Belt Scale Controller JY500B2
Belt Scale Controller JY500B2
,LTD
Email:szlongxin@gmail.com Website : www.meterforall.com
JY500 B2
MANUAL
PLEASE READ THIS MANUAL VERY CAREFULLY
BEFORE OPERATING
CONTENT
PART ONE SUMMARIZE.................................................................................... 4
1.1 Main Functions and Characteristics........................................................ 4
1.2 Technical Specification............................................................................5
1.3 Cautions in Use........................................................................................ 5
1.4 Panel View............................................................................................... 6
PART TWO INSTALLATION AND CONNECTION........................................ 8
2.1 Dimension................................................................................................ 8
2.2 Meter Box Installation Opening Size.................................................... 8
2.3 Operation Flow...................................................................................... 9
2.4 Meaning of the Code..............................................................................10
2.5 Load Cell Connection............................................................................ 11
2.6 Port Define............................................................................................. 11
PART THREE BASIC PARAMETERS SETTING........................................... 12
3.1 Zeropoint Setting................................................................................... 12
3.2 Span Setting........................................................................................... 13
3.3 Speed Mode Selection........................................................................... 13
3.4 Flux Unit Selection................................................................................ 14
3.5 Max. Flux Setting.................................................................................. 14
3.6 Display Delay Setting............................................................................ 14
3.7 Weight Unit Selection............................................................................15
3.8 Decimal Setting......................................................................................15
3.9 Pulse Width Setting............................................................................... 16
3.10 Number of Belt Cycle Setting............................................................. 16
3.11 Time of Belt Cycle Setting.................................................................. 17
3.12 Belt Length Setting.............................................................................. 17
3.13 Auto Zero Track Setting...................................................................... 18
3.14 Auto Zero Track Rang Setting.............................................................18
3.15 Flux Dead Zone Setting....................................................................... 19
3.16 Flux Dead Zone Rang Setting............................................................. 19
2 Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd
2.4
10 Meaning of the Code CODE MEANING
Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd
2. Six-wire Connection
Press key if you need to change zeropiont, then a small numberic LED
flashes, it means you can change the number, then press key the next left
numberic LED will flash. Then press key or key to add or minuse
the flashing number. Press key to acknowledge, or press key to
Press key if you need to change speed mode, then all small numberic
LEDs flash, then press key or key to choose simulated
( ) or extanal ( ) speed mode. Press key to
acknowledge, or press key to cancel.
The default value of speed mode is external, the rang of speed mode is simulated
and external.
14 Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd
Press key if you need to change flux unit, then the small numberic LEDs
flash, then press key or key to choose t/h ( ), kg/h ( )
The default value of flux unit is t/h, the rang of flux unit is t/h, kg/h and kg/s.
The default value of max. flux is 100.0, the rang of max. flux is 0~9999999.9.
The default value of display delay is 3, the rang of display delay is 0~59999.
The default value of weight unit is t ( ), the rang of weight unit is t ( ) and kg
( ).
The default value of decimal bit is 1 (right four numberic LEDs diplay ),
the rang of decimal bit is 0~3 (right four numberic LEDs diplay ,
, and respectively).
The default value of pulse width (unit: ms) is 100, the rang of pulse width is
0~59999.
The default value of number of belt cycle is 1, the rang of number of belt cycle is
0~99.
The default value of time of belt cycle is 30, the rang of time of belt cycle is
0~59999.
The default value of belt length is 100.00, the rang of belt length is
1.00~999999.99.
The default value of auto zero track is off ( ), the rang of auto zero track is
off ( ) and on ( ).
When auto zero track is on, if the value that A/D value minuse zeropoint is in the
range of auto zero track, and the time the value keeps is bigger than time of belt
cycle (the belt keeps running and it is in the main menu at this time), zeropoint
will be reset. When it is tracking zeropoint, the first big numberic LED will flash
as .
The default value of zero track rang is 5%, the rang of zero track rang is 0%~99%.
The default value of flux dead zone is off ( ), the rang of flux dead zone is
off ( ) and on ( ).
When flux dead zone is on, if the real-time flux is smaller than the value that max.
flux multiply flux dead zone rang, then the flux tracks to zero.
The default value of flux dead zone rang is 5%, the rang of flux dead zone rang is
0%~99%.
Make sure the belt is running steadily at max. speed at this time, then press
key to start, it displays as follow.
The upper line displays the remain time and the lower displays the amounts of the
pulses that has been checked by the equipment. When calibration has been
completed, press key to save the amounts of the pulses. Press key
to cancel saving the data if the belt runs unsteadily during calibration. Please do
calibration at least thrice if it is the first calibration, if it is not the first calibration
and the belt speed and length are unchanged, you can do zero calibration and span
calibration directly.
Make sure the belt is running steadily at max. speed at this time, then press
key to start, it displays as follow.
The upper line displays the remain time and the lower displays the accumulation.
When calibration has been completed, it displays as follow.
The upper line displays the new zeropoint and the lower displays the error. Then
press key to save the new zeropoint, press key to cancel.
The upper line displays the remain time and the lower displays the accumulation.
When calibration has been completed, it displays as follow.
The upper line displays the actual accumulation. Then input the actual weight in
the lower line via pressing key, key and key, for an example,
input 8.700, it displays as follow.
The upper line displays the error and the lower displays new span.
Let the belt run at this time, then load the material on the belt, it displays as follow.
The upper line displays the real-time flux and the lower displays the accumulation.
Press key when the material goes across the load cell, then it displays as
follow.
The upper line displays the actual accumulation. Then input the actual weight in
the lower line via pressing key, key and key, for an example,
input 8.700, it displays as above. Press key to save or press key
to cancel.
The default value of baud rate is 9600, the rang of baud rate is 4800, 9600, 19200,
38400.
The default value of data format is 8n1, the rang of data format is 8n1, 8n2, 8o1,
8o2.
The default value of time printing is off, the rang of time printing is off, on.
Press key when it is in TIME PRINTING menu, it displays TIME 1 FOR
PRINTING menu.
The format of time is HH-MM-SS. There are four times (default: 00-00-00) for
printing in all. When time printing mode is on, if you just need to print the report
once a day, please set time 1~time 4 as the same. If you just need to print the
report twice a day, please set time 2~time 4 as the same. If you just need to print
the report thrice a day, please set time 3~time 4 as the same.
If you want to print the report at any time, please press key in main menu
(it has not to do with time printing mode).
Press key to confirm, then it displays as follow and returns to main menu.
Press key if you want to load default parameters, then it displays as follow.
Press key to confirm, then it displays as follow and returns to main menu.
Warning: the parameters you have set before in the equipment will be lost
can’’t be used till you
after loading default parameters, then the eqiupment can
input the parameters before and do all calibrations.
The format of time is HH-MM-SS. Change the time with key, key
and key. Press key to confirm, or press key to cancel, then it
displays DATE menu.
The format of date is YY-MM-DD. Change the date with key, key
and key. Press key to confirm, or press key to cancel.
34 Copyright© 2008 Shenzhen Lonxinzhiye Technology Co.,Ltd
Inputing the number of one numberic LED, then press key to switch the
numberic LEDs, and the number you just input will be shielded at the same time,
it displays 0 instead. When you input the correct password, press key to
enter the new password interface.
Input the new password again then press key, it displays as follow.
The default value of analog current is 4~20 mA, the rang of analog current is 0~20
mA, 4~20 mA.
The default value of control current is 4~20 mA, the rang of control current is
0~20 mA, 4~20 mA.
The default value of flux controling mode is auto ( ), the rang of flux
controling mode is auto ( ), serial ( ).
The default value of high flux alarm is 120%, the rang of high flux alarm is
0~200%
The default value of high flux alarm is 0%, the rang of high flux alarm is 0~99%
Meter (Respond)
Address Read Number Word ... Word Check Check
of meter of bytes H L H L
XX 03 XX XX ... XX CRCH CRCL
Note:
If the register address of sending to the equipment is error or CRC check is wrong,
no data will return from the equipment.
e.g.
We know the address of totalization is 9C59H. For an example, totalization =
100.84. In fact, the data type of totalization is long integer, it is 2764H with hex
format, and the value of decimal is 2. If the address of the equipment is 1, then the
read command and the respond from the equipment are as follow.
PC Read: