RAPIDLab® 1200 Series

User Guide

Answers for life.

Step-by-Step

1. Enter or scan password if prompted. 5. Enter the demographics if prompted,


followed by the green continue button.

2. Mix the sample thoroughly. 6. When prompted remove the sample device
as shown on the screen and press the green
continue button.

3. Check on the screen the sample device 7. View the results. When completed the
you are using is highlighted, if not touch results will be printed. Press the green
the device to change it. continue button when done.

4. Insert the sample device as shown on the


screen and press the green analysis button.

Troubleshooting
1. If there is an obstruction (possible clot in the sample), you will be
required to change the sample port as shown on the screen.

2. To view a video of the procedure press the grey camera icon next to point 1.

3. When you have attached the new sample port, press the green button.

Never use an old sample

For more information, please visit Siemens at www.usa.siemens.com/bloodgas