Professional Documents
Culture Documents
Servo-U Startup Guide
Servo-U Startup Guide
Startup guide
1 System overview...................................................5
2 Alarms.....................................................................20
3 Trends.....................................................................27
4 NAVA and NIV NAVA.........................................28
5 Views...................................................................... 30
6 Media......................................................................37
7 Neonatal patient category............................ 38
Note
This guide is intended for Some modes and functions
hospital personnel as start up are options and might not be
training using the SERVO-U ® included.
ventilator. It does not cover all
aspects of the SERVO-U ventila-
tor. Please see the user’s manual
for more information.
S E R V O - U S TA R T U P G U I D E 3
1 System overview
The user interface
can swivel/tilt.
ON/OFF switch
4 S E R V O - U S TA R T U P G U I D E
Modules are interchangable between modules. An alarm “Missing Battery” will be
SERVO-i®/n and U ventilator models (except Y activated if less than 2 batteries are mounted.
sensor modules). The SERVO-U can host 2-6
battery modules and several different plug-in
Y sensor module
CO2 module
Edi module
Batteries
S E R V O - U S TA R T U P G U I D E 5
1. Expiratory outlet 13. Lock for inspiratory channel cover
13 16
2 20
3 14 17
4 15 18
5 21
6 22
7
8 19
10 11 12
6 S E R V O - U S TA R T U P G U I D E
1. Alarm indicator, light frame
6. Loudspeaker
9. VGA port
1
1
6
3
4 7
5
2
11 10 9 8
S E R V O - U S TA R T U P G U I D E 7
1.2 Workflow to start ventilation
1. Connect the ventilator to the mains power. 7. Select invasive or non-invasive
ventilation.
2. Connect pressurized 02 and air tubes.
8. Select ventilation mode and adjust
3. Connect the patient circuit and filter. settings.
If active humidification add water to the
water chamber. 9. Check and adjust alarm limits.
8 S E R V O - U S TA R T U P G U I D E
The patient circuit test is included in the
pre-use check but can also be selected
separately.
S E R V O - U S TA R T U P G U I D E 9
1.4 Modes and settings
There are three different patient categories,
adult, pediatric and neonatal. The neonatal
patient category is optional on SERVO-U and
will therefore not be shown if not included.
10 S E R V O - U S TA R T U P G U I D E
Slide the bar to the right or left to increase
or decrease the settings. Confirm the setting
by tapping .
Exit settings without changing by tapping
the .
The bar displays the safety scale, which normal use again tap the . Note that this
is the range that represents normal use. can only be done if the current value is within
To access the full settings range tap the . the normal settings range.
To only show the range that represents
S E R V O - U S TA R T U P G U I D E 11
1.5 Modes
Tap to select mode. The current mode tile is
always highlighted and the previous mode
tile is marked PREVIOUS, together with the
date and time it was last used.
Non invasive modes: All non invasive modes atric and neonatal patient categories,
are optional and may therefore may not be and NIV PS is available in adult and pediatric
available. Nasal CPAP is available in the pedi- patient categories.
12 S E R V O - U S TA R T U P G U I D E
The mode settings are divided into
supported and controlled settings.
S E R V O - U S TA R T U P G U I D E 13
Dynamic images are presented for some of
the settings. A dynamic image illustrates the
effects of changes made.
14 S E R V O - U S TA R T U P G U I D E
1.7 Trigger settings
When triggering is based on flow, to the left When triggering is based on pressure, to
on the scale, the ventilator system senses the right on the scale, the ventilator system
deviations in the bias flow delivered during senses deviations in the pressure below
expiration. The further to the left on the PEEP created by the patient. The pressure
scale, the less effort the patient has to make. below PEEP required to initiate a breath is
At the far left of the scale, there is a risk of displayed when the setting is made. The fur-
auto-triggering, and the scale and value are ther to the right on the scale, the greater the
therefore marked in red. patient effort required to trigger.
S E R V O - U S TA R T U P G U I D E 15
1.8 VT&PBW
In adult patient category enter patient patient circuit compensation is on to receive
gender and height, in pediatric and neonatal accurate VT/PBW.
categories enter weight. It is important that
The ventilator monitors the ratio of tidal right of the volume. VT/PBW (ml/kg) is
volume to predicted body weight (VT/PBW). continuously trended and measured.
In volume controlled modes the VT/PBW For pediatric/neonatal patient category
(ml/kg) is calculated and presented to the body weight (BW) is used.
16 S E R V O - U S TA R T U P G U I D E
1.9 User interface
Status bar
Numerical values
Quick menu/
Extended menu
Direct access bar
S E R V O - U S TA R T U P G U I D E 17
1.10 Active/inactive modes
and settings
The grey text and settings indicate an
inactive mode and settings. If a mode is
changed then the other mode becomes
white.
18 S E R V O - U S TA R T U P G U I D E
2 Alarms
Light frame for 360° visibility. Alarm indica- High prioirty
tion is shown in two ways; blinking value Medium priority
(measured or calculated) and alarm message Low priority
in the alarm message area.
With some alarms the audio can be turned ical values area and a message is displayed in
off by tapping . Audio off is displayed in the status bar.
the corresponding parameter in the numer-
S E R V O - U S TA R T U P G U I D E 19
Settings for alarm limits include upper Autoset alarm limits are available in
and lower limits settings and the current controlled modes.
measured value.
Two alarms active
Alarm management
checklist
20 S E R V O - U S TA R T U P G U I D E
2.1 Batteries
If the ventilator system is running on battery estimated remaining battery time in minutes
power, the battery symbol turns yellow and is always displayed, regardless of the power
the mains power symbol disappears. The supply in use.
2.2 Maneuvers
When the MANUAL BREATH is tapped,
the ventilator system will initiate a new
breath cycle according to the current
ventilator settings.
S E R V O - U S TA R T U P G U I D E 21
2.3 Static measurements
22 S E R V O - U S TA R T U P G U I D E
2.6 Nebulization
Nebulization can be either activated for
a certain period of time (5–30 minutes) or
continuous (only use Aerogen Solo).
2.7 O2 boost
When tapped, O2 boost delivers the interrupted by tapping the red cancel symbol
oxygen setting displayed here for a period in the O2 boost timer window anytime during
of 1 minute. The O2 boost function can be the 1 minute interval.
S E R V O - U S TA R T U P G U I D E 23
By tapping O2 BOOST LEVEL, it is possible to O2 boost level is 30 %, the O2 boost function
change the desired level for the O2 boost will, when tapped, deliver 70 % O2 . The O2
function. It is possible to lock the O2 boost boost function figure displayed will change
level to 100 %. The value entered under O2 accordingly. Since the minimum O2 concen-
boost (%) level specifies the number of per- tration is 21 %, the O2 boost (%) level scale
centage units that will be added to the value goes from 0 to 79 %.
set for the O2 concentration. For example: if
the current O2 concentration is 40 % and the
24 S E R V O - U S TA R T U P G U I D E
2.8 Disconnect/suction support
DISCONNECTION/SUCTION enables When using a closed-suction system,
automatic inhibition of the ventilator system DISCONNECTION/SUCTION should not be
during a tracheal suction procedure or when used. The O2 boost function should be used
briefly pausing ventilation in invasive modes. instead for oxygenation purposes. Consider
The ventilator system is prevented from pre-silencing the alarms before suctioning.
cycling without activating alarms.
Important
Use DISCONNECTION/SUCTION for all discon-
nections of the patient circuit when ventilating
in NAVA or in the neonatal patient category
when leakage compensation is active.
S E R V O - U S TA R T U P G U I D E 25
3 Trends
Trend values are stored every 60 seconds figure and an explanation appears to the left
and retained for a maximum of 72 hours. of this ring.
Stored events and system changes are also If a recording is saved at a time corre-
displayed here. sponding to the cursor position, a recorder
The time valid for the cursor position is is displayed. To view the recording, tap this
displayed. If events have been stored, their recorder.
number is displayed in the ring shown in the
26 S E R V O - U S TA R T U P G U I D E
4 NAVA and NIV NAVA
The NAVA® and NIV NAVA workflows facili-
tate the set-up and use of these modes.
In the verticle field both the symbols on the tected, there will be no such indications.
representation of the Edi catheter and the Note that the color highlights on the leads
color highlights on the leads show where the change more rapidly than the symbols on the
Edi signal is strongest. If no Edi signal is de- Edi catheter.
S E R V O - U S TA R T U P G U I D E 27
The pressure waveform (yellow) and the Pest Edi catheter positioning window showing
wave (grey) are at the bottom of the position- error and the catheter should be moved up
ing window. The waveforms can be used to slightly.
evaluate synchrony.
Important
The positioning arrows can only be used to fine
tune the catheter position when there is a reli-
able Edi signal. If the Edi signal is low or absent,
the arrows should not be used to adjust the
catheter position.
28 S E R V O - U S TA R T U P G U I D E
5 Views
The ventilator system offers different views
to suit different needs. They are accessed via
the quick menu during ventilation.
S E R V O - U S TA R T U P G U I D E 29
5.2 Advanced view
The view consists of two to five waveforms. if desired and the Edi and CO2 waveforms,
Pressure and flow waveforms are always if available are presented in two columns of
present, together with the volume waveform numerical values.
30 S E R V O - U S TA R T U P G U I D E
5.4 Compass view
SERVO COMPASS® visualizes volume and If the driving pressure or total pressure
pressure in relation to set targets in invasive exceeds the pressure limit, the pressure limit
modes. the pressure animation will change color.
S E R V O - U S TA R T U P G U I D E 31
5.6 Family view
Displayed information is minimized to: – A dynamic representation (moving bub-
– One column of numerical values. bles) shows that ventilation is in progress.
– Alarms and messages in the status bar. To exit the family view you can tap anywhere
– The direct access settings bar. on the screen.
Choose filled or
non-filled waveforms
32 S E R V O - U S TA R T U P G U I D E
Filled waveforms
S E R V O - U S TA R T U P G U I D E 33
5.7 Panel lock
Possible to lock the screen for example for
cleaning. Tap and hold to unlock.
34 S E R V O - U S TA R T U P G U I D E
5.8 High Flow therapy
High Flow therapy can be selected in both Possibility to switch directly from invasive
invasive and non-invasive ventilation as well ventilation.
as in Standby.
S E R V O - U S TA R T U P G U I D E 35
6 Media
Screenshot
Recording
6.1 Recording
Recording highlighted to left and screenshot The recording will be stamped with the
to right. date and time that it was initiated and will be
A 30 second long recording will be made saved under the Recordings tab in the MEDIA
starting 15 seconds before, and lasting until Iibrary. Forty recordings can be saved on
15 seconds after the time the recording the SERVO-U.
was initiated.
6.2 Screenshots
The screenshot will be stamped with the All screenshots and recordings can be
date and time it was taken and saved under transferred to a USB memory stick. The USB
the Saved screens tab in the MEDIA window. port is under the user interface.
Forty screenshots can be saved.
36 S E R V O - U S TA R T U P G U I D E
7 Neonatal patient category
7.1 Invasive leakage compensation
Invasive leakage compensation is available will show the symbol for leakage compen-
in the neonatal patient category to compen- sation.
sate for the leakage around uncuffed endo- Leakage compensation maintains PEEP,
tracheal tubes common when ventilating compensates for trigger sensitivity, end in-
neonatal patients. spiration and measured values, as well as
SERVO-U has real volume guarantee in adding volume in PRVC and VS to reach
PRVC and VS, the leakage loss is added to targeted tidal volume.
inspiratory volume.
Compensated values in the user interface
S E R V O - U S TA R T U P G U I D E 37
7.2 Volume restriction
For the neonatal patient category, the volume than set by this limit will not be
ventilator system has a volume restriction. delivered. This applies in: PC, PS, AUTO-
This means that the alarm limit set for VTi MODE® PC PS and all SIMV modes.
restricts volume delivery, so that a higher
7.3 Y sensor
For tidal volumes below 10 ml, it is recom- To guarantee that waveforms and mea-
mended to use a Y sensor to increase the sured values are always displayed on the
accuracy of gas delivery and monitoring. screen, the internal pressure and flow sensors
When the Y sensor is active, the flow are at all times active as backup. Their read-
through the sensor will replace the bias ings are compared with the Y sensor measure-
flow as source for flow based triggering. The ment. The Y sensor is disabled if there is a
patient pressure is measured at the Y piece significant deviation or malfunction.
via a pressure line.
To interrupt the Y sensor measuring,
disconnect the Y sensor module.
38 S E R V O - U S TA R T U P G U I D E
S E R V O - U S TA R T U P G U I D E 39
© Maquet Critical Care AB 2017 · Servo-u, Servo-i , Ssevo compass, NAVA and Automode are trademarks of Getinge or any of its subsidaries · mx-5948 Rev05 · English
This document is intended to provide information to an international audience outside of the US. Servo-u may be pending
regulatory approvals to be marketed in your country. Contact your Getinge representative for more information.
Getinge is a global provider of innovative solutions for operating rooms, intensive care units, sterilization departments and
for life science companies and institutions. Based on our firsthand experience and close partnerships with clinical experts,
healthcare professionals and medtech specialists, we are improving the everyday life for people, today and tomorrow.
Manufacturer · Maquet Critical Care AB · Röntgenvägen 2 SE-171 54 Solna · Sweden · +46 (0)10 335 73 00
www.getinge.com