Autotransfusion Setup Instruction
Autotransfusion Setup Instruction
I In-line connector
P Patient connector
P P P
Q Easy-to-grip handle
O O O
A
Setup for continuous ATS with a blood compatible infusion pump
B
For direct reinfusion of shed autologous blood via a blood compatible infusion pump, a microemboli blood filter
C E and non-vented, blood compatible I.V. administration set must be used. Please refer to all pump manufacturer’s
D Directions for Use and Warnings and Cautions prior to use.
F
A I.V. set to patient For priming infusion pumps, filter and I.V. set can be primed by aspirating air out of the I.V. circuit with a three-
B Access line clamp way stopcock and syringe or pre-primed with saline prior to attachment to the blood filter.
C Microemboli blood filter
D Drip chamber
1. Clamp ATS access line. 6. Connect 60 cc Luer-lock 11. When drip chamber is ¼ full,
E Infusion pump 2. Place ATS access line around syringe to side port turn “spike up” and continue
ATS
F I.V. set to pump access hanger or patient line before of stopcock. purging air from line.
line
G ATS access line spiking filter. 7. Open free flow setting on 12. When completed, insert I.V.
3. Remove spike port cap and pump and cassette. cassette into pump.
Microemboli
blood filter
insert blood filter. 8. Turn filter in “spike down” 13. Purge all air prior to patient
4. Spike a non-vented blood position. connection.
administration I.V. set 9. Unclamp ATS access line 14. Set pump to desired “volume
Drip
chamber
into filter. and I.V. tubing. to be infused” and “ml
5. Attach 3-way stopcock to 10. Use 60 cc syringe to per hour”.
patient end of I.V. set. aspirate blood.
Have a question or need help in a hurry? Call Getinge toll free at 1-800-528-7486.