BELT SCALE CONTROLLER

JY500 B2

MANUAL
PLEASE READ THIS MANUAL VERY CAREFULLY
BEFORE OPERATING

Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd


1

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

CONTENT
PART ONE SUMMARIZE.................................................................................... 4
1.1 Main Functions and Characteristics........................................................ 4
1.2 Technical Specification............................................................................5
1.3 Cautions in Use........................................................................................ 5
1.4 Panel View............................................................................................... 6
PART TWO INSTALLATION AND CONNECTION........................................ 8
2.1 Dimension................................................................................................ 8
2.2 Meter Box Installation Opening Size.................................................... 8
2.3 Operation Flow...................................................................................... 9
2.4 Meaning of the Code..............................................................................10
2.5 Load Cell Connection............................................................................ 11
2.6 Port Define............................................................................................. 11
PART THREE BASIC PARAMETERS SETTING........................................... 12
3.1 Zeropoint Setting................................................................................... 12
3.2 Span Setting........................................................................................... 13
3.3 Speed Mode Selection........................................................................... 13
3.4 Flux Unit Selection................................................................................ 14
3.5 Max. Flux Setting.................................................................................. 14
3.6 Display Delay Setting............................................................................ 14
3.7 Weight Unit Selection............................................................................15
3.8 Decimal Setting......................................................................................15
3.9 Pulse Width Setting............................................................................... 16
3.10 Number of Belt Cycle Setting............................................................. 16
3.11 Time of Belt Cycle Setting.................................................................. 17
3.12 Belt Length Setting.............................................................................. 17
3.13 Auto Zero Track Setting...................................................................... 18
3.14 Auto Zero Track Rang Setting.............................................................18
3.15 Flux Dead Zone Setting....................................................................... 19
3.16 Flux Dead Zone Rang Setting............................................................. 19
2 Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com
PART FOUR CALIBRATION........................................................................... 21
4.1 Test Cycle Calibration........................................................................... 21
4.2 Zero Calibration..................................................................................... 22
4.3 Span Calibration.................................................................................... 24
4.3.1 Analog Calibration......................................................................25
4.3.2 Material Calibration.................................................................... 26
PART FIVE COMMUNICATION SETTING................................................... 28
5.1 Baud Rate Setting.................................................................................. 28
5.2 Address Setting...................................................................................... 28
5.3 Data Format Setting............................................................................... 29
5.4 Time Printing Setting............................................................................. 29
PART SIX CLEAN TOTALIZATION............................................................... 31
PART EIGHT CHANGE TIME..........................................................................33
PART NINE CHANGE PASSWORD................................................................ 34
PART TEN CONTROLATION SETTING........................................................ 36
10.1 Analog Current Setting........................................................................ 36
10.2 Control Current Setting........................................................................36
10.3 Flux Setting.......................................................................................... 37
10.4 Flux Controling Mode Setting............................................................. 37
10.5 Coefficient P Setting............................................................................ 38
10.6 Coefficient I Setting............................................................................. 38
10.7 Coefficient D Setting........................................................................... 39
10.8 Flux Controling.................................................................................... 39
10.9 High Flux Alarm.................................................................................. 40
10.10 Low Flux Alarm.................................................................................40
PART ELEVEN MODBUS PROTOCOL.......................................................... 42
11.1 Communication Format....................................................................... 42
11.1.1 Read Data.................................................................................. 42
11.1.2 Write Date................................................................................. 43
11.2 Read-only Register Address................................................................ 44

Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd


3

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com
11.3 Read-write Register Address............................................................... 44

4 Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

PART ONE SUMMARIZE


JY500B series belt scale adopts highest precision ∑ -Δ A/D conversion chip
and SCM with strong anti-jamming capability, the designer considered the
complexity in industrial field fully, designed with aborative software and
hardware, to make production and management to get a beneficial guarantee. This
product applied widely to measure and control industries, such as building
materials, chemical engineering, food, steel and mining.
For using this equipment safely and correctly, and fully exerting the function
of this equipment, please read this manual in detail before using this equipment.

1.1 Main Functions and Characteristics


◆ Gathers to weighing, display, communication, measurement at the integral
whole
◆ Internal integrate switch power supply can resist influence of voltage wave
◆ Has analog-digital dual filter function
◆ Diagnoses mistakes automatically, discovers and solves the problem easily
◆ Has RS-485 communication port and is able to carry on long-distance
correspondence
◆ Has function of setting password purview, it is convenient to manage
◆ Adapts to resistance strain mode load sensor
◆ Adapts to belt speed examination equipment whose output type is pulse signal
◆ Can correct belt self-weight, belt speed and totalization automatically via
inputting the structure parameters
◆ Can save all process parameters automatically when powering down
◆ Provides MODBUS mode, can transmit various parameters from long distance

Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd


5

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

1.2 Technical Specification


Ext. Power supply AC 220V ±20%, 50Hz ±2%
Power waste ≤30W
Working temperature -10-40℃
Humidity ≤90%RH(none coagulate water)

Keyboard 6 soundable key in total


Display mode 14 numeric bits
Measurement rang 0~59999
Linearity 0.01FS
Accuracy 0.25
Load cell signal range 0~24mV
Belt Speed pulse 0~5000Hz
Load cell excitation DC9V 250mA
Speed sensor DC24V 120mA

Modbus interface RS485≤1000m, baud rate and communication format


optional
Dimension 158.7(W)×166.6(D)×80.8(H)mm
Protected class IP54

1.3 Cautions in Use


※ Please don't open this equipment no metter it is power-on or power-off, so as
not to endanger your body safety or bring damage to this equipment.
※ Please install this equipment indoors, or it may be caused to short circuit while
it is raining.
※ Don't cut in or pull out the plugs from the backboard of the equipment, or
replace an insurance tube or a sensor when it is powered up.
※ The cable of sensor should be as short as possible, and should be kept off
other power cables and control wires to avoid possible jamming.
6 Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com
※ This equipment must be connected to the ground, and make sure all
connections are correct, firm and credible before being powered up.
※ Please don't use the same power cable(include the ground wire) or power plug
between this equipment and other electricity equipments, so as not to affect
the functions of this equipment.
※ Please don't press or knock the keyboard of the equipment heavily or in
disorder, so as not to bringing damage to this equipments.
※ Non-members of Longxin Company or other qualified personnel, don't
calibrate this equipment or set parameters, otherwise doing these will effect it
to work normally.
※ If this equipment is idle, it must be powered up at least once each month, and
more than an hour each time to drive out its internal moisture.
※ This equipment can't be cleaned with the strong sour or the strong alkali, or
the organic solvents like hydrocarbons, alcohols, ketones, so as not to damage
the chassis, the panel and the internal components of the equipment.
※ We won't accept that repairing or modifying this equipment by yourself. If the
equipment goes wrong, please solve the problems according to this manual or
contact us, otherwise you will lose the right of after-sales service.

1.4 Panel View

Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd


7

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com
1. LED for load cell fault or disconnecting load cell
2. LED for high flux alarm
3. LED for flux unit (lit when flux unit that the upper line displays is t/h)
4. LED for flux unit (lit when flux unit that the upper line displays is kg/h)
5. LED for flux unit (lit when flux unit that the upper line displays is kg/s)
6. LED for speed unit (lit when speed unit that the upper line displays is m/s)
7. LED for belt load unit (lit when belt load unit that the upper line displays is
kg/m)
8. LED for totalization unit (lit when totalization unit that the lower line displays
is t)
9. LED for totalization unit (lit when totalization unit that the lower line displays
is kg)
10. LED for date (lit when the lower line displays date)
11. LED for time (lit when the lower line displays time)
12. LED for A/D code (lit when the lower line displays A/D code)
13. MENU key
14. SWITCH key
15. ESC key
16. UPPER SCROLL key
17. LOWER SCROLL key
18. ENTER key

8 Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

PART TWO INSTALLATION AND CONNECTION


2.1 Dimension

2.2 Meter Box Installation Opening Size

Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd


9

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

2.3 Operation Flow

2.4
10 Meaning of the Code CODE MEANING
Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

Set parameters CODE MEANING


Zeropoint Clean weight
Span (totalization)
Speed Clean weight
Flux unit (totalization) confirm
Max. flux Load default parameters
Display delay
Weight unit Load default parameters
Decimal confirm
Pulse width Change time
Number of belt cycle
Time of belt cycle Change time
Belt lenth Change date
Auto zero track Change password
Zero track rang
Flux dead zone Input old password
Flux dead zone rang Input new password
Calibration New password confirm
Test cycle calibration Control menu
Zero calibration Analog (flux) current
Span calibration Control current
Communication Flux setting
setting Flux controling mode
Baud rate Coefficient P
Address Coefficient I
Data format Coefficient D
Flux controling

2.5 Load Cell Connection


Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd
11

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com
1. Four-wire Connection

2. Six-wire Connection

2.6 Port Define

12 Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

PART THREE BASIC PARAMETERS SETTING


Chang the parameters in the menu. If the password is activated at this time (not all
zero, reload default parameters will be cleaned), it will appear password inputting
dialog box (please read PART NINE), you can chang the parameters after
inputting the correct password. The following operations is base on the situations
that inputting the correct password or the password is unneed.

3.1 Zeropoint Setting


Press key after powering up, it displays SET PARAMETERS menu.

Press key when it is in SET PARAMETERS menu, it displays


ZEROPOINT menu.

Press key if you need to change zeropiont, then a small numberic LED
flashes, it means you can change the number, then press key the next left
numberic LED will flash. Then press key or key to add or minuse
the flashing number. Press key to acknowledge, or press key to

Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd


13

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com
cancel. Change other parameters are as the same with above, it will be not
introduced as follow.
The default value of zeropoint is 5000, the rang of zeropoint is 0~59999.

3.2 Span Setting


Press key when it is in ZERO menu, it displays SPAN menu.

The default value of span is 100000, the rang of span is 0~99999999.

3.3 Speed Mode Selection


Press key when it is in SPAN menu, it displays SPEED menu.

Press key if you need to change speed mode, then all small numberic
LEDs flash, then press key or key to choose simulated
( ) or extanal ( ) speed mode. Press key to
acknowledge, or press key to cancel.
The default value of speed mode is external, the rang of speed mode is simulated
and external.
14 Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

3.4 Flux Unit Selection


Press key when it is in SPEED menu, it displays FLUX UNIT menu.

Press key if you need to change flux unit, then the small numberic LEDs
flash, then press key or key to choose t/h ( ), kg/h ( )

or kg/s ( ). Press key to acknowledge, or press key to cancel.

The default value of flux unit is t/h, the rang of flux unit is t/h, kg/h and kg/s.

3.5 Max. Flux Setting


Press key when it is in FLUX UNIT menu, it displays MAX. FLUX menu.

The default value of max. flux is 100.0, the rang of max. flux is 0~9999999.9.

3.6 Display Delay Setting


Press key when it is in MAX. FLUX menu, it displays DISPLAY DELAY
menu.

Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd


15

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

The default value of display delay is 3, the rang of display delay is 0~59999.

3.7 Weight Unit Selection


Press key when it is in DISPLAY DELAY menu, it displays WEIGHT
UNIT menu.

The default value of weight unit is t ( ), the rang of weight unit is t ( ) and kg

( ).

3.8 Decimal Setting


Press key when it is in WEIGHT UNIT menu, it displays DECIMAL menu.

16 Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

The default value of decimal bit is 1 (right four numberic LEDs diplay ),
the rang of decimal bit is 0~3 (right four numberic LEDs diplay ,
, and respectively).

3.9 Pulse Width Setting


Press key when it is in DECIMAL menu, it displays PULSE WIDTH menu.

The default value of pulse width (unit: ms) is 100, the rang of pulse width is
0~59999.

3.10 Number of Belt Cycle Setting


Press key when it is in PULSE WIDTH menu, it displays NUMBER OF
BELT CYCLE menu.

Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd


17

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

The default value of number of belt cycle is 1, the rang of number of belt cycle is
0~99.

3.11 Time of Belt Cycle Setting


Press key when it is in NUMBER OF BELT CYCLE menu, it displays
TIME OF BELT CYCLE menu (unit: s).

The default value of time of belt cycle is 30, the rang of time of belt cycle is
0~59999.

3.12 Belt Length Setting


Press key when it is in TIME OF BELT CYCLE menu, it displays BELT
LENGTH menu (unit: m).

18 Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

The default value of belt length is 100.00, the rang of belt length is
1.00~999999.99.

3.13 Auto Zero Track Setting


Press key when it is in BELT LENGTH menu, it displays AUTO ZERO
TRACK menu.

The default value of auto zero track is off ( ), the rang of auto zero track is
off ( ) and on ( ).
When auto zero track is on, if the value that A/D value minuse zeropoint is in the
range of auto zero track, and the time the value keeps is bigger than time of belt
cycle (the belt keeps running and it is in the main menu at this time), zeropoint
will be reset. When it is tracking zeropoint, the first big numberic LED will flash
as .

3.14 Auto Zero Track Rang Setting


Press key when it is in AUTO ZERO TRACK menu, it displays ZERO

Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd


19

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com
TRACK RANG menu.

The default value of zero track rang is 5%, the rang of zero track rang is 0%~99%.

3.15 Flux Dead Zone Setting


Press key when it is in ZERO TRACK RANG menu, it displays FLUX
DEAD ZONE menu.

The default value of flux dead zone is off ( ), the rang of flux dead zone is
off ( ) and on ( ).
When flux dead zone is on, if the real-time flux is smaller than the value that max.
flux multiply flux dead zone rang, then the flux tracks to zero.

3.16 Flux Dead Zone Rang Setting


Press key when it is in FLUX DEAD ZONE menu, it displays FLUX
DEAD ZONE RANG menu.

20 Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

The default value of flux dead zone rang is 5%, the rang of flux dead zone rang is
0%~99%.

Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd


21

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

PART FOUR CALIBRATION


The equipment must be calibrated when it is used for the first time, and belt length
and time of belt cycle must be known before calibration. Input the parameters in
SET PARAMETERS menu, and input number of belt cycle.

4.1 Test Cycle Calibration


Press twice in main menu, it displays CALIBRATION menu.

Press key when it is in CALIBRATION menu, it displays TEST CYCLE


CALIBRATION menu.

Press key when it is in TEST CYCLE CALIBRATION menu, then it asks


whether to start test cycle calibration.

22 Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

Make sure the belt is running steadily at max. speed at this time, then press
key to start, it displays as follow.

The upper line displays the remain time and the lower displays the amounts of the
pulses that has been checked by the equipment. When calibration has been
completed, press key to save the amounts of the pulses. Press key
to cancel saving the data if the belt runs unsteadily during calibration. Please do
calibration at least thrice if it is the first calibration, if it is not the first calibration
and the belt speed and length are unchanged, you can do zero calibration and span
calibration directly.

4.2 Zero Calibration


Press key when it is in TEST CYCLE CALIBRATION menu, it displays
ZERO CALIBRATION menu.

Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd


23

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

Press key when it is in ZERO CALIBRATION menu, then it asks whether


to start zero calibration.

Make sure the belt is running steadily at max. speed at this time, then press
key to start, it displays as follow.

The upper line displays the remain time and the lower displays the accumulation.
When calibration has been completed, it displays as follow.

24 Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

The upper line displays the new zeropoint and the lower displays the error. Then
press key to save the new zeropoint, press key to cancel.

4.3 Span Calibration


Press key when it is in ZERO CALIBRATION menu, it displays SPAN
CALIBRATION menu.

Press key to select analog salibration or material salibration when it is in


SPAN CALIBRATION menu, the span calibration mode can be switched via
pressing key or key.

Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd


25

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

4.3.1 Analog Calibration


If you select analog calibration mode (it is suitabe for chain calibration and weight
calibration), then after pressing key, it displays as follow.

Press key to start, it displays as follow.

The upper line displays the remain time and the lower displays the accumulation.
When calibration has been completed, it displays as follow.

26 Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

The upper line displays the actual accumulation. Then input the actual weight in
the lower line via pressing key, key and key, for an example,
input 8.700, it displays as follow.

Then press key, it displays as follow.

The upper line displays the error and the lower displays new span.

4.3.2 Material Calibration


If you select material calibration mode, then after pressing key, it displays
as follow.

Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd


27

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

Let the belt run at this time, then load the material on the belt, it displays as follow.

The upper line displays the real-time flux and the lower displays the accumulation.
Press key when the material goes across the load cell, then it displays as
follow.

The upper line displays the actual accumulation. Then input the actual weight in
the lower line via pressing key, key and key, for an example,
input 8.700, it displays as above. Press key to save or press key
to cancel.

28 Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

PART FIVE COMMUNICATION SETTING


There is only one protocol -- modbus protocol in this equipment, please consult
PART TEN.
5.1 Baud Rate Setting
Press key thrice in main menu, it displays SET COMMUNICATION menu.

Press key when it is in SET COMMUNICATION menu, it displays BAUD


menu.

The default value of baud rate is 9600, the rang of baud rate is 4800, 9600, 19200,
38400.

5.2 Address Setting


Press key when it is in BAUD menu, it displays ADDRESS menu.

Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd


29

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

The default value of address is 1, the rang of address is 0~255.

5.3 Data Format Setting


Press key when it is in ADDRESS menu, it displays DATA FORMAT
menu.

The default value of data format is 8n1, the rang of data format is 8n1, 8n2, 8o1,
8o2.

5.4 Time Printing Setting


Press key when it is in DATA FORMAT menu, it displays TIME
PRINTING menu.

30 Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

The default value of time printing is off, the rang of time printing is off, on.
Press key when it is in TIME PRINTING menu, it displays TIME 1 FOR
PRINTING menu.

The format of time is HH-MM-SS. There are four times (default: 00-00-00) for
printing in all. When time printing mode is on, if you just need to print the report
once a day, please set time 1~time 4 as the same. If you just need to print the
report twice a day, please set time 2~time 4 as the same. If you just need to print
the report thrice a day, please set time 3~time 4 as the same.
If you want to print the report at any time, please press key in main menu
(it has not to do with time printing mode).

Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd


31

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

PART SIX CLEAN TOTALIZATION


Press key four times in main menu, it displays CLEAN TOTALIZATION
menu.

Press key if you want to clean totalization, then it displays as follow.

Press key to confirm, then it displays as follow and returns to main menu.

32 Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

PART SEVEN LOAD DEFAULT PARAMETERS


Press key five times in main menu, it displays LOAD DEFAULT
PARAMETERS menu.

Press key if you want to load default parameters, then it displays as follow.

Press key to confirm, then it displays as follow and returns to main menu.

Warning: the parameters you have set before in the equipment will be lost
can’’t be used till you
after loading default parameters, then the eqiupment can
input the parameters before and do all calibrations.

Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd


33

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

PART EIGHT CHANGE TIME


Press key six times in main menu, it displays CHANGE TIME menu.

Press key to change the time, it displays as follow.

The format of time is HH-MM-SS. Change the time with key, key
and key. Press key to confirm, or press key to cancel, then it
displays DATE menu.

The format of date is YY-MM-DD. Change the date with key, key
and key. Press key to confirm, or press key to cancel.
34 Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

PART NINE CHANGE PASSWORD


Press key seven times in main menu, it displays CHANGE PASSWORD
menu.

Press key to enter the old password (default: 000000) interface.

Inputing the number of one numberic LED, then press key to switch the
numberic LEDs, and the number you just input will be shielded at the same time,
it displays 0 instead. When you input the correct password, press key to
enter the new password interface.

Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd


35

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com
Input the new password then press key, it displays as follow.

Input the new password again then press key, it displays as follow.

Or press key to rechange the password, it displays as follow.

36 Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

PART TEN CONTROLATION SETTING


10.1 Analog Current Setting
Press key in main menu, it displays CONTROLATION menu.

Press key when it is in CONTROLATION menu, it displays ANALOG


(flux) CURRENT menu.

The default value of analog current is 4~20 mA, the rang of analog current is 0~20
mA, 4~20 mA.

10.2 Control Current Setting


Press key when it is in ANALOG (flux) CURRENT menu, it displays
CONTROL CURRENT menu.

Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd


37

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

The default value of control current is 4~20 mA, the rang of control current is
0~20 mA, 4~20 mA.

10.3 Flux Setting


Press key when it is in CONTROL CURRENT menu, it displays FLUX
SETTING menu.

The default value of flux is 100.0, the rang of flux is 0~9999999.9.

10.4 Flux Controling Mode Setting


Press key when it is in FLUX SETTING menu, it displays FLUX
CONTROLING MODE menu.

38 Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

The default value of flux controling mode is auto ( ), the rang of flux
controling mode is auto ( ), serial ( ).

10.5 Coefficient P Setting


Press key when it is in FLUX CONTROLING MODE menu, it displays
COEFFICIENT P menu.

The default value of coefficient P is 50, the rang of coefficient P is 0~99.

10.6 Coefficient I Setting


Press key when it is in COEFFICIENT P menu, it displays COEFFICIENT
I menu.

Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd


39

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

The default value of coefficient I is 2, the rang of coefficient I is 0~99.

10.7 Coefficient D Setting


Press key when it is in COEFFICIENT I menu, it displays COEFFICIENT
D menu.

The default value of coefficient D is 0, the rang of coefficient D is 0~99.

10.8 Flux Controling


Press key when it is in COEFFICIENT D menu, it displays FLUX
CONTROLING menu.

40 Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

The default value of flux controling is stop ( ), the rang of flux

controling is stop ( ), start ( ) and (it is max.

flux and is used for calibration)

10.9 High Flux Alarm


Press key when it is in FLUX CONTROLING menu, it displays HIGH
FLUX ALARM menu.

The default value of high flux alarm is 120%, the rang of high flux alarm is
0~200%

10.10 Low Flux Alarm


Press key when it is in HIGH FLUX ALARM menu, it displays LOW
FLUX ALARM menu.

Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd


41

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

The default value of high flux alarm is 0%, the rang of high flux alarm is 0~99%

42 Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

PART ELEVEN MODBUS PROTOCOL


MODBUS is a network protocol of Main-Sub format. This equipment adopts hex
RTU format and is as a slave to call by the PC or host.
And the JY500B2 controller adopts CCITT-16/N (G(x)=x16+x13+1) to check
sum.

11.1 Communication Format


11.1.1 Read Data
Host (Send)
Address Read Address Address Register Register Check Check
of meter of of Number Number H L
register register H L
H L
XX 03 9C XX XX XX CRCh CRCL

Meter (Respond)
Address Read Number Word ... Word Check Check
of meter of bytes H L H L
XX 03 XX XX ... XX CRCH CRCL
Note:
If the register address of sending to the equipment is error or CRC check is wrong,
no data will return from the equipment.
e.g.
We know the address of totalization is 9C59H. For an example, totalization =
100.84. In fact, the data type of totalization is long integer, it is 2764H with hex
format, and the value of decimal is 2. If the address of the equipment is 1, then the
read command and the respond from the equipment are as follow.
PC Read:

Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd


43

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

01H 03H 9CH 59H 00H 02H 3AH 48H


Respond from the equipment:
01H 03H 04H 00H 00H 27H 64H E0H 28H

11.1.2 Write Date


Host (Send)
Address Write Address Address Word ... Word Check Check
of meter of of H L H L
register register
H L
XX 10 XX XX XX ... XX CRCH CRCL
Meter (Respond)
Address Write Address Address Word ... Word Check Check
of meter of of H L H L
register register
H L
XX 10 XX XX XX ... XX CRCH CRCL
When the data is transmiting, the high byte is front and the low byte is back.
e.g.
If the flux I want to set is 100 (float format), at first we know the address of
setting flux is 9C4EH, then we know the four-byte format of 100 is 42C80000H,
so sending data is as follow:
01H 10H 9CH 4EH 00H 02H 04H 42H C8H 00H 00H 1AH 93H
If the data the equipment receives is correct, then the equipment will answer from
the high byte to the low byte of register and checkout. The data is as follow:
01H 10H 9CH 4EH 00H 02H 0FH 8FH
When setting address is over the range or not existent, the function code that the
equipment responds will return 90H instead of 10H. And the data order is from
the first-byte to the low byte of register and checkout.

44 Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

11.2 Read-only Register Address


Register Function Remarks
address
9C53H 4 bytes, belt load Float number, unit: kg/m
9C55H 4 bytes, speed Float number
9C57H 4 bytes, real-time flux Flaot number
9C59H 4 bytes, totalization Long integer, it has to do with the
decimal setting (consult 3.8), e.g. the
code that PC read is 10084, but the
value of decimal is 2, then the actual
totalization is 100.84.
9C5BH 4 bytes, symbol of D7 D6……D0, D7 is high flux alarm
alarm bit, D3 is low flux alarm bit. When it is
alarming, the corresponding bit is 1,
otherwise is 0.

11.3 Read-write Register Address


Register Function Remarks
address
9C40H 2 bytes, zero point Integer, rang: 0~59999
9C41H 4 bytes, span Long integer, rang: 0~99999999
9C43H 4 bytes, max. flux Float, rang: 1.0~9999999.9
9C45H 2 bytes, display delay Integer, rang: 0~59999
9C46H 2 bytes, decimal Integer, rang: 0~3
9C47H 2 bytes, pulse width Integer, rang: 0~59999ms
9C48H 2 bytes, auto zero track Integer, rang: 0: off, 1: on
9C49H 2 bytes, zero track range Integer, rang: 0~99%
9C4AH 2 bytes, flux dead zone Integer, rang: 0: off, 1: on
Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd
45

All Rights Reserved. www.meterforall.com


SHENZHEN LONGXIN ZHIYE TECHNOLOGY CO.,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com

9C4BH 2 bytes, flux dead zone rang Integer, rang: 0~99%


9C4CH 2 bytes, high flux alarm Integer, rang: 0~200% (correspond
to flux setting)
9C4DH 2 bytes, low flux alarm Integer, rang: 0~99% (correspond to
flux setting)
9C4EH 4 bytes, flux setting Float, rang: 0~9999999.9
9C50H 2 bytes, flux controling mode Integer, rang: 0: auto, 1: serial
9C51H 2 bytes, control current output Integer, rang: 0~100%
percentage
9C52H 2 bytes, flux controling Integer, rang: 0: stop, 1: start, 2: max.
flux (20mA)
9C59H 4 bytes, clean totalization Long integer, rang: 0: clean
totalization, others: noneffective
9C61H 2 bytes, coefficient P Integer, rang: 0~99
Note: the continuous address parameter can be muti-read or muti-write. When you
use the equipment to communicate with kingview 6.53, the address of the register
must be added 900001. e.g. we know the address of flux setting is 9C4EH (it is
40015 in decimal system), so the address of register that setting in kingvie 6.53 is
940016 (decimal system).

46 Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd

All Rights Reserved. www.meterforall.com