Application of Rapid Microbiological Met

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 11

Journal of Applied Microbiology ISSN 1364-5072

ORIGINAL ARTICLE

Application of rapid microbiological methods for the risk


assessment of controlled biopharmaceutical environments
T. Sandle, C. Leavy, H. Jindal and R. Rhodes
Bio Products Laboratory, Elstree, UK

Keywords Abstract
cleanrooms, environmental monitoring,
IMD-Aâ, particle counting, real-time Aims: To assess the different operational states within a biopharmaceutical
monitoring, spectrophotometric, viable but grade clean room, using a rapid microbiological method. The method was a
nonculturable. novel system, based on spectrometry, designed for sampling, discriminating,
and enumerating airborne particles. Central to the study was the aim to
Correspondence
determine the microbiological levels as a clean room went from standard use
Tim Sandle, Bio Products Laboratory, Dagger
through maintenance and shutdown, disinfection, and then back to standard
Lane, Elstree, Hertfordshire, UK WD6 3BX.
E-mail: timsandle@btinternet.com use. The objective was to evaluate whether a rapid method could replace
conventional environmental monitoring using growth-based media.
2014/0020: received 5 January 2014, revised Methods and Results: The instrument evacuated was a BioVigilant IMD-Aâ
16 February 2014 and accepted 20 February System, which is a real-time and continuous monitoring technology based on
2014 optical spectroscopy that can differentiate between biological particles and inert
ones (biological particles expressed as bio-counts based on the detection of
doi:10.1111/jam.12487
microbial metabolites). The results indicated that certain activities lead to a
high generation of biological particles and in showing an increase over the
baseline, would be regarded as presenting a microbiological risk to the
cleanroom. These activities include removing HEPA filter grilles, turning off an
air handing unit, and tasks which requires an active personnel presence, such
as cleaning and disinfection.
Conclusions: The optical instrument can be used to process sufficient
information, so that clean rooms can be returned to use following a period of
unexpected downtime or following maintenance without the need to wait for
the results from growth-based methods. As such, this type of rapid
microbiological method is worth exploring further for clean room air
monitoring.
Significance and Impact of the Study: Few studies have been undertaken
which examine air-monitoring devices that can both enumerate and
discriminate particulates, in a volume of air as ‘inert’ or ‘biological’. This study
extends this limited field. Furthermore, the data collected in relation to
cleanrooms is of interest in helping microbiologists understand that risks posed
by different activities in relation to clean air-handling systems and personnel
particle shedding.

contamination into a critical area and to protect pharma-


Introduction
ceutical products. Whilst the emphasis should also be
Cleanrooms and controlled environments within pharma- with environmental control (through an appropriately
ceutical facilities need to have robust systems of environ- designed heating, ventilation and air conditioning system
mental control to minimize the generation of particulate (HVAC) and strict personnel behavioural disciplines)
and microbiological contamination. Such control mea- (Sandle and Saghee 2013), understanding the extent
sures are also required to prevent the transfer of any to which control is being maintained can only be

Journal of Applied Microbiology © 2014 The Society for Applied Microbiology 1


Application of rapid microbiological methods T. Sandle et al.

determined through a robust environmental monitoring Due to the variables of length of incubation and subse-
programme. quent time-to-result, most cleanroom activities are car-
Conventional environmental monitoring programmes ried out with a degree of uncertainty given that the
use a combination of nonviable particle counting and tra- environmental conditions are unknown at the time of
ditional microbiological sampling methods (including processing. Uncertainty increases when unusual or less
volumetric samplers and settle plates; surface sampling well-practiced events happen, including those which
using swabs and contact plates; and, where appropriate, require a high level of personnel input or where air han-
samples of personnel suits and gloved hands). To address dling is not maintained. An example, and one upon
these considerations, many sections of the industry have which this paper is based, is when a cleanroom shuts
moved towards the adoption of risk-based approaches down. The closure of a cleanroom can occur due to an
(Sandle 2012a). unplanned event, such as power cut; or due to a planned
Risk assessment and risk management have been event, where a cleanroom is shutdown to allow for repair
advanced through International Conference on Harmoni- work. With these events, a cleanroom is restored for nor-
sation guidelines, most prominently ICH Q9 ‘Quality mal processing through re-activation of the air system
Risk Management’ (ICH 2011). Different analytical tools and by personnel performing cleaning and disinfection.
can be used for the identification and risk assessment of At the point when the cleanroom is ready to be handed
hazards. The most widely used are failure modes and back to operations staff, it will typically be subject to
effects analysis (FMEA) (Asefzadeh et al. 2013) and haz- environmental monitoring. At this point, the cleanroom
ard analysis and critical control points (HACCP) (Osi- manager has a choice: whether to wait for the results of
mani et al. 2013). The use of such risk guidance allows the growth-based environmental monitoring results or to
for the identification of hazards within the cleanroom use the cleanroom at risk.
(aspects that could potentially cause product harm) and To complement growth-based microbiological meth-
then to the evaluation of the risks posed by such hazards ods, particle counting can be performed. A concern with
through considering the severity of the hazard, should particle counting is that only a small proportion of parti-
contamination occur, and the probability of the hazard cles detected within a cleanroom will be indicative of
occurring. micro-organisms and there is no means to differentiate
The primary hazards within cleanrooms are micro- between ‘viable’ and ‘non-viable’ particulates (some
organisms. Most micro-organisms within cleanrooms are authors have attempted to show a correlation between
found in the air and the risk is the extent to which those the two (Raval et al. 2012); however, correlative relation-
present could contaminate a product, either through air ships are dependent upon too many variables for any
currents or through gravitational effects. Few micro- universal ratio to be provided) (Oxborrow et al. 1975; Li
organisms are found free-floating; most attached to and Hou 2003). To address this, rapid microbiological
‘microbial carrying particles’ such as dust or skin detritus air-monitoring methods have been developed (Miller
(Whyte and Hejab 2007); notwithstanding that some 2012).
reports suggest that bacteria can occur in the airstream in One recent type of technology is optical spectrophoto-
combination, forming larger particles (Tham and Zuraimi metric technologies, designed to simultaneously detect
2005). the number and size of particles from a volume of air
Once a hazard has been identified and where it is con- and to additionally detect micro-organisms within the air
sidered to pose an unacceptable level of risk, the objective sample (Sandle 2012b). The advantage of these systems is
should always be to eliminate the hazard. Where a hazard that by counting the microbial content of an aerosol sam-
cannot be completely eliminated, then some form of ple on a particle-by-particle basis, they provide a continu-
detection system (in this case environmental monitoring) ous, real-time analysis of the microbial content of a given
should be in place to assess occurrence. Whilst environ- volume of air.
mental monitoring can provide important information A second advantage is with improved detection and
relating to changing trends over time, the most widely accuracy. Because the detection of micro-organisms is
used methods are growth based and thus take some time not reliant upon conventional culture-based microbiolog-
to produce a result. The incubation times required vary; ical methods, the data relating to microbiological activity
however, there is no time shorter than 2 days that will are more representative than conventional growth-based
give a meaningful assessment of the level of bacterial con- methods that are capable of detecting. This relates to the
tamination. Moreover, most regimes require far longer phenomena of the so-termed ‘viable but nonculturable’
incubation times, especially those which use a general micro-organisms (Weichart 1999), which is a reference to
purpose agar and dual incubation for the assessment of micro-organisms that will not grow on culture media, yet
both bacteria and fungi (Arduino et al. 1991). retain metabolic activity and viability. Moreover, there

2 Journal of Applied Microbiology © 2014 The Society for Applied Microbiology


T. Sandle et al. Application of rapid microbiological methods

are organisms that might grow on media under one set by a certain wavelength of light, these compounds emit
of conditions but will not do so because they have an intrinsic fluorescence signal that serves as a marker for
become stressed or because they are sublethally damaged. biological content of the sample. The assessment of
The adoption of rapid microbiological methods within NADH and riboflavin is necessary for the detection of
the pharmaceutical sector has gained regulatory and vegetative bacteria and fungi, whereas the detection of
industry acceptance (Miller 2008). Examples include the DPA is required when assessing the presence of endosp-
2013 revision of the Parenteral Drug Association Techni- ores. Fluorescence detection is suitable for the detection
cal Report 33, which provides revised guidance on equip- of microbial contamination as it has a high sensitivity
ment and method validation and regulatory expectations and requires a relatively short collection time (Bjerner
(PDA 2013). In keeping with regulatory concerns, the et al. 2012). With the generated bio-counts, because more
instruments, by collecting data in ‘real time’ are in keep- than one micro-organism passing through the instrument
ing with the driver initiated by the US Food and Drug would be counted as ‘one count’, one bio-count does not
Administration for pharmaceutical manufacturers to uti- necessarily correlate with one micro-organism.
lize the concept of process analytical technology (PAT), For the study, an EU GMP Grade D area (14644 class 8
where measurements are made during processing (FDA in operation) was selected. The cleanroom had a surface
2004). area of approximately 200 square metres and it was used
The dilemma discussed in the opening section of this for the processing and preparation of equipment. The
paper—the unknown risks associated with using clean- occupancy level was relatively high (up to ten staff) and
rooms for processing without an assessment of the the area included a wash-bay. Samples were taken during
microbial levels—can be addressed with rapid microbio- the summer of 2013. The cleanroom was located in a phar-
logical methods. An example of a ‘real-time’ air counter maceutical facility based in the south-east of England.
is the Azbil BioVigilant IMD-Aâ system. The instrument With the operation of the IMD-Aâ, the instrument
is detects ‘bio-counts’. These are biological counts, based was located two metres from the area where vessels are
on the detection of microbial metabolites and chemical handled and personnel activity was greatest; this was
compounds, enabling the counting of micro-organisms within the theoretical sampling range of the instrument
(Miller et al. 2009a). and was at the most representative area within the clean-
This paper describes a study that was undertaken to room. For sampling, a continuous monitoring mode was
assess whether the IMD-Aâ system could be used to com- used and the sample size was set at one cubic foot of air.
pare the microbiological levels in the air of a cleanroom The particle sizes measured were ≥05 and ≥50 lm,
after it had been reinstated following a period of shutdown which met GMP recommendations for particle monitor-
and maintenance with the microbiological levels prior to ing in pharmaceutical cleanrooms.
the shutdown occurring. The aim was to assess whether Over the course of the study, the selected cleanroom
the reconditioned cleanroom was suitable for biopharma- was assessed over different periods of time and under dif-
ceutical processing by drawing this comparison. ferent conditions, for particle counts and bio-counts.
These conditions were as follows:
Materials and methods • Cleanroom assessed during normal operations (pre-
shutdown).
The instrument used for the study was an IMD-Aâ 350
• Cleanroom during shutdown and maintenance.
System. The IMD-Aâ system simultaneously detects the
• Cleanroom during cleaning and disinfection.
size, number and biologic fluorescence of airborne partic-
• Cleanroom returned to normal operations (postshut-
ulates within a 05–15 lm range (Miller et al. 2013).
down).
With the counting of nonviable particles, the counter
functions in a similar way to conventional particle coun- With the conventional particle counting, EU GMP
ters, utilizing Mie scatter (Bohren and Huffmann 2010). Grade C/ISO 14644 class 8 (in operation) limits were
However, the basis of the technology pertaining to the applied (although the cleanroom was graded D, there are
detection of micro-organisms is quite different. Microbial no specified ‘in operation’ limits within EU GMP).
detection is via a laser light source (405-nm) where the The particle levels were (Table 1) as follows:
presence or absence of intrinsic fluorescence determines In terms of setting limits, given that one of the objec-
whether a particle is biologic or inert (Miller et al. tives of the exercise was to assess the bio-counts for the
2009b). This is based on micro-organisms containing cer- preshutdown period and to compare these to the post-
tain metabolites and chemical compounds, such as nico- shutdown period, to assess whether the norm had been
tinamide adenine dinucleotide (NADH), riboflavin restored, it was decided that the preshutdown period
(vitamin B2), and dipicolinic acid (DPA). When excited would provide the limit for acceptability. To determine

Journal of Applied Microbiology © 2014 The Society for Applied Microbiology 3


Application of rapid microbiological methods T. Sandle et al.

Table 1 Table showing particle count limits for the studied clean- remained considerably below the maximum recom-
room mended values as indicated by GMP guidance (in relation
Maximum permitted number of to both mean counts and the highest recorded counts);
particles per cubic metre moreover, the spread of the particle counts was relatively
small (in relation to the standard deviations).
EU GMP Grade 05 lm 50 lm
With the bio-counts, the typical value was 27 bio-
C (in operation) 3 520 000 29 000 counts per minute and the highest recorded bio-count
was 220. These bio-counts had a relatively narrow spread,
as indicated by the standard deviation of 30. Although
whether the limit had been exceeded when the postshut- the exercise did not seek to measure any direct correla-
down period was examined, provided that the bio-counts tion between bio-counts and the numbers or incidents of
between the two periods were within one logarithm (base colony-forming units recovered from conventional envi-
10) of each other, then equivalence would be taken as ronmental monitoring methods, the data from standard
having been met. microbiological monitoring remained in control for the
Over the course of the study, conventional microbiolog- assessment period, with the active air sampler counts
ical monitoring was undertaken at periodic intervals. averaging at 20 CFU m 3.
However, due to the limitations with conventional micro- The assessment of the IMD-Aâ instrument represented
biological air-sampling methods, a direct comparison with the use of novel technology and the cleanroom had not
such methods was not made. Air samplers have design been previously examined using any comparable device;
limitations which affects their ability to capture all of the hence, there was no direct reference point for the bio-
micro-organisms within the air, and, even when such counts. Nonetheless, the bio-counts obtained can be
micro-organisms are captured, damage or desiccation can regarded as relatively low. Moreover, the collected data
occur which can prevent them from being recovered via provided a suitable benchmark for the subsequent com-
the agar medium (Jensen et al. 1994). On this basis, there parative analysis, as discussed below.
was little value in seeking a direct comparison between via- Exploring the basis of this benchmark further, with
ble air samplers and the IMD-Aâ method. both particle counts and bio-counts, there was some vari-
ation according to different activities taking place within
the cleanroom at different time points. For some of the
Results
different activities, the data were trended. Figures 1 to 2
The data collected from the different monitoring condi- display selected data plots for the preshutdown period of
tions were examined, and the analysis is presented below normal operations.
in tabular and graphical formats. The above chart displays some data collected on Day 1
for a selected period of time (15:10 h to 18:46 h). The per-
iod chosen is intended to be illustrative of the fluctuation
Cleanroom during normal operations
that can occur with both particle counts and bio-counts.
To assess ‘normal operations’, the cleanroom was exam- The chart shows that four peaks in relation to 05 lm par-
ined more than a 4-day period to achieve a baseline read- ticle counts occurred. One of these coincided with a rela-
ing. During this period, the cleanroom was used more tively high peak for bio-counts (at around 16:40 h).
than a 24-h period, with a medium-level occupancy These peaks coincided with equipment being moved in
(averaging at six members of staff). The activities taking and out of the area; in addition, at around 16:40 h, there
place primarily consisted of the processing of equipment was considerable movement of personnel with six people
vessels (cleaning, drying and storage) and initial formula- being present in the area. From this, the data suggest that
tion of drug products. During this time, equipment was the personnel-related activities have some association
moved in and out of the room and wash bays were used. with bio-counts and that with the equipment, there was a
The particle counts and bio-counts produced from this ‘nonviable’ particle count association but not a bio-count
period of the study have been summarized in Table 2. association. These events were affirmed through the use
The table provides the mean counts per minutes, the of a video camera.
standard deviation (as an indicator of variance) and the Figure 2 illustrates a second period of the preshutdown
highest recorded counts in relation to both particle operations from Day 2 of monitoring (time period B). The
counts and bio-counts. To aid the review, the data have chart has the same scale as per Fig. 1, for comparative
been subdivided into periods of days. purposes.
Table 2, which summarizes data collected more than a Figure 2 indicates that during the period of time
96-h period, indicates that the nonviable particle counts assessed, the particle counts and bio-counts remained

4 Journal of Applied Microbiology © 2014 The Society for Applied Microbiology


T. Sandle et al. Application of rapid microbiological methods

Table 2 Data summary for the assessment of the cleanroom during normal operations, preshutdown

Mean counts/minute Standard deviation Maximum count/minute

≥05 lm Bio- ≥50 lm ≥05 lm Bio- ≥50 lm ≥05 lm Bio- ≥50 lm


Day Room condition count count count count count count count count count

1 Normal operations, preshutdown 143 38 5 252 46 4 1372 268 22


2 Normal operations, preshutdown 63 22 3 78 22 4 556 157 23
3 Normal operations, preshutdown 119 34 5 222 33 4 2151 280 30
4 Normal operations, preshutdown 37 12 2 75 19 4 604 175 29
Mean counts for days 1–4 91 27 4 157 30 4 1171 220 26

Day 1 preshutdown, selected time period A


1400

1200

1000
Count

800

600

400
Figure 1 Chart displaying a selected time
period (A) from the preshutdown period, Day 200
1. The graph shows counts per minute 0
(equivalent to counts per cubic foot). ( )
15
15 0
15 0
15 0
15 0
16 0
16 0
16 0
16 0
16 0
16 0
17 0
17 0
17 0
17 0
17 0
17 40
18 0
18 0
18 0
18 0
18 0
:1
:2
:3
:4
:5
:0
:1
:2
:3
:4
:5
:0
:1
:2
:3
:
:5
:0
:1
:2
:3
:4
≥05 lm; ( ) Bio and ( )≥50 lm.

0
Day 2 preshutdown, selected time period B
1400

1200

1000

800
Count

600

400
Figure 2 Chart displaying a selected time
200
period (B) from the preshutdown period, Day
2. The graph shows counts per minute 0
14:11
14:21
14:31
14:41
14:51
15:01
15:11
15:21
15:31
15:41
15:51
16:01
16:11
16:21
16:31
16:41
16:51
17:01
17:11
17:21
17:31
17:41
17:51
18:01
18:11
18:21
18:31
(equivalent to counts per cubic foot). ( )
≥05 lm; ( ) Bio and ( )≥50 lm.

relatively low. The main bio-count peak, at around particulate air (HEPA) filters which made up the air-han-
15:55 h, coincided with operators moving a large vessel. dling units (AHU). The maintenance works included the
Thus, as with the Fig. 1 data, the primary trigger for the following:
bio-counts was the degree of operator movement.
• The removal of HEPA filter grilles.
The overall assessment from the preshutdown period
• The testing of HEPA filters using an aerosol smoke test.
was that the impact of the standard activities was mini-
• The deactivation of the AHU and the replacement of
mal in terms of both particle and bio-count generation
a HEPA filters.
and that the cleanroom remained in a state of control.
• Low level of activity following HEPA filter replace-
ment (rooms not clean, although with minimal per-
Cleanroom during shutdown and maintenance sonnel activity).
On Day 5 of the study, the cleanroom was formally shut- The removal, testing and replacement of the HEPA fil-
down and handed over to the maintenance department ters took 3 days to complete (this was not a continuous
for the testing and replacement of the high-efficiency activity). This was followed by a period of low level of

Journal of Applied Microbiology © 2014 The Society for Applied Microbiology 5


Application of rapid microbiological methods T. Sandle et al.

activity, which lasted for fourteen days (this long period The Day 7 activity, where the AHU was deactivated and
of time was due to other works being undertaken the HEPA filter replaced, led to a large rise in both particle
throughout the facility which resulted in the cleanroom and bio-counts. This is unsurprising because the clean-
not being needed and hence cleaning and disinfection room was no longer supplied with ‘clean’ (low particulate)
was not performed until the seventeenth day of the shut- air. Furthermore, the air was not circulating turbulently
down). neither was the air being removed from the cleanroom at
The particle and bio-counts relating to the different a rapid rate. Under these conditions, the bio-counts aver-
activities across the shutdown have been summarized in aged at their highest level (at 730) and peaked at 2665.
Table 3. Figure 4, below, illustrates the time point when the
The data displayed in Table 3 indicate that relatively AHU was deactivated.
higher particle and bio-counts were obtained compared With Fig. 4, the AHU was deactivated at 13:31 h. The
with the preceding period of normal operations. During chart has the same scale as that of Fig. 3 to illustrate the
the period of normal operations, the level of bio-counts increase in bio-counts (and for this reason, the particle
had averaged at 27 counts per minute. With the first type counts rise above the chart scale). The bio-counts showed
of maintenance activity (the removal of the outer cover- an initial increase to the mid-2000s and then reduced
ing grille for the HEPA filters), the mean bio-counts slightly to the 1000-level and then remained around this
more than doubled to a value of 58 and peaked at 2555 level whilst the AHU was offline. The spike with the bio-
during the main part of the activity. This high count is counts related to an increase in traffic from process oper-
shown in Fig. 3, below. ators who entered the room at 14:15 h and remained in
The main peak on the chart, for the selected time per- the room until 15:07 h. Again, bio-counts tallied with
iod (C), relates to the removal of the grille to the HEPA human activity.
filter and its subsequent replacement. During this time, With the bio-counts recorded during this part of the
two maintenance staff were present in the area. The activ- shutdown period, when maintenance work was occurring,
ity was of a relatively short duration (under 6 min); nev- these counts were considerably higher than those
ertheless, it led to both the highest recorded bio-counts recorded during the period of normal operations. Con-
and the highest recorded 05 lm size particle counts. sidering the ‘normal’ baseline, the maintenance period
For the Day 6 activity, the period immediately follow- could, therefore, be considered as presenting a greater
ing smoke testing, the bio-counts and particle counts risk to the cleanroom and its operations.
remained relatively low and, in terms of mean counts, Following the re-activation of the AHU, the cleanroom
close to the levels recorded during standard operations. was closed and no personnel entered to room. This
Here, the air-handling unit would be running at normal period produced the lowest level of particle and bio-
capacity, and any residual smoke used for HEPA filter counts yet recorded. Therefore, despite the room surfaces
leak testing would have been removed. Monitoring was being ‘unclean’, the restoration of HEPA filtered air along
not undertaken during the use of the smoke because this with other cleanroom parameters, such as adequate room
could have caused damage to the instrument. air changes and air flow, the levels of particles and

Table 3 Data summary for the assessment of the cleanroom during the shutdown period

Mean Standard Maximum


counts/minute deviation count/minute

≥05 lm Bio- ≥50 lm ≥05 lm ≥50 lm ≥05 lm ≥50 lm


Day (s) Room condition count count count count Bio-count count count Bio-count count

5 Works included the 164 58 9 702 252 35 7098 2555 350


removal of HEPA filter grilles
6 Works included testing of 51 21 5 96 25 7 695 169 46
HEPA filters using an aerosol
smoke test
7 Operations included the 1720 730 116 2202 647 88 12 682 2665 403
deactivation of the AHU and
the replacement of a HEPA filters
8–17 Shutdown period 9 6 1 16 10 3 96 55 14
continued, minimal room use
Mean for shutdown period 486 204 33 754 234 33 5143 1361 203

6 Journal of Applied Microbiology © 2014 The Society for Applied Microbiology


T. Sandle et al. Application of rapid microbiological methods

Day 5 - shutdown - AHU grille removal (time period C)


7000
6000
5000

Counts
4000
3000
Figure 3 Chart displaying the first work 2000
undertaken during the shutdown: removal of
the HEPA filter grills. The graph shows counts 1000
per minute (equivalent to counts per cubic 0
foot). ( ) ≥05 lm; ( ) Bio and

12
12 9
12 5
12 1
12 7
12 3
12 9
13 5
13 1
13 7
13 3
13 9
13 5
13 1
13 7
13 3
13 9
14 5
14 1
14 7
:1
:2
:3
:3
:4
:4
:5
:0
:0
:1
:1
:2
:3
:3
:4
:4
:5
:0
:0
:1
( )≥50 lm.

3
Day 7 - shudown AHU deactivated (time period D)
7000
6000
5000
Counts

4000
3000

Figure 4 Chart displaying the time prior to 2000


and at the point of the deactivation of the 1000
cleanroom AHU. The graph shows counts per 0
minute (equivalent to counts per cubic foot).
13

13

13

13

13

14

14

14

14

14

14

14

14

14

14

15
:3

:3

:4

:5

:5

:0

:0

:1

:2

:2

:3

:3

:4

:5

:5

:0
( ) ≥05 lm; ( ) Bio and ( )≥50 lm.
2

2
micro-organisms remained low. These data arguably sup-
Cleanroom returned to normal operations
port the general assumption in the literature that people
are the main sources of microbial carrying particles within The final stage of the evaluation was to assess the clean-
cleanrooms; moreover, personnel movement would also room postcleaning, following the handover of the clean-
stir up some settled particles (McIntosh et al. 1978). room back to the process area staff. The period following
handover was monitored for 4 days with the objective of
assessing whether the counts were similar to those of the
Cleanroom during cleaning and disinfection
preshutdown period. These data from the postshutdown
The third part of the exercise was an assessment of the period are illustrated in Table 5, below.
particle and bio-count levels generated during the clean- The data from the postshutdown period compared rela-
ing (use of detergent) and disinfection of the cleanroom. tively favourably to the preshutdown period, although the
This was a standard exercise undertaken following a per- bio-counts were notably higher at the outset. This suggests
iod of production downtime to restore a cleanroom for that the area required a ‘rest’ period following the comple-
production use. The cleaning exercise was undertaken tion of the cleaning and disinfection prior to the com-
more than a 2-day period. The data collected during this mencement of standard operations. By the second day of
activity are summarized in Table 4. the postshutdown period, the counts had lowered and by
The data displayed in Table 4 indicate that the bio- the end of the assessment, the data were within the normal
counts were around twice the value seen when the clean- pattern established prior to the shutdown.
room was running for standard operations (as shown in The changing bio-counts for each of the main events
Table 2). However, the bio-counts were at a much lower are summarized in Fig. 5:
level than those seen for the maintenance works. The Figure 5 displays the changing bio-counts for the dif-
inference from these data is that, whilst the cleaning ferent periods of activity; however, the data are relatively
activity itself causes a relatively high level of particle gen- distorted by the AHU going offline. When this event is
eration, the operation of the AHU and clean air supply removed from the data set, the relative values appear as
provide a strong degree of protection for the cleanroom presented in Fig. 6.
and AHU operation is relatively effective at removing The presentation of the data in Fig. 6 allows a clearer
microbial carrying particles. comparison to be made of the relative bio-counts. Here,

Journal of Applied Microbiology © 2014 The Society for Applied Microbiology 7


Application of rapid microbiological methods T. Sandle et al.

Table 4 Data summary for the assessment of the cleanroom during cleaning and disinfection

Mean counts/minute Standard deviation Maximum count/minute

≥05 lm ≥50 lm ≥05 lm ≥50 lm ≥05 lm ≥50 lm


Day Room condition count Bio-count count count Bio-count count count Bio-count count

18 Detergent cleaning 390 43 6 656 63 8 3011 364 43


19 Disinfection cleaning 315 50 8 630 109 10 3770 1281 59
Mean for cleaning period 352 47 7 643 86 8 3391 823 51

the largest rise in bio-counts is between the end of clean- a degree of confidence that the likelihood of contamina-
ing and the point of room handover. A day later, the val- tion would be low, provided that the data from the two
ues returned to the pattern seen prior to the shutdown periods were comparable.
period. Given that the rudimentary criterion to compare The exercise also provided interesting data about a
the period prior to and after the shutdown was a one log cleanroom as it is taken through different phases of use.
difference, the bio-counts for these two periods easily fall Events such as removing HEPA filter grilles and turning
within this criterion. Thus, as a crude measure, there is off an air handing unit are significant activities which
no noticeable difference and the cleanroom—postshut- lead to a high generation of biological particles and, in
down—can be considered to have similar environmental showing an increase over the baseline, would be regarded
conditions compared with the preshutdown period. as presenting a microbiological risk to the cleanroom. As
a signal of confidence, the restoration of the correctly
functioning air-handling unit returns a cleanroom to a
Discussion
level of low biological activity relatively quickly (at least
The study showed that the IMD-Aâ system was capable for the distributed air, surfaces will probably remain con-
of detecting, sizing and enumerating particles from con- taminated until cleaning and disinfection has been under-
tinuous air sampling and that the instrument was addi- taken).
tionally able to determine whether these collected Another interesting observation related to the activity
particles were biological or nonbiological in origin and of cleaning and disinfection of a cleanroom by opera-
thus able to obtain a measure of the free-floating micro- tors using the classic mop and bucket methods. Whilst
organisms and micro-organisms attached to microbial this activity is more likely to generate nonviable parti-
carrying particles in the air. cles than bio-counts, the time taken for the area to
In terms of the exercise, the instrument was able to settle down and return to the norm following cleaning
provide an indication of cleanroom conditions during a took longer than the time required to return the area
period of normal operation and allow a comparison to to a ‘safe’ level following the reactivation of the air-
be made of conditions following extensive maintenance handling system. This adds further weight to the dis-
works and reinstatement activity. With this exercise, the cussions that people present the biggest risk factor to
bio-counts indicated that there was a similar state of cleanrooms and also that the activity of cleaning
microbiological control postshutdown as compared with appears to re-deposit biologic and nonbiologic particles
its condition prior to the shutdown. On this basis, a into the airstream.
cleanroom user could use the IMD-Aâ instrument to An important consideration for the study was the zone
assess the risk of an area following maintenance and have within the cleanroom where the IMD-Aâ instrument was

Table 5 Data summary for the assessment of the cleanroom during normal operations, postshutdown

Mean counts/minute Standard deviation Maximum count/minute

≥05 lm ≥50 lm ≥05 lm ≥50 lm ≥05 lm ≥50 lm


Day Room condition count Bio-count count count Bio-count count count Bio-count count

20 Normal operations, postshutdown 619 80 11 1097 162 15 5505 2005 110


21 Normal operations, post 122 28 7 153 29 7 1311 218 70
22 Normal operations, post 187 46 7 324 50 6 1879 277 33
23 Normal operations, post 22 10 3 36 14 5 371 138 49
Mean for days 20–23 238 41 7 402 64 8 2267 660 66

8 Journal of Applied Microbiology © 2014 The Society for Applied Microbiology


T. Sandle et al. Application of rapid microbiological methods

Bio count mean values/day


800
700
600

Bio count
500
400
300
200
100
0

in nt c wn
ng

et t-sh line
1

Po n c ing

Po hut ing

Po hut n 1

td 2
hu n 3

4
d
ok ove
y
ay

do ay

hu n

n
AH etsi

o
Pr utd da

da

w
ow

ow
-
td
f
d

ea

ea
of

do

do
Sm em

u
t
sh wn

sh wn

sh wn

td
l

l
e

U
w

r
o

s
Pr utd

td

io

-s

-s

-s

-s
s
ge
le
ut

po

ct
u

st

st

st

st
ril
Figure 5 Chart showing changes for mean sh

er

fe

Po
rg

ng
e-

e-

e-

e-
bio-counts during the study (selected periods

te

is
ni
Pr

Pr

fil

D
ai
em
of operation). The graph shows counts per

A
EP

R
minute (equivalent to counts per cubic foot).

H
( ) Bio-count. Cleanroom event

Bio-count/mean values per day (excluding AHU)


90
80
70
60
Bio count

50
40
30
20
10
0
n
g

ng
1

1
2

hu n 2

hu n 3

4
ng oke ed

w
in

in
y

n
y

n
ni
o
v

ts

n
da

da

da

da

ow

ow

ow

ow
d
Sm mo

ea

ea
te

ut

td
n

td

td

td
n

cl

Po n cl
h
re
w

hu

hu
-s

nt
do

do

do

do

io
st

-s

-s

-s

-s
e
le
ut

ut

ut

ut

ct
po

Figure 6 Chart showing changes for mean


st

st

st

st
ril

er
sh

sh

sh

sh

fe

Po

Po

Po
rg

et

in
e-

e-

e-

e-

bio-counts during the study (selected period


te

is
ni
Pr

Pr

Pr

Pr

fil

D
ai

of operation). The graph shows counts per


em
A
EP

minute (equivalent to counts per cubic foot).


H

( ) Bio-count. Cleanroom event

placed. In a nonunidirectional flow environments, local would be the case with a conventional particle counter);
airflow dynamics, particle settling velocity and sampler on this basis, the counter was sited within a few metres
flow rate can each affect the size of the ‘zone’ and how of the intended target area.
quickly, or indeed whether, a particle originating some Whilst the data gathered were of great interest and there
distance away from the sampling instrument may be were clearly identifiable trends, the study had some limita-
detected. Thus, the area selected should be representative tions. Firstly, although the study ran for several weeks, the
of the cleanroom and ideally has some corresponding cleanroom was studied on just one occasion. Further stud-
basis to the primary risk to the product or process; this ies would be ideally required to demonstrate the repeatabil-
could be, for example, near open processing or close to ity of the results. Nonetheless, the results obtained were
where the highest level of personnel activity is. The risk is consistent across the time window evaluated.
with selecting an area that is unrepresentative of the main Secondly, with any study using novel technology, there
personnel or equipment activities, thus leading to an will always be a degree of uncertainty with the measure-
underestimation of the particles generated. There is a ments obtained. With the case of bio-counts, the correla-
degree of uncertainty as to the size of the sampling zone tion with ‘colony-forming units’ and microbial carrying
that an instrument such as the IMD-Aâ can sample (as particles could not be shown directly, and microbial levels

Journal of Applied Microbiology © 2014 The Society for Applied Microbiology 9


Application of rapid microbiological methods T. Sandle et al.

could instead only be inferred as ‘low’ or ‘high’ depending particles. These could falsely increase the number of
upon the degree of biological activity. Nevertheless, a level assumed viable micro-organisms (Eaton et al. 2012). This
of uncertainty exists with conventional environmental potential concern can be overcome, to an extent, through
monitoring methods and given that the IMD-Aâ is capable the collection of a large number of samples and by set-
of continuous monitoring, unlike an active air sampler, this ting a baseline. With the study described in this paper,
level of uncertainty can be assumed to be less with the the materials that might cause this effect will have been
IMD-Aâ system (despite the uncertainty not being quanti- present within the cleanroom before it was shut down for
fiable). In time, further studies will build upon the study maintenance and when it was restarted; thus, by under-
outlined in this paper and this will further understanding standing the baseline, the effect of nonmicrobial fluoresc-
of the significance of different levels of bio-counts. ing particles can be captured within the normal operating
It follows that the most difficult aspect when evaluating a bio-levels within the cleanroom. Thus, the presence of
system such as the IMD-Aâ is in understanding the rela- any nonmicrobial fluorescing particles does not detract
tionship between bio-counts and micro-organisms. This is from the finding that the pre- and postshutdown periods
especially so because it cannot be quantified how many were similar.
micro-organisms are present in cleanroom air but not These limitations, whilst noteworthy, did not detract
detected by conventional methods, so direct parallels can- from the study’s findings. Based on the exercise, real-time
not be made. Comparative analysis can be undertaken using monitoring of airborne microbial concentrations within
a bioaerosol chamber and a mixed population of micro- cleanrooms and controlled environments provides con-
organisms; however, this is somewhat artificial, given that siderable advantages. This is primarily as a means to
such chambers cannot mimic cleanroom air patterns, and assess whether an area is ‘safe’ or ‘clean’, thereby avoiding
for most cleanroom users, the interest will be upon using the issues arising from the delay in obtaining results
the instrument in situ (as this study attempted). caused by the slow growth of micro-organisms measured
A way around this, and an area for future study, would by conventional sampling methods. For assessing such
be by building up voluminous data and showing that very activities, the device will prove useful for microbiologists.
high or numerous excursions with conventional microbio- The instrument can be run continuously and provided
logical methods are reflected in the recording of high bio- sufficient data can be gathered, so that operational limits
counts. More exact causality is not possible, given that can be set, and the instrument offers advantages over tra-
comparing any two instruments—such as comparing the ditional microbiological methods.
IMD-Aâ system with an active air sampler—is difficult. As with describing any novel technology, the results of
This is because the two instruments would sample different this paper add to a developing database of studies. Fur-
volumes of air (within which micro-organisms are not nor- ther work will be required to enhance this, including a
mally distributed). Furthermore, the instruments would replication of the study outlined in this paper as well as
have different ranges, be calibrated with different d50 (col- other studies, such as an analysis of different types of
lection efficiency) values; they would ‘register’ microbial cleanrooms, including unidirectional air flow worksta-
activity through different means (Whyte et al. 2007). This tions used for the aseptic filling of products.
latter point, about different means of detection, is likely to
be the reason for the largest variation in results should the Acknowledgements
IMD-Aâ be compared with an active air sampler.
Another limitation is with not knowing what species of The authors would like to thank Biovigilant and Techno-
micro-organisms the bio-counts relate to. For this study, path for the use of the IMD-Aâ instrument and for
this was not a direct concern. Nonetheless, in most situa- advice during the study. This paper has been written
tions, the microbiologist needs to know this; therefore, at independently of these companies.
present, it is unlikely that real-time technology will wholly
replace growth-based method. The technology remains a
Conflict of Interest
destructive method in that the instruments will indicate
whether micro-organism is present, but they cannot not No conflict of interest declared.
determine the species.
A third limitation, remaining with the concept of
References
uncertainty of measurement, is that some of the bio-
counts may not relate to micro-organisms. Various mate- Arduino, M.J., Bland, L.A., Aguero, S.M. and Favero, M.S.
rials, such as clothing fibres and potentially some alcohol (1991) Effects of incubation time and temperature on
sprays, used within the cleanroom can potentially microbiologic sampling procedures for hemodialysis fluids.
disperse unknown numbers of nonmicrobial fluorescing J Clin Microbiol 29, 1462–1465.

10 Journal of Applied Microbiology © 2014 The Society for Applied Microbiology


T. Sandle et al. Application of rapid microbiological methods

Asefzadeh, S., Yarmohammadian, M.H., Nikpey, A. and optical spectroscopy technology for the continuous and
Atighechian, G. (2013) Clinical risk assessment in real-time environmental monitoring of viable and
intensive care unit. Int J Prev Med 4, 592–598. nonviable particles. Part I: review of the technology and
Bjerner, G., Kene, V., Akerlund, E. and Andersson, K. (2012) comparative studies with conventional methods. In
Real-time microbiological air monitoring. Eur J Parenter Encyclopedia of Rapid Microbiological Methods ed. Miller,
Pharm Sci 17, 52–55. M.J. pp. 213–234. Volume 4. USA: PDA and Davis
Bohren, C.F. and Huffmann, D.R. (2010) Absorption and Healthcare International Publishing.
Scattering of Light by Small Particles. New York: Wiley- Osimani, A., Aquilanti, L., Tavoletti, S. and Clementi, F.
Interscience. (2013) Microbiological monitoring of air quality in a
Eaton, T., Davenport, C. and Whyte, W. (2012) Airborne university canteen: an 11-year report. Environ Monit Assess
microbial monitoring in an operational cleanroom using 185, 4765–4774.
an instantaneous detection system and high efficiency Oxborrow, G.S., Fields, N.D., Puleo, J.R. and Herring,
microbial samplers. Eur J Parenter Pharm Sci 10, 339–351. C.M. (1975) Quantitative relationship between
FDA (2004) Guidance for Industry: PAT - A Framework for airborne viable and total particles. Health Lab Sci 12,
Innovative Pharmaceutical Development, Manufacturing and 47–51.
Quality Assurance. Food and Drug Administration, PDATechnical Report No. 33, Revised (2013) Revised 2013
Rockville: MD, USA. (TR 33) Evaluation, Validation and Implementation of
International Conference on Harmonisation (ICH). (2011) Alternative and Rapid Microbiological Methods. Bethesda,
Quality Risk Management (ICH Q9), EMA/INS/GMP/ MD, USA: Parenteral Drug Association.
79766/2011. Raval, J.S., Koch, E. and Donnenberg, A.D. (2012)
Jensen, P.A., Lighthart, B., Mohr, A.J. and Shaffer, B.T. (1994). Real-time monitoring of non-viable airborne particles
Instrumentation used with microbial bioaerosol. In correlates with airborne colonies and represents an
Atmospheric Microbial Aerosols: Theory and Applications acceptable surrogate for daily assessment of cell-
eds Lighthart, B. and Mohr, A.J. pp. 226–284. New York, processing cleanroom performance. Cytotherapy 14,
NY: Chapman & Hall. 1144–1150.
Li, C.-S. and Hou, P.-A. (2003) Bioaerosol characteristics in Sandle, T. (2012a). Environmental Monitoring: a practical
hospital clean rooms. Sci Total Environ 305, 169–176. approach. In Environmental Monitoring: A Comprehensive
McIntosh, C., Lidwell, O.M., Towers, A.G. and Marples, R. Handbook ed. Moldenhauer, J. Volume 6, pp. 29–54. River
(1978) The dimensions of skin fragments dispersed into Grove, USA: PDA/DHI.
the air during activity. J Hyg 81, 471–479. Sandle, T. (2012b) Real-time counting of airborne particles
Miller, M.J. (2008). Rapid Microbiological Methods. In and microorganisms: a new technological wave? Clean Air
Microbiology Pharmaceutical Manufacturing ed. Prince, R. Containment Rev 9, 4–6.
pp. 171–221. Vol. 2, Bethesda, MD: DHI/PDA. Sandle, T. and Saghee, M.R. (2013). Cleanroom certification
Miller, M.J. (2012) New year, old challenges. Eur Pharm Rev and ongoing compliance. In Cleanroom Management in
17, 2–5. Pharmaceuticals and Healthcare eds Sandle, T. and Saghee,
Miller, M.J., Walsh, M.R., Shrake, J.L., Dukes, R.E. and Hill, M.R. pp. 169–184. Passfield, UK: Euromed
D.B. (2009a) Evaluation of the BioVigilant IMD-Aâ, a Communications.
novel optical spectroscopy technology for the continuous Tham, K.W. and Zuraimi, M.S. (2005) Size relationship
and real-time environmental monitoring of viable and between airborne viable bacteria and particles in a
nonviable particles. Part II: case studies in environmental controlled indoor environment study. Indoor Air 15(Suppl
monitoring during aseptic filling, intervention assessments 9), 48–57.
and glove integrity testing in manufacturing isolators. PDA Weichart, D. (1999). Stability and survival of VBNC cells -
J Pharm Sci Technol 63, 258–282. conceptual and practical implications. In Proceedings of the
Miller, M.J., Lindsay, H., Valverde-Ventura, R. and O’Conner, 8th Symposium on Microbial Ecology ed. Bell, C.B.-G.
M.J. (2009b) Evaluation of the BioVigilant IMD-Aâ, a Halifax, Canada: Atlantic Canada Society for Microbial
novel optical spectroscopy technology for the continuous Ecology.
and real-time environmental monitoring of viable and Whyte, W. and Hejab, M. (2007) Dispersion of particles and
nonviable particles. Part I. Review of the technology and microorganisms from people. Eur J Parenter Pharm Sci 12,
comparative studies with conventional methods. PDA J 39–46.
Pharm Sci Technol 63, 244–257. Whyte, W., Green, G. and Albisu, A. (2007) Collection
Miller, M.J., Lindsay, H., Valverde-Ventura, R. and O’Connor, efficiency and design of microbial air samplers. J Aerosol
M.J. (2013) Evaluation of the BioVigilant IMD-A, a novel Sci 38, 101–114.

Journal of Applied Microbiology © 2014 The Society for Applied Microbiology 11

You might also like