Professional Documents
Culture Documents
BME4322 Week3-6 UltrasoundHardware 31032022
BME4322 Week3-6 UltrasoundHardware 31032022
ULTRASOUND TECHNIQUES
ULTRASOUND HARDWARE
2
Ultrasound imaging is based on the ‘pulse-
echo’ principle
𝑐𝑡
𝑑=
2
3
A MODE
Lmax
Amplitude
Gain
Amplitude
Amplitude
5
Time (depth)
Amplitude (A) mode ultrasound displays the
ultrasound echoes along one beam, or ‘A Line’
• Ultrasound scanners assume c = 1540 m/s to locate reflections.
• Due to attenuation in tissue, echoes from similar reflectors will have
different amplitudes depending on their depth.
• For tissue imaging, it is desirable to remove this depth dependence so
that, for example, an image of the liver will appear uniform.
• This is achieved using time gain compensation (TGC), which is a gain
applied to the radiofrequency (RF) amplifier and increases with time
after the pulse is transmitted.
• A-mode places a restriction on pulse repetition frequency such that
enough time can elapse before the next pulse is transmitted for the
echo from the maximum imaging depth of penetration Lmax to return.
Amplitude (A) mode ultrasound displays the
ultrasound echoes along one beam, or ‘A Line’
2L max
TP =
c
7
Brightness (B) mode uses each individual
echo strength to build up a 2D image
Amplitude
More reflective structures appear brighter
Time (depth)
Amplitude
Time (depth)
Amplitude
Time (depth)
𝑐
PRF =
2𝑑
• The frame rate, or number of images produced per
second, is then dependent on the number of scan lines
needed to make up the B-mode image :
!
Frame rate = "#$
9
Block diagram of a typical ultrasound imaging
system and typical operator controls
Beam Signal
Transducer
former processing Image
Produces Display
Focus the Minimize memory
ultrasound pulse
pulse artifacts
Control Function
Output power Increases pulse amplitude (and associated
energy deposited in patient)
Receiver gain Increases size of received signal (but also
noise)
Time gain compensation Compensates for non-uniform acoustic data
Focusing Improves lateral resolution
Frame rate / line density / Product is constant, illustrating important
field of view trade-off
10
The transducer
11
An ultrasound transducer is composed of three
main parts
• A pulse of sound waves is
transmitted into the body, and
the echo is detected with the
time of flight used to determine
the depth of the reflector that
generated the echo.
• Ultrasound waves are detected
using a transducer made from a
piezoelectric material.
• Damping material determines
bandwidth.
• Acoustic lens focuses generated
waves and matching layer
maximizes transmission of
produced ultrasound into the
tissue.
Lawrence (2007) Crit Care Med 12
The piezoelectric effect is the appearance of
surface charge in response to applied pressure
~ ~
The application of a short (~ 1μs) pulse of high voltage (~ 150 V) causes PZT
contraction and subsequent vibration at a natural resonant frequency
14
The crystal thickness (l) determines the ultrasound
frequency (f) produced
16
The pulse duration determines ultrasound axial
imaging resolution
• Axial resolution – Ability to discern two closely separated objects in the direction of
beamline.
• Spatial pulse length (SPL) is the product of wavelength and the number of wave cycles
constituting the pulse.
• Axial resolution is equal to ½ spatial pulse length (SPL).
• Axial resolution is limited by SPL.
• The shorter the SPL, the better the axial resolution.
• Typically, the SPL of a pulsed beam of ultrasound is of the order of 1 -3 wavelengths of
the beam.
• Since frequency and wavelength are inversely related, the SPL will decrease with
increasing frequency (e.g., the higher the beam frequency, the better the axial
resolution.
17
Example: Axial resolution
• What is the axial resolution in soft tissue of a 5 MHz transducer
producing a pulse of 3 cycles duration?
• How much better would this be at 10 MHz?
Frequency = 5MHz
1 cycle takes 1/5000000, 3 cycles take 3/5000000
So SPL = 1540 x (3/5000000) = 0.9 mm
Best possible spatial resolution is SPL/2 = 0.45 mm
At 10MHz ???
18
The pulse frequency bandwidth is an important
consideration in transducer design
Long spatial pulse length Short spatial pulse length
Poor axial resolution Good axial resolution
Narrow frequency bandwidth Wide frequency bandwidth
Pulse bandwidth Δ𝑓
~ 1/pulse duration
SPL long SPL short
Energy Energy
f/f0 f/f0
0.8 1 1.2 0.8 1 1.2
20
Impedance matching determines the intensity
of ultrasound emitted Constructive Interference
A
B
A+B
Destructive Interference
Matching 2
Matching 3
Matching
Piezo
(PZT)
(Al)
A
B
A+B
λ λ
L= L=
2 4
The material should have an acoustic impedance of: 𝑍)*+!, = 𝑍%&' + 𝑍+-../0
- Offers the interface between the transducer element and the tissue,
- Minimizes the acoustic impedance differences between the transducer and the patient.
- Optimal impedance matching occurs when the thickness is one quarter of the wavelength – this keeps waves that were
reflected within the matching layer in phase when they exit the layer preventing destructive interference 21
Damping determines the bandwidth of
ultrasound emitted
Matching 2
Matching 3
Matching
Backing (
epoxy)
Piezo
(PZT)
W in
(Al)
Energy
Damped
High sensitivity
Optimal axial resolution
f/f0
0.8 1 1.2 22
Q factor describes how damped an oscillator
is
Matching 2
Matching 3
Matching
Backing (
epoxy)
Piezo
(PZT)
W in
(Al)
𝑓( energy stored per cycle
𝑄= or 𝑄=
Δ𝑓 energy lost per cycle
(𝐵𝑎𝑛𝑑𝑤𝑖𝑑𝑡ℎ)
High Q transducer is lightly damped and has a narrow bandwidth, so good for continuous wave ultrasound.
Low Q transducer is highly damped and has a wide bandwidth, so good for pulse echo imaging ultrasound.
23
The ultrasound beam
24
The simplest ultrasound case is a continuous wave
created by a circular disk of lead zirconate (PZT)
• The pressure (and intensity) field can be calculated using
Huygen’s principle for superposition of wavelets.
25
Considering the axial behaviour along the z
axis normal to the centre of the disk:
Length of near field
r 2 2 a2
a >> λ ⇒ z'max =
λ
Transducer radius a Iz
z
Last axial max
Iz "π %
I0
= sin 2 $
#λ ( )
a2 + z2 − z '
& z
• The ultrasound beam propagates as a longitudinal wave from transducer surface into
propagation medium, and exhibits two distinct patterns : a slightly converging beam out to a
distance speficied by the geometry and frequency of the transducer (the near field), and a
diverging beam beyond that point (the far field).
• The beam is cylindrical in shape close to the transducer (Fresnel zone) then becomes conical
further away (Fraunhofer zone).
27
In the far field regime, the cylindrical
ultrasound beam diverges
Beam divergence angle
θ
Lateral behaviour
θ
On axis:
Iz 1
∝ 2
I0 z
I z 2J1 ( kasin θ )
Off axis: ∝
I0 kasin θ where J1 is a Bessel function of the first kind
29
The ultrasound beam can therefore be
shaped by adjusting transducer geometry
a2
z'max =
λ
a = 1cm, f = 2 MHz, c = 1540 ms−1 ⇒ λ = 0.77mm
z'max = 13cm, θ = 2.7!
a = 1cm, f = 4 MHz, c = 1540 ms−1 ⇒ λ = 0.39mm
z'max = 25.6cm, θ = 1.4!
a = 0.5cm, f = 2 MHz, c = 1540 ms−1
z'max = 3.2cm, θ = 5.4! 30
Lawrence (2007) Crit Care Med
The ultrasound beam can therefore be
shaped by adjusting transducer geometry
a2
z'max =
λ
a = 1cm, f = 2 MHz, c = 1540 ms−1 ⇒ λ = 0.77mm
z'max = 13cm, θ = 2.7!
a = 1cm, f = 4 MHz, c = 1540 ms−1 ⇒ λ = 0.39mm
z'max = 25.6cm, θ = 1.4!
a = 0.5cm, f = 2 MHz, c = 1540 ms−1
z'max = 3.2cm, θ = 5.4! 31
Lawrence (2007) Crit Care Med
The ultrasound beam can therefore be
shaped by adjusting transducer geometry
a2
z'max =
λ
a = 1cm, f = 2 MHz, c = 1540 ms−1 ⇒ λ = 0.77mm
z'max = 13cm, θ = 2.7!
a = 1cm, f = 4 MHz, c = 1540 ms−1 ⇒ λ = 0.39mm
z'max = 25.6cm, θ = 1.4!
a = 0.5cm, f = 2 MHz, c = 1540 ms−1
z'max = 3.2cm, θ = 5.4! 32
Lawrence (2007) Crit Care Med
Side lobes are much weaker than the main
lobe but can be responsible for artefacts
• The strong reflecting surface of the central structure lies in the side lobes of the beam directed at angles
away from the perpendicular. The echoes from the side lobes are misplaced (arrows).
• Side lobe artifacts occur where side lobes reflect sound from a strong reflector that is outside of the
central beam, and where the echoes received are mapped into the main beam, creating unwanted
artifacts.
33
Some terminology
• Transmission Field - the “corridor” along
which transmitted waves travel and the
associated intensity pattern.
• Reception Zone - the region in which a point
source of ultrasound must lie if it is to produce
a detectable signal at the receiving transducer.
• Ultrasound beam - the product of the
transmission field and the reception zone.
• In the simple case of a single disc transducer
the transmission field and reception zone
have the same shape and size.
• The ultrasound beam will have a narrower
width than both the transmission field and the
reception zone.
34
Using a curved transducer or acoustic lens
allows for focusing of the ultrasound beam
a2 a2
z'max = z'max =
λ λ
a d
z z
λd
∝
2a
35
Lawrence (2007) Crit Care Med
The beam former
(and array designs)
36
Transducer arrays can be used to build up B-
mode images • Transducer elements in a linear array that are fired
simultaneously produce an effective transducer width
equal to the sum of the widths of the individual
elements.
• Individual beams interact with constructive and
destructive interference to produce a collimated beam.
• In a linear array, we typically have 256 – 512 discrete
transducer elements, grouped into 15 – 20 adjacent
elements that are activated simultaneously.
• By activating all groups at the same time but introducing a time delay between pulsing of the
groups, we can generate an angled wavefront for beam steering (planar) or we can focus. This is
approach is called phased array ultrasound.
• With a phased-array transducer, the beam is formed by interaction of the individual wave fronts
from each transducer, each with a slight difference in excitation time.
• Minor phase differences of adjacent beams form constructive and destructive wave summations
that steer of focus the beam profile.
38
http://www.olympus-ims.com/en/ndt-tutorials/transducers/
Phased arrays enable electronic beam steering,
removing mechanical complexity
39
http://www.olympus-ims.com/en/ndt-tutorials/transducers/
Phased arrays enable beamforming (focusing)
on the transmit or receive signals
40
Side lobes must be reduced for high quality imaging
• Side lobes are present in the transmission fields and reception zones of arrays or occurs in a
forward direction along the main beam,
• Apodization, a beam-forming process, is used to reduce these:
• On transmission, inner array elements are excited more than outer.
• On reception, different amplifications are performed in each array element.
• Grating lobes can arise from the periodic nature of the linear and phased array designs, emitted
from the array surface at very large angles.
𝑛𝜆 Where m = 1, 2, 3 …, 𝜆 = wavelength, and 𝑥 =
𝑠𝑖𝑛𝜃 = centre − centre distance between elements
𝑥
41
Several different scanning geometries can be
employed
• Annular arrays consist of concentric
annuli of PZT
• Axially symmetric focusing
• Single scan line requires mechanical
scanning
• Curvilinear array
• Similar operation to linear array
• Smaller contact area required
• Intra-cavity probes
• Avoid problems with strong reflections
e.g. air in the bowel
• Intra-vascular probes
• Operate at high frequencies to give high
resolution
42
Several different scanning geometries can be
employed
43
GE Healthcare
The signal processor
44
Pre-processing is typically performed on the
analogue signals
45
http://199.116.233.101/index.php/Ultrasound_Instrumentation
Digital signal processing includes
demodulation and compression
46
http://199.116.233.101/index.php/Ultrasound_Instrumentation
Summary (1)
• Impedance mismatch causes acoustic reflections
• Ultrasound can undergo reflection, refraction, absorption and
scattering in tissue
• Depends on the angle of incidence, size of the object relative to the
ultrasound wavelength, acoustic impedance
• Resolution must be traded against penetration depth because
• High frequency ultrasound provides better spatial resolution
• but high frequency ultrasound is strongly attenuated in tissue
47
Summary (2)
• The pulse echo approach is used to form a brightness (B) mode ultrasound
image
• Transducers are composed of:
• A piezoelectric element to generate and detect acoustic waves
• Matching elements to maximise coupling of acoustic waves to the piezoelectric
element
• Backing material for damping to create a short pulse length and improve axial
resolution
• The finite transducer size results in a far field divergence of the beam and
addition of side lobe imperfections
• Focusing can partially compensate for this
• Transducers are commonly combined into arrays for imaging
48
Next up: ultrasound imaging
modalities!
49