Professional Documents
Culture Documents
Corrosion Inhibitor For Beginner
Corrosion Inhibitor For Beginner
Corrosion Inhibitor For Beginner
https://www.linkedin.com/in/bony-budiman-39886031/
COMMON QUESTIONS
• OBJECTIVE
• INTRODUCTION
• INTERNAL CORROSION MITIGATION
• FILM FORMING CORROSION INHIBITOR
• SELECTION
• FIELD APPLICATION
• MONITORING
• FIELD CASES
• SUMMARY
OBJECTIVE, SCOPE AND DISCLAIMER
• OBJECTIVE
• INTRODUCTION
• INTERNAL CORROSION MITIGATION
• FILM FORMING CORROSION INHIBITOR
• SELECTION
• FIELD APPLICATION
• MONITORING
• FIELD CASES
• SUMMARY
CYCLE OF IRON
+ C, Ni, Mn, Cr, V etc
Fe3C.”X”
FexOy
• Loss of production
• Economic : maintenance and replacement
• Environment
• Fatality
TESORO FATALITY
• OBJECTIVE
• INTRODUCTION
• INTERNAL CORROSION MITIGATION
• FILM FORMING CORROSION INHIBITOR
• SELECTION
• FIELD APPLICATION
• MONITORING
• FIELD CASES
• SUMMARY
INTERNAL CORROSION
MITIGATION
Mechanical Chemical
Material Regular
Selection Cleaning
CHEMICAL TREATMENT STRATEGY
• OBJECTIVE
• INTRODUCTION
• INTERNAL CORROSION MITIGATION
• FILM FORMING CORROSION INHIBITOR
• SELECTION
• FIELD APPLICATION
• MONITORING
• FIELD CASES
• SUMMARY
FILM-FORMING CORROSION INHIBITORS (FFCI)
• Formula will depend on fluid composition that will be treated, field condition (temperature
and pressure), injection facilities and price as well
• Common composition :
• Active content such as imidazoline and or quaternary amine
• Surfactant to increase absorption performance
• pH adjuster such as glacial acetic acid (GAA) and dimer trimer acid (DTA) to increase its
effectiveness
• Mutual solvent to increase solubility in water such as isopropyl alcohol (IPA) and butyl cellosolve
(EGMBE)
• Solvent : commonly water as the cheapest solvent, to increase stability at high temperature
condition, formulator may add glycol.
FORMULA FOR GAS WITH WATER(2)
• For these pipelines with water, a volatile amine should be mixed with a water dispersible
film forming amine
• An example of formulation :
• Imidazoline : 15 – 25 %
• GAA : 7.5 – 12.5 %
• Methoxypropyl amine : 14 – 18 %
• Solvent (include mutual solvent) : Rest
FORMULA FOR GAS, OIL AND WATER(2)
• For these systems the two phase inhibitor as shown below is the obvious choice.
• Formula may be as below :
• Imidazoline : 15%
• DTA : 7.5%
• GAA : 6.0%
• Solvent : 71.5%
FORMULA FOR WATER ONLY(2)
• It looks simple since just need to ensure the FFCI soluble in water but it may tricky in
the effectiveness since the presence of hydrocarbon will reduce the corrosion rate.
Hydrocarbon will thicken the film protection.
• The formulation may be as follow :
• Imidazoline : 15 – 25 %
• Surfactant NP-100 : 2 - 5%
• IPA or EGMBE : 15 – 25 %
• GAA : 12 - 14%
• Water : Rest
OUTLINE
• OBJECTIVE
• INTRODUCTION
• INTERNAL CORROSION MITIGATION
• FILM FORMING CORROSION INHIBITOR
• SELECTION
• FIELD APPLICATION
• MONITORING
• FIELD CASES
• SUMMARY
QUALIFICATION
TEST
Field
Storage Toxicity Biodegradability Bioaccumulation
Condition
• Material Compatibility
• Metal
• Non Metal
• Fluid Compatibility
• Production Fluid
• Chemical
• A series of tests should be carried out utilizing the CI and a variety of specimens that may contact
from the storage tank, daily tank, injection line, injection point and pipeline-flowline.
• Testing should focus first on the highest expected temperature to assess possibility of attack in the
most aggressive environment.
• The metal material could be as carbon steel, stainless steel, cooper, or alloys.
• The non metal material could be as plastic, rubber or composite.
MATERIAL COMPATIBILITY - PROCEDURE
• The compatibility of CI shall be check with any fluid that may contacted such as produced fluid or others
chemicals
• Produced fluid especially with produced water and liquid hydrocarbon
• Compatibility shall be observed between CI and any chemicals (such as Scale Inhibitor, Biocide, Oxygen
Scavenger, Reverse Demulsifier etc) that injected in the same stream/line and has possibility to contact
each other
• Observed parameter : change of color, precipitation, emulsion, heat increase, chemical change (by finger
print test)
STABILITY - DEFINITION
Stability means that the properties of CI not significant changes under certain condition.
The properties could be phase separation, color, viscosity, pour point and effectiveness
• Efficiency means how effective the CI in the fluid could inhibit corrosion rate compare to non
treated fluid.
• The corrosion rates can be measured using mass loss or electrochemical methods.
• Using the methodologies, several variables, compositions of material, composition of
environment (gas and liquid), temperature, pressure, and flow, that influence the corrosion
rate in the field can be simulated in the laboratory.
• The performance of a corrosion inhibitor is influenced primarily by the nature of inhibitor,
operating conditions of a system, and the method by which it is added. The lab tests shall
consider this such as the temperature, pressure, shear stress, oil-water composition, corrosive
gas concentration (CO2 and H2S) and ions composition in the brine water.
STATIC IMMERSION TEST
Courtesy pictures
: https://www.researchgate.net/figure/The-setup-of-the-rotating-cylinder-electrode-
RCE_fig2_256913871 and https://pineresearch.com/blog/researchers-corrosion-electrochemistry/
FLOW LOOP TEST
• .In this test, the inhibitor was only added after the corrosion rate had
stabilized (4 days), but in subsequent tests the inhibitor was added
immediately before coupling to save time.
• In all tests, within an hour of adding the inhibitor the polarity had reversed
so that the array electrodes were now anodic to the external electrode. In
Test 2, a peak average corrosion rate of around 4 mm/y was reached after
about 1 day.
• The severe anodic polarization of the array results from the more rapid
adsorption of the inhibitor on the external electrode, stabilizing the FeCO3
film, and the restricted access to the array because of the sand.
• It took about 12 days for the average corrosion rate of the array electrodes
to reach the acceptable value of 0.1 mm/y, which is related to the transport
properties of the inhibitor through the sand.
TEST MATRIX
JIT
OIL-WATER PARTITIONING
• Most of CI contain surfactant to enhance its effectiveness. Surfactant may cause problem
such as create emulsion and foaming.
• The emulsions formed can be quite difficult to remove and can lead to separation
difficulties in the production facilities.
• Shake flask tests are used to evaluate whether the CI will cause the water/hydrocarbon
mixture to form an emulsion.
• Volumes of 100-mL (or other suitable volume) mixtures of oil phase and water phase
are tested. The ratios an be used depending on the system, for example.
• Inject CI at various dosage (from 0 ppm as blank until the 2 or 3 times of field dosage).
Hard shake minimum 100 times.
• Observe the separation, the emulsion layer and time required to separate.
• Is the capability of a corrosion inhibitor to form a stable and protective film able to
mitigate corrosion without being renewed. A high persistency is mandatory for corrosion
inhibitors to be used in the batch mode; however, persistency is a useful property also for
inhibitors for continuous treatments.2
• Persistency can be assessed in laboratory tests measuring the corrosion rate after
removal of the solution containing the inhibitor and replacing with the same solution free
of corrosion inhibitor; the time the system increases up to the baseline corrosion value is
a measure of the inhibitor persistency.
ENVIRONMENT EVALUATION
• OBJECTIVE
• INTRODUCTION
• INTERNAL CORROSION MITIGATION
• FILM FORMING CORROSION INHIBITOR
• SELECTION
• FIELD APPLICATION
• MONITORING
• FIELD CASES
• SUMMARY
APPLICATION
• Continuous
• Non continuous
• Batching
• Squeeze
• Stick
CONTINUOUS INJECTION
Courtesy pictures Handout Oilfield Production Chemicals and Microbiology, HTS Consultants, 2007
CONTINUOUS INJECTION : DISPERSION
• This is common to applied in close system or semi close system when the water base liquid
mostly recirculated with minimum lose such as at Close Cooling Water System
• Also known as slug dose method.
• This method may also applied as additional treatment beside the continuous in order to provide
thicker protection layer.
• For Close Cooling Water with nitrite as active content, the injection rate is 1-5 liter CI per 2,500
liter water with interval batching every 1-4 week(s)
• For oil and gas field with amine as active content, the dosage is between 1000 – 5000 ppm for 3-6
hours injection
BATCHING : USING PIG
Courtesy pictures Handout Oilfield Production Chemicals and Microbiology, HTS Consultants, 2007
VOLUME OF SQUEEZE TREATMENT
• OBJECTIVE
• INTRODUCTION
• INTERNAL CORROSION MITIGATION
• FILM FORMING CORROSION INHIBITOR
• SELECTION
• FIELD APPLICATION
• MONITORING
• FIELD CASES
• SUMMARY
MONITORING
Reference : NACE STD SP0775-2018 : Preparation, installation, analysis and interpretation of corrosion coupons in oilfield operations.
UNIT CONVERSION OF CORROSION RATE
• Water Analysis
• Residual CI
• Total iron
CORROSION COUPON (1)
* https://inspectioneering.com/tag/pigging#:~:text=Intelligent%20Pigging%20is%20an%20inspection,inner%20walls%20of%20the%20pipe.
Courtesy pictures : https://www.openpr.com/news/2095424/intelligent-pigging-market-is-expected-to-see-growth-rate
RESIDUAL CORROSION INHIBITOR
• Most of the CI’s molecules will form layer with the inner side of
metal, some molecules will be degraded due to reaction with
others chemicals or by relative high temperature and the others
will be as free molecule in water known as residual CI (the rest of
CI’s molecules).
• High residual means the amount of CI is enough or even
excessive. Low residual means the amount of CI is not sufficient
or CI may not compatible with other chemicals or not stable with
the field condition
• Residual CI at field may different depend on type of active content
and treated system. For example residual amine at pipeline may
vary from 5 to 20 ppm while residual nitrite at close cooling water
system shall be keep between 400 – 1000 ppm
• Residual CI could be measured by spectrophotometric method.
TOTAL IRON
*NACE Standard RP0192-98 Item No. 21053 Monitoring Corrosion in Oil and Gas Production with Iron Counts
OUTLINE
• OBJECTIVE
• INTRODUCTION
• INTERNAL CORROSION MITIGATION
• FILM FORMING CORROSION INHIBITOR
• SELECTION
• FIELD APPLICATION
• MONITORING
• FIELD CASES
• SUMMARY
1. COMPARISON OF TWO TYPE OF CI(10)
• The FF-NCI was injected in well A, meanwhile, the FFCI was used in
the well B.
• Each inhibitor was injected in the first pipe (testing pipe)
downstream of the choke using a dosing pump as shown in picture
beside
• The continuous treatment is carried out in two stages:
• The 1st is to make a shock treatment required for forming a layer of
the inhibitor on the testing pipe surface (the period of this step
reached 24 h with an injection rate of 20 g/m2/d),
• The 2nd step is to maintain the inhibitor film; the concentration of
the inhibitor is reduced to a low dose in this mode of treatment in
order to optimize its injection rate
1. COMPARISON OF TWO TYPE OF CI(10)
• A French refinery faced tremendous corrosion and safety issues in the overhead system at their sour
water stripper.
• Severe corrosion was found in the overhead, but was also seen in the overhead lines, water drum, and
reflux line.
• Corrosion in the reflux line, especially downstream of the reflux pump, measured, on average, 40 mils per
year.
• 10-15% concentration of NH4HS found in the reflux line sample. Typically, any concentration greater
than 4% indicates the potential for corrosion.
• Deposit analysis confirmed that the problem in this case was primarily due to NH4HS salts and the
associated under deposit corrosion.
2. CORROSION INHIBITOR FOR SOUR SYSTEM(11)
• OBJECTIVE
• INTRODUCTION
• INTERNAL CORROSION MITIGATION
• FILM FORMING CORROSION INHIBITOR
• SELECTION
• FIELD APPLICATION
• MONITORING
• FIELD CASES
• SUMMARY
SUMMARY
• Internal corrosion at oil and gas field usually caused by dissolved gas such as O2, CO2 and H2S
• Film Forming Corrosion Inhibitor used to reduce corrosion rate by produce dynamic film protection
against corrosion dissolved gas.
• FFCI is a specialty chemicals. Composition may different for each field depends on corrosion gas
concentration, fluid composition, temperature, pressure and materials used.
• Selection of Corrosion Inhibitor usually include the performance, stability and compatibility
• Application of Corrosion Inhibitor depends on corrosion location, type of corrosion, previous field
experiences and the injection facilities
• Monitoring corrosion will provide data for continues improvement in term of quality and cost.
• No single monitoring method could be used to explain the corrosion phenomenon at each field.
Combination of some method will provide better understanding
REFERENCES
1. Production Chemicals for the Oil and Gas Industry, Malcolm A.Kelland, CRC Press, page 195-217
2. Product Bulletin of Akzo Nobel Surfactants, Corrosion Inhibitors for Oilfield Production
3. Investigation Report Catastrophic Rupture Of Heat Exchanger (Seven Fatalities), U.S. Chemical Safety And Hazard
Investigation Board, 2014
4. ASTM G170 Standard Guide for Evaluating and Qualifying Oilfield and Refinery Corrosion Inhibitors in the Laboratory
5. NACE Item No 24259 Laboratory Evaluation of Corrosion Inhibitors Used in Oil and Gas Industry
6. https://shodhganga.inflibnet.ac.in/bitstream/10603/13315/7/07_chapter%203.pdf
7. https://www.icorr.org/wp-content/uploads/2020/06/2011-03-15-Clariant_Corrosion-Inhibitors-and-Corrosion-Inhibitor-
Selection.pdf
8. https://core.ac.uk/download/pdf/140429.pdf
9. https://www.researchgate.net/publication/273755924_Testing_of_inhibitors_for_underdeposit_corrosion_in_sour_conditions
10. https://doi.org/10.1007/s13202-021-01287-y
11. Case Study from SUEZ : pHilmPLUS* 5K15 Filming Inhibitor Resolves Corrosion and Safety Concerns in Sour Environment