August 2017


ImplantDentistry

mechanical damage of the restoration/ scratching the implant components.33 assessment are required to discover such
prosthesis, fracture of porcelain may be conditions. This necessitates recall visits and
repaired intra-orally using the Co-Jet® Conclusion check-ups which allow the dental personnel
system (3M ESPE, St Paul, Mn, USA) and to intervene in the proper time and to rescue
Dental implants are widely
composite resin material. It is considered the implant and its restoration/prosthesis.
used and considered as one of the options
as a reliable method for such repairing. Therefore, the dental personnel should be
by which missing teeth are replaced. They
Fracture of acrylic may also be repaired prepared and able to diagnose and to deal
are used successfully to replace single and
using composite resin materials. However, with such complications and to refer the
multiple missing teeth as well a completely
when the metal frame-work is fractured, patients when required.
edentulous jaw. The use of dental implants
the only solution is its removal.
are increasing and the dental professionals
RISO attachment failure and Acknowledgement
are more likely to see patients who have
complications are mostly of a mechanical The authors would like to thank
implant-supported restorations/prostheses.
nature and include:35 Mr Emmet Ryan (Dublin Dental University
 Fracture of the acrylic base, teeth and Therefore, basic knowledge of dental
implants is necessary for dental personnel. Hospital) for providing the images in Figure
retentive clip; 11 and Dr Brendan Grufferty (Dublin Dental
 Reduction of retention as a result of Several factors are known to affect success
of any implant system. These factors may University Hospital) for providing the images
wear of the retentive elements or in Figure 13.
loosening of matrices and screws; be related to features locally, such as bone
 Fracture or wear of the clip and matrix; quality and quantity. Other factors are
 Fracture of solder joints; and related to the surgical method by which References
 Dislodgement of the attachments. an implant is placed or which are related 1. Gokcen-Rohlig B, Yaltirik M, Ozer S, Tuncer
to the implant system used, such as length ED, Evlioglu G. Survival and success of ITI
Wear of the attachment implants and prostheses: retrospective
component is a problem that may and diameter of the implant. Furthermore, study of cases with 5-year follow-up. Eur J
reduce the overdentures’ retention and, dental implants are affected by peri-implant Dent 2009; 3: 42−49.
diseases which, if not treated, can cause 2. Baig MR, Rajan M. Effects of smoking
consequently, a replacement of the worn on the outcome of implant treatment: a
attachment becomes a necessity. Less the implant to fail. It requires continuous literature review. Indian J Dent Res 2007;
prosthetic maintenance was required with monitoring, regular check-ups and may 18: 190−195.
require professional interventions, the time 3. Zupnik J, Kim S-W, Ravens D, Karimbux
the splinted (bar/clip) designs than with N, Guze K. Factors associated with dental
the unsplinted ones.124 Nevertheless, the of intervention being vital. implant survival: a 4-year retrospective
use of bars may complicate the hygiene The success of any implant- analysis. J Periodontol 2011; 82:
supported restoration/prosthesis is 1390−1395.
matter125 and it may be associated with 4. Abt E. Growing body of evidence on
a misfit of the framework, which has the dependent on the interaction between survival rates of implant-supported fixed
potential to generate unwanted stress on the patient and the dental personnel. prostheses. Evid Based Dent 2008; 9:
Therefore, maintaining good oral hygiene 51−52.
the attachment, the implant, the retained 5. Han HJ, Kim S, Han DH. Multifactorial
screw and also the peri-implant bone. and committing to regular check-ups are the evaluation of implant failure: a 19 year
Relining of the denture is responsibility of the patient. On the other retrospective study. Int J Oral Maxillofac
hand, it is the responsibility of the dental Implants 2014; 29: 303−310.
also required regularly and may need 6. Warreth A, Boggs S, Ibieyou N, El-Helali
to be carried out every few years to personnel to examine the implants and R, Hwang S. Peri-implant diseases: an
compensate for the changes in the the restorations/prostheses clinically and overview. Dent Update 2015; 42: 166−184.
radiographically. It is also the responsibility 7. Renvert S, Giovannoli J-L. Peri-implantitis.
alveolar ridge that may occur. Failures Paris, France: Quintessence International,
of the implant-supported fixed dental of the dental practitioner to demonstrate 2012.
prosthesis also occur. The failures and educate the patient on how to look after 8. Buser D, Weber HP, Donath K, Fiorellini
the implant and to tailor check-up recall JP, Paquette DW, Williams RC. Soft tissue
include screw loosening and fracture reactions to non-submerged unloaded
of the superstructure. Speech may be visits according to the patient’s needs. titanium implants in beagle dogs.
affected when tissue loss is severe. Mechanical failures associated J Periodontol 1992; 63: 225−235.
The compensation of lost tissue with with implant-supported restorations/ 9. Weber HP, Buser D, Donath K, Fiorellini JP,
Doppalapudi V, Paquette DW, Williams RC.
acrylic or porcelain is usually required. prostheses, such as screw loosening or Comparison of healed tissues adjacent
This compensation may lead to an fracture and chipping of porcelain veneer to submerged and non-submerged
increase in plaque accumulation and and fracture of the superstructure, are not unloaded titanium dental implants. A
histometric study in beagle dogs. Clin Oral
tissue inflammation as the oral hygiene uncommon. Loss of retention of the implant- Implants Res 1996; 7: 11−19.
procedure is compromised. Meticulous supported overdenture are common clinical 10. Cochran DL. The scientific basis for and
effort from the patient is required. findings which may make the patient seek clinical experiences with Straumann
implants including the ITI Dental Implant
Calculus deposition once formed cannot treatment. On the other hand, plaque System: a consensus report. Clin Oral
be removed by a daily oral hygiene. accumulation and mucosal hyperplasia Implants Res 2000; 11(Suppl. 1): 33−58.
Therefore, professional intervention is in the per-implant site do not necessarily 11. Tarnow DP, Cho SC, Wallace SS. The effect
of inter-implant distance on the height
necessary. This intervention consists of promote the patient to look for treatment. of inter-implant bone crest. J Periodontol
the use of scalers with plastic tips to avoid Consequently, professional evaluation and 2000; 71: 546−549.

July/August 2017 DentalUpdate 617


ImplantDentistry

12. Brånemark P-I. Osseointegration and 29. Schwarz F, Alcoforado G, Nelson K, 44. Le Guehennec L, Goyenvalle E, Lopez-
its experimental studies. J Prosthet Dent Schaer A, Taylor T, Beuer F, Strietzel FP. Heredia MA, Weiss P, Amouriq Y, Layrolle
1983; 50: 399−410. Impact of implant-abutment connection, P. Histomorphometric analysis of the
13. Sennerby L, Ericson LE, Thomsen P, positioning of the machined collar/ osseointegration of four different implant
Lekholm U, Astrand P. Structure of the microgap, and platform switching on surfaces in the femoral epiphyses of
bone-titanium interface in retrieved crestal bone level changes. Camlog rabbits. Clin Oral Implants Res 2008; 19:
clinical oral implants. Clin Oral Implants Foundation Consensus Report. Clin Oral 1103−1110.
Res 1991; 2: 103−111. Implants Res 2014; 25: 1301−1303. 45. Renvert S, Roos-Jansåker AM, Claffey N.
14. Clokie CML, Warshawsky H. 30. Romanos GE, Basha-Hijazi A, Gupta B, Nonsurgical treatment of peri-implant
Morphological and radioautographic Ren YF, Malmstrom H. Role of clinician’s mucositis and peri-implantitis: a literature
studies of bone formation in relation to experience and implant design on review. J Clin Periodontol 2008; 35:
titanium implants using the rat tibia as a implant stability. An ex vivo study in 305−315.
model. Int J Oral and Maxillofac Implants artificial soft bones. Clin Implant Dent 46. Misch CE. Short dental implants: a
1995; 10: 155−165. Relat Res 2014; 16: 166−171. literature review and rationale for use.
15. The glossary of prosthodontic terms. 31. Menicucci G, Pachie E, Lorenzetti M, Dent Today 2005; 24: 64−68.
J Prosthet Dent 2005; 94: 38. Migliaretti G, Carossa S. Comparison of 47. Monje A, Fu JH, Chan HL, Suarez F,
16. Warreth A, Fesharaki H, McConville R, primary stability of straight-walled and Galindo-Moreno P, Catena A, Wang HL.
McReynolds D. An introduction to single tapered implants using an insertion Do implant length and width matter
implant abutments. Dent Update 2013; torque device. Int J Prosthodont 2012; 25: for short dental implants (<10 mm)? A
40: 7−17. 465−471. meta-analysis of prospective studies.
17. Thomas KA. Hydroxyapatite coatings. 32. Alves CC, Neves M. Tapered implants: J Periodontol 2013; 84: 1783−1791.
Orthopaedics 1994; 17: 267−278. from indications to advantages. J 48. Misch CE, Steignga J, Barboza E, Misch-
18. Sykaras N, Iacopino AM, Marker VA, Periodont Rest Dent 2009; 29: 161−167. Dietsh F, Cianciola LJ, Kazor C. Short
Triplett RG, Woody RD. Implant materials, 33. Jacobs SH, O’Connell BC. Dental Implant dental implants in posterior partial
designs, and surface topographies: their Restoration: Principles and Procedures edentulism: a multicenter retrospective
effect on osseointegration. A literature 1st edn. New Malden, UK: Quintessence 6-year case series study. J Periodontol
review. Int J Oral Maxillofac Implants 2000; Publishing, 2001. 2006; 77: 1340−1347.
15: 675−690. 34. Misch CE. Contemporary Implant Dentistry 49. Bahat O. Treatment planning and
19. Higuchi KW, Folmer T, Kultje C. Implant 2nd edn. St Louis: Elsevier, 2008. placement of implants in the posterior
survival rates in partially edentulous 35. Warreth A, McAleese E, McDonnell P, maxillae: report of 732 consecutive
patients: a 3-year prospective multicenter Slami R, Guray SM. Dental implants and Nobelpharma implants. Int J Oral
study. J Oral Maxillofac Surg 1995; 53: single implant-supported restorations. J Ir Maxillofac Implants 1993; 8: 151−161.
264−268. Dent Assoc 2013; 59: 32−43. 50. Winkler S, Morris HF, Ochi S. Implant
20. Geckili O, Bilhan H, Geckili E, Cilingir A, 36. Steigenga J, Al-Shammari K, Misch C, survival to 36 months as related to length
Mumcu E, Bural C. Evaluation of possible Nociti Jr. FH, Wang H-L. Effects of implant and diameter. Ann of Periodontol 2000; 5:
prognostic factors for the success, thread geometry on percentage of 22−31.
survival, and failure of dental implants. osseointegration and resistance to reverse 51. Jokstad A. The evidence for endorsing
Implant Dent 2014; 23: 44−50. torque in the tibia of rabbits. J Periodontol the use of short dental implants remains
21. Lekholm U, Zarb, GA. Patient selection 2004; 75: 1233−1241. inconclusive. Evid Based Dent 2011; 12:
and preparation. In: Tissue-Integrated 37. Rabel A, Köhler SG, Schmidt-Westhausen 99−101.
Prostheses. Osseointegration in Clinical AM. Clinical study on the primary stability 52. Monje A, Chan HL, Fu JH, Suarez F,
Dentistry 1st edn. Brånemark P, Zarb of two dental implant systems with Galindo-Moreno P, Wang HL. Are short
G, Albrektsson T. New Malden, UK: resonance frequency analysis. Clin Oral dental implants (<10 mm) effective? A
Quintessence Publishing Co, 1985: Investig 2007; 11: 257−265. meta-analysis on prospective clinical
195−205. 38. Yoon HG, Heo SJ, Koak JY, Kim SK, Lee trials. J Periodontol 2013; 84: 895−904.
22. Jones AA, Cochran DL. Consequences of SY. Effect of bone quality and implant 53. Lee JH, Frias V, Lee KW, Wright RF. Effect
implant design. Dent Clin North Am 2006; surgical technique on implant stability of implant size and shape on implant
50: 339−360. quotient (ISQ) value. J Adv Prosthodont success rate: a literature review. J Prosthet
23. Kim TH, Lee DW, Kim CK, Park KH, Moon 2011; 3: 10−15. Dent 2005; 94: 377−381.
IS. Influence of early cover screw exposure 39. Cochran DL. A comparison of endosseous 54. Allum SR, Tomlinson RA, Joshi R. The
on crestal bone loss around implants: dental implant surfaces. J Periodontol impact of loads on standard diameter,
intra-individual comparison of bone level 1999; 70: 1523−1539. small diameter and mini implants: a
at exposed and non-exposed implants. 40. Oue H, Doi K, Oki Y, Makihara Y, Kubo T, comparative laboratory study. Clin Oral
J Periodontol 2009; 80: 933−939. Perrotti V, Piattelli A, Akagawa Y, Tsuga K. Implants Res 2008; 19: 553−559.
24. Javed F, Romanos GE. The role of primary Influence of implant surface topography 55. Misch CE, Qu M, Bidez MW. Mechanical
stability for successful immediate loading on primary stability in a standardized properties of trabecular bone in the
of dental implants. A literature review. osteoporosis rabbit model study. J Funct human mandible: implications for dental
J Dent 2010; 38: 612−620. Biomater 2015; 6: 143−152. implant treatment planning and surgical
25. Kim YS, Lim YJ. Primary stability and 41. Novaes AB Jr, Souza SL, de Oliveria PT, placement. J Oral Maxillofac Surg 1999;
self-tapping blades: biomechanical Souza AM. Histomorphometric analysis of 57: 700−706.
assessment of dental implants in the bone-implant contact obtained with 56. Renouard F, Nisand D. Impact of implant
medium-density bone. Clin Oral Implants 4 different implant surface treatments length and diameter on survival rates.
Res 2011; 22: 1179−1184. placed side by side in the dog mandible. Clin Oral Implants Res 2006; 17(Suppl 2):
26. Morris HF, Winkler S, Ochi S, Kanaan A. Int J Oral Maxillofac Implants 2002; 17: 35−51.
A new implant designed to maximize 377−383. 57. Ivanoff CJ, Sennerby L, Johansson C,
contact with trabecular bone: survival 42. Klokkevold PR, Johnson P, Dadgostari S, Rangert B, Lekholm U. Influence of
to 18 months. J Oral Implantol 2001; 27: Caputo A, Davies JE, Nishimura RD. Early implant diameters on the integration of
164−173. endosseous integration enhanced by screw implants. An experimental study in
27. Davies JE. Mechanisms of endosseous dual acid etching of titanium: a torque rabbits. Int J Oral Maxillofac Surg 1997; 26:
integration. Int J Prosthodont 1998; 11: removal study in the rabbit femur. Clin 141−148.
391−401. Oral Implants Res 2001; 12: 350−357. 58. Langer B, Langer L, Herrmann I, Jorneus
28. Warreth A, Ibieyou N, MacCarthy D. 43. Wong M, Eulenberger J, Schenk R, L. The wide fixture: a solution for special
Bisphosphonates, oral implants and Hunziker E. Effects of surface topology bone situations and a rescue for the
osteonecrosis of the jaw: a review and on the osseointegration of implant in compromised implant. Part 1. Int J Oral
guidelines. J Dent Oral Hyg 2010; 11: trabecular bone. J Biomed Mater Res 1995; Maxillofac Implants 1993; 8: 400−408.
155−162. 29: 1567−1575. 59. Davarpanah M, Martinez H, Tecuciana

618 DentalUpdate July/August 2017


ImplantDentistry

J-F, Celletti R, Lazzara R. Small- bioindicator. Implant Dent 2015; 24: a function of time. Int J Oral Maxillofac
diameter implants: indications and 37−41. Implants 2004; 19: 124−132.
contraindications. J Esthet Dent 2000; 12: 74. Andreiotelli M, Wenz HJ, Kohal RJ. Are 91. Winkler S, Ring K, Ring JD, Boberick KG.
186−194. ceramic implants a viable alternative to Implant screw mechanics and the settling
60. Suba C, Velich N, Turi C, Szabó G. Surface titanium implants? A systematic literature effect: overview. J Oral Implantol 2003; 29:
analysis methods of biomaterials used in review. Clin Oral Implants Res 2009; 242−245.
oral surgery: literature review. J Craniofac 20(Suppl. 4): 32−47. 92. Kim KS, Han JS, Lim YJ. Settling of
Surg 2005; 16: 31−36. 75. Özkurt Z, Kazazoğlu E. Zirconia dental abutments into implants and changes in
61. Mouhyi J, Sennerby L, Wennerberg implants: a literature review. J Oral removal torque in five different implant-
A, Louette P, Dourov N, van Reck Implantol 2011; 37: 367−376. abutment connections. Part 1: Cyclic
J. Re-establishment of the atomic 76. Fuentealba R, Jofré J. Esthetic failure in loading. Int J Oral Maxillofac Implants
composition and the oxide structure of implant dentistry. Dent Clin North Am 2014; 29: 1079−1084.
contaminated titanium surfaces by means 2015; 59: 227−246. 93. Tan KB, Nicholls JI. Implant-abutment
of carbon dioxide laser and hydrogen 77. McGlumphy EA, Mendel DA, Holloway JA. screw-joint pre-load of 7 hex-top
peroxide: an in vitro study. Clin Implant Implant screw mechanics. Dent Clin North abutment systems. Int J Oral Maxillofac
Dent Relat Res 2000; 2: 190−202. Am 1998; 42: 71−89. Implants 2001; 16: 367−377.
62. Cooper LF. A role for surface topography 78. Maeda Y, Miura J, Taki I, Sogo M. 94. Lang LA, May KB, Wang RF. The effect of
in creating and maintaining bone at Biomechanical analysis on platform the use of a counter-torque device on the
titanium endosseous implants. J Prosthet switching: is there any biomechanical abutment-implant complex. J Prosthet
Dent 2000; 84: 522−534. rationale? Clin Oral Implants Res 2007; 18: Dent 1999; 81: 411−417.
63. Blumenthal NC, Cosma V. Inhibition 581−584. 95. Gratton DG, Aquilino SA, Stanford CM.
of apatite formation by titanium and 79. Gracis S, Michalakis K, Vigolo P, Vult von Micromotion and dynamic fatigue
vanadium ions. J Biomed Mater Res 1989; Steyern P, Zwahlen M, Sailer I. Internal properties of the dental implant-
23(A1 Suppl): 13−22. vs. external connections for abutments/ abutment interface. J Prosthet Dent 2001;
64. Siddiqi A, Payne AG, De Silva RK, Duncan reconstructions: a systematic review. 85: 47−52.
WJ. Titanium allergy: could it affect dental Clin Oral Implants Res 2012; 23(Suppl. 6): 96. Haack JE, Sakaguchi RL, Sun T, Coffey JP.
implant integration? Clin Oral Implants Res 202−216. Elongation and pre-load stress in dental
2011; 22: 673−680. 80. Akça K, Cehreli MC, Iplikçioğlu implant abutment screws. Int J Oral
65. Javed F, Al-Hezaimi K, Almas K, Romanos H. Evaluation of the mechanical Maxillofac Implants 1995; 10: 529−536.
GE. Is titanium sensitivity associated characteristics of the implant-abutment 97. Quek HC, Tan KB, Nicholls JI. Load fatigue
with allergic reactions in patients with complex of a reduced-diameter morse- performance of four implant-abutment
dental implants? A systematic review. Clin taper implant. A nonlinear finite element interface designs: effect of torque level
Implant Dent Relat Res 2013; 15: 47−52. stress analysis. Clin Oral Implants Res 2003; and implant system. Int J Oral Maxillofac
66. Alhag M, Renvert S, Polyzois I, Claffey N. 14: 444−454. Implants 2008; 23: 253−262.
Re-osseointegration on rough implant 81. Segundo RM, Oshima HM, da Silva IN, 98. Goheen KL, Vermilyea SG, Vossoughi J,
surfaces previously coated with bacterial Burnett LH Jr, Mota EG, Silva LL. Stress Agar JR. Torque generated by handheld
biofilm: an experimental study in the dog. distribution of an internal connection screwdrivers and mechanical torqueing
Clin Oral Implants Res 2008; 19: 182−187. implant prostheses set: a 3D finite devices for osseointegrated implants. Int J
67. Renvert S, Polyzois I, Maguire R. element analysis. Stomatologija 2009; 11: Oral Maxillofac Implants 1994; 9: 149−155.
Re-osseointegration on previously 55−59. 99. Khraisat A, Abu-Hammad O, Al-Kayed
contaminated surfaces: a systematic 82. Keating, K. Connecting abutments AM, Dar-Odeh N. Stability of the implant/
review. Clin Oral Implants Res 2009; to dental implants: ‘an engineer’s abutment joint in a single-tooth external-
20(Suppl 4): 216−227. perspective’. Irish Dentist 2001; July: hexagon implant system: clinical and
68. Gineste L, Gineste M, Ranz X, Ellefterion 43−46. mechanical review. Clin Implant Dent Relat
A, Guilhem A, Rouquet N, Frayssinet 83. Jung SW, Son MK, Chung CH, Kim HJ. Res 2004; 6: 222−229.
P. Degradation of hydroxylapatite, Abrasion of abutment screw coated with 100. Monje A, Pommer B. The concept of
fluorapatite, and fluorhydroxyapatite TiN. J Adv Prosthodont 2009; 1: 102−106. platform switching to preserve peri-
coatings of dental implants in dogs. J 84. Jo JY, Yang DS, Huh JB, Heo JC, Yun implant bone level: assessment of
Biomed Mater Res 1999; 48: 224−234. MJ, Jeong CM. Influence of abutment methodologic quality of systematic
69. Förster Y, Rentsch C, Schneiders W, materials on the implant-abutment joint reviews. Int J Oral Maxillofac Implants
Bernhardt R, Simon JC, Worch H, Rammelt stability in internal conical connection 2015; 30: 1084−1092.
S. Surface modification of implants in type implant systems. J Adv Prosthodont 101. de Almeida FD, Carvalho AC, Fontes M,
long bone. Biomatter 2012; 2: 149−157. 2014; 6: 491−497. Pedrosa A, Costa R, Noleto JW, Mourão CF.
70. Wie H, Herø H, Solheim T, Kleven E, Rørvik 85. Satterthwaite J, Rickma L. Retrieval of a Radiographic evaluation of marginal bone
AM, Haanaes HR. Bonding capacity fractured abutment screw thread from level around internal-hex implants with
in bone of HIP-processed HA-coated an implant: a case report. Br Dent J 2008; switched platform: a clinical case report
titanium: mechanical and histological 204: 177−180. series. Int J Oral Maxillofac Implants 2011;
investigations. J Biomed Mater Res 1995; 86. Drago CJ. A clinical study of the efficacy 26: 587−592.
29:1443−1449. of gold-tite square abutment screws in 102. Strietzel FP, Neumann K, Hertel M. Impact
71. Cook SD, Salkeld SL, Gaisser DM, Wagner cement-retained implant restorations. of platform switching on marginal peri-
WR. The effect of surface macrotexture Int J Oral Maxillofac Implants 2003; 18: implant bone-level changes. A systematic
on the mechanical and histologic 273−278. review and meta-analysis. Clin Oral
characteristics of hydroxylapatite-coated 87. Byrne D, Jacobs S, O’Connell B, Houston Implants Res 2015; 26: 342−358.
dental implants. J Oral Implantol 1993; F, Claffey N. Pre-loads generated with 103. Linkevicius T, Apse P, Grybauskas S, Puisys
19(4): 288−294. repeated tightening in three types of A. Influence of thin mucosal tissues on
72. Osman RB, Elkhadem AH, Ma S, Swain screws used in dental implant assemblies. crestal bone stability around implants
MV. Titanium versus zirconia implants J Prosthodont 2006; 15:164−171. with platform switching: a 1-year pilot
supporting maxillary overdentures: 88. Binon PP. Evaluation of three slip fit study. J Oral Maxillofac Surg 2010; 68:
three-dimensional finite element analysis. hexagonal implants. Implant Dent 1996; 2272−2277.
Int J Oral Maxillofac Implants 2013; 28: 5: 235−248. 104. Maeda Y, Satoh T, Sogo M. In vitro
e198−208. 89. Binon PP, McHugh MJ. The effect of differences of stress concentrations
73. Kajiwara N, Masaki C, Mukaibo T, Kondo eliminating implant/abutment rotational for internal and external hex implant-
Y, Nakamoto T, Hosokawa R. Soft tissue misfit on screw-stability. Int J Prosthodont abutment connections: a short
biological response to zirconia and metal 1996; 9: 511−519. communication. J Oral Rehabil 2006; 33:
abutments compared with natural tooth: 90. Cantwell A, Hobkirk JA. Pre-load loss 75−78.
microcirculation monitoring as a novel in gold prosthesis-retaining screws as 105. Cimen H, Yengin E. Analyzing the effects

July/August 2017 DentalUpdate 619


ImplantDentistry

of the platform-switching procedure with simulated subgingival margins. 120. Adell R, Lekholm U, Brånemark PI, Lindhe
on stresses in the bone and implant- J Prosthet Dent 1997; 78: 43−47. J, Rockler B, Eriksson B, Lindvall AM,
abutment complex by 3-dimensional fem 112. Burns DR. Mandibular implant Yoneyama T, Sbordone L. Marginal tissue
analysis. J Oral Implantol 2012; 38: 21−26. overdenture treatment: consensus and reactions at osseointegrated titanium
106. Hebel KS, Gajjar RC. Cement-retained controversy. J Prosthodont 2000; 9: 37−46. fixtures. Swed Dent J 1985; 28(Suppl):
versus screw-retained implant 113. Sakka S, Baroudi K, Nassani MZ. Factors 175−181.
restorations: achieving optimal occlusion associated with early and late failure of 121. Adell R, Lekholm U, Rockler B,
and aesthetics in implant dentistry. dental implants. J Investig Clin Dent 2012; Brånemark PI, Lindhe J, Eriksson B,
J Prosthet Dent 1997; 77: 28−35. 3: 258−261. Sbordone L. Marginal tissue reactions at
107. Newsome P, Reaney D, Owen S. Screw- 114. Albrektsson T, Zarb G, Worthington P, osseointegrated titanium fixtures (I). A
versus cement-retained crowns. Irish Eriksson AR. The long-term efficacy of 3-year longitudinal prospective study. Int
Dentist 2011; March: 22−25. currently used dental implants: a review J Oral Maxillofac Surg 1986; 15: 39−52.
108. Warreth A, Ramadan M, Bajilan MR, and proposed criteria ofsuccess. Int J Oral 122. Kracher CM, Smith WS. Oral health
Ibieyou N, El-Swiah J, Elemam RF. Maxillofac Implants 1986; 1: 11−25. maintenance dental implants. Dent Assist
Fundamentals of occlusion and 115. Socransky SS, Haffajee AD. Dental 2010; 79: 27−35.
restorative dentistry. Part I: basic biofilms: difficult therapeutic targets. 123. Lang NP, Berglundh T, Heitz-Mayfield
principles. J Ir Dent Assoc 2015; 61: Periodontol 2000 2002; 28: 12−55. LJ, Pjetursson BE, Salvi GE, Sanz M.
201−208. 116. Marsh PD. Dental plaque: biological
109. Feine JS, Carlsson GE, Awad MA, Chehade significance of a biofilm and community Consensus statements and recommended
A, Duncan WJ et al. The McGill consensus life-style. J Clin Periodont 2005; 32(Suppl clinical procedures regarding implant
statement on overdentures. Mandibular 6): 7−15. survival and complications. Int J Oral
two-implant overdentures as first choice 117. Mombelli A, Lang NP. The diagnosis and Maxillofac Implants 2004; 19(Suppl):
standard of care for edentulous patients. treatment of peri-implantitis. Periodontol 150−154.
Montreal, Quebec, May 24−25, 2002. 2000 1998; 17: 63−76. 124. Stoumpis C, Kohal RJ. To splint or not
Int J Oral Maxillofac Implants 2002; 17: 118. Lang NP, Berglundh T. Periimplant to splint oral implants in the implant-
601−602. diseases: where are we now? − Consensus supported overdenture therapy? A
110. British Society for the Study of Prosthetic of the Seventh European Workshop on systematic literature review. J Oral Rehabil
Dentistry. The York consensus statement Periodontology. J Clin Periodontol 2011; 2011; 38: 857−869.
on implant-supported overdentures. 38(Suppl 11): 178−181. 125. Warreth A, Byrne C, Alkadhimi AF,
Eur J Prosthodont Restor Dent 2009; 17: 119. Adell R, Lekholm U, Rockler B, Brånemark Woods E, Sultan A. Mandibular implant-
164−165. PI. A 15-year study of osseointegrated supported overdentures: attachment
111. Agar JR, Cameron SM, Hughbanks implants in the treatment of the systems, and number and locations of
JC, Parker MH. Cement removal from edentulous jaw. Int J Oral Surg 1981; 10: implants − Part II. J Ir Dent Assoc 2015; 61:
restorations luted to titanium abutments 387−416. 144−148.

620 DentalUpdate July/August 2017


View publication stats

You might also like