2200 Calibration Procedure

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 1

Calibration Procedure

1. Press menu key and hold for 4 seconds

2. Display – Batch No. (Curr 000001)

3. Press MR/MC Keys two times.

4. Display – Press enter for authorized menu.

5. Press menu key

6. Display – Enter Password XXXXXXXXXXX

7. Press Esc, Menu, Start, MR/MC, M+, Start.

8. Display – Wt Calibration.

9. Press Menu Key.

10. Display – Calibration by weight.

11. Press Menu Key, Display – Zero Calibration EX. curr. 052775

12. Press menu key, Display – Ent for Zero Calibration EX. 052781.

13. Press menu key, Display – Cal point 2 Ex.007938, Press menu key.

14. Display – Place wt. & Enter EX.0007939

15. Put some weight on platform say 20.00kg, Press menu key.

16. Display – Place wt & Enter EX.0000.00

17. Set for 0020.00kg using numeric keys 0 to 9.

18. Press menu key, Display – Calibration by Edit.

19. Press ESC key 2 times, Save display.

20. Press menu key.