The following procedures should be performed in the following order every day.
1. Perform a Full Shutdown of the Infinia application (if not done at end of
previous day). This closes the Infinia application and logs off the current
user.
Note It is important that the acquisition console is shutdown via the Full shutdown
option from the Infinia application. Do not close the Infinia application via the
Windows 2000 options (Start > Shut Down...) or by using the Power OFF button
of the computer.
4. Using the RCU, verify that the detectors can be moved “IN” and “OUT”.
8. Check Disk Space as described on page 8-1 and clean up the To Do list.
If a study is no longer needed in the To Do list, press
[Study Acq. Completed].
9. Go to the Data Management list and verify that all data has been
transferred to Xeleris. Clean up the Data Management screen. Delete or
archive all studies that have been transferred and are no longer necessary
on Infinia. It is recommended to delete all studies that are more than one
week old from the local Data Base.