Professional Documents
Culture Documents
Recent Developments For On Line Partial
Recent Developments For On Line Partial
Recent Developments For On Line Partial
net/publication/253763684
CITATION READS
1 34
4 authors, including:
All content following this page was uploaded by Trevor Blackburn on 03 April 2014.
The user has requested enhancement of the downloaded file. All in-text references underlined in blue are added to the original document
and are linked to publications on ResearchGate, letting you access and read them immediately.
RECENT DEVELOPMENTS FOR
ON-LINE PARTIAL DISCHARGE DETECTION IN CABLES
Abstract: The viability of on-line partial discharge (PD) monitoring on high-voltage power
cables is investigated with particular emphasis on methods to overcome the interference problem.
These include the use of a multi-sensor monitoring system, choice of detection frequency and
signal discrimination based on the waveshapes. To reduce the noise level, measurements are
carried out at high frequencies (above 2MHz). An added advantage is better pulse resolution that
enables more accurate locating of the PD source. Comparison is made between two different types
of sensors used for detecting the PD signals.
To filters, amplifiers,
and CDA system 3 phase
HF-CT system
( ratio )
0.6 HFCT
to the clamping sensor [2]. This is done by breaking clamping
0.4
the continuity of the metal ring with a thin rubber
0.2
washer. Fig.3 shows the capacitively coupled inductor
sensor, the washer is where the two ends of the ring 0
1 10 100 1000
are joined. ( metres )
3. CHARACTERISTICS OF PD SIGNALS
Fig.5: Clamping sensor vs HF-CT.
The rise-times of the PD pulses vary in the order of One possible method for interference rejection is by
nano-seconds to several tens of ns. The pulsewidths examining the characteristics of the received signals. It
can extend from a few ns to some hundreds of ns. is expected that discharges would produce signals that
Thus, the bandwidth of the PD signal can extend up to have different waveshapes from those associated with
1 GHz. Conventional electrical PD detection is limited other types of disturbances. To this end, a test circuit
to the lower frequency range (<1MHz) where the noise was set up in the laboratory. As shown in Fig.6, the
level is relatively higher than that in the upper same PD detection circuit as used for on-line
frequency band. Therefore, the high frequency monitoring consists of the HF-CT, high-pass filter,
detection technique used here (>2MHz) would give a amplifier and digital storage oscilloscope.
better SNR.
signal attenuation
( using clamping sensor )
1
0.8
XLPE
( ratio )
0.6
0.4 paper
0.2
0
1 10 100
( metres ) Fig.6: Test circuit setup in the laboratory.
Fig.4: Attenuation in paper and XLPE cables. To obtain the discharge magnitude in pC as specified
in the IEC-60270 or AS1018 Standards, the
Frequency response measurement of the HF-CT sensor conventional PD measurement circuit was also
shows an upper limit of ~300MHz. The upper cut-off included where the signal was picked up through a HV
of the clamping sensor is much higher, at least coupling capacitor Cb. Note that the conventional PD
>500MHz. However, the high frequencies in both detector has a detection bandwidth of 10kHz-300kHz.
paper/oil and extruded cables are quickly attenuated.
This is shown in Fig.4 for the clamping sensor. The Different types of faults were artificially created in the
attenuation is much higher in paper/oil cable. It should cable samples and a collection of PD characteristics in
be noted that the allowable PD level in paper cables is the form of time-domain waveforms and frequency
also much higher. Also, since the HF-CT can measure spectra was obtained. In Fig.7 are shown some of the
the lower frequency components, its signal attenuation typical examples: (a) external corona, (b) external
surface discharges, (c) discharges in cavity in cable the cables. To demonstrate the effect, a detachable
and (d) discharges at the cable termination. discharge source was attached to a live 66kV Yellow-
phase bus bar. Fig.9 shows the disturbances were
picked up by the other two phases. Note the longer
time-base for displaying the pulses in relation to the
50Hz cycle and the relative magnitudes of the signals.
(a)
(b)
(a)
(c)
(d)
0.2µs/hor. div 10MHz/hor. div
50.0mV/vert.div 100mV/vert. div (b)
Fig.7: Typical PD signal characteristics. Fig.8: Waveshape and spectrum of signal recorded (a)
on-site and (b) in laboratory.
There are numerous factors that can influence the
signal characteristics. For example even for a simple Simultaneous recording of signals from sensors
void defect, the PD signal is dependent on the void attached to the three phases can facilitate the
shape, its size, electrical stress, dielectric composition, recognition of external interferences. Fig.10 is an
ageing stage, etc. Also if the PD source is remote, the example of different signal patterns recorded during a
received signal will be distorted since the attenuation site visit. In Fig.10(a), the signals occur on all three
varies with frequency. Therefore, it would be channels exactly at the same time. Furthermore, the
practically very difficult to identify the type of defects waveshapes and their relative magnitudes are very
based on the signal waveshapes. This type of pattern similar. Thus, it can be reasonably concluded that this
recognition is mainly helpful in discriminating against is an external radiative interference, equally picked up
the interference. Even in this case, it will fail if the by the three cables. Fig.10(b) also corresponds to
interference is a real external PD signal from other another interference. Note the difference in the
connecting HV components. waveforms between these two interference signals and
also against those shown in Fig.7.
4. ON-SITE MEASUREMENT RESULTS
Sensor 1
Black Sensor 2
Blue
Joint
(d)
Fig.10: Rejection of external interference.
Sensor 3
Red
Fig.10(c) shows a pulse detected by the Yellow-phase
sensor (top trace) and at that instant nothing occurs on
the other two sensors. This suggests an internal
discharge in the Yellow-phase cable. Fig.10(d) is
another example of an internal discharge pulse on the Fig.11: Location of the three clamping sensors.
Blue-phase cable (middle trace).
In this case study, the subject is a three-core XLPE
Thus, a possible solution for interference rejection is 11kV cable which is approximately 1km long and has
the use of a multi-channel monitoring system that can a joint in the middle. Separate source tests gave an
record simultaneously signals from the three phases. unacceptable high reading of 3000-8000pC at 2Uo
Rejection is achieved by comparison of the recorded (12.7kV). In order to determine whether the cable joint
signals. A pulse is a valid internal PD if it occurs on is sound, three capacitively coupled clamping sensors
one channel only. However, if the cable discharges were positioned near the joint as shown in Fig.11.
near the termination, the signal can be picked up by Sensors 1 and 2 are 2.05m and 0.65m, respectively, to
other channels but the coupling is somewhat reduced. the left of the joint whereas sensor 3 is 2.15m to the
In other words, the pulse on the source cable is much right. The signals recorded simultaneously from each
larger than the pick-up signals on the neighbouring sensor pair are shown in Fig.12. The sampling rate of
cables. Thus the largest signal should be kept and the the DSO was set at 10Gs/s. It can be seen that the PD
others rejected. The criterion for comparison of the pulsewidths are <5ns and the clamping sensors are
able to capture these very fast PD signals. The time
delay between each pair of sensor signals can be 5. CONCLUSIONS
measured with accuracy in the order of 1ns. Based on
the propagation velocity of 0.2m/ns, the PD location The following are main points covered in the paper:
can be resolved to within 20cm.
¾ The PD signatures (time-domain waveforms and
spectra) of different types of defects on cable
samples were obtained from laboratory tests. Such
a database can facilitate PD recognition.
¾ Examples of interference and PD waveforms as
recorded on-site at a number of substations were
presented. A case of matching between laboratory
and on-line signal waveforms was given.
(a)
¾ Comparisons were made between attenuation
characteristics of paper and XLPE cables and also
between the clamping sensor and the HF-CT.
¾ The clamping sensor enables measurement at even
higher frequencies and thus it gives a much better
time resolution for discharge location. This was
demonstrated with a case study where a fault was
successfully located at the joint of an 11kV cable.
(b) 6. ACKNOWLEDGMENTS
7. REFERENCES