Professional Documents
Culture Documents
UNISTERI HP - NP - en 2018-09 - V4.00
UNISTERI HP - NP - en 2018-09 - V4.00
Steam sterilizer
The Instructions for use is protected by copyright law, the owner of the copyright being the
company BMT Medical Technology s.r.o. The Instructions for use is designed for personal use
only and it is not allowed to make it available to public in any way, as a whole or in parts. Breach
of the above stated regulations is considered to be a breach of the owner´s copyrights.
3 STEAM STERILIZATION.................................................................................................... 8
3.1 performance check..........................................................................................................................8
3.1.1 check by chemical process test.....................................................................................................8
3.1.2 check by chemical test of sterilization........................................................................................8
3.1.3 check by means of the bowie-dick test (bd-test).........................................................................8
3.1.4 check built in the unit......................................................................................................................8
3.1.5 check by means of biological indicators......................................................................................9
3.2 preparation of the sterilizer load................................................................................................9
3.2.1 textiles...............................................................................................................................................9
3.2.2 instruments.......................................................................................................................................9
3.2.3 rubber load........................................................................................................................................9
3.2.4 loads of special material.................................................................................................................9
3.2.5 liquids...............................................................................................................................................10
3.3 what may be sterilized....................................................................................................................10
5 SPECIFICATIONS ............................................................................................................ 12
5.1 technical specifications.................................................................................................................12
5.2 indicating and operating elements.................................................................................................12
5.2.1 operation and material loading side............................................................................................12
5.2.2 the side of unloading the material .................................................................................................13
6 PROGRAMS - BASIC....................................................................................................... 13
6.1 heating .............................................................................................................................................13
6.2 universal .........................................................................................................................................13
6.3 universal containers .....................................................................................................................13
6.4 rubber...............................................................................................................................................13
6.5 bd - test ...........................................................................................................................................13
6.6 vacuum test vt.................................................................................................................................13
6.7 instruments quickly .......................................................................................................................14
6.8 other programs .............................................................................................................................14
6.9 programs use and validation.........................................................................................................14
7 ATTENDANCE OF THE DEVICE..................................................................................... 14
7.1 basic operations of the attendance............................................................................................14
7.1.1 touch display...................................................................................................................................14
7.2 everyday putting into operation .................................................................................................14
7.3 the device operation.......................................................................................................................15
7.3.1 inserting the material and program start..................................................................................15
7.3.2 unloading of the sterilized material ..........................................................................................16
7.4 switching off the apparatus.........................................................................................................16
7.5 program interruption....................................................................................................................17
7.6 information - display.......................................................................................................................17
7.6.1 actual operational information, monitoring of the input media.............................................17
7.6.2 course of the cycle - information...............................................................................................17
7.7 menu - users setting and services...............................................................................................18
7.7.1 display settings...............................................................................................................................19
7.7.2 date and time....................................................................................................................................19
7.7.3 print - setting..................................................................................................................................20
7.7.4 automatic start...............................................................................................................................20
7.7.5 change of the program parameters............................................................................................21
7.7.6 review of the inputs and outputs................................................................................................22
7.7.7 system versions..............................................................................................................................22
7.7.8 air detector.....................................................................................................................................22
7.7.9 chip card...........................................................................................................................................22
7.7.10 steam source - change of steam supply manner........................................................................24
7.7.11 supplementary commentary............................................................................................................24
7.7.12 protocol history, memory card....................................................................................................25
7.7.13 errors history................................................................................................................................25
7.7.14 automatic door opening.................................................................................................................26
7.7.15 automatic device switching - off..................................................................................................26
7.8 access rights (log on)...................................................................................................................26
7.8.1 open user.........................................................................................................................................26
7.8.2 individual access rights.................................................................................................................26
7.8.2.1 log on - change of the level.........................................................................................................26
7.8.2.2 administrator...................................................................................................................................27
7.8.2.2.1 setting the mode - individual access rights/open user............................................................27
7.8.2.2.2 password administration and access securing..........................................................................27
7.8.2.2.3 daily charge counter.....................................................................................................................29
7.8.2.2.4 choosing a third variable in the recording table.....................................................................29
7.9 audit trail.........................................................................................................................................29
8 OPERATION FAILURES................................................................................................... 29
8.1 technical safety devices................................................................................................................29
8.1.1 error messages, unloading of the charge in case of failure.................................................29
8.1.2 operational message......................................................................................................................30
8.2 electric energy outage.................................................................................................................30
8.3 outage of input media pressure (media monitoring)..................................................................30
8.4 blocking the operation in case of electrical energy deficiency............................................30
14 WAY OF LIQUIDATION..................................................................................................... 36
14.1 package.............................................................................................................................................36
14.2 unit....................................................................................................................................................37
WARNING
2001 - 2400 m 87 °C 72 °C 82 °C
• The steam sterilizer UNISTERI® HP may be 2400 - 2800 m 86 °C 71 °C 81 °C
operated only by persons, who, on basis of their >2800 m 85 °C 70 °C 80 °C
training or knowledge and practical experience,
guarantee correct and proper operation. Workers
in charge of using the sterilizer shall be trained for
this activity and acquainted with the Instructions
for use. The keeper is held responsible for proper
training of the personnel in operating the sterilizer.
• The unit shall be used in accordance with the
Instructions for use to secure safety.
!
WARNING he steam sterilizer UNISTERI® HP is a device
T 2.1.1 CHECK BY CHEMICAL
intended for use in the healthcare for moist heat
sterilization of unwrapped and wrapped medical
PROCESS TEST
devices including invasive devices according to the
EU directive 93/42/EEC, intended by their • Chemical process tests react by colour change already
manufacturer for sterilization by moist heat. on the presence of the sterilization media and serve for
distinguishing between goods prepared to sterilization and
goods already sterilized.
1.2 INSTALLATION OF THE UNIT • Every package unit must be provided with a chemical
process test indicator (it is also possible to use
a selfadhesive chemical process test indication tape, which
Assembly of the unit, its installation and first putting into
is stuck onto the package containing goods to sterilize).
operation may be performed only by an authorized person
• Use only wrapping material and indication tape destinated
in accordance with the instructions contained in Chapter 12
for steam sterilization.
„Installation of the unit“.
• This test cannot be used to prove that the sterilized load is
really sterile!
1.3 STERILIZATION
2.1.2 CHECK BY CHEMICAL TEST
Sterilization is a process which kills any viable microorganisms OF STERILIZATION
inclusive of their especially resistant spores.The sterilization
inactivates viruses irreversibly. Sterilization kills also organisms, • The chemical test of sterilization reacts by a colour
whose do not directly induce any disease of the patient, but
change in case the required values of some or all
might endanger him by weakening his organism.
sterilization cycle parameters have been reached.
• Use only chemical tests destinated for steam sterilization
and the test evaluating methodology (colour change
evaluation) recomended by the producer.
2 STEAM • This test cannot be used to prove that the sterilized load
is really sterile as well!
STERILIZATION
2.1.3 CHECK BY MEANS OF THE
During steam sterilization, saturated water steam condensates BOWIE-DICK TEST (BD-TEST)
on the material being sterilized, what causes its heating.
The condensate penetrates through the cellular shell of the
microorganism directly to the cell nucleus and destroys it due to • Perform the BD-test regularly every day before the
thermal energy transfer. From the said it is obvious that, inspite beginning of sterilization, after having run the first
of a relatively short sterilization exposure and low temperature, sterilization cycle with an empty chamber.
the sterilization effect of saturated steam is very high. Steam • The BD-test confirms the correct course of the sterilization
sterilization is suitable for sterilizing textiles made of cotton and program (venting, fractional course of vacuum, sterilization
heat resistant textile blends, metal goods with sufficient corrosion temperature during the exposure, steam quality...).
resistance and many articles made of rubber, plastics, ceramics • Perform the BD-test by strictly following the instructions
and glass. The used sterilization temperature is mostly 134 of the indicator manufacturer.
°C. Sensitive rubber articles, parts made of plastics, glass and • If you sterilize common porous materials in the device, it
sensitive mechanical appliances are sterilized at 121 °C. To suffices to perform the BD-test according to EN ISO 11140-4.
assess objects that are to be sterilized in saturated water vapour, • If you sterilize hollow objects, the BD-test must be
follow the instructions for use of such objects manufacturers. In performed using a process challenge device (PCD)
questionable cases, you should make sure at first whether such according to EN 867-5.
objects are resistant to increased temperature, moisture, and
temperature and pressure changes resulting from the sterilization
procedure. If this is not known in individual cases, you should 2.1.4 CHECK BUILT IN THE UNIT
make sure by a testing sterilization.
2.3 WHAT MAY BE STERILIZED The pressure devices of the stabiliser are designed, produced
and tested according to the procedures given by the EU
Directive No. 2014/68/EU. Pressure devices of a sterilizer,
! are made as an assembly that is defined by piping schemes.
Safety of operation of this assembly is connected with the
WARNING Inside the apparatuses UNISTERI® HP it is possible working safety of the complete sterilizer and it is described
to sterilize only such material, which is permitted for in the appropriate chapter of these instructions. Procedures
a steam sterilization by the manufacturer: for maintenance, cleaning, attendance of pressure parts,
• instruments, inclusive recommended intervals are also included in these
• textile, instructions. Inspection and periodical checks of the sterilizer
• objects made form rubber, pressure devices at the user’s place are regulated by national
• objects made form plastics (withstanding the sterilization regulations for operation of pressure devices.
temperature),
• water solutions in opened bottles (only by corresponding
programme for solution sterilisation!).
• qater solutions in closed bottles (only by a programmewith 3.2.1 ASSEMBLY OF THE STERILIZER
forced cooling for solution sterilization and after consulting PRESSURE DEVICES
the sterilizer manufacturer!).
• pressure chamber of the sterilizer (according to the
required volume and number of doors);
• electric steam generator;
• piping;
3 DESCRIPTION OF THE • safety and pressure equipment.
UNIT
3.2.1.1 PRESSURE CHAMBER
The safety and pressure equipment is described in the 3.7 SIMCON (ECOSOFT)
pipeline connection diagrams and consists mainly of:
• Safety valve V2; UNISTERI® HP can use an optional SimCon (Simple Connect)
• Level switch B90 (HW) – regulation of the operating function. It allows the sterilizer integration in the Hospital24
water level in the generator; network by means of SimCon Netbox. SimCon NetBox is
• Level switch B91 (NW) – emergency heating switch-off if a device which is connected to the sterilizer by means of a serial
the water level in the generator is low; RS232 line. The other device terminal is connected to the local
• PE3 – pressure sensor JUMO – regulation of pressure in computer network. The network server saves all sterilization
the generator; cycle records and generates the sterilization protocols. The
• Thermal fuse TE1 – protection of the heating elements procedure is uniform for all different types of devices integrated
in the generator (e.g. in case of an emergency operation in the network. The SimCon function can be switched off from
without water); the service menu and its serial channel is then free for print data
• PE1.1, 1.2, 2 – pressure sensors JUMO – regulation of sending and archiving e.g. by means of PrinterArchiv software.
pressure in the sterilization chamber jacket and chamber;
• PT1.1, 1.2 – temperature sensors PT100, JUMO – NOTE If SimCon is switched on but disabled for any
sterilization chamber temperature checking. reason (e.g. a disconnected cable etc.), neither the printer
will print nor the data will be sent to an external device. If
this happens, the cause must be corrected and the sterilizer
3.3 DOORS must then be switched off and on.
5.4 RUBBER
5.5 BD - TEST
For the termination of this screen, push the button for return:
• Basic screen of the running sterilization cycle: • Digital record of pressure and temperature:
On this screen there are emphasized all the services, which are
accessible to the attendance. For the selection of the required
6.7.2 DATE AND TIME
service, push the button of the appropriate selection:
• Push the button MENU/Date And Time.
6.7.1 DISPLAY SETTINGS NOTE The password is pre-set to empty line by the
manufacturer, in this case, a confirmation by the “Enter”
• Push the button MENU/Display Settings. button is sufficient. The administrator is allowed to change
On the screen it is possible to select subsequently by the names and passwords.
pushing the appropriate button: • By pushing you activate the line of time and by means of
the keyboard you set the actual time.
• By pushing you activate the line of time and by means of
the keyboard you set the actual date.
Output Direction: • User Color Profile - this function permits the printing of
• Printer - activation of the built in printer(additive). values of custom adjustment of the colour environment.
• PrinterArchiv - activation of the output to PrinterArchiv For the termination of this screen, push the button for return:
(archiving of the output data from the sterilizer in the
computer).
Pressures And Temperatures Record NOTE In case of a failure the printer lists the appropriate
• Graphic - activation of the graphic form (line) record failure reporting. In case of a long time archiving it is
of pressures and temperatures. After pushing the button necessary to keep the record from the printer in dry and
Print period it is possible to set the required frequency of cold environment without a direct light.
the appropriate points of the graphic record (interval).
• Digital - activation of the digital form of record of
pressures and temperatures. After pushing the button 6.7.4 AUTOMATIC START
Print period it is possible to set the required interval of
listing of the values of the digital record.
This service makes possible to program the automatic start
• None - record of pressures and temperatures is switched off.
of the apparatus.
Immediate Print: It enables the selection of a vacuum test and another
• Last Protocol - this function makes possible the sterilization program.
additional printing of the last protocol from the memory • Push the button MENU/Automatic Start.
of the apparatus. (printing the data will be performed
according to the actually set configuration)
• Protocol History - this function enables choosing the
required record (last 10 records) from the history and
its printing, or displaying the record of pressure and
temperature sensors in a graphical or numerical form on
the display.
For selection of the second program (B), push the button 6.7.5 CHANGE OF THE PROGRAM
Program B. From the following screen, please, select in PARAMETERS
case of need the required sterilization program. By pushing
the button None Program the program B is omitted.
This service makes possible to adapt in the user´s mode the
selected parameters of the sterilization cycle (additive).
!
WARNING The editing may be performed only by a qualified
person! By this procedure it is possible to
significantly influence the sterilization efficiency of
the edited program! The manufacturer does not
responsible for the parameters of the program
edited in such a way!
• Push the button MENU/Program Parameter Change.
• Use arrows to select the required parameter. The air detector (aditivum) operates automatically and in case
of possible exceeding of threshold of the noncondensed gases
• Use the +/- button to change the parameter: +/- in the sterilizer chamber, it announces the appropriate error
• Insert the new value of the given parameter and confirm message, which interrupts the sterilization cycle. In case of
it by pressing the Enter button. (The permitted range of instruments with the activated Air detector, the corresponding
the adjusted parameter is shown in the brackets.) information is printed on the printer too. The eventual adjusting
of the Air detector is described in the service manual.
• Enter the name (only if it is requested in given mode • Press button for copying.
of operation).
• Enter the password and confirm by button Enter:
NOTE
↵
The password is pre-set to empty line by the
manufacturer, in this case, a confirmation by the “Enter”
button is sufficient. The administrator is allowed to change
the names and passwords. The administrator can change
the password.
Programmes deleting:
• Press the buttons of the programmes you want to delete
6.7.10 STEAM SOURCE - CHANGE OF
(on one side only). STEAM SUPPLY MANNER
• Close the manual valve of the steam supply from the el.
generator and open the valve for the steam supply from
an external source.
Supply from the own el. generator
Check, if the valve for demi-water supply is open - DW into
the device.
Individual programmes transferring within the device: • After water is topped up, the generator starts working
• Press the button of the appropriate programme you want automatically. The electrical steam generator is ready
to transfer. to operate as soon as the working pressure is reached
• Press the button of a free (Not Defined) programme (approx. 340 kPa).
where the given programme is to be transferred. • Close the external valve of the steam supply from the
• Press the Transfer button and then the Confirm button. central source, open the manual valve of the steam
To terminate the service, press the Return button: supply from the own (integrated) generator.
NOTE
How to write the commentary: a graphical or numeric way. The button Display Period [s] n
• Press the MENU/Additional Comment. allows selection of the sampling period for graphic and table
expression of temperature and pressure parameters of cycles.
• To end the mode, press the Return button:
SD memory card
The memory card is used for large-extent history saving
of sterilization protocols. The memory card allows also the
Audit Trail recording.
• Press Menu/Protocol History /Memory Card button:
The memory card can be then used for any individual saved
protocol displaying.
NOTE The commentaries to individual programmes are
stored in the device memory also if the activation is off. The
6.7.13 ERRORS HISTORY
commentary is displayed together with the parameters during
the programmes selection and is printed by the printer.
Press the buttons of the service MENU/Error History.
Setting the function of automatic sterilizer door opening The device could be used in two modes of operation:
after the end of the sterilization cycle. • Open User;
The administrator can activate this function for selected
programmes. • individual access rights.
The device is pre-set for free use (this status is displayed on
• Press MENU/Automatic Door Opening. the display by the inscription “Open User“).
The device enables to set access rights for the user on the
levels:
• user (device operation);
Activate the automatic door opening for selected programmes. • administrator (working place head).
• Then press the button for return:
Name: admin
• Enter the required period of time. (The pre-set name “admin” will be changed by the
administrator.)
Password: *****
(The pre-set Password: “*****” will be changed by the
administrator.)
6.8.2.2 ADMINISTRATOR
The administrator has pre-set name to “admin” and Select the requested mode by the check button
password to “heslo”. The administrator (after log on) has, in Individual Access Rights in the upper part of the display
addition, some rights, as. (Administrator).
• Setting the mode “Individual access rights” or “Open User”.
• Passwords management and access securing.
• Daily Charge Counter. 6.8.2.2.2 PASSWORD ADMINISTRATION
• Optional analogue variable displayed in the third column in AND ACCESS SECURING
the table of variables displayed during the programme run.
If you are not a person with administrator rights to your
device, skip the whole chapter Administrator. On the right, on the Administrator button level, there is the
Default button, by which could be set the administrator name
If you are a person with administrator rights to your and password to initial values so, as they were pre-set.
device, log on as the administrator and press the MENU/ By the Administrator button, you can reach the
Administrator. “Administrator – Access securing” screen for administrator
name and password setting.
The User 1 will log in under the name “tom” and password
“1212”. His name will be displayed as “tomas”. Such user is
entitled to run the programmes except for P5; he may cancel
the error messages, set the device time, interrupt the running
programme, and change the programme parameters.
Setting for the mode - Open User
CAUTION T
he safety valves function shall be checked by
Soporte de
la obturación an authorized person according to the national
regulations.
Obturación Half-yearly
de la puerta It is necessary to put the steam sterilizer after 800 cycles,
but no later than within every six months, to periodical
service inspection.
Instrumento
a quitar la Service inspection - contents:
obturación • After 800 cycles, but every six months at the latest,the
steam sterilizer must undergo a periodic serviceinspection
performed by a service technician.Service inspection –
description.
• Check the tightness of screwing, pipe joints, hoses andtanks.
• Check visually whether the sterile aeration filter fillinghas
CAUTION The sealing must not be damaged while beingtaken not been damaged and replace it if necessary.
out.Check the sealing for mechanical damage, wear • Check and set, if necessary, the anti-vacuum valve ofthe
and tearand hardness changes. In case of vacuum pump.
establishing anay of thefault, replace the sealing. • Check the lowest underpressure (vacuum) of thevacuum
pump.
• Check the temperature measurement equipment.
32/44 UNISTERI HP_np_en 2018-09_V4.00
Instructions for use
The steam generators are supplied with water by means ofa The consumables can be ordered under the following order
pump from a tank with regulated water level. The waterlevel numbers:
in the storage tank is controlled and monitored by means of Component name Order number
a level indicator S10. If there is not enough water,the supply
Door sealing 336 fixed 010094748
pump of the generator is blocked.
Door sealing 559 fixed 010446568
Electrical steam generator (ED) Door sealing 636 fixed 010108622
Water level in the boiler is checked by means of levelswitches.
Door sealing 336 – movable / flexible 010120942
The heating elements protection is doubled. Thelowest water
level is safeguarded by means of a safety levelswitch B91 Door sealing 559 – movable / flexible 010446762
(switch-on blocking).The supply water level is safeguarded Door sealing 636 – movable / flexible 010120943
by the level switchB90 which has a protective function as Aeration filter 000538626
well. If the level isnot automatically filled within the time limit Cleaning
(approx. 40 s),the generator heating is switched off reversibly. Cleaning means: 000544308
If the switchB90 is not submerged (enough water is pumped) „ Ultrapur sprey „- face panels 000644375
for morethan approx. 4 minutes, a breakdown is announced.
The generator is pressure-controlled by means ofa pressure „ Cif „- chamber 010096216
converter.The generator heating elements are protected Termosensitive paper DPA-038-MG10 010447229
againstoverheating (e.g. operation without water) by means ofa for printing
thermostat TE1. Chip card UNISTERI® HP 010445887
Adjusted supply water must be used for the steamgenerator.
The limit conductivity value of completelydesalted water
withdrawn from the water-treatment plantis max. 15 μS/cm at
water temperature of 20 °C (seeChapter 12.2.2). The values are
generally read directlyInstructions for use from the measurers
placed at the water desalting facility.Even though the generator
11 STERILIZER
is supplied with treated water,salt concentration in the generator
increases gradually dueto the boiler water evaporation. Such
INSTALLATION
salt concentrationwhich is harmful to the generator pressure
vessel material,causes corrosion (so called pitting corrosion).
It is thereforenecessary to desalt the generator regularly. Partial
desalting of the generator is carried out automaticallyat each CAUTION T
he sterilizer may be installed and put into
supply water filling when the desalting valve Y99 isopened for
operation by an authorized person only.
about 3 seconds.The generator sludging (complete emptying)
must beperformed every six months at least.
11.2.3 DRAIN - A
CAUTION Not keeping the quality requirements on service The waste piping must be provided with a disconnecting
media negatively influences the unit‘s functionality trap. The waste pipes material must withstand the
and consequently causes the need of more temperature of 100 °C; see EN 285. This condition must be
frequent servicing. This may lead to a limitation of complied with only in the variant without the waste water
the warranty period or even to refusal of the aftercooling equipment.
warranty as a final consequence.
The device assembly and its first putting into operation may
!
only be done by an authorized person. WARNING
• It is recommended to demount the side and bottom tiling • The upper part of the device is heavier. Beware of
before the device is put down from the pallet. overturning!
• When putting the device down from the pallet and handling • Depending on the model, the device can be equipped
it, hold only the fixed parts – the frame or the chamber. with travelling wheels (optional equipment) for easier
When handling the device, do not hold the face panels! moving. Such wheels are not intended for long
• The device can be moved on a hard and flat floor by means transport but for local moving only!
of integrated wheels. • When putting the device down from the pallet and
• Connect the hose connectors. Roughen the threads and handling it, hold only the fixed parts – the frame. When
wrap teflon tape around them except for the first thread. The handling the device, do not hold the face panels!
drain must be connected first. • The device has adjustable legs by means of which
• Place the device accurately to its position and install it it can be levelled at the site and fixed. Pull out the
by means of adjustable legs (the screws M10 should be legs by means of screws (M10) so as the travelling
secured against loosening by means of nuts). If the bottom wheels were approx. 1-2 mm above the floor.
face panels were demounted, remount them on the device.
• Connect the electric cable plug to the socket.
• Hoses must be positioned in the device to form large bends;
the discharge hose must not warp above the discharge
elbow-pipe on the storage tank.
• Open the closing valves. Check the installation junctions 12 TRANSPORT
sealing.
• Switch on the main electric switch on the outer power supply. AND STORAGE
• Check whether the device is in the “Ready-to Start” status.
• Switch the device on.
• Take out immediately the objects stored during the transport
in the chamber (instructions for use, revision log, etc.)!
• Check the basic device setting in the service menu. 12.1 TRANSPORT
• Check the door closing function.
• Wait till the generator is filled and pressurized to approx. The unit is being prepared for transport by an authorized
3.4 bar (absolute pressure). Manual filling of the generator person, who also dismounts the service connections. The
is recommended under concurrent monitoring of the level unit shall be transported in its original package.
indicator NW, HW by means of service menu.
• Carry out the V-test.
• Carry out the test sterilization programme.
!
WARNING The upper part of the device is heavier. Beware
• Check the setting of all temperature and pressure sensors
(in heated condition); recalibrate them if necessary. of overturning!
• Carry out the test sterilization programme.
12.2 STORAGE
11.3.1 BEFORE THE FIRST DEVICE USE
For storage the unit is being prepared by an authorized
person who dismounts the service connections, drains off
the pipework and storage tanks and dries them by pressure
CAUTION air. The sterilizer shall be put in its original package. The
The technical documentation, sterilizer storage temperature shall not drop below 4 °C.
components and equipment in a plastic pack are
placed in the sterilization chamber for transport.
Do not forget to take them out immediately after
the sterilizer is placed to its working site. When
opening the sterilization chamber, follow the
instructions of Chapter 7.1, 7.2 and 7.3.
13 WAY OF LIQUIDATION
13.1 PACKAGE
Valid only for devices provided with another large display Output protocol example:
(LCD) at the unloading side (optional). This modification
allows full comfortable control of the device as well as the
sterilized material flow from both sides (according to the
device setting).
After pressing the button the service technician can suppress errors for up to 999 seconds (repeated pressing).
Door 1 Unsealed The door on the loading side is not sealed 1. Check the door sealing Operator /
2. Check the valve Y09 Service
3. Check the pressure switch B20
Door 2 Unsealed The door on the unloading side is not 1. Check the door sealing Operator /
sealed 2. Check the valve Y09 (Y09.1) Service
3. Check the pressure switch B20
(B20.1)
Door Sealed The door is sealed 1. Wait for the door to be unsealed Operator /
2. Check B20 and B20.1 Service
High Steam Generator High Steam Temperature in the Generator 1. Check the pressure sensor PE3 Service
Temperature 2. Check the temperature sensor
PT4
3. Check the printed circuit board
VP_CPUS
4. Check the contactors K4-7
High Chamber High steam temperature in the chamber 1. Check the door sealing Operator /
Temperature 2. Check the inlet steam Service
3. Check the valve Y07
4. Check temperature and pressure
sensors
Low Chamber Low temperature in the chamber 1. Perform the vacuum test. Start Operator /
Temperature again the cycle Service
2. Remove untightness
3. Check the condensate trap
4. Check the process parameters
5. Check the temperature and
pressure sensors
Phase Too Long The specified phase of the sterilization 1. Check the input media Operator /
cycle is longer than permitted 2. Check the inputs and outputs Service
Air In The Chamber - The quantity of gases unable to condensate 1. Check the sealing Operator /
Failed is above the limit 2. Check the tightness of joints Service
Voltage 24 V The voltage 24 V decreased 1. Check the fuse F4 Service
below 17 V 2. Check the door seal
3. Check the door motors
4. Check the pressure sensors
5. Check the printed circuit board
VP_CPUS and VP_PDC
Pressure Sensor Pressure sensor disconnected 1. Check the sensor and its Service
Disconnected connection
2. Check the printed circuit board
VP_CPUS
Pressure Sensor Over Pressure sensor is above limit 1. Check the sensor and its Service
Limit connection
2. Check the printed circuit board
VP_CPUS
Demi-Water Level Switch B86 is not flooded for over 180 s 1. Check the demi water supply Operator /
2. Check the sensor B 86 Service
Methylene Level B102 Switch B102 is not flooded for over 5 s 1. Check the methylene supply Operator /
2. Check the sensor B 102 Service
EEPROM Error It is not possible to read or write to 1. Switch the device off using the Operator /
EEPROM on some electronics board main switch and switch it on Service
again after 10s
2. Check connections and
electronics boards
Door 1 Not Closed Door 1 not closed 1. Close the door Operator /
Properly 2. Check the door sealing Service
3. Check the door switches
Door 2 Not Closed Door 2 not closed 1. Close the door Operator /
Properly 2. Check the door sealing Service
3. Check the door switches
Door Not Closed Door not closed 1. Close the door Operator /
Properly 2. Check the door sealing Service
3. Check the door switches
Water In The Chamber Water remains in the chamber after the 1. Check the waste piping including Service
cycle termination valves
2. Check the level meter S1
Watch Dog At Process VP_CPUM processor board function failed 1. Re-start the cycle Operator /
Time 2. Check the VP_CPUM… board Service
Solitary Slv Reset There was a processor reset on the 1. Check the VP_CPUS board Operator /
VP_CPUS board without resetting the 2. Check the supply voltage Service
VP_CPUM board 3. Remove a potential source of
electromagnetic interference
Fatal Error Slv VP_CPUS processor board function failed 1. Re-start the cycle Operator /
2. Check the VP_CPUS board Service
File Creating Error It is not possible to write on the SD card 1. Check the SD card Service
2. Check the VP_CPUM… board
3. Check the VP_KON board
Manual Valve V36 Or Valve V36 or V37 is not closed 1. Close the valves V36 a V37 Operator /
V37 Not Closed 2. Check the switches SV36 and Service
SV37
Door Sensor S11 The door switch S11 changes the value in 1. Open and close the door and start Operator /
the course of the cycle a new cycle Service
2. Adjust the switches S11
Door Sensor S13 The door switch S13 changes the value in 1. Open and close the door and start Operator /
the course of the cycle a new cycle Service
2. Adjust the switches S13
Door Driving Mechanism The door does not reach the limit switch 1. Check (lubricate) the door sealing Operator /
within specified time period 2. Check the door sensors Service
Unauthorized System Repeated trial of unauthorised person to An authorised person to unblock the Operator
Access access device using a valid name and
password
Mismatch In Mst/Bck Discrepancy in the memory of processors 1. Login as an Admin, change Operator /
Flash Memory - CFG Master and Back - offsets, configuration, anything in access rights and Service
access rights save it
2. Save configurations again
3. Save offsets again
Mismatch In Mst/Bck Discrepancy in the memory of processors Change some parameter in Operator /
Flash Memory - PRG Master and Back - program parameters the "Program Download" (e.g. Service
fast select) and save it
High Pressure In The At the time when the cycle is not running Check the valves Y80 and Y80.2 Service
Chamber the pressure in the chamber rises above
200 kPa
Temperature Sensor Temperature sensor disconnected 1. Check the sensor and its Service
Disconnected connection
2. Check the printed circuit board
VP_CPUS
Temperature Sensor Temperature sensor short circuit 1. Check the sensor and its Service
Short Circuit connection
2. Check the printed circuit board
VP_CPUS
Fo Parameter The value of the Fo parameter is outside Eliminate the source of interference on Service
the expected range the temperature transmitter
Program For User The user is not authorised to start the A user with sufficient authorisation Operator /
Disabled program must log in Service
Invalid Content ! The protocol has invalid content, the x Operator /
original protocol has been changed Service
See 'System Version' For a more detailed specification, see the x Operator /
'System version' Service
Vacuum Pump Circuit Breaker The Q1 breaker is switched off 1. Switch on the breaker Service
2. Check the vacuum pump
Water Pump Circuit Breaker The Q2 breaker is switched off 1. Switch on the breaker Service
2. Check the pump
Methylene Pump Circuit The Q3 breaker is switched off 1. Switch on the breaker Service
Breaker 2. Check the pump
Steam Generator Turned Off The generator is switched off Switch on the generator Operator
Steam Generator Turned On The generator is switched on Switch off the generator Operator
Saving Failed, Repeat Action! Saving failed 1. Repeat saving Operator / Service
2. Check connection and
electronics boards
Door Sealed The door is sealed 1. Wait for the door to be Operator / Service
unsealed
2. Check B20 and B20.1
Service Door Is Open Service door open 1. Close the door Operator / Service
2. Check the door switches
Door 1 Security Bar Activated Door 1 Security Bar Activated Check the protective strip Operator / Service
Door 2 Security Bar Activated Door 2 Security Bar Activated Check the protective strip Operator / Service
Door Security Bar Activated Door Security Bar Activated Check the protective strip Operator / Service
Atmospheric Pressure Error, Atmospheric Pressure Error, Open The Open the door Operator / Service
Open The Door Door
Program For User Disabled The user is not authorised to start the A user with sufficient Operator / Service
program authorisation must log in
Insufficient Rights For This The user is not authorised to start the A user with sufficient Operator / Service
Action action authorisation must log in
Manufacturer: Distributor: