Download as pdf or txt
Download as pdf or txt
You are on page 1of 14

<<Contents>> <<Index>>

General Model EJA210 and EJA220


Specifications Flange Mounted Differential Pressure
Transmitters
GS 01C22C01-00E [Style: S2]

The model EJA210 and EJA220 Flange Mounted Differen-


tial Pressure Transmitter can be used to measure levels or
densities of solidifying or precipitating liquids. The
transmitter outputs a 4 to 20 mA DC signal.
The model EJA210 and EJA220 also features remote setup
and monitoring through communications with the model
BT200/100 BRAIN TERMINAL, or CENTUM CS/XL system
etc.

■ STANDARD SPECIFICATIONS
Refer to GS 01C22T02-00E for Fieldbus communication
type marked with “䉫.”
Measurement Ranges:

Capsule Measurement Span Measurement Range


2.5 to 100 kPa -100 to 100 kPa
M {250 to 10000 mmH2O} {-10000 to 10000 mmH2O}
25 to 500 kPa -500 to 500 kPa Ambient Temperature Effect:
H {0.25 to 5 kgf/cm2} {-5 to 5 kgf/cm2} See Table 1-1, 1-2 and 1-3.
T01E.EPS
Process Temperature Limits:
-40 to 120 ⬚C (-40 to 248 ⬚F) (general use type)
Output Signal “䉫”: (See ‘Optional Specifications’ for Explosion-protected
4 to 20 mA DC, 2-wire system with digital communication types)
Failure Alarm: Ambient Humidity Limits:
Output status at CPU failure and hardware error; 5 to 100 % R.H.(at 40 ⬚C)
Up-scale: 110%, 21.6 mA DC or more(standard)
Down-scale: -5%, 3.2 mA DC or less Working Pressure Limits:
-2.5%, 3.6 mA DC or less (Optional code /F1) 2.7 kPa abs{20 mmHg abs} to flange rating pressure
Note: Applicable for Output signal code D and E For atmospheric pressure or below, see Figure 2.
Supply Voltage “䉫”: Static Pressure Effect:
10.5 to 42 V DC for general use and flameproof type See Table 1-1, 1-2 and 1-3.
10.5 to 32 V DC for lightning protector (Optional code /A) Power Supply Effect “䉫”:
10.5 to 30 V DC for intrinsically safe, Type n, ⫾0.005 %/V (21.6 to 32 V DC, 350 ⍀)
nonincendive, or non-sparking type
Mounting Flange Rating:
10.5 to 28 V DC for TIIS intrinsically safe type
See ‘Model and Suffix Codes.’
Conditions of Communication Line “䉫”: Flange conforming to ANSI are serration-worked on the
Power supply voltage; 16.4 to 42 V DC gasket surfaces (ANSI B16.5)
Load resistance; See Figure 1. Note: For model EJA210 wetted parts material code H or T,
Note: In case of an intrinsically safe transmitter, external no serration is worked.
load resistance includes safety barrier resistance.
Degrees of Protection:
Communication distance;
2 km, when CEV polyethylene-insulated PVC-sheathed IP67, NEMA4X
control cables are used. Explosion-protected Construction:
Note: Communication distance varies depends on kind of See ‘Optional Specification.’
cable.
Electrical Connection:
Load capacitance; 0.22 ␮F or less.
See ‘Model and Suffix Codes.’
Load inductance; 3.3 mH or less.
Spacing from power line; 15 cm or more. Low Pressure Process Connection:
Input impedance of receiver connected receiving resistance; See ‘Model and Suffix Codes.’
10 k⍀ or more at 2.4 kHz Wetted Parts Material:
(See Optional Specifications for Intrinsically safe type) High pressure side diaphragm;
Accuracy: See ‘Model and Suffix Codes.’
See Table 1-1, 1-2 and 1-3 and Figure 3. Capsule gasket; Teflon-corted JIS SUS316L
Ambient Temperature Limits: Process connector gasket; PTFE (Teflon)
-40 to 85 ⬚C (-40 to 185 ⬚F) (general-use type) Flange Bolts Material:
-30 to 80 ⬚C (-22 to 176 ⬚F) (with integral indicator) See ‘Model and Suffix Codes.’
(See ‘Optional Specifications’ for Explosion-protected types)

Yokogawa Electric Corporation GS 01C22C01-00E


2-9-32 Nakacho, Musashino-shi, Tokyo, 180-8750 Japan ©Copyright July 1994
Phone: 81-422-52-5690 Fax: 81-422-52-2018 28th Edition April 2012
<<Contents>> <<Index>> 2
Amplifier Housing: External Zero Adjustment “䉫”:
Cast aluminum alloy or JIS SCS14A stainless steel Continuously adjustable Resolution; 0.01 % of span
(optional) Zero Adjustment Limit:
Painting: Zero can be fully elevated or suppressed as long as low
Polyurethane resin baked finish and high range value are within the measurement range
Deep sea moss green (Munsell 0.6GY3.1/2.0) limits of the capsule.
Integral Indicator: Tag Plate:
LCD digital indicator (optional) JIS SUS304 or SUS316 (optional).
Damping Time Constant: Weight:
(Sum of time constants for amplifier assembly and capsule 9.4 kg (20.8 lb) Model EJA210 with 80 mm JIS 10K
assembly) flange, without integral indicator and process connector.
Amplifier assembly time constant; 13.9 kg (30.7 lb) Model EJA220 with 100 mm JIS 10K
Can be set in 9 increments from 0.2 to 64 sec flange, X2= 100 mm without integral indicator and process
Capsule assembly time constant; connector.
Add 1.4 kg (3.1 lb) for JIS SCS14A stainless steel
Model EJA210 EJA220 amplifier housing.
Capsule M H M H
Approx. Approx. Approx. Approx.
EMC Conformity Standards “䉫”: ,
Time Canstant
(sec) 0.4 0.4 0.4 0.4 EN61326-1 Class A, Table2(For use in industrial loca-
T02E.EPS
tions)
EN61326-2-3
European Pressure Equipment Directive 97/23/EC:
Sound Engineering Practice

< Settings When Shipped > “䉫”

Tag Number As specified in order *1 Calibration Range


Lower Range Value As specified in order
Output Mode ‘Linear’
Calibration Range As specified in order
Display Mode ‘Linear’ Higher Range Value
Operation Mode ‘Normal’ unless otherwise specified in order Selected from mmH2O, mmAq, mmWG,
Calibration Range mmHg, Pa, hPa, kPa, MPa, mbar, bar,
Damping Time ‘2 sec.’ Units gf/cm2, kgf/cm2, inH2O, inHg, ftH2O, or psi.
Constant (Only one unit can be specified)
T05E.EPS
*1: Up to 16 alphanumeric characters (including - and · ) will be entered in the amplifier memory.

100{750} Atmospheric
600 pressure
Working
E-10.5 pressure
R=
0.0236 Digital KPa abs
External Communication {mmHg abs}
load range
Applicable range
resistance BRAIN and HART
10{75}
250

R (⍀)
2.7{20}

1{7.5}
10.5 16.4 24.7 42 -40 0 40 80 120
Power supply voltage E (V DC) Process temperature (⬚C)
F02E.EPS F02E.EPS
Figure 1. Relationship Between Power Supply Voltage and Figure 2. Working Pressure and Process Temperature
External Load Resistance

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 01C22C01-00E Oct. 01, 2008-00
<<Contents>> <<Index>> 3
Table 1-1 . Accuracy, Ambient Temperature Effect and Static Pressure Effect (As Percent of 'x') *1
[For model EJA210 3-inch (80 mm) flange and model EJA220 4-inch (100 mm) flange
(wetted parts material code S )]

Capsule M, H
⫾0.2 % For x ≥ Pref
Accuracy Pref
⫾(0.1 + 0.1 ⫻ x ) % For x < Pref

Zero 2Pref
shift ⫾(0.3 + 0.35 ⫻ x ) % / 50 ⬚C
Ambient
temperature
effect
Total ⫾1.0 % / 50 ºC For x ≥ 2Pref
shift *3 2Pref
⫾(0.5 + 0.5 ⫻ x ) % / 50 ⬚C For x < 2Pref

Zero ⫾0.1 % / 0.98 MPa{10 kgf/cm2} For x ≥ Pref


shift Pref
Static ⫾(0.1 ⫻ x ) % / 0.98 MPa{10 kgf/cm2} For x < Pref
pressure
effect
Total ⫾0.14 % / 0.98 MPa{10 kgf/cm2} For x ≥ Pref
shift *2 *3 Pref
⫾(0.04 + 0.1 ⫻ x ) % / 0.98 MPa{10 kgf/cm2} For x < Pref

T06E.EPS

*1: ‘x’ is the highest value among the absolute values of lower range value (LRV) and higher range value (HRV), and the
span value in calibration range.
Percent of span = Percent of x ⫻
x
span
*2: Values for measurement span 4.9 kPa {500 mmH2O} or below are estimated values.
*3: Combined zero and span shift

0.4

Accuracy
% 0.2

0
Value of 'x'
M capsule 0 3.3 {330} 10 {1000} 100 kPa{10000 mmH2O}
H capsule 0 33 {0.33} 100 {1} 500 kPa{5 kgf/cm2 }
(Pref)
F03E.EPS

Figure 3. Accuracy

Table 2. Value of ‘Pref’

Capsule Pref
10 kPa
M {1000 mmH2O}
100 kPa
H
{1 kgf/cm2}
T07E.EPS

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 01C22C01-00E Mar. 23, 2006-00
<<Contents>> <<Index>> 4
Table 1-2 . Accuracy, Ambient Temperature Effect and Static Pressure Effect (As Percent of 'x') *1
[For model EJA210 2-inch (50 mm) flange and model EJA220 3-inch (80 mm) flange]

Capsule M, H
⫾0.2 % For x ≥ Pref
Accuracy Pref
⫾(0.1 + 0.1 ⫻ x )% For x < Pref

Zero 2Pref
shift ⫾(0.3 + 0.45 ⫻ x ) % / 50 ⬚C
Ambient
temperature
effect *2
Total ⫾1.2 % / 50 ºC For x ≥ 2Pref
shift *4 2Pref
⫾(0.5 + 0.5 ⫻ x ) % / 50 ⬚C For x < 2Pref

Zero ⫾0.1 % / 0.98 MPa{10 kgf/cm2} For x ≥ 2Pref


shift Pref
Static ⫾(0.1 ⫻ x ) % / 0.98 MPa{10 kgf/cm2} For x < 2Pref
pressure
effect
Total ⫾0.14 % / 0.98 MPa{10 kgf/cm2} For x ≥ Pref
shift *3 *4 Pref
⫾(0.04 + 0.1 ⫻ x ) % / 0.98 MPa{10 kgf/cm2} For x < Pref

T08E.EPS

*1: ‘x’ is the highest value among the absolute values of lower range value (LRV) and higher range value (HRV), and the
span value in calibration range.
Percent of span = Percent of x ⫻
x
span
*2: Below -10 ˚C (14 ˚F), value is twice that given in the table.
*3: Values for measurement span 4.9 kPa {500 mmH2O} or below are estimated values.
*4: Combined zero and span shift

Table 1-3. Accuracy, Ambient Temperature Effect and Static Pressure Effect (As Percent of 'x') *1
[For wetted parts material code H(Hastelloy C) and T(Tantalum) : Model EJA210]
Capsule M, H
⫾0.2 % For x ≥ Pref
Accuracy Pref
⫾(0.1 + 0.1 ⫻ x )% For x < Pref

Zero 2Pref
shift ⫾(0.5 + 0.5 ⫻ x ) % / 50 ⬚C
Ambient
temperature
effect
Total ⫾1.3 % / 50 ºC For x ≥ Pref
shift *3 2Pref
⫾(0.8 + 0.5 ⫻ x ) % / 50 ⬚C For x < Pref

Zero ⫾0.3 % / 0.98 MPa{10 kgf/cm2} For x ≥ Pref


shift Pref
Static ⫾(0.3 ⫻ x ) % / 0.98 MPa{10 kgf/cm2} For x < Pref
pressure
effect
Total ⫾0.4 % / 0.98 MPa{10 kgf/cm2} For x ≥ Pref
shift *2 *3 Pref
⫾(0.1 + 0.3 ⫻ x ) % / 0.98 MPa{10 kgf/cm2} For x < Pref

T09E.EPS

*1: ‘x’ is the highest value among the absolute values of lower range value (LRV) and higher range value (HRV), and the span
value in calibration range.

Percent of span = Percent of x ⫻


x
span
*2: Values for measurement span 4.9 kPa {500 mmH2O} or below are estimated values.
*3: Combined zero and span shift

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 01C22C01-00E Mar. 23, 2006-00
<<Contents>> <<Index>> 5
■ MODEL AND SUFFIX CODES
● Model EJA210 [ Process Flange size : 3-inch (80 mm)]

Model Suffix Codes Description


EJA210 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Flange-mounted differential pressure transmitter
(Flush diaphragm type)
Output Signal -D . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 to 20 mA DC with digital communication (BRAIN protocol)
-E . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 to 20 mA DC with digital communication (HART protocol) *1
-F . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Digital communication (FOUNDATION Fieldbus protocol) *3
Measurement span M . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.5 to 100 kPa {250 to 10000 mmH2O}
(capsule) H . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 to 500 kPa {0.25 to 5 kgf/cm2}
High pressure side [Diaphragm] [Others]
(Process flange side) S# . . . . . . . . . . . . . . . . . . . . . . . . . . . . JIS SUS316L JIS SUS316 *7
wetted parts material *2 *5 H# . . . . . . . . . . . . . . . . . . . . . . . . . . . Hastelloy C-276 *6 Hastelloy C-276 *6
T . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Tantalum Tantalum
Process flange rating J1 . . . . . . . . . . . . . . . . . . . . . . . . . . JIS10K
J2 . . . . . . . . . . . . . . . . . . . . . . . . . . JIS20K
A1 . . . . . . . . . . . . . . . . . . . . . . . . . . ANSI Class 150 P1 . . . . . . . . JPI Class 150
A2 . . . . . . . . . . . . . . . . . . . . . . . . . . ANSI Class 300 P2 . . . . . . . . JPI Class 300
D2 . . . . . . . . . . . . . . . . . . . . . . . . . . DIN PN10/16
D4 . . . . . . . . . . . . . . . . . . . . . . . . . . DIN PN25/40
Process flange 夽 D . . . . . . . . . . . . . . . . . . . . . . . . . 3-inch (80 mm) / JIS S25C
size / material E . . . . . . . . . . . . . . . . . . . . . . . . . 3-inch (80 mm) / JIS SUS304 *8
F . . . . . . . . . . . . . . . . . . . . . . . . . 3-inch (80 mm) / JIS SUS316 *9
Low pressure side process 0...................... without process connector (Rc1/4 female on the cover flange)
connection 1...................... with Rc1/4 female process connector
2...................... with Rc1/2 female process connector
3...................... with 1/4 NPT female process connector
4...................... with 1/2 NPT female process connector
夽 5...................... without process connector (1/4 NPT female on the cover flange)
Cover flange bolts material 夽 A . . . . . . . . . . . . . . . . . . . . . JIS SCM435
B . . . . . . . . . . . . . . . . . . . . . JIS SUS630
Installation -9 . . . . . . . . . . . . . . . . . . Horizontal impulse piping type, left side high pressure
Electrical connection 0 . . . . . . . . . . . . . . . . . G1/2 female, one electrical connection
夽 2 . . . . . . . . . . . . . . . . . 1/2 NPT female, two electrical connections without blind plug
3 . . . . . . . . . . . . . . . . . Pg 13.5 female, two electrical connections without blind plug
4 . . . . . . . . . . . . . . . . . M20 female, two electrical connections without blind plug
5 . . . . . . . . . . . . . . . . . G1/2 female, two electrical connections and a blind plug
7 . . . . . . . . . . . . . . . . . 1/2 NPT female, two electrical connections and a blind plug
8. . . . . . . . . . . . . . . . . Pg 13.5 female, two electrical connections and a blind plug
9 . . . . . . . . . . . . . . . . . M20 female, two electrical connections and a blind plug
A . . . . . . . . . . . . . . . . . G1/2 female, two electrical connections and a SUS316 blind plug
C . . . . . . . . . . . . . . . . . 1/2 NPT female, two electrical connections and a SUS316 blind plug
Integral indicator D . . . . . . . . . . . . . . . Digital indicator
E . . . . . . . . . . . . . . . Digital indicator with the range setting switch *4
夽 N . . . . . . . . . . . . . . . (None)
N . . . . . . . . . . . . . Always N
Optional codes /䊐 Optional specification
T10E.EPS
The ‘夽’ marks indicate the most typical selection for each specification. Example: EJA210-DMSA1D5A-92NN/䊐
The ‘#’marks indicate the construction materials conform to NACE material recommendations per MR01-75. For the use of
SUS316 material, there may be certain limitations for pressure and temperature. Please refer to NACE standards for details.
*1: Refer to GS 01C22T01-00E for HART Protocol version.
*2: Low pressure side wetted parts material: Cover flange; JIS SCS14A, Process connector; SCS14A,
Capsule; JIS SUS316L or ASTM grade 316L (Diaphragm; Hastelloy C-276 or ASTM N10276), Vent plug; JIS SUS316 or ASTM grade 316.
*3: Refer to GS 01C22T02-00E for Fieldbus communication.
*4: Not applicable for Output signal code F.
*5: Users must consider the characteristics of selected wetted parts material and the influence of process fluids. The use of inappropriate materials
can result in the leakage of corrosive process fluids and cause injury to personnel and/or damage to plant facilities. It is also possible that the
diaphragm itself can be damaged and that material from the broken diaphragm and the fill fluid can contaminate the user's process fluids.
Be very careful with highly corrosive process fluids such as hydrochloric acid, sulfuric acid, hydrogen sulfide, sodium hypochlorite, and high-
temperature steam (150°C [302°F] or above). Contact Yokogawa for detailed information of the wetted parts material.
*6: Hastelloy C-276 or ASTM N10276.
*7: JIS SUS316 or ASTM grade 316.
*8: JIS SUS304 or ASTM grade 304. Forged version may be used.
*9: JIS SUS316 or ASTM grade 316. Forged version may be used.

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 01C22C01-00E Jul. 08, 2009-00
<<Contents>> <<Index>> 6
■ MODEL AND SUFFIX CODES
● Model EJA210 [ Process Flange size : 2-inch (50 mm)]

Model Suffix Codes Description


EJA210 ................................... Flange-mounted differential pressure transmitter
(Flush diaphragm type)
Output Signal -D . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 to 20 mA DC with digital communication (BRAIN protocol)
-E . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 to 20 mA DC with digital communication (HART protocol) *1
-F . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Digital communication (FOUNDATION Fieldbus protocol) *3
Measurement span M . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.5 to 100 kPa {250 to 10000 mmH2O}
(capsule) H . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 to 500 kPa {0.25 to 5 kgf/cm2}
High pressure side [Diaphragm] [Others]
(Process flange side) W# . . . . . . . . . . . . . . . . . . . . . . . . . . . . HastelloyC-276 *6 JIS SUS316 *7
wetted parts material *2 *5
Process flange rating J1 . . . . . . . . . . . . . . . . . . . . . . . . . . JIS10K
J2 . . . . . . . . . . . . . . . . . . . . . . . . . . JIS20K
A1 . . . . . . . . . . . . . . . . . . . . . . . . . . ANSI Class 150 P1 . . . . . . . . . . JPI Class 150
A2 . . . . . . . . . . . . . . . . . . . . . . . . . . ANSI Class 300 P2 . . . . . . . . . . JPI Class 300
D2 . . . . . . . . . . . . . . . . . . . . . . . . . . DIN PN10/16
D4 . . . . . . . . . . . . . . . . . . . . . . . . . . DIN PN25/40
Process flange 夽 A......................... 2-inch (50 mm) / JIS S25C
size / material B......................... 2-inch (50 mm) / JIS SUS304 *8
C......................... 2-inch (50 mm) / JIS SUS316 *9
Low pressure side process 0....................... without process connector (Rc1/4 female on the cover flange)
connection 1....................... with Rc1/4 female process connector
2....................... with Rc1/2 female process connector
3....................... with 1/4 NPT female process connector
4....................... with 1/2 NPT female process connector
夽 5....................... without process connector (1/4 NPT female on the cover flange)
Cover flange bolts material 夽 A.................... JIS SCM435
B.................... JIS SUS630
Installation -9 . . . . . . . . . . . . . . . . . Horizontal impulse piping type, left side high pressure
Electrical connection 0................. G1/2 female, one electrical connection
夽 2................. 1/2 NPT female, two electrical connections without blind plug
3................. Pg 13.5 female, two electrical connections without blind plug
4................. M20 female, two electrical connections without blind plug
5................. G1/2 female, two electrical connections and a blind plug
7................. 1/2 NPT female, two electrical connections and a blind plug
8................. Pg 13.5 female, two electrical connections and a blind plug
9................. M20 female, two electrical connections and a blind plug
A................. G1/2 female, two electrical connections and a SUS316 blind plug
C................. 1/2 NPT female, two electrical connections and a SUS316 blind plug
Integral indicator D............... Digital indicator
E............... Digital indicator with the range setting switch *4
夽 N............... (None)
N............. Always N
Optional codes /䊐 Optional specification
T11E.EPS
The ‘夽’ marks indicate the most typical selection for each specification. Example: EJA210-DMWA1A5A-92NN/䊐
The ‘#’marks indicate the construction materials conform to NACE material recommendations per MR01-75. For the use of
SUS316 material, there may be certain limitations for pressure and temperature. Please refer to NACE standards for details.
*1: Refer to GS 01C22T01-00E for HART Protocol version.
*2: Low pressure side wetted parts material: Cover flange; JIS SCS14A, Process connector; SCS14A,
Capsule; JIS SUS316L or ASTM grade 316L (Diaphragm; Hastelloy C-276 or ASTM N10276), Vent plug; JIS SUS316 or ASTM grade 316.
*3: Refer to GS 01C22T02-00E for Fieldbus communication.
*4: Not applicable for Output signal code F.
*5: Users must consider the characteristics of selected wetted parts material and the influence of process fluids. The use of inappropriate materials
can result in the leakage of corrosive process fluids and cause injury to personnel and/or damage to plant facilities. It is also possible that the
diaphragm itself can be damaged and that material from the broken diaphragm and the fill fluid can contaminate the user's process fluids.
Be very careful with highly corrosive process fluids such as hydrochloric acid, sulfuric acid, hydrogen sulfide, sodium hypochlorite, and high-
temperature steam (150°C [302°F] or above). Contact Yokogawa for detailed information of the wetted parts material.
*6: Hastelloy C-276 or ASTM N10276.
*7: JIS SUS316 or ASTM grade 316.
*8: JIS SUS304 or ASTM grade 304. Forged version may be used.
*9: JIS SUS316 or ASTM grade 316. Forged version may be used.

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 01C22C01-00E Jul. 08, 2009-00
<<Contents>> <<Index>> 7
● Model EJA220

Model Suffix Codes Description


EJA220 .................................. Flange-mounted differential pressure transmitter
(Extended diaphragm type)
Output Signal -D . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 to 20 mA DC with digital communication (BRAIN protocol)
-E . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 to 20 mA DC with digital communication (HART protocol) *1
-F . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Digital communication (FOUNDATION Fieldbus protocol) *5
Measurement span M............................. 2.5 to 100 kPa {250 to 10000 mmH2O}
(capsule) H............................. 25 to 500 kPa {0.25 to 5 kgf/cm2}
High pressure side [Diaphragm] [Pipe] [Others]
(Process flange side) S *3 # . . . . . . . . . . . . . . . . . . . . . . . . . JIS SUS316L JIS SUS316 JIS SUS316 *9
wetted parts material *2 *7 W *4 # . . . . . . . . . . . . . . . . . . . . . . . . Hastelloy C-276 *8 JIS SUS316 JIS SUS316 *9
Process flange rating J1 . . . . . . . . . . . . . . . . . . . . . . . . . . JIS10K
J2 . . . . . . . . . . . . . . . . . . . . . . . . . . JIS20K
A1 . . . . . . . . . . . . . . . . . . . . . . . . . . ANSI Class 150 P1 . . . . . . . . JPI Class 150
A2 . . . . . . . . . . . . . . . . . . . . . . . . . . ANSI Class 300 P2 . . . . . . . . JPI Class 300
D2 . . . . . . . . . . . . . . . . . . . . . . . . . . DIN PN10/16
D4 . . . . . . . . . . . . . . . . . . . . . . . . . . DIN PN25/40
Diaphragm extension 夽 2......................... X2=50 mm
length (X2) 4......................... X2=100 mm
6......................... X2=150 mm
Process flange 夽 G...................... 4-inch (100 mm) / JIS S25C
size / material H...................... 4-inch (100 mm) / JIS SUS304 *10
J...................... 4-inch (100 mm) / JIS SUS316 *11
D...................... 3-inch (80 mm) / JIS S25C
E...................... 3-inch (80 mm) / JIS SUS304 *10
F...................... 3-inch (80 mm) / JIS SUS316 *11
Low pressure side process 0.................... without process connector (Rc1/4 female on the cover flanges)
connection 1.................... with Rc1/4 female process connector
2.................... with Rc1/2 female process connector
3.................... with 1/4NPT female process connector
4.................... with 1/2NPT female process connector
夽 5.................... without process connector (1/4 NPT female on the cover flanges)
Cover flange bolts 夽 A................. JIS SCM435
material B................. JIS SUS630
Installation -9 . . . . . . . . . . . . . . . Horizontal impulse piping type, left side high pressure
Electrical connection 0.............. G1/2 female, one electrical connection
夽 2.............. 1/2 NPT female, two electrical connections without blind plug
3.............. Pg 13.5 female, two electrical connections without blind plug
4.............. M20 female, two electrical connections without blind plug
5.............. G1/2 female, two electrical connections and a blind plug
7.............. 1/2 NPT female, two electrical connections and a blind plug
8.............. Pg 13.5 female, two electrical connections and a blind plug
9.............. M20 female, two electrical connections and a blind plug
A.............. G1/2 female, two electrical connections and a SUS316 blind plug
C.............. 1/2 NPT female, two electrical connections and a SUS316 blind plug
Integral indicator D . . . . . . . . . . . . . Digital indicator
E . . . . . . . . . . . . . Digital indicator with the range setting switch *6
夽 N . . . . . . . . . . . . . (None)
N........... Always N
Optional codes /䊐 Optional specification
T12E.EPS
The ‘夽’ marks indicates the most typical selection for each specification. Example:EJA220-DMSA12G5A-92NN/䊐
The ‘#’marks indicate the construction materials conform to NACE material recommendations per MR01-75. For the use of
SUS316 material, there may be certain limitations for pressure and temperature. Please refer to NACE standards for details.
*1: Refer to GS 01C22T01-00E for HART Protocol version.
*2: Low pressure side wetted parts material: Cover flange; JIS SCS14A, Process connector; SCS14A,
Capsule; JIS SUS316L or ASTM grade 316L (Diaphragm, Hastelloy C-276 or ASTM N10276), Vent plug; JIS SUS316 or ASTM grade 316
*3: Applicable for 4-inch (100 mm) flange size (Process flange size/material G and H).
*4: Applicable for 3-inch (80 mm) flange size (Process flange size/material D and E).
*5: Refer to GS 01C22T02-00E for Fieldbus communication.
*6: Not applicable for Output signal code F.
*7: Users must consider the characteristics of selected wetted parts material and the influence of process fluids. The use of inappropriate materials
can result in the leakage of corrosive process fluids and cause injury to personnel and/or damage to plant facilities. It is also possible that the
diaphragm itself can be damaged and that material from the broken diaphragm and the fill fluid can contaminate the user's process fluids.
Be very careful with highly corrosive process fluids such as hydrochloric acid, sulfuric acid, hydrogen sulfide, sodium hypochlorite, and high-
temperature steam (150°C [302°F] or above). Contact Yokogawa for detailed information of the wetted parts material.
*8: Hastelloy C-276 or ASTM N10276.
*9: JIS SUS316 or ASTM grade 316.
*10: JIS SUS304 or ASTM grade 304. Forged version may be used.
*11: JIS SUS316 or ASTM grade 316. Forged version may be used.

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 01C22C01-00E Jul. 08, 2009-00
<<Contents>> <<Index>> 8
■ OPTIONAL SPECIFICATIONS (For Explosion Protected types “䉫”)
For FOUNDATION Fieldbus explosion protected type, see GS 01C22T02-00E.
Item Description Code
FM Explosionproof Approval *1 *3
Applicable standard: FM3600, FM3615, FM3810, ANSI/NEMA250
Explosionproof for Class I, Division 1, Groups B, C and D
Dust-ignitionproof for Class II/III, Division 1, Groups E, F and G
FF1
Hazardous (classified) locations, indoors and outdoors (NEMA 4X)
Division 2, 'SEALS NOT REQUIRED',
Temp. Class: T6
Amb. Temp.: –40 to 60⬚C (–40 to 140⬚F)
FM Intrinsically safe Approval *1 *3
Applicable standard: FM3600, FM3610, FM3611, FM3810, ANSI/NEMA250
Factory Mutual (FM) Intrinsically Safe for Class I, Division 1, Groups A, B, C & D, Class II, Division 1,
Groups E, F & G and Class III, Division 1 Hazardous Locations.
Nonincendive for Class I, Division 2, Groups A, B, C & D, Class II, Division. 2,
Groups E, F & G, and Class III, Division 1 Hazardous Locations.
FS1
Enclosure: “NEMA 4X”, Temp. Class: T4, Amb. Temp.: –40 to 60⬚C (–40 to 140⬚F)
Intrinsically Safe Apparatus Parameters
[Groups A, B, C, D, E, F and G]
Vmax=30 V, Imax=165 mA, Pmax=0.9 W, Ci=22.5 nF, Li=730 ␮H
[Groups C, D, E, F and G]
Vmax=30 V, Imax=225 mA, Pmax=0.9 W, Ci=22.5 nF, Li=730 ␮H
Combined FF1 and FS1 *1 *3 FU1
ATEX Flameproof Approval *2
Applicable standard: EN 60079-0, EN 60079-1
Certificate: KEMA 02ATEX2148
II 2G Ex d IIC T4, T5, T6 KF21
Amb. Temp.: T5; –40 to 80⬚C ( –40 to 176⬚F), T4 and T6; –40 to 75⬚C ( –40 to 167⬚F)
Max. process Temp.: T4; 120⬚C (248⬚F), T5; 100⬚C (212⬚F), T6; 85⬚C (185⬚F)
ATEX Intrinsically safe Approval *2 *3
Applicable standard: EN50014, EN50020, EN500284, EN50281-1-1
Certificate: KEMA 02ATEX1030X KS2
II 1G EEx ia IIC T4, Amb. Temp.: –40 to 60⬚C (–40 to 140⬚F)
Ui=30 V, Ii=165 mA, Pi=0.9 W, Ci=22.5 nF, Li=730 ␮H
Combined KF21, KS2 and ATEX Type n *2 *3
ATEX Type n
Applicable standard: EN 60079-15, EN 60079-0
II 3G Ex nL IIC T4 Gc, Amb. Temp.: –30 to 60⬚C (–22 to 140⬚F)
Ui=30 V DC, Ci=22.5 nF, Li=730 ␮H
Dust
[For combined with II 2G]
Applicable standard: EN 61241-0, EN 61241-1 KU22
II 2D Ex tD A21 IP6X Max. surface Temp. for dust-proof :
T85⬚C (Tamb: –40 to 75⬚C, Tp:85⬚C), T100⬚C (Tamb: –40 to 80⬚C, Tp:100⬚C),
T120⬚C (Tamb: –40 to 75⬚C, Tp:120⬚C )
[For combined with II 1G]
II 1D maximum surface temperature T65⬚C (149⬚F) {Tamb.: 40⬚C (104⬚F)},
T85⬚C (185⬚F) {Tamb.: 60⬚C (140⬚F)}, T105⬚C (221⬚F) {Tamb.: 80⬚C (176⬚F)}

T14E.EPS
*1: Applicable for Electrical connection code 2 , 7 and C (1/2 NPT female).
*2: Applicable for Electrical connection code 2, 4, 7, 9 and C (1/2 NPT and M20 female).
*3: Applicable for Output signal code D and E. For intrinsically safe approval, use the safety barrier certified by the testing laboratories
(BARD-400 is not applicable).

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 01C22C01-00E April 23, 2012-00
<<Contents>> <<Index>> 9
Item Description Code
CSA Explosionproof Approval *1 *3
Applicable standard: C22.2 No. 0, No. 0.4, No. 25, No. 30, No. 94, No. 142
Certificate: 1089598
Explosionproof for Class I, Division 1, Groups B, C and D
Dustignitionproof for Class II/III, Division 1, Groups E, F and G
Division2 ‘SEALS NOT REQUIRED’ , Temp. Class: T4, T5, T6 Encl Type 4x CF1
Max. Process Temp.: T4; 120⬚C (248⬚F), T5; 100⬚C (212⬚F), T6; 85⬚C (185⬚F)
Amb. Temp.: –40 to 80⬚C (–40 to 176⬚F)
Process Sealing Certification
Dual Seal Certified by CSA to the requirement of ANSI/ISA 12.27.01
Canadian Standards No additional sealing required. Primary seal failure annunciation: at the zero adjustment screw
Association (CSA)
CSA Intrinsically safe Approval *1 *3
Applicable standard: C22.2 No. 0, No. 0.4, No. 25, No. 30, No. 94, No. 142, No. 157, No. 213
Certificate: 1053843
Class I, Groups A, B, C and D Class II and III, Groups E, F and G
Encl Type 4x, Temp. Class: T4, Amb. Temp.: –40 to 60⬚C (–40 to 140⬚F) CS1
Vmax=30 V, Imax=165 mA, Pmax=0.9 W, Ci=22.5 nF, Li=730 ␮H
Process Sealing Certification
Dual Seal Certified by CSA to the requirement of ANSI/ISA 12.27.01
No additional sealing required. Primary seal failure annunciation: at the zero adjustment screw
Combined CF1 and CS1 *1 *3 CU1
IECEx Intrinsically safe, type n and Flameproof Approval *3 *9
Intrinsically safe and type n
Applicable Standard: IEC 60079-0:2004, IEC 60079-11:1999, IEC 60079-15:2005,
IEC 60079-26:2005
Certificate: IECEx KEM 06.0007X
Ex ia IIC T4, Ex nL IIC T4 Enclosure: IP67
Amb. Temp.: –40 to 60⬚C (–40 to 140⬚F), Max. Process Temp.: 120⬚C (248⬚F)
IECEx Electrical Parameters: [Ex ia] Ui=30 V, Ii=165 mA, Pi=0.9 W, Ci=22.5 nF, Li=730 ␮H
SU2
[Ex nL] Ui=30 V, Ci=22.5 nF, Li=730 ␮H
Flameproof
Applicable Standard: IEC 60079-0:2004, IEC60079-1:2003
Certificate: IECEx KEM 06.0005
Ex d IIC T6...T4 Enclosure: IP67
Max.Process Temp.: T4;120⬚C (248⬚F), T5;100⬚C (212⬚F), T6; 85⬚C (185⬚F)
Amb.Temp.: –40 to 75⬚C (–40 to 167⬚F) for T4, –40 to 80⬚C (–40 to 176⬚F) for T5,
–40 to 75⬚C (–40 to 167⬚F) for T6
TIIS Flameproof Approval, Ex do IIC T4X *3 *4 *5 *7 *8
Certificate: TC15296 (Without integral indicator)
JF3
TC15297 (With integral indicator)
TIIS certification Amb. Temp.: –20 to 60⬚C, Process Temp.: –20 to 120⬚C
TIIS Intrinsically safe Approval, Ex ia IIC T4 *6 *8
Certificate: TC14632 JS3
Amb. Temp.: –20 to 60⬚C, Process Temp.: –20 to 120⬚C
Attached flameproof Electrical connection: G1/2 female 1 pc. G11
packing adapter *5 Applicable cable: O. D. 8 to 12 mm 2 pcs. G12
T17E.EPS
*1: Applicable for Electrical connection code 2, 7 and C (1/2 NPT female).
*2: Applicable for Electrical connection code 2, 4, 7, 9 and C (1/2 NPT and M20 female).
*3: Applicable for Output signal code D and E. For intrinsically safe approval, use the safety barrier certified by the testing laboratories
(BARD-400 is not applicable).
*4: Not applicable for Electrical connection code A and C.
*5: If cable wiring is to be used to a TIIS flameproof type transmitter, add the YOKOGAWA-assured flameproof packing adapter.
*6: Applicable for Output signal code D. See <Safety Barrier for TIIS Intrinsically Safe Type>.
*7: In case that the ambient temperature exceeds 50°C or that the ambient temperature exceeds 45°C with the process temperature of
90°C or above, use heat-resistant cables with maximum allowable temperature of 75°C or above.
*8: TIIS (The Technology Institution of Industrial Safety) Certification is a new notation for the explosionproof approval in Japan
instead of JIS.
*9: Applicable for Electrical connection code 2, 4, 7 and C (1/2 NPT and M20 female).

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 01C22C01-00E April 23, 2012-00
<<Contents>> <<Index>> 10
■ OPTIONAL SPECIFICATIONS
Item Description Code
Amplifier cover only P
Color change
Painting Amplifier cover and terminal cover, Munsell 7.5 R4/14 PR
Coating change Epoxy resin-baked coating *14 X1
Exterior parts on the amplifier housing (name plates, tag plate, zero-adjustment screw, stopper
316 SST exterior parts HC
screw) will become 316 SST *5
Transmitter power supply voltage : 10.5 to 32 V DC (10.5 to 28 V DC for TIIS intrinsically safe
Lightning protector type, 10.5 to 30 V DC for intrinsically safe type other than TIIS, or 9 to 32 V DC for Fieldbus A
communication type.)
Allowable current : Max. 6000 A (1✕40 ␮s), Repeating 1000 A (1✕40 ␮s) 100 times
Degrease cleansing treatment K1
Oil-prohibited use Degrease cleansing treatment with fluorinated oil filled capsule.
Operating temperature -20 to 80°C K2

Oil-prohibited use Degrease cleansing and dehydrating treatment K5


with dehydrating treatment Degrease cleansing and dehydrating treatment with fluorinated oil filled capsule.
Operating temperature -20 to 80°C K6

No serration *1 No serration work on the flange gasket surface ( for ANSI flange only ) Q
Diaphragm protection from sticky process fluid by FEP Teflon film attached with fluorinated oil.
Teflon film *2 Operation range: 20 to 120°C, 0 to 2 MPa(Not applicable for vacuum service). T
Update time: 0.125 sec
Fast response *11 Amplifier assembly damping time constant: 0.1 to 64 sec in 9 increments. F1
Response time (with min. damping time constant): max. 0.5 sec (excluding level unit)
Failure alarm down-scale *3 Output status at CPU failure and hardware error is –5%, 3.2 mA or less. C1
Failure alarm down-scale: output status at CPU failure and
C2
Output signal limits: hardware error is –5%, 3.2 mA or less.
NAMUR NE43 compliant *3 *12 3.8 mA to 20.5 mA
Failure alarm up-scale: output status at CPU failure and
C3
hardware error is 110%, 21.6 mA or more.
Data configuration at factory Description into “Descriptor” parameter of HART protocol CA
Amplifier housing material : JIS SCS14A stainless steel
Stainless steel amplifier housing *4 E1
(equivalent to JIS SUS316 cast stainless steel or ASTM CF-8M)
High pressure side : Process flange, Block
*6 For M03
Low pressure side : Cover flange
model
High pressure side : Process flange, Block EJA210
*7 M13
Low pressure side : Cover flange, Process connector
Mill Certificate
High pressure side : Process flange, Block, Pipe, Base For
*6 M04
Low pressure side : Cover flange model
High pressure side : Process flange, Block, Pipe, Base EJA220
Low pressure side : Cover flange, Process connector
*7 M14

( Flange rating ) ( Test Pressure )


JIS 10K 2 MPa {20 kgf/cm2 } T31
Pressure test/
JIS 20K 5 MPa {50 kgf/cm2 } T32
Leak test
3 MPa {29.8 kgf/cm2 } Nitrogen ( N2 ) Gas *10 T36
Certificate *13 ANSI/JPI Class 150
Retention time: 10 minutes
ANSI/JPI Class 300 7.7 MPa {77 kgf/cm2 } *8 T37
ANSI/JPI Class 300 7 MPa {70 kgf/cm2 } *9 T38
T13E.EPS
*1: This item cannot be applied to model EJA210 Wetted parts material code H and T, whose gasket contact surface are not serrated as
standard specifications.
*2: Teflon film can only be specified for model EJA210.
*3: Applicable for Output signal code D and E. The hardware error indicates faulty amplifier or capsule.
When combining with Option code F1, output status for down-scale is –2.5%, 3.6 mA DC or less.
*4: Applicable for Electrical connection code 2, 3, 4, A and C. Not applicable for Option code P䊐, X1, and JF3.
*5: 316 or 316L SST. The specification is included in option code /E1. Not applicable with option code /JF3.
*6: Applicable for Low Pressure Side Process connection code 0 and 5.
*7: Applicable for Low Pressure Side Process connection code 1, 2, 3, and 4.
*8: Applicable for model EJA210.
*9: Applicable for model EJA220.
*10: Pure nitrogen gas is used for oil-prohibited use (Option code K1, K2, K5, and K6 ).
*11: Applicable for Output signal code D and E. Write protection switch is attached for output code E. Not applicable for Integral
indicator code E.
*12: Not applicable for Option code C1.
*13: The unit on the certificate is always MPa regardless of selection of option code D1, D3, or D4.
*14: Not applicable for color change option.

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 01C22C01-00E July 01, 2011-00
<<Contents>> <<Index>> 11

Item Description Code


Gold-plate *2 Inside of isolating diaphragms(fill fluid side) are gold plated, effective for hydrogen permeation. A1
Wired tag plate Stainless steel tag plate wired onto transmitter N4
P calibration ( psi unit ) D1
Calibration units *1 ( See Table 3. ) D3
bar calibration ( bar unit )
M calibration ( kgf/cm2 unit ) D4
Sealing treatment to JIS Sealant(liquid silicone rubber) is coated on JIS SUS630 cover flange mounting nuts against
SUS630 nuts stress corrosion cracking. Y

T16E.EPS
*1: The unit of MWP (Max. working pressure) on name plate of a housing is the same unit as specified by Optional code D1, D3, and
D4.
*2: Applicable for Wetted parts material code S, H and W .
Issue Tokuchu in case gold-plated disphragm is required for low pressure side.

Table 3. Calibration Units


Measurement Optional Code
Span and
Range D1 ( psi Unit ) D3 ( bar Unit ) D4 ( kgf/cm2 Unit )
Span 10 to 400 inH2O 25 to 1000 mbar 250 to 10000 mmH2O
EJA210/220

M
Range -400 to 400 inH2O -1000 to 1000 mbar -10000 to 10000 mmH2O
Span 100 to 2000 inH2O 250 to 5000 mbar 0.25 to 5 kgf/cm2
H
Range -2000 to 2000 inH2O -5000 to 5000 mbar -5 to 5 kgf/cm2
T15E.EPS

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 01C22C01-00E Oct. 01, 2008-00
<<Contents>> <<Index>> 12
■ DIMENSIONS
● Model EJA210
Unit : mm (approx.inch)
174 (6.85) 30 (1.18)
110 (4.33)
External indicator
conduit connection 133 (5.24) 72 Conduit Zero 35
connection adjustment (1.38)
(Optional) (2.83)
25 Internal Terminal side
(0.98) indicator ø78
t (Optional) (3.07)
162 146
(6.38) (5.75)
Shrouding bolt *3 Ground
terminal
*1
øD øC øg ød
Process
connection Vent plug

46 Drain plug
Process
(1.81) connector n- øh
(Optional) 54 37 *2
Process flange (2.13) (1.46)

*1: Indicates inside diameter of gasket contact surface.


*2: When Optional code K1, K2, K5, or K6 is selected, add 15 mm(0.59 inch) to the value in the figure.
*3: Applicable only for ATEX, IECEx, and TIIS Flameproof type.

Flange size: 3-inch (80 mm)


Flange Nominal Bolt Holes
Diameter and øD øC øg ød t
Rating No.(N) Dia.(øh)
JIS 10K 185(7.28) 150(5.91) 130(5.12) 90(3.54) 18(0.71) 8 19(0.75)

JIS 20K 200(7.87) 160(6.30) 130(5.12) 90(3.54) 22(0.87) 8 23(0.91)

ANSI Class150 190.5(7.50) 152.4(6.00) 130(5.12) 90(3.54) 23.9(0.94) 4 19.1(0.75)

ANSI Class300 209.6(8.25) 168.1(6.62) 130(5.12) 90(3.54) 28.5(1.12) 8 22.4(0.88)

JPI Class150 190(7.48) 152.4(6.00) 130(5.12) 90(3.54) 24(0.94) 4 19(0.75)

JPI Class300 210(8.27) 168.1(6.62) 130(5.12) 90(3.54) 28.5(1.12) 8 22(0.87)

DIN PN 10/16 200(7.87) 160(6.30) 130(5.12) 90(3.54) 20(0.79) 8 18(0.71)

DIN PN 25/40 200(7.87) 160(6.30) 130(5.12) 90(3.54) 24(0.94) 8 18(0.71)

Flange size: 2-inch (50 mm)


Flange Nominal Bolt Holes
Diameter and øD øC øg ød t
Rating No.(N) Dia.(øh)
JIS 10K 155(6.10) 120(4.72) 100(3.94) 61(2.40) 16(0.63) 4 19(0.75)

JIS 20K 155(6.10) 120(4.72) 100(3.94) 61(2.40) 18(0.71) 8 19(0.75)

ANSI Class150 152.4(6.00) 120.7(4.75) 100(3.94) 61(2.40) 19.1(0.75) 4 19.1(0.75)

ANSI Class300 165.1(6.50) 127(5.00) 100(3.94) 61(2.40) 22.4(0.88) 8 19.1(0.75)

JPI Class150 152(5.98) 120.6(4.75) 100(3.94) 61(2.40) 19.5(0.71) 4 19(0.75)

JPI Class300 165.1(6.50) 127(5.00) 100(3.94) 61(2.40) 22.5(0.89) 8 19(0.75)

DIN PN 10/16 165(6.50) 125(4.92) 100(3.94) 61(2.40) 18(0.71) 4 18(0.71)

DIN PN 25/40 165(6.50) 125(4.92) 100(3.94) 61(2.40) 20(0.79) 4 18(0.71)


F04E.EPS

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 01C22C01-00E Mar. 23, 2006-00
<<Contents>> <<Index>> 13
● Model EJA220
Unit : mm (approx.inch)
174 (6.85) 30 (1.18)
110 (4.33)
External indicator
conduit connection 133 (5.24) 72 Conduit Zero 35
connection adjustment
(Optional) (2.83) (1.38)

Internal Terminal side


indicator
ø78
(Optional) (3.07)
162 146
(6.38) (5.75)
Shrouding bolt *2 Ground
terminal
øD øC øg øA
Process
connection Vent plug
46
Drain plug
(1.81) Process
connector n- øh
(Optional) 54 37 *1
t Process flange
(2.13) (1.46)
25 (0.98)
X2

*1: When Optional code K1, K2, K5, or K6 is selected, add 15 mm(0.59 inch) to the value in the figure.
*2: Applicable only for ATEX, IECEx, and TIIS Flameproof type.

Flange size: 4-inch (100 mm)


Flange Nominal Bolt Holes
Diameter and øD øC øg øA t
Rating No.(N) Dia.(øh)

JIS 10K 210(8.27) 175(6.89) 155(6.10) 96±0.5(3.78) 18(0.71) 8 19(0.75)

JIS 20K 225(8.86) 185(7.28) 155(6.10) 96±0.5(3.78) 24(0.94) 8 23(0.91)

ANSI Class150 228.6(9.00) 190.5(7.50) 155(6.10) 96±0.5(3.78) 23.9(0.94) 8 19.1(0.75)

ANSI Class300 254(10.00) 200(7.87) 155(6.10) 96±0.5(3.78) 31.8(1.25) 8 22.4(0.88)

JPI Class150 229(9.02) 190.5(7.50) 155(6.10) 96±0.5(3.78) 24(0.94) 8 19(0.75)

JPI Class300 254(10.00) 200.2(7.88) 155(6.10) 96±0.5(3.78) 32(1.26) 8 22(0.87)

DIN PN 10/16 220(8.66) 180(7.09) 155(6.10) 96±0.5(3.78) 20(0.79) 8 18(0.71)

DIN PN 25/40 235(9.25) 190(7.50) 155(6.10) 96±0.5(3.78) 24(0.94) 8 22(0.87)

Flange size: 3-inch (80 mm)


Flange Nominal Bolt Holes
Diameter and øD øC øg øA t
Rating No.(N) Dia.(øh)

JIS 10K 185(7.28) 150(5.91) 130(5.12) 71±0.5(2.80) 18(0.71) 8 19(0.75)

JIS 20K 200(7.87) 160(6.30) 130(5.12) 71±0.5(2.80) 22(0.87) 8 23(0.91)

ANSI Class150 190.5(7.50) 152.4(6.00) 130(5.12) 71±0.5(2.80) 23.9(0.94) 4 19.1(0.75)

ANSI Class300 209.6(8.25) 168.1(6.62) 130(5.12) 71±0.5(2.80) 28.5(1.12) 8 22.4(0.88)

JPI Class150 190(7.48) 152.4(6.00) 130(5.12) 71±0.5(2.80) 24(0.94) 4 19(0.75)

JPI Class300 210(8.27) 168.1(6.62) 130(5.12) 71±0.5(2.80) 28.5(1.12) 8 22(0.87)

DIN PN 10/16 200(7.88) 160(6.30) 130(5.12) 71±0.5(2.80) 20(0.79) 8 18(0.71)

DIN PN 25/40 200(7.87) 160(6.30) 130(5.12) 71±0.5(2.80) 24(0.94) 8 18(0.71)

Diaphragm Extension Length Code X2

2 1.97 inch (50 mm)

4 3.94 inch (100 mm)

6 5.91 inch (150 mm)


F05E.EPS

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 01C22C01-00E Mar. 23, 2006-00
<<Contents>> <<Index>> 14
● Terminal Configuration ● Terminal Wiring
+
SUPPLY Power supply and output terminal

CHECK + External indicator(ammeter) terminal*1


Communication Ground terminal


Terminals(BT200 etc.) SUPPLY CHECK CHECK METER
Connection hook*1 *1: When using an external indicator or a check meter,
Connection hook
the internal resistance must be 10 ⍀ or less.
Not available for Fieldbus communication(Output signal
code F).
F05E.EPS

< Ordering Information > “䉫” < Safety Barrier for TIIS Intrinsically Safe Type >
Specify the following when ordering
1. Model, suffix codes, and optional codes Supplier Type Model
2. Calibration range and units: MTL3046B
MTL
1) Calibration range can be specified with range value MTL4041B
Isolator
specifications up to 5 digits KFD2-STC3-Ex 1
(excluding any decimal point) for low or high range limits P+F
KFD2-STV3-Ex 1-1, 2, 3
within the range of -32000 to 32000. T001E.EPS
2) Specify only one unit from the table, ‘Settings when
shipped.’ Note: Requirements of capacitance and inductance for cable
3. Select linear or square root for output mode and display wiring.
mode. Cw ≤ Co – 11[nF]
Note: If not specified, the instrument is shipped in normal Lw ≤ Lo – 730[µH]
operation mode. (Co: Max. external capacitance)
(Lo: Max. external inductance)
4. Display scale and units (for transmitters equipped with
integral indicator only) < Reference >
Specify either 0 to 100 % or engineering unit scale and
‘Range and Unit’ for engineering units scale: 1. JIS SUS316L stainless steel; Equivalent to AISI 316L.
Scale range can be specified with range limit specifications 2. JIS SUS316 stainless steel; Equivalent to AISI 316.
up to 5 digits (excluding any decimal point) for low or 3. JIS SUS304 stainless steel; Equivalent to AISI 304.
high range limits within the range of -19999 to 19999. 4. JIS S25C carbon steel; Equivalent to AISI 1025.
5. Tag Number (if required) 5. JIS SECC Carbon steel.
6. Teflon; Trademark of E.I. DuPont de Nemours & Co.
< Related Instruments > “䉫” 7. JIS SUS630 stainless steel; Equivalent to ASTM 630.
8. Hastelloy; Trademark of Haynes International Inc.
Power Distributor: Refer to GS 01B04T01-02E or
9. JIS SCS14A stainless steel; Equivalent to JIS SUS316
GS 01B04T02-02E
cast stainless steel or ASTM CF-8M.
BRAIN TERMINAL: Refer to GS 01C00A11-00E
10. HART; Trademark of the HART Communication
Foundation.
11. FOUNDATION; Trademark of Fieldbus Foundation.
12. Other company names and product names used in this
material are registered trademarks or trademarks of their
respective owners.

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 01C22C01-00E Mar. 23, 2006-00
Subject to change without notice.

You might also like