System Serial Number: 31046

Biograph 40
Customer ID: 01

General Information

Standard:

Equipment
Tester Name PET CT Customer Name PET CT
Product Name Biograph 40 Customer ID 01
Software Version VG51C Hospital Name UI za Pozitronsko Emisiona
Tomografija
Serial Number 31046
Street Bledski dogovor 10
Handed over 2016-11-29
Zip Code 1000
City Skopje
District District
Country MK
Customer Phone Number +389 2 0000 000

Serial Number
Component Serial Number Revision Level
Gantry 1528 00
DMS A 19760 02
DMS B - -
PHS n.a. n.a.
Tube Assembly A 621341474 00006
Tube Assembly B - -
Tube Collimator A 5429 03
Tube Collimator B - -
Generator Electronic A 8531 03
Generator Electronic B - -
HV Tank A 8539 03
HV Tank B - -

2022-09-05 09:25:43 Page: 1 / 6


System Serial Number: 31046
Biograph 40
Customer ID: 01

Table of Contents

1. Quality Daily Noise OK................................................................................................................. 3

2. Quality Daily Homogeneity OK..................................................................................................... 4

2022-09-05 09:25:43 Page: 2 / 6


System Serial Number: 31046
Biograph 40
Customer ID: 01

1. Quality Daily Noise

Description
The measurement is performed for two sets of scan parameters, representing a typical head-mode and a typical
body mode. The image noise is evaluated measuring the standard deviation of CT numbers in a Region Of Interest
(ROI) in the image of a cylindrical water phantom with diameter 20 cm. The ROI is located in the phantom center.
To suppress the influence of image inhomogeneity and ring artifacts, the evaluation is performed for the difference
image of two succeeding scans.

Common Parameters
Voltage 120 kV
UHR no
Shaped filter Wide
Window width 500 HU
Window center 0 HU
FoV 250 mm

Test: Typical head mode


Test Parameters Test Results OK
System A
Scan Type Rotation
Current 350 mA
Slice width reconstructed 4.8 mm
Number of images 3
Collimation 12x1.2 mm
Scan Time 1.0 s
Rotation time 1.0 s
Body Region Head
Kernel H30s
Radius of ROIs 40 mm

Result
Results for System A
Slice 1 3.55 HU
Slice 2 3.45 HU
Slice 3 3.49 HU
Tolerance 3.22 ... 3.94 HU

2022-09-05 09:25:43 Page: 3 / 6


System Serial Number: 31046
Biograph 40
Customer ID: 01

Test: Typical body mode


Test Parameters Test Results OK
System A
Scan Type Sequence
Current 420 mA
Slice width reconstructed 5.0 mm
Number of images 2
Collimation 20x0.6 mm
Scan Time 0.5 s
Rotation time 0.5 s
Body Region Body
Kernel B30f
Radius of ROIs 40 mm

Result
Results for System A
Slice 1 4.06 HU
Slice 2 4.06 HU
Tolerance 3.78 ... 4.62 HU

2. Quality Daily Homogeneity

Description
The measurement is performed for two sets of scan parameters, representing a typical head-mode and a typical
body mode. The water value and homogeneity is evaluated by measuring the mean CT number in five Regions Of
Interest (ROI) in the image of a cylindrical water phantom with diameter 20 cm. The ROI are located in the center
and near the edge of the water phantom at 12 o'clock, 3 o'clock, 6 o'clock and 9 o'clock. The water value is the CT-
number of the central ROI, the homogeneity is the difference between the outer ROI and the central ROI each.

Common Parameters
Voltage 120 kV
UHR no
Shaped filter Wide
Window width 500 HU
Window center 0 HU
FoV 250 mm

2022-09-05 09:25:43 Page: 4 / 6


System Serial Number: 31046
Biograph 40
Customer ID: 01

Test: Typical body mode


Test Parameters Test Results OK
System A
Scan Type Sequence
Current 420 mA
Slice width reconstructed 5.0 mm
Number of images 2
Collimation 20x0.6 mm
Scan Time 0.5 s
Rotation time 0.5 s
Body Region Body
Kernel S80f
Radius of central ROI 10.00 mm
Radius of peripheral ROI 10.00 mm

Water Value Results


Water Value Results for System A
Slice 1 -1.61 HU
Slice 2 -2.25 HU
Tolerance -4.00 ... 4.00 HU

Results for System A


System A
Center Diff.3 Diff.6 Diff.9 Diff.12
Slice 1 -1.61 HU 1.29 HU 1.49 HU 1.06 HU 0.96 HU
Slice 2 -2.25 HU 1.69 HU 1.81 HU 1.77 HU 1.35 HU
Tolerance -4.00 ... 4.00 HU -4.00 ... 4.00 HU -4.00 ... 4.00 HU -4.00 ... 4.00 HU -4.00 ... 4.00 HU

Test: Typical head mode


Test Parameters Test Results OK
System A
Scan Type Rotation
Current 350 mA
Slice width reconstructed 4.8 mm
Number of images 3
Collimation 12x1.2 mm
Scan Time 1.0 s
Rotation time 1.0 s
Body Region Head
Kernel S80s
Radius of central ROI 10.00 mm
Radius of peripheral ROI 10.00 mm

2022-09-05 09:25:43 Page: 5 / 6


System Serial Number: 31046
Biograph 40
Customer ID: 01

Water Value Results


Water Value Results for System A
Slice 1 -0.63 HU
Slice 2 -1.30 HU
Slice 3 -1.69 HU
Tolerance -4.00 ... 4.00 HU

Results for System A


System A
Center Diff.3 Diff.6 Diff.9 Diff.12
Slice 1 -0.63 HU 0.58 HU 0.39 HU 0.18 HU 0.55 HU
Slice 2 -1.30 HU 0.94 HU 0.98 HU 0.99 HU 1.03 HU
Slice 3 -1.69 HU 1.00 HU 1.23 HU 1.07 HU 0.90 HU
Tolerance -4.00 ... 4.00 HU -4.00 ... 4.00 HU -4.00 ... 4.00 HU -4.00 ... 4.00 HU -4.00 ... 4.00 HU

2022-09-05 09:25:43 Page: 6 / 6