Professional Documents
Culture Documents
Murakami 2006
Murakami 2006
that might interfere with the results. All the specimens using a special device. The tensioner (Alfresa pharma,
were then frozen at 20 1C until the day of testing. After Osaka, Japan) had a graded scale to apply tension of
thawing to room temperature for 24 hours, the surround- 200 N to the cable (Fig. 3B), (3) Double tape from T7 to
ing soft tissue and muscles were dissected meticulously to T10 was placed sublaminarly and fixed to the rectangle
preserve the osseous and ligamentous structures before and tightened to 200 N of tension using a special device
testing. The specimens were kept moist throughout the (Fig. 3C). Every test was carried out twice and successive
entire testing procedure. load/deformation curves were nearly identical (this is
The intact specimens, which were placed in pots usually the case because the specimens were precondi-
containing dental cement, were aligned in the cement to tioned). From the load/deformation curves, stiffness
make the cranio-caudal ends of the pots parallel. These values were calculated by measuring the slope of the
were then clamped on a biomechanical testing machine linear part of the curve. This method is explained more
(MTS 858 Mini-Bionix Test System, Minneapolis, MN) fully elsewhere.7 Normalized stiffness values were calcu-
in the upright position. The intact specimen was first lated by dividing the stiffness obtained from the experi-
cyclically preloaded by axial compression (150 ± 50 N at mental specimen by the stiffness obtained from the intact
1 Hz for 15 min) to remove excess fluid from the discs and specimen (ie, ratio of instrumented to intact).
ligaments. This preconditioning is a normal procedure in Statistics: A 1-way repeated measures analysis of
biomechanical testing of cadaveric spines in our labora- variance was used to compare the results of the intact
tory; it allows the disc height to get back to the normal specimens and those fixed by steel wire, cable, or tape in
physiologic height and also to yield reproducible stiffness each loading sequence, followed by Tukey’s multiple
curves during subsequent loading. Biomechanical testing comparison procedure. The overall level of significance
of the intact specimen was then carried out according to was set at a P<0.05.
the following sequence: axial compression (250 N), flexion
(7.5 Nm), extension (7.5 Nm), left lateral bending RESULTS
(7.5 Nm), right lateral bending (7.5 Nm), left axial torsion
(10 Nm), and right axial torsion (10 Nm). This biomecha- Experiment 1: Laminar Cut-through Test
nical testing procedure is explained in more detail On the basis of the results from the 2-factor analysis
elsewhere.5 Destabilization was then performed by of variance, there was no statistically significant interac-
sectioning the supraspinatous ligament, interspinous tion between wire type and whether the laminar was
ligament, ligamentum flavum, facet joint capsules, poster- decorticated. Overall, the average cut-through force
ior longitudinal ligament, and the posterior part of the whether the lamina was decorticated or not was margin-
anulus fibrosus.6 The specimens were then restabilized ally insignificant (P = 0.06). However, there was a
using a Hartshill rectangle (with rod 5 mm diameter) statistically significant difference in the average cut-
(Surgicraft, Redditch, UK) plus 1 or other of the same 3 through force depending on the steel wire (P = 0.002).
different types of sublaminar wires as described in On the basis of Tukey’s pairwise comparison using a 0.05
Experiment 1. The implants used with the Hartshill error rate, steel wire was significantly different from both
rectangle were constructed as follows: (1) Single steel cable and tape. There was no significant difference
wires from T7 to T10 were placed sublaminarly and fixed between cable and tape.
to the rectangle by twisting the ends together (Fig. 3A), The average BMD, thickness of the laminar cortex,
(2) Cable from T7 to T10 was placed sublaminarly and and the average value of force at the point of 5 mm cut-
fixed to the rectangle and tightened to 200 N of tension through on the lamina in each of the 3 groups tested using
the 3 types of sublaminar wires are shown in Table 1. In
the test using steel wire, the average BMD for the 5
vertebrae was 0.79 ± 0.04 g/cm2. The average thickness of
the 5 laminar cortices was 3.72 ± 0.72 mm, and the
average cut-through force when the laminar cortex was
intact was 566.8 ± 82.2 N, and 501.0 ± 55.4 N when the
laminar cortex was decorticated.
In the test using cable, the average BMD for the
5 vertebrae was 0.79 ± 0.03 g/cm2. The average thickness
of the 5 laminar cortices was 3.96 ± 0.48 mm, and the
average cut-through force when the laminar cortex was
intact was 808.6 ± 144.4 N, and 610.8 ± 153.8 N when
FIGURE 3. Experiment 2: Biomechanical stiffness test. The the laminar cortex was decorticated.
specimens were restabilized from T7 to T10 using a Hartshill
rectangle plus sublaminar wire. A, sublaminar single steel In the test using tape, the average BMD for the
wires fixed to the rectangle by twisting the ends together; 5 vertebrae was 0.78 ± 0.05 g/cm2. The average thickness
B, Sublaminar cable fixed to the rectangle and tightened of the 5 laminar cortices was 3.67 ± 0.72 mm, and the
to 200 N of tension using a special tightening device; C, average cut-through force when the laminar cortex was
Sublaminar double tape fixed to the rectangle and tightened intact was 1164.7 ± 57.6 N, and 911.3 ± 7.0 N when the
to 200 N of tension using a special tightening device. laminar cortex was decorticated.
Experiment 2: Biomechanical Stiffness Test values. In axial torsion, we averaged the results for right
In none of the experiments did the bone or the and left.
sublaminar wire break. All data are shown normalized as
a relative ratio to the results obtained from the intact DISCUSSION
specimens (Fig. 4). In axial compression, the average The main results to emerge from these experiments
normalized values of stiffness were all less than intact are as follows: (1) Experiment 1; in the laminar cut-
(0.88 steel wire, 0.76 cable, and 0.99 tape). There is no through test, the cut-through force for steel wire was
significant difference between any of these 3 values. significantly lower than that for either tape or cable. The
In flexion, the average normalized values of stiffness BMD and thickness of the laminar cortex was similar for
were 1.11 steel wire, 1.20 cable, and 1.04 tape. There was each of 3 experimental groups. Further, whereas the
no significant difference between any of these 3 values. average cut-through force for the intact lamina was
In extension, the average normalized values of higher than that for the decorticated lamina for all 3 types
stiffness were 1.11 steel wire, 1.05 cable, and 1.07 tape. of wires it was not statistically significant at the 0.05 level.
There was no significant difference between any of these 3 (2) Experiment 2; in the biomechanical stiffness test, there
values. was no significant difference between the 3 different
In lateral bending, the average normalized values of sublaminar wiring constructs in any of the loading modes
stiffness were 1.08 steel wire, 0.98 cable, and 1.0 tape. tested.
There was no significant difference between any of these 3 Butler et al8 reported that BMD correlated with
values. In lateral bending, we averaged the results for strength and stiffness of the sublaminar wire construct in
right and left. osteoporotic spines. Andingoz et al9 reported that there
In axial torsion, the average normalized values of was positive linear relationship between BMD and
stiffness were 1.76 steel wire, 2.36 cable, and 1.48 tape. laminar strength in sublaminar wiring. Biomechanical
There was no significant difference between any of these 3 studies have reported that the strength of the decorticated
lamina was less than that of the intact lamina in ‘‘pull-
through’’ force using sublaminar wire.9,10 Actually, in the
setting of osteoporotic bone, especially in the decorticated
lamina, migration of sublaminar wire through soft
laminar bone occurs frequently. We considered BMD
and the thickness of laminar cortex as major factors
determining the fixation strength of a sublaminar wiring
construct. Normalizing the properties of the vertebrae
(eg, BMD and thickness of the laminar cortex) as we did
in each of the 3 groups is necessary to make fair
comparisons between the 3 different types of sublaminar
wiring constructs. Thus, the results from the cut-through
test (Experiment 1) suggest that tape has the advantage in
securing the osteoporotic lamina, as compared with steel
wire or cable. The narrow contact surface between the
steel wire and cable produces a high-local stress on the
lamina. This high-local stress results in lower values of
cut-through force. Tape, on the other hand, is broad and
FIGURE 4. Results for Experiment 2. All data are shown so distributes the force better (ie, lower local stress), thus
normalized as a relative ratio to the results obtained from the allowing the lamina to resist a higher cut-through
intact specimens. force. The results from the biomechanical stiffness test
(Experiment 2) revealed no difference (in terms of between tape and cable could be established. In the
stiffness) between the 3 types of sublaminar wire. That biomechanical stiffness test, there was no significant
may be a true result or may be because of the lack of difference between the 3 different sublaminar wiring
power with such a small sample. It is difficult to obtain configurations in any of the loading modes tested. The
large numbers of cadaver spines that are suitable for results of both experiments suggest that tape is as good, if
experiments such as these. Further, it could be that not better, than steel wire and, perhaps, cable as a
stiffness may not be a good indicator for showing sublaminar wiring implant material.
differences between spinal constructs, unless the differ-
ences are large.
REFERENCES
The following is the best information we could
1. Eysel P. Biomechanical principles of ventral and dorsal instrumen-
obtain on the mechanical properties of the 3 different tation correction in scoliosis. Orthopade. 2000;29:507–517.
types of sublaminar wires. Dickman et al11 reported that 2. Yazar T, Gurkan I, Yilmaz C. A new approach to scoliosis. Eur
polyethylene cable exhibited creep at loads that were Spine J. 1999;8:86–92.
much lower than the static failure loads, namely 3. Hu SS. Internal fixation in the osteoporotic spine. Spine.
1997;22:43S–48S.
polyethylene cable 1.0 mm diameter stretched 43% at 4. WHO. Assessment of fracture risk and its application to screening
800 N force, although it had similar strength to stainless for postmenopausal osteoporosis. Report of a WHO Study Group.
steel cable in the tensile testing. Also they reported that World Health Organ Tech Rep Ser. 1994;843:1–129.
titanium cable 1.2 mm diameter stretched 6% at 800 N 5. Murakami H, Boden SD, Hutton WC. Anterior lumbar interbody
forces, and monofilament steel wire 0.8 mm diameter fusion using a barbell-shaped cage: a biomechanical comparison.
J Spinal Disord. 2001;14:385–392.
stretched 9% at 250 N. Creep can cause a sublaminar wire 6. Abumi K, Panjabi MM, Kramer KM, et al. Biomechanical
construct to loosen. We used tape made of woven ultra- evaluation of lumbar spinal stability after graded facetectomies.
high molecular weight polyethylene fibers. This tape Spine. 1990;15:1142–1147.
allows little stretch and creep under tensile force, as 7. Volkman T, Horton WC, Hutton WC. Transfacet screws with
lumbar interbody reconstruction: biomechanical study of motion
compared with standard polyethylene cable; the material segment stiffness. J Spinal Disord. 1996;9:425–432.
properties for the ultra-high molecular weight polyethy- 8. Butler TE Jr, Asher MA, Jayaraman G, et al. The strength and
lene tape were supplied by Alfresa pharma. Furthermore, stiffness of thoracic implant anchors in osteoporotic spines. Spine.
the tape we used is soft and flexible; this property helps to 1994;19:1956–1962.
9. Aydingoz O, Bilsel N, Botanlioglu H, et al. Effect of decortication
avoid neural damage while passing the tape under the on laminar strength during sublaminar wiring: an experimental
lamina. study. J Spinal Disord Technol. 2004;17:498–504.
10. Crawford RJ, Sell PJ, Ali MS, et al. Segmental spinal instrumenta-
CONCLUSIONS tion. A study of the mechanical properties of materials used for
sublaminar fixation. Spine. 1989;14:632–635.
In the laminar cut-through test, the cut-through 11. Dickman CA, Papadopoulos SM, Crawford NR, et al. Comparative
force for tape was significantly higher than for wire in mechanical properties of spinal cable and wire fixation systems.
these osteoporotic vertebrae, but no significant difference Spine. 1997;22:596–604.