Professional Documents
Culture Documents
F 6 Control Panel
F 6 Control Panel
User’s Guide
Range.................................................................................................................................... 2-10
Sense/Timer.......................................................................................................................... 2-11
Input ............................................................................................................................... 2-11
Timer .............................................................................................................................. 2-12
Pulse Width .................................................................................................................... 2-13
Timer Unit ...................................................................................................................... 2-14
Sense Condition Indicator .................................................................................................... 2-14
Reset ..................................................................................................................................... 2-14
Frequency Settings ............................................................................................................... 2-14
Battery Simulator Settings.................................................................................................... 2-14
Source Configuration............................................................................................................ 2-15
Ramp/Set/Pulse Controls...................................................................................................... 2-16
Ramp/Set/Pulse .............................................................................................................. 2-16
Rate/Delta Step............................................................................................................... 2-17
Ramp/Set/Pulse Procedure ............................................................................................. 2-17
File Save and Open............................................................................................................... 2-18
Timing Test Procedure ............................................................................................................... 2-19
Zoom/Unzoom...................................................................................................................... 5-35
Phase..................................................................................................................................... 5-35
Input/Output Meter............................................................................................................... 5-35
Digital and Energy Recorder ................................................................................................ 5-36
Output ID Field .............................................................................................................. 5-37
Values Are Field ............................................................................................................ 5-37
Mode..................................................................................................................................... 5-37
Device Settings for Digital Output Transducers......................................................................... 5-37
Nameplate Data .................................................................................................................... 5-38
Watthour Tab........................................................................................................................ 5-39
Varhour Tab.......................................................................................................................... 5-39
Volt-Amphour Tab ............................................................................................................... 5-40
Digital Output Transducer Tabs ........................................................................................... 5-41
Auto Settings .................................................................................................................. 5-42
Input Mask ..................................................................................................................... 5-44
Battery Simulator Settings.......................................................................................................... 5-45
File Save and Open ..................................................................................................................... 5-45
Reports ........................................................................................................................................ 5-47
Viewing ................................................................................................................................ 5-47
Report Configuration............................................................................................................ 5-49
Saving ................................................................................................................................... 5-50
Opening ................................................................................................................................ 5-51
C. Vector Calculator..................................................................................C-1
Vector Table .................................................................................................................................C-1
Keypad..........................................................................................................................................C-2
Function Keys ........................................................................................................................C-2
Operator Keys ........................................................................................................................C-3
Number Keys .........................................................................................................................C-4
Recalculate Formulae Key .....................................................................................................C-4
Viewing Vectors...........................................................................................................................C-5
Menu Bar ......................................................................................................................................C-5
File..........................................................................................................................................C-5
Edit .........................................................................................................................................C-7
View .......................................................................................................................................C-7
Formulae ........................................................................................................................C-10
Help ......................................................................................................................................C-11
Toolbar .......................................................................................................................................C-12
Performing Calculations.............................................................................................................C-12
Source Name Column ..........................................................................................................C-12
Formula Column ..................................................................................................................C-13
Performing Formula Box Calculations.......................................................................................C-13
Formula Operators ...............................................................................................................C-13
Save the Answer in the Vector Table...................................................................................C-14
Features
F6ControlPanel emulates front-panel controls on the F6150. It has flexible
data-entry procedures to accommodate the wide range of test configurations
possible.
F6ControlPanel contains four different control panel types: QuickTest,
Generic, Dynamic, and Transducer. These control panels are used for manual
and automatic control of F6150 sources. Choose a suitable control panel to
coordinate with the test instrument you are using:
The Startup Options dialog box appears, superimposed on the splash screen
(Figure 1-2). Select your preferences:
Language Options. Choose the language you want to use with the software
(English, Chinese, Portuguese, or Spanish). Default is English.
Startup Mode. Choose whether F6ControlPanel opens with the QuickTest
control panel or the interface for the Generic and Dynamic devices. Default is
QuickTest.
Ask Again. Choose whether you wish this screen to appear each time you
open the software. Default is Ask Again.
If you select Don’t Ask Again and you later want to show the Startup Options
dialog box, you can access it by clicking the Select Language icon in the
QuickTest device (see the QuickTest toolbar icon descriptions at the beginning
of Chapter 2).
Click OK. Check the Don’t ask again check box if you prefer not to have the
dialog box appear again.
The window you have selected opens—either the QuickTest control panel
(Figure 1-9) or the application window for the Generic and Dynamic control
panels, with the Open File dialog box (Figure 1-5).
Figure 1-5 Generic and Dynamic Devices Application Window with Open
File Dialog Box
You can also open the desired application by clicking Cancel and then
clicking the appropriate launch button on the toolbar (Figure 1-8). The launch
buttons are the same on the toolbars of all three devices.
After opening the control panel, if you are using it for the first time, proceed to
the Setup instructions at the beginning of the chapter for that control panel.
The Setup dialog box for the Generic and Dynamic control panels is the same.
The QuickTest control panel has a different Setup dialog.
The setup parameters for any of the devices are replicated in the other devices.
Some of the Generic and Dynamic parameters, however, are not used in the
QuickTest control panel and do not appear in that Setup dialog.
Figure 1-9 shows the QuickTest main screen, Figure 1-10 shows the Generic
main screen, and Figure 1-11 shows the Dynamic main screen.
The QuickTest control panel (with touchscreen keypad if using the F6010
Handheld Controller) provides simple, rapid, manual operation of an F6150
Power System Simulator. QuickTest gives you individual control of voltage,
current, phase angle, and frequency. It provides one sense circuit, one
programmable timer, and one logic input circuit. QuickTest enables
programmable linear ramps, step changes, and pulse mode operation.
Figure 2-1 shows the main screen of the QuickTest control panel.
Title Bar
The title bar (Figure 2-2) displays the Doble logo, the title F6ControlPanel,
and the name of the active control panel. Click on the logo to access a
drop-down menu that includes:
• Restore
• Move
• Size
• Minimize
• Maximize
• Close
Status Bar
Figure 2-3 shows the status bar at the bottom of the main screen. The status
bar displays messages related to program function, such as the menu item
selected and whether caps lock, number lock, and the battery simulator are
engaged.
Menu Bar
The only item in the QuickTest menu bar (Figure 2-4) is an icon that displays
the same drop-down menu as the logo in the title bar.
Menu bar options are disabled or grayed out if they are not applicable to the
current context window.
Shortcut keys are:
Close Ctrl + F4
Exit Alt + F4
Toolbar
Figure 2-5 shows the QuickTest toolbar. Table 2-1 describes icon
functionality.
Setup
Click the Setup button (Figure 2-7) to open the QuickTest control panel.
Setup
Connect with
• Ethernet — Enter the IP address of the F6150 in the IP Address field
of the Setup dialog box. The IP address of the instrument appears in
the front panel display when the F6150 is turned on and the firmware
loads.
Use the Network connector on the F6150 (see Appendix B, “Ethernet
Communications” for connection information).
• Wireless — Enter the IP address of the F6150 in the IP Address field
of the Setup dialog box. The IP address of the instrument appears in
the front panel display when the F6150 is turned on and the firmware
loads.
• Serial — Use the COM button in the Comm Port field of the Setup
dialog box to select the appropriate port for your system. Your choices
are COM1 to COM20. If your computer does not communicate with
the F6150 on COM1, make sure the communication port in the Setup
display matches the port assigned in Windows.
Use the RS-232 connector on the F6150 (see Appendix B, “Ethernet
Communications” for connection information).
Simulation/Instrument
To run in simulation mode, select Simulation. Simulation mode is useful for
training and for configuring tests that will be conducted later. Also use
simulation mode if your computer is not connected to an instrument or if the
instrument is switched off.
LL/LN button
Select line-to-line or line-to-neutral.
Battery Selection
The Fixed/Variable button enables you to select one of three fixed battery
voltages or to enter an arbitrary voltage. Fixed gives you the option of 48, 125,
or 250 volts on the main screen. Variable enables you to enter a battery
voltage between 6 and 300 volts, in 1-volt increments, on the main screen. The
Fixed/Variable button is available only when the Simulation setting is selected
with the Simulation/Instrument button. The 6875 Variable Option Battery
Simulator option must be installed in the F6150 for variable battery settings to
be available.
After all the settings in the Setup dialog box are changed or confirmed, click
OK to accept the modifications and close the display or Cancel to close
without change.
Source Table
The Source table (Figure 2-9) at the left of the panel contains five columns.
The first column enables you to select the sources to be controlled by the
Ramp/Set/Pulse controls. The middle three columns contain settings for each
source. The last column is an Enable/Disable switch for each source. If you
click Master, all sources are enabled, including those that have not been
selected.
Source
The Source column (which has a computer icon on the top button) shows the
configured sources. The Source button background is yellow when a source is
selected. The selected sources can then be controlled by the Ramp/Set/Pulse
controls.
In Figure 2-9, the Source column shows that six out of eight sources are
selected: three voltage sources (VA, VB, and VC) and three current sources
(I1, I2, and I3).
Amplitude
Amplitude indicates the voltage or current value of a source. Maximum
amplitude is determined by the range settings (see “Range” on page 2-10). If
the entered amplitude exceeds the maximum range value, an error message
appears. To correct the error, reduce the amplitude or increase the range.
Phase
Phase indicates the phase angle in degrees. Phase angle may be entered in
positive or negative values (0+359.9o)
Frequency
The source frequency can be changed in multiples of the system frequency or
DC values using the Hz Up and Hz Down buttons or in 0.001 Hz increments
using the Ramp/Set source function.
Use the Setup dialog box to change the system frequency. Use the Hz Up and
Hz Down buttons to select an AC harmonic (1–20) or a DC+ or DC– range.
Master
The Master button enables all configured sources with a single click,
regardless of whether they have been selected. The Master and Enable buttons
are green when enabled. Individual sources can be disabled by clicking the
Enable button for each source. No power is applied, however, until you click
the System Output button.
System Output
Clicking the System Output button (Figure 2-10) turns all enabled sources on
simultaneously at the desired amplitude and phase angle. Clicking the button
again turns off all enabled sources.
Abort
To turn off all enabled sources during a test, click Abort (Figure 2-10).
Operating Controls
The right side of the main screen contains operating controls (Figure 2-10)
that enable you to perform tests with the instrument.
Range
The range buttons determine the maximum value for the amplitude of a
source. Available ranges are determined by the source configuration and type
of source (AC or DC). The selected range determines the available resolution
of the source values.
Sense/Timer
F6ControlPanel enables you to control any of the eight logic inputs and eight
logic outputs on the F6150. Each source on the instrument maps to one logic
output and one logic input on F6ControlPanel. Logic outputs are controlled by
the source’s change of state. The Sense/Timer dialog box enables you to
control logic inputs.
Click the Sense/Timer button to open the Sense/Timer dialog box
(Figure 2-11). This dialog box enables you to specify a timer with individually
selected start and stop conditions.
Only one timer is available in QuickTest. For additional timers, use the
Generic or Dynamic device.
Input
• Input — Select an input corresponding to the F6150 logic inputs
(Input1 through Input8).
• Type — Select the type:
Potential: The device under test supplies the sense potential for
the input.
Contact: The F6150 supplies the sense potential for the input. For
Contact, the red input terminal is positive with respect to the white
terminal.
• Sense Condition — Select the transition that must occur for the input
to be true:
Potential:
On Off: Sense condition is satisfied when logic input records a
change from potential to no potential.
Off On: Sense condition is satisfied when logic input records a
change from no potential to potential.
Contact:
Open Closed: Sense condition is satisfied when logic input
records a change from open contact to closed contact.
Closed Open: Sense condition is satisfied when logic input
records a change from closed contact to open contact.
Whether the input type is potential or contact, the first sense condition is
always false and the second sense condition is always true (Table 2-2).
Sense
Input Type Description
Condition
Timer
• Timer — Select a timer corresponding to the F6150 timer (Timer1
through Timer8).
• Start On Src/Input — Select a configured source or a defined logic
input to start the timer. The button will cycle through the sources
selected in the Source Configuration dialog box as well as logic inputs
LN1 through LN8.
If a source is selected, the timer starts when the source start condi-
tion is met.
If a logic input is selected, the timer starts when the logic input
condition goes from false to true.
• Stop On Src/Input — Select a configured source or a defined logic
input to stop the timer. The button cycles through the sources selected
in the Source Configuration dialog box as well as logic inputs LN1
through LN8.
If a source is selected, the timer stops when the source start condi-
tion is met.
If a logic input is selected, the timer stops when the logic input
condition goes from false to true.
• Sense Latch — Select Latch to have the sense condition indicator (to
the right of the Sense/Timer button) seal-in after the sense condition is
satisfied. Latching is used primarily when the Ramp/Set/Pulse button
is in Pulse mode. Without latching, the flash of the indicator is so brief
that it typically cannot be seen.
• Start Condition — Select the logic state to start the timer at the
beginning of the test (this button is available only when Start On
Src/Input is a source and not a logic input):
On Off: Timer starts when the selected source turns off.
Off On: Timer starts when the selected source turns on.
On Change: Timer starts when the up or down arrow is pressed
and the source is on.
• Stop Condition — Select the logic state to stop the timer at the end of
the test (this button is available only when Stop On Src/Input is a
source and not a logic input):
On Off: Timer stops when the selected source turns off.
Off On: Timer stops when the selected source turns on.
On Change: Timer stops when the up or down arrow is pressed
and the source is on.
Pulse Width
Enter a value for the pulse width (duration) in cycles, or click in this field to
open a keypad that allows you to enter a value. Default is 100. When done,
press Enter on the keypad.
Timer Unit
Select the desired increments for the timer: ms (milliseconds), s (seconds), or
cycles.
When done with the Sense/Timer dialog box, click OK to save your settings.
Reset
Click the Reset button to clear the timer field above it and unlatch the Sense
Condition Indicator.
Frequency Settings
Use the Hz Up and Hz Down buttons to select an AC harmonic (1–20) or a
DC+ or DC– range. The buttons are available only when the cursor is in a
frequency field. The frequency changes only in multiples of the system
frequency. Use the Setup dialog box to change the system frequency.
Source Configuration
Click the Source Configuration button (to the right of the battery simulator
buttons) to open the Source Configuration dialog box (Figure 2-12).
Ramp/Set/Pulse Controls
The Ramp/Set/Pulse button and Rate/Delta Step button enable you to rapidly
change the values in the source table.
Ramp/Set/Pulse
Click this button to select the method to vary the source table values.
• Set — Varies the source table by discrete steps each time the up or
down arrow button is pressed or the control knob is turned (if using the
F6010 Handheld Controller).
• Ramp — Varies the source table values continuously as long as the up
or down arrow button is pressed or the control knob is turned (if using
the F6010 Handheld Controller).
• Pulse — Limits the output duration for the selected source to the Pulse
Width specified in the Sense/Timer dialog box, to limit potential
damage due to excessive current.
The values will change by the increment set with the Rate/Delta Step button or
entered in the field at the right.
Rate/Delta Step
Click this button to select the increment used to vary the source table values.
The default values are 0.001, 0.01, 0.1, 1, 10, and 100. You can also enter a
custom value by typing the value in the box. The value must be between the
smallest and largest default values.
Ramp/Set/Pulse Procedure
1. Click the Ramp/Set/Pulse button to select the mode you wish to use.
2. Click the Rate/Delta Step button to select the desired increment.
3. Click the Amplitude, Phase, or Frequency button at the top of the
appropriate column to select the values you wish to vary.
The button turns yellow. The changes apply to the selected sources in this
column. Selected sources are indicated by yellow source buttons in the
first column.
4. Click the Up or Down arrow button at the bottom of the panel to vary the
selected sources simultaneously.
In Ramp mode, hold the button down for continuous change.
The frequency of the first source in the source table varies independently of
the other seven sources. The frequencies of sources 2 through 8 vary together.
For example, if VA is the only selected source in the source table, the
frequency for VA is the only variable that changes when the up or down arrow
is pressed. If any other source is selected, however, the frequencies for all
other sources change at the same rate.
Filenames for saved files in the QuickTest control panel all begin with
“QuickTest.” Stored configuration files for all four control panels,
however, will appear as you cycle through the list. If you load a file you
saved in the Generic or Dynamic control panel, the main screen for that
device, with the values loaded, will appear when you click OK.
If using the F6010 Handheld Controller, you can use the control knob to
change values instead of the up and down arrows.
1. On the main screen, select the desired sources by clicking the Source
Configuration button and selecting the desired configuration.
2. Click Sense/Timer to open the Sense/Timer screen.
3. Select the inputs, types, and sense conditions.
4. Select the timers for testing and set the start and stop conditions.
5. Click OK to save the Sense/Timer settings.
6. Click the Ramp/Set/Pulse button to reach Set.
7. Change the source table values to the desired settings, using the up and
down arrows.
8. Click the Rate/Delta Step button to select a value, or enter a value in the
entry field.
9. Click the Phase, Amplitude, or Frequency button to select the column of
values you wish to change.
10. Click the Up or Down arrow to make appropriate changes to the source
values.
11. Click Reset.
Ensure that the timer field resets.
12. Enable the selected sources by clicking the Off button for each source (to
enable the sources individually) or the Master button (to enable all
selected sources simultaneously).
13. When the timer, sources, and parameters are chosen, click System
Output.
14. The timing result is displayed in the timing fields.
The Generic control panel enables you to conduct basic manual relay tests.
The Generic control panel gives you control of up to twelve sources, with
eight timers and eight logic inputs and outputs, for more advanced manual
control than QuickTest. It provides graphical vector representation for
positive, negative, and zero-sequence quantities. An analog meter displays the
action output of the selected source. Figure 3-1 shows the main screen of the
Generic control panel.
Title Bar
The title bar (Figure 3-2) displays the Doble logo, the title F6ControlPanel,
and the name of the active control panel. Click the logo to access a drop-down
menu that includes:
• Restore
• Move
• Size
• Minimize
• Maximize
• Close
Status Bar
Figure 3-3 shows the status bar at the bottom of the main screen. The status
bar displays messages related to program function, such as the menu item
selected and whether caps lock, number lock, and the battery simulator are
engaged.
Menu Bar
Each item in the Generic menu bar (Figure 3-4) is discussed below. Some
items are available only in one of the control panels or when F6ControlPanel
is running but the devices are closed. To close a control panel, click the small
gray X in the upper right corner. Menu bar options are disabled or grayed out
if they are not available in the current window.
Open Ctrl + O
Save Ctrl + S
Close Ctrl + F4
Exit Alt + F4
Copy Ctrl + C
Cut Ctrl + X
Paste Ctrl + V
Find Ctrl + F
The first icon displays the same dropdown menu as the logo in the title bar
(Figure 3-4).
File
File menu (Figure 3-5) items are:
• Open — Opens a file. An Open dialog box appears, from which you
can select a file or browse to the desired file. Shortcut key: Ctrl + O
• Close — Closes the current control panel but leaves the application
open. Shortcut key: Ctrl + F4
• Save — Saves the control panel configuration into the currently
opened file. Shortcut key: Ctrl + S
• Save As — Saves the control panel configuration as a .f6x file under a
name you enter.
• Print Setup — Opens the standard Windows Print dialog box.
• Recent Files — Displays recently used files.
• Exit — Exits the program. Shortcut key: Alt + F4
Edit
Edit menu (Figure 3-6) items are:
• Cut — Cuts the selected file or data and retains it on the clipboard.
Shortcut key: Ctrl + X
• Copy — Copies the selected file or data and retains it on the
clipboard. Shortcut key: Ctrl + C
• Paste — Inserts data from the clipboard. You can do multiple pastes
from one cut or copy. Shortcut key: Ctrl + V
View
View menu (Figure 3-7) items are:
Device
Figure 3-8 shows the Device menu with the control panels closed.
Device menu (Figure 3-9) items with one of the devices open are:
Tools
Tools menu (Figure 3-10) items are:
Setup
The Setup menu (Figure 3-12) items are:
For information on the Setup menu options, refer to “Setup” on page 3-11.
Summary
Summary menu items (available only in the Dynamic control panel) enable
you to view, save, and configure reports of test results.
Help
Help menu (Figure 3-14) items are:
Toolbar
Figure 3-15 shows the Generic and Dynamic toolbar with F6ControlPanel
running, but the control panels closed. Some icons are available only in one of
the devices or when both devices are closed. To close a control panel, click the
small gray X in the upper right corner of the screen. Toolbar options are
disabled or grayed out if they are not available in the current window.
Table 3-1 describes icon functionality.
Shortcut
Icon Name Key Function
Setup
Open the Setup dialog box (Figure 3-16) by clicking the Setup icon in the
toolbar. This dialog box presents the same options as the Setup menu
(Figure 3-12).
Setup is not accessible when a control panel is open. To access the Setup
dialog box for the Generic and Dynamic devices, click the small gray X in
the upper right corner of the screen. This closes the control panel without
closing the F6ControlPanel application. Then, select Tools Setup or the
Setup icon on the toolbar to open the Setup dialog box.
This dialog box enables you to select the control panel you want to load as a
default, with its associated settings.
Database File
Not available in this version.
Communication
The Communication option enables you to set the communication parameters.
The On and Enable switch icons represent toggle switches; the position of the
switch changes when you click it to turn a source on or enable it. The default
position for the switch is down for off and up for on. You can reverse this by
selecting Up from the drop-down menu.
Source Parameters
This option enables you to set the Source Parameters dialog box (Figure 3-22).
Figure 3-23 Generic and Dynamic Devices — Assign Source Dialog Box
Test Defaults
Not available in this version.
Source Table
The Source table (Figure 3-25) at the upper left contains seven columns. The
first five columns contain the settings for each source; the last two columns
contain On/Off and Enable/Disable switches for each source.
Figure 3-25 Source Table with Three Voltages and Three Currents
Configured
If a source error occurs, the alarm is visible in the source table. The name
of the source affected changes to ER and blinks. The Amplitude and
Phase fields for that source also blink and an audible alarm sounds from
the speakers of the control PC.
Source
The Src column shows all configured sources. The source background is green
when a source is enabled and red when a source is on.
The Src column in Figure 3-25 shows that six out of twelve sources are active:
three voltage sources (VA, VB, and VC) and three current sources (I1, I2, and
I3). Configure your sources with the F6000 Configuration dialog box (see
“F6000 Configuration” on page 3-21).
Amplitude
Amplitude indicates the voltage or current value of a source. Maximum
amplitude is determined by the range settings (see “Range” on page 3-20). If
the entered amplitude exceeds the maximum range value, an error message
appears. To correct the error, reduce the amplitude or increase the range.
Right-click menu selections are:
• Undo — Undoes a change to a value in a data field. Available only if
selected before clicking outside the data field.
• Cut — Cuts a value in a data field. Available only if the value is
highlighted by clicking Select all or double-clicking with the cursor.
• Copy — Copies a value in a data field. Available only if the value is
highlighted by clicking Select all or double-clicking with the cursor.
• Paste — Pastes the contents of the clipboard into a data field at the
insertion point.
• Delete — Deletes a value in a data field. Available only if the value is
highlighted by clicking Select all or double-clicking with the cursor.
• Select All — Highlights the value in a data field (all digits).
• Apply Nominal Values to All Sources — Applies the Nominal
amplitude from Setup to all sources.
• Apply Nominal Value to This Source — Applies the Nominal
amplitude from Setup to the selected source.
• Apply Amplitude to Same Source Type — Applies the amplitude of
the selected source to all other sources of that type.
Phase
Phase indicates the phase angle in degrees. Phase angle may be entered in
positive or negative vales (0+359.9o).
Phase right-click menu selections are:
• Undo
• Cut
• Copy
• Paste
• Delete
• Select All
• Apply Default Source Angles to All Sources — Applies the default
source angles from Setup to all sources.
• Apply Default Source Angles to This Source — Applies the
appropriate default source angle from Setup to the selected source.
The first six right-click selections are the same as for Amplitude.
Frequency
The source frequency can be changed only in multiples of the system
frequency. Use the Setup dialog box to change the base frequency. Click the
up or down arrows to the right of the Frequency column to select the AC
harmonic (1–20) or a DC or –DC range.
Frequency right-click menu selections are:
• Undo
• Cut
• Copy
• Paste
• Delete
• Select All
These are the same as the first six right-click selections for Amplitude and
Phase.
Range
The range setting determines the maximum value for the amplitude of a
source. Available ranges are determined by the source configuration and type
of source (AC or DC). The selected range determines the available resolution
of the source values. Use the up or down arrows to the right of the Range
column to select the desired range.
The Range right-click menu selection is:
• Apply This Range to Same Source Type — Applies the range of the
selected source to all other sources of that type.
On
On/off is typically used for single-point calibration of voltage, current, phase
angle, or frequency.
Click the On switch to activate a source manually. The button turns red and
the System Output button blinks red. The Abort button is solid red when any
source is on (except the battery simulator). Click the On switch again to turn a
source off and the switch turns gray.
When the System Output button blinks red, sources must be turned off
manually by clicking their On buttons or the Abort button. The System
Output button stops blinking when all sources have been turned off.
When sources are turned on with the On button, they must be turned off
by clicking their On buttons again or the Abort button.
Enable
Enable and Disable are typically used for timing functions: when a time value
is reached, the sources shut off.
Click the Enable button to enable a source. The Enable button for that source
turns green.
To enable or disable the Enable switches for all configured sources, right-click
on the Enable column head and select Enable All or Disable All from the
popup menu, or use the Master Enable or Master Disable checkboxes (see
below).
Enabling multiple sources allows you to apply those source settings
simultaneously.
Master Enable
When this box is checked, clicking the Enable button for any configured
source will enable all configured sources.
Master Disable
When this box is checked, clicking the Disable button for any configured
source will disable all configured sources.
To turn off all active sources during a test, click Abort. Clicking Abort
turns off the battery simulator only if Battery Off on Abort is checked in
Setup.
F6000 Configuration
Use the F6000 Configuration dialog box (Figure 3-27) to configure your
sources.
1. Open the dialog box by selecting Tools F6000 Configuration or by
clicking the F6000 Configuration icon on the toolbar.
2. Click the User-Defined or Preset option button.
Ramp/Set Controls
The Ramp/Set Sources section (Figure 3-28) enables you to control the values
in the source table.
Ramp/Set Sources
The Ramp/Set Sources section has twelve checkboxes, with one active
checkbox per configured source. To change a variable (phase angle,
amplitude, or frequency) for a source, check the box for that source.
Alternatively, right-clicking on the space between the two columns of
checkboxes opens a menu that enables you to select or deselect all current
sources, all voltage sources, or all sources.
To avoid altering the values for a source during a test, make sure the box
for that source is not checked.
Change
To change parameters, choose a variable in the Change field to vary:
• Click the Phase option button to vary the phase angle of the selected
sources.
• Click the Amplitude option button to vary the voltage or current of
the selected sources.
• Click the Frequency option button to vary the frequency of the
selected sources.
The frequency of the first source in the source table varies independently
of the other sources. The frequencies of sources 2 through 12 vary
together.
Up/Down Arrows
The up and down arrows enable you to change selected source parameters.
The up arrow increments and the down arrow decrements the selected source
parameters. The type of change depends on the mode selected (see Mode
below).
When the up arrow is pressed, the checked sources increment by the value in
the box above the arrows or increase at the rate specified in the box. The
setting cannot increase beyond the limit set in the Range column of the source
table. The setting stops at the last valid value and remains there.
If Phase is selected, the up arrow rotates phasors counterclockwise; the down
arrow rotates them clockwise.
Mode
The settings in the source table can be varied in discrete steps in Set mode or
continuously in Ramp mode:
• Set — Select a value from the list or enter the delta step manually. The
selected source variables increment or decrement by this amount each
time the up or down arrow is pressed.
• Ramp — Select a value from the list or enter the rate (value/second)
manually. The selected source variables increment or decrement at this
rate when the up or down arrow is pressed and held.
For both modes, the values in the dropdown menu are 0.001, 0.01, 0.1, 1, 10,
and 100. You can also enter a custom value by typing the value in the box. The
value must be between the smallest and largest default values.
The Store and Recall functions are available only for the current session. Once
the control panel or application is closed, the values are lost.
2. Enter a name for the file or accept the suggested name. To save the settings
as the default for future sessions, select default.f6x.
3. Click Save.
To load the file:
1. Select File Open.
The Open F6 Control Panel File dialog box appears (Figure 3-30).
Each source maps to one logic output. The mapping of sources to outputs
depends on the source configuration in effect. The mapping rule assigns
outputs to voltage and current sources in ascending order, first from left to
right, then from top to bottom. Figure 3-32, Figure 3-33, and Figure 3-34
illustrate how the rule works for three common source configurations.
Sources V1 V2 V3 I1 I2 I3
Logic nos. 1 2 3 4 5 6
Figure 3-32 Output Indicators for Three Voltages and Three Current
Sources
Sources V1 V2 V3 I1 I2 I3
Logic nos. 1 2 3 4 5 6
Sources VN IN
Logic nos. 7 8
Figure 3-33 Output Indicators for Four Voltages and Four Current Sources
Sources I1 I2 I3
Logic nos. 1 2 3
Sources I4 I5 I6
Logic nos. 4 5 6
The logic output is programmable and changes state when its associated
source goes on or off. The output is in its normal state when the source is off.
A normally open output closes when the source is turned on, and its
corresponding status indicator illuminates. The logic output indicators are
gray when open and green when closed.
A logic input is a signal that originates with the device under test and is sent to
the instrument. The logic input indicators are gray for a false state and red for
a true state (i.e., when the input type and sense condition are satisfied).
Logic Settings
Click Logic Settings to open the Logic Settings window. The Generic device
has four tabs.
Inputs Tab
The Inputs tab contains the settings for eight logic inputs, one for each input
terminal on the instrument’s front panel. The Inputs tab also supports controls
for the AutoSense facility and the threshold value. Figure 3-35 shows settings
for I/O 1 and I/O 5 boards. Figure 3-36 shows settings for I/O 2 and I/O 4
boards.
For each input, choose the Type and the Sense Condition:
Type
Select Potential or Contact from the drop-down menu:
• Potential: The device under test supplies the sense potential for the
input.
• Contact: The F6150 supplies the sense potential for the input. With
Contact, the red input terminal is positive with respect to the white
terminal.
Sense Condition
Select the transition that must occur for the input to be true:
• Potential:
On Off: Sense condition is satisfied when logic input records a
change from potential to no potential.
Sense
Input Type Description
Condition
Inputs cannot directly start or stop timers—only triggers can. Triggers are
made up of logic inputs using Boolean logic.
AutoSense
AutoSense enables you to stop a manual source ramp when a logic input goes
from false to true. Choose AutoSense by clicking the option button for the
appropriate input.
Only one logic input can be chosen as the AutoSense input. If you require
multiple inputs to have this special function, they must be jumpered together
with the single defined AutoSense input.
Use the Off checkbox to disable AutoSense.
Threshold
The default threshold setting is 1.5 V. The option for 15 V is provided to
reduce sensitivity to circuit noise. Threshold applies only when the input type
is Potential.
Outputs Tab
The Outputs tab enables you to individually configure up to eight output
contacts as either normally open or normally closed (Figure 3-37).
To configure outputs, click the appropriate option button (Normally Open or
Normally Closed) for each contact and click OK.
The Logic I/O display on the main screen reflects the configuration. A green
status indicator designates a normally closed contact. As an example, the
configuration shown in Figure 3-37 results in the main screen display in
Figure 3-38.
Timers Tab
The Timers tab enables you to establish up to eight timers with individually
selected start and stop conditions (Figure 3-39).
Sources will not turn off automatically until all timers are stopped. Use
only the number of timers needed for each test.
Sources turn off when timers stop, but only when the sources are turned
on using the Enable and System Output buttons.
Setting Timers
1. In the Set Timer column, select each required timer by its option button.
2. In the Set Timer Conditions pane, Select a configured source or defined
trigger in the Start on Source drop-down menu:
• If a source is selected, the timer starts when the source start condition
is met.
• If a trigger is selected, the timer starts when the trigger condition goes
from false to true.
3. Select a Start Condition (this is active only when Start on Source is a
source and not a trigger):
• On Off: Timer starts when the selected source turns off.
• Off On: Timer starts when the selected source turns on.
• On Change: Timer starts when the up or down arrow is pressed and
the source is on.
4. Select a configured source or defined trigger in the Stop on Source/Trigger
drop-down menu:
• If a source is selected, the timer stops when the source start condition
is met.
• If a trigger is selected, the timer stops when the trigger condition goes
from false to true.
5. Select a Stop Condition (this is active only when Start on Source is a
source and not a trigger):
• On Off: Timer starts when the selected source turns off.
• Off On: Timer starts when the selected source turns on.
• On Change: Timer starts when the up or down arrow is pressed and
the source is on.
6. Click Set to update each timer’s start and stop conditions. To clear a
timer’s start and stop conditions, click Reset.
7. Click Apply to update all the settings on the tab without closing the dialog
box.
8. Click OK to close the dialog box with updated settings from all tabs.
Inputs cannot directly start or stop timers—only triggers can. Triggers are
made up of logic inputs using Boolean logic.
Define Triggers
Click Define Triggers to open the Triggers dialog box and set each Trigger
Name and define Trigger Logic (Figure 3-40).
The eight Set buttons correspond to eight possible triggers. You can accept the
default trigger name or enter a unique name. The default trigger names and
logic input pairings are:
• Trigger1: Logic Input 1 (In1)
• Trigger2: Logic Input 2 (In2)
• Trigger3: Logic Input 3 (In3)
• Trigger4: Logic Input 4 (In4)
• Trigger5: Logic Input 5 (In5)
• Trigger6: Logic Input 6 (In6)
• Trigger7: Logic Input 7 (In7)
• Trigger8: Logic Input 8 (In8)
Click Set to open the Set Trigger Logic window. Several logic operators are
available per input (Figure 3-41 and Figure 3-42).
When the And mask operator is selected, the inputs within a group are
Anded together. For example:
1. Select an And mask operator.
2. Click 1, and In1 appears in the logic statement.
3. Click 2, and *In2 is appended to the statement, to produce In1 *
In2. This means that both In1 and In2 must be true for trigger 1 to
be true.
When the Or mask operator is selected, the inputs within a group are
Ored together. For example:
1. Select an Or mask operator.
2. Click 1, and In1 appears in the logic statement.
3. Click 2, and + In2 is appended to the statement, to produce In1 +
In2. This means that either In1 or In2 must be true for trigger 1 to
be true.
• Group Operator — Group Operator defines how the logic groups are
combined. The logic statement allows a maximum of two groups. A
group is created when Group Inputs is clicked. A group consists of the
logic elements between the open and close parentheses.
The Mask Operator and Group Operator must be different. For
example, if the Mask Operator is equal to And, the Group Operator
must be Or.
When the And group operator is selected, the logic statement groups
are Anded together. For example:
1. Select an And mask operator and an Or group operator.
2. Click 1, and In1 appears in the logic statement.
3. Click 2, and *In2 is appended to the statement, to produce In1 *
In2.
4. Click Group Inputs; open and close parentheses appear around
the existing statement to produce (In1 * In2).
5. Click 3, and + In3 is appended to the statement. The statement
now reads (In1 * In2) + In3. This means that either (In1 * In2) or
In3 must be true for trigger 1 to be true.
When the Or group operator is selected, the logic statement groups are
Ored together. For example:
1. Select an Or mask operator and an And group operator.
2. Click 1; In1 appears in the logic statement.
Notes Tab
Use the Notes tab (Figure 3-43) to document any part of a test setup or test
procedure by entering text such as a test sequence or the reasons for key
settings. When in simulation mode, the settings for a test can be entered and
saved. A field technician can use the information in the Notes tab to set up and
conduct the test.
Phasor Diagram
The phasor diagram (Figure 3-44) provides a visual representation of the
amplitude and phase values in the source table. The diagram is based on polar
coordinates. Each phasor represents the amplitude and phase angle of a
source. The distance from the origin to the endpoint of the phasor represents
the source’s amplitude. The angle formed by the phasor and the positive half
of the horizontal axis represents the source’s phase angle.
The source table and phasor diagram interact with each other. Any single
source’s amplitude and phase angle can be changed by clicking its arrowhead
in the diagram and dragging and dropping it to a new location. The source can
be on or off. Source table values are continuously updated as phasors are
dragged to new locations, but the change is not sent to the F6150 until the
phasor is dropped.
90(-270)
180(-180) 0
270(-90)
Figure 3-44 Phasor Diagram
Scale Settings
The upper left corner of the phasor diagram contains the highest current
setting from the Range column of the source table. The upper right corner of
the phasor diagram contains the highest voltage setting from the Range
column. These settings determine the scale of the phasor diagram.
For example, if the amplitude for current source I1 is 15 A and the highest
range setting for the current sources is 15 A, the I1 phasor reaches to the
perimeter of the circle in the phasor diagram.
Similarly, if the potential for voltage source VA is 50 V and the highest range
setting for the voltage sources is 75 V, the length of the VA phasor is
two-thirds the radius of the circle in the phasor diagram.
Multi-Rotation
Right-click in the phasor diagram and select Multi-Rotation. The phasors for a
set of voltage or current sources can then be moved by dragging and dropping
any one of them. Each phasor maintains its position relative to the other two.
When phasors are being moved, the source table is continuously updated, but
the new phase angle values are not sent to the F6150 until the selected phasor
is dropped.
Setup
To configure the appearance of each phasor, right-click in the phasor diagram
and select Setup. The Properties dialog box appears (Figure 3-45). In the
dialog box, select the color, thickness, and line type for each phasor and circle.
You can also set the color for the axes and background. By default, current
phasors are thicker than voltage phasors.
Vector Selection
To select the phasors you want displayed, right-click in the phasor diagram
and select Vector Selection. The Vectors dialog box appears (Figure 3-46).
Check or uncheck the box for each active source. To select all the phasors,
click Select All. To show sequence information on the phasor diagram, click
Show Sequence Component.
Click Apply to change the vectors selected in the phasor diagram without
closing the dialog box. Click OK to close the dialog box with the new vector
selection settings in effect.
Sequence Selection
Right-click in the phasor diagram and select Sequence Selection. The
Sequence dialog box appears (Figure 3-47).
Matching Sources
Select the source for each phase. The sources can be mapped in a different
order, if desired.
Select the voltage and current sequence phasors to be displayed:
• Positive Sequence: V1 voltage and I1 current
• Negative Sequence: V2 voltage and I2 current
• Zero Sequence: V0 voltage and I0 current
Negative and zero vectors appear in the phasor diagram only if there is an
imbalance in the applied voltages and currents.
Zoom/Unzoom
Zoom/Unzoom enlarges an area of the phasor diagram. To use the feature,
right-click and select Zoom/Unzoom. The pop-up menu closes with the zoom
feature in effect.
To select an area, click the left mouse button in the diagram and drag the
cursor diagonally. A dashed rectangle appears around the selected area. When
the mouse button is released, the selected area is enlarged.
To deselect Zoom, right-click in the phasor diagram and select
Zoom/Unzoom. When the pop-up menu closes, the phasor diagram returns to
its normal state.
Action Meter
The Action meter (Figure 3-48) indicates the action output of the selected
active source.
Use the drop-down menu directly below the meter to select the active source.
The sources in the list correspond to those checked in the Ramp/Set Sources
section.
If a voltage source is selected, meter readout is in volts for amplitude, degrees
for phase, and hertz for frequency.
If a current source is selected, meter readout is in amps for amplitude, degrees
for phase, and hertz for frequency.
Depending on the parameter selected in the Change section of the main screen
and the scale setting of the meter (see Figure 3-49 below), segments of the
semicircular bar under the scale may be green, yellow, or red. When the meter
displays volts or amps:
• Scale values for a green segment are less than or equal to the Nominal
value entered in Setup.
• Scale values for a yellow segment are from nominal to 125% of
nominal for voltages and from nominal to 200% of nominal for
currents.
• Scale values for a red segment are greater than 125% of nominal for
voltages and greater than 200% of nominal for currents.
The Action meter’s right-click menu provides three options:
• Change Scale
• Display Values Only When the Selected Source Is On
• Display Values Always
To change the scale of the Action meter:
1. Right-click in the meter and click Change Scale. The Scale Settings
dialog box appears (step 3).
2. Click the Manual option button and enter the desired minimum and
maximum values for the meter scale.
3. Click OK.
Click the BAT button to toggle the battery simulator on and off. The battery
simulator has the following operating characteristics:
• It provides continuous output while you conduct tests or change the
logic settings.
• It provides continuous output when the main screen is open or when a
third-party application is running.
• It switches off if F6ControlPanel is closed or if communication with
the F6150 is lost.
When the battery simulator is on, a red BT icon appears in the lower right
corner of the main screen.
F6ControlPanel supports two types of battery simulator options for the F6150:
• Fixed Battery Simulator (FBS), which has three voltage levels: 48,
125, and 250 volts DC. If Fixed has been selected on the Setup dialog
box, a voltage selector button will appear on the main screen. Click the
button to choose the desired voltage level.
• Variable Battery Simulator (VOBS), which can output between 6 and
300 volts DC in 1-volt increments. If Variable Battery has been
checked in the Setup dialog box, a data field will appear on the main
screen. Enter the desired voltage level. If VOBS is installed in the
F6150, you will see “6875” scrolling in the instrument’s option
display.
The Fixed and Variable battery simulator settings are relevant only in
simulation mode. When connected to an instrument, F6ControlPanel will
automatically detect whether the unit is fitted with FBS or VOBS and will
display accordingly.
Timing Results
The timing results pane (Figure 3-51) shows timing results from selected
sources, logic settings, and parameters.
Use the Dynamic control panel to run a three-state dynamic test: pre-fault,
fault, and post-fault. The Dynamic device provides real-life fault simulation
and cycle-by-cycle control of the F6150. Logic outputs can be controlled on a
state-by-state basis, independent of assigned sources. The eight timers can be
started on the beginning of any of the three states and stopped by any of the
eight logic inputs.
Many of the features of the Dynamic control panel are the same as those of
the Generic control panel. To avoid repetition, only the differences in the
Dynamic device are discussed in this chapter.
Figure 4-1 shows the main screen of the Dynamic control panel.
Setup
Setup for the Dynamic control panel is the same as for the Generic control
panel (see Chapter 3, “Generic Control Panel.”).
Click the button for the desired state (Figure 4-2) to view the source table (and
phasor display) for that state.
Frequency, Range
Frequency and Range can be set or changed only in the pre-fault source table
(Figure 4-3).
Source Buttons
The On and Enable buttons and their associated functions are not available in
the Dynamic device.
The Dynamic device uses the source buttons shown in Figure 4-4.
Click Run to download the dynamic test to the instrument and perform the
test.
Ramp/Set Controls
In the Dynamic device, the Ramp/Set Sources section (Figure 4-5) is used
only to configure the Action meter (see “Action Meter” on page 3-42). Only
the Set mode is available.
Logic/State Settings
Figure 4-2 shows the control buttons for the logic I/O functions. Click the
Settings button to open the Logic Settings dialog box.
Inputs Tab
AutoSense is not available in the Dynamic device.
Outputs Tab
The Outputs tab contains sections for pre-fault, fault, and post-fault logic
states.
Timers Tab
Figure 4-6 shows the Timers tab. The Start Condition and Stop Condition
drop-down menus are not available in the Dynamic device.
Setting Timers
1. In the Set Timer column, select each required timer by its option button.
2. In the Set Timer Conditions pane, select the Start On State from the
drop-down menu.
3. Select the Stop On Trigger from the dropdown menu.
4. Click Set to update the selected timer’s start and stop conditions.
To clear the start and stop conditions for selected timers, click Reset.
Define Triggers
In the Dynamic device, only the trigger name can be modified. Each trigger is
set to its corresponding logic input. For example, Trigger 1 is set to Logic
Input 1 (In1).
Timers start at the beginning of the selected state and run until one of the
following occurs:
• The timer encounters the stop event. If the stop event occurs, the
elapsed time is displayed.
• The timer encounters the end of the test. If the end of the test occurs
before the stop event, NoOp (No Operation) appears in the results
field on the main screen.
State Settings
The State Settings tab (Figure 4-8) enables you to specify the conditions under
which a state ends and transitions to the next state or to the end of the test.
Device Settings
The Device Settings dialog box (Figure 4-9) enables you to enter information
for the summary reports that document the test results.
To enter device settings:
1. Select Device Settings to open the dialog box.
2. From the manufacturer’s documentation, enter the Manufacturer Name for
the device, the Model Number, and the Serial Number of the device. These
entries help identify the summary reports.
3. The general test data section is prefilled from the Setup dialog box and the
Max Voltage and Max Current data fields in the lower right corner of the
main screen, or you can enter different values:
• Nominal Voltage: Defines the beginning of the yellow zone in the
action meter. Voltages greater than this appear in the yellow zone (see
“Action Meter” on page 3-42).
• LL or LN from the dropdown menu.
• Nominal Current: Defines the beginning of the yellow zone in the
action meter. Currents greater than this appear in the yellow zone (see
“Action Meter” on page 3-42).
• Max Voltage: Defines the beginning of the red zone in the action
meter. Voltages greater than this appear in the red zone (see “Action
Meter” on page 3-42).
• Max Current: Defines the beginning of the red zone in the action
meter. Currents greater than this appear in the red zone (see “Action
Meter” on page 3-42).
• Voltage transformer ratio (VTR): Not available in this version.
• Current transformer ratio (CTR): Not available in this version.
Changing a value in the Max Voltage or Max Current field will change that
value at the bottom right of the main screen. When you close the Dynamic
control panel, however, the value reverts to the value in Setup the next time
the control panel is opened. Changing the Nominal Voltage or Nominal
Current will not change the value in Setup.
Function Graph
The Dynamic function graph is coordinated with the source table buttons
(Figure 4-2). The graph shows the waveforms of the phasors you select with
the Pre-Fault, Fault, or Post-Fault button. Figure 4-10 shows a pre-fault
dynamic function graph.
Click the “+” button in the upper right corner for a comprehensive view
(Figure 4-11) of pre-fault, fault, and post-fault information.
Select the desired viewing options in the Properties pane on the right:
• Display Type: Check the Analog or Digital checkboxes, or both.
• Superimposed: Check the Voltage or Current checkboxes, or both, to
superimpose the selected sources.
• Analog and Digital selector buttons: Click the buttons for the
desired sources.
• Apply: Applies the selected settings.
• Hide Properties: Hides the Properties pane. To display the pane
again, click the “–” button in the upper right corner, then click “+” to
reopen the display, or right-click in the graph (when maximized) and
select Show/Hide Properties.
Right-clicking in the graph provides additional functions:
When minimized:
• Maximize. Expands the graph to the comprehensive view
(Figure 4-11).
• Pan. Releases the graph from the axes, allowing you to drag it to any
position by left-clicking the mouse.
When maximized:
4. In the timing results pane, select the unit of measurement: ms, s, or cy.
5. Click Reset.
6. When all the timers, sources, and parameters have been set, click Run.
Run cannot be clicked while a value is displayed. You must clear all timers
by pressing Reset.
The results of the test are displayed in the appropriate fields.
Reports
Test results can be saved for later viewing or printing as reports.
Viewing
To view test results, select Summary Test Results. The Test Results
window opens. The window has two tabs: Test Parameters (Figure 4-14) and
Test Results (Figure 4-15).
Report Configuration
To select the items that appear in the test results:
1. Select Summary Report Configuration.
The Report Configuration dialog box opens (Figure 4-16).
2. Make the appropriate selections and click OK.
Saving
To save test results:
1. Select Summary Test Results.
2. Click Save.
The Save As dialog box opens (Figure 4.17).
The default filename is prefilled in the File name field:
DynamicResult.xml. Selecting this filename will overwrite any other test
results with that name. To save more than one set of results, rename the
file.
3. Click Save.
If the filename already exists, a dialog box appears asking if you wish to
replace the existing file. Click Yes to replace the file or No to cancel.
Opening
To open saved test results files:
1. Select Summary Open Results.
The Open dialog box appears (Figure 4-18). This is the same as the File
Open dialog box, except that it shows only .xml files.
Figure 5-1 highlights the portion of F6ControlPanel used to test analog output
transducers.
The analog output portion of the Transducer F6Control Panel contains these
parts:
• Input/Output Meter
• Action Meter
• DC Meter Output
• Function graph and the F6150 DC meter input range to the right of the
graph.
By right-clicking on the graph, you can select which graph you want to
display. The following list is available for a Watt transducer:
% Full Scale
Watt
Watt vs. mAmp
% Error
Complex Power
• Selecting tolerance will add the specified tolerance lines to the graph.
Device Settings
Use the Device Settings window to enter the transducer device details. From
the menu bar, select Device Settings to open the window (Figure 5-2). The
settings in this window affect the test conditions applied in the Control Panel.
For a summary of the settings in the Device Settings window, see “Settings for
Analog Output Transducers” on page 5-13.
3. Click OK to close the Device Settings window and return to the Control
Panel.
4. Select the required transducer function for testing in the Transducer
drop-down list.
The entries in the Transducer drop-down list correspond to the transducer
function selected in the Device Settings window.
Test Type
The F6CP can perform two types of tests on transducer functions:
• Step Response
• Input/Output
Selecting Step Response changes the test to to display the transducer output
vs. time. With the sources turned on, selecting the up or down control arrow
causes the transducer’s output to be recorded. The recording starts when the
arrow is clicked and stops once the sum of the Response Time and the settling
time elapses.
The resulting transducer response is displayed on the function graph as output
verse time.
Selecting Input/Output changes the test to display the transducer output vs.
input. With the source(s) turned on, a timer starts each time an up or down
control arrow is clicked. After the sum of the Response Time and the settling
time elapse, the transducer’s output is recorded. The resulting transducer
response is displayed on the function graph as transducer input verse output
for each step.
To conduct a Step Response or Output/Input transducer test, select either test
from the Test Type drop-down menu.
Scale Setting
Right-click in the function graph and select Change Scale from the pop-up
menu to open the Scale Setting dialog box (Figure 5-6). The scale style for
both the Input and the Output is Auto by default. To set the minimum and
maximum values for each axis manually, click Manual in both the Input
section and Output section, enter the desired values in the Min and Max fields,
and click OK to close the dialog box and to implement the new settings.
Graph Setup
Right-click in the function graph and select Setup to open the Properties
dialog box (Figure 5-7). Use the dialog box to set the color of the actual data
points, the expected values, the two axes, the grid lines, and the background.
Then set the width and the line type for the actual and expected values. Click
OK to close the dialog box and to implement the new settings.
Zoom/Unzoom
The Zoom/Unzoom feature enlarges an area of the Input/Output graph. To use
the feature, right-click and select Zoom/Unzoom in the pop-up menu or click
the the plus sign (+) button to the right of the function graph. The pop-up
menu closes with the Zoom feature in effect.
To select an area, click the left mouse button in the diagram and drag the
cursor down and to the right. A dashed rectangle appears around the selected
area. When the mouse button is released, the selected area is enlarged.
To deselect the Zoom feature, right-click in the Input/Output graph and
uncheck Zoom/Unzoom in the pop-up menu. When the pop-up menu closes,
the Input/Output graph returns to its normal state.
To expand the Output Response graph, click the plus sign (+) button to the
right of the function graph (Figure 5-9).
Meters
Three meters are provided on the F6Control Panel (Figure 5-10) when the
Transducer control panel is selected.
The range and settings of the meters are determined by the device settings,
including the input/output range and the device nominal and max settings.
There are three combo-boxes (located under the meters) common to all three
meters. These are:
• Output ID — Specifies the transducer output selected for the test if
multiple device outputs are configured.
• Values are — Indicate whether the values are shown in Secondary or
Primary quantities.
• Mode — Choices from this box are Manual and Auto. Select Manual
to configure parameters to your requirements. Select Auto to
configure the control panel parameters automatically based on the
specified device parameters.
Input/Output Meter
The Input/Output meter uses a red needle to indicate the transducer input and a
purple needle to indicate the transducer output (Figure 5-11).
The indications in the meter correspond to the values in the Input and
Output fields below the meter.
Select Manual, enter the minimum and maximum values for the
meter, and click OK. The new settings appear in the Input/Output
meter.
Action Meter
The Action meter (Figure 5-13) indicates the output of the selected action
source and the actual value is displayed in the field below the meter.
Use the drop-down menu directly below the Action meter to select the active
source. The sources in the list correspond to the sources checked in the
Ramp/Set sources section. If a voltage source is selected, the meter readout is
in Volts for Amplitude mode, Degrees for Phase mode, and Hertz for
Frequency. If a current source is selected, the meter readout is in Amps for
Amplitude mode, Degrees for Phase mode and Hertz for Frequency.
When displaying Volt or Amps, values above the green bar are at or less than
the Nominal value entered in Device Settings. Values above the yellow bar are
greater than the Nominal value but less than the Maximum value. Values in the
red bar are greater than the Maximum value. The user is warned on entering
the red zone.
To change the scale of the Action meter:
1. Right-click in the meter. The following menu appears.
DC Output Meter
The DC Output meter (Figure 5-16) indicates the transducer output measured
by the F6150 DC Meter Input; the actual value is displayed in the field below
the meter.
Mode
Two modes of operation are provided:
• Manual — Enables the user to manually configure the source table
and ramp functions, as required.
• Auto — Automatically configures the source table and ramp functions
according to the specified device settings.
To conduct a transducer test in Auto mode, select Auto from the Mode
drop-down menu. The confirmation box in Figure 5-18 appears in the Control
Panel. Click OK to close the warning box and reconfigure the Control Panel
for Auto mode.
• Manual Mode — The required ramp step size can be entered by the
user as either an amplitude, phase, or frequency quantity. The step size
is entered using the Rate or Delta Step field on the Control Panel.
Clicking the up and down arrow buttons causes the source amplitude,
phase, or frequency quantity to increase or decrease for the sources
selected in the Ramp/Set Sources check boxes according to the
required change.
• Auto Mode — The step size is automatically calculated based on the
transducer function Auto Settings in Device Settings. Clicking the up
and down arrow buttons causes the source amplitude, phase, or
frequency quantity to increase or decrease according to the parameters
specified in the Auto Settings of the Device Settings.
In Auto mode, the auto step size appears in the Step Size field of the
Control Panel. In Manual mode, the auto step size display is not
applicable and N/A appears.
Nameplate Data
Use the device nameplate details to enter the name of the manufacturer, the
model number, and the serial number of the transducer (Figure 5-20). These
entries help to identify the summary reports that document the test results.
From the device nameplate and specification enter the relevant device rating
details:
• Nominal Voltage – the rated voltage selected as either line to neutral
or line to line from the drop down box selection.
• Nominal Current – the rated current in amps.
• Maximum Voltage – the maximum specified continuous line to
neutral voltage.
• Maximum Current – the maximum specified continuous current.
• Maximum Power – This field is automatically calculated for Watt,
VAR, and VA transducers. All others are equal to zero.
If testing in primary values is preferred, enter the relevant voltage and current
transformer ratios:
• Voltage transformer ratio (VTR) – enter the ratio of primary to
secondary voltages.
• Current transformer ratio (CTR) – enter the ratio of primary to
secondary currents.
• Calibration Quantity Over Range — enter a specified device over
range quantity; a warning is displayed if this value is exceeded.
• Frequency
• Power Factor
• AC Phase
A separate device tab is provided to add the individual function parameters.
The Watt tab (Figure 5-21) is used as a example.
Auto Settings
The fields and possible setting values for a transducer being tested in
Automatic mode are described below. The control panel source table will be
updated automatically, according to the specified parameters.
• Action — Select the action source type: Voltage or Current, to be
varied during auto mode testing for the selected transducer function.
For transducer functions with both Current and Voltage selections, the
amplitude, phase angle, and frequency of the non-action sources are
fixed during Auto mode testing to the nominal device rating. The
choices vary based upon the transducer function. For example, a watt
transducer has a Voltage and a Current selection; a voltage transducer
only has a Voltage selection.
• Parameter — Select the source parameter: Amplitude, Phase or
Frequency, to be varied during Auto mode testing for the selected
transducer function. The choices vary based upon the transducer
function. For example, a watt transducer has an Amplitude and Phase
parameter selection; a frequency transducer only has a frequency
selection.
• VI Phase Angle — For Power and Energy transducers enter the phase
angle for the required power factor during Auto mode testing.
Entering a VI Phase Angle value causes the Power Factor field to
automatically update relative to the entered phase angle.
• Power Factor — For Power and Energy transducers enter the power
factor required during Auto mode testing. Power factor may be
specified as lagging or leading values as required. Entering a Power
Factor value causes the VI Phase Angle to automatically update
relative to the entered power factor.
• Step Size — Enter the value of the required test quantity increments
for each step during Auto mode testing. The step quantity changes
based upon the transducer function. For a Watt transducer, the step
quantity is Watts. For a voltage transducer, the step quantity is volts.
The entries in the Source Configuration show how the source table in the
Control Panel is configured when Auto mode is selected. Sources and the
values are calculated automatically based on transducer device settings.
The following bulleted list describes the Source Configuration options and
details.
• Source — The Source column displays the F6150 source names
configured during Auto mode testing. The default source names are
defined in the Source Name Scheme in the Source Parameters setup
menu. To change the name of a source, click the source name and
select the new name from the drop-down list that appears.
• Amplitude — The Amplitude column displays the F6150 source
amplitudes.
The source's amplitude is set to the Nominal value if the source is
not an Action quantity or Amplitude is not the action parameter.
The source's amplitude is calculated according to the step size
value if the source is an Action quantity and Amplitude is the
action parameter.
• Phase — The Phase column displays the F6150 source phase values.
The default phase values are defined in the Source in the Source
Parameters Setup menu.
The source's phase is set to the default value if the source is not an
Action quantity or Phase is not the action parameter.
The source's phase is calculated according to the step size value if
the source is an Action quantity and Phase is the action parameter.
Input Fields
The F6150 Input section contains fields for selecting the required DC Meter
Input (mAmps or Volts) and setting the input range. Table 5-1 lists the
available settings for the DC meter input voltage and current ranges.
Units Ranges
mAmps 25 ... + 25
10 ... + 10
1 ... + 1
Volts 12.5 ... + 12.5
5.0 ... + 5.0
0.5 ... + 0.5
Characteristics
The Characteristics section contains fields for defining the transducer's
characteristic for the selected function.
Fields and possible values for the transducer's output characteristic settings are
as follows:
• Type — Watt, Var, Volt-Amp, Voltage, Current and Frequency
transducer functions support Linear, Compound and Quadratic
characteristic types. The Power Factor transducer function supports
Linear, Compound and Cosine characteristic types. The AC Phase
transducer function supports Linear and Compound characteristic
types.
Transducer characteristics can be Unidirectional (starting or offset
from the origin) or Bidirectional (symmetrical with the origin).
• Rated Output — Enter the Rated Output for the transducer function.
The Rated Output is entered by first selecting the output units (mAmps
or Volts). Then select the output range. Changing the units alters the
F6150 Input units and range to match the selection. Refer to the
following table for details on selecting the Rated Output range.
Unidirectional Bidirectional
mAmps mAmps
0 1 -1 0 1
1 5 1 5
4 20 4 20
Custom (Select Custom to enter a custom Custom (Select Custom to enter a custom
range) range)
Volts Volts
01 –1 0 1
1 10 1 5.5 10
Custom (Select Custom to enter a custom Custom (Select Custom to enter a custom
range) range)
• Min Value (Input) — Enter the minimum test value of the transducer
input for the required test.
• Min Value (Output) — This field displays the minimum test value of
the transducer output calculated from the transducer characteristic
type and rated output.
• Max Value (Input) — Enter the maximum test value of the transducer
input for the required test.
• Max Value (Output) — This field displays the maximum test value
of the transducer output calculated from the transducer characteristic
type and rated output.
The Cal Qty/Element is calculated based on the values entered for the Input
Min Value and Input Max Value.
Characteristic Graph
The device setting characteristic graph displays the selected device
characteristic as input verses output based on the entered device parameters.
Items in the right-click menu for the Characteristic Graph are shown below:
• Grid — Displays grid lines parallel to the X-Axis, Y-Axis, or both.
• Properties — Opens the Properties dialog box to select colors.
• Zoom/Unzoom — Enables or disables the Zoom feature.
Watt Tab
The Watt tab contains the settings used to test a watt transducer in Auto mode
(Figure 5-26). Enter the device details and parameters from the nameplate or
manufacturer's specification. Select the Auto settings required to configure the
control panel source table for the test to be performed.
Values for the Auto settings in the Watt tab are shown below:
• Action — Current or Voltage, choose from drop-down menu.
• Parameter — Amplitude or Phase, choose from drop-down menu.
• VI Phase Angle — 0.00°... ±360°
• Power Factor — 0.000 (Lag) ... ±1.000
• Step Size — Watts
Var Tab
The Var tab contains the settings used to test a Var transducer in Auto mode
(Figure 5-27). Enter the device details and parameters from the nameplate or
manufacturer's specification. Select the Auto settings required to configure the
control panel source table for the test to be performed.
Features and values for the Auto settings in the Var tab are shown below:
• Action — Current or Voltage. Choose from drop-down menu.
• Parameter — Amplitude or Phase. Choose from drop-down menu.
• VI Phase Angle — 0.00°... ±360°.
• Power Factor — 0.000 ... ±1.000
• Step Size — Var
Volt-Amp Tab
The Volt-Amp tab contains the settings used to test a Volt-Amp transducer in
Auto mode (Figure 5-28). Enter the device details and parameters from the
nameplate or manufacturer's specification. Select the Auto settings required to
configure the control panel source table for the test to be performed.
Features and values for the Auto settings in the Volt-Amp tab are shown
below:
• Action — Current or Voltage. Choose from drop-down menu.
• Parameter — Amplitude or Phase. Choose from drop-down menu.
• VI Phase Angle — 0.00 ... ±360°.
• Power Factor — 0.000 ... ±1.000
• Step Size — VA (Volt-Amps)
Voltage Tab
The Voltage tab contains the settings used to test a Volt-Amp transducer in
Auto mode (Figure 5-29). Enter the device details and parameters from the
nameplate or manufacturer's specification. Select the Auto settings required to
configure the control panel source table for the test to be performed.
Features and values for the Auto settings in the Voltage tab are shown below:
• Action — Voltage
• Parameter — Amplitude
• VI Phase Angle — Disabled
• Power Factor — Disabled
• Step Size — Volts
Current Tab
The Current tab contains the settings used to test a current transducer in Auto
mode (Figure 5-30). Enter the device details and parameters from the
nameplate or manufacturer's specification. Select the Auto settings required to
configure the control panel source table for the test to be performed.
Features and values for the Auto settings in the Current tab are shown below
• Action — Current
• Parameter — Amplitude
• VI Phase Angle — Disabled
• Power Factor — Disabled
• Step Size — Amps
Frequency Tab
The Frequency tab contains the settings used to test a frequency transducer in
Auto mode. (Figure 5-31). Enter the device details and parameters from the
nameplate or manufacturer's specification. Select the Auto settings required to
configure the control panel source table for the test to be performed.
Features and values for the Auto settings in the Frequency tab are shown
below:
• Action — Current or Voltage, choose from drop-down menu.
• Parameter — Frequency
• VI Phase Angle — Disabled
• Power Factor — Disabled
• Step Size — Hz
Features and values for Auto settings in the Power Factor tab are shown
below:
• Action Quantity — Current or Voltage, choose from drop-down
menu.
• Parameter — Phase
• VI Phase Angle — 0.00°... ±360°
• Power Factor — 0.000 ... ±1.000
• Step Size — Degrees
AC Phase Tab
The AC Phase tab contains the settings used to test an AC phase transducer in
Auto mode (Figure 5-33). Enter the device details and parameters from the
nameplate or manufacturer's specification. Select the Auto settings required to
configure the control panel source table for the test to be performed.
Features and values for the Auto settings in the AC Phase tab are shown
below:
• Action — Current or Voltage (select from drop-down menu)
• Parameter — Phase
• VI Phase Angle — 0.00°... ±360°
• Power Factor — 0.000 ... ±1.000
• Step Size — Degrees
The digital output portion of the F6Control Panel contains these parts:
• Complex energy function graph, with information on:
P – active power
Q – reactive power
S – complex power
• Input/Output meter.
• Digital recorder, with readouts for pulses counted and counts per hour
or energy/pulse or pulse/energy.
• Selection lists for the transducer output ID, the value type, and the test
mode.
Use the Device Settings window to set up a transducer test. Select Device
Settings to open the window (Figure 5-35).
The settings in this window affect the test conditions applied in the F6Control
Panel.
See “Device Settings for Digital Output Transducers” on page 5-37.
To configure the F6 Control Panel for testing a selected type of transducer:
1. Select Device Settings from the Control Panel’s menu bar.
The Device Settings window opens.
2. Click Add Function in upper left corner of the window. A drop-down
menu listing 11 types of transducers appears (Figure 5-36).
3. Click OK to close the Device Settings window and return to the Control
Panel.
4. Select the required transducer function for testing in the Transducer
drop-down list.
The entries in the Transducer drop-down list correspond to the transducer
function selected in the Device Settings window.
The right-click menu items for the complex power graph are shown below:
• Setup — Opens the Properties dialog box.
• Zoom/Unzoom — Enables or disables the Zoom feature.
• Phase — Opens a submenu for a selection containing these options:
System
A Phase
B Phase
C Phase
Setup
Select Setup in the right-click pop-up menu to open the Properties SetUp
dialog box (Figure 5-39). Use the box to select colors for each quantity in the
complex energy graph, as well as the axes, the circle, and the background.
Also use the dialog box to set the line width and line types for all elements of
the graph. Click OK to close the dialog box.
Zoom/Unzoom
The Zoom/Unzoom feature enlarges an area of the complex power graph. To
use the feature, right-click and select Zoom/Unzoom from the pop-up menu.
The pop-up menu closes with the Zoom feature in effect.
To select an area, click the left mouse button in the diagram and drag the
cursor down and to the right. A dashed rectangle appears around the selected
area. When the mouse button is released, the selected area is enlarged.
To deselect the Zoom feature, right-click in the complex power graph and
select Zoom/Unzoom from the pop-up menu. When the pop-up menu closes,
the complex power graph returns to its normal state.
Phase
Select Phase to display the calculated Active Power, Reactive Power, and
Complex Power for:
• System
• Phase A
• Phase B
• Phase C
Input/Output Meter
The Input/Output meter (Figure 5-40) uses a red needle to indicate the
transducer input and a purple needle to indicate the transducer output.
Right-click in the Input/Output meter and select Change Scale to open the
Scale Setting dialog box (Figure 5-41). Select Manual and enter the minimum
and maximum values for the meter scale. Click OK to close the dialog box.
Output ID Field
Selects which transducer outputs is displayed in the Input/Output meter's
output quantity and the Output field. The number of outputs, and the names of
the outputs in the selection list, are set in the Device Settings window.
Mode
Two modes of operation are provided:
• Manual — Enables the user to manually configure the source table
and ramp functions, as required.
• Auto — Automatically configures the source table and ramp functions
according to the specified device settings.
Figure 5-43 shows the Device Settings window when the function is selected a
Watthour transducer.
Nameplate Data
Use the device nameplate details to enter the name of the manufacturer, as
well as the model number and serial number of the transducer (Figure 5-44).
These entries identify the summary reports that document the test results.
Enter the general test data. The fields in this section of the Device Settings
window are the same as those discussed in “Nameplate Data” on page 5-13.
Watthour Tab
The Watthour tab contains the settings used to test a Watthour transducer in
Auto mode (Figure 5-45). Enter the device details and parameters from the
nameplate or manufacturer's specification. Select the Auto settings required to
configure the control panel source table for the test to be performed.
Features and values for the Auto settings in the Watthour tab are:
• Action Quantity — Current or Voltage,choose from the drop-down
menu.
• VI Phase Angle — 0.00°... ±360°
• Power Factor — 0.000 ... ±1.000
Varhour Tab
The Varhour tab contains the settings used to test a Varhour transducer in Auto
mode (Figure 5-46). Enter the device details and parameters from the
nameplate or manufacturer's specification. Select the Auto settings required to
configure the control panel source table for the test to be performed.
Features and values for the Auto settings in the Varhour tab are:
• Action Quantity — Current or Voltage. Choose from the drop-down
menu.
• VI Phase Angle — 0.00°... ±360°
• Power Factor — 0.000 ... ±1.000
Volt-Amphour Tab
The Volt-Amphour tab contains the settings used to test a Volt-Amphour
transducer in Auto mode (Figure 5-47). Enter the device details and
parameters from the nameplate or manufacturer's specification. Select the
Auto settings required to configure the control panel source table for the test to
be performed.
Features and values for the Auto settings in the Volt-Amphour tab are shown
below:
• Action Quantity — Current or Voltage. Choose from the drop-down
menu.
• VI Phase Angle — 0.00°... ±360°
• Power Factor — 0.000 ... ±1.000
Auto Settings
Features of the Automatic mode settings are shown below. The control panel
source table will be updated automatically according to the specified
parameters.
• Action — Select the action source type: Voltage or Current, to be
calculated to provide the required energy during auto mode testing for
selected transducer function.
• VI Phase Angle — Enter the phase angle for the required power
factor during Auto mode testing. Entering a VI Phase Angle value
causes the Power Factor field to automatically update relative to the
entered phase angle.
• Power Factor — Enter the power factor required during Auto mode
testing. Power factor may be specified as lagging or leading values as
required. Entering a Power Factor value causes the VI Phase Angle to
automatically update relative to the entered power factor.
• Test Type — Select the Auto Test type: Meter Constant (Full Load),
No Load, Creep or User Defined.
• Test Type % — Enter the percentage of the test quantity for the Auto
Test. For a Meter Constant (Full Load) test the Test Type % is fixed to
100%. For a No Load test the Test Type % is fixed to 0%.
• Sys. Cal. Qty. — The System Calibration Quantity equals the
Calibration Quantity/Element multiplied by No. of Elements
multiplied by Test Type %.
• Max Test Time — Represents the calculated time for the sum for the
Pulse to Start and Pulse For Test to occur with the specified system
calibration quantity.
• Max On Time — This value can be user entered and defaults to 125%
of the Max. Test Time. In auto test mode, all sources are turned off
when all counters have counted the specified number of pulses or
when the maximum on time is reached.
Click Source Configuration in the Auto Settings section to open the Source
Configuration window (Figure 5-48).
The entries in the Source Configuration show how the source table in the
Control Panel is configured when Auto mode is selected. For a description of
these settings, refer to “Auto Settings” on page 5-17.
Input Mask
The rows of the Input Mask correspond to a transducer output. The number of
rows equals the No. Of Outputs selection. The columns are numbered 1
through 8 and correspond to logic inputs 1 through 8 on the F6150. Individual
check boxes are provided to assign each transducer output to an F6150 logic
input.
If the outputs are defined as unipolar, check one input in each row. If the
outputs are bipolar, check two inputs on each row. When the required number
of inputs for an output have been checked, the background of the rest of the
inputs in that row turns dark gray.
Figure 5-49 shows a typical setup for three bipolar outputs.
Click the button to the left of the drop-down menu to toggle the battery
simulator on and off. The battery simulator has the following operating
characteristics:
• When the battery simulator is on, it provides continuous output while
the user conducts tests or changes the logic settings.
• The battery simulator provides continuous output when either the
Control Panel or the Configuration display is open, or when a third
party application is running.
• When the battery simulator is on, a red BT icon appears in the lower
right corner of the F6Control Panel.
• The battery simulator switches off if the F6Control Panel is closed or
if communication with the F6150 Instrument is lost.
Reports
Test results can be saved for later viewing or printing as reports.
Viewing
To view test results, select Summary Test Results. The Test Results
window opens. The window has two tabs: Test Parameters (Figure 5-54) and
Test Results (Figure 5-55).
Report Configuration
To select the items that appear in the test results:
1. Select Summary Report Configuration.
The Report Configuration dialog box opens (Figure 5-56).
Saving
To save test results:
1. Select Summary Test Results.
2. Click Save.
The Save As dialog box opens (Figure 5-57).
The default filename — TransducerResult.xml — is prefilled in the
File name field.
Selecting this filename will overwrite any other test results with that name.
To save more than one set of results, rename the file.
3. Click Save.
If the filename already exists, a dialog box appears asking if you want to
replace the existing file. Click Yes to replace the file or No to cancel.
Opening
To Open saved test results files:
1. Select Summary Open Results.
The Open dialog box appears (Figure 5-58). This is the same as the
File Open dialog box, except that it shows only .xml files.
To Ethernet Port
on F6150
To Same Setup on
PC’s Ethernet Port
50-ohm Terminator
CPU2 Instruments
If connecting to the F6150 directly from a PC, plug the crossover adapter
(401-0303) supplied with the instrument into the Ethernet connector on the
F6150 and then use the supplied standard Ethernet cable between the adapter
and the PC.
If connecting to the F6150 through a hub, use the standard Ethernet cable
supplied with the instrument between the hub and the instrument. The
crossover adapter is not necessary.
1. Right-click the Network icon on the desktop (or go to the Start menu,
select My Network Places, and right-click) and select Properties
(Figure B-2).
3. In the General tab, scroll down the list of network components and select
the appropriate component:
• For CPU2-based instruments: Internet Protocol (TCP/IP)
• For CPU1-based instruments: the Ethernet card for the control PC
4. Click Properties underneath the list of network components.
The TCP/IP Properties dialog box appears (Figure B-4). If there is data in
any of these fields, copy it down in case you wish to restore it later.
5. In the General tab, click the option button for Use the Following IP
Address (Figure B-5).
6. In the IP Address field, enter an IP address in the same subnet as the IP
address displayed on the F6150 when it is powered on.
7. Enter the Subnet Mask.
8. Click OK, then click OK again.
9. For systems using Windows 95 or Windows 98, restart the computer for
these changes to take effect. For systems using Windows 2000 or
Windows XP, you do not need to restart.
Figure B-5 IP Address and Subnet Mask in TCP/IP Properties Dialog Box
2. Click Get.
• If you are using a serial connection (the default), the utility obtains the
IP address of the F6150 via the specified COM port. The IP address
and subnet mask appear in the top two fields.
• If you are using an Ethernet connection, the utility attempts to connect
to the IP address specified in the Communications field of the Set
F6000 IP Address dialog box. If it is successful, the IP address and
subnet mask of the F6150 appear in the top two fields.
The Flash Loader and Key Code Update utilities are not accessible when a
control panel is open. To access them, click the small gray X in the upper
right corner of the screen. This will close the control panel without closing
the F6ControlPanel application. Then select Tools and the desired
application.
When the F6150 powers up, the current firmware revision number and the
options installed appear in the display on the instrument’s front panel.
Flash Loader
Operation of the F6150 requires the correct firmware version:
• 2.11 (or higher) for CPU1 instruments
• 4.11 (or higher) for CPU2 instruments
The F6150 requires a firmware package file, *.pkg. This file must contain
F6150 in its name.
Flash Loader installs revised firmware in the F6150. Select Tools F6000
Flash Loader to open the program. (The control panels must be closed to do
this.) The Flash Loader dialog box appears (Figure A-1).
Table A-1 lists and describes the fields and buttons in the Flash Loader dialog
box.
Field/Button Description
Check the Marketing Release Notice to ensure that the firmware enables
all required options and enhancements and is compatible with the
installed version of F6ControlPanel.
If the flashload fails, see the next sections for procedures to attempt to
remedy the problem.
Communications Parameters
If the revised firmware does not load successfully, confirm that you have
communication between the computer and the F6150.
To verify or change the communication settings:
1. Click Change in the Flash Loader dialog box (or any of the other Tools
menu utilities with a Change button).
The Set Communications Parameters dialog box appears (Figure A-2).
b. Click Yes.
You are returned to the initial dialog box.
4. Click OK.
5. Click Get.
• If you are using a serial connection, the utility obtains the IP address
of the F6150 via the specified COM port. The IP address and subnet
mask appear in the top two fields.
• If you are using an Ethernet connection, the utility attempts to connect
to the IP address specified in the Communications field of the Set
F6000 IP Address dialog box. If it is successful, the IP address and
subnet mask of the F6150 appear in the top two fields.
If the utility is not successful, you will receive an error message.
Make sure you have an Ethernet cable connected between the Ethernet
connection on your PC and the Ethernet (Network) connection on the
F6150.
6. To change the IP address and subnet mask, enter the desired settings in the
top two fields.
7. Enter an IP address in the same subnet as the desired IP address.
8. Click Set.
9. To enable the new IP address and firmware, turn the unit off and then back
on.
These procedures to work only with a serial connection; do not try them
with an Ethernet connection.
CPU1 Instruments
1. Connect the serial cable.
2. Run Flash Loader and reload the firmware.
3. Turn the instrument off and then back on.
CPU2 Instruments
1. Turn off the instrument.
2. Connect the serial cable.
3. Open HyperTerminal on your computer.
4. Enter these parameters:
• Bits per second: 57,600
• Data bits: 8
• Parity: None
• Stop bits: 1
• Flow Control: None
5. Turn on the instrument.
A series of messages appears. The series ends with a prompt.
6. At the prompt, type f6xs and press Enter.
7. Shut down HyperTerminal or use the disconnect on the toolbar.
8. Run Flash Loader and reload the firmware.
If the bootloader is defective, the CPU card or the entire unit must be
returned to Doble.
The Key Code utility is required only when options are to be installed in
the field. All options ordered with the unit are installed before delivery.
a. Click Yes.
A message appears, indicating a successful update (Figure A-7).
b. Click OK.
a. Click OK
b. Return to step 2 on page A-7.
4. If necessary, click Change to enter new communications parameters. (See
“Communications Parameters” on page A-4 for the procedure.)
The Vector Calculator consists of a vector table and keypad. Vectors are
displayed graphically in the vector (or phasor) view area. A vector formula
you enter with the keypad is displayed in the Formula Box. If the formula is
evaluated, the result is shown in the Answer Box.
Vector Table
The vector table contains five columns:
• The Vector column contains data entry fields for vectors V1 through
V9.
• The second and third columns define the vector values. The headings
for these columns depend on whether you select Polar or Rectangular
form on the keypad. For Polar form, the columns are labeled
Magnitude and Angle; for Rectangular form, the columns are labeled
Real and Imaginary.
• The Graph column contains buttons that enable you to select whether
each vector appears in Normal View. If a vector entry has been saved
from a formula created in the Formula Box, the formula appears in the
Formula column of the vector table.
• The Source Name column enables you to add a source name to relate
the vector to a power system source (VA, I1, etc.)
Keypad
The Keypad consists of the following key groups, each of which is described
below:
• Function keys
• Vector keys
• Operator keys
• Number keys
• Recalculate Formulae
Function Keys
The following function keys appear in the first two rows of the keypad.
Key Description
Polar and Rect These keys switch between Polar and Rectangular
form for all vector values, including the Answer
Box. This allows values to be entered in one form
and redisplayed in the other form.
Back (Backspace) Deletes the preceding value in the cell being edited.
Same as Backspace on a keyboard.
CFB (Clear Formula Clears the formula in the Formula Box and the result
Box) in the Answer Box.
Key Description
SV (Save Vector) The next vector selected by a vector key (see below)
will be overwritten by the value in the Answer Box
and the corresponding formula will also be saved in
the table.
CV (Clear Vector) The next vector selected by a vector key will be
cleared.
CAV (Clear All Clears all values in the vector table.
Vectors)
Vector Keys Click keys V1 to V9 to specify a vector table entry.
These are used in two ways:
• When creating a formula in the Formula Box,
clicking a vector key inserts a vector name into
the formula, as an alternative to typing the vector
name.
• Clicking SV and then a vector key saves the
Answer Box result to that vector. Clicking CV
and then a vector key clears the contents of that
vector.
Operator Keys
Click an operator key to enter it in the Formula Box. You can enter values
from the keyboard or the keypad. Table C-2 describes the operator keys.
Table C-2 Operator Keys
Keys Description
Keys Description
= Evaluate the formula in the Formula Box and place in the Answer Box. The
Answer is shown in either the polar or rectangular form, depending on whether the
Polar or Rect function key is selected.
SQRT (square Enter a number whose square root is to be calculated. The function SQRT (nnnn)
root) appears in the Formula Box.
CONJ Changes the sign of an angle from positive to negative or vice versa. Click CONJ
(conjugate) to enter the operator and opening parenthesis before an expression. Then insert the
closing parenthesis at the end of the expression.
a Enters “a*” operator into the Formula Box at the insertion point, to rotate the
following vector 120°. If the following vector is entered from the keyboard as a
polar or rectangular value, it should be enclosed in parentheses:
a*(10@0)
which evaluates to 10@120.
a2 Enters “a2*” operator into the Formula Box at the insertion point, to rotate the
following vector value 240°. Note that “a*a2*” rotates the vector 360°.
Number Keys
Table C-3 describes the number keys.
Keys Description
0-9 Use to enter values in the Formula Box. You can also enter values from the
keyboard. Negative values are also valid (do not space on either side of the -).
Viewing Vectors
Two graphic displays show the vector table and Formula Box as vectors:
• Normal View. If the button in the Graph column displays a red
checkmark, the vector is graphed on a circular plot. Vector tails begin
at the origin.
• Formula View. If a vector formula in the Formula Box has been
evaluated and a result is displayed in the Answer Box, the vector
operations are graphed, showing the resultant vector in red. Vector
tails begin at the origin.
To evaluate a formula, press the equal sign (=) on the keypad.
Menu Bar
Menu bar options are:
• File
• Edit
• View
• Formulae
• Help
Each of the options is discussed below.
File
File Menu (Figure C-2) items are:
Edit
Edit menu (Figure C-4) items are as follows. They apply to highlighted data,
either cells in the vector table or data in the Formula Box.
• Cut — Cuts the selected data and retains it on the clipboard. Shortcut
key: Ctrl + X
• Copy — Copies the selected data and retains it on the clipboard.
Shortcut key: Ctrl + C
• Paste — Inserts data from the clipboard. You can do multiple pastes
from one cut or copy. Shortcut key: Ctrl + V
• Delete — Deletes the selected data.
View
View menu (Figure C-5) items are:
For example, in Figure C-6, with Scaling off, all vectors are scaled to the
largest magnitude (20), with the others proportionately smaller. In Figure C-7,
with Scaling on, voltage vectors are scaled to the largest magnitude (20), and
current vectors are shown to the same scale, even though the largest current
vector is only 15.
1. Click just outside the area. Holding down the left mouse button,
drag diagonally downward to create a dashed box outlining the
area (Figure C-8).
Alternate method:
1. Select View Zoom In Normal View or Formula View.
A red cross appears at the origin of the graph.
2. Outline the area to be expanded:
Formulae
Formulae menu (Figure C-10) items are:
To remove a formula:
1. Place the cursor in the line containing the formula or comment.
2. Click Remove.
Any entries below the deleted formula move up one row.
3. Click OK.
To insert a formula in the Formula Box:
1. Place the cursor in the line containing the formula or comment.
2. Click Export.
The formula appears in the Formula Box.
3. Click OK.
The Formulae menu also contains a list of the five most recently used
formulae, which you can enter by selecting the desired formula.
Help
Help menu (Figure C-12) items are:
Toolbar
A vertical toolbar appears at the upper left of the Vector Calculator screen.
Figure C.13 shows the toolbar with icon labels.
Cut
Copy
Paste
Delete
Help
Exit
Performing Calculations
To enter a vector value in the vector table:
1. Select Polar on the function keypad to enter magnitude and angle or select
Rect to enter real and imaginary values.
2. Click in the desired cell of the vector table.
3. Enter the value, using the Vector Calculator keypad or the keyboard.
4. Press Enter to complete the entry or use the mouse to select another cell.
5. Click the Graph button to view the vector in Normal View.
To clear a value in the vector table, place the cursor in that cell and click CV.
To clear all values, click CAV.
Formula Column
The Formula column contains read-only data fields, showing a vector formula
from the Formula Box that was used to evaluate the vector in the vector table.
For example, if “VA-VB” is entered in the Formula Box and evaluated by
clicking the equal sign, the value and the formula can be saved as a vector in
the table. To do this, click SV on the keypad, followed by a vector key — for
example, V7.
Formula Operators
Use formula operators as follows:
• Standard Operators — Enter the addition (+), subtraction (-),
multiplication (*), and division (/) operators from either the keypad or
the keyboard.
Use parentheses around vector values entered in polar or rectangular
form:
a*(-33.5-j58.02).
• a, a2 — Click a on the keypad as a multiplier to rotate the following
vector 120°:
a*V1
A
AC Phase tab 5-29 Digital Output Transducer tabs
Action meter 5-9 measurement units 5-41
Analog Output Transducer tabs 5-14 Digital Output Transducers
Auto settings 5-17 device settings 5-37
device function settings 5-15 Nameplate data 5-38
Analog Output Transducers Varhour tab 5-39
Nameplate data 5-13 Volt-Amphour tab 5-40
settings 5-13 Watthour tab 5-39
Auto mode 5-12
Auto settings 5-17
G
C GENERIC Virtual Device
changing parameters 3-23
Characteristic graph control arrows 3-24
using 5-21 Graph Setup 5-5
Complex Power Graph 5-33
Current tab 5-26
I
D Input/Output function graph 5-4
Input/Output graph
DC Output meter 5-11 Graph Setup 5-5
default settings 5-41 Scale Setting 5-4
Device settings 5-2 Zoom/Unzoom 5-5
Digital 5-36 Input/Output meter 5-8
Digital and Energy Recorder
Values Are field 5-37
Digital Output Transducer M
Complex Power Graph 5-33 Manual mode 5-12
Digital and Energy Recorder 5-36 Meters 5-7
Input/Output Meter 5-35 Input/Output 5-8
Mode 5-37 Mode 5-12
Phase 5-35
Setup 5-34
Zoom/Unzoom 5-35
N W
Nameplate data 5-13 Watt tab 5-22
Q Z
QuickTest virtual device Zoom/Unzoom 5-5
setup components 2-6
S
Scale Factor
calculating 5-16
Scale factor calculations 5-16
Scale Setting 5-4
Setup menu
Battery Off on Abort 3-17
Communication 3-13
Database File 3-12
Source Parameters 3-15
Switch Default Position 3-14
Test Defaults 3-17
Source Configuration settings
viewing 5-18
Step Response Function graph 5-6
Summary of device settings
auto settings 5-42
input mask 5-44
T
Test Type 5-3
Transducer device
Analog Output 5-1
TRANSDUCER Virtual Device
meters 5-7
test type 5-3
V
Var tab 5-23
Voltage tab 5-25
Volt-Amp tab 5-24