Professional Documents
Culture Documents
Simultaneous Multiwavelength Detection
Simultaneous Multiwavelength Detection
SUMMARY
The potential of the use of the diode-array detector in conjunction with flow injection
analysis is outlined. Methods for multicomponent resolution can be based on the formation
of complexes absorbing at different wavelengths. The example given is the determination
of mixtures of copper(I1) and iron(II1) with a mixed l:lO-phenanthroline/neocuproine re-
agent. Amplification and dilution methods are based on the sum of the absorbances at
several wavelengths and on monitoring absorbances at wavelengths away from the absorp-
tion maximum, respectively. The determination of nitrite via the Griess reaction is used as
an example; the viable determination range is extended to 0.002-60.0 fig ml-’ nitrite.
Software for the implementation of the suggested methods is outlined.
EXPERIMENTAL
Apparatus
A Hewlett-Packard 8451A diode-array detector was used with the HP-9121
floppy-disk unit, HP-98155A keyboard and HP-7470A plotter. Other equip-
ment included a Gilson Minipuls-2 peristaltic pump, Hellma 178.12-QS flow
cell (inner volume 18 PI), Tecator L-100-1 injection valve and Tecator TM-II
chemifold.
Nitrite determination
The reagent solutions contain 0.95 g of sulphanilamide dissolved in 10 ml
of concentrated hydrochloric acid and diluted to 250 ml with water, and
281
SAMPLE
NaAdNHgOH
o-PHENANTHROLINE
+
NEOCUPROINE
(bl
SAMPLE ! ,
WATER
SULPHANILAMIOE DIODE
+HCl ARRAY
N-II-NAPHTHYLI
ETHYLENEDIAMINE _ !
.NaCI W
Fig. 1. Manifolds used: (a) for resolution of Fe(III)/Cu( II) mixtures, (b) for nitrite deter-
mination. (For q, V, and L values, see text.)
Simultaneous determination
The resolution of the mixture of iron(II1) or iron(I1) and copper(I1) ions is
based on the use of mixed selective reagents; l,lO-phenanthroline and neo-
cuproine were chosen for Fe(II1) and Cu(II), respectively. The working con-
ditions are similar for both complexes, whereas their u.v.-visible absorption
spectra are different.
The calibration equations obtained at the maximum absorption wave-
lengths for the two complexes (454 and 512 nm) were: for iron(I1) or
282
TABLE 1
Variable Meaning
It can be seen that the sensitivity for the Cu(I1) determination is much
lower than that of the conventional method. This is due partly to the fact
that in the conventional method, extraction of the chelate into chloroform
is used to increase sensitivity [ 131 and partly to the fact that the presence of
l,lO-phenanthroline dramatically diminishes the absorbance of the Cu(I)/
neocuproine system (by about 55%), because it competes for the Cu(1).
In the resolution of the mixture, it was necessary to take into account
283
Fig. 2. Flow diagram of the MIXFIA program used for resolution of Fe(III)/Cu(II) mix-
tures. Variables are defined in Table 1.
that the individual absorbances are not additive in the mixture; the overall
absorbance is slightly higher than the sum of the individual values (8-10%).
This synergic effect required mathematical study and clarification. For this
purpose, calibration graphs were constructed for copper(I1) in the presence
of different iron(II1) concentrations. The slopes and intercepts of these
straight lines were a linear function of the iron(II1) concentration. Finally
expressions were established similar to those found in other cases of synergy
[14, 151, but this time at the two different wavelengths:
A 454 = -0.0025[Fe3’] [Cu”‘] + 0.1307[Fe3’] + 0.03407[Cuz’] - 0.2776
A 512 = -0.0033[Fe3’] [Cu”‘] + 0.1628[Fe3’] + 0.01493[Cuz’] - 0.2836
284
TABLE 2
TABLE 3
Variable Meaning
Fig. 3. Flow diagram of the DINAFIA program for normal, dilution and amplification
methods. Variables are defined in Table 3.
oh), then it requests the injection into the flow-injection system of each of
the standards for the three calibration graphs (dilution, normal and amplifi-
cation method in the above-mentioned ranges). Afterwards, the features of
each calibration graph are calculated and, finally, the program evaluates the
nitrite concentration in the unknown samples, using the appropriate calibra-
tion graph in each case.
The calibration graphs used for nitrite determination were obtained by
monitoring the reaction product at 450 nm (dilution method), 542 nm (nor-
mal method) and the set of Z:h15 = 528, 530, . , .554, 556 nm (amplification
method). Table 4 lists the characteristic of these graphs and several others
for the dilution and amplification methods. The following conclusions can
be drawn. The range of determinations is extended from 0.2-4.0 pg ml-’ for
the conventional method to 0.002-60.0 pg ml-’ for the method suggested
here. Although the possible range of determination is considerably shortened
as a result of using the amplification plots instead of the dilution plots, the
linearity ratio (quotient between the upper and lower determination limits)
is very similar in every case. The sensitivity of the procedure, as well as the
manipulation factor (ratio between the slopes of the dilution or amplifica-
tion calibration graphs and the normal graphs) in the dilution calibration
increases with the proximity of hdll to h,,,, whereas in the amplification
calibration it increases with the number of wavelengths considered.
The relative errors found on injecting 15 samples with concentrations ly-
ing over a very wide range into the flow-injection system, are listed in
Table 5. Such errors are small for all the samples, but even smaller in the
normal method (average relative error +0.28%) than in the dilution method
(?1.58%) or the amplification method (?1.50%). The precision of the
method was calculated separately from 11 different samples for each curve.
The relative standard deviations obtained and the concentrations used were
TABLE 4
aThe regression coefficient was 0.999 in all cases. bZh, = 540, 542, 544 (nm). CCh,, =
528, 530,. .., 554, 556 (nm). d xkz5 = 518,520,. . . ,564,566 (nm).
287
TABLE 5
as follows: dilution, +0.40% (25 pg ml-‘); normal method, *0.13% (2.0 c1g
ml-‘); amplification method, +O.lO% (0.15 pg ml-‘).
REFERENCES