GE MEDICAL SYSTEMS GE MAGNET & CRYOGENS SUBSYSTEM

REV 10 DIRECTION 15093

FUNCTIONAL CHECKS
TABLE OF CONTENTS
SECTION TITLE PAGE

2 MAGNET ELECTRICAL CHECKS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–1


GE MEDICAL SYSTEMS GE MAGNET & CRYOGENS SUBSYSTEM
REV 5 DIRECTION 15093

SECTION 2 - MAGNET ELECTRICAL CHECKS

This section provides go - no go tests for internal magnet circuitry faults.

ELECTRICAL CHECKS CAN ONLY BE PERFORMED WITH THE COILS ( MAIN & SHIM )
RAMPED DOWN ( 0 AMPS). DO NOT MAKE CONTACT AT ANY CONNECTOR WITH COILS
RAMPED UP.

Note

Resistance Table 2-1 is for cold superconducting coils ( 4.2 K ).

1. Locate the Connector Pins using Table 2-1 and Schematics/ Interconnects, Illustration 1-1.

2. Use a digital meter to measure the resistance across the Identified Connector Pins.

3. Measure the resistances and record them in Table 2-1, comparing them to acceptable range.

2–1 FUNCTIONAL CHECKS


REV 11 DIRECTION 15093

TABLE 2–1
MAGNET CIRCUITS RESISTANCE CHECK COLD ( 4.2 K )

2-2 FUNCTIONAL CHECKS