Professional Documents
Culture Documents
Lunar DPX1
Lunar DPX1
Lunar DPX1
, .-1"
,,'
'\ ,
.
Excellence I n Imaging
DPX-IQ
Bone Densitometer
"
Service Manual
t I
-------
Part #3939
Rev.E
3/98
Sales: 1'~800-445-8627
/. . " .. ',~
" Service: 1-800-334~5831 '
',. Fax: 608-274.;.8210
...http://www.lunarcorp.com
COrpOtate Headquarter's' LUNAR Europe LUNAR Germany LUNAR Asia! Pacific
313 W Beltline Hwy. . Sterrebeekses1eenweg 101 Homelier SIr. 33 - 54 Taylor Street .
Madsion WI 53713 3078 Everberg 0-50858 Kotn W~ Pennant Hills. NSW 2125
800-334-5831 Belgium ' Germany Australia
r ';"
608-274-2663 32 2 759.73.73' (49) 615 2234 954 590 61 29873 1091 ;...
608-274-8210 AX 3227S9.57.58 AX 49 615 2234273 153 FAX 61 2 9 871 3753 AX
<
DPX-IQ Service Manual Rev. E (3/98)
A person who will be performing service work on the DPX-IQ should use this manual in the
following manner:
Chapter 4 contains technical deSCriptions of the circuit boards. These should b~(ead and
understood for the best understanding of how the scanner works, but can be use~r
troubleshooting as the need arises. . "~;, :i!~~~Ji;< ;
':f~~~ '.;t~~~. ~~~ !:'
.~ 'l'''~'~)~.. ",;r.
Section 5.1 should be understood completely as it explains the Diag~\~tTOOLS) Diskette.
'111;., B~;l~;:\~ 'r~'
The remainder of Chapter 5, all of Chapter 6 and Chapte~l . ". ," tead as needed, but are
good sources of information. .~. ~~:
~.
.,.
When requesting assistance from Lunar, please provide the following information:
System Number
Institution or Doctor's name
Location
Complete list of symptoms
Names and part numbers of parts needed for service
In addition to the information above, a Quality Assurance History printout listing the following
parameters will also help to improve the accuracy of our diagnosis:
A Ji~~~~~
High Channe l Air Counts (AIR_HIGH)
Low Channel Air Counts (AIR_LOW)
Spillover (SPILLOVER)
Peak (PEAK)
t:f. \~. w~~p
Tissue Value (TISS_LOW and TISS_HIGH) ~, ~ ..."",
,~,~ ~~~,
Large BM (0 BS_L_MASS) '1i~~:&
Large Width (OBS_L_WDTH) ~~\\-" ;> .
If time allows, send Lunar the file namE!.d D''"')~~
'"''' ' ' '"' '",... BF from the subdirectory C:\LUNAR\ANC\ })0< l)\':.1
, This file is a complete QA History al e above data and more.
'---"" It is highly recommended of the Air Counts and Spillover be viewed using the
Lunar software. Any parameters will be apparent from the graphs. For more
information on see "Viewing Quality Assurance Trends", section 8.1Q_
~~ ) I
For proble patient scans, it is recommended that you copy the scan files to
diskette, a description of the problem, to the Applications Department at Lunar.
.-..
.
', '
0-3
Information Request
LUNAR<!9 DPX-IQ Service Manual Table of Contents
Ta Ie of Contents
Chapter 1
Safety
1.1 General Safety ..................... ... ..... ..... ................... .. ....... ...... .............. ..... ....... .. ....... .. 1-1
1.2 Emergency Stop Button .... .. ................. ...... .................. ............ ........... .................. ... 1-1
1.3 Laser Exposure ...... ... ... ... .. ..... ... ..... .. ........ .. ... .. ........ ............. ... ...... ........... ..... ...... .... 1-2
1 .4 X-Ray Shutter Open Symbol ..... ..... ....... ... ......... ........... ... ......... .. ........... .. ........ .... ... .. 1-2
1.5 Cautions, Warnings, and Notes ..... .. .. ....... .... ...... ... ..... ...... .... ..........:.... ...... ... ........... 1-3
1.5.1 Caution Statements .. .......... .. .............. ..... ... .. ..... .... ... ... ....... ........ .. .... ............ ..... 1-3
1.5.2 WARNING Statements ..... ..... ... .. .... ..... ... ... ...... ....... .... .. .... ....... .. ...... .. .. .... ....... ... 1-3
1.5.3 Note Statements .... ..... ...... ......... .... ...... .. ... .... ....... ... ..... ........................ ... ..... ..... 1-3
1.6 Safety Concerns and Labels ..... .... .... ...... .... .... ...... ~ .. ......... ........ .. .. ... ...... .. ... ..... .. ...... 1-3
1.7 Keys .... ......... ................................ ........ ..................... ............ ... .............. ........ .. ...... . 1-5
1.8 Leakage Radiation .... .. ....... ......... .. ...... ..... ............. .... ......... ........... ........................... 1-5
1.9 Controlling Computer and Accessories .......... ..... .... .. .. .... .................. ............ .. ......... 1-5
.
Chapter 2
System Overview
2 System Overview ................... ....... .. ............... ......................... ....... ... ..... ............ ......... 2-1
2.1 Electronics .... .............. ......... .......................... ... ........ ........ ........ ... ..... ....... .... ............2-1
2.2 Single Board Controller (SBC-03) .. ..... .... ....... .. ...... .. ..... .. ......... .. ..... ..... ...... ...... ....... .2-1
2.3 High Voltage Power Supplies .......... .. ......... .......................... ...................................2-2
2.4 Tube Head and MAX Board ..... ............................................................. ...... .. ...........2-2
. 2.5 Transient Protection ...... ..... ... ........ .. ... ... ... .. .... ................ .. ....... ..... ... ........ .. ...... ... .... ..2-2
2.6 Audible X-RAY OFF Signal ........ .. ...... ... ...... ............................. .. .. ........ .. .............. ....2-2
2.7 X-Ray Collimator Subsystem ............,.............. ..... .. ......... ......... ..... ..........................2-3
2.8 System Detector Circuit ... ...... .... .......... .. ....... .. ............ ..... ................ ... .... .. .............. .2-3
2.9 Motion Control ... ....... ....... ................................... ........ .. ..... ......................................2-3
'. '2.10 Control Panel ............................. ..... ................................. ........................................ 2-4
',· 2.11 External Covers ... ..... ... ......... ........ .... .... ... .... .. ....... .... ...... ..... ................... .. .. .... .... .... ..2-4
2.12 DPX-IQ Fusing .... .............................. ....... ................. ........... ....... ·............................ 2-5
Chapter 3
Technical Specifications
3.1.1 X-Ray Generation:" LUNAR Part Number 2288 ........ .. ....... ................ ........ ... .....3-1
3.1.2 X-Ray Tube Head ............ ... .... ...................... ... '" .. ...... .......................... ............ 3-5
3.1.3 High Voltage Power Supplies .... .. ..... ......... .. .......... ....................... ..................... 3-7
3.2 Collimator Subsystem ............. .......... ....................................................................... 3-8
LUNAR~ DPX-IQ SeNice Manuai Table of Contents
4.6.2 Address Selection .. ... ......... .... .. .. .. .... ..... .. ... ... ...... .. .... .. ... .... ..: ..... .. .... ...... .. .... .. .4-23
4.6.3 RS-232-C Interface (U7 & U8) ..... .. ........... ....... ..... .... ... ... ..... ,.. .... .. ... ... .. ........ .. .4-24
4.6.4 System Reset (U8, SWi) ..... ... ... ....... ...... ... .. .... ... ..... .... .. .... ..... .... ..... .. .. .. .. .... .. .4-25
4.6.5 System Clocks (Xi , X2, U15) ...... .... ... ... .... ..... .. ... ... ..... ..... ... .. .... .. ..... .... ...... .... .4-25
4.6.6 Event Counting (U13, Ui4, U17, U18, Ui9, U20, J1 , J2) ..... .... ...... .. ..... ..... .....4-27
4 .6.7 Stepper Motor Control (U9, U10, U3i , U32) ..... .......... ............. ..... .. .. ..... ... ... ... .4-27
4 .6.8 Limit Switch Sensing (U27, U31) ... ...... ... ... ... .... ..... ... .. .... ... .. .... .. ... .. ..... ... .... .... .4-27
4 .6.9 1/0 Ports (U23, U41) ... ... ....... ... .. .... ...... .. ......... ... ... ... ......... ..... ........ .. ... ..... ...... ..4-28
4.6.10 Watch Dog Timer (U48) ... .. .... ... ... ... .. ....... .... .... .... ... .. .. .... ...... ..... .. ........ ... .. ..... .4-31
4.6.11 D/A (Remote Detector Amplifier High Voltage) Control (U44, U45) .. .. ... .... ... .. .4-31
4.6.12 DIA (Remote X-ray High Voltage and Current) Control (U36, U37) .. ....... .. .... ..4-31
4.6.13 AID (Remote X-ray High Voltage, Current and Amplifier High Voltage Feedback
Sense) (U35, U40, U42, AND U43) .... ... ... .... .... .. ....... ... .... .. .... ..... ... .. .... .... ...... .. .. ... ... .. .4-31
4.6.14 Thermocouple Sense (U38, U39) .... ... ... ....................... ..... ... .... ...... .. .. .... .........4-32
4.7 The OINK Board ... .. ... ... ..... ........ .. ... .. ... ...... .... .... .. ... .... .. .. ... ........ ...... .. .... ... ...... .... ...4-35
4.7.1 Motion System ... .... ... ....... ...... .. .. ... ..... .... ... ... ... .. .... ............ ...... .. ........ ... ..... .. ... .4-35
4.7.2 Patient locator Light .... .......... ....... ... ..... ... ........... .... ......... ..... ..... .. ... .... .... ....... .4-36
4.7.3 Collimator Control ...... ...... ... ... ... ... .. .... ...... .. .... .:... ... ... ... ...... .... ........... .. .... ... ... ..4-37
4.7.4 Fans ...... .... ...... .. ... ...... ....... .... ....... .... ... .... .. ...... ..... ..... .. ... ..... .... ....... ... .... .... ...... 4-37
4.7.5 Indicators ... ...... .... .... .... .... ... ..... .. ...... .. .. ... .... ... ..... ........ ... ...... ...... ... ..... ............. 4-37
4.7.6 Emergency Stop ...... .... ..... ...... ... ... ...... .. .. .......:..... ..... .. .... .. ... .... .... ............ .... ... .4-38
4.7.7 OP/CAl ..... .... .. ........ ....... ... ......... .. ....... ........ .. .... ...... .. ...... .............. ............. ..... 4-38
4.7.8 Error Detection .......... ............... ...... .... ........ .... .... ..... .. .. .... .... ...... .... ... .. .......... ...4-38
4.8 Optical Motion Interrupt (OMI) Board ........ .... .. ... ..... ... .. .. .. .. ..... ............... ... .. .... ..... ..4-41
4.9 The Laser Board .... ... .... ....... ... ... .... ....... ..... ...... ....................... ....... ..... .. .. .... ..... .. .. ..4-43
4.10 DPX-IQ Mechanics .... ... ... ..... .... ...... .... .... ... .. ... .. ... ... .. .................. .. ........ .... .... ....... ..4-45
Appendix 4.A
AMP Board Schematics
Appendix 4.8
AGS Board Schematics
Appendix 4.C
DCA Board Schematics
, Appendix 4.D
MAX Board Schematics
Appendix 4.E
XORB Board Schematics
Appendix 4.F
SBC Board Schematics
iii
D:-')'.. -,'e;; Service Mar.ual TaDle ot (;ontem~
8.7.3 Limit Switch Tripped During Scan ...... .. .. ..... ..... ... .. ........... .... .. .. ... .. .. ....... .... ...... 8-39
8.8 Imaging Problems .. .... .... ... ..... ...... ............... ... .......... .. ... ........ .... .... .... ....... .. ........ ... .8-41
8.9 Femur Scan Problems ..... .. .. .... ..... .... ........ ..... .. ..... ... ....... .. .... ...... ..... ...... .. ... ..... ... .. .8-47
8.10 Failing Air Matrix Results ... .... .. .... ... ..... ...... .. ... ... .. ....... ... ... ...... .... .. ... .... .... .... .. ...... ..8-49
8.11 Indicator Failures ... :.. .... ................ ................. ........ ................... .. ........................ 8-51
8.12 Communications Failures .... .. .. .. .... .. ... ... ... ..... .... .. .. ..... .. .. .... .. .... .. ... ... ...... .... .... ... ... .8-55
8.13 Start Up Software .... ..... .... ... ..... ... ... ... .......... .. ............... .... .. ...... ... ... ..... .. .. ........ ...... 8-57
8.14 Viewing Quality Assurance Trends .............. .......... .... .......... ...... ...... ..... ..... .... ..... ... 8-61
Appendix B.A
AGS Board Troubleshooting
Appendix B.B
AMP Board Troubleshooting
Appendix B.C
DCA Board Troubleshooting
Appendix B.D
MAX Board Troubleshooting
Appendix B.lE
OINK Board Troubleshooting
Appendix B.F
'OMI Board TroubJeshooting
Appendix B.G
SBC Board Troubleshooting
Appendix B.H
XORB Board Troubleshooting
',.
vii
DPX-IQ Service Manual Rev. E (3/98)
Safety
This chapter highlights safety devices and features one should know before using the DPX-IQ
system.
,/
Safety Chapter 1
DPX-IQ Service Manual Rev. E (3/98)
BLANK INTENTIONALLY
Chapter 1 Safety
DPX-/O Service Manual
Manual Rev. E (3/98)
------------------~~~~~~
Lh.SAFETY
1.1 General Safety
• DO NOT attempt to service the DPX-/O without first reading this manual.
• DO NOT attempt any repairs without prior instructions from authorized LUNAR
personnel.
•• In order to maintain electrical safety and electromagnetic compatibility, the Lunar DPX-
DPX
10 is only to be connected to a computer, printer, and peripherals that are certified to be
compliant with IEC 950/EN 60950 Safety of information technology equipment, including
electrical business equipment and IEC 601-1-2 Medical electrical. equipment, Part 1:
General requirements for safety, 2. Collateral Standard: Electromagnetic compatibility-
compatibility
Requirements and tests.
.'
g
... LEFT RIGHT"
"-
BACK
FRONT
T i
DPX-IQ SeiVice Manual Rev. E (3/98)
i
\
1.3
,
Laser Exposure
The DPX-IQ is equipped with a Class \I Laser device. This laser is usee ~:Jr t=a:lent positioning.
A Class 11 rating indicates a low power visible laser that is not normally l,dZ"rdous to eyesight
but has the potential to be hazardous if viewed directly for an extended per .od of time. Because
of this potential hazard, DO NOT stare directly into the be'am while the !2':.er is in operation, and
DO NOT allow the beam to shine directly into the patients' eyes. No spedfic eye protection is
re:juired with a Class II laser.
A yellow laser-on indicator, located on the front of the scan arm, is lit wr ,en the laser is on. The
program activates the laser during positioning for an image acquisitior The program then turns
off the laser when you begin the scan. The emergency stop button will turn off the laser.
There is a caution label (Figure 1.2) on the top of the scan arm above the controtpanel.
O ;,< This symbol is used to indicate an open-shutter condition in accordam;e with the
safety standards established by the International Electrotechnical Commission
(lEC).
This symbol appears near the Amber X-ray shutter-open indicator light. The X-ray shutter-open
indicator light is located on the control panel on the scan arm near the front.
/ .
When this symbol appears in literature associated with the DPX-IQ scanner, it will be used to
indicate that the procedure being described results in an open-shutter condition. During these
times personnel should exercise caution to avoid excessive exposure to the X-rays.
1-2 Safety
DPX-IQ Service Manual
Manual RE:v. E (3/98)
CAUTION:
CAUTION: A caution statement reflects a condition that, if not avoided, could
cause equipment or property damage.
WARNING:
WARNING: A warning statement reflects. a potentially hazardous condition that, if
not avoided, could result in serious injury.
NOTE:
NOTE: This symbol turns, th,~leader's attention to important information which may
otherwise be ol[ef/obked.
Because the DPX-IQ Densitometer contains moving parts, there are places on the SCclnner
where there is a danger of pinching. Operators should be aware of these pinch points to avoid
injury to the patient or themselves. Labels applied at the LUNAR factory indicate the location of
th~ pinch points. The pinch points and their labels are shown in the figure 1.2.
Safety 1-3
LLINARCB DPX-IQ Service Manual Rev. E (3/98)
Lt.~
----;'9
£/~,
1-4 Safety
1.7 Keys
Keys are provided for the front and side panels on the DPX-JQ. These keys are used during
installation to secure those panels. After installation, all keys MUST be removed from the
scanner and safely stored by the installer. Electrical safety codes require that these keys be
removed from the DPX-JQ when scanning to prevent unauthorized access to hazardous internal
components.
Safety 1-5
DPX-IQ Service Manual Rev. E (3/98)
21'
20'
mRlhr",,...... _ - -_ . . . 15"
I' "
0.1 mRlhr 0.2 mRlhr
/ \
\ 7-
7
\
I I I I I
20' 10'
I
12" 20'
27'
o o
L -__~--------I~--.---~I----~--~I------_r---J
36' 10' 14- 38
38-
Front View
LEGEND
SCALE
Approximately
--- sg~n<fiteJ~~~vaf9'(ih:l~~'iid
-Scanner 4.75mA
1: 18
1-6 Safety
DPX-IQ Service Manual Rev. E (3/98)
------------------------~~
Chapter 2
System Overview
This chapter provides an overview of the DPX-IQ system. In addition the chapt~~ntains a
brief discussion and illustrations of the DPX-IQ power system. ~. #~4t·~~
i> ~:I-i~ fi;',i'!;~ f~)
~~;.",~.~
2.1 Electronics ...........................................................................,~~'~5? .... .................. 2-1
2.2 Single Board Controller (SBC-03) ................................. ~.,'5'.~:~{:~~\ ..~ ........................... 2-1
High Voltage Power Supplies .................................~:~~~'i .....................................2-2
. • .r!;~~"tttit;
2.3
2.4 Tube Head and MAX Board ...........................~;~.~.~~.~;. ...........................................2-2
2.5 Transient Protection ...................................... ~~;~~~~;;..............................................2-2
2.6 Audible X-RAY OFF Signal ..................... ~J:~~~:.:................................................... 2-2
2.7 .
X-Ray Collimator Subsystem ........:.: ...~~.\,.;;~ ......... ~ ..... ............................................ 2- 3
. .it~· 'i,""
1
\
' \,
2.1
I
Electronics.
The electronic components of the DPX-IQ are mounted on the grounded Electronics Mounting
Chassis which is horizontally fastened inside the frame. There are three low-voltage linear DC
power supplies (under 30VDC), and two very-high-voltage dc power supplies (to supply 76kV to
the x-ray tube) on the pan. One medium-high-voltage power supply (1000VDC) is located in the
upper arm near the X-ray detector and supplies power to the Photo-Multiplier Tube (PMT).
In addition to the power supplies, the electronics mounting chassis holds seven printed circuit
boards, a stepper motor controller, the audible alarm, an AC entrancelline filterlfuse holder, an
isolation transformer and a terminal strip for power distribution.
if"" .
'I';t)}
2.2 Single Board Controller (SBC-03) ~~,~$;'~~'
The microprocessor-based Single Board Controller (SSC) provides overta!t~~tion and control
of the scan table. I~ ha~dles r:notion control, pulse counting, and . i~1.i~crfsenSing, ~s w~1I as
... 1
RS-232-C communications with the host computer. The S032.:bas~tl't8c
\'!h(. .... .,....;.
has 56K of" available
~~~
Event counting for kes place on the SSC, after pulses have been properly
am plified and . discriminated by other circuitry. The SSC also provides
independe trol of the two stepper motors, and senses limit switch actuation at
the limits
Two 12-bit converters are used to set high voltage and current for the X-ray source. A 12-
bit AID converter is used in the feedback path to sense the actual high voltage and current
settings.
',.
Although the SSC sets the high voltage and current, the MAX board actuallYl!re , ,\)'lS current to
the filament transformer In the Tube Head. The X-ray Insert converts thl? ~., · .,~ .,. 1r1to x-rays by
accelerating the electrons with a high voltage from the cathode into a tu~t~ arget on the
anode. The tube current can range anywhere from 0.15 rnA to 4.75«)\'9epending on the
acquisition type. ,f!..'f'~~.~p'
2.5
The XORB printed rovides protection to various circuits from transients within the
system. There ionai static discharges within the Tube Head. The transients
caused by ='~ ="'--'1o"'=''''''''''':'charges (arcs) are shunted to ground through the 'array of transorbs
present on . Although there is no safety hazard to anyone, the electronics
could be da were no protection prOvided.
notifies the operator that there is a "safe" condition. The OINK board senses the closure of the
Shutter or the cessation of current through the X-ray Insert and switches on the audible signal if
/ either of these occur during an exposure.
',.
Limit switche - ; ach end of the scanner and each end of the arm signal the SSC whenever
motion reaches either end or side of the scanner. If a limit switch is tripped during a patient
scan, the motor associated with the tripped limit switch is halted and the shutter is closed.
There is a dedicated DC power supply for the motion control circuitry, supplying +26VDC. The
+26VDC ground is isolated from the logic ground to reduce noise.
",'
2. ~ 0 Control Panel
The control panel of the DPX-IQ, located on the front face of the upper arm, is the main operator
interface outside the host computer. Four indicators display status. A green Light Emitting
Diode (LED) indicates a power-on condition; it lights when the +26VDC power supply comes
on. Another yellow LED indicates if the laser is on. An amber X-ray On LED indicates that x-
rays are being produced (current is flowing through the X-ray Insert), though exposure is not
necessarily taking place. Only when the amber Shutter Open LED comes on is the Shutter
open and exposure possible.
Rocker switches on the panel allow operator control of the beam position.
~~.
T.he s.quare. red button may be used for emergency stops, and is to be use~ ~\o~~mergency
situations; It resets the scanner and halts x-ray production. d~~~:~" '.
.,~,.;;:.~"~~-t-_''f.;:~
2.11 External Covers ~\ 1~~~11:t.:,'~+'·
. "':t.{ , !,:~-:-.fl.
The Internal components of the scanner are safely secure,$ ~Jlt~mber of panels and doors
including the scanner's table top. The Front and Side ~,~secured by locks, for which the
customer should have keys. The Front Door gives ~:~ . ',,;Jj'(a"'the electronics panel, but for
_
o S§.
It is not u to remove the front and back panels for most service needs.
However, needed to the Front and Rear Longitudinal Carriages, these can be
removed. panel is secured by hex socket head-head screws and must be slid out of
the way, for it IS between the Arm Column and the frame (see section 6.11 if the scanner has an
arm slot cover). The front panel is held by two hex socket head-head bolts and lifts off the
frame.
If access is needed to the detector, Transverse Limit Switches or the other components
mounted above in the arm, the covers of the arm must be removed (these covers are grounded
r to the arm). The bottom cover can be removed by loosening the eight screws holding it in
place. The top cover is held in place by four screws, two in the front and two in the back. The
top panel cannot be completely removed because the control panel is attached to it. Instead,
support the top panel with the shroud support brackets.
',.
nm
J3
AGS
J8 J7
J41---1
J
1DPX..,~
.-~
HIGH ENERGY
r-~~----------------------~
co ~
Q? PMT .~
£1 AMP AMP~IFY
~'':'i;;;''''!''iH' 1-
w
~
;>
Q)
HVPS
\JENUA"
iEl OMI,:('" 8:·!" R~·Nl~~:X);:
>'=.'
{:'.~"'~'},) :;~:·j :·i:"'· #: ~\
TRANSVERSE MECHANICS
g!
o
E
Q)
LONGITUDINAL MECHANICS
PMTI LASER
~~S~~E~.
ti'." . Ir---------~
Detector & J15 J17 J16 J21 J20 J1
LASER
. ALARM
BOARD
cu SCINTILATIO
:J
C N CRYSTAL
cu
~
~ ~%~ ";p"l1eNt ,,.,.'ic
a .::;:.>\,;.{t:< :'::~':;"'~
. ,;:";'" DP~V
A SERIES
.~
ill
. ,. . . . . ..., ';nrrL.r':·:{"~~BLE TOP BLOCK
.
CJ)
o li~~~~; DIAGRAM
~
gs '~~;':. ~
K-EDGE FIL TI
0.~;'O~:::~~~I~~C~~~:~S
Nt·
/1ft:\.
/ \\
HF~-c'CP:" 1 § ," ;·······. J \. \ ~ I ,
_
~;.
FILAMENT
TRANSFORMER
.. : ,
ry."': 1".1
.
.:;::': •....•.
_ NEGATIV
E HVPS
MJ
FILAMENT
CURRENT
CONTROL
J7 J2 J4
II
I
a..
.
/ _• -
' .._st:1..1-_
@
a:::
<{
z
~
-I
THfM~COUPL
~---------------------~I
XORBJ~~~~~~~~~~~~~~~~~;~;~~~~~~~~~~~~~~~~~~~~~f~==iF~tJ
J5rRANSIENT J
PROTECT/O~4
J3
J15t=========~::~::::~~~llQ~
J21 -kV MONITOR
__________________________====:j=========== <0I
J13 N
L:_fPl~~K~' DP,)(-JQ Service ManL:a! Rev. E (3/98)
----------------------- - _. -_. __._ ---------'---.....:....
/':
/ /
~
)
POWER BLOCK
DIAGRAM
OINK
FUSES
100/110 2301240
+26VDC FUSES
POWER F7 160mA
F22.5A 1.25A
SUPPLY F8 160mA
F42.5A 1.25A
I
i
fUSES +5VDC ! HV/AMP I
1001110 2301240 +1-12VDC : MODULE
,.-....
F1125A O.63A POWER
I i I
f31 .25A O.63A SUPPLY
I I I
+28VDC
POWER
~
~ I
I1 ~
SUPPLY 1 0
I :::0
i
MAX I
I
F1 O.SA
4
-+ 1...
I
I
-HV I
POWER
SUPPLY
XORS
+HV
POWER
SUPPLY
'"
',.
1
Chapter 3
Technical Specifications
This chaprter includes specific technical information perstdddaining to the various corrWnents and
subsystems of the DPX-IQ System. ~ .~:t·~2:~~,
". ~ ..~;.;... '" ;fi...i-':"
3.1.1 X-Ray Generation - LUNAR Part Number 228B ............. ~~~.:~; ::': ~~ ....................3-1
3.1.2 X-Ray Tube Head .................................................. :~:~~t:~:~.·:~· ...........................
3-5
3.1.3 High Voltage Power Supplies ............................~:.~;-:.:;~~.~~~ .................................. 3~7
3.2 Collimator Subsystem ....................................._.:!A,•. ::~~~;~::: .. ............ ........ ................ 3-B
3.PLnI~~~~t~ers7~:Y@~:~~::::::::::::::::::.: :::.:.::::::::::::::::::::::::::::::::::::::::::::::::.::::::~~l
3.B.1 I~' -:-: _ . . a (Long. x Trans.) ........... ... .................................... ........... ... .... .. .. .3-12
3.B.~~P~: - ... ' ··s(· ............................................................................................ ............ 3-12
3.9 ~~:.. ,'. ~frtal Specifications ..................... ............................................ ..... ........... 3-13
3.9.1 ..~bientSpace ...................................... .......................................................... 3-13
3.9.2 Dust, Fumes, and Airborne Debris .................................................................. 3-13
3.9.3 Humidity .. ,............................................ ..... ...................................................... 3-13
3.9.4 Static Electricity ........................................................................................ ....... 3-13
3.9.5 Temperature....................................................................................................3-13
3.9.6 Shock and Vibration ........................................................................................3-13
3.10 OPX.-tO F1JSeS ......... .. . .... .. ... ........... ...... ..... . ....... .......... . .. .......... .. ....... ......... .3-14
The foilO+.'ir;g tabiB wmams tr,e soedficaticns f~; trre lEe X-Ray generator for DPX-IQ devices.
J. *:~~~~.~\~'i-
max. into surrounding air through
forced air convection. Flow rate:
~ ~6~' 36m3/h (approx.) Temp. rise of air
~ ~~"~ stream: 2S0C (approx.)
Allowable hj~ge power Spellman X21121X2113 rev. K and IEC 601-2-7 6.8.1 &
supplies higher 50.2.101-102
Bertan 2411 P & 2411 N rev. A and
higher
Allowable tube head assemblies Lunar model 2288 or equivalent IEC 601-2-7 6.8.1 &
50.2.101-102
Original language of English IEC 601-2-7 6.8.1
accompanying documents
Max. continuous kV,mA @ 76kV,4mA IEC 601-2-7 6.8.2 1)
nominal rated kV
Max. intermittent kV, mA @ 76kV, SmA IEC 601-2-7 6.8.2 1)
nominal rated kV
Max. continuous kV, rnA @ max. 61 kV, SmA IEC 601-2-7 6.8.22)
mA
Max. intermittent kV, mA @ max. 76kV, SmA lEG 601-2-7 6.8.22)
rnA
Continuous kV, mA for max. 76kV,4mA lEG 601-2-7 6.8.23)
electric output power
4<'>"'~"f'
~i-
Serial Load Rating 361 W (4.750 rnA, 76 kV) for up to 4 IEC 613 I 1989
i;,~";;;~<.,,
min and 59 sec with a 10 min cool
.
o@;~,:: ~~,. down time between scans.
X-Ray Tube Assembly Heating I see Figure 1.7 IEC 613 11989
Cooling Curves
OISlance
Reference
Longitudinal
---Axis
300.00
::;-
~
C
4.1 200 .0
C
0
0
iV
4.1
::I:
"0
!O
4.1
::I: 100.00
4.1
.0
76kV, 1 rnA (l6W)
::::l I
I-
50.00
0.00
0 5 10 15 20 25 30 35 40 45 50
Time (minutes)
", ..
k\ ..
2.4
. . . ..
~ {aJ<;(~ -'7'jc=-
!
,.,.. ~
~TtN5AZ
I - -' _. -
2.3 3
... ~
. G.O:/
Curren!. -
.' "/------- ---- -----
- _. - 1/)
2.2
'/
"";/
2.8 ~
$ Voltage .2:
C 41
e Cl
:; ~
:,"/
"0
() 2 .1
,/ 2.6 >
C C
41
E I 0 __
41
E
.!!! I
u:: 2
~~
".
:"/ I
-' ,
~:,..
/
1.9 2.2 sre fJ O. 02.
I
0/1 ~
i
I
1.8 .",,":'-. ':0:-:. 2
1000 4000 5000
Maximum Anode Heat Content 6000 Joules X ~'J6; 8 -i~6a IEC 613/1989
Anode Heating and Cooling Curves see Figure 1.9 IEC 613/1989
6000.00 T
5000.00
0 .00
0 5 10 15 20 50 55 60
I Reference Axis
,.'
#~~.~~'
Switching Frequency High Frequency (50 kHz)" ,'~~0 .:;I~" ):·
Table 3.3
';-;. ~
Goilimator 5uosystem
-:-he collimator subsystem contains a beam coliimator and a beam shutter.
3.2.'1 Collimator
The collimator consists of two circular openings; fine and mectium. The medium setting is used
as default.
3.2.2 Shutter
An X-ray shutter occludes the radiation beam during tube warm up and certain QA tests. The
shutter is a lead plug mounted to a linear solenoid on the collimator assembly. I~. opened
electronically by the system software during acquisitions and is closed aut'?T~pY- at all other
times. ~i;~\~"Wf
3.3 Detector ~. t~;·~~J' .'
. . 'i:.~,-~~;',1'
The DPX-IQ uses a Nal(TL) crystal photo-multiplier tu~~~~,.75" in diameter and 6.S25'!
long. It is connected to the AMP board by a lemo conJ1~~~{, The PMT uses a positive high
vo Itage .Input b etween 600-
800 VDC
. ;v..&~
~:'., ~~~. ,y,~,
",.
VO "'<
Wavelength 670 nm
.'
DPX-/Q Service Manual Rev. E (3/98)
Specifications DPX-IQ240
Table Hei ht
37.1 Transverse
The tube head and detector assemblies are mounted (.~ r arriages whk ) move transversely on
the arm. This motion type is required for positioning, dally QA's, and ~I' image acquisitions.
DPX- 180
Transverse Software
Limit
197.5 em (77.8")
196.69 em (77.4")
. . *-~ .
Longltudl~~,..
Software LiffiW~
3.8.2 Programs
• AP Spine acquisition and analysis.
• Femur acquisition and analysis.
• Lateral Spine acquisition and analysis.
• Total Body acquisition and analysis .
• Forearm acquisition and analysis.
• Orthopedic acquisition and analysis.
• Pediatric acquisition and analysis.
• Small Animal acquisition and analysis. ,~~~"':i;. .
r'~()"%> .
~...."(,. stJ·
" $-
Do not block the cooling vents on the computer and scan table. Allow 6" between the console
table and the wall to ensure adeq'late clearance for the cables and pJugs at the back of the
computer.
.~&,"
• • ~~. ~er~1,~5>
3.9.2 Dust, Fumes, and Airborne Debns &';"1- %~ 41~>"
Install the system in a clean, well-ventilated area. The accumulation ~~.~"nd other airbome
contaminants can impair the diskette drive heads and other s~. ns~~~<;mponents. Lunar
recommends a "No Smoking" policy for the system area . .,. ~~.~:it ~
£"" ' ,. ·~;,. ·W
'i'(#.~' ~~~
3.9.3 Humidity
.:ft.h ,~· ~ .'
~~,~,
'!;'::o w"
The humidity level in the system area shoul~ee~fo, non-condensing.
. ~c:e:
~
3.9.4 Static Electricity .~~
The system should be in a st . ~. '~''-f~ ·ronment. This can be accomplished by using anti-
static mats and by not all<;> ..-~. tive humidity to drop below 20%.
}'§,r:if- .
3.9.5 Tempe~q .
The acceptable " t. . ..~:~Ture range for the system is 65DF to 81 OF (180C to 27DC).
'~".
\t~
3.9.6 Shoc({7and Vibration
Shock to the imaging table may not exceed 1 G for a period of more than 1 millisecond. Do
NOT subject the imaging table to vibration in excess of 0.25 G at 5 Hz.
NOTE: Whenever the system is turned off or there is a power failure~ it must be warmed
up f.or one hour before .acquiring images. again.
F3 1.2SA,2S0V .63A,2S0V
F4 2.SA,2S0V 1.2SA,2S0V
.,; ':
FS .63A,2S0V .63A,2S0V ~ ...
~
~ ~- );.iI. ~".
MAX PCB F1 .SA,2S0V & :+: ~~~A, 2S0V
Table 3.8 "
Chapter 4
Technical Description
This chapter includes discussions of the various interface circuit boards of the D~-lQ.
" .~~' ~
: ~.1 ~he~i~:Ui~~::~;;ii~~ ••••••.•• • •: •••••: ••••• •• ••• ••••• •• •••• • •••• ••~~i\.~t.~·~ .. .. 1:i
.2 The AutomatIc Garn StabIlizer (AGS) ...... ... ..... ... .... .. ........,.,, ~ .~~: ..... ... ........ .... .. ..... ...4-5
".".~ '1"
4.6.12 DIA (Remote X-ray High Voltage and Current) Control (U36, U37) ... .... ....... .. .4-30
4.6.13 AID (Remote X-ray High Voltage, Current and Amplifier High Voltage Feedback
Sense) (U35, U40, U42, AND U43) ........... ...... ........ .. ...... ... ................ ... .... ...... ........... .4-31
4.6.14 Thermocouple Sense (U38, U39) ........ ..... ..... ... ....................... ........ .............. ..4-31
4 .7 The OINK Board .......... .................... .......... .. ... .. .... ..... .............. .......... .... .... ....... ... ..4-33
4.7.1 Motion System ....... ..................... ...... ..... ..... .... ... ... ............. .. ....... .. ... .... .......... .4-33
4.7.2 Patient locator Light .. .. .... ............ ..... ........... ... .. ............. .. ....... ............ .. .... ..... .4-34
4.7.3 Collimator Control ........................ ..... ... .. .................. .. .... ... ....... ........... .. .. ... .... .4-34
4.7.4 Fans .............. ... ... ......................... ... .. ... ... ........ ... ........... .............. .... ................4-34
4.7.5 Indicators ...... ..... ..... .......... .............. ...................... ................... .. ..... ..... ........... 4-35
4.7.6 Emergency Stop ... .................. ................. ............ ... .................. ..... .. ...... .. .. ..... .4-35
4.7.7 OP/CAl ........ .......... .... ... ....... .. .................. .... ..... .. .... .............. ........ ~ ...... ........ .4-35
4.7.8 Error Detection .............. ... .. ... .. ..... ......... ....... .. ..... ... ........... ..... .. .~;;,"t~~~ ........... .4-36
4.8 Optical Motion Interrupt (OMI) Board .......... ....... .. ........... .. ...... ... .'~:;.,..'~l~.~....... ..... .4-39
"i::",'C;.,
Circuit Functions
•. ..Amplify charge pulses from the PMT
• Shape the signal into a stable bipolar pulse
• Drive the pulse down a 50 Q coaxial cable to the AGS board
The input to the AMP-04 circuit is a PMT/i'Jal charge pulse and output of the circuit is a
700 nsec bipolar pulse (3 Vpp) into 50 n.
i ">
~t$i~~~~~_
4.1.1 Circuit Description . 'i~, 5~}:t "
':r.1!.:. ·k;·~, "" -'t.:!"
The AMP-04 amplifies charge pulses from the detector system opera~~~OO to +800 Volts ,
.-6.1...:.. -::J."e -
shapes the signal into a stable bipolar pulse, and drives the pulse_~wp'a 5.1 meter 50 Q
coaxial cable. Maintaining constant pulse shape at high coun ' ~s is a prime goal.
,,1'$- ¢~ \. "<,f••••
Pulses of charge from the PMT are passed through a 2 ~' · . · -'. voltage blocking capacitor
located in the PMT high voltage power supply. Th.~~fi3ulses go through the virtual ground of
the current to voltage ~on~erter formed from~.1 ,~'ti;scharges C5 ~hen the amplifier is in the
power-off state. The nse time of the pulse ~ s~-uythe 6 MHz bandwidth of the current to
voltage converter, and the fall time ~30 nsec decay of the Nal crystal. The
amplitude of the pulse is determ the 3K D value of R3 and the voltage of the PMT
(approximately 700 Volts for ignal with 100 KeV photons). The components of the
current to voltage conve , C12) were chosen to maximize the output signal while
retaining stability . These values provide a flat 6 MHz bandwidth for input
capacitances of u
The pulse rk composed of R6, R7 and C13 shortens the pulse. C13 and R7 set
the decay to 100 nsec. The addition of R6 to this circuit cancels the 230 nsec
decay of the crystal. The Signal is then amplified by U2 , an amplifier with a voltage gain of
16 and a bandwidth of 6 MHz.
C14 shapes the unipolar pulse into a bipolar pulse. This AC coupling also removes the effect of
after glow in the Nal crystal and DC drift problems in U4. The bipolar signal is amplified with a
voltage gain of 4.3 in U3, and passed to U4, a 50 Q line driver. Two voltage amplification
stages were used to provide low DC drift with good high counting rate performance. Capacitor
clamping prevents fluctuations in the power supply level when the output Signal is clipped .
~
.~ ~~:\.-
'<~ &.4y.~
5tl ~, .=
Figure 1 shows the unipolar output of the first stage of amplification CI~~'-Frgure 2 shows the
output after the second stage of amplification (TP5). The last fig~~iie output of the line
driver and the output of the board (TP6) . ., , ~ .\;;:::"
4:..... ~.
.vl\\,\ J"
".:\-=?
Test Points
Test Point
TP2
Output
+12 VDC
TP9 Ground
Table 4 .11
'~~~.
,,:~.~
BLANK INTENTIONALLY
Circuit Functions
• Amplifies high channel pulses
• Attenuates low channel pulses
Operations
The analog section of the AGS-05 consists of an input divider, an input unity-gain buffer, a
variable gain stage through the use of an N-Channel JFET, an amplification stage followed by
an AC coupling network, and an output unity gain current driver. The digital section consists of
an "AMPLIFY" and "ATIENUATE" pulse qualifier, an "AMPLIFY/ATIENUATE#,..~~e indication
circuit, two 4-bit binary UP/DOWN counters, a "ROLLOVER/RELOAD" circbil ;<>:~ a D/A
converter to control the gate voltage to the JFET for gain control. _ ~:~i;·~>(?
+J!..~i'.~o
A signal entering the AGS (TP4) sees 50 0 of input im '. :j~nd is divided by 4.24 before it
passes through a unity-gain high-output current buffe _.~0002CN). The signal then passes
through two JFETs connected in series with their .~ ' . orted to their sources (Vgs=O). This
way they have a constant drain to source r~ ,3~, ~nd are the equivalent of series resistors.
The series JFET's (01 and 02) are uSE?d t~s"ie a thermally stable voltage divider with 03 .
.... ~:;~,.~
When an "AMPLIFY" pulse comes in from the AGSbCA it is qualified by U2 (74LSOO) and the
counter pair (U3 & U4 - 74LS193) is incremented 'or "CLOCKED" up . When an "ATIENUATE"
pulse comes in it is qualified and the counter pair is decremented or "CLOCKED" down. The
counter outputs 00-07 are connected to a D/A converter (U7 - AD558KN), the counter range is
from OOH to FFH with SOH being the center of the range. Two potentiometers (R20 & R1S) and
an op amp circuit (US - LF351 N) scales the DAC's output voltage and inverts it to drive the JFET
gate of 03. If "ATIENUATE" pulses continue to clock the counters, they will clock down to OOH.
There the JFET gate voltage will be maximum (-4 to -6 VDC) and the AGS will be at a minimum
gain of O.SO as set by R20. Unity gain of the AGS should be adjusted at a counter state of SOH
(center range) by R1S. If "AMPLIFY" pulses continue to clock the counters, they will clock up to
FFH where the JFET gate voltage will be minimum (OVDC) and the AGS will be at a maximum
gain of about 1.50; this is not adjustable and will vary slightly because of differences in forward
transfer characteristics between .JFETs.
,"
4.2.3 Rollover x..
When the input signal to the AGS increases or decreases until it is out of range the AGS
counters rollover, and the CARRY or BORROW outputs of the MSD counter are gated back into
the LOAD inputs of the counters . If the counter rolls over at either end of the count interval (OOH
or FFH), the counters will re-Ioad 80H and the AGS will be reset to unity gain.
10 x 0.75 = 1.00
4.2.6
The gain-adjusted signal is output driven by a unity gain current driver (U10-LH0002CN)
through a 10 n resister for short circuit protection. The current amplifier power pins are heavily
decoupled to ground to maintain clean power supply rails during peak driving conditions.
I DPXDCA
I AGSDCA
I
•• • 0
jJf>'
TP4 1.200 2 .020
These voltages are the settings which are currently used in the Lunar factory.
Use the static counter from the diagnostics software to set the operating voltage to 76 kV and
the current to 1500A. This test may be perfonned with or without the table top in the x-ray
beam, but nothing else should be in the path of radiation. Open the shutter and observe the
counts. If the AGS is operating correctly, the low channel counts should be greater than the
high channel counts and both counts should be stable (approximately ±3000 counts from the
mean value). If the low channel counts are not stable or are lower than the high channel
counts, begin to adjust the AGSDCA (the DCA board in the rear of the scanner) .
If an oscilloscope is available, adjust DCA-T4 until AGS-TP12 does not rollover. An example of
a stable AGS signal as seen at TP12 is shown in figure 1.
~;~;
,~~~i;;.f~'/~~·
~~;
Use the aluminum wedge to perform the AGS Test (see section 6.2) and check the scanner's
ability to give reasonable counts.
TP2 GND
TP3
.'t"l.;;,
-12 VDC .~.':".
TP4
TP7
TP8
TP11
TP12
+5VDC
GND
Output
GND
Circuit Functions
• Discriminate energy levels
• Send TIL pulses to SBC board (DPX~)
\
< \
'-. ' \
Input Circuit
The input i rement of the circuit is 90 D. This is achieved by R 1, a shunt to
signal input is current limited by R2 and clamped to a range from +4 VDC to -
0.7 VDC by R3 and R4, diodes CR 1 and CR2 and capacitor CO. (The differential input
voltage for the LM306 is ±7 VDC max.) The signal is then fed to the inverting input of each
comparator.
Comparators
Each ULD comparator in the circuit has a 220 n resistor in series with the non-inverting input
and a 1Mn resistor (e.g . R26 and R36 for U2) in the feedback loop to provide the necessary
hysterisis. A 150 kD resistor and a 1N5817 diode provide the hysterisis for the LLD
comparators. A 0.11lF ceramic filter capacitor is attached from each reference divider to analog
ground to improve noise immunity when using potentiometers.
4.3.4
1.200 2.020
These voltages are the settings which are currently used in the Lunar factory.
TP16 Amplify
Attenuate
.1$'
".
EFT BLANK INTENTIONALLY
Circuit Functions
.. Control tube current
.. current to
Operations
The MAX and the filament transformer in the Tube Head form a power supply that
current in the X-ray Insert by controlling the filament As the filament
temperature there are more electrons available to be accelerated to t~ anode.
MAX Board
LUNAR® DPX-IQ Service Manual Rev. E (3/98)
2. Power FET's and Snubbing Network: The IRF 530 power FET's, 01 and 02, are
power switches driven by U4. They pull the filament transformer leads to ground, thus
driving the transformer. The snubbing network is connected across the drains of the two
power FET's. This network consists of R1 , C7, and the two MOV's , VR1 and VR2 . This
network is intended to absorb the transient spikes created by the rapid switching of
current in the primary of the filament transformer.
'>1-'••
. ~~.
dc·.
3. FET Driver: The FET driver consis~f (4:{~"rt SG3525 switch mode power supply
controller. However, on the MAX b$.'arcgiti S used as an oscillator providing two 180
degree phase shifted dri~~1 and 02. It has open collector totem-pole
outputs connected to a 1 urce that is separate from the rest of the board to keep
switching noise out . R15 and C16 set the operating frequency of
the MAX at 16-1 R3 set the symmetry (i.e. on versus off times) of the
set both outputs at about 49% on time, 51% off time, slightly
R16 sets the dead time between one output switching off and
This ." ime prevents 01 and 02 from tuming on simultaneously, shorting the positive
center ap of the filament transformer to ground. R10 and C3 provide some power
supply decoupling. R18 and R19 slow the charging/discharging time of the gate
capacitance of 01 and 02. This causes them to tum on and off more slowly and helps
reduce some of the switching noise a faster riselfall time would produce . The wave
forms at the gates of 01 and 02 are seen at TP1 and TP2.
4. Feedback Buffer AMP: The mA input buffer amp (U2B) receives its input from J2, the
mA feedback input connector when. This signal comes from the mA MONITOR output
of the Positive High Voltage Power Supply. It is in the form of an 0 to 5 VDC signal ,
corresponding to a 0 to 5 rnA current output. Op amp U2B buffers this signal. The
output goes to J5, the .mA feedback output conn~ctor, is inverted by U3A and then goes
to the input of the error amp. The output at this connector goes to the SBC to allow it to
monitor the tube current. C8 filters AC components in the mA FEEDBACK input signal.
~.
The mA input buffer amp receives its input from J2, the mA feedback input connector.
This signal comes from the mA MONITOR output of the Negative High Voltage Power
Supply. It is in the form of an 0 to 1 VDC signal, corresponding to a 0 to 1 mA current
output. U3A is used to buffer this signal. The output of the amp goes to the input of the
error amp and to J5, the mA feedback output connector. The output at this connector
goes to the SBC to allow it tq mOl1itor the tube current.
Error AMP: The other half of U3 (U3B), 03 and 04, comprise the error amp. In normal
operation Uumper JB2 in the normal position) a summing junction is applied to the
inverting input of the error amp. This voltage comes from the SBC via J6, the mA control
input. The mA CONTROL input voltage can be measured at TP11. The signal from the
output of U3A is applied to the summing junction of the error amp. The dWerence
between these two signals is amplified by U3B. The gain of this stage ...~at about 24
by R29 and R31 . The error signal and is fed to the base of 03 ~i~~'<;·the base of
emitter follower 04. It is 04 that controls the voltage on the cent~ap. the Filament
Transformer and therefore the insert current. ~~l.,::l(f/>
m
"''., t,-';lf '~, '<:J
o
J.A'~'7.,(
~. ~; . .J."~ -
<'2-. ~~,
~<-~~- ¥- ,~"
El
~"~;";'· -t;t~
r- ~1;i:~~
~~. ~
"{~*?:
As an example of loop operation , assume the insert current starts to drop from its 0.750
mA set point. The voltage at the input of U28 would start to drop. This would cause its
output to drop also. The error amp (U38) compares the mA feedback voltage with the
mA control voltage and generates an error signal. This signal would be the difference
between the two voltages amplified by the gain of the error amp. This voltage is fed to
the base of 03, causing it to start to tum off. As this happens, the base of 04 increases
causing it to tum on further and increase the voltage on the center tap of the filament
transformer. This in tum increases the filament current and therefore the insert current
until it returns to its set point.
As for the rest of the components ; R11 , C6, R13, R27, R28, C10, C17 and C26
decouple U3 and U2 from noise on the +12 volt power supply; R6 provid~ base current
to 04, turning it on unless 03 holds it off; R7 limits current through 03; .c~~d C14
filters ripple and noise on the Filament Transformer center tap. ..~~~~" ,,-,
. ~. :~r'~~
6. Logic and Driver Power Supplies: There are two +12V po~~l$"plies which consist
of two LM7812 volt~ge regulators . On~ power ~upplyg~,f~fbrthe logic circuitry,
and the other supplies power (TP6) stnctly to dnv~~~ FETs. They have
separate chassis and signal grounds in an effo~~switching noise out of the
control circuitry, and share a common groun~y~'fthe 28 volt input connector, J1
(TP10). ~ .",
p .:~ .
7. kV Lockout: This circuit preve[1ts ~~nt drive until the kV is high enough to assure a
stable start of control loop .' ' . ' .,,~~
As the trip point is reached, U2 changes state and the slow start circuit allows 05 to tum
off slowly (10-20 ms). As 05 tums off, control of the loop is transferred to U3B and 03.
During normal operation of the control loop, 05 is turned off and thus is out of the circuit.
Final Notes
1. There is a MAX board "test mode" which bypasses SBC control and external current
sensing. Then the insert current is controlled by adjustment of R4. To configure the
MAX board for the "test mode", set jumpers JB1, J82, and JB3 to short pins 1, 2, and 3.
2. There are two light-emitting-diode indicators on the board. If the red LED is lit and the
greell LED is out, the'filament power fuse (F1) is blown. If both the red and green
indicators are out, there is no +28 VDC power to the board.
Test Points
~
TP3 Ground
TP5
TP8
TP9
TP10
TP11
TP12
Set Point
Clamp Out
TP19 kV Sample
Circuit Functions
Provide protection to the system electronics from transients
4.5.1 Operations
The XORB board is designed to prevent high voltage transients, generated in the x-ray system ,
from entering the rest of the DPX-IQ system electronics. The transient energy is decoupled to
the local chassis ground by three types of solid-state devices, described below. Since the high
voltage power supplies are tightly coupled and in close proximity to the XORB bg&rd, the energy
does not find its way into the rest of the system. <4:;
~. ~~~~~~~>
The XORB board also functions as an terminal for the signals that contr~~1¥ra'nitor the High
Voltage Power Supplies. Test points for all of the important high vOI~,,~trol signals are
located on the XORB board . ~ :!,"'7t,,\,~ j
~~. ~~ff'
Normally, the XORB board has no effect on any of the ~~at pass through it. All of the
signals (including signal grounds) that pass through the~af<:j are connected to chassis ground
through one or more transient suppressors. _. ~r'\i}, ...
~. r~ .~.
If a high voltage transient appears on one lTlQrE: of the signal lines, and it exceeds the
threshold of the transient suppress, it ~e shunted to the chassis ground. Since the
source of the energy for the tran High Voltage Power Supplies) is tightly tied to the
chassis, the transient ene . retumed to these supplies. This shunted energy
never reaches the where it could cause damage.
MOV's are bipolar suppressors which will clamp at the same voltage above and below ground .
The 1N4004's are used in reverse - parallel combinations to clamp signal grounds to ± 0.7 volts
of chassis ground. They were chosen for their current handling ability.
Test Points
TP1 - OA Monitor
TP4 -V Reference
.'\
TP5 + OA Monitor . ~~
.;);) ' :.-~
TP6 + kV M 0)~Olk.j
~'~ ,.
TP7 ~~1:f:'ii~~
+kV
TP8 ~~'ference
...Q~,
.~ .
~ >i'T
.;;.t:;;;':: . I~,it:
Use the anode side of one of the solid stat~s\1srent
: "':~"W'~.
voltage suppressors for ground .
Circuit Functions
• Count pulses from the DPXDCA
• Tube head high voltage and current control
• .. RS-232-C communications with host computer
• Motion control
• Limit switch and thermocouple sensing
~~.
· ~;.;;;~. ~:k
O perat Ions ,,,1> ~?;,~iS;.. " ,'
The LRC Single Board Controller VER-03 (SBC-03) consists of a single ~tib6ard that uses
the Intel 8032 microprocessor with a RS-232-C interface and up to 59!<-~~"i;'ory (8k EPROM,
8k EEPROM, 8k NVRAM, and 32k RAM) . The EEPROM is utiliz~~1ibw easy program
modification and the I\JVRAM is used to retain data during .po~t~.
~~+~. ··Y;ji .
....,."'''';;;,.~
~~, ~~" .
U21 is designated as the . ince the 8032 vectors to OOOOH upon reset. This must
be nonvolatile. U22 co re that operates the SBC. U1 and U2 may be used to
store data or expand
n
U6 (74LS1 ured as the memory and 110 address decoder. Outputs YO-Y6 are used
to access mory while output Y7 is used to select yet another series of decoders U11 and
U34(74LS138's) . U11 and U34 are arranged to operate as a 4-to-16 address decoder which
are used to access 110. The following are pertinent addresses:
U6P13 U1 (NVRAM)
4000H - 5FFFH
6000H - DFFFH U33P6 U2 (RAM)
110 DEVICES
EAXXH U11P10
FOXXH
"X" Represents a "Don't Care" condition; any hexadecimal character will yield a valid address to
select these devices.
Table 4.1.7 summarizes the signals routed between the acquisition computer and the serial
interface
Input Pin# Output Pin # Signal
~t.~
~ .
Table 4.1.7 - Senal Interface Signals Ji,:t.:.t;J:1::>:"
SSC Communications .. '~" ''IS';
The Acquisition Computer and the EXPO-SBC board communicate..~~llindard RS-232 serial
communications link. Instructions toifrom the computer are se'Rt ~ffis link through the
optically isolated serial 1/0 interface, and fromlinto the 8~3£.. ~Pf6cessor chip located on the
EXP-Oaughter board . ~ '¥~ "<-'--
·~rj~ ~ .
are set using special registers within the 8~ ~~serial port is set as COM2 and the interface
serial communications parameters used ar~ollows :
8
1
Even
&~:~~1J:
A 2.0 MHz quarts crystal (X2) in a DIP generates TIL I~~ ~~
7:.').•.
~ie~~'
at
.
"\.:">.'
4.6.6 Event Counting (U13, U14, U17, U18, U19, U20, J1, J2)
Two Intel 8254's (16-bit timer/counters) U13 and U14 are used to generate the count interval
and count the individual events from the LRC DCA-03 boards. U 13 is used to generate the
count interval and to count events from the PMT through J2. High speed opto-couplers (HCPL-
2200's) U17-U20 are utilized for isolation and noise reduction.
Counter 0 of U13 is operated in mode 3 to generate the count interval by interrupting the 8032
(INT1) causing it to vector to the data acquisition routine. The clock line of Counter 1 (mode 2)
U13 is used to count events from Channel 1 (J2P3) . Likewise Counter 2 (mode 2) U13 is used
to count events from Channel 2 (J2P1). Counters 1 and 2 on U14 could be used in the same
configuration as counter 1 and 2 on U 13 to count the other events. The 8254 can count up to 8
MHz, which is well above the limit dictated by the DCA output pulse width of 0~,(3.3 MHz) .
0/ t:~:-'~~1~_ ",f'-
">:~." ':;'':: fi:..::~ •.",.
4.6.7 Stepper Motor Control (U9, U10, U31, U32) .z,JJ.%~~? "-
Independent open-loop control of two stepper motors is accompli~~irprogramming the
8254's (U9 for the transverse motor and U10 for the 10ngitudiQ~. ~9J6r) to generate square
wave pulses (in an a stable multi-vibrator configuration);i~Mo~lection and direction is
controlled by Port A of U23 (8255). B4 of Port B of U?z~~d to determine whether U9 is
generating motor pulses and B5 of Port B of U23 i~ ~~_fO determine if U1 0 is generating motor
.~~~ ~~;
pu Ises. ~, ~
'0>
, . ff,~:~~'
The three counters of each 8254 ( l.f1t):f operate in the following manner. Counter 0
outputs the motor pulse while C unter 2 is connected to OUT 0 and counts the pulses
(their gates are common). aded with 10, and when it reaches 0, OUT 2 goes
low enabling Counter 1 Counter 1 (gate enabled by an output on U23) reaches
the preloaded num 1 goes low and disables both Counter 0 and Counter 2.
Hence Counter er (x10). If Counter 1 is loaded with 300 steps, the 3000
microsteps output from Counter O.
When U15P ~. ~w, closure of limit switch 3 or 4 will stop the transverse motor (U12P6) as well
as close the s'5femoid. When U15P38 is low closure of limit switch 1 or 2 will stop the
longitudinal motor (U12P3) as well as close the solenoid. When U15P1/U15P38 is high, the
state of the limit switch circuitry (U27 and U31) is "overridden" and the motors will continue to
step. This is necessary to move the carriage off any limit switch such as when the scanner is
"homed".
PA2 2
E8X3H CONTROL WORD = 83H PA OUT, PB IN, PCL IN, and ~CH.~~4'"
F4XOH POD 4
PD1 3
0=OVERRIDE,1=NORMAL
UNUSED
O=ON, 1=OFF
"X" in the address indicates a "Don't Care" condition; any hexadecimal character will yield a
valid address for the device.
An 8-bit A0558 (U44) sits on the data bus and latches the .v~e output when U44P10 is
strobed by the decoder. An integral on chip-op amp~h ~im~e non-inverting input of the
OP-77 (U45P3) which is configured as a differential • .,. ·· )~ R43 provides the offset voltage to
's
U45P2 which sets the output offset of U45. A D~~' required in order to operate the high
voltage supply in the 600-800 VDC region ~t~j~::""R44 , R45, R47, R48, R49, and R51 were
selected to yield approximately a 1 . inciim.if1ffor each single O/A set point increment (e.g.,
a O/A transition of212 to 213 wou as'W-ffie high voltage by approximately 1VDC). 02
clamps the output at 8.2VDC to voltages (9VDC maximum input control to the HV
supply).
U37 is used to provide control for the X-ray Insert current. The 0 to 5 volt output of U37
corresponds to a 0 to 5 mA output in the insert current (Le. a 1 bit change on U37P2 is a 1.22
IJA change in the insert current) .
4.6.13 AiD (Remote X-ray High Voltage, Current and Amplifier High Voltage
Feedback Sense) (U35, U40, U42, AND U43)
U35 is a 12-bit AID that is used to sense the high voltage setting, the X-ray Insert current, and
the Detector amplifier high Voltage. U40 is a 5 volt reference for U35. U41 is a 8 to 1 analog
multiplexer that is controlled by ~o bits on Port 0 of U41. Depending on the setting of these
bits, U42 allows either the positive high voltage feedback, negative high voltage feedback, X-ray
Insert current feedback, or the Detector amplifier high voltage feedback to pass through to
U43P3. U43 is a unity gain op amp that passes its signal to the input of U35P13 . U35 then
senses this voltage and places its digital value on the data bus when the 8032 requests it.
, C/:f 'ii. ~
'-----'.
Test Points
TP1 +5 VDC
TP4 Gi'JD
TP7
TP8
TP9
TP14
TP15
PMT HV FEEDBACK
AMP HV PROGRAM
RESET
TP25 ISOGND
TP26 ISO+V
TP27 ISO-V
NOTE: This discussion covers revision E and higher OINK boards. The theory holds for
02C boards but the components have changed. For compatibility issues please
see OXSE0003, OXSE0025, and OXSE0026 in Chapter 6.
Circuit Functions
• Motor Control
• Laser Control
• Fan Control
• Indicator Control
• Error Detection
Motor Control
Motor control signals (TRANS & ULSE, RUN/HOLD, AND REV/FWD) are applied to
the OINK at J1, a dual-row 1 ctor which comes from the SBC (J3). Each signal is
buffered through one 7407. Buffering assures that these signals can sink 16
mA when LOW (i.e
The the OINK through J6 and J7, and connect directly to the micro
ers (Centents) . R35 and R36 are bias resistors , matched to the stepper
motors, s n the Centent manual. The OINK adds optical isolation to the RUN/HOLD
lines. Please refer to DXSE0003 (Chapter 6) for replacement compatibility infonnation.
Position Controls
Signals from the control panel rocker switches are brought in to the OINK via J12, optically
isolated by IS05, and sent to the SSC via J2. C31 through C34 provide debouncing.
Motion Detection
If the Tube Head Carriage should stop during a scan, undesired exposure may occur. Circuitry
on the OINK senses loss of motion and can cut the +28VDC supply via K1. Slotted disks
mounted on the motor gears interrupt an infra-red beam from an optical sensor. The pulses
created by the sensor arrive at J13 and J14. If they should time out at U7, K1 shuts down the
+28 VDC power to the MAX board .
U8 and U9 provide delay to allow use of both the rising and falling edges of the pulses for better
resolution. SBCO-RIDE (active LOW) enables U10, and pulses are counted at U7, a one-shot.
U7 is reset by the pulses from the optical sensors . After each reset, pin 7 or pin 15 drop down
to 0.7V and then C27 or C31 begin to charge. If either capaCitor reaches roughly 2.1V before
the next pulse from the optical sensor arrives , a LONG INT signal (TP8) or TRANS INT signal
(TP7) is created . The transverse time out is set at 2.5 seconds at U7A. The longitudinal time
out is set at 20 seconds at U7B.
When SBCO-RIDE is asserted , U4B gates an interrupt signal should either half of U7 time out.
A motion failure or a dropout of the +28VDC supply, sends an interrupt (US) to the SBC shutting
down the Tube Head . When SBCO-RIDE is LOW, the longitudinal and transverse motion
detection systems are prevented from interrupting the SBC. This is the state of the scanner
except when a patient scan is in progress.
Revision E and higher OINK boards require that R1,on the OMI board be removed . R1 is
required for revision 02C OINK boards however. Please refer to DXSE0003 (Ch~ter 6) for
further information. '''''~\*l '
~ ~~t~ !~~~.;~~.
Motor Kill ~: ~;,';i"
The MOTOR KILL signal comes from the SBC via J2P9. It is active~nd held LOW
during OINK operation. When LOW, it turns on 07, which in ~~~i~es relay K3, closing
solid state contacts, keeping +26VDC applied to J6 and Jl (~~motor controllers) . In case
of undesirable operation, MOTOR KILL would be set HI~~+26VDC would be killed to the
Centents only, and motor action would stop. A 1 N62aa:~n'er diode on the 26VDC line on the
motor side of the motor kill relay suppresses transieNi>$i~'Used by the collapsing field in the
$j~ \,",,~"1 '
.r'~~~ 'rI:>'"
This signal is applied to the 1. It controls one of IS04's isolators, which turns on
06, a PNP transistor us nt drive. The signal leaves at J8 to the Laser PCB.
Solenoid C
The two solen Ids (shutter and collimator) are actuated in response to signals from the SBC
which enter the OIN K at J 1, As with the locator light, one of the ISO's is used to control a PNP
transistor. Current is controlled by an RC network conSisting of a 470uF capaCitor and an 75
ohm resistor. The resistor is twice the working resistance of the solenoid, and allows only a
9VDC final drop across the solenoid. The capacitor allows an initial inrush of current when 05
or 04 delivers +26VDC before the capacitor charges. RC = 38ms.
4.7.4 Fans
There are two fans on the Tube Head, wired in series, powered by the +26VDC power supply.
They hook up at J26. The fans are on at all times.
4.7.5 Indicators
This signal, in the form of current feedback indicating current ~, t0B'e head, is applied to the
OINK from the SBC at J1P10. Determination of Insert 9J!!~"~ater than ;::A11lA is made at
U11 , an LM111 comparator. If current is detected, a si~fi§"'T'assed through IS04 and leaves
on J8. ~t1§;,c'¥':" •
•.~ . '"T$.;--: ~o')
The Control Panel Emergency Stop button's normally closed contacts come to the OINK at J17,
and apply power to relay K1. When K1 is on, +26VDC is applied at J16 to a relay which
controls the +28VDC power supply. The relay coil is also wired in series with the Tube Head
Thermostat. Should either the Thermostat open up or the E-stop button be pressed, power
drops out from the relay, and +28VDC drops out to the HV power supplies . These conditions
are also fed to the SBC via IS01 , U8, Ul, and finally J18, another dual-row 10-pin header.
4.7.7 OP/CAl
The OP/CAL signal is also routed through the OINK from the SBC to the AGS.
0 1 1 1 1 15 Tran~4~l~~btion
1 0 1 1 1 23 oXe·
~l~~;-;;~I Motion
{}<:! '.' . <
,.ti:.,~'
1 1 0 1 1 27 ,§i <~~!, ~8 VDC Power Supply
1 1 1 0 1 29 /;.J;
vr~':
.c'~~.'-W ",
~<~
-
.
~~~':::. ~~'; .
Emergency Stop Condition
Yest Points
TP6 TP12
TP7 TP1
TP8 TP2
TP9
Longitudinal Interrupt
4-37
DPX-IQ Servic€ Manual Rev. E (3198)
Circuit Functions
Reports Motion to the OINK board
Operations
Optical Motion Interrupt (OMI) boards are used to detect motion in both the X and Yaxes. As
the carriage moves transversely, a slotted disk attached to the pulley near the front of the
scanner rotates through a slot in a OMI board, interrupting an infra-red beam of light. This
creates the signal in figure 1 that is sent to the OINK board. Another slotted disk is attached to
the foot-end pulley to detect longitudinal movement (see figure 2). If the approp~te signal is
absent while there is supposed to be movement, an emergency stop cond~:~::~~~ccur .
.". e~~:~.'"
,",·G .!:i.":::
c/:)<~"
4-40 OM18oard
LUNAR® DPX-IQ Service Manual Rev. E (3/98)
Circuit Functions
.•.. Regulation of the power to the Laser
• Send power to Laser indicator
Operations
. The Laser Board is mounted next to the detector in the upper arm. The board has two J
.. connectors, VIN and VlDEO OUT. VIN is the -power to the board and is COlltrolted by transistor
01 on the OINK board . VIDEO OUT connects to the LED on the front panel. B~ the LED and
the laser are lit when there is power to the board. The laser itself is hardwiredJ,~1\: board and
cannot be ordered separately. i;.::~,'ft;~;- .
.... ~....~~\~.
It is possible to destroy transistor 01 on the OINK board by groundi~f~C1 on the Laser
board when the power to the laser is on . There is a groundin~w' " ,;Jtached to the laser board
with one of the mounting screws, if any additional ground(Bg . " ""e attached at this point, it is
possible to short D1 and C1 to ground as the screw i~,,!t, ':'
Astigmatism 11 microns
ppendix4~A
AMP Board Schematics
This appendix contains the schematics for the latest version of the AMP board.?So, .
.~'. c~~~~.r.\~~
NOTE: The revision of these schematics is current as of the timeGPiit;JfiiiSe of the last
Service Manual or Service Manual Update. Older sca~~y have earlier
versions of this board.~.,<t:!-:~;;'<~i .
Appendix 4~ B
AGS Board Schematics
This appendix contains the schematics for the latest version of the AGS board ..;.,
.u·:·, ~~;~JE~~~)
NOTE: The revision of these schematics is current as of the time!fi-f~Je;ase of the last
Service Manual or Service Manual Update. Older sca!Jfi,~(liFc1Y have earlier
versions of this board, ~,,"t;~;&'·'j
Appendix-4-_ C
DCA Board Schematics
This appendix contains the schematics for the latest version of the DCA board .,,::"
...~J\~. ' 2...~.
'.' )..:.," r ':'<" -; "
NOTE: The revision of these schematics is current as of the timeff2f1f:J/ease of the last
Service Manual or Service Manual Update. Older SCCl;~*ill7ay have earlier
versions of this board. .:',~~~~~ ."
Append; 4.D
MAX Board Schematics
This appendix contains the schematics for the latest version of the MAX board. '\,-,~
A ~.~~Si:" ... ~j
NOTE: The revision of these schematics is current as of the timeJPi..~/&asce of the last
Service Manual or Service Manual Update. Older sca~~y have earlier
versions of this board. $'k~~~~'· ;::·"
Appendix 4.E
XORB Board Schematics
This appendix contains the schematics for the latest version of the XORB board::;~_
-;":.
~ r~~;:~I.·:;~:·>
NOTE: The revision of these schematics
.
is current as of the time .~~/ea:se of the last
,'.;~. :. "~'.
Appendix 4.F
SBC Board Schematics
This appendix contains the schematics for the latest version of the SBC board. l[::,.
~. ~ ~~f~~h·~'
NOTE: The revision of these schematics is current as of the time,£Qi1riJeaSe of the last
ServIce Manu~1 or Service Manual Update. Older Sc~~y have earlier
versions of this board. . .
...-:,~.....,. . "'.-
..,.."""",,''<t\i,
Appendix 4. G
OINK Board Schematics
This appendix contains the schematics for the latest version of the OINK board. '<:}1.::, ,
.~ .r~~,'~~
NOTE: .The revision of these schematics is current as of the time,i!i1!eJease of the last
;J'.'t-.~:"' ."......
( (
. Z1 Z2 l3 Z4
~ ~ ~ ~
,
, UNCONTROLLED DOCUMENT. ,
-
Verify Latest Revision
Before Using
.
,"
UNUSED PARTS PWB1
I
uoo
7"lS32 lJ6C
,
~ ~
LPN G>5 REV E f>MI
P\MI
I ,
704lSOO
-
tnD roo. o.\TI! DESCI\1PTlOH Of' awm rmOREV SHEff CURfI£HT
7 .... lSM
CCAREV
~~
3370 WtW7 e ALl. e
3m 4'1M17 R3510 47K1o ""PPO'1-_. ~ ALl. ~
~~
4073 lCW27J117 " RlI&RI8102K7,RllIo2IQ.
° ALl.
°
PS~1A.-04
, J f--t4<
0EP1' . ~N.:. LUNAR CORP.
:::f~~
~J e: 0
~~ 313 w. BEln..a-.aE HIOI--N-JAY
~
eLK "'-"OlSON. WI 153713
USWOH:
K d ll" ~ PS2:501A.-04
OPX
USA
I''''
1
DQCU.AEHT CINHER: DPX ON( CCA <me ~nc DWlANoI
U4A
jo
741S70 R&D
OAAWNIIY: 81~
M8OCIot.TI!D I'U.3 I 0300_0D6H
I TONY DeROSA
A
• C
LUNAR® DPX-IQ Service Manual Rev. E (3/98)
Appendix 4.H
OMI Board Schematics
This appendix contains the schematics for the latest version of the OMI board.~~<
."- £:~~
NOTE: The revision of these schematics is current as of the timeZi(ffB1ease of the last
Service Manual or Service Manual Update_ Older sca~~y have earlier
versions of this board. . ,~:~~'-~>
I
\
/
+5 VDC
1
~'""- ~
SIGNAL 2
A GND 3
'--- R2 ..
Not Pop 220
I I
I > R1
<.
I
I 10k I
I I
- - - -
r~~
~ OPT01
f- H21B1
B B
QPTO-lNTERRUPTER
....
""
I LPN <396 REV
8
c
ECO. DATE DESCRIPTION OF CHANGE DWG REV SHEET CURRENT
CCAREV I
PVIIB1 LPN 4396 c
3581 1120197 Initial C ALL C
r--
-
313W. BELTLINEHIGHWAY
USED ON: MADISON, WI 53713
j
0 0
USA
IF)(
rntJe
DOC. OWNER: DPX OMI CCA2817 SCHEMATIC. .
R&D ISlze -f Document Number .,
DRAWN BY: A 2817-SCH
r~
,
ASSOCIATED FILES I 2817_c.DSN I 4 -t
TONY DEROSA Pate:
I
Tuesday. August 06 1996 ISheet 1 of 1
2 3 5 e 7 I 8
LUNAR® DPX-IQ Service Manual Rev. E (3/98)
Appendix 4.1
Laser Board Schematics
This appendix contains the schematics for the latest version of the Laser board. ,%;,..
NOTE:
.
of the last
Service Manual or Service Manual Update. Older scaI*..~Y have earlier
versions of this board. .. tlfj"~'
\ ,.~. 2 3 4 5 6 7 \
i '\ i
\.. i
./
,...
lASER DIODE ~
~D2
A DIRECT TO BOARD A
N
SOLDER CONNECTIONS
~~7805 LLED
~I 1
VlN VOUT
_2
;j2
2t::: LASER ON INDICATOR
1 1~4004 1
2X. 100 Header
~
;:r:;: C1 ;:~ C2
0.1 uF
220uF !'l
B B
NOT POPULATED
, -=~
,,r- ---------,,
,
, ,
,
I LS ,
II ,
I
~ , ,
I j.
DIRECT TO BOARD , 0iIVE I
SOLDER CONNECTIONS I
--------- - - I
LASER ON AUDIABLE
ECOII DATE DESCRIPTION OF CHANGE DING REV SHEET CURRENT
CCAREV
c c
8f7196 Initial A ALL
3562 A
PVVB1
Ii
U1A
AM!
I
Heatsink
313W. BELTLINEHIGHWAY
MADISON, WI 53713
USED ON:
0 USA
, j D
a:?< [Title .
DOC. OWNER: DPXLASER
R&D !SIze I Document Number
"
ChapterS Olagnostcs
LUNAR® DPX-IQ Service Manual Rev. E (3/98)
c apter 5
Diagnostics
This chapter contains brief discussions of the tools diskette and the error log. ~t.
ti ~:.(~'\
'i~
. ' '!):;::.~ ,~
~~ " . :~.;: "
~~
Diagnostics . ChapterS
DPX-IQ Service Manual Rev. E (3/98)
The Tools diskettes haV€ thr-ee utilities. The software fDr -each utmty resides in a separate
subdirectory on the Tools diskette. The subdirectories are named: Tools, Diags and Monitor.
Tools Directory
There are 6 files in the TOOLS subdirectory.
ERRDUMP EXE . . ~
. ~.
CR ~i . $~~~'
FF ~~ ~~ ~.
PRCONFIG EXE ~"",~" ?i
EDITSCAN EXE & ;r4p.::- W
WINDPMI 386 A ~
~ 1" ~ ~
Most of these files are loaded in to the customer's~~~';;n the system software is
loaded. % .~
. t"~<,~.
The files that are not stem software are CR and FF. CR and FF stand for "camage
retum" and
The Edit S lets you edit a patient's mandatory and optional information, unanalyze
a file, and the Scan Type and Side information in femur and orthopedic hip files. These
are instruction for installing and using the Edit Scan pro.gram:
NOTE: These instructions assume that the DPX-IQ software is installed on the C Drive in
the \LUNAR subdirectory. If the DPX-IQ software is not installed in this location,
substitute the drive letter and subdirectory in which the software is installed in
. place of "C:l.LUNAR" in these instructions.
copya:\tools\editscan.exe c:\lunar
Press [Ent~r]. The EDITS,?AN.EXE file should now be copied to the C:\LUNAR~ubdirectory,
and the EdIt Scan program IS ready to use. . .\~~.>
r(}. ~~~ \0'
~"'.¢' ~.
Next type:
cd\lunar[Enter]
3. A message appears asking for the scan file name. Enter the drive letter, the name of
the subdirectory in which the copy of the file is located, and the name of the file. For
example, to edit a copy -of Jane Smith's first femur scan (SMiTHJOO.FOO} with the copy
located in the C:\STUDY subdirectory, type:
c:\study\smithjOO.fOO[Enter]
4. Use the function keys listed in the menu to edit the file. The function keys are-
described in the section entitled "Function Keys" below.
NOTE: If the Scan Type or Right Side information are changed, be sure to unanalyze the
file {[F-3]) .before exiting the progKJm~ There will be problems apening the file if it
is not unanalyzed.
5. When finished editing, press [Home] from the Scan File Editor menu to exit the
program. A message appears asking if you want to save your changes.
Function Keys
F1 - Patient Information
F2 - Selected Parameters
F3 - Unanafyze File
Home - Previous
[F1]-Patient Information , ,.
Press [F1], Patient Info e patient's information. The patient's mandatory and
optional information ap
Use only E characters when entering information into the Ethnic Code and Base
Filename fie you do not use UPPER CASE characters in these fields, problems may
occur during ana1ysis or database rebuilding.
[F2]-Selected Parameters
Press [F2], Selected Parameters, to change the scan type and side. The Selected Parameters
fields appear (Figure 2).
DO NOT change the scan that are not femur or orthopedic hip scans (for
example, do not change scan to a totaJ body scan). Making changes other than
femur and orthoped may destroy data.
Use the a from field to field. Change the scan type by highlighting the Scan
Type field 9 one of the following number keys:
pine
2 - Femur
3 - Lateral
4 - Tot~Body
5 - Forearm
6 - Orthopedic
For example, to change a femur scan to an orthopedic hip scan, highlight the Scan Type field
and press [6]. To change an orthopedic hip scan to a femur scan, press [2].
Edit the Right Side field by highlighting the Right Side field and pressing either [1] or [2].
1 - True
2 - False
If ''True'' appears in the Right Side fietd, the scan ts a fight f-emur scan. U "Fatse" appears in the
Right Sille field, the scan is a left femur scan .
[F3]-Unanafyze FHe
Press [F3] , Unanalyze File, to unanalyze a file.
A new screen does not appear when [F3] is pressed because the program unanalyzes the file
immediately after it is pressed. To check that the file is unanalyzed, press [F2] , Scan
Parameters. The word "False" appears in the AutoAnalyzed and Preprocessed fields when the
file is unancrlyzed.
These fi1es contain and load the monnor program into t~~M of the scanners SSC
board. This software is a1so part of the System DiSk~\l(fjs automaticaHy loaded as part of
the software loading procedures... " 0'.
¥.
The monitor program may be run by tYPin~Ommand:
DL [Enter]
the latest version of the tools diskette is not used, the correct
monitor version may be overwritten with an earlier version! When in
doubt, contact the Lunar Service Department.
If you wish to see which version of the sse monitor program is currently programmed into the
SSC, type:
FIRMVER [Enter]
Diags Directory
The Diags directory contains 2 files and one subdirectory. They are:
ANC (DIR)
DIAGNOS . ' BAT
DIAGS EXE
The Diagnostic software performs various tests used to verify proper operation of the scanner
as well as to troubleshoot scanner failures. The diagnostic tests are discussed later in this
document.
DIAGNOS.BAT is a batch program to transfer the diagnostic files to or remove ~m from the
computer's hard drive. The loading or erasing includes all of the files found in. r\1'JC
subdirectory within the Oiags directory. The loading or €rasing does not)~"", W"" . files found
in the Tools or Monitor d i r e c t o r i e s . ... {~~ .. '"
~1f. ~"
DIAGS. EXE is the file containing the software to run the diag~S~~'On the scanner.
~f- . ~';~'1'"
DIAGS\ANC Directory ~'\:t~-;l';
The ANC directory within the DIAGS directory contaiJ~t1irectories. These directories are:
'!f?~ ~.I¥"
t~::~~, '<;-~
ENGLISH ~~: .,.~~:::
FRENCH
GERMAN
V·
SPANiSH
Each of these diref"'i'f"'I,nc'es:: software to provide the diagnostic screen displays in one
of the four languag At this time not all screens have been translated into the other
languages.
a:\diags\diagnos load (subdirectory) [Enter]. (all characters must be lower case), where
(subdirectory) is where the executable program is located (Le., dpx*.exe, orthoped.exe, or
sa.exe).
Once the loading procedure is complete, the diagnostic software can be access~, by typing
DIAGS from the C:\(subdirectory) prompt. ? Pf7J;,\
Diags begins witb ssurance Tests Menu selected. To return to this part of the
program frof\l t grams, press (F4).
• ;I...... ~
Tools Diskette
LUNAR® DPX-IQ Service Manual Rev. E (3/98)
NOTE:
~. ~.
Press (Esc) to begin the sets of QuaJity Assurance Tests. Th~Qs'bf the scans will not be
printed out until the completion of ALL the scans. If the~e.@6b'een turned on in the Print
Manager, then the printouts will begin to be printed at1"~nated time. .
os:~~
W~ 'f,.~
3. Perform Air Matrix Scan .. ~,
If (3) is selected, the "Matrix" menu /aP1Qec:lrsr I~mally all of the default values should be used.
Remove the table top and insert aluminum (or an attenuation insert) must be
between the tube head and the If this is not done, the results will not be valid.
If it is not desired that th rea be scanned, smaller numbers may be entered for
Scan Lines (longitudi ) and Scan Points (transverse scan length). When installing
the scanner, al It settings so that the alignment over the -entire scanning
surface is exa
If "Stop" is ~aea~a r Detector Motion, all of the data will be collected in one place.
Press (Esc) to begin the scan. The "Matrix-Acquisition" screen will appear. Write down the file
name in the upper right hand comer.
F1 Overall Statistics
F2 Line Statistics
F3 Column Statistics
F4 Deviation Image Home - Previous
Selecting (F1) will display the statistical analysis and ask if a printout is desired. The F2 and F3
options are not implemented at this time. The CF4) option shows a color pattern of each pixel's
deviation (in units of standard deviations) from the mean value of all pixels. There is a choice of
High Channel Counts, Low Channel Counts, or Worst Case (the computer selects the most
deviant of the two channets). The Deviation 1mage can be very hetpfu~ 1n determ1ntng how to
take corrective action when the Air Matrix Scan fails (see "Failing Air Matrix Results" section
8.10).
5. Static Counter
The Static Counter is one of the most useful of the Diagnostic programs. The Static Counter
program was PI il Ilarify designed 10 take sampie counts of the 38 and 7D keV channels, .and
display the counts on the monitor screen, print them on the printer, or dump the results to a file.
However, this program can do much more.
CAUTION: Do not operate at over 1000 .~.f .' '. ~",. 'oW?than 5 minutes. And if
LL ,
•
o~rated at hig~ current for ~~s, allo~ a 1~ minu~e cooli.n g .
penod. Operation at 75n;~,~lower dunng thiS cooling penod JS
acceptable. ~~:~-
NOTE: When kV and mA pa~n'J~t{.:af'CS!~ set in the Static Counter X-rays are being
made regardless . is being done. All time limitations start when kV and
mA are set, they are lowered to 750 pA or less, not when
sampling I
~~------------------------------------------
Using the arrow e cursor to the Operating Voltage or Tube Current entries. Use
the backspace some spaces, and then type the appropriate number. The voltage
and curren as soon as the (Enter) key is pressed.
Press (F1) an use the front panel position switches to move the Tube Head and detector to
any location. This feature is useful for positioning the x-ray beam for the alignment pictures.
Press (Home) to leave the scanner at the present position and retum to the scanner control
screen, or choose (Esc) to aHow the scal11"1ef to automatically adjust for the Pat1ent Locator Light
offset.
Use the arrow keys to move the cursor to the Collimator entry. Press (1) for medium
collimation, and (2) for small collimation.
The Sampling Parameters determine how often the SSC reads the 38 and 70 keV counts and
reports these counts to the program. There should not be any need to change the Samples per
Second or Number of Samples value. It may be useful to change the number of repeats to a
large value if an x-ray beam or counts are needed for a long period of time. Otherwise 50
repeats (the default value) is perfect for making alignment pictures.
Data Corrections should be nonnally left OFF. When Data Corrections are switched ON-,- the
software will correct the counts for signal superpositions caused by multiple photons striking the
detector simultaneously, and for dead time (the inability of the counting electronics to react to
two photons arriving at the detector nearly simultaneously).
The Data Output may be directed to -a printer Of a fHe 1n add1t1on to the disp1ay-on the monitor
screen. To print the results, move the cursor to ''To Printer" and press (1) to tum this feature
ON. To save the results to a file, move the cursor to ''To File" and press (1) to tum this feature
ON. Move the cursor down one more line to set the file name, or allow the data to be stored in
the default file C:\STATIC. LOG.
After the Scanner Parameters, Sampling Parameters, and Data Output modes h~ye been set,
the sampling process can be started by pressing the (ESC) key. ~~~,
<t> a{f..~~~t.
it .-~.. i[,f,,~ ,"
6. Qwck CoUimation Ratio ~~. '%» ~.
This test runs part of the daily QA to test the collimation ratio in oU,r",f~'i\'
~. A~~';>'~,"
~4;~~~' '
-J!1s.~;i,\''''''
"'~"~~o.. ~.""~'
~~~; --=
,;fi;.
,7'~$h.::~
~~~~..
,~
.". " O"~~~' ,:,
j?r--.-.- "'."'~
fi
r.\:)o'
~.
~k';r
~,....
.\ /\V,":J"
:0 \ {\ ' '-',,- ~~, ~~'~~" . ():"'L' . ~-
The setting and verification of the Limit Switches can o~~~lle with the Table Top removed. '.
Once the desired switch is selected, press (1) to s~w~:~S~o verify. When \1) is pr~ssed, the
program asks that the scanner be moved to Ho~~the Front Panel Position SWitches.
Wrth the End Doors and Front Panel remov,5; ' 1 t~,Jcrme Position Fixture, which is part of the
Alignment Kit, should be positioned in ~e . . comer of the scanner. Use the Front Panel
'-- e
Position Switches to move the s . ' ... until -' large collimator hole is directly beneath the hole
. -j in the Home Position Fixture. and the scanner will move to the exact position
where the Limit Switch sho t the Limit Switch, and press (Esc) .
Once the Limit _'.. ',Tr- .. program allows for verification immediately. The verification is
the same as se . The Position Switches are used again to align the large
collimator h ome Position Fixture h01e. When (Esc) is pressed, the scanner will
move until Limit Switch. As soon as it closes the Limit Switch, the program will
displayw r not the limit Switch has been set within aUowed tolerances.
4. Home Scanner
Moves arm to home position.
F2 Motor Diagnostics
1. Pos1tion Contr{)~s
If (1) is pressed, the Front Panel Position Switches may be tested.
2. Distance Accuracy
If (2) is selected, the source and detector will be moved either 100 mm transversely or
longitudinally.
3. Speed Accuracy
If (3) is selected, the transverse or longitudinal speed is calculated and displayed on the screen.
No speed tolerances are given or have been set. ~
~.~~,
~-:., i~~ ....
4. C able Stress B um-In , ~~~ ~'-;;~Ji"
This program runs the scan arm several ways to stress the cables. ~ ~~;> -.,.
. ~~ ~iifJf'
..-..:.t"'7,. ~.~;\
F5
17 of the Version 3.4 Operator's Manual.
F6 Select Language
NOTE: If the computer is rebooted after an error has occurred, the error is not usually
saved to the error log.
15
23
27
29
30
66
71
Faulty Negative Power Supply
NOTE: Although error codes 78 and 80 usually mean the above, they can mean that
either power supply is bad or that they both are.
The program used to access the error log is ERRDUMP.EXE in the C:\LUNAR directory. This
program is the only way to display the information in the ERROR LOG. To use this program,
exit the DPX-IQ system software to the C:\LUNAR> prompt. Then the options for using
ERRDUMP are:
This will send the information to the plinter for a hard copy w~J(jjttTfen~may be faxed to
LUNAR for analysis. ¢l)!~,~~,
~_ ~J;;.""
~
~-'l-:"~'
~~~~P
&'~~O'
~~~.
-'
~~.
~ ~\~~
',. V
f} ,~~
Chapter 6
Procedures
This chapter contains procedures for testing and replacing parts. The appendic~are all
uncontrolled copies of GMP documents and as such are subject to change~wi~~~otice.
~~~t~'
6.1 Peak Adjustment Procedure ......................................... '" .. ~..f~~~";~:'¥: ....................... 6-1
6.2 AGS Test Procedure .....................................................~'-.~~~ .............................. 6-7
6.3 Reloading LUNAR Software ...................................~~ .~.:~.:" ................................... 6-9
6.4 Tube Head Replacement. ..............................~"~\. C,;.. ::~~ .....................................6-11
,,'" ~J.<. ""~B).......................... ".................. 6- 13
6.5 Lower Cable Bundle Rep Iacement Proced ur~.~::
6.6 Upper Cable Bundle Replacement Proce~};~ ....................................................6-17
6.7 Tube Head Control Cable Replaceme~1Q:cedure ............................ ......... .......... 6-19
.* .~""'i
6.8 Patient Compressor Strap.: ........fI~.:t;~z.0':"' ........................................... ".................. 6-23
6.9 Returning Parts To LUNAR (For ~atjonal Distributors) ....................................6-25
--- 6.10 Tests To Be Performe , ~:, .............................................. ! .. . ............................... 6-27
Procedures Chapter 6
DPX-IQ Service Manual Rev. E (3/98)
Chapter 6 Procedures
DPX-IQ Service Manual Rev. E (3/98)
The last 'peak value obtained during a Quality Assurance scan is stored in the Quality
Assurance History database file. If the next peak value obtained differs from the last value
stored in the database by more than 50 units, the Quality Assurance scan will fail. If the peak
must be shifted by more than 50 units to return it to 500 units :!:25, the peak must be adjusted in
steps of less than 50 units (40 unit steps are recommended). A complete QA must be run
between each step to update the Quality Assurance History database file.
I.f) n
FIG 6.1 DPX-IQ sse (t p e 0\ 1')
'\\ .
/\' V\\) \'~\V
"
{J\\
.(\, '
\
)-\
).
J\\~
tG ~'
J...~ I ~
;'11
',., !,.1
/ ;~':cr
\. f,
I ~ .~
,\...J':::)
(
~
Jl}
J)/)
r\'\J
,.
..- \. -...
'v\~ \}l'j
DPX-IQ Service Manual Rev. E (3/98)
A . If the software peak setting corresponds to where both the high and low channel counts are
at a maximum, (see fig. 6.3) follow the steps outlined below. The graph on the QA will look
similar to the one shown in fig 6.3 with the dashed vertical line (where the software thinks
the peak is) on the graph at a point where both the high and low energy plots are maximized
(the dashed curved lines). If the graph on the QA is unlike fig 6.3, go to part B of this
procedure.
1. Load tt"!e ,~~f'~"'software located on the TOOLS diskette (see seGtion 5.1).
:~~'~:':';~~'" ~t?}~·:. '~-':> ':\ '
2. T~DtA$$/:~nd from the menu, select "Peak / Set AMP Gain", then select "Perform
PeaJ\:~ahd [Esc] to run the peak. Observe the results generated by thE~ QA to
determine if the software has selected the Amplifier Setting where both high and low
channel counts are at a maximum. If not go to part 8 .
3. If the peak requires adjustment, (is not within +/- 25 units of 500) connect a voltage
meter to the SSC at TP18 (see fig 6.2). Put the meter to the VDC scale and record the
voltage reading (e.g. 0.703VDC)
!)~a,
I:
Using the voltage recorded in step 3 and the peak value recorded in step 2 determine
the appropriate voltage adjustment for R43.
If Peak <500
*00 Inot attempt to adjust the peak by more than 40 units as an adjust~.nt
1;",
without a
complete passing QA in the LUNAR software will cause a peak fa~lur~~e QA. For
Pea~~ settings that are more than 50 units away from 500 mUltiP.J.~~~ents are
required. . ~ ~'?
. Mt,,~
7. With your voltmeter attached to TP 18, adjust R43 (~e~j~Lmtil
your voltmeter
reads what you determined in step 6. ~'. ,~~~
8. From the main menu press F3 System QU~'\:~ce and place the System
".. ' rt up and at head end of table. Press
Standard on the table silk-screen with ~~~.
[Esc] and allow the QA to run in itr:Wand to print out. Verify that QA passes.
9. If peak value is not withi ," nits~O, record peak value from the QA that was just
run and the current 18. Using these values repeat steps A. 3-8 until
Pea~~ is within 25 u d a complete passing OA has been run.
S. If software is unable to identify the correct peak (the dashed line does not line up with high --
points of both the high and low energy plots as in the figure below), use the following
procedure:
b£Tt.CtOlt,
". · 1
'.%50 -----
laO .
. 3';0'
400
.-;; -, '~;';<' ...so '. '
;:::.::;:?,:;~:\;" .1:::~, ~::~ ~~"M+M ~",n'" ....• 9:0
:\'::::"':':'/'.: - - '.. - ,,:fWII:,~~-aJtl!'A_al,HIQlUlI1U11I S!SO
1. Determine the Amplifier Setting where both high and low channel counts are at a
,maximum, for ~xample in fig. .f.6 .4,at ~OO\mits. (LoW keV = 397258 High keV=28B363)
'00 DO,'s Yf7f,-t,5fic r O::; n,~ rt1 e5 YJ;-LJ t~/YJ lij· .
i
2. Measure the DC voltage~atSBG TP18 for example 690mV (see fig. 6.2).
:..... ." . . . . ..
3. Use the foliowingJorn1~la to adjust the peak using R43 (see fig 6.2) with your voltmeter
attached to. TP18on,tne
. ,:-::.'.
.
SSC. All voltages are in millivolts.
Ne~ Vms ~ ··='Old V TP1S + {Estimated Correct Current Peak Location - SOO)/S
Usin~fthe sample readings in steps 1 and 2 above, the new voltage at TP18 that will
correspond to a Peak Setting of 500±25 units should be:
4. If the formula above yields an answer that requires the peak be moved more than 50
units, (the absolute value of [300-500] = 200 >50) the peak must be moved in ste!ps.
Start with the peak setting of the last passing QA. Move the peak setting first toward
the Estimated Correct Current Peak (in a step less than 50 units) until the scanner
recognizes the Estimated Peak as the correct peak and a QA passes. (The Low and
High Channel counts are maximized at the amplifier setting which matches or nearly
matches (a difference of less than 50 units) the peak setting on the QA).
Remember that a complete passing QA must be run between each step.
5. Run the peak and ensure the software is able to identify the setting where both high
and low channel counts are at a maximum. If this peak setting is not within 25 units of
500, retum to part A of this procedure to make the final adjustment.
Peak Adjustment
LUNAR® DPX-IQ Service Manual Rev. E (3/98)
C: [enter]
CD\LUNAR [enter]
Insert the TOOLS Diskette in the A: drive and load the diagnostic software. Type:
Adjustment of the AGSDCA may cause a change in BMD results. Scan the aluminum
spine phantom three times before and after artY adjustment, and compare the averages
of the three scans. Run 5 Quality Assurance scans prior to the spine phantoml scans to
properly calibrate the densitometer.
Before adjusting the AGSDCA to improve AGS performance! please review the discussions on
Performance, Operation and Adjustment in section 4.1. Lunar strongly recommends that if an
adjustment is to be made, all of the DCA c;md AGSDCA voltages (T4 - T7) should be reset to the
values in table 4.1. This adjustment procedure will only work correctly from the proper starting
point.
1. As you watch static counter counts on the monitor display" adjust the potentiometer
nearest to AGSDCA-T4 up by 10 mV. () ~j) .' '._~ .}!')j C r= ;:O! I)'/)/.l-ru-
'}J::s" 5.>
i
'./le -"-:' t: ",\ . :.... , . / ..:: / "': , ';" ~~\)'\ J~ t"'~ ..::_-)',...~t-
u,
.
b. the low channel counts are more than the high channel co~~e counts are
stable, and the counts pass the following verification te~, .~' .
Verify proper operation with the aluminum wedge. Press tbe ~~;and
note where the
'''':t'.:~Qtable top. Press the <Esc>
center of the red Patient Localizer lamp spot is posiiltlon
key and the collimator will move into the localizer la ""s.iftOno. Press <Esc> again and the
shutter will open. Allow a few samples to appea~. ~ onitor, and then move the Aluminum
Wedge slowly through the radiation beam, pti_~ifh the thin edge of the wedge. Move the
wedge very slowly until the beam reacges yminum thickness of approximately 2 mm.
In the results displayed on the the sample number where the high channel air
counts begin to be reduced. the counts in both channels should be reduced for the
following samples. In oth the high and low channel counts must decrease
consistently as the beam.
If the low increase at any point instead of decreasing, the AGS is not properly
adjusted. use "bone points" or white pixels in the image around the outside of a
patient's T y scan.
The Main Menu of the Lunar Software shows which version is presently on the system.
Begin by making sure the System Diskette is write-protected (or READ-ONLY). ~. the default
drive is the A: drive when A:INSTALL is typed, the software loading procedurl:'~ttempt to
e:ase all ~e executable files on the A: diskette. If the A: diskette is. not Wi~ ' -:" ~ ~ed, the
diskette Will be permanently corrupted. Both square holes on the dlske .' ,. ' e open. If only
one hole is open, move the slide on the other hole to open it also. T "-' '>.''_ diskette is write-
protected. ~.. . tJi:-;. ~'
. J
~.~~
When reloading software, you must be in the root dire~~~ C: drive. Get there by typing
at any C: prompt: :A ~~' "ie~
P~~t~: .
CD\ [enter] p t ~~~-#<
Insert the Lunar System Diskette i '·" e A~!e'~nd type:
A:install [enter]
When new '~!:"~I'" software are released they may not be loaded in the above manner. BE
SURE TO F THE LOADING INSTRUCTIONS THAT ACCOMPANY ALL SOFTWARE
UPDATES.
CERIUM FILTER
DPX X-RAY ~
TUBEHEAD
ASSEMBLY
LEAD RING
----..,0~
---'U
COLLIMATOR . , /,. ," ::: WAR N I N j(; I
STUD ,'~
"..~. :
w .''.-.
'. ! \ DO NOT
BRASS REMOVE
PLATE WHENTHE .
,,: • REPLACING THE TUBEHEAD
~ .
-.-------BRASS WINDOW PLATE
SPRING
(BILLIVILLE
WASHERS
. .. .. {}~,: .'~;
?';;~;~)/I
TOOEHEAD
ADJUSTMENT
SCREW o
Shelf Life
Many distributors have asked about the shelf life on the Tube Heads in their stock. The shelf
life of thE~ Tube Head is not accurately known~ but we feel at this time that it is over two years.
Perform the following Tube Head inspections on all replacements before mounting them:
1. Check the shelf for oil spots which would indicate that the Tube Head has a leak. If oil
is discovered, retum the Tube Head for warranty replacement.
2. Shake the Tube Head and listen for the sound of an air pocket or bubble inside. If a
bubble is discovered, retum the Tube Head for warranty replacement.
If a Tube Head has been on the shelf for three weeks or more, perform the wa~p procedure
described below. " 'i.~~'::.\ .,
3. kV is reached.
Necessary tools: 1/8" and 5/32" Hex Wrench, Small Screwdriver, Diagonal Cutters, 3/8"
Wrench, Large and small wire ties.
2. Remove the head and foot end panels. At the foot end of the scan table unplug the
AlC power cable to the scanner.
3.
4.
::::::::::b~:O:::~::::::~o~rt::r::~e(mo at each en:,~~7:'
5. On your new lower cable bundle, use a diagonal cutters to ~~V;ire ties holding the
cable together, also unfold and read the LUNAR SE~y~~TES attached to the .
new cable bundle. _~ .~:;:l;;it!9
~~~~~.
6. Lay the new cable bundle flat out on the fl?.-(~~e the end with the (2) 16 pin
connectors and the yellow and orange~~nnectors, this is the end that connects
to the bulkhead, this is the end to .§.lart_UnstalI action from .
.. _~ ~" !t.l
~~t;"'
7. Locate the bulkhead on - nnenhere are (3) wire ties that hold thE~ cable bundle
in place between the nd the cable track. Before cutting any wire ties note
the way the cable down and the way the cable lays before it enters the
cable track. T must be fastened in the same position and lay the same
way.
8. (3) wire ties between the bulkhead and the cable track with your
. Disconnect the two Lemo connectors (orange and yellow) by
gra the Lemo connector between the bulkhead and the colored strain relief and
pulling outward.
9. Next disconnect the two white 16 pin molex connectors. These connectors have a
catch lock located in the center of the housing which must be depressed to disconnect
them from the bulkhead.
10. If you try to remove the cable bundle by pulling it back towards the cable track you will
notice a 3 conductor gray wire cable tied to the high voltage cable bundlie. This wire
carries the transverse optical interrupt signal from the front carriage (we will refer to
this wire as the OMI cable). To find the location of the OMI board and the front carriage
follow the OMI cable. On your new Cable bundle the OMI wire is connec;ted through
the bulkhead.
11. To remove the OMI cable push the x-ray tube head all the way to the outside of the
scanner (patients left side), then notice the clearance between the front panel and the
high voltage cable bundle connecting to the x-ray tube head and the way the cables
are tied down as you will need to wire tie them back the same way.
12. Now you can cut the (4) large wire ties and the (2) small wire ties that hold down the
high voltage cable bundle and OMI cable. Then you can pull the lower cable bundle
back to where it enters the cable track.
13. Next take your small screw driver and pry open (20) of the cable track plastic retaining
clips; that hold the cables into the track. These clips pivot at the back of the cable track
and can either be removed or left in the open position so the cables ca~e removed .
.~~~
14. Pull the old cable bundle out of the cable track towards the foot,~~·table.
15. Now install the (3) conductor OMI cable from the bUlkhead~\;i\t1I BOARD. To.do
this you must remove the white 16 pin connector frog}, t~}IJlead labeled bulkhead
2, this connector is a part of the upper cable bu~l~ new OMI cable will
connect into this housing. Note the position of~~ that holds the upper cable
bundle in place and then cut it in order to p~:ctnnector out to install the OMI
~bl ,e. Note th~ orientation of the .e. xisti~ ~!ilals in this bulkhead. connector as they
. e
slide! and lock Into the connector I~ . ~ctlon only. Where the wires enter the
connector each hole is numerjcal~..#.red. (Please read the attached LUNAR
SERVICE NOTES) Put n wire into hole # 11, and the black wire into hole # 9,
and the red wire into
16. the bulkhead braCket the way it came out and fasten the
uppHr a large wire tie the way it was before the original wire tie was
cut.
17. OMI cable in the same position as the previous cable and connect it to
the rd. Then tie the high voltage cable bundle and the OMI cable down to the
high voltage cable trough,(there are three large wire ties near the bulkhead, two large
wire ties put together to tie the high voltage cables down at the center of the high
voltage cable trough, and two small wire ties to anchor the OMI cable down to the
Bracket before it connects to the OMI board). Before tightehing the wire ties assure
that the proper clearances and no stress is applied at any of the connectors or wires.
Tighten the center cable ties first (two large wire ties put together and located at the
centler of the high voltage cable trough). Be sure the high voltage cable bundle does
not hit the front panel when the tube head is an the way toward the outside of the
scanner (patients left). Then tighten the three wire ties near the bulkhead making sure
that lthe high voltage cables do not hang over into the cut out space on the high
voltage cable trough as they may hit the longnudinal motor bracket and cause
positioning problems.
\ ,
18. Take the new cable bundle and connect it to the bulkhead by routing the cable bundle
the same way as the replace cable bundle, (note the connectors are labeled bulkhead
1 and bulkhead 2) and fasten the cable bundle down in the three spots between the
cable track and the bulkhead with wire ties. Lay the new bundle down into the cable
track making sure aI/ the cables lay parallel to each other. Start replacing the clips to
the cable track as far as you can. Continue the process: remove a few clips, extract
the old cable bundle, add in the new cable bundle, and replace the clips.
19. To fin'lsh removing the old and installing the new cable bundle in the lower portion of
the cable track you must remove a screw located at the end of the cabll~ track. To do
this use a 1/8" hex wrench and a 3/8" wrench. Once you remove the screw you can lift
the cable track out of the channel it rides in and finish prying open the q~aining clips.
Remove the old cable bundle until it reaches the cable entrance~ hle~e all the
cables enter the electrical pan area. Then finish installing the ne,t.. : . ~ ~undle until
you reach the cable entrance hole, finish clipping all the cliP.S at«:~, . replace the
cable track hold down screw (do not tighten the screw at ~w.
20. Locate where a/l the wires from the old cable bUgdl~aT: on the electronics pan.
All of the wires are labeled except for the wire~~~o the terminal block. Here is a
list of wires that must be disconnected befo~~ng the cable bundle from the pan
area. SBC(J-5), OINK(J-17), OINK(J-1ft~(J-14), OINK(J-5), OINK(J-8),
PMT/AGS(J-11)(J-11 is the yello~rinector), the orange Lemo connector
(connected to the H.V. Lemo .J.)o~.~ the last three wires connect to the Terminal
block. Write down the n . where the orange, blue and yellow wires go to on the
terminal block and di m with your small screwdriver. Carefully pull the old
cable bundle thro entrance hole from the pan area. The old cable bundle
21. connector at a time) the new cable bundle wires through the cable
the pan area. Connect the wires to their proper locations.
22. the arm all the way down to the foot end of the scan table. At this time
make sure that the Lower cable bundle and the high voltage cable bundle have some
slack where the cables enter the cable track near the bulkhead area. Then make sure
the cable track has at least 1/8" of clearance from hitting the frame. To adjust this
slide the cable track in the trough it rides in to get the proper clearance, then tighten
the screw that holds the cable track in place with a 1/8" hex wrench and a 3/8" wrench.
Move the arm up and down the scan table to be sure there is nothing wrong with the
cable track motion. Leave the arm at the foot end of the table.
23. Leave a little slack in the cables from the cable track to the pan area to allow all of the
cables to be routed and wire tied exactly like the original cables.
24. Reconnect the AlC power cord, put the table top, pad, and end panels bClck on the
scanner.
25. Turn on the scan table, computer, monitor and printer. Do a daily calibration to verify
thalt everything passes. Do a phantom scan and try the position control buttons for
proper operation. Verify that the patient positioning light is working. continue to verify
that the scanner will also pass a phantom scan.
ools Necessary: Small Phillips Screwdriver, Small Slot Screwdriver, Diagonal Cutt~r's, Hex
Wrench Set, Mini Molex Pin Extractor, Large and Small Wire Ties.
1. Tum off the power to the scan table. Remove the table pad and both head and foot
end panels . Next, remove the table top by removing the two allen head screws at
each end (use 1/8" hex wrench). Lastly, remove the column cover located at the back
of the arm held in place by (10) Phillips Head screws.
2. Remove the bottom arm cover by removing the (8) screws that hold it i~place (use
9/64 hex wrench). Next, remove the top cover (4 screws, 3/32 hex w~J. Remove
the column cover lastly. i-, ,~~:,::~ ,
.~~~.
3. Open the cable track and cut the wire ties at the end of it. ~~ect all the cables
coming from the cable track and remove the cable b~~~cfr as the top of the
column (through the hole). -$' -:- ., ?~.
£:~A~O~
4. Place the new cable bundle in the cable tr~~1tQ~nnect all cables to their
appropriate positions. Close the cable , ~itrci~~ire tie the cables at the end of it.
Route the cable bundle throu;h v y t n e top of the column.
Next locate the Laser and Oisconnect it. Remove this cable to the top of the
column cutting any n 're ties.
7. -'f the arm column follow the cables vertically down and cut all the wire
ti upper cable bundle to the other cable assemblies.
3. Carefully pull the upper cable bundle out towards the back of the arm and down
through the holes in the column.
). Next cut the wire ties that hold the cables down near the bulkhead. these are located
in the lower rear portion of the arm.
10. Make sure the bulkhead connectors are labeled on both sides of the bulkhead bracket
and then disconnect them.
1. Next remove the two (16) pin Molex housing connectors that are attached to the upper
cable assemblies from the bulkhead bracket. Also disconnect the yellow Lemo
connector from the bulkhead bracket. Pull the housings and the Lemo connectors out
to work on them.
12. Use the mini Molex pin extractor tool and remove from bulkhead 1 pin16 and pinS.
From bulkhead 2 (the lower connector) remove pin15, pin16, pin6, pin?, and pinS.
13. Now remove the old cable bundle by pulling the cable out through the back of the
column. The old cable may now be discarded.
14. Next route the new cable bundle through the column wire tieing as you go.
15. Route the taped end of the cable through the hole at the bottom of the column to the
bulkhead area (follow the other cables that go to the bulkhead).
16. Carefully remove the masking tape from the cable exposing the mini MC?!~x pins and
the Lemo connector. Note the orientation of the existing terminals in~lkhead
connectors and the numeric labels at each hole of the connectqr ..(~ ~~:\~':.-
~~;"$
17. Connect the wires as follows: ~ ~~t~ ~
.~'-#:'"
"'C'"~~_~"
.------------.-----------~~k4:f'
Color
Yellow
Brown
White
18. Insert the connectors into the bulkhead bracket the same way they were removed and
fasten the upper cable bundle with wire ties as Original.
19. From the wire tie just fastened, the cables going into the arm column must not have
any slack in the cables (pull the cables from the column area to remove any slack).
Make sure that none of the cables hit any of the mOving parts in the narrow area where
the cables come from the back of the column out to the bulkhead.
20. Mount the column cover back in place and replace the top and bottom arm covers.
Replace the table top, table pad and the head and foot end panels back on to the
scanner.
21. Tum on the power to the scan table. Run 2 daily QA's, 2 phantom scans, and an air
matrix and verify that all tests pass.
2. Remove the head and foot end panels. At the foot end of the scan tablE! unplug the
AlC power plug to the scanner.
3. Take off the table top by removing the 4 screws (two at each end).
5. Take the black tape off the high voltage cable bundle at each end of ~~eving
(open any cable track clips that are in the way as in step 8) as wetk~~~'wire ties
holding the bundle in place. Next, slide the sleeving away fro~~e~ head and
remov~ the tape beneath the sleeving. Note the position; ~~cable before
remoVIng the tape. The cables must be returned to ~:eositlons.
4 · ~'t
6. Disconnect the military-style connector and th~~wire (if one exjsts) from the
tube head and slide the cable out of the bla,~~'1mg toward the cable track.
~' 'Y
~. "iii?J.
7. Unplug J3 and J5 on the XORB b~.~o~ trace the ground wire (if one exists) from
the control cable and discon~ct W
tne power supply. Cut the wire ties holding the
control cable to the pan.
8. Pry open the plasti ips that hold the cables into the cable trclck. These clips
pivot at the ba track and can either be removed or left in jthe open
position les. The screw located at the end of the cable track near
the removed. Once this screw is removed the cable track can be
li~~ nne! it rides in and the remaining clips can be opened. Mark the
p-. cable track screw on the frame to retum the track to its original position.
9. Remove the old .control cable. Note the position of all the cables in the cable track.
They will need to be returned to the same position during this procedurE~ .
~
.~! _C_A_U_T_I_O_N_:__I_f_th_e---.;;...ca_b_l_es_a_re-n_ot_r_e_tu_m_e_d_to_th_e_ir_o_n_'9_in_a_I__P_O_s_it_io_n_s__
damage will result.
, ca_b_le_ _ __
10. Feed the new cable, military connector first, through the black sleeving on the high
voltage cable bundle starting near the bulkhead. The sleeving can be expanded by
pushing two parts of it together to better fit the connector through.
11. Once the cable is through the sleeving, connect it to the tube head and connect the
ground wire (see figures 1 and 2). Position the cables next to the high voltage cable
bracket as in figure 1. The collimator control cable should come out of the cable
bundle between the two high voltage cables toward the back of the scanner. Tape the
cables together for several inches with three layers of electrical tape starting as close
to the bracket as possible. Continue to spiral the tape around the cables as it was
origlinally and finish by taping three layers for several inches near the bulkhead. Tape
the sleeving near the bulkhead as it was originally. Pull the sleeving tight along the
bundle and tape near the tube head. Wire tie the cables to the tube head high voltage
cable bracket.
n 9 CDANGERJ@
High Voltage! ~
Top View of
Tube Head
I I
"- /
~
••(J\<
/
.
0«;-'
~.~4,:.
(i; Ji91~, ,~
...( "'" ~;.. ~':"':II'
.f~ "\t,~~
1t~: ~,
~;o~~r
t!\~,,~
~'\s'the
12. Lay the control cable in the cable track in the same original one. Start
replacing wire ties between the bulkhead and tQ~!~ead. Before tightening the
wire ties assure that there is proper clearance~~~tress is applied at any of the
connectors or wires through the entire ran~J~~~tion. Tighten the center cable ties
first (two large wire tie~ put together a~~d at the ce~ter of the high voltage cable
trough). Besure the high v9Itag~.l~ndle does not hit the front panel when the
tube head is all the way th~t of the scanner. Make sure the high voltage
cables do not hang into out space on the high voltage cable trough. Tighten
the three wire ties n ead.
13. ble track and reattach the track to the scanne!r frame.
14. e foot end of the scan table. Check that the lower Galble bundle and
e cable bundle have some slack where the cables enter the cable track
khead. Check that the cable track has at least 1/8" clearance of the frame.
st this, slide the track in its trough, then tighten the screw that holds the track in
place. Move the arm up and down the table to verify that the cable trac!k moves
properly.
15. Reconnect the AlC power cord, and replace the table top and end panels.
3. Stand behind the scanner with the Patient Compressor Strap extending away through
the back ledge and towards the front of the scanner. Wind the strap on the Roller
Assembly, winding with the top of the roller turning towards yourself. .~~
4. When the entire strap has been rolled onto the Roller AssemblY",. ~~ROlier
Assembly into the end blocks. The left block has a slot in it to ~~e spring axle.
Tum the top of this block towards yourself for 20 turns. 'S'~~>'~'
~ '~i'
5. Bolt the end blocks to the scanner. -Zt
Ik"t\~<'~"
~\t~[;~ :?
Shipments of defective parts being retumed to LUNAR from intemational distributors should add
the following instructions to the air bill and proforma invoice:
Parts must also in 90 days of LU NAR opening a service report for them (usually
the day they a All parts not retumed within 90 days (warranty or not) will be
invoiced fo butor price.
If it is the inte on of a distributor to keep a part for their spare parts stock, they should notify
the LUNAR Customer Service Department. This way, LUNAR may immediately invoice for such
parts and close service reports in a more timely manner.
.' ,
Centent, Transverse 1
Centent, Longitudinal 1
Collimator 5
Laser
,(Positioning
1 1
5 .1 3
PMT, 5 1 3 1
Tube Head '(Counts
./Air Ratio
./Spillover
,IBM Values
SBC 2 1 2
Tests 6-27
DPX-IQ Service Manual Rev. E (3/98)
ppendix 6.A
EXA Installation Work Instructions
Work Instructions [DXAP0003] ,.
~,
i appendix contains the latest work instructions for installing a DEXA sc;.an~~ease
~fully read .and f?lIow the steps described in the "PROCEDU RES" se~~ne document
m performing thiS procedure. . ~~' :Si,;> .
£T~
.,
;;VI; .
.c..n:..,~~-
< .f;.~ ~ ,.,.
TE: The revision of this document is current as, gf~ , : ' ,:.~ffifJI of release of the last
Service Manual or Service Manual Upd ~~er, this is an 'uncontrolled'
copy and is subject to change at any ti. " ~. . ,1i:rout notice, The LUNAR Service
Drs
Department will periodically notify .' ,' _'. - . of any changes and provide copies
of the latest revisions in futu~v . pdates.
The distributor is re '. ~enSUri~g that all copies of the Service Manual
are updated with documents,
~~--------------------------------.-----
------------------------------------------------------------------.
'EXA Installation Work Instructions Appendix 6.A.
WNAR~ DOCUMENT 10: DXAP0003 REV:G DATE: 11/1!0/97 CO: 4501
TITLE: DPX Series Installation Work Instructions
DOCUMENT TYPE: Work Instructions
DOCUMENT OWNER: Applications WRITTEN BY: Jeff Franz
AUTHOR/MODIFIER APPROVAL:
PURPOSE
This procedure describes the steps necessary to assemble a OEXA scanning system at
a customers site, and to assure that the system is operating within LUNAR's
specifications and in compliance with applicable Federal and State regulations.
APPLICATION
This procedure applies to all employees performing installation of a DEXA scaoning
system.
RESPONSIBILITY
LUNAR Corporation maintains a staff of individuals specifically trained to:
b. instruct operators at the cust?mers site in the safe and proper use
of the system.
ASSOCIATED MATERIALS
DXAP0001 OPX Arm Grounding Brush Installation
DXAP0012 DEXA External Options Installation
UNCON.TROllED DOCUMENT
DXAP0004 DEXA Installation Report Vtgrrfy Latest Revision
DXAP0008 DPX-Series Installation Tes~ Procedures Before Using
DXAP0009 Source/Detector Alignment
DXAP0010 DPX-IQ Arm Slot Cover Installation
DXAP0011 DPX-IQ Arm Grounding Brush Installation
DXQC0022 Aluminum Spine Phantom Scan Procedure
SYS_INST System I nsta.lJation Record
SYS HSTO System Change History
"'--' .
PROCE:DURES
1. PREPARATION
B. Begin the installation by filling out the initial information on the DPX-Series
Installation Report.
2. EQUIPMENT CHECK
A. Upon arrival at the customer's site, verify all of the equipment listed on the
packing slip has arrived. If there is evidence any equipment has been
damaged in shipment, notify the LUNAR Customer Service Department
immediately. Fill out the System Identification section of the Installation
Report. Verify all compliance labels are affixed to the system are legible,
and completely filled out:
III. system, high voltage "'power supplies (2), tube head assembly,
collimator assembly and British Type B labels on the foot end panel
of scanner.
B. Unlock the end panels and remove them by hinging them down and
sliding them toward the front of the scanner (toward the rear of the
scanner for DPX-IQ).
NOTE: Note that all panels are grounded with green/yellow wires with
protective earth labels (p/n 17231) near the connection points. For
safety reasons, it is imperative that these connections be re-
established when the panels are re-attached. Each connection should
have its own protective earth label and should only be rT,ade with a
green/yellow cable. If this is not the case, notify the Lunar Customer
Service Department.
C. Remove table top (four #1 0-32 screws from the underside, or from the
upper rim channel on the scanner frame) and back panel (four 1/4-28
screws and five #10-32 screws) from the table assembly.
D. Unlock both ends of the front door and carefully swing it down. Remove
the upper front panel by removing the two screws holding it to the front
legs and lift it upward and off. For IQ systems, remove the upper front rail
by removing the phillips head screws.
E. Remove the arm assembly from its crate. Remove the high voltage cable
trough (four screws) and the front longi~udinal carriage (three flathead
screws) from the arm. Slide the rollers·on the front longitudinal carriage
onto the front longitudinal track located on the frame assembly. Keep the
carriage as far as possible towards the head end of the table assembly.
Separate the Longitudinal Drive Cable at the rear of the scanner, being
careful not to lose the attached ground cable. You will have to loosen the
jam nut before the turnbuckle. Tape the two ends and the k)wer cable
slack to the frame so it is out of the way and will not be broken when the
arm is hung.
F. If you wish to have an extra hand hold, remove the arm's back cover
panel by removing the phillips head machine screws.
G. Clip the wire-tie that is holding the source carriage to the back of the arm.
Also, remove the two lower adjustable wheels from the rear longitudinal
carriage plate using 2-7/16" wrenches. At the head end of the table
where the cutout in the rear trough is, hold the arm upright and position
the arm's lower transverse rail through the back of the table. Lift the arm
and hang the four wheels of the rear longitudinal carriage onto the upper
longitudinal extrusion (this is easier if you tilt the front of the arm slightly
downward). Be careful not to damage longitudinal and transverse belts.
NOTE: DPX- a arms must be carefully lifted over their longitudinal carriage as
they have no column extending to the floor.
The rear longitudinal carriage must NOT be bent (see illustration, next
page). Have someone then balance the arm in an upright position while
you slide the front longitudinal carriage into position and attach it to the
end of the lower transverse extrusion.* Then re-attach the two adJustable
wheels to the rear carriage and adjust them using two 7/16 wrenches and
a 9/16 wrench.
Reconnect the longitudinal drive cable after the arm is secured. Be sure
to include ground cable in assembly. TIghten ground cable ring terminal
between turnbuckle and jam nut.
*00 not tighten screws until the transverse belt is run onto the pulleys.
3 2 1
Long Column
Short Column
Figure 1
H. Re-attach the high voltage cable trough to the front and rear longitudinal
carriages.* Then unbolt the tube head from the tube head shipping
bracket and attach the tube head to the source carriage, using the
Belleville washers at the top and the tube head tilt bar at the bottom (use
the same hardware used to attach the tube head to the shipping bracket).
Put the hardware back on the tube head the same way it ca.me off.
Remove the shipping bracket from the pan (four bolts and nuts) and
reattach it to the pan, with one bolt and nut, laying it down flat. Attach the
other 3 bolts and nuts to the mounting bracket. This is done for
transporting the Tube Head in the Mure should the scanner be moved.
* Run transverse belts onto pulleys. Tighten the 3 flathead screws which attach the
front longitudinal carriage to the arm's lower transverse extrusion. Re-check the
adjustment of the eccentric wheels.
I. Mount the OMI board on the end of the arm. Be sure it is properly
adjusted so that the wheel between the slots fully blocks thE~ infra-red
beam of the transducer. As the wheel rotates, each wheel slot should be
aligned with the centerline of the transducer window. Also Attach the Arm
Grounding Brush and the Arm Slot Cover if applicable (see DXAP0001,
DXAP0010, DXAP0011, and DXAP0012).
CAUTION
Throughout the following steps, wire ties will be used. Do not snu~J them too
tightly, as damage to the cables will result.
Do not apply power to any of the electronics until connections havte been checked
at least twice for accuracy.
Do not connect or disconnect any wires with power applied to system.
QUALITY SYSTEM DOCUMENT Pau e 5 of 12
This document contains confidential or proprietary Infonnatlon of Lunar Corp. Neither the document nor the information themn Is to b• ...produced, dlstl1buted,
used or disclosed, elther In whole or In part, except .. apeclflcaJly authorized by Lunar Corp.
IWNAll" DOCUMENT 10: DXAP0003 REV: G DATE: 11/10/97 CO: 4501
L. Cut the wire-tie that is holding the black cable track to the +26VDC power
supply.. Attach the cable track (cables will be protruding) to the rear
longitudinal carriage plate using the bolt which threads into the tab of the
carriage. This bolt will have to be removed first, so remove the nut from
the end of the bolt, and remove the bolt. Move the end of the cable track
into position and insert the bolt from the back side, through the (outer) tab
and the cable track, and into the threaded (inner) tab. Do not over
tighten; the cable track should still be able to pivot. Install the lockwasher
over the end of the bolt and fasten snugly with the nut. For the DPX-IQ,
the cable track attaches to the rear longitudinal carriage bracket using (2)
10-32 x 1/2 SHeS.
0111
Caw.
Figure 2
I. a stiff red cable which goes to the transverse Centent (top side of
the high voltage cable trough),
II. a small gray cable which goes to the transverse motion detector
from the bulkhead.
III. a large black wrapped bundle which is connected to the tube head,
and
N. Run.. t.he longitudinal belt onto its idler pulleys. Lay the red cable (see
paragraph 3.M.1. above) to the Centent (make sure red cable is routed
ab?ve longitudinal drive belt and on top of the high voltage cable trough)
and make connections. Also hook up the transverse motor ~,ires (on top
of high voltage cable trough and above longitudinal belt) to the Centent in
the proper config uration (see below).
[ CENTEN T CN014
POWER SUPPLY GND. L
Figure 3
-- - - - -
--- - . -- ~ - ---~- ...
There are three transverse motor types that can be used with DEXAI scanners.
When replacing a transverse stepper motor, refer to the following tables for
proper motor configuration.
CAUTION: If a transverse stepper motor is not wired properly, it may staN or cause
other motion errors.
5 '/
v Green ;
Wire-tie the red cable and the motor cable to the high voltage cable
trough.
P. Lay the small gray cable (see paragraph 3.M.I!. above) in place along the
inside of the high voltage cable trough and connect it to the transverse
Route the four cables (see paragraph 3.M.IV. above) along the
longitudinal carriage bracket, to the bulkhead bracket and wiretie the
cables in place to the two anchor positions on the rear longitudinal
carriage. The cable should not be strained when the armis at the head ...
and foot ends of the scanner. Connect the LEMO coaxial cable and the''''
Molex connectors lat1eled "Bulkhead 1'f and "Bulkhead 2" to the
appropriate bulkhead connector. Connect the green/yellow ground cables
to the rear carriage near the protective earth label (p/n 17231). This label
must be in place and only green/yellow wires may be attached near it.
I
r'--"'"
S. Assure that all cables and bundles lie neatly, do not bend sharply, and are
securely in place. Wireties should be placed where provision is made
(anchors mounted to frame); wireties should also be used to neatly
bundle wires. DO NOT over tighten the wireties.
T. Re-check all electrical connections made. Check for bindin!~ as the arm is
moved to the extreme comers of the table.
U. Re-attach the back panel being sure to reconnect the green/yellow ground
wire near the protective earth label (p/n 17231). Move the scan table to
its final position in the room, with six to eight inches between the wall (2 to
3 inches minimum) and the bpck of the ·arm.
.. '"
V. Unscrew the casters from the borlom of the legs and replace them with
glides. Store the casters inside the scanner in the front trough behind the
door.
w. Level the scan table by placing a torpedo level alternately on the front,
rear, and sides of the frame and adjusting the glides to mak,e the table
level. Once the table has been leveled, it must not be moved.
x. Move the arm to the foot end of the table assembly, near the access hole
in the main frame . The longitudinal drive cable passes through the front
and rear parts of the mainframe twice, creating an "upper" and a "lower"
longitudinal drive cable. Attach longitudinal ground cable to clamp via
clamp screw. You will have to remove (and reinstall) the screw to do so.
Place the upper rear longitudinal drive cable into the clamp on the
longitudinal carriage keeping the turnbuckle on the head end of the clamp.
Move drive cable relative to clamp so as to take up any slack in the
ground cable. Tighten the clamp. Place the lower front drive cable in the
clamp mounted on the front longitudinal carriage. Verify the arm is
perpendicular to the centerline of the table and that this is the natural
position (Le., no force is necessary to maintain this position). Then tighten
the front clamp, accessing it through the hole in the left side of the
mainframe's top strut.
Y. Check to see each printed circuit board is firmly mounted and properly
connected.
Z. Check the other electrical assemblies to make sure they are firmly
mounted and properly connected.
AA. Before applying power to the scanner connect one lead of an ohmmeter
to the ground lug of the scanners AG..power cord. Connect the other lead
to the detector ground screw: The measured resistance shall be less than
10 ohms. If not, check all electrical.connections and remeasure. If
continuity cannot be achieved, contact the LUNAR Service Department.
AB. Determine the regional AC line levels by setting a meter for AC voltage
measurement and at least 250 VAC full scale. Carefully measure the line
voltage at the outlet.
I. Verify the voltage specified on the fuse rating label near the power
supply isolation transformer agrees with the measured voltage.
Check the scanner wiring at the AC input. Verify the wiring at the
isolation transformer is correct (Le. the "secondary" wires are
attached to the side of the transformer with two terminals marked 0
and 115, and that the "primary" wires are attached to the
appropriate input terminals for the regional AC conditions). Verify
the white and black wires from the fuse blocks go to the correct
10caHohs on the terminal strip, and that the shorting blocks are in
their correct position for the regional AC line voltage (see drawings
for ECO 931 in section 7.C. of the Service Manual)
A. Remove all components from the boxes and instruct the customer to save
the boxes, if possible, in case a component needs to be returned for
service.
B. Position the computer, monitor, and printer on the console table. Remove
any shipping insert that may be in the A: drive.
C. Connect the monitor, keyboard, mouse, and printer to the computer. The
computer, monitor and keyboard shall be placed at least 1 rTleter from the
side of the table. Feed the supplied paper into the printer.
E. Using the cables supplied, connect the scan table to the conlputer. Make
sure that all 110 cables are fastened to their respective units (by screws or
clamps). The 110 cable on the scan table connects at the P1' connector of
the SSC.
A. Plug each of the AC power cables into the outlet strip on the underside of
the console. Make sure that the AC cable from the CPU is first plugged
into a "surge suppressor plug" before insertion into the outlet strip.
B. Plug the outlet strip into the wall outlet that was checked previously.
Verify the outlet strip is on.
C. Turn on the scan table, COrTlputer, monitor, printer and any peripherals, in
that order. Use the Setup program (Ctrl-Alt-Enter) to ensure Serial2 and
Parallel ports are correctly set to Auto. The non-cached address region
must be changed to START=O, LENGTH=1024.
A. Verify that all Scan Types on the Software Order form can be accessed.
S. Press Alt-S and observe the system number programmed into the
software. Verify it matches the system number recorded on the disk.
C. Load the LUNAR Diagnostic software. Insert the LUNAR ools Diskette.
The version number of the System Diskette and Tools Diskette must
match exactly. Type:
xxxxx is the directory where software was loaded. This is essential for
DPX-IQ diags to operate properly.
8. RECORDS
A. (U.S. Installs Only) - Completely fill out form FDA 2579 - Report of
Assembly of a Diagnostic X-Ray System. Leave the appropriate copy
with the customer and return the other three copies to LUNAR. These
forms must be submitted to the federal and state agencies within 15 days
of the install. It is imperative they are completely filled out and returned to
the Customer Service Department as quickly as is possible.
B. (U.S. Insta"s Only) - Have the customer sign the delivery receipt and the
Software License agreement.
Appendix 6.B
DEXA Split Frame Installation Proce,dure
Work Instructions [DXAP0005]
~'.
,1"
This appendix contains the latest work instructions for moving a DEXA SY~~~h openings
that would not be large enough for the scanner to go through whole. PI.~~UIIY read and
fo~low the steps described in the "PROCEDURES" section of the do~~~en performing
this procedure. ;_ ~~,,~, '"''I'
,)\~~,*
NOTE: The revision of this document is current ¥;SJ~ime of release of the last
Service Manual or Service Manual Up ~ '''~''' '_ 'FIowever, this is an 'uncontrolled'
copy and is subject to change at a .' ,,' without notice. The LUNAR Service
Department will periodically ~. ,.. f9!1'l1butors of any changes anci provide copies
of the latest revisions in/ut~~nual Updates.
PURPOSE
This procedure is for making possible to move a DEXA system through openings that
otherwise would not be large enough for the scanner to go through whole.
APPLICATION
This procedure applies to all situations where it is necessary or more convenient.
RESPONSIBILITY
All personnel installing or performing moves should be trained in this procedure.
PROCEDURES
1. Remove the two end doors.
6. Mark the position of the longitudinal away limit switch (at the foot end) and
remove.
7. Mark the positions of the longitudinal idler and drive brackets on the frame of the
scanner either with a pencil or by scratching the paint around them. This will aid
in repositioning them when the frame is reassembled. Remove the upper bolt
from the brackets and push each bracket to the side (the brackets ~'ill pivot on
the lower bolt). Secure the brackets to the frame in this position with electrical
tape so they do not swing freely when the lower half of the scanner is moved.
8. Remove the eight (8) 5/16" hex head bolts from the frame legs. There are two
bolts per leg located about half way up each leg of the frame.
9. Now the two halves of the scanner may be separated. Two people are required
to lift the head and foot ends simultaneously taking the upper half off of the lower
half. The upper section is much lig hter than the lower half (with the Electronics
Mounting Chassis). Take care not to twist or bend either half of the frame as this
may cause alignment problems.
10. Reassemble the frame in reverse order. Use the marks on the frame to
reposition the longitudinal brackets, longitudinal away limit switch and verify that
the longitudinal belt runs properly on its pulleys.
Appendix 6. C
Source-Detector Alignment Procedure
Work Instructions [DXAP0009] ,
~
This appendix contains the work instructions for aligning a DEXA scanner.~~~\arefully ~.\ ~.,.
'''iT ..
read and follow the steps described in the "PROCEDU RES" section of ~.~ent when
performing this procedure. . ~,·~""'it~~·~
.,
~" ~'"~
~(...~j~~.. ~~
NOTE: The revision of this document is current as of ' ~. :;l" of release of the last
"'"
PURPOSE
This procedure describes the necessary steps to check and align the source-detector
assembly of DEXA Densitometers.
APPLICATION
This procedure applies to all employees installing and servicing DEXA systems.
Associated document: DEXA Installation Procedures.
RESPONSIBILITY
Technicians who align DEXA systems are responsible for following these 'work
instructions. The Production, Applications, and Service Managers are responsible for
ensuring all DEXA systems are aligned per these work instructions and that personnel
performing these alignments are adequately trained.
PROCEDURES
1. Turn the scan table off and remove the tabletop by unscrewing the four 10/32 x
3/8 S.H.C.S. screwed in from the bottom of the tabletop. Remove screws and
hinge the hood cover up. Remove the screws on the left hood side panel. For
DPX-IQ, remove the lower arm shroud
2. Remove the detector collimator and lead cup (if applicable), and photomultiplier
tube. Insert an upper alignment bushing into the detector holder, keeping it in
place with the detector chuck nut or service nut (DPX-lQ).
UNCONTROLLED DOCUMENT
Verify latest Revision
Before Using
Remove the source collimator and lead wrap (shipments after May, 1997 have
the collimator already removed and packaged separately). Slide the dial
indicator onto the lower alignment rod, and attach the lower alignment rod base
to the top of the Tube Head (see drawing, Dial Indicator Setup, above). Put the
upper a.lignment rod into the upper alignment bushing so that the points of the
two rod can be easily lined up. Using the tilt bar and the upper spring loaded
7/16 nut, align the two points on alignment rods so that they are inline in the
transverse and longitudinal planes. Note: The tilt bar adjustn1ent screw is
unidirectional, i.e. tightening it tilts the Tube Head toward the front; loosening it
has no effect (see drawing, Tilt Bar Operation, at right). So if the Tube Head is
tilted too far toward the front, it must be repositioned toward the rear and brought
forward again. At this time tighten the lower two nuts that mount the Tube Head,
and the nut that is located on the tilt bar adjustment screw.
0 C
0
FOOT L
U
..- END
M
r'l N
____Tilt Bar
Adjustment Screw _~I'---"
4. At this point remove all the alignment tools. Install the lead wrap and source
collimator on the Tube Head. Install the Detector (WITH POWER OFF!!!), lead
cup and detector collimator (DPX and DPX-L). Install the film holder into the
detector collimator rather than the brass collimator insert. For DPX-IQ, install the
detector (with power off), lead cup, service nut,. and film holder. Qround the PMT
at the threaded hole on the front side of the detector holder. Note that the
protective earth label (p/n 17231) should be present wherever a green/yellow
ground wire is attached to a chassis piece. Notify Lunar Customer Service
- Department if this is not the case.
5. Turn the scanner and computer on, from the <C:\LUNAR> prompt, load the
diagnostic software by inserting the Tools Diskette into the A: drive and typing:
NOTE: The word "lunar" may be substituted for other working directories such as
"ped".
6. Note: Refer to section IX.A.2 in DXAP0003 for instructions on Tube Head warm
up before performing this step. Then insert a piece of film into the holder (large
number "57" on film should face up) and take an initial picture using the DIAGS
static counter program. For the DPX and DPX-L use 76 kV, 750 mA, 30 repeats,
and large collimator (leave all other parameters at defaults). This picture is your
starting point for final alignment. HINT: If the collimator is manually positioned
to minimize counts when sampling through the film holder, this should center the
beam and minimize time and film usage. Sample without a piece a film until the
counts are minimized, then insert a piece of film and expose for at least 30
repeats.
7. Follow the directions on the film holder for developing the film (move the lever to
'L', insert the film, move lever to 'pI, remove film, wait 20 seconds, peel apart) If,
when the film is developed, the beam spot is not in the exact center, adjust the
collimator position:
A. Determine from the images on the film which of the two collimator
jackscrews must be adjusted and the amount of adjustment necessary.
One complete turn of the jackscrew equals 4 mm of correction on the film.
8. Take one more film, positioning the tube head in the four corners of the table and
in the center, moving the film in the film holder to mimic the location on the table
and making an exposure at each of the five locations. Do a large and small
collimator shot in the middle of the film.
9. Examine the four corner pictures and the center shot. All the crosshairs should
be visible in all 5 spots and be no more than 2 mm in either direction from center.
The only criteria for the small collimator shot is that the crosshairs should be
visible at all. If these conditions are not met, it may be necessary to redo the
runout procedures (0 and E) and make adjustments to the scanner. If, after
adjustments, the pictures are not satisfactory, contact the LUNAR Customer
Service Department.
10. Label the film with the systenl number, the date, and the relative location of
home position. Coat the film using the chemical sponge supplied with the film.
Save this film and include it with the rest of the installation paperwork.
11. Remove the film holder from the detector collimator and insert the detector
collimator insert in its place. For DPX-IQ, replace the service nut with the
detector chuck nut. Re-install the lower arm shroud. For safety reasons, it is
imperative that protective earth connections be re-established when re-installing
certain components such as panels or the PMT. Each connection should have
its own protective earth label and should only be made with a green/yellow cable.
If this is not the case, notify the Lunar Customer Service Department.
12. Confirm the alignment by running an Air Matrix scan. For new installations this is
the nE~xt step in the install procedure, DXAP0003. Note: Remember that the
Detector must be peaked prior to performing an Air Matrix.
ppendix 6.D
lPX-IQ Arm Slot Cover Installation
Work Instructions [DXAP001 O] ~t.
*,
~~ ~,."
'.~,~
"
s appendix contains work instructions for installing the arm grounding ~~~ifle DPX-IQ.
PURPOSE
To document the DPX-IQ Arm Slot Cover Installation.
RESPONSIBILITY
The Applications Manager is responsible for ensuring that this work instruction is current and
that all people who perform this process are trained.
PROCEDURES
DPX-IQ Arm Slot Cover Installation
The arm slot cover must be installed for TOV G.S. compliance after the scan arm is attached.
1. Insert the lower spring guide screws (6-32x%" flathead) into the lower spring
guide (part number 1943 or 1848). The screw heads must go between the two
strips of felt on the inner sides of the guide. Slide the guide along the top edge
of the back panel until all the screw holes line up. Tighten the screws.
2. Line the upper spring guide (part number 1944 or 1847) up with the screw holes
in the back rail, slot down. Fasten the guide on with 10-32x%" screws and
lockwashers.
3. Insert a stainless steel strip spring (part Dumber 1839) into each of the end
canisters (part numbers 3560 and 3561 ) with each tab sticking out through the
side slot in each canister (see figures 1 and 2).
UNCONTROllED DOCUMENT
Verify latest Rev'rSron-
.
Before Using
,;
\
Head Canister
Head-End Cannistet
_____ PIN l560 -------.
,/ .........
o0 o0
1 2
Figure 1
Foot Canister
~
o 0 o 0
o
~ ~
o0 o0
1 2
Figure 2
Figure 4
5. The foot end strip spring should now be ready to be attached to the arm. Extend
the spring until it reaches the foot side arm spacer (part number 1856). Then
attach the spring bracket to the arm spacer with two 8-3x3/8" button head screws
with #8 lock washers (see figure 5).
i---~/
Scan
'
'' .
Arm
Figure 5
6. The Arm Slot Block (pIn 3847) is normally assembled to the Arm at Luna~ If not,
attach the Arm Slot Block to the head side of the extrusion near the column (see
figure 6). Next attach the head end strip spring bracket to the arm slot block in
the same manner as above.
Figure 6
Back Panel
:~II~C:
~Slotted holes
/ ~
Figure 7
NOTE: There must be at least 20mm of space between the column and
the canister covers when the ann is at its longitudinal limits.
7. Test the arm slot cover by running the scan arm from head to foot several times.
Verify that the arm slot spring does not bind or "ride out" of the grooves created
by the upper and lower spring guides. If necessary, raise or lower the back
panel of the scanner by loosening the screws holding the panel to the scanner
frame.
j ,
Appendix 6.E
DPX-IQ Arm Grounding Brush Installation
Work Instructions [DXAP0011] _
.~~..,.
This appendix contains work instructions for installing the arm grOunding.bpJ~~e DPX-IQ.
~'1.,"~ ~~~,
NOTE: The revision of this document is current as of the time . '. ~ se of the last
Service Manual or Service Manual Update. Hope~ ., :~
..,'tiS an 'uncontrolled'
copy and is subject to change at any time wjt~~e. The LUNAR Service
Department will periodically notify distri?~ . i'(':Why
' changes and provide copies
of the latest revisions in future Manua/~fes.
PURPOSE
This procedure describes how to install the arm grounding brushes required for CE
compliance on DPX-IQ scanners after the scan arm is attached .
APPLICATION
This document applies toward the installation of all DPX-IQ systems.
RESPONSIBILITY
The Customer Support Manager is responsible for ensuring that this work instruction is
current and that all people who perform this process are trained. Personnel who perform
this task are responsible for following these instructions .
. ASSOCIATED MATERIALS
DXAP0003 - DPX-Series Installation Work InstruC?tions
UNCONTROLLED DOCUMENT
Verify Latest Revision
Before Using
PROCE:DURES
1. Attach the grounding brushes (Lunar part number 3717) to the grounding brush
support (part number 3716) as in figure 1. Use the 6-32 3/8" screws, the #6 lock
washers, and the 6-32 lock nuts provided.
Brushes
~(p/n3717)
o Brush Support
(pIn 3716)
Figure 1
2. Locate the head end Arm Spacer (part number 1856) near the arm slot (see
figures 2 and 3). Attach the grounding brush support to the arm spacer (on the
foot side) with the button head screws and lock washers making sure that the
brushes are pointing up. Verify that both the brushes contact the bottom of the
back rail.
I I
I
Figure 2
Figure 3
3. Verify that the brushes retain contact with the rail as the arm moves fran1 limit to
limit.
GENERATED RECORDS
None
':,,'
! I
i
Appendix 6.F
DEXA External Options Installation
Work Instructions [DXAP0012] .
'\.
This appendix contains work instructions for installing the arm slot cover;~-\~-IQ.
The distributor is responsibt . ' ~ffng that all copies of the Service Manual
are updated with the curren . o~~nts.
PURPOSE
This procedure specifies the methods used to install external options for a DPX-Series
system.
RESPONSIBILITY
The Customer Support Manager is responsible for ensuring that this work instruction is
current and that all people who perform this process are trained. Personnel who
perform this task are responsible for following these instructions.
PROCEDURES
New hardware must be added to the DPX for TOV GS mark compliance. This
hardware facilitates two important external safety options:
External Options
E-Stop and Indicator
I
~
,
. ...
Neutral IN r------r--,--....,---r----,.-----,
, V A2 I A2 I A1 · A2 I A2 I A1
o LED 0 LED
u. E-Stop
F9 F10 Line Neutral • r- --Jumper
Relay Relay
11 j 14 I 12 11 I 14 I 12
Figure 1
A. The e-stop jun1per must be removed and a normally closed dry contact or
switch is wired in series with the system's e-stop circuitry. The external e-
stop must be a normally closed dry contact switch rated for 26 VDC or
greater and 1 amp or greater. Without external emergency stop, the e-
stop jurrlper must be inserted as indicated in the diagram above.
3. INSTALLATION PROCEDURE
) I
A. Fasten the 6" DIN rail (3273) to the PEM nuts located on the pan using 2
I
I
8-32 X 3/8" screws each with a #8 flat washer and lock washer. Some
older systems may not have the PEM nuts on the pan . In that case, drill
two holes in the pan 5" apart as in figure 1.
100-------25"--------.:
5·'
o
+5i+121·12VPS
[SSl
o
~
Termnal Blocks
1 Relay I
~
D Terminal Block I
AC ENTRANCE
Figure 2
B. Cut the 4 5/8" piece of grommet (1029) in half. Place a small aillount of
cyanoacrylate based glue on the inside edge of the grommets and insert
around the edge of the 3/4" diameter holes. If these holes are not
present, one or both will have to be drilled. The location of each hole can
be seen in figure 1.
C. Connect the cable assembly (3682) to the solid state x-ray relay located
on the pan. Connect the black wire terminal ring to the (+) termlinal on the
solid state relay and the red wire terminal ring to the (-) terminaL Route
the cable from the solid state relay to the external options terminal block,
wire tying the cable to existing wires along the way. Insert the black wire
into A 1 of the line relay and the red wire into A2 of the neutral relay (see
figure 2).
D. For OINK Revision E and higher, connect cable 4339 to J27. RO~lte the.
cable from the OINK board to the external options terminal block, wire
tying the cable to existing wires along the way. Insert the cable ends into
the terminal block (see attached diagram).
For OINK revision 02C, locate the gray wire coming from J17 of the OINK
board. Cut the gray wire approximately two inches from the OINK board
connector. Strip 1/4" of insulation from both ends of the gray wire. Cut
the J connector off of cable 4339 and strip 1/4" of insulation from the black
and red wires. Cut the 1 1/2" piece of heat shrink tubing (5190) in half
and slide it onto the red and black wires. Solder the red wire end to a
gray wire end, twisting the two wires together before soldering; repeat for
the black wire end and the remaining gray wire end. Place the heat shrink
over the exposed wire.
E. Place the External options label (3680) on the pan. The label should be
placed in the opening to the upper left of the external option terminal block
assembly. Leave the data boxes on the label blank.
External Options
Terminal Block Assembly
BlK RED
Black
~. ----
E-Stop
F10 ~ Line Neutral -Jumper
Relay Relay
. RED BlK
Cable 4339
From OINK Board
Figure 3
J I
I
I
NOTE: The revision of this document is current as of iliJ/'ii of release of the last
SeNice Manual or SeNice Manual Upd ~~e" this is an 'uncontrolled'
copy and is subject to change at any ti ),. ~ .)1fJtout notice. The LUNAH SeNice
Department will periodically notify . ' .. ' drs of any changes and provide copies
of the latest revisions in futup1:' fbpdates.
~'.;
PURPOSE
This procedure details the tests required to be run on a DEXA system upon its initial
installation at a customer's site.
RESPONSIBILITY
All personnel performing installations of DEXA systems shall be trained on this procedure.
PROCEDURES
These test procedures shall be performed following the assembly of the DPX-Series
scanner according to procedure DXAP0003 .
I\, Use the DIAGS static counter program to verify the kV and rnA are
being properly set. Use the following parameters:
C. Using the DIAGS software and the control panel positioning switches,
carefully move the scan arm to all areas of the table to check for binding
points. In particular: Tube head and solenoids must clear tabletop and
carriage bracket bolts. Collimator and shutter wiring should be correct and
wiretied so they do not catch on anything. Especially watch for binding while
the scan arm is moving to home position. The Molex connector of the
collimator assembly may hang up, causing the scanner to jerk when moving
off of the limit switch. Verify the collimator limit switches operate correctly.
Make sure the red, "SHUTTER OPEN" light illuminates when the shutter is
open.
~),!5
(
Tfl;lIJ.SA.6
~.,
' I
'~
•
.
-_/h'lrt:"'
1'/) I '1' 1\.....,;.....
q';) - -/," n" "~-6'VY 1
I U,- /..1 1--, .
G
.
, <'_
,
( ......
.,... , (/:.
(QJ1I')/r:>
.u71') 0/\ -'"
en'Y') 01 . .) //I
~t....1
n
..... f
_ /J ' -, /'1. t"""
/-://111 , .... ...r { ~ I , .
/
..,.~':.,:'" j'
E. Calibrations
I. Unpack the standard and verify that the serial number on the delivery
receipt and/or packing list matches the number stamped on the
standard. Record the serial number on the installation report .
VII. After the calibration has been run, exit to DOS by pressing F10 from
the Main Menu.
NOTE: The following two tests should be run immediately following the QA.
If not, the scanner should be repeaked (either by running another
QA, or by performing a Peak in DIAGS) before doing them.
II. With nothing in the beam (except perhaps the tabletop), press Esc to
open the shutter. Take about 5 repeats, then begin to slide the
wedge into the beam (narrowest edge first). While sampling, slowly
push the wedge through the beam. Take about 10 repeats in the
wedge. This will leave five repeats with nothing in the beam at the
end.
III. After the shutter closes, the collected data will be printed. Examine
the printout. Note the high channel counts. At the repeat where the
high channel counts begin to decrease, the low channel should also
decrease. The low channel should then steadily decrease for all other
samples taken through the wedge.
NOTE: The sign of a failing AGS Test is where the low channel counts
decrease, then increase brie'ny, then decrease again. See the AGS
Test in the Service Manual for more information.
G. Air Matrix
III. Press Esc to beg .in the scan. The scan will take approx. 30 minutes to
complete. NOTE: If anything passes through the beam during this
test, IT Will FAll.
IV. Analyze the Air Matrix scan and print the overall results. All
categories must pass. If NOT, check the deviation imagt~ for possible
electronic or alignment problems. If an alignment or an electrical
adjustment (or replacement) is needed after a failing Air Matrix, rerun
a calibration from the system software, or a Peak from the diagnostic
software, before retrying the Air Matrix.
V. Once an acceptable Air Matrix scan is printed, delete the Air Matrix
files (these files begin with "XM" or "XP" plus the System Number, and
they are generally the only files beginning with "X" in the C:\lUNAR
subdirectory). Return the printout of the passing Air Matrix with the
install report.
NOTE: Notify the site's Radiation Safety Officer that the sc;a nner will be
generating x-rays while unattended. Post signs (W;a rning, X-
Ray Radiation) on the door of the room and on the scanner.
II. Check the overnight results the following morning. Make sure all 10
scans ran. Air values should be within 2 0/0, air ratios should not vary
by more than 0.02, and the peak units should not range l1l0re than 25
units. If the air ratio, counts, or peaks do not fall within these specs
contact the lU NAR Customer Service Department.
II. Run the aluminum spine phantom scans according to the Procedure
DXQC0022. When setting up the phantom scan, verify the Patient
Positioning Light and the Laser On light operate properly.
III. Run 3 scans at Medium (750uA) speed. Average the results from the
3 medium scans, and verify that the mean is within 2% of the
expected value for the phantom. Record the average results of the
.medium scans and the expected result on the Install Report.
NOTE: Follow the analysis instructions included with the Aluminum Spine
Phantom. If not analyzed properly, the results may deviate from the
expected value.
J. Test Completion
".--.
I I
j
Appendix 6.H
Aluminum Spine Phantom Scan Proce(,jure
Work Instructions [OXQC0022] l '.
~,
This appendix contains the latest work i.nstructions for scanning an alumi.n~~J~Phantom.
Please carefuHy read and before scanning. _~~J.~~1-
.~,A"~·
NOTE: The revision of this document is current as oftf)# ~:f¥?elease ofth~~ last
Service Manual or Service Manual Update. H~ this is an 'uncontrolled'
copy and is subject to change at any tim ~: ~otice. The LUNAh~ Service
PURPOSE
This procedure details the process for scanning aluminum spine phantoms on DPX-Series
scanners at the time of system installation. . .
UNCONTROLLED DOCUM ENT
Verify Latest Revision
APPLICATION Before Using
This policy applies to all aluminum spine phantoms used in the field.
RESPONSIBILITY
All personnel that scan phantoms at customer installations will be trained on this procedure.
PROCEDURES
The scans will be run as part of the quality assurance procedures.
A. Verify that five (5) calibrations have been run on the scanner. These shall
have been run within the previous week. If a calibration has not been run
within the last two days run a calibration. If all of the calibrations are more
than a week old, rerun all five.
B. From the LUNAR Main Menu, verify that you are in the AP Spine scan
program. Select option <F1> for a Patient Spine Scan and then <F4> to add a
new patient.
Enter as the first name "Spine" followed by the phantom serial number, and
"Phantom" as the last name.
Weight : 70 kg
Sex : Male
*** When properly positioned. the detector will remain over the water bath
during the entire scan. ***
E. The next screen displays the scan parameters. Verify or change that the scan
parameters are set as follows:
kV : 76.0
uA : 750
Collimation : Medium
NOTE: The scan length 'can be shortened to no less than 120 mm if desired.
If left at 170 mm, the scan should be aborted after 3-5 lines into T12.
r iG. Press <ESC> and the scanner ramps up the tube head voltage and current,
and the arm moves to a location where the Patient Localizer Light is
positioned over the transverse center of the scan table.
H. Use the positioning switches to center the patient localizer light in the middle of
the L5 section of the phantom. All scan lines must be completely in water.
I. Press <ESC> and the scan will begin. Verify the scan does not go outside of
the tub (the confines of the water) and starts scanning below L4, otherwise,
abort, reposition, and restart the scan.
J. At the completion of the scan, the arm will move to the home position. Press
<ESC>. The scan process is complete and the user is returned to the Main
Menu.
B. Choose the patient name (or appropriate -identification) of the phantom scan
by using the up and down arrow keys, and press <ESC>. Select the
appropriate scan and press <ESC>.
D. Press <F2> Auto Analysis and adjust the intervertebral markers. Adjust the
markers using the histogram as the Operator Manual describes for a normal
spine scan. L2-L4 should contain 88 lines; L 1-L4 should contain 108 lines.
Verify the labels are correct on the phantom for L 1 through L4. If not, correct
them using option <F3>.
If performing repeat scans, use the Compare scans option for optimum
reproducibility.
,I I
.
Appendix 6.1
DEXA Installation Report
Work Instructions [DXAP0004]
The distributor is respon.;S/DI'a1IjOf8,U ring that all copies of the SeNiCf.~ Manual
are updated with
.. '-..-
ASSOCIATED MATERIALS
DXAP0003 - DEXA Installation Work Instructions UNCONTROLLED DCICUMENT
DXAP0008 - DEXA Installation Test Procedures Verify Latest Revision
Before Usinl~
CHECK LIST
SYSTEM NUMBER:,_ _ _ _ _ _ _ _ _ INSTALL DATE:
FACILITY: '-------------
ADDRESS:_______________________________________
CITY:,__________________,STATElCOUNTRY:_ _ _ _ _ _ _";ZIP:-,_ _ _ __
CONTACT: PHYSICIAN:
PHONE#: FAX#:
----------------------
----------------_.--------
SYSTEM IDENTIFICATION
SCAN TABLE SIN: STANDARD S/N:,___________TUBE HEAD S/N: ______
CONTROLLER MODEL I HD SIZE: SIN:
PRINTER TYPE:, _______________--~--------------------------
---------------
SOFnNAREVERSION:,___________~-~----------------------~
STANDARD BATCH: 6, 7, or 8 (Circle One) SOFTwARE CONFIGURED FOR STANDARD BATCH? YIN
SETUP
SOURCE/DETECTOR ALIGNED?
LIMIT SWITCHES VERIFIED?
VERIFY PROPER AGS OPERATION (WEDGE TESn?
RUN AIR MATRIX TEST. PASSED?
RUN 10 OVERNIGHTS. PASSED?
PMT DETECTOR HIGH VOLTAGE (SBCTP18): VDC PEAKSETTING:, _ _ _ __
RESULTS OF 3 PHANTOM SCANS: I 1_____
AVERAGE: PHANTOM S/N:_ _ __
EXPECTED VALUE: ±2% DIFFERENCE: (%)
kV I ~ VERIFICATION:
Test Point kV /-lA Limits Actual
XORB TP1 76 750 0.74 - 0.76
XORB TP 5
XORB TP 1 3000 2.97 - 3.03
XORB TP 5
XORB TP 1 4750 4.70 - 4.80
XORBTP5
XORBTP2 3.760 - 3.840
..
XORBTP6
REPORTS
__ VERIFY SYSTEM NUMBER IN SOFTWARE (ALT-S)
__ VERIFY SCAN TYPES AVAILABLE MATCH SALES ORDER
__ AIR MATRIX RESULTS
__ PRINTOUTS FROM 10 STANDARD SCANS (OVERNIGHTS)
__ PRINTOUTS FROM 3 PHANTOM SCANS (INCLUDING ANCILLARY RESULTS)
__ SIGNED (BY CUSTOMER) DELIVERY RECEIPT
__ SIGNED (BY CUSTOMER) SOFTWARE RELEASE FORM
__ SIGNED (BY CUSTOMER) DPX-SERIES TRAINING OUTLINE
__ COMPLETED FDA FORM 2579 (AND ANY REQUIRED STATE FORMS) USA 0
___ SIGNED (BY DOCTOR) Responsibility for Adm. of Rad. to Human SUb. FORM
...
COMMENTS/ISSUES
SALES/ORDER:
OUT OF BOX:
THAIN/SITE:
(print name)
I ,
I
I
Appendix 6.J
System Installation Record
[SYS_INST]
~),
This appendix contains the a system installation record. This record Shou~,.~d for each
system installed. ~ ~. (~r
~.. ~
...w6tl;, ~~).
NOTE: The revision of this document is current as oft t.iIRe~~release ofth€~ last
Service Manual or Service Manual Update.. H . ,~ . ~";~ this is an 'uncontrolled'
copy and is subject to change at any tim ~' .. ~ . ~ otice. The LUNAR Service
Department will periodically notify diS. )~.~
.'-"', . ~'bf any changes and provide copies
of the latest revisions in future M~~teS.
resPo'1sible;~rfSUring
The distributor is that all copies of the ServiCE! Manual
are updated with the ",:: nt~uments.
~----------------------------------------
INSTALLATION INFORMATION
INSTALLED
BY:
----------------------------------DATE:------------
INSPECTED
BY:
---------------------------------DATE:------------
SOFTWARE INFORMATION
LUNAR SOFTWARE VERSION: DOS VERSION: -----
SOFTWARE OPTIONS INSTALLED (check all that apply) AP Spine___
Femur Lateral Spine Total Body Forearm _ _
Orthopedic SA Peds ___
PC KWIK PowerPak Serial Number:
-------------------
Optical Drive Software (Name & Version): __________________________
Appendix 6.K
DEXA Operator Training Procedure
Work Instructions [DXAP0007] .
~i'
This appendix contains the latest version of the DPX Series Operators Tr~,i~~edure.
Please review before training. ,\..
'\.~I
NOTE: The revision of this document is current as of t!Jf! ~ ~ . .' lease of the last
Service Manual or Service Manual Update. H.~ this is an 'uncontrolled'
copy and is subject to change at any tifl1~otice. The LUNAR Service
Department will periodically notify dist~fany changes and provide copies
of the latest revisions in future Ma ~t~s.
The distributorJs resP01sib,c,f su;ng that all copies of the Service Manual
are updated wIth the ,,,,.. ' ~~ments.
~-------------------------------------------
'.
RESPONSIBILITY
All personnel performing DEXA operator training shall be trained in this procedure. Such
personnel will use the most current DEXA Operator Training Outline.
UNCONTROLLED DOCUMENT
ASSOCIATED MATERIALS Verify Latest Hevision
Before Using
SYS_HSTO - System Change History
DXAP0006 - Responsibility for Administration of Radiation to Human Subjects
OX_TROUT - DPX Series Training Outline
DEFINITIONS
The installation consists of assembling the device, verifying proper operation, and training
'- the operator(s) in a) operation of the device, b) analysis of acquired data, and c) device
specific radiation safety. Training of scan acquisition will take place with phantoms, unless
a licensed physician has authorized specific patients to be scanned during training. The
Lunar employee conducting the training will obtain a signed copy of the Responsibility for
Administration of Radiation to Human Subjects form (DXAP0009) from the physician, prior
to permitting the scanning of patients for training purposes. This form will be filed with the
systems Device History Record at Lunar. If patients are scanned during training the Lunar
employee may assist in positioning, but the customer must actually operate the scanner.
PROCEDURES
A. Determine if patients will be scanned during training. If so, obtain a si~~ned copy of
the Responsibility for Administration of Radiation to Human Subjects form before
proceeding. REMEMBER: You may help with positioning, but the actual scan must
be initiated by the user.
B. Train the operator in the use of the system following the latest version of the DPX
Series Training Outlin~ (DX_TROUT).
1. An individual who has completed Lunar Corporation training, holds the Lunar
Corporation training certificate and has a valid California Diagnostic
Radiology certificate.
or
2. An individual who has corrlpleted Lunar training, holds the Lunar training
certificate, and has a valid California Torso, Skeleton Limited Permit.
or
D. Obtain the operator's signature and the training date on the cover page of the
appropriate training outline.
, I
i
Appendix 6.L
DPX Series Operator Training Outline
Work Instructions [DXAP0013] ,
Thi~ appendix ~ntains the latest version of the DPX Series Operator Traiqjn~e.
reVIew dunng training. ~',. ,~.[.,
Please
~ ..
1!;~'.
~.
.....
~; '7~-r ~~
NOTE: The revision of this document is current as of ~ release of the last
Service Manual or Service Manual Update, HCJl~ter:· this is an 'uncontrolled'
copy and is subject to change at any ti >W'I1·I"'IAlA,I'iJ1,,.,tlf~l=l. The LUNAf{ Service
Department will periodically notify G/S1:n/iU
of the latest revisions in future MaflJ.J.~~lIOOrate~s
The distributor is reC"""'r'lC""""O"'~""~"'~C"1 'ng that all copies of the Service Manual
are updated with
I 0Ne) have reviewed the attached outline and u erstand the topics discussed, and I
verify that the training outlined is complete. In addition, I will read (or have read) the
appropriate Manual(s), including Radiation Safety.
( )
Signature Print Name Date
( )
Signature Print Name Date
( )
Signature Print Name Date
( )
Signature Print Name Date
Facility System#
UNCONTROlleD DOCUMENT
Verify Latest RevfsfOft
Before Using
A. High Voltage 1 Current Settings - 76 kV, current variable (150, 750, 3000 or
4750mA) depending on scan mode (all set in software).
B. X-Ray Unit Location - X-ray tube under tabletop. Place the computer >1
meter from the side of the scan table for operator exposure of <0.1 mremlhr
or 1.5 meter if unit has TUV as mark.
G. Pregnant Women - Even though the dosage level is low, LUNAR does not
recommend scanning pregnant women.
Check with your State Radiological Board to see what safety regulations your state has
regarding X-Ray equipment, and what the state's operator requirements are (e.g. is an RT
required to operate the scanner?).
2. SYSTEM COMPONENTS
A. Scan Table
I. Location of on/off switch and POWER ON (GREEN) light.
Always leave the power on.
VI. Keys, Front Panel Locks, Side Panel Locks. Keys MUST be
removed from panel locks and stored in a safe place. Do not
open the scanner without LUNAR authorization.
B. Computer
I I
I". Printer
a. On/off switch; turn off overnight.
i b. Paper loading.
c. Ink/Cartridge loading.
; 3. SYSTEM SOFTWARE
A. All LUNAR software is loaded on the hard drive and a backup (SYSTEM
DISKETTE) is left on-site.
D. Do not load any non-LUNAR memory resident programs onto your DPX
computer. Load any non-LUNAR non-memory resident softwarE~ onto any
available partition other than C:. If only C: is available, load othE~r software
into H:s own directory.
IV. Save all Quality Assurance printouts. Both passing and failing
QA's are stored in the database. Only passing QA's are used
for system calibration.
NOTE: Training of scan acquisition will take place with phantoms. unless a licensed physician has
authorized specific patients to be scanned during training. The LUNAR employee conducting the training will
obtain a signed copy of the Responsibility for Administration of Radiation to Human Subjects form from the
physician, prior to permitting the scanning of patients for training purposes. This form will be filed at Lunar.
The lunar A~pplications Specialist may assist the customer with patient positioning, but the actual
scan must be perfonned by the customer.
F. AP Spine Scan
V. Begin Scan
G. AP Spine Analysis
V. Auto Analysis.
a. Check computer placement of edges and baselines; "adjust if
aberrant (exclude artifacts).
b. Patient should lie on their left side with their back flat against
the rear of the lateral positioning device and buttocks against
the lower portion of the device. Legs should bend so that back
is flat against positioning device. Shoulders should not roll
forward. Head should be supported by cylindrical foam pad.
v. Begin scan.
I I
a. Watch for the iliac crest in the first few scan lines.
V. Auto Analysis.
a. Check intervertebral markers and adjust if needed.
V. Begin scan
a. The first few scan lines should include the shaft of the femur
centered in the scan path. The beam should not scan outside
of the patient's body. The scanner should follow the contour of
the femur, slightly more tissue will be acquired on the medial
side below the ischium.
e. Press [F1] to end the scan if the auto-stop does not invoke.
K. Femur Analysis
V. Auto Analysis.
a. If necessary, rotate ROI box so that it is perpendicular to the
femoral neck.
v. Position patient.
a. Parallel to long axis of table.
NOTE: WhilE~ the patient is restrained, do not leave them unattended. Make slire they are
comfortable ;and observe them frequently.
V. Verify Values.
a. Image parameters.
VII. Use Manual Analysis option to exclude any artifacts or for user
! I
defined ROJ analysis.
IX. Compare
X. Save changes.
IV. Verify Values to enter/change patient arm length, arm side and
scan mode, width and length.
V. Position patient.
a. Remove table top pad and install forearm scan platform.
NOTE: While the patient is restrained, do not leave them unattended. Make sure they are
comfortable and observe them frequently.
c. If the scan was started too far distally (includes carpel bones),
abort the scan [F1], and reposition. Continue scan at end for
additional length to include full 33%) site if auto stop appears to
have invoked too soon.
V. Auto Ana,lysis
a. Check edges, baselines and Reference Point for proper
placement.
P. Database Utilities
) I
I. Patient Results - Offers options to view, print or graph patient
scan results.
/ IV. Restore - Allows you to return archived files to the hard drive.
/
! This is NOT necessary to reanalyze scans that are on Archive
/
I
diskettes.
J
I V. Copy Files - Allows you to copy scan files from one location to
another, such as the C:\DB_DATA\ directory to a micro disk in
the A:\ drive.
VII. Delete Patient - Allows you to delete a patient's name from the
patient directory (also deletes mandatory and optional
information). You can delete a patient only if the patient has
\\
not been scanned.
\.
~,
-"'\ VIII. Backup Data - This routine is used to copy all non-archived
files to diskette. It may be used in conjunction with Archiving
for better data security.
R. Choose Scan Type - Permits the user to switch between AP Spine, Lateral
Spine, Femur, Total Body and Forearm programs.
VERIFY LATEST REVISION BEFORE USE Page 15 of 21
this document contains confidential or proprietary Information of Lunar Cofp. neither the document nor the Infonnatlon Is to be reproduced, distributed,. used or
disclosed, either In whole or In part, except as specifically authorized by Lunar Corp.
~IK DOCUMENT 10: DXAP0013 REV: D DATE: 10122197 co: 4494
S. Print Manager - Permits the user to analyze files and print them at a later
time.
II. Select All and Mark Report allows the operator to print reports
at times other than that specified under delayed print time.
Press <ESC> after marking reports to print them immediately.
III. Delete All allows the operator to erase all reports stored in the
Print Manager.
T. Select Language - Screen displays can appear and reports can be printed
out in English, French, Spanish or German.
V. Exit To DOS - Permits user to leave the DPX series software. This should be
done before turning off the computer.
4. RESULtS DEFINED
The purpose of the Orthopedic software is to provide densitometry of the proximal femur in
which a prosthetic device has been implanted in order to monitor the changes in an
individual patient. The software has operating similarities to the standard software but is an
optional external package. With some versions this package may share databases with
the DPX series software.
1. SYSTEM SOFTWARE
At the C:\ORTHOPED> prompt type ORTHOPED, and press the Enter key.
S. Quality Assurance (option F3) is not necessary and not recommended. The
ORTHOPED software package can use the calibrations from the DPX series
software. Use a batch program to copy the DPXS.DSF file from the LUNAR
software to the Orthopedic software.
NOTE: While the patient is restrained, do not leave them unattended. Make sure they are
comfortable and observe them frequently.
V. Begin scan
a. The first few scan lines should include the shaft of the femur
centered in the scan path. The beam should not scan outside
of the patient's body. The scanner should follow the contour of
the ferrlur, slightly more tissue will be acquired on the medial
side below the ischium.
b. The implant should appear only after 10-20 lines have been
scanned.
c. Abort and restart if scanning into air on the outside of patient's
hip. Select a 750mA scan mode and/or use bags of tissue
equivalent (rice) jf necessary.
d. Scan should proceed 5 lines beyond the top of the greater
trochanter and auto-stop. (continue scan if auto-stop invoked
too early).
e. Press F1 if auto-stop does not invoke.
E. Database Utilities
If databases are not being shared with the DPX series software, a separate
backup and archive is recommended. See the DPX series Database Utilities
Outline or the Orthopedic Software User Manual
F. Separate but identical Default Settings exist for the ORTHOPED software.
Customized Report Headings are selected from here, as well as Reference
Data and Reference data adjustments. Reference data must be switched to
turned "on" and the Report Headings must be re-entered any time you reload
or update the LUNAR software from diskette.
I I
DPX-Series Scan Mode Selection for AP
Spine and Femur Scans (modepicbsd - Rev. 0 - ECO
1650)
oPXIDPX-A
DEFINITIONS:
Use the MEDIUM
describes patient thickness in scan region
3000uA Mode Use the SLOW
750uA Mode
Hi-Res. - 10 Only
THIN = 12 -15cm
@LARGE= 22 - 30cm
&LARGE =
24 - 28cm
=
-NORMAL 15 - 22cm
#NORMAL = 12 - 24cm
Other factors may influence which scan mode will yield optimal results.
Please contact LUNAR Applications if you have questions.
. ~
) J
The distributor is for ensuring that all copies of the Service Manual
are updated documents.
Your Lunar instrument emits ionizing radiation and Federal and State law requires that it
only be used by or on the order of a licensed physician. The Lunar Applications Specialist
can NOT authorize the use of the device on human subjects. If you wish to scan patients
during the training session, please read the following statement and sign this document.
The undersigned licensed physician, accepts all responsibility for the administration of
radiation to human subjects with the indicated Lunar device.
Signature Date
Address
City/State/Zip
Applications Specialist
I I
Appendix 6.N
High Voltage Repacking Procedure
Work Instructions [DXSEOOOO]
NOTE: The revision of this document is current as of ~~~ of release of the last
Service Manual or Service Manual UPdatt:~r, this is an 'uncontrolled'
copy and is su?ject .to ~hange. a.tif:;n
..Y ti '. ~1#;Ut notice. The LUNAf? ~ervice.
Department will penodlcally notIfy , ors of any changes and provIde copIes
of the latest revisions in futu~; [!j , pdates.
The distributor i s ' . Mensuring that all copies of the Service Manual
are updated with documents.
~~-------------------------------------
PURPOSE
This procedure outlines how to repack the high voltage connections during a preventative
maintenance visit or when replacing the tube head, high voltage power supplies, or high
voltage cables. . .
UNCONTROLLED DOCUMENT
Verify Latest Revision
APPLICATION Before Using
This document applies toward the repacking of all DEXA high voltage connections.
RESPONSIBILITY
Personnel who replace high voltage components ore perform preventative maintenance
visits on DEXA systems are responsible for following these work instructions.
SPECIAL PRECAUTIONS
Thorough Cleaning:
It is very important that all of the High Voltage connections be thoroughly cleaned. The
reason for repacking the connections is to assure that the entire joint between mating
connectors is filled with the special dielectric Insulating Compound. Any gaps, lint or
finger oil can be a site for moisture. Water is a polarized molecule that helps break down
the electric field through a spark discharge. Trapped air can be ionized by the large
potential. This corona discharge causes the insulation of the Tube Head and cable
connectors to deteriorate.
This operation must be performed around very sensitive circuitry. Move the scan arm
down to the foot end of the scanner, turn off and unplug the scanner. Be sure to spread
out the large absorbent pads supplied with the Repacking Kit underneath the Tube Head,
to cover the power entrance filter and power supplies.
Give yourself as much slack as possible in the cable to allow the cable to be straight away
from the connector. Position the connector ring so that it appears to be straight in line with
the Tube Head.
Begin threading with your fingers. Generally the first three or four revolutions, as the
connector screws into the Tube Head, may be accomplished with your fingers. If it begins
to turn hard, immediately back the connector out of the threads and try again .
PROCEIJURES
1. Turn off the scanner main power switch on the bottom front panel at the foot end of
the table. Disconnect the scanner from the wall or outlet strip.
2. Grasp the upright column of the arm assembly and pull the scan arm down to the
foot end of the scanner.
3. Unlock and lower the front door. Unlock, drop, and slide off the foot end panel.
4. Observe how the cables are fastened to the cable mounting bracket beneath the
Tube Head : It will · b~e necessary to fasten these cables in the sarn~ way after the
repacking procedure. Use a di'agonal (iitters to cut the wire ties holding the cables
to the mounting bracket to insure that the high voltage cables are not cut or
damag ed . It is important to have these cables free during the cleaning procedure,
so that it is not necessary to bend the cables sharply.
6. Disconnect the ground wires from the High Voltage Cable connectors at the High
Voltage Power Supplies located near the head end of the scanner. Remove the
cable connector from the power supply by twisting the outer ring counter-clockwise
until loose. Once the ring is loose, pullout the connector.
7. Use the "lint-free towel supplied with the Repacking kit to remove all the old
Insulating Compound from both ends of the cable and the mating surfaces on the
power supplies and the Tube Head. Use the cotton swabs supplied with the
Repacking Kit to clean the hard to reach areas (e.g., the brass center sockets on
the Tube Head connectors, and the inside of the connectors on the Hi~~h Voltage
Power Supplies). Next clean the cable ends and the receptacles with alcohol pads.
DO NOT TOUCH THE CONNECTORS OR RECEPTACLES AFTER GLEANING.
8. Inspect all of the connector surfaces for any signs of arc tracks. Arc tracks will be
black or brown streaks or burned areas. IF ARC TRACKS ARE PRESENT, THE.
CONNECTION WILL EVENTUALLY ARC AGAIN. Superficial arc tracks may be
removed by carefully excising the discolored piece of rubber from the plug. An arc
track on a Tube Head receptacle will necessitate eventual replacement of the Tube
Head.
9. Put on new plastic gloves (do not use powder coated gloves).
" 1b., Inject 1.5 to 2.0 cc into the inside end of each High Voltage Power Supply
connector as shown in figure 1.
11. Insert the cable connector into the High Voltage Power Supply. Slowly rotate each
plug as it pushes into place. The rotation will help to evenly spread the Insulating
Compound as it extrudes up the sides of the connector. Once the insertion of the
plug is started, do not back the plug out or remove it from the connector or air will
be introduced into the connection .
12. Reconnect the grounding system between the chassis, the power supplies and the
cable connectors.
13. Squeeze approximately 2cc of insulation compound onto the tip of each tube head
connector as in figure 2. Also, fill the tube head receptacles with 3cc of insulation
compound.
14. Insert each connector into the tube head, n1ake sure the polarity is correct. Slowly
rotat~~the plug as it pushes into place. Once the insertion of the plug is started, do
not back the plug out or remove it from the connector or air will be introduced into
the connection.
15. Tighten the connector ring into the Tube Head with your fingers. It should be
poss!ble to turn the ring three full turns with the fingers before it is necessary to use
the tube head wrench. If this is not possible, the connector ring may be cross
threaded.
16. Replace the 6-32 o-ring sealing screw into the connector after the excess Insulation
Compound is sq ueezed out.
17. Use two large wire ties to tie the cables to the cable bracket.
18. Wipe off any excess Insulating Compound from around each connector.
Figure 1
Figure 2
, I
Appendix 6.0
Laser Replacement Procedure
Work Instructions [DXSE0001]
.'€;.,
This appendix contains the latest work instructions for replacing a laser PO<?i~i~~lease
carefully read .and f?lIow the steps described in the "PROCEDURES" s~OMne document
when performing thiS procedure. ...~~~.
•~~~~~ ":!'"
NOTE: The revision of this document is current as 9f ,~_ of release of the last
Service Manual or Service Manual UP~~ .:; ~~~er, this is an 'uncontrolled'
copy and is subject to change at any ~i~~/H1out notice. The LUNAR Service
Department will periodically n.otify~ .' .~... : ors. of any changes and provide copies
of the latest revisions in futu!fPilJ!lVTtJPdates.
The distributor is
are updated with "'. ~, '
Z.yb/~~nSUring
. nt documents.
that all copies of the SeNice Manual
------------:::::::::::--':' " .. :,
UNCONTROLLED DOCUMENT
Verify Latest Revision
PURPOSE Before Using
This procedure describes the method of replacing defective lasers in Lunar DEXA
scanners.
APPLICATION
This procedure applies to any employee involved in the servicing of Lunar scanners
equipped with lasers. The laser compliance procedure is referenced by this procedure
and is to be used as the means of testing the replacement laser.
RESPONSIBILITY
. Defective lasers are to be replaced by field service technicians. The entire assembly is
replaced with a new assembly.
SAFETY PRECAUTIONS
DO NOT STARE INTO THE LASER BEAM at any time during the test or replacement of a
laser. The light reflected from the table top or shutter/collimator assembly should be
sufficient to determine if the laser is correctly operating.
PROCEDURES
1. Remove the foot end arm cover.
2.' Disconnect the two cables from the laser printed circuit board.
3. Remove the four screws from the laser printed circuit board. Raise the board until
the wires to the laser underneath the board are exposed. Pull gently on these
wires to slide the laser out of the holder.
4. Slide the new laser into the holder, and fasten the laser printed circuit board into
place with the four screws.
5. Plug the connector from the front panel indicator lamp into the 'YIDEO OUT"
connection on the circuit board. Plug the cable from the upper cable bundle into
the 'Yin" connection on the circuit board.
6. Verify that the OINK board has a jumper wire across R25.
7. Test the operation of the laser according to procedure LAS_CMP1 using form
number LAS_CHK2. DO NOT LOOK DIRECTLY INTO THE BEAM.
System Tests
1. Using the systems software, start an AP Spine scan. Verify the Laser is turned
on when the "Adjust Start Position" message is displayed.
3. Position the scan arm and press <Esc> to start the scan. Verify the Laser is
turned off and the emission indicator is no longer visible.
4. Using the systems software, start an AP Spine scan. Verify the Laser is turned
on when the "Adjust Start Position" message is displayed. Press the
Emergency Stop button and verify the laser is turned off.
I J
Appendix 6.P
DEXA Label Replacement
Work Instructions [DXSE0002] _
.~",
~~
This appendix contains work instructions for replacing labels for certified cQm~~.
.. . . . , ~~~~~~
NOTE: The revIsIon of thIs document IS current as of the fJme~se of the last
Service Manual or Service Manual Update. Howe.~~~is an 'uncontrolled'
copy and is subject to change at any time wit~~e. The LUNAR Service
Department will ~rio~ica"y notify distrib~~,-\ilfny changes and provide copies
of the latest revisIOns In future Manu~~ff1s.
The distributor is responsibl . G"7tn~ that all copies of the Service Manual
are updated with the curren .o~ments.
.~
PURPOSE
This procedure is to ensure that labels for certified components are replaced on the foot
panel of a scanner at the time of the component replacement.
APPLICATION
This Procedure applies to all field personnel involved in component replacement.
ASSOCIATED MATERIALS
0441 DPX Service Manual UNCONTROLLED DOCUMENT
3939 DPX-IQ Service Manual Verify Latest Revision
. Before Using
PROCEDURE
Replace the label(s) on the foot panel with the extra label(s) provided when replacing any
of the following certified components:
I I
Appendix 6. Q
OINK Board Modification
Work Instructions [DXSE0003]
'\.
. This appendix contains work instructions for replacing an OINK board in ttw>f~~'
~" '~'~
NOTE: The revision of this document is current as of the tim , .,' se of the last
SeNice Manual or Service Manual UPdate.~ H~~1, is an 'uncontrolled'
copy and is subject to change at any time wit liiice. The LUNAR SeNice
Department will periodically notify d"is~n
. 'b , ~ , " ' y ~hanges and provide copies
of the latest revisions in future Manual ,,' .
(: ;..
The distributor is responsibl,- A 'ng that all copies of the 8eNice Manual
are updated With the cur:enfIfJlI3·1fttnts.
PURPOSE
This document addresses compatibility issues between the current OINK board and older
systems.
UNCONTROLLED DOCUMENT
APPLICA TION Verify Latest Revision
Before Usi ng
This document applies to the DEXA series of bone densitometers.
RESPONSIBILITY
Lunar authorized personnel who replace an OINK board in the field are responsible for
following these instructions.
PROCEDURES
1. LASER POSITIONER VS. LED POSITIONER
A. For Revision E OINK boards and higher, the current limiting resistor required
for the original light emitting diode patient locating lamp is no longer required
with the laser positioning light. A jumper must be installed on JB2 in order
for the laser diode to operate correctly.
CAUTION: If the jumper is installed on a system with the light emitting diode patient
locator, the diode will be destroyed.
NOTE: Revision 02C OINK boards have a red jumper over R25 that must be
removed for use with systems with the LED positioning light.
J1
+5VDC
SIGNAL Not Populate Jar..OINK
R2
Rev. E an Higher
GND 220
LRC-OMI-01A
(c) 1987 J1 0 0 0
W R2
LUNAR CORP
-(=Hilllilf
+
o o o R1
CAUTION: If a transverse stepper motor is not wired properly, it may stall or cause other
motion errors.
Wiring Configuration
--- I I
Appendix 6.R
Transverse Motor Service Replacement
Work Instructions [DXSE0025]
, $.
~,
This appendix contains work instructions for replacing transverse motor in #1~~
s, ':. ~~c 4:EJf
NOTE: The revision of this document is current as of the time - se of th€~ last
Hll'
i ,
The distributor is responsibl '~nng that all copies of the Service Manual
are updated with the curren ,' p~~~ts.
PURPOSE
To document the wiring requirements for transverse stepper motor field replacements
necessary for use with various OINK board revisions.
RESPONSIBILITY
Lunar authorized personnel who replace transverse motors are responsible for following
these instructions.
PROCEDURES
There are three transverse motor types that can be used with DEXA scanners. When
replacing a transverse stepper motor, refer to the following tables for proper rnotor
config uration.
CAUTION: If a transverse stepper motor is not wired properly, it may stall or cause
other motion errors.
Wiring Configuration
Oriental Motor AMP Motor Superior Motor
OINK - 02C Parallel Serial, add Resistor* Serial, add Resistor*
OINK - E Parallel Serial, add Resistor* Serial, add Resistor*
OINK - F Serial Serial Serial
*The resistor to be added is 47 kQ between pins 11 and 12 on the transverse Centent.
I I
Appendix 6.S
OMI Board Service Replacement
Work Instructions [DXSE0026] ~,
~i~;)
~.;,
This appendix contains work instructions for replacing an OMI board in t~a~l~:-;~
..~ 't"C~';':.....;.·~~
NOTE: The revision of this document is current as of the tim . . ~, se of th€~ last
• •. ~Io.,..
The distributor is responsibl , if" ~ffng that all copies of the ServiCf.~ Manual
are updated with the curren ..oc1iJ~nts.
~. ":;,,.
PURPOSE
To document modifications to the OMI board necessary for use with various OINK
board revisions.
UNCONTROLLED DOCUMENT
- APPLICATION Verify Latest Revision
Before Using
This document applies toward OMI boards replaced in the field.
RESPONSIBILITY
Lunar authorized personnel who replace OMI boards are responsible for following these
instructions.
PROCEDURES
The OMI board is no longer compatible with all revisions of OINK boards. R1 is not
populated for new scanners with revision E or higher OINK boards and must be
removed on both transverse and longitudinal boards when used as replacements.
Inversely, R1 nlust be populated for both boards for use with Revision ·02C OINK
boards and lower. Service replacement OMI boards will be shipped with R1 populated.
J1
+5 VDC
Not Populate tor OINK
SIGNAL Rev. E an Higher
R2
220
GND
R1 :
10k
LRC-OMI-01A ~
(C) 1987 Jl 0 0 0
LUNAR CORP
-ellill =r ~-- R2
+
o D 0
R1
Appendix 6. T
System Change History Record
[SYS_HSTO]
This appendix contains the a system change history record. This record
each system worked on. ~~, ~'$
slil9~oPied for
~~ ~"j;.
, ..~;i>,."'~.
NOTE: The revision of this document is current as of '" release of th(3 last
Service Manual or Service Manual Update. H(j~'tei:',' this is an 'uncontrolled'
copy and is subject to change at any ti·,m~~x.r,;f:HG!~~nn1~ir'l=J. The LUNAR Service
Department will periodically notify di any changes and provide copies
of the latest revisions in future Ma tes.
The distributor is re5;f)O,nShO/a~0~ie~1S~ ring that all copies of the Service Manual
are updated with
.
~ ,
'--
LUNAR® DPX-IQ Service Manual Rev. E (3/98)
Appendix 6. U
DPX Certified Components
Specification [OXRAOOOO]
This appendix contains the DPX Certified Components list.
NOTE: The revision of this document is current as of the timt;.~~se of the last
Service M~nua/.or Service Manual UP?ate ..f-!~.~e~.,;Af'4ftiJS is an 'uncontrolle?' .
copy and IS subject to change at any tIme 'l!ltfJ?!J/iiii#ce. The LUNAR ServIce
Department wiII.~rio~ically notify distrib"~;Wny changes and provide copies
of the latest revIsIons In future Manua/~tes.
~·~0·'\:'~
PURPOSE
This listing specifies all certified components in the DPX-Series of x-ray bone
densitometers. This includes the .LUNAR DPX, DPX-L DPX-ALPHA (DPX-<x,), DPX-A,
DPX-SF and DPX-IQ. It is Lunar's policy to certify all x-ray controllers, bearn limiting
devices, high voltage power supplies, and tube housing assemblies. All such devices used
the in above mentioned bone densitometers shall be listed here.
APPLICATION
All personnel installing, servicing, and generating labeling for DPX Systems (Applications,
Service, Sales, Production, and Documentation) will follow this specification. This
specification applies to all Lunar DPX Series systems.
...
RESPONSIBILITY
The Regulatory Affairs Manager is responsible 'for ensuring this list of compatible
components is kept current and that all such components are certified, as necessary, with
the FDA through the Initial Report and Report of Model Change processes.
Personnel who install and service Lunar systems are responsible for being familiar with
this list and ensuring systems installed and serviced use components on this list.
UNCONTROLLED DOCUMENT
Verify lates1 RI;,vision
Before Using
isk
el 35SX - 40MB Disk -
Model L40SX - 60MB - 20 M
Model40SX
40MB Disk - 20 MH
- 80MB Disk - 20 M
Model50z
- 30MB Disk
Model 60
- 44MB Disk 3105 3105
-70MB Disk 3106 31
odel 70 - i486 -
- i486 - 120 B Dis
- 80MB - 20 MHz
- 160MB - 20 MHz
- 80MB - 25 MHz
- 160M MHz
Model 90
- 80MB - 25
33 z
Disk 20 MHz
- 160MB Dis 20 MHz
- 80MB - 25 Hz
- 1601V1B - 25 MHz
- 160MB - 33 MHz
- 400MB-
- 230MB - 33 MHz
- 80MB - 33 MH
Model 433/NP
- 120MB - 33 MHz 1684 1684
- 170MB - 33 MHz 1823 1823
- 210MB - 33 MHz 1899 1899
- 230MB - 33 MHz 1688
- 270MB - 33 MH 1780
33 MHz 2701 2701
Model 450/M - 500MB - 50 MHz 17003 17003
Model 450/MX
- 230MB - 50 MHz 1686
- 500MB - 50 MHz 1682
Mod I 466/MX- 230MB - 66 MHz 17034'
Model 466/LE
- 270MB, 4MB - 66 MHz
- 270MB, 8MB - 66 MHz
- 528MB, 4MB - 66 MHz
- 528MB, 8MB - 66 MHz
- 365MB, 8MB - 66 MHz
- 365MB, 4MB - 66 MHz
- 1GB, 4MB - 66 MHz
Model 466/MTE, 270 MB Tower - 66 MHz
Model 486P
-100 - 100MB - 33 MHz 1318 1318
-200 - 200MB - 33 MHz 1309 1309
Model G575 - 528MB, 8MB - 75 MHz 3808 3808
DPX -IQ
X-Ray Controller (De1l 4 based, sorted by Component):
COMPONENT LUNAR MODEL#
PART #
Model 51201XM, Pentium,, 1GM, 8MB -120 MHz 3828 3828
Model 5120/XM, Pentium, 1GM, 8MB.:120 MHz · 3828 . XRC1
Model 5120/GM - 1GB - 16MB 4075 4075
Model 5120/GM - 1GB - 16MB 4075 XRC1
DPX -IQ -
X-Ray Controller ACER s based, sorted b
COMPONENT LUNAR MODEL#
PART#
Model 2021 - 2 ,16MB - 166MHz
Model 2021 .- 2GB, 16MB - 166MHz
Model 2181 .- 6.4GB, 32MB - 233MHz
Model 2341
- 8MB RAM
- 8MB RAM
Model 2341+
- 16 WIB RAM
- 16 WIB RAM
- 1.6 B, 32MB - 1 MHz (blk)
- 1.6GB, 32MB - 133 MHz blk
Model 2731
- 2GB, 16MB - 166MHz 4928 XRC1
- 2GB, 32MB - 166MHz 4928 XRC1
Model 2831 .. 1.6GB, 16MB - 133MHz 4713 2831
M el2831 ,16MB - 133MHz 4713 XRC1
DXRAOOOO \
J
, DPX, DPX-L DPX-ALPHA (DPX-a, DPX-A), DPX-SF
X-Ray Controller (18M 3 based, sorted by Part #):
COMPONENT LUNAR MODEL#
PART #
Model 70 - 60MB Disk - 16 MHz 0126 0126
Model 70 - 120MB Disk - 20 MHz 0127 ·01"27
Model 50z - 30MB Disk 0140 0140
Model 50z - 60MB Disk 0141 0141
Model 70 - 120MB Disk - 25 MHz 0157 . "0157
Model 55SX - 30MB Disk 0233 0233
Model 55SX - 60MB Disk 0234- 0234
Model 70 - 60MB Disk - 25 MHz 0235 0235
Model 70 - 60MB Disk - 20 MHz 0248 0248
Model 30 - 30MB Disk - 8 MHz 0425 0425
Model 25 - 30MB Disk - 10 MHz 0437 0437
Model 30 - 20MB Disk-10 MHz 0527 0527
Model 30 - 40MB Disk 0710 0710
Model 70 - i486 - 60MB Disk - 25 MHz 0711 0711
Model 70 - i486 - 120MB Disk - 25 MHz 0712 0712
Model 90XP - 80MB Disk - 20 MHz 0713 0713
Model 90XP - 160MB Disk - 20 MHz 0714 0714
"'---.
Model L40SX - 60MB - 20 MHz 0735 0735
~
...
Model 35SX - 40MB Disk - 20 MHz 0754 0754
Model 40SX - 40MB Disk - 20 MHz . 0755 0755
Model 40SX - 80MB Disk - 20 MHz 0756 0756
Model 57SX - 80MB Disk- 20 MHz 0757 0757
Model 57SX - 160MB Disk- 20 MHz 0758 0758
Model 70 - 80MB - 20 MHz 0759 0759
Model 70 - 160MB - 20 MHz 0760 0760
Model 70 - 80MB - 25 MHz 0761 0761
Model 70 - 160MB - 25 MHz 0762 0762
Model 55SX - 40MB Disk 0789 0789
Model 55SX - 80MB Disk 0790 0790
Model 90 - 80MB - 25 MHz 0800 0800
Model 90 - 320MB - 33 MHz 0801 0801
Model 95 - 160MB - 25 MHz 0802 0802
Model 95- 320MB - 25 MHz 0803 0803
Model 95- 320MB - 33 MHz 0804 0804
Model 95XP- 160MB - 20 MHz 0805 0805
Model 95XP - 400MB- 20 MHz 0806 0806
Model 90XP - 80MB - 25 MHz 1150 1150
Model 90XP - 160MB - 25 MHz 1151 1151
DPX -IQ
X - Ray C on t ro II er (D e 114 b ase d ,sorte dbIy Part #)
COMPONENT LUNAR MODEL#
PART #
Model 5120/GM 3828 5120GM
Model 5120/GM 3828 XRC1
lodel5120/GM -1GB-16MB 4075 5120GM
Model 5120/GM -1GB-16MB 4075 XRC1
- -
Model 575GM-528GB-16MB ·085 575GM
Model 575GM-528GB-16MB 4085 XRC1
Model 5133/GM
- 1GB, 8MB 4398 5133GM
- 1GB, 8MB 4 398 X 1
- 1GB, 16MB 4415 5.133GM
1GB, 16 B 4415 XRC
- 2GB, 16MB 133 Mhz 4579 5133G
- 2GB, 16MB - 133 Mhz 4579 XRC1
Model 5200GXM - 2GB - 32MB w/1 0/1 00 Ether Card 4634 5200GXM
Model 5200GXM - 2GB - 32MB w/1 0/1 00 Ether Card 4634 XRC1
Model 5133/GM - 2GB, 2MB 4647 5133GM
Model 5133/GM - 2GB, 32MB 4647 XRC1
Model 5200GXM - 2GB - 32MB 4724 OOGXM
..
4
DPX -IQ
X-Ray Controller (ACER 5 based, sorted by Part #):
COMPONENT LUNAR MODEL#
PART #
Model 2341+ - 16 MB RAM 4013 2341+
Model 2341+ - 16 MB RAM 4013 XRC1
Model 2341 - 8 MB RAM 4115 4115
Model 2341 - 8 MB RAM 4115 XRC1
Model 2341 + - 1.6GB, 32 MB RAM (blk) 4491 2341+
Model 2341 + - 1.6GB, 32 MB RAM (blk) 4491 XRC1
Model 2831 - 1.6GB, 16MB - 133MHz 4713 2831
Model 2831 - 1.6GB, 16MB - 133MHz 4713 XRC1
Model 2021 - 2GB, 16MB - 166MHz 4761 2021
Model 2021 - 2GB, 16MB - 166MHz 4761 XRC1
Model 2731
- 2GB, 16MB - 166MHz 4928 XRC1
- 2GB, 32MB - 166MHz 4928 XRC1
Model 2181 - 6.4GB, 32MB - 233MHz 5524 XRC1
NOTE: The revision of this document is current as of the tim~~ftet§ase of the last
Service M~nua/.or Service Manual UP?ate. .JH . - ~ ''fflis is an 'uncontrolle?' ,
copy and IS subject to change at any ftm~s.w·1f ~~~, lIJiifdfJce. The LUNAR Service
Department will periodically notify distribll?~p ' ny changes and provide copies
of the latest revisions in future Manu~fJates.
The distributor is responsi ring that all copies of the Service Manual
are updated with the Cu·~n~!U:\AI~''''1.Qn~C'
~--~----------------------------------
PURPOSE
This procedure describes the configuring and testing Iomega drives for the Expert and
DEXA imaging systems. UNCONTROLLED DOCUMENT
Verify Latest Revision
Before Using
APPLICA TION
This procedure applies to any employee involved in the final system setup or testing.
Electrical Production personnel and Quality Personnel will setup, inspect, and record
results of the final system check of LUNAR devices. The Quality Assurance Manager
will review all documents and tag systems as "Accepted" for release, if appropriate.
RESPONSIBILITY
The Production Manager is responsible for ensuring all Iomega drives are setup per this
procedure.
PROCEDURES
1. Setup Iomega Zip Drive
B. Unpack the Zip drive. Connect its cable to the parallel (printer) port. If
setting up with a printer, connect the printer to the Zip drive. Connect the
zip drive to its transformer and plug it in. Insert the supplied Zip Tools
disk into the drive.
E. Because LUNAR uses both DOS and Windows95, some cautions must be
used when starting either operating system.
II. The other scenario is exiting Windows95 to DOS. In this case, the
Zip drive will still be present as drive 0:, but the CD-ROM will not
be present. This is not critical as we do not require the CD-ROM in
DOS at all, but this too can be avoided by rebooting (or restarting)
the computer and choosing MS-DOS from the LUNAR menu.
III. Applications should train all users to reboot or restart the system to
change from DOS to Windows95, or vice-versa.
A. First, obtain a Jaz Jet PCI SCSI card. Change the jumper from the MAC
to the PC position. PCI cards go into the smaller white adapter slots in
newer computers.
B. Upgrades for older PC's will use an ISA Jaz Jet card instead of aPC1
card. This should be specified on the Sales Order. The installation is the
same except the card is inserted into the first available ISA slot. The ISA
slots are usually black in color and are longer than the PCI slots. The Jaz
Jet ISA card is set to IRQ 11 by default, and so no changes need to be
made to its switch settings.
C. Setup the Acer or Dell computer first WITHOUT the Jaz Jet card installed.
Complete the onscreen setup of Windows 95 as is necessary.
E. Insert the supplied Windows/DOS Jaz Jet PCllnstall diskette, and type
A:INSTALL [enter]. Follow the instructions for an Express Setup. ' A
C:\IOMEGA directory will be created. At the end of the setup an
AUTOEXEC.IOM and a CONFIG.IOM file will be created with lines that
need to be added to AUTOEXEC.BAT and CONFIG.SYS. Those files
should be modified to look as below:
I. AUTOEXEC.BAT
PATH =C: \; C:\WI N DOWS; C:\WI N DOWS\COM MAN D; C: \LUNAR
@rem LUNAR Corporation, 313 West Beltline Hwy, Madison, WI 53713.
(608) 274-2663
@rem File: autoexec.bat DOS version: 7.0
:win
cal! C:\LUNAR\setdpx
win
:end
echo.
C:
cd \
switches = If
[menu]
menuitem=dos,Start MS-DOS
menuitem=win,Start Windows 195
menuitem=dpx,Start Lunar DPX
menudefault=dpx,8
[common]
files = 50
buffers = 45
device=C:\windows\himem.sys Itestmem:off
device=C:\windows\emm386.exe noems
dos=high,umb
shell=C:\windows\command.com Ip
DEVICE=C:\IOMEGA\ADVASPI.SYS
DEVICE=C:\IOMEGA\SCSICFG.EXE N
DEVICE=C:\IOMEGA\SCSIDRVR.SYS
for Acer: device=c:\dosdrvrs\d91 Ov31 O.sys Id:CDIDE01
for Dell: device=C :\cdrom\nec_ide.sys Id:mscd001
[dos]
[dpx]
[win]
F. Reboot the system and return to the DOS prompt. Check that the Jaz
drive is drive D: in DOS and the CD-ROM is drive E: .
NOTE: CD-ROM drives are assigned names. In the examples above
the Dell CD-ROM is labeled MSCD001; in the Acer example it is called
CDIDE01. You may find other names for the CD-ROM on other computers.
This should not be a problem. The name stated in CONFIG.SYS must
match the name in AUTOEXEC.BAT (at the end of the MSCDEX.EXE line).
G. Reboot the system and select LUNAR DPX-IQ. Verify the Jaz drive and
CD-ROM drivers load successfully.
H. Reboot once again and start Windows 95. With the card installed and the
drive connected, Windows 95 will automatically detect the Jaz system.
You will have to insert the supplied Windows/DOS Jaz Jet PCllnstall
diskette when prompted.
I. Next, use the supplied Installation Guide for IBM-Compatible PC's that
comes with the Jaz drive for setup instructions. You will need the
supplied Windows/DOS Jaz Install diskette, and the supplied 1.0GB Jaz
Tools disk. Follow the enclosed instructions at step 12.
The Jaz drive should be seen by Windows 95 as the 0: drive (the CO-
ROM drive will be moved to E:).
4. General Note
A. The Jaz Tools disk comes with the drive and IS NOT just a blank disk! It
contains useful programs for running the Jaz drive. That is why the Jaz
Drive Assy includes an extra disk for the customer to use for their data
while the Tools disk is set aside as the Jaz-software backup.
A. Insert a scratch disk into the drive and copy files from the Iomega drive ·
(test file subdirectory) to the hard disk (use the C:\SCRATCH directory).
Verify the files have all copied and that the files s'izes are the same. Then
copy from C:\SCRATCH to the root directory of the Iomega drive (0:\).
Again, verify the files have all copied and that the files sizes are the same.
NOTE: The Tools disk comes with the drive and IS NOT just a blank disk! It
contains useful programs for running the Iomega drive. That is why the Drive
Assy includes an extra disk for the customer to use for their data while the Tools
disk is set aside as the Iomega-software backup.
. ....--....
Appendix 6.W
Limit Switch Adjustment
[LUMA0004]
,~,
This appendix contains instructions for adjusting the collimator limit switc~h~: ~~~~~t~
~~ ~~.qg.
NOTE: The revision of this document is current as of the time~ se
. .. ~~- " of the last
Service Manual or Service Manual Update. H~O eo;~/~ is an 'uncontrolled' .
copy and is subject to change at any time wJt ''''f _ ,Y:ilJHce. The LUNAR Service
Department will periodically notify distrib~~< · ~ - y changes and provide copies
of the latest revisions in future Manu~l~fes.
r.T~I~T~L~E~:7L7im~it~S~w~it~c7h~A~d~ju-s~t-m-e-n~t--------~~----~------------~------------
rD=-=O-::C:-:-U-::M:-=E=N:-:=T-=TY~P=-=E=-:~W-:-:-o-:-rk-:'-n-:st~ru-c-:-:ti:-o-n- - - - - - - - - - - - - - - - - -.".-
APPLICATION
This work instruction applies to the adjustment of any limit switch which does not have some
external means of position adjustment.
RESPONSIBILITY
The Production Manager is responsible for ensuring that personnel adjust limit switches
according to these work instructions.
UNCONTROLLED DOCUMENT
Verify latest Revision
Before Using
PROCEDURES
Ideal Adjustment Positions
The Collimator or
Collimator or shutter
shutter slide slide ~
should strike the If 0 degrees .,
actuator at 0 '-----~\~·U V
degrees. This is
the touch angle.
"--' Bend .the actuator adjustment on the limit switch as noted in these diagrams:
Plier Tips
Page 3 of 3
QUALITY SYSTEM DOCUMENT
Thtl document contains confldential or proprietary Infonnation of wnw Corp. ~ther the document nor the information thefwin II to be reproduced. distributed, used or
dlae$oaed, eHher in whote or In part. except .. apecH\cally authOfUed by Lunar Corp.
DPX-IQ Service Manual Rev. E (3/98)
PURPOSE
This procedure covers the field tear down procedures for all DPX-Series. Teardown is the
disassembly of the scanner, to prepare the scanner for transit to another site.
APPLICATION
These procedures apply to the teardown of all DPX systems listed above. All personnel
moving any DPX system shall be trained on this procedure.
RESPONSIBILITY
Personnel who tear down DPX systems for system moves are responsible for following this
procedure.
ASSOCIATED MATERIALS
DXAP0010 DPX-IQ Arm Slot Cover Installation
DXAP0003 DEXA Installation Work Instructions UNCONTROLLED OOCUMENT
DXAP0011 DPX-IQ Arm Grounding Brush Installation Verify Latest Revision
DXAP0001 DPX Arm Grounding Brush Installation Before Using
DXAP0002 DEXA External Options Installation
DXAP0012 DPX Arm Slot Cover Installation
DXAP0008 DEXA Installation Test Procedures
DXQC0022 Aluminum Spine Phantom Scan Procedure
DXAP0004 DEXA Installation Report
DXAP0005 DEXA Split Frame Installation Procedure
DXAP0009 Source Detector Alignment Procedure
PROCEDURE
1. PANEL I PMT REMOVAL
B. Remove side doors from scanner. Pull scanner out from wall if possible, be
careful not to damage customers floor. Placing pieces of cardboard under the
glides will allow you to slide the scanner on almost any surface. Remove table
top, Rear panel (if possible) and drop front access door. If the scanner is
equipped with an Ann slot Cover remove it as well (see procedure DXAP0010).
C. For 10 systems remove the upper front panel rail (3 Phillips head screws) and
then the front panel (two bolts through frame).
For non-I a systems remove the front panel (two bolts through frame)
D. For 10 systems remove the lower arm shroud (8 bolts) be careful not to
puncture the plastic film in the window of the shroud and be careful for the
panel ground in the back of the shroud.
For non-IO systems remove the three bolts that hold the bJack upper ann
access door closed and open the hatch.
E. Remove the PMT from arm, and wrap PMT in foam, place in phantom tub.
F. Stuff a small roll of foam in place of PMT so that the lead ring does not come
out or remove lead ring and place in phantom tub.
For Non-IO units lower the black top back down into position between both
hoods. Use three bolts to screw black top back in place.
NOTE: Leave out fourth bolt on backside of hood, which would be on the inside
of the arm column. All steps are the same on the DPX-a.
A. Slide arm all the way to the foot end of the scanner.
B. Disconnect cable clamps located on either carriage using 5/32" Allen wrench.
There is a small hole on the front side of the scanner frame where the Allen
wrench must be inserted to loosen the front carriage plate from the cables.
C. Remove optical interrupt bracket from the underside of front carriage using a
1/8-inch Allen wrench. Place bracket in 5 x 5 cleaJ plastic zip lock bag with
scan ner keys and place in phantom tub.
D. Clip wire ties holding the tube head harness and optical interrupt wire lead to
the H.V. cable trough.
E. Disconnect the 11 wires going to the centent located on H.V. cable trough
using a small straight blade screwdriver.
F. Unplug the three connectors and unscrew ground wire (replace screw) located
on the rear longitude carriage bracket. Wire tie a bag around these.
G. For DPX-IQ, DPX-Alpha, and DPX-SF, the cable track is attached to the rear
longitudinal carriage bracket with (2) 10-32 x 3/8 SHeS in the bottom of the
track remove these and replace screws into longitudinal carriage.
For non IQ systems: Remove the 5/16 x 3-1/2-inch bolt locknut holding the
cable track to rear longitudinal carriage using a 1/2 inch socket. Wrth the cable
track disconnected from the carriage, replace the 3-112 inch bolt and locknut on
the carriage.
H. Remove the longitudinal belt from the timing pulley and fold five times, then
wire tie. Place one loop over transverse cable pulley.
I. Remove four bolts holding H.V. cable trough in place using 5/32" Allen wrench.
There are two on the small carriage in front and two on the rear longitudinal
carriage. While in the front of the machine, loosen the three bolts that are
holding the lower extrusion to front carriage and remove the transverse bett
from the pulleys. Fold transverse bett and wire tie to the bottom of the tube
head carriage.
J. Slide arm to the head end of scanner so tube head can be removed.
r
Rear
Electronics Mounting Panel
DPX
la/
CJ ~PX_ r;t
AC
Entrance
tI pexmex-I ~
SBe !
Front
Figure 1
K. Unmount the Tube Head Mounting bracket from the storage position. Place a
piece of foam (1/2 x 2 x 5) in slot under tube head mounting bracket.
L. To mount the tube head bracket, use four 114 x 28 x 1/2 inch sockethead botts,
four 1/4" flat washers, and four 1/4 x 28 inch kep nuts. Mount in place as
shown in Figure 1.
A. Disconnect the collimator Molex connector. Remove the collimator and pack it
in foam and place in phantom tub. Replace the lead shield and collimator
mounting hardware onto the tube head.
B. Use a 7/16-inch socket with a 6-inch extension. Remove lower two tube head
mounting nuts. While holding tube head with left hand, remove the third nut
from topside of tube head. Using both hands, flip the tube head 112 a turn
straight back, then 1/4 tum so the bolts are aligned with bracket. DO NOT lose
Bellville Washers on upper mounting stud. Place foam under the tube head
being careful of all wires. Be careful not to damage the tube head stud and
collimator mounting clasps when mounting the tube head onto the shipping
bracket. Slide tube head into the bracket. Replace all three nuts. Tighten top
two nuts until snug, and then an additional 112 tum. DO NOT over tighten the
nuts or the tube head bolts could break. Tighten bottom nut so the washers on
the backside just come together. DO NOT compress washers. (See figure 2.)
TUBE HEAD
Figure 2
4. CABLES
A. Connect the end of the cable track to the 26-volt power supply using a wire tie.
Push the Lema plugs from the plastic bag between the 26-volt power supply
and the frame. Separate wires on the transverse motor control cable to avoid
electrical damage during installation.
QUALITY SYSTEM DOCUMENT Page 4 of6
This document contains con1IdentJa/ or proprletaty Infonnatlon of Lunar Corp. Netther the document nor the Information therefn Is to be reproduced. distributed, used or
dlsclond. either In whole or In part, except .. specffic&l1y authoriz1td by Lunar Corp.
REV: A DATE: 12/9/97 co: 4580
B. Disconnect I/O cable from controller. Retrieve the end of the I/O cable through
the exit hole in the scanner. Coil the remainder of the cable and wire tie to the
top of the HVPS.
C. Retrieve power cable and either remove from systems where it attaches at
bottom, or pull cable inside of scanner, coil neatly, wire tie and place in front
access door. ~
5. REMOVAL OF ARM
A. ~ Wire tie the transverse belt to the tube head carriage. Wire tie the optical
interrupt wire and remaining centent cable to the longitudinal carriage.
C. Remove the bottom two wheels (eccentric V-wheels) on rear carriage, using
7/16" socket. Be sure to remove the washers with the wheels on the back side
of the carriage as well, they have a tendency to stick in place, but will fall off if
they are not removed.
D. Remove the arm grounding brush mounted on the arm as shown in DXAP0011
for 10 systems or DXAP0001 for non 10 systems. Place grounding brush in
plastic bag with other parts. The brush does not have to be disassembled. A
Chapman set works well for removing the two Phillips head screws.
E. Tape Longitudinal drive cables to frame neat the cut out at the head end of the
scanner so that they are out of the way for arm removal.
F. Remove the three bolts from lower extrusion that were loosened in step 2. I.
Holding the arm horizontally, move to the rear side of ann and support it from
there. Hold right arm under the two hoods and place left hand in the arm
column with a good grip.
E. Let the front of the arm tilt forward about 3 inches and then lift straight up and
out and down to cJear the back rail. After clear of the back rail and the scanner,
place the arm on its back on the floor, the extrusion should be pointing up. Use
a wood support for the column base on the DPX-A's. Be careful, as the tube
head carriage will move about freely on the arm. If it is not pushed to the back
of the arm it will free fall down the extrusion. Wire tie the tube head carriage to
the rear longitudinal carriage.
F. Re-assemble arm.
I. Rerrlove the carriage from front rail by pulling the carriage towaJd the
front of the scanner and the bottom inward.
II. Put the front carriage onto the lower arm extrusion using the three
bolts, do not over tighten. If too tight, they will lock into the
aluminum rail. If any shims (if present) come free from the extrusion,
replace between extrusion and the carriage.
III. Replace the H.V. trough the same way as it was removed.
IV. Replace the two wheels and washers that were removed earlier.
6. REPLACE CASTERS
A. Jack up or lift and support the scanner by the frame and remove glides.
Replace glides with casters. Place glides in plastic bag with other parts.
7. REPLACE PANELS
A. Make sure the lip on back panel is outward and replace back panel.
8. PACK CONTROLLER
Chapter 7 Maintenance
DPX-IQ Service Manual Rev. E (3/98)
Chapter 7
Maintenance
This chapter contains brief discussions of the Preventative Maintenance procedure and Check
Sheet as well as a discussion on unusual sounds and the wire list and a listing of change orders
associated with the DPX-IQ scanner.
Maintenance Chapter 7
DPX-IQ Service Manual Rev. E (3/98)
There are a number of items to be checked, cleaned and tested during a PM visit to a site. See
the DPX-IQ PM / Service Call Checklist in Appendix A.
QA History
An important aspect of the PM is checking the QA History (see section 8.14). This is important
because when a QA is run during the PM, it should be consistent with the past QA's. Copy the
DPXS.DBF file to a diskette and send to the LUNAR Customer Service Department.
Wear
The scanner should also be checked for signs of wear. The Shutter/ Collimator Assembly
especially should be inspected. Wear on these parts can cause lead shavings to fall on the filter
in the x-ray beam. This can cause a drastic decline in Air Counts (see section 8.5), causing
QA's to fail.
Also check all the scanner's moving parts for signs of wear. This would include wheels, pulleys,
gears, bearings, belts, cables and,rnptor assemblies. INSPECT CAREFULLY cables that enter
the plastic cable track. These cable~ <are subjected to a good deal of motion and thus have the
potential for breakage! Also,' when performing the various tests listed below, listen for any
unusual noises that could be indications of wear (see section 7.2).
Cleaning
Cleaning thescanne(is important. Remove dust and any debris from the electronics mounting
chassis. Clean the transverse and longitudinal extrusions. Small bits of debris from the wheels
may accumulate and impede motion. Clean the Shutter/Collimator Assembly and the filter, if
necessary.
Inspection
Inspect the scanner internally and externally for obvious damage and loose connections. Check
all PCB's for tight mounting screws.
Tests
Run all other tests according to the DPX Series PM check sheet.
If the high pitched sound only occurs during motor operation, check the optical interrupt disks
and verify that they are not rubbing against the black plastic transducer of the OMI board. The
disk should turn in the middle of the transducer's slot.
For other sounds of an apparent mechanical nature, check all appropriate gears and pulleys.
Verify that the set screws are tightened and that the gears and pulleys are not out of position.
With the scanner turned off, move the affected parts by hand to determine the exact source of
the sound. Also look for rusted or worn bearings and examine the V-wheels forwear. With age,
the longitudinal and transverse drive belts will dry out and start to cause noises. Treating these
belts with belt dressing is a short term solution until they can be replaced.
Clicking or clunking sounds made during motor operation can be caused by belts that are riding
up on their pulleys, or by a too tight bearing support. The bearing support is held to the carriage
by a set screw from the rear. If this set screw is so tight that the bearing and support have no
play (i.e. can be rocked 1-2 mm), it can cause a clunking sound during transverse motion. If this
screw is loosened, it should be secured with Lock·:Tite so it cannot work its way loose.
Pulleys that have become defective can also cause clicking and squeaking/squealing sounds.
Try to isolate the movement of individual pulleys to locate the defective part.
Scraping sounds can be caused by any of the moving carriages, cables or cable track rubbing
against an adjacent -panek ·Localize the sound and investigate the appropriate areas .
. . ..> . •
CO 3536
Field Modification NO
This CO changes the length of the return flange on the back of the column and decreases the
width of the column cover plate to match it.
CO 3557
,:',:,
Implemented '
,August 15,1996
This ECO changes the revision of Spellman high voltage power supplies from N to N 1. This
change will increase the life of the power supplies.
CO 3671
Implemented 2/17/97
Field Modification NO
This CO changes the length of the leg posts on the scanner and the routing of the power cable.
This will lower the scanner table height at installation.
CO 3690
Implemented 1/15/97
Field Modification NO
CO 3805
Field Modification NO
. This CO adds two bolts to the corner of the tube head cover to add strength.
CO 3937
Implemented
Field Modification NO
This CO changes the mounting hardware for the front transverse limit switch to allow a greater
range of movement.
CO 3978
Implemented 2/14/97
Field Modification NO
This CO changes the tissue equivalent material from Delrin to Acetron to improve durability.
CO 4005
Implemented 3/11/97
Field Modification NO
CO 4054
Implemented 5/21/97
Field Modification NO
CO 4111
Implemented 5/19/97
Field Modification NO
This CO adds new Dell P166 computers to be standard controllers for white IQ systems.
CO 4097
DPXIQ180 (White)
Implemented 5/6/97
Field Modification NO
This CO adds new Dell P133 GSM computers to replace Dell P133 GM computers.
CO 4171 ....
Serial Number(s)
Implemented 8/14/97
Field Modification NO
This CO creates part kit for field replacement of the transverse shaft. Simplify field replacement
of transverse shaft and associated parts.
CO 4232
Serial t':Jumber(s)
,.
DPXIQ240 (Black) 5555+
DPXIQ180 (Black)
Implemented 6/20/97
I Field Modification I NO
This CO changes the standard controller for black systems from Acer to Citus with internal
Iomega Zip drives.
. ...-.....
/ . ...--...,
Appendix 7.A
DPX Series Preventative Maintenance
Quality Record [DXSE0012]
This appendix contains the DPX-IQ Preventative Maintenance Check Sheet.
NOTE: The revision of this document is current as of the time of release of the last
Service Manual or Service Manual Update. However, this is an 'uncontrolled'
copy and is subject to change at any time without notice. The LUNAR Service
Department will periodically notify distributors of any changes and provide copies
of the latest revisions in future Manual Updates.
The distributor is responsiblefor ensuring that all copies of the Service Manual
are updated with the current documents.
. UNCONT~OLLED ~DOCUME~T
, Veri~lat?si: ~~vislon ...
.',.' ·., c~h:~;~~2I~::Y~!~~ ::_~:;~.~·_: ::
CHECKLIST , .
BY: LOCATION:
THINGS TO DO:
o Review the QA history. Verify that the Detector has not lost significant resolution
by checking that the high and low energy air counts have not decreased by more
than 10 % • Verify that the air ratio has not changed by more than .02. Copy the
QA database file (C:\LUNAR\ANC\DPXS.DBF) to disk and send it to Lunar.
y
0 Review the error log. Contact Lunar Service Department if any hardware errors
are discovered.
0 Delete contents of C:\SCRATCH and C:\PRNTMNGR directories. Delete *.CHK
files in C:\ directory.
0 Run SCANDISK and DEFRAG.
0 Run CHKDSK and verify the "total bytes memory" = 655,360.
0 Scan the hard drive for viruses.
0 Verify operation of the computer, keyboard, monitor, and printer (and any
optional peripherals).
0 Clean scanner interior, inspect mechanical parts and cables for WHar and
breakage, and verify all electrical connections.
0 Repack High Voltage Connections.
0 Peak Detector; Adjust if necessary to 500 units ±25.
PMT High Voltage @ SBC TP18=_ _ __
0 Run the AGS wedge test and verify that it passes.
0 Verify Limit Switches; Be certain Home Position is consistent with the tabletop
silk-screen.
0 Run a Daily QA and verify that it passes. Make sure the four indicator lamps all
are functioning.
QUALITY SYSTEM DOCUMENT Page 1 of 2
VERIFY LATEST REVISION BEFORE USE
DXSE0012 DATE: 2/18/97 CO: 3972
"" .
\ '
Appendix 7.B
DPX-IQ Wire List
This appendix contains the wires of a DPX-IQ system.
NOTE: The revision of this document is current as of the time of release of the last
Service Manual or Service Manual Update. However, this is an 'uncontrolled'
copy and is subject to change at any time without notice. The LUNAR Service
Department will periodically notify distributors of any changes and provide copies
of the latest revisions in future Manual Updates.
The distributor is responsible for ensuring that all copies of the Service Manual
are updated with the current documents.
, II
\
,\( 1 White OINK J8 Pin 2 1094
""
,
';' " . "-
)( \\ _:;,' ';" 3 Brown Power On Light Pin 2 1095
" ,
5 Black OINK J8 Pin 4 1094
'"
7 Yellow Emergency Stop Button 1095
. ~~D
rc:;J \ .''.
,\Y<> \. -('\~ . 3 Green SBC J5 Pin 3 2697
~rl
.~ " I
~
~;~
12 Red OINK J12 Pin 4 1094
3 Black XORBJ8Pin 3
I~~
""j 3 White Longitudinal Centent Terminal
\
4 Orange Longitudinal Centent Terminal
"
,
1'
OlNKJ8 ~ , Orange Bulk Head #1 Pin 9 1094
,"
'2 White Bulk Head #1 Pin 1
t78-6 W re List
DPX-IQ Service Manual Rev. E (3/98)
2 Black Bell
..
5 White .. : Bulk Head #2 Pin 6
SBC J1 Unused
-
Board and Pin # Color Goes To Part #
Connector
SBe J6 Ribbon Cable OlNKJ2 2696
" - , '"
sse J7 Red LEMO OINK J19 2489
4 Unused
5 Unused
,.", n
il} ~ v \ ~-*U
Board and
Connector
Pin # Color Goes To ,:>
\IL __-
t~r1
- .~ e~ ~
:j Part #
'--- .
Wire List
78-12
OPX-IQ Service Manual Rev. E (3/98)
Chapter 8
Troubleshooting
This chapter contains troubleshooting techniques for diagnostic failures, failing quality
assurance tests, mechanical failures and imaging problems.
Troubleshooting Chapter 8
DPX-IQ Service Manual Rev. E (3/98)
Chapter 8 Troubleshooting
DPX-IQ Service Manual Rev. E (3/98)
If the transverse motion stops during. the scan, see section 8.7. In Total Body scanning, a
defect in the transverse mechanics may cause one of the transverse limit switches to be closed.
Closure of a limit switch prevents further operation of the motor. The shutter will close, the
shutter open lamp will go out and the end of exposure alarm will sound. Seconds later the error
message will appear on the screen. For further information, see section 8.7.
The transverse and longitudinal motion detection system is only operational during patient
scans. If the problem occurs during a quality assurance scan, for instance, then it is probable
that the high voltagesystem is arcing.
If the error occurs consistently on the first line of a patient scan, and the scanner is moving in
the transverse direction, then check the following:
Remove the table top and the front panel. Verify that the slotted disk at the front of the lower
arm rail is in the middle of the slot between the photo diode and photo transistor. The slotted
disk must be completely flat and remain in the center of the sensor slot during its entire rotation.
If the slotted disk has been in contact with the optical sensor, disassemble this mechanism, and
clean the sensor and the slots of the disk.
The pulses that normally enter the OINK board at J14 (the black yvire at the center of the
connector) may have stopped. These pulses are necessary to keep the OINK board from
sending an interrupt to the SSC. These pulses can be seen with an oscilloscope (see figure 1).
Use a voltmeter to measure the voltage on the positive side of C16 (33 ~f capacitor). This
voltage should never approach 2 Volts DC during a scan if the pulses are present to reset
U11A. If this voltage stays well below 2VDC during the scan, but the error 15 still occurs, the
interrupt was invalid. Check for arcing in the high voltage system or replace the OINK and SSC.
~Stopped: 5 Acquisitions
[-J
+
T
t
.. +i ... .
T
T
-t . ...
T
I
. .L .
m-+-H-+-f-t-H-t-I-+-++-<-I-+-I-+-t-T-I-+++-I-+++-++H-++-~~-l-'-;-r~
"
"
. .
··f ' · . . ... .
T
. . . . . . .. . ..
!:
+
.. . ...L .. . .. ... .
±
• . • • • . . . -1- . . • • • • .•
.!.
~Stopped: 1 Acquisitions
£J T-i
~
t·
1
.i.
. . . .. ..L .
t ... .....
If the longitudinal motion stops in the middle of the scan, see "Longitudinal Motion Failure" in
section 8.7.
This supply is turned on by the X-ray Relay, so verify that the Relay is closing. If not, then either
the Relay is bad or it is not receiving the signal from the SBC via the OINK (see section 4.5) .
The Tube Head Thermostat is wired in series with the Relay, so if it has opened, the Relay will
not be able to close . (see Failure #30 below)
The 28VDC should also be measured at the Terminal Block. If not, check the continuity of the
wiring and refasten all connections. Also, check the wire tie-downs for excess tension they may
be putting on the wires .
It may be necessary to check the wiring from the Terminal Block to the MAX board and to the
High Voltage Power Supplies .
If any of the QA test results fail, none of the results are considered valid. The results will be
stored in the Quality Assurance History file (typically C:\LUNAR\ANC\OPXS.DBF), but these
values will not be averaged with the other results for calibration purposes. In addition, after two
days the operator will find messages appearing on the screen warning that no valid Quality
Assurance Test has been run within the last 48 hours. After two weeks without a valid Quality
Assurance Test, the software will prevent patient scans. Recent valid QA's are necessary for
accurate results.
{ill£~QJeL(}P1f~IIQJn
Just before the QA begins, the operator is asked if the Shutter Open, X-Ray On, and Power-On
lights are illuminated. If the Green Power-On light has burned out and yet the other two lights
are lit, it is recommended to ignore the Green light and answer "yes", allowing the QA to pass.
Answering "no" will fail the QA regardless of all the other tests passing, and failed QA's are not
stored in the database.
However, it is not recommended that scanner operation be continued if the X-ray On light fails,
and the scanner will not operate if the Shutter Open light fails. These LED's are the only
indicator of exposure to x-rays and must be replaced as soon as possible.
~ ea ~. ~lRiIP
1:;a~to~;rs will note slight fluctuations in the Peak Setting, but only a shift of >50 units in
consecutive QA's will give a "fail" evaluation. Drift in the Peak setting should be monitored by
the customer.
t._ iimltQrtl1
The background test will fail if other sources of radiation are located in the room or nearby on
the other side of walls.
Verify that the lead cup is installed immediately underneath the PMT and that other sources of
radiation are not causing the failure. Any other sources of radiation in the vicinity of the scanner
which causes failure of this background measurement must be removed for proper operation of
the scanner.
eam op on test verifies the ability of the lead shutter to atten'uate x-rays from the
tube head by verifying that the Background counts when the shutter is closed are lower than
counts taken through the brass piece with the shutter open.
I~ e in the Spillover percentage over time is an indication that the PMT is losing
resolution, even if the Spillover test does not fail. Trends in excess of those specified on in
section 8.15 indicate service is required. A large change in the Spillover can be explained in
some cases by standard positioning (or a small change in the home position of the scanner). If
the beam is not fully eclipsed by the brass piece during the Spillover test, the Spillover value
increases dramatically (see section 8.3)
IS is a stat stical test to measure the spread of the Air Count data. Under normal conditions,
the number of counts obtained in a large number of samples should form a Gaussian
distribution. The Chi Square test is a measure of how the data fits to a Gaussian distribution,
based on a certain confidence factor. There is a small statistical probability that this test will fail
even though there is no problem. Rare failure of this test parameter should not cause concern .
. .J. ,. [0
~ '~-Tne 'os'portant numbers are the High and Low channel Air Counts and the Air Ratio.
J ,.
During this test the baseline is established which will later be used for comparing with the values
obtained during the scan of the standard. The High and Low Air Count numbers should be
checked against the Quality Assurance history to verify that the numbers are close to numbers
obtained in the past. Care should be taken that t~e x-ray beam is not missing the standard, nor
being obstructed by the brass piece on the standard during the High and Low Air Count test.
The Air Ratio value should remain very constant over time. Check to see that the Air Ratio has
not varied from the value at the time ofinstallation by +0.02 or greater. So if the Air Ratio was
0.59 at installation, then 0.58 and 0.60 are also acceptable, but 0.61 or 0.57 would be a cause
for concern. .
Version 4.2a and higher software revisions have an additional test for air counts called "True Air
Ratio". ThisJestmoves the beam off the standard block and into unobstructed air, collects data,
and calculates the count ratio. The normal air count ratio is sensitive to rollunder but not
. rollover. The True Air Ratio is sensitive to rollover but not rollunder. The combination of the two
should detect any problems with the AGS system.
Unfortunately, Air Ratio changes are usually not noticed until the test results fail, and sometimes
the customer will not even notice the problem until they are warned that no valid Quality
Assurance Test has been run within the last 48 hours. Therefore, it is important that the Quality
Assurance trends be inspected during each service visit (see section 8.15).
It is not possible to view the Air Ratio trends in the Quality Assurance History file, but it is
possible to view the Low Channel Air .Counts. A change in the Low Channel Air Counts will
usually be accompanied by a change in Air Ratio (see "Decreasing Air Counts" in section 8.5).
If the graph of the Low Channel Air Counts shows a trend, view the Quality Assurance History,
and calculate the initial and final Air Ratio by dividing HIGH by LOW. Compare these two
values to determine if the Air Ratio has changed 0.02 or more.
... -:
These are very difficult to diagnose by a method other than substitution of new components until
the Air Ratio returns to normal. The Scintillation Detector deterioration or unstable High Voltage
Power Supply, will usually cause the Air Matrix Scan (see section 5.1) to fail. The specification
for XORB Board transorbs is that they must allow less than 0.5 micro amperes reverse bias
current. At LUNAR, each transorb is measured by applying a 5 Volt reverse bias to the transorb
and a 100 kQ 1% resistor connected in series. The voltage measured across the 100 kQ
resistor must then be less than 50 mV.
Homing Problem
The scanner may not be finding the correct "Home" position. The correct "Home" position
aligns the center of the x-ray beam with the center of the Brass Piece when the Standard is
correctly positioned. The Air Counts must be obtained through the plastic part of the Standard
next to the Brass Piece. If the Air Counts are obtained with the x-ray beam passing through the
Brass Piece or missing the standard entirely, the Quality Assurance Scan will fail. For additional
information on these symptoms, see "Standard Positioning Problems", section 8.3.
~ :\ - , .~... U Ilt1lt~§
Failure of this test is caused by one ormore of the follo~ing:
• Incorrect Limit Switch positioning
• Defective .Limit Switch
• A mechanical constraint
• A defective Motor
• A defective Centent Motor Controller
The values recorded for these tests should remain fairly constant over time. Variations between
QA's of under 25 steps should not be a cause for concern as 1 transverse step = 0.05 mm and
1 longitudinal step = 0.1 mm, so the actual variation is only a few millimeters.
If the number of steps continually increases from QA to QA, this could indicate an impediment to
the scanner's motion and should be rectified.
Another cause of falling %CV can be arcing during the QA. This can be detected by examining
the QA History graphically and looking for variation of the Large BM values (see Figures 1 and 2
below). Note the few data points that vary. These are the early signs of arcing.
OBS·L·MASS
f" ! "
--~
--1i
...!
0
--i
-1i
This will also be apparent (but not obvious) on the QA Results printout. The arc occurred in the
third standard scan line and elevated the BM values. If a customer reports a failing QA because
of a %CV being too high, be aware that this could be an early warning of arcing. Obtain from
the site the DPXS.DBF file for analysis. Look for variation of the Large BM values. Arcs of
greater magnitude can also affect more than one line of the standard scan.
It is very important to notice these early warning signs of arcing so that the system can be re-
greased before any damage is done to the high voltage cable connectors or the tube head.
Limits
(High keV) 0 50
Chi Square 0 40
DETECTOR
Luzes Passou
Ajuste de Pico 695 unidades Pas sou
Fundo(keV Baixo) 3 cps
(keV Alto) 3 cps Passou
AC9FO de paragem do feixe Pas sou
Percentagem excedente 8.67 % Passou-Bom
Qui Quadrado 1 Passou
Contagem do ar(keV Baixo) 402266 cps Falhou
(keV Alto) 245468 cps Falhou
Taxa de ar 0.61 Pas sou
Mecanica transversal 10822/10824 passos Pas sou
Mecanica longitudinal 13293/13294 passos Pas sou
Valor do Tecido 1.315 Passou
Taxa de Colima9Fo 4.093 Passou
1 2 3 4 5 MEDIA SD %CV
Revised 4/99B
Part Number: 0222
The aluminum spine phantom is used during service site visits; it may also be used for
troubleshooting. This phantom is supplied with a plastic water container. Read
through these instructions before you perform a phantom measurement for the first
time.
The phantom represents the typical density range and size of a normal human spine.
The spine phantom (Figure 1) approximates vertebrae L 1 through L4. Parts of
vertebrae T 12 and L5 are landmarks used by the program during analysis.
T12
L1
L2
L3
L4
L5
2. Secure the lid to the top of the container and place the water bath on the scanner table.
4. Put the aluminum phantom in the middle of the plastic container bottom. Position the
phantom so TI2 is toward the head of the scanner table.
I"
5. measurement. Have you l1'lP~~H'I"Pr1
First Name:
Sex: Male
Ethnic \Vhite
2. Localizer
NOTE: It would be prudent in both of these cases to take a positive and negative power
supply, AND a tube head. Also, the XORB test points may look normal
(indicating 76 KV) in each case as well. This usually occurs where the power
supplies are defective and so the monitor voltages returned to the XORB may not
show the true voltage. When the tube head a;cs causing the voltage to go too
high, then the XORB test points may indeed show the actual voltages.
The clearest indicators of these problems are the shape of the Peak, the Air Counts, and Air
Ratio. Note how these differ from cases where the standard block is positioned incorrectly (see
section 8.3).
Distribution of Points
Observed Expected
Class Interval Frequency Frequency
Distribution of Points
Observed Expected
Class Interval Frequency Frequency
In the first example of low kV (see Figure 4), defective power supplies are most likely
responsible.
I DETECTOR
I
C I AMPLIFIER COUNTS COUNTS
0 /t-\ , SETIING LOWkeV HIGH keV
I I \ /
U I I \ I
N ".- .... _/1 I \ / 250 29555 390
I \ /
T
I 300 39718 554
/'''1', \ /1 350 40026 1312
S / I '\ 400 36234 6506
/ I \ 450 48899 22634
/ I \, 500 67661 38630
I
I " "- , 550
600
68608
49011
41050
34429
650 26595 42352
700 11142 60496
AMPLIFIER SETIING 750 4269 70912
1 2 3 4 5 MEAN
LARGE BM260.9 264.0 262.4 259.5260.1 261.41.640.63
WIDTH 503 507 507 498 503 504 3.320.66
MEDIUM BM192.7 194.6 193.7194.6194.3 194.00.730.37
WIDTH 437 437 436 433 434 435 1.620.37
SMALL BM137 . 7 13 8 . 5 13 9 . 6 13 7 . 5 138 . 6 138. 4 0 . 73 0 . 53
WIDTH 370 370 374 369 369 370 1.850.50
In the case of high voltage (Figure 5), an arcing tube head is the most likely cause although
defective power supplies could cause this as well.
I DETECTOR
I
/'l AMPLIFIER COUNTS COUNTS
C ,
0 I
""'-,
'J/ I '\ / ..j. \
I
/
SETI"ING LOWkeV HIGH keV
U / , \ I
I I I
I ,.,.. 250 87594 1747
N I \
I 300 97331 2522
T I I \ 350 78506 9101
/ \
S / I , 400 54877 34186
I I " 450 59094 75590
... / I " 500 76557 96291
I 550 78016 80230
I 600 55782 53878
650 31110 53238
700 13338 72266
AMPLIFIER SETTING 750 4963 80477
Lights Pass
Peak Setting 500.units Pass
Background (Low keV) 2 cps Pass
(High keV) 1 cps Pass
Beam Stop Action Pass
Percent Spillover 7.12% Pass
Chi Square 2 Pass
Air Counts (Low keV) 816197 cps Pass
(High keV) 692936 cps Fail
Air Ratio 0.85 Fail
Transverse Mechanics 12398/12400 steps Pass
Longitudinal Mechanics 19752 /19754 13ar.B~S
Tissue Value 1.309 Pass
Collimation Ratio 3.806 Pass
1 2 3 4 5 MEAN
LARGE BM2 61 . 0 261. 6 261. 9 263 . 6262 . 5 262 .1 0 . 870.33
WIDTH 493 495 495 495 494 494 0.800.16
MED IUM BM1 9 6 . 1 1 9 5 . 4 1 9 6 . 9 1 95 . 0 1 95 . 2 195. 7 0 . 71 0 . 3 6
WIDTH 425 423 427 425 424 425 1.330.31
SMALL BM140.7 139.4 139.5 140.0139.3 139.80.530.38
WIDTH 362 358 359 360 358 359 1.500.42
Example 1
Figure 1is a normal passing QA with the standard block properly placed.
DETECTOR
Lights Pass
Peak Setting 510 units Pass
Background (Low keV) 1 cps Pass
(High keV) 1 cps Pass
Beam Stop Ac~ion Pass
Percent Spillover 7.59% Pass
Chi Square 6 Pass
Air Counts (Low keV) 789683 cps Pass
(High keV) 457478 cps Pass
Air Ratio 0.58 Pass
Transverse Mechanics 12398/12400 steps Pass
Longitudinal Mechanics 19755 /19755 ~ar.a~s
Tissue Value 1.309 Pass
Collimation Ratio 3.799 Pass
1 2 3 4 5 MEAN
LARGE BM264.6 265.2 265.4264.4264.7264.90.350.13
WIDTH 499 502 505 500 502 502 2.060.41
MED I UM BM1 9 9 . 3 197. 5 196. 8 1 9 5 . 8 197 . 0 1 9 7 . 3 1 . 13 0 . 57
WIDTH 432 430 428 428 428 429 1.600.37
SMALL BM142.7 140.4 140.0141.3142.0141.30.980.70
WIDTH 364 362 361 364 364 363 1.260.35
8-15
Standard Positioning Problems
DPX-IQ Service Manual Rev. E (3/98)
Example 2
Figure 2 is a QA in which the standard was not on the table. Note the errant Spillover value and
Air Counts (the Low KeV counts indicates the counters rolled over). In this case Spillover will
usually be approximately 1100/0. However, this is unpredictable as individual AGS boards try to
handle the superimposed signals from the unattenuated x-ray beam.
I DETECTOR
.+" -,
C /I \ // \ AMPLIFIER COUNTS COUNTS
0
/ .l-" \ / SETIING LOWkeV HIGH keV
,.., / / I "\./
U / ,_/ / I \
N / I \ 250 283312 46448
/ I \ 300 293533 62419
T / I \ 350 261792 107250
S / I \ 400 281987 190755
/ I \ 450 352064 273731
/ I \ 500 374781 287616
/
I \ 550 311120 259901
- ... /
I
\
...... 600
650
700
204298
116512
64794
276326
344339
380410
AMPLIFIER SETIING 750 40192 337149
Lights Pass
Peak Setting 495 units Pass
Background \Low keV) 2 cps Pass
(High keV) o cps Pass
Beam Stop Accion Pass
Percent Spillover 129.64% Fail
Chi Square 6 Pass
Air Counts (Low keV) 14087 cps Fail
(High keV) 3593432 cps Fail
Air Ratio . 255.10 Fail
Transverse Mechan~cs 12398/12401 steps Pass
Longitudinal Mechanics 19755 /19755 st6rnffis
Tissue Value 3.350 Fail
Collimation Ratio 2.119 Fail
2 3 4 1 5 MEAN S
---- ---- ---- ----
LARGE BM 61.7 44.6 90.0 35.6 70.4 60.519.2131.79
WIDTH 53 61 85 62 70 6610.8316.36
MEDIUM BM 6.5 49.2 35.1 38.7 27.5 31.414.2645.44
WIDTH 42 71 47 42 65 5~12.2122.8~
SMALL BM 47.7 1.3 38.2 78.5 33.3 39.24.8462.3
WIDTH 42 42 58 112 43 5926.9945.44
Example 3
Figure 3 is a QA run with the standard upside down and backwards. Spillover was measured
through the white delrin plastic plug and Tissue Value was measured through the brass
Spillover piece.
DETECTOR
1 2 3 4 5 MEAN
LARGE BML974. 41986.71968.41966.0.983.21975.1fl 8.10 0.41
WIDTH 385 390 379 383 382 384 3.66 0.95
MEDIUM BM2357. 72362.02359.02343. ~340 .12352.lt! 8.90 0.38
WIDTH 450 447 451 449 454 45~ 2.32 0.51
SMALL BM2 746 . 22755 . 52729 . 82741 . ~ 728 . 92742 . 3 8 . 44 0 . 3 1
WIDTH 526 523 518 520 519 521 2.93 0.56
Example 4
Misplacing the standard by approximately 1" toward the foot end will give results as in figure 4 .
Spillover is measured through the air and the Air Counts and Peak are measured through the
brass.
DETECTOR
1 2 3 4 5 MEAN S
---- ---- ---- ----
LARGE BM 0.2 0.1 -0.4 1.6 4.1 1.1 1 . 61142.4~
WIDTH 1 1 1 1 1 1 0.00 0.00
MEDIUM BM 0.2 0.1 -0.4 1.6 4.1 1.1 1.61142.4~
1 1 1 1 0.00 0.00
o .~ 0.00 0.00
WIDTH 1
SMALL BM 0.0 0.0 0.0 0.0 0.0
WIDTH -48 -40 -22 -13 -13 -2714.33 0.00
Example 5
The QA in figure 5 is what would happen in the rare instance that someone would inadvertently
remove the standard during the QA (after the Peak) and then replace it having realized the QA
was in progress (after the Tissue Value Test). Although the five lines are scanned across the
block as normal, the errant Air Counts will cause the 8M values and widths to deviate from
expected.
I DETECTOR
I
C /i'\ ",. AMPLIFIER COUNTS COUNTS
/-
"- I /~ \ /
SETIING LOWkeV HIGH keV
0 /
"- ..... ,/ // I \ \ /
/
I I
U I \\ /
N I
I I "'-,/ 250 57565 768
I \
300 66736 1245
T I
I
\
I \ 350 56262 4445
S I I \ 400 44794 21453
/ I '- 450 56982 49398
",
/
I " ..... 500 75050 66630
I " 550 73011 57766
I 600 49197 43491
650 25197 50310
700 10186 69146
AMPLIFIER SETIING 750 3846 75318
1 2 3 4 5 MEAN S
---- ---- ---- ---- ----
LARGE BM 0.2 0.3 0.2 0.2 0.2 0.2 0.2 10.9i
WIDTH 1 1 1 1 1 1 0.00 0.00
MEDIUM BM 0.2 0.3 0.2 0.2 0.2 0.2 0.2 10.9i
WIDTH
SMALL BM
1
0.2
1
0.3
1
0.2
1
0.2
1
0.2 oj 0.00 0.00
0.2 10.9i
WIDTH 1 1 1 1 1 1 0.00 0.00
Example 6
Placing the standard backwards with the brass piece toward the rear and the delrin plug
forward, yet both toward the foot end, will result in Spillover and Tissue Value being taken
through the black plastic of the block and Air Counts and Peak are taken through the delrin plug
(see figure 6). .
I DETECTOR
I
C I /'
AMPLIFIER COUNTS COUNTS
/'\ /
SETIING LOWkeV HIGH keV
' ' IIIIII \ ~_/' /
0 ".- /
/
U I ..... 1' I 250 40941 480
N {
I
\
\
T
I 300 47661 579
/ I \ 350 40710 2317
S I \
400 30253 13773
", .....
/ \
I I 450 34890 34957
-'"
".
I 500 44995 47395
I .....
.... 550 43933 40989
I 600 29830 29962
650 14794 31216
700 5702 41411
AMPLIFIER SETIING 750 2272 45334
1 3 4 5 2MEAN SD
---- ---- - - - - ----
LARGE BM 5.0 592.9 621. 3 -5.8 645.3 371.lJ304.3!B1.89
WIDTH 42 400 493 42 454 2 86 2 0 1 . 51>0 . 4 3
MEDIUM BM 12.0 717.8 657.3 687.4728.3 560.5275.43*9.15
WIDTH 42 501 485 502 533 4 1 ~ 185 . 9435 . 0 7
SMALL BM -0.2 898.3 109.3 95.0 55.2 231. 335.59.44. ~
WIDTH 42 608 94 76 46 173218.2$26.01
If a customer reports a QA failure on a scanner that seemed to be operating fine and showed no
signs of mechanical problems while running the QA, the QA results should be examined for
characteristics like the above examples. Ask the customer to check the position of the standard
if it is still on the table, or have them rerun the QA making sure the block is placed correctly.
Check the signal at TP4 of the AGS board with an oscilloscope. Look for noise on the wave
form. This test-point is the input for the AGS , so noise here means a problem, somewhere
towards the PMT. Also, check TP19 on the AGS with an oscilloscope. This is the output of the
AGS and would be a good indication of the AGS board's condition. If a storage oscilloscope is
used to look at these wave forms, they will look similar to the channel 1 signal (see Figure 1). If
a normal oscilloscope is used , the signal will appear as a superimposed multitude of bipolar
signals with different amplitudes.
~ Stopped :
! I ,-+]
8 Acqui sitions
T~
.:> : 10 . 38).1S
@: -15 . 12).1S
r
l .... :
t -
i
~ .j i l il l ~~-~
r -
l
r'
i
f
If the TP19 wave form looks abnormal, replace either the AGS or the AGSDCA printed circuit
board assemblies. If the spare part is not immediately available, then disconnect J8 on the AGS
and run a calibration . This will run the signal through the AGS at unity gain. If normal results
reappear after this modification, the cause of the problem is confined to a defective AGS or
AGSDCA board.
-E-Ie-c-tr-ic-a-l-N-o-is-e-----------------------------;8;--2~f
DPX-IQ Service Manual Rev. E (3/98)
If the "End of Exposure Alarm" rings during the time the Air Matrix scan is running, see "Alarm
Pings During Scan" in section 8.11 . If the Shutter Open or the X-ray On lamps on the front
panel blink, during the test, see section 8.11 also.
While the Air Matrix scan is running, measure the voltage at test points 1, 2, 5 and 6 of the
XORS board. They should have approximately the following values respectively:
0.150VDC, 3.8VDC, 0.150VDC, and 3.8VDC. A pure DC voltage on an oscilloscope at test
point 1 and 5 of the XORB will verify that there is constant current through the filament of the X-
ray Insert. If there is AC ripple in excess of 0.2 Vpp, the power supplies should be replaced.
Start and stop the x-rays several times while observing the voltage at TP3 and TP7 of the
XORS board. This is the programming voltageJrom the SSC, and although this voltage is
dependent on the feedback information returned to the SSC, the voltages at the XORB board
test points should be approximatelythesame each time the x-rays are produced. The AC ripple
on this signal must be less thanO.2Vpp. Replace the SSC board ifthe proper control signal is
not present.
When the x-rays are off between scans, verify that the shutter and collimator slide apertures line
up with the aperture to the Tube Head beneath. Take some alignment pictures and verify that
the image is a bright, round image. If the image is slightly elliptical, the collimator, shutter and
Tube Head apertures are not aligned. Open and close the shutter and collimator to make sure
that they return to the same position each time. Turn the x-rays back on, and make sure that
you obtain nearly the same count rate each time the shutter is opened or the collimator is
cycled. Either the shutter or the collimator slide may be adjusted by loosening the set screw on
the solenoid mounting bracket and rotating the solenoid. Tighten the set screw after
adjustment.
8.5.2 No Counts
When the table to the right of the peak graph on the Quality Assurance Results printout is
entirely filled with zeros, use the Static Counter program in the Diagnostics (see section 5.1) to
create x-rays at 76 kV and 150 uA and open the Shutter for sampling. Then check the following
items:
B. If not, the voltage or current ramping has probably failed . Are both the red and
green LED's on the MAX Board illuminated?
2. If the red and green MAX Board LED's are illuminated, verify the following
test point voltages:
c. X-ray On LED
Test the X-Ray On LED by inserting it into the Power On
receptacle. Replace if defective.
If either of the test points is at 1.0 VDC, the High Voltage Power
Supply is delivering as much current as it possibly can, and has
automatically limited the voltage.
Check TP2 and TP6. If the voltage is approximately 3.8 VDC, the
voltage has been set properly.
Feel the heat sinks on the back of the High Voltage Power
Supplies. If one of the power supplies is 'cold, this is usually the
defective one . If one is warm and the other is hot, replace the hot
one. The best troubleshooting technique may be to substitute
power supplies.
If either TP2 or TP6 are incorrect, the possibility exists that one of
the High Voltage Cables or the Tube Head is shorted. The short
may be possible to find with an ohm meter, but often it takes
several kV to break down the defective component. It will be
difficult to troubleshoot this problem by any method other than part
substitution.
~
~! p pp
_C_A_U_T_I_O_N_:__D_o_n_o_t_att_e_m_p_tt_o_r_a_m_p_t_h_e__o_w_e_r_s_u__l_y_a_b_0_v_e_5_0_k_V_w_ith_o_u_t_a_ _
cable connected, or arcing will occur near the connector.
~Stopped: 14 Acquisitions
:' 1 T-i
... ... . ·t
A: 10 . 381..1S
~
[
± @: -15.12I..1s
r ' .. .
t T
.....
.
+.
I
[
f
L ... .. . .. ... ... ·f ..
t
f ±
.1
i Jl j
t
l) 1 1 I J
-; .
T
..
..f. .
I
t
.-f -.,
!
CAUTION : Never disconnect any connectors from the AM P board when the
power is on, or AMP-U1 will be destroyed.
it.
blister a finger.
~ Stopped: 8 ACQuisitions
:: '-+1 T-l
a: 10 .381Js
@: -IS .1 2IJS
.. . ......
.
+ .
~
t
~
If there is a signal, check the continuity of the coaxial cable with the
yellow strain reliefs. Remem.ber that this cable pas?es through a
connector at the Arm Bulkhead, and a poor connection there could be the
cause of the problem.
Mark the position of the large and small collimator apertures on the slide channel
edge. Then lift the Collimator, taking care not to damage the limit switch, and
mark the open and close position of the shutter aperture center. Lift the shutter
slide, taking care not to damage the limit switch, and visually determine if the
apertures above match up with the bottom aperture.
Take another picture with the alignment film holder. The x-ray image should be
round. If the image is even slightly elliptical, the apertures are not aligned.
Remove the Collimator assembly from the Tube Head, and hold it up to a light.
Look through the aperture with the Shutter open and closed and the Collimator in
each position.
Verify that the SSC is properly controlling the current. TP11 of the MAX board should be
approximately 0.75 volts . This voltage is dependent on feedback information from the
power supplies, which makes it difficult to troubleshoot by any means other than SSC
substitution.
Check the current through the X-ray Insert. The absolute value of the voltage at'test points 1
and 5 on the XORB board is proportional to the current through the X-ray Insert. 1 millivolt is
equal to 1 micro-ampere of current. A current setting of 750 IJA should give a reading of 0.750
VDC at test points 1 and 5. If either of these voltages vary from the expected by more than 10
millivolts, the MAX board could be at fault. If TP1 and TP5 are more than 15 millivolts apart,
substitute new high voltage power supplies.
Air Counts
8-28
DPX-IQ Service Manual Rev. E (3/98)
8~6 Arcing
The X-ray Tube Head Insert is an evacuated glass enclosure. An AC current is applied to the
filament inside the insert. It glows like the filament in a light bulb, and electrons are boiled off
into the evacuated space. A high voltage is applied between the anode and the cathode
causing electrons to rush toward the anode, striking it and creating x-rays. As long as the insert
is properly evacuated, there can be no internal arc. However, no insert can be totally evacuated
and impurities can be ionized creating a lightning like effect; arcing. During the arc the
resistance of the insert is dramatically decreased and a large amount of current flows.
The significant amounts of electromagnetic energy released inside the scanner can cause
problems with the electronics of the scanner system and may result in abnormal operations of
the scanner. An arcing scanner will have one or more of the following symptoms:
• complete lock up of the system requiring rebooting the computer,
• a diagnostic error code message which is undefined or inappropriate,
• receiving a screen that says an error has occurred refer to appendix I in the manual,
• a stripe or artifact in the image of the patient's scan, or
• 15-4 Transverse Motion Failure message.
To specifically locate the source of an arc,it is necessary to find out which high voltage power
supply provides the excessive current. This :may be done by recording the power supply current
. monitors at TP1 and TP5 of the XORB board with a storage oscilloscope. An arc typically
draws 1.5 mA from the power supply for about 1 ms. The trigger voltage should therefore be
set at 1.5V for TP1 and for TP5.
Arcing can also occur inside the high voltage connectors. This will usually result in a plainly
visible black or brown carbon track through the grease on the connector. In any case, once the
high voltage connectors have been removed from the Tube Head and the power supplies, they
should not beTe~connected without first being cleaned and regreased (see procedure
DXSV0002 in chapter 6 appendix).
After cleaning the old grease off of the connectors, they should be carefully inspected for carbon
tracks. Look for these tracks on both the rubber cable connectors and on the phenolic sockets
of the Tube Head. If such tracks are found after cleaning, the following are the options for
returning the scanner to service:
c lf ~arbontracks are found on the rubber cable connectors, they can be removed by excising the
dam~ged section with a sharp blade. Severe tracks can burn quite deep into the rubber, so
care must taken to remove all the damaged rubber. After the carbon tracks have been
removed, additional grease must be used when repacking the connection to fill in the volume of
the removed rubber.
If the above procedure requires too much rubber to be removed, or the carbon track looks very
severe, another option is to replace the high voltage cable.
~'Arctracks will be impossible to see on the sockets in the Tube Head. Therefore, if arc tracks
are seen on the rubber cable connectors, the sockets should be sanded with emory cloth as a
precaution. After sanding the socket, remember to flush the socket with cleaning solution to
remove any particles.
If the tracking inside the socket is too severe or the carbon track cannot be removed, another
option is to replace the Tube Head.
'"
Arcing
8-30
DPX-IQ Service Manual Rev. E (3/98)
The High Voltage Cables can also impede transverse motion toward the front Limit Switch. This
is caused when the cables have been tied down without enough play for the Tube Head and
Detector to move all the way forward. These problems should be investigated by manually
tripping both Limit Switches while inspecting for cable conflicts, binding, or tension problems.
On Total Body scans; a'limit switch out of position could allow the Source/Detector Carriages to
hit the frame or panels'b efore the Limit Switch is actuated.
Wiring
In rare instances, the wires from the Shutter Solenoid and/or Fans can snag on the bolts that
protrude through the frame on the foot end of the scanner. This is solved by properly tying
down these wires.
Transverse Motor
Check the Transverse Motor for a broken wire in one of the internal coils, or a bad electrical
connection to its Centent Motor Controller.
Transverse Centent
The Centent Motor Controller, if defective, will cause transverse motion problems. Sometimes
the controller works well enough to acquire some scans, but it will not provide enough torque to
complete every fast scan. The CURRENT SET voltage at terminal 11 on a properly operating
controller should be 14 to 16 VDC (transverse motor wired in parallel).
Transverse Belt
The Transverse Belt should not be excessively tightened or this will cause excessive binding in
the transverse mechanism. It should be possible to deflect the belt by 4 cm when it is properly
tightened. Sometimes the spare belt material near the clamp on the Tube Head Carriage
comes into contact with the forward gear and prevents the scanner from going all the way to
Home position.
Drive Wheels
The wheels that support the Tube Head and Detector Carriages must be adjusted so that they
come into perfect contact with the Transverse Rails. These wheel are best inspected with the
table top, front panel and arm covers off so one can sight down the extrusions. However, it is
possible to test the wheels by preventing any wheel from turning and seeing if the carriage will
still move. By preventing any wheel from turning, you should be able to tell that it slides along
the extrusion while the others roll. This indicates that the wheel has not been excessively
tightened down. This is least likely to be the source of transverse motion problems, as it is
unlikely that the adjustment of the wheels would have become any tighter over time. They .
would be more likely to loosen over time and fail alignment tests (see Air Matrix Test, section
5.1). Adjustments can be made by loosening and rotating the eccentric bearings of any of the
lower wheels.
Cable Track
The major impedimenttolongitudinal motion is the plastic Cable Track that runs through the
trough at the rear of the scanner.
This track is attached at two points: at the Rear Longitudinal Carriage and to the scanner frame
(low, rear and center) each spot by 4 bolts. Also, adequate slack must be left in the cables
inside the Cable Track or they will stop the arm from moving fully to the foot end.
Should the Cable Track detach from the scanner frame, it will slide freely in the trough and will
eventually cause trouble. This can allow the Cable Track to get in between the Rear
Longitudinal Carriage and the scanner frame on the foot end preventing the tripping of the limit
switch.
Slip Clutch
A slip clutch is part of the longitudinal motion system to limit torque. This is a feature to protect
the patient should he/she pinch an arm or leg between the back side of the scanner and the
Arm Column. If the Slip Clutch is set too loose, it will fail to move the belt and will just "slip" as
the motor turns. This may produce the following symptoms:
• failure of Quality Assurance 'scan Mechanics test
• compressed image in limited areas of the image
• a Diagnostic Failure #23-3
On Total Body scans, a limit switch out of position could allow the Source/Detector Carriages to
hit the frame or panels of the scanner before the Limit Switch.
If the mechanical stop is reached before the Limit Switch is actuated, check carefully to see
which part of the Arm Assembly is in contact with the Table Assembly. The front part of the
Lower Transverse Extrusion is clamped to the Longitudinal Drive Cable at the front of the
scanner. If the Lower Transverse Extrusion is not clamped in such a way that it forms a 90
degree angle with the length ofthe table, the rollers at the front end of the Lower Transverse
Extrusion may strike the 'end of the scan table before the Limit Switch is actuated.
Longitudinal Motor,
Check the motor,for a broken wire in one of the internal coils, or a bad electrical connection to
the Centent 'Motor Controller.
Longitudinal Centent
The longitudinal Centent Motor Controller may be the cause of a failure. The current set voltage
should be roughly 9-11 volts at terminal 11 on a properly operating controller. If this voltage is
not correct, replace the Centent Controller.
Longitudinal Belt
The Longitudinal Belt should not be tightened too much or this will cause the brackets holding
the gears to deform at either end of the scanner. When the belt is properly tightened, it should
be possible to deflect the upper and lower sides of the belt so that they touch within 8 cm of the
gears at either end.
'--
Mechanical Failures 8-33
LUNAR® DPX-IQ Service Manual Rev. E (3/98)
Drive Wheels
The rollers in front and the wheels in back that support the Arm must be adjusted so that they
come into perfect contact with the Longitudinal Rails. Test them by preventing any wheel from
turning and see if the carriage will still move. By preventing any wheel from turning, it should be
possible to slide the carriage along the rail with one wheel dragging while the others roll. This
indicates that the wheel has not been excessively tightened down. Adjustments can be made
by loosening and rotating the eccentric bearings of any of the lower wheels.
If the transverse mechanics of the scanner become imprecise, it may cause a Limit Switch to be
tripped during a scan. This will usually only happen during a total body scan or during an Air
Matrix test since both of these scans involve full-width scanning. Once the switch is closed, the
motors stop, the End-of-Exposure Alarm sounds, the amber SHUTfER OPEN lamp turns off
and after a few seconds a Diagnostic Failure #15-4 appears on the screen.
During full-width scanning, the Detector Carriage comes very close to the Limit Switches, so any
imprecision will cause a Limit Switch to be tripped and the scan will be aborted with a Diagnostic
Failure message for Transverse Motion Failure.
A cause of this problem is a loosening of the first drive Reduction Belt which connects the
Transverse Motor to the first Reduction Pulley. This loosening causes the belt to "walk" on the
pulleys causing enough imprecision inthe motion to trip a switch. To tighten the belt, first
remove the Pulley Shroud and loosen all four nuts that h"old the motor in place. Then, while
holding the motor such that the belt is pulled taught, tighten the nuts to secure the motor in
place. Replace the shroud and test the scanner (an Air Matrix test works well).
If the Tube Head cables come into contact with the cable bundle entering the cable track, it may
be impossible for the scanner to complete all of the necessary transverse steps away from the
operator. Consequently, on each scan line the detector will move closer to the front transverse
limit switch, and the switch may be eventually closed. Form the cable bundle exiting from the
cable track into an arch such that the Tube Head cables will move under the arch rather than
running into the bundle. This arch must not be too high. If it is, there will not be sufficient slack
in the Tube Head cable bundle to allow the Tube Head to move to the front transverse limit
switch. Also, if the arch is too small, it may cause too much slack when the Tube Head is at the
forward side of the table. This causes the Tube Head cable bundle to rub against the inside of
the front panel causing a scraping noise to be heard. If the above does not solve the problem
then check the other mechanical components. Electrical components that could cause such a
failure are the Centent Motor Controller, the Transverse Motor, the OINK board or the SBC
board.
These lines are caused by Automatic Gain Stabilizer (AGS) trying to adjust the input signal
which is at an excessively high count rate. However, the AGS is not at fault, and the correct
action is for the customer to attenuate the x-ray beam. For an x-ray beam generated at 3 mA
the tissue areas need to be the equivalent of 12 cm. If the patient does not have 12 cm of
tissue, the operator must make up the difference with rice. Lunar normally recommends placing
a rice bag (on its side to create 12 cm of rice in the x-ray beam) along the side of the patient's
thigh for thin or osteoporotic patients:
In the case the last test fails, perhaps the voltage at AGSDCA-T4 needs to be increased 10 or
20 mV. Check the rollover at AGS-T12 with an oscilloscope and an x-ray beam at 150 ~ to
verify that there ·is no roll,;.over or roll-under when this voltage is increased. If stability can not be
maintained in air at 150 ~ and in 5-18 cm of water at 3 mA then it will be necessary to replace
some parts to determine which item is causing the instability. The following items can cause
this stability problem : detector, AGS board, both DCA boards, and the AMP-03 board.
If the signal changes suddenly at AGS-TP4, the cause must be the circuitry that controls the
high voltage to the Scintillation Detector, the Scintillation Detector itself, the circuitry on the
AMP-03, or the coaxial cables carrying the signals up the Scanner Arm . Measure the signal at
AGS-TP4, while taking voltage measurements at AMP-TP1 (the high voltage to the Scintillation
Detector) and SSC-TP20 (the control voltage from the SSC to the Bertan PMT-10A).
If the AMP-TP1 voltage changes at the same time the AGS-TP4 amplitude changes, but the
SBC-TP20 signal does not change, then the Bertan PMT-10A power supply or the coaxial cable
from the power supply to the AMP-03 must be faulty.
If the SBC-TP20 voltage changes at the same time the AGS-TP4 amplitude changes (AMP-TP1
will also change), replace the SBC.
Shutter Failures
Certain problems can cause the shutter to close intermittently completely attenuating the x-rays.
The shutter-open lamp should turn off and the end of exposure alarm will sound. A IIShutter
ll
Failure error message should appear on the monitor screen. Scanning should cease at this
moment, but if it does continue, blue lines should be expected in the image. Always ask the
operator if a "pingll sound (the end of exposure alarm) is heard, and whether the Shutter Open
or X-Ray On lamp turns off. In the case of a shutter problem, the Shutter Open' light will turn off.
Unstable AC Line
If the customer's AC line conditions are unstable, voltages below 95 VAC will begin to drop the
high voltage on the x-ray tube. The counts will drop significantly with only a change of a few kV.
However, this normally causes black or dark grey lines in the scan image.
Arcing
If the tube head is arcing, the arcs will be more likely to occur at the higher current settings.
-
Arcing generally also causes other strange symptoms such as:
• Scanner stops in the middle of a patient or QA scan with no error message or with
Diagnostic failure 15-4 or 31-6 message.
• Scanner stops during a QA scan with the error message "Print Buffer Error"
• The QA scan fails intermittently because the "Mean value deviates from expected" or
because there is "Excessive variation in standard values"
• The scanner periodically leaves the message on the screen "Starting X-rays Please
Wait" for a longer than normal period of time (more than 10 seconds)
1. View the Quality Assurance History (see section 8.15). In particular note the trend in
BM Values as discussed in the Failing QA's section of this chapter.
3. Replacement of the tube head if no arc tracks are discovered, or if the repacking of the
connectors does not eliminate all arcing symptoms.
Summary
Except for the x-ray relay, the cause of the problem will be difficult to 'find. Try to obtain some
additional information. If the customer is hearing the end-of-exposure alarm, or seeing the
yellow x-rayon lamp flickering, you can deduce the tube head control cable is bad. Likewise, it
is possible to decide on the shutter/collimator/fan cable. In the case of no alarm or lamp
_indications, orin the case where large deviations occur in the peak, the upper and lower cable
bundles should be replaced, as problems with the coaxial cable will be too expensive to
troubleshoot.
Check the size of the trochanter, neck and Ward's areas and verify that the sizes are
reasonable. The software may not have correctly selected the baseline causing the bone edges
to be improperly fixed. This false baseline could be due to one or more of the following
problems:
• There are not enough lines below the pelvis and above the trochanter (>25 lines) for an
adequate baseline to be established. The Operator's manual recommends starting the
scan 5.1 cm below the trochanter, and verifying that there are 15-20 lines before the
ischium appears in the image.
• If the x-ray beam passes into air during the femur scan, the software will incorrectly set
the baseline. White lines on the edge of the image, or unreasonable bone areas are
indications that the beam has passed into air. If this happens, the patient must be
rescanned with a soft tissue equivalent material (such as bages of uncooked rice) placed
against their hip.
• New systems will be delivered with LUNAR Tissue Bags (pIn 0784). These bags are to
be used as described above. Extra bags may be ordered through the L.UNAR Customer
Service Department. .
• Be sure the customer places the ROI box near the pelvis end of the neck with nearly
equal amounts of soft tissue in each end of the box and none of the ischium or
trochanter. The software moves the box down the neck towards the trochanter during
the search for the Ward's Triangle area.
To insure proper baselines, Lunar recommends that tissue equivalent material (as mentioned
above) be placed along side of and on top of the hips of very thin or osteoporotic patients.
In the case of a patient having unusually low bone mass, the program may have difficulty in
locating bone edges as it scans. In this case the only option is to slow down to Slow or even
Detail mode, and the automatic width option may have to be changed to manual if the bone
edges are still not tracked properly.
If the white lines that occur in the image when the beam passes into air continue to be white
when the beam passes back into the patient's body, the Automatic Gain Stabilizer (AGS) may
not be properly adjusted. There is a test for proper AGS operation (see ASG Calibration Test).
If manual analysis is used, the ROI boxes must be inside bone image only to obtain a correct
BMD . Include no soft tissue points in manual ROI's.
Category Limit
Chi Squared 45
Skew ±.O7
Kurtosis ±.O7
NOTE: All six categories MUST PASS during testing for new installations!
Chi Squared values ~45 that fail should be considered in the light that the Air Matrix test is a
$tatistical evaluation, and therefore, there is a statistical chance that a properly functioning
scanner could fail the Air Matrix test. Skew -and Kurtosis values will pass between -0.07 and
0.07. Anything outside of these limits is failing. These limits are valid only for scans made with
.1/4" (6 mm) of aluminum in the beam~ Limits for scans made with more or less aluminum have
not been determined. .
If the failing Air Matrix hasvaluesthat are close to passing (as described above), check the
Deviation Image (see below).Jf that looks good, run another Air Matrix (perhaps a shortened
test depending on 'time limits). The second test should pass. If not, a problem exists in the
scanner.
A misadjusted AGS DCA could be the cause of a failing Air Matrix. To determine if this is the
cause before parts SUbstitution is used to solve the problem, refer to the AGS Adjustment
Procedure in Section 6.2. NOTE: This test and adjustment procedure will not be of use.if the
PMT Detector is defective. Verify that the Detector is good by examining QA History trends in
%Spillover, Air Ratio and Air Counts (see Section 8.14).
Otherwise, the following parts should be substituted until passing Air Matrix results can be
obtained:
• ',' Scintillation Detector
'. Automatic Gain Stabilizer Board (AGS)
• AGS DCA
• DPX DCA
A poor Deviation Image will have colored lines, streaks or spots (indicating counts higher or
lower than expected). Colored lines or streaks going longitudinally in the image indicate an
alignment problem. Recheck the alignment of the scanner including the levelness of the
scanner. Also, check the scanner frame for any bending that may have occurred at the site or
during shipment.
Colored lines or streaks going transversely across the image indicate a problem with the AGS,
the AGS DCA or the DPX DCA (also see section 6.2). Again, replace parts until a passing Air
Matrix is obtained.
/' '
The amber LED is controlled by a comparator on the OINK Board. This LED illuminates when
approximately 40~A (or greater) of current is passing through the X-ray Insert. On Spellman
system scanners the amber light may appear to fade out slowly when x-rays are ramped down.
This due to the Spellman power supplies bleeding off current and is not a problem.
If the voltage ramping fails, the High Voltage Power Supplies must be shut completely off before
a second ramping attempt is made. If the amber X-RAY ON LED turns on momentarily, then
turns off for a few seconds before coming on steadily, the system is having difficulty ramping the
high voltage. One of the High Voltage Power Supplies may be defective.
The SBC does not monitor the Insert current once the proper level has been reached.
Therefore, if the current to the Insert is interrupted, the SBCwili not recognize the problem and
will not alert the computer, or terminate the exposure.
If the X-RAY ON LED blinks once the x-ray high voltage has been set, there must be a problem
in one of the following areas:
• Tube Head Control Cable-The most likely cause of an intermittent problem in this
circuit is a broken wire in theTube Head Control Cable. The Red wire is most likely the
broken one, as a broken blue or black wire should blow the MAX Board Fuse. Turn the
scanner off ,and test for continuity between MAX Board test points TP4, TP5, and TP13.
• MAX "Board-The MAX Board may be operating intermittently. Verify that the wave
formson'T P4, and TP5 match the wave forms on page 9.17 and 9.18. Substitute a
replacement MAX Board.
• OINK Board-Noise may be present on the line extending from pin 3 of U5 to the SBC
board. This noise would cause the bell to ring after every scan line during Total Body
scans at 1500A current modes. Adding a 0.1 OF capacitor between pins 1 and 3 of U5
on the OINK Board should attenuate this noise (see section 7, ECO #1393).
If the Shutter is actually opening and closing intermittently and uncontrollably, the problem is
either on the OINK, or a broken wire between the OINK and the Shutter/Collimator Assembly.
Remove connector J11 from the OINK Board and measure the resistance between pin 5 and
pin 6. This will be a measurement of the resistance of the Shutter Solenoid and the cable
through the Cable Track. If the problem occurs on every scan, run a scan while making this
continuity check. If the cable and solenoid seem to be working properly, substitute a new OINK
Board.
If no Diagnostic Failure Code is reported, the scanner may still be in motion continuing with the
scan. Abort the scan and remove the patient from the table. In any case, note the status of the
SHUn'ER OPEN light and the X-RAY ON light.
• Both the SHUTTER OPEN light and the X-RAY ON light are on and steady. This would
indicate a faulty OINK board. Replace it.
• The SHUTTER OPEN Iightis' off. This could indicate the following:
.. The LED becomes defective during the ,scan and since the shutter solenoid and this
light are wired in series, the Shutter closed and the Alarm sounded .
•, The cable lothe amber Shutter Open light broke during the scan with the same
result as above.
• The shutter solenoid failed and the Shutter closed followed by the Alarm sounding.
• The OINK is faulty.
• The SSC is faulty.
• The X-RAY ON light is off. This could indicate the following:
• X-ray production has halted. This turns off the Xray On light and sounds the Alarm.
Check the 110 cable connections at the serial port on the back of the computer and at the SSC.
Se sure both connections are tight and that the thumb screws are used to hold the connectors
firmly together. Also, inspect the 25 pin connector on the SSC. This connector is very fragile
and may have been damaged. Se very careful when connecting the I/O cable to the SSC.
If the ports are properly configured and communication still fails, then the probable causes are:
• A faulty 1/0 cable.
• A faulty SSC.
• A faulty system board async (serial) port.
Replacement of these parts is the best troubleshooting method. The system board async port is
a part of the computer's system board.
Another possibility is that the SSC has lost its firmware or the firmware has become corrupted.
This can be remedied by downloading the program again. See Monitor Directory, section 5.1,
for instructions.
DOS Installation
CONFIG.SYS
rem LUNAR Corporation, 313 West Beltline Hwy, Madison, WI 53713. (608) 274-2663
rem File: config.sys DOS version: 6.20
files = 50
buffers = 45
stacks = 0,0
dos = high,umb
shell = C:\command.com C:\ IP le:900
device = C:\dos\himem.sys
device = C:\dos\emm386.exe 64 ram
AUTOEXEC.BAT
PATH=C:\;C:\DOS
@rem LUNAR Corporation, 313 West 8eltline Hwy, Madison, WI 53713. (608) 274-2663
@rem File: autoexec.bat . DOS version: 6.20
@echo off
break off
prompt $p$g
C:\dos\smartdrv.exe 512
call C:\LUNAR\setdpx
: - (check disk integrity and)
cis
echo RUNNING CHKDSK ON DRIVES
for %%f in (C:) do if EXIST %%f\nul chkdsk %%f IF
: - (Create a temporary timestamp file for todays date)
echo. I date> C:\defrag.tmp
:-- (Check to see if the official timestamp file exists)
if not exist C:\defrag.dat goto CheckDate
: - (A timestamp file exists, check if this is a new day by comparing files)
fc C:\defrag.tmp C:\defrag.dat I find "no differences encountered" > \nul
CONFIG.SYS
rem LUNAR Corporation, 313 West Beltline Hwy, Madison, WI' 53713. (608) 274-2663
rem File: config.sys DOS version: 7.0
switches = If
[menu]
menuitem=dos;Start MS-DOS
menuitem=win,Start Windows '95
menuitem=dpx,Start Lunar DPX
menudefault=dpx,8
[common]
files = 50
buffers = 45
device=C:\windows\himem.sys Itestmem:off
dos=high,umb
she!i=C:\windows\command.com Ip
[dos]
device=C:\windows\emm386.exe ram highscan noems
device=C:\cdrom\necjde.sys /d:mscd001
[dpx]
[win]
AUTOEXEC.BAT
PATH=C:\;C:\WINDOWS;C:\WINDOWS\COMMAND;C:\LUNAR
@rem LUNAR Corporation, 313 West Beltline Hwy, Madison, WI 53713. (608) 274-2663
@rem File: autoexec.bat DOS version: 7.0
@echo off
break off
prompt $p$g
goto %config%
:dpx
:dos
Ih C:\windows\smartdrv 512
call C:\LUNAR\setdpx
Ih C:\mouse\mouse.exe /0
Ih C:\windows\command\doskey
if %config%==dos goto end
C:
cd \LUNAR
call dpx
goto end
:win
call C:\LUNAR\setdpx
win
:end
echo.
C: .
cd \
MSDOS.SYS
[Paths]
WinDir=C:\WINDOWS
Win BootDir=C:\WI N DOWS
HostWinBootDrv=C
[Options]
Logo=O
BootGUI=O
'---- .
BootMulti=O
I
;The following lines are required for compatibility with other programs.
;Do not remove them (MSDOS.SYS needs to be >1024 bytes).
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxa
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxb
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxc
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxd
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxe
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxf
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxg
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxh
;xxxxxxxxxxxxxx:xxxxx:xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxi
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxj
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxk
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxXxxxxxxxxxxxx I
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxm
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxn
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx xxxxxxxxxxx xxxxxxxxxxxxxxxxxx 0
;xxx xxxxxxxxxxxxxxxxxxx xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxp
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxq
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxr
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxs
To use these features press F4 at the Main Menu to select Database Utilities and then press F2
to select QA History.
The Quality Assurance History screen displays three columns of data results that are stored in
the database. The 11 highlighted categories are selected by default. Using thE~ arrow keys and
F4, F5 and F6 keys, the various categories can be selected or deselected depending on what
needs to be examined. Since only one category can be graphed at a time, simply move the
highlight to the item to be graphed and press F3.
Pressing F1, F2 or F3 selects View, Print or Graph and the next option is to enter the starting
and ending dates for the time period to be examined. Once that is done to view the data, press
[Esc] to start the printout or generate the graph.
When viewing the 11 default categories, or more, all the information will not fit on the screen.
Use the arrow keys to view the data off the screen to the right. All 11 default categories will fit
on one printed page. Printing more data will simply place the overflow on a separate page. In
graphing, it is possible to use the arrow, PgUp and PgDn keys to highlight a particular date's
QA. Note that the point on thegraphfor the highlighted date is also highlighted. The exact
value of the category is also displayed by the date.
LUNAR tries to analyze every system's QA History database file (DPXS.DBF) in order to
discover systems in need of service. This database file is converted to a Lotus '1-2-3
spreadsheet so that the data can be easily analyzed. LUNAR has determined, through these
analyses, acceptable limits for the variability observed for three important factors: Low Channel
Air Counts, Air Ratio and % Spillover.
The acceptable limit of change for Low Channel Air Counts can be expressed three ways: 1) no
more than 1% change/month, 2) no more than 100/0 change from the value at install or 3) no
change greater than 300 counts/QA.
The limits for Air Ratio are: 1) no change greater than 0,002/month, 2) no more than 0.02
change from the value at install or 3) no more than 0.0001 change/QA.
The limits for % Spillover are: 1) no change greater than 0.05/month or 2) no more than 0.002
change/QA.
The Peak setting normally drifts up or down. A long term upward trend is expected as the
Detector ages, but a downward trend is not unusual. The ideal operating peak is 500 units ±25.
However, Peaks between 400 and 600 units are acceptable. If the Peak is found to be further
from 500 than ±1 00 units, it should be adjusted (see section 6.1) until the Peak is 500 units
again (±25).
NOTE: The changes per month and per QA must be consistent over a three month
period or 63 QA's to be considered valid. Changes from time of install are invalid
if major parts were replaced since install, such as the Tube Head or PMT.
Changes should be compared to values after these replacements.
If any sudden jumps are noticed in the values of any of the categories, determine first if these
are due to service work such as a Tube Head replacement. If service work causes a shift in the
QA values, then compare present values to those obtained only after the service.
For complete analysis of the QA History database, Lunar suggests periodically sending a copy
of the DPXS.DBF file to the Lunar Customer Service Department.
Appendix B.A
AGS Board Troubleshooting
/'
',.
AGS Board
OPX-IQ Service Manual Rev. E (3/98)
Appendix 8.B
AMP Board Troubleshooting
Appendix 8. B
OPX-IQ Service Manual Rev. E (3/98)
It is located next to the Detector in a metal case in the arm assembly of the OPX-IQ.
Electrical Tests
TP1 HIGH VOLTAGE to Detector.
TP2 INPUT SIGNAL
TP6 BIPOLAR AMP OUTPUT (peak amplitude of approximately 2.4V)
TP7 +12Vdc
TP8 -12Vdc
TP9 GND
-A-M-P-B-oa-r-d---------------------------8~8~-1'
DPX-IQ Service Manual Rev. E (3/98)
Appendix 8. C
DCA Board Troubleshooting
Appendix B.C
DPX-IQ Service Manual Rev. E (3/98)
',.
DCA Board
8C-2
DPX-IQ Service Manual Rev. E (3/98)
There are two DCA PCBls in the DPX-IQ. The AGSDCA and the DPXDCA. The AGSDCA is
used in conjunction with the Automatic Gain Stabilizer (AGS) PCB to control the AMPLIFY and
ATTENUATE function of the AGS. The DPXDCA discriminates and detects the photon energy
levels for the SBC, where the actual event counting takes place.
The two DCAls are NOT interchangeable, although they share the same part number (2640) as
ADJUSTABLE DCA PCB. .
Appendix 8.D
MAX Board Troubleshooting
',-
Appendix 8.0
DPX-IQ Service Manual Rev. E (3/98)
The current at a given kV across the insert is a function of insert filament temperature . The
MAX board adjusts the current by regulating the filament temperature to provide the proper
current as set by the S8C. The relationship is Insert IJA X 1000 = SBC input (in volts DC) , so
for an SSC input of 5VDC, Insert IJA = 5000.
The Positive High Voltage Power Supply has a current monitor output (mA MONITOR); the
voltage at this output is proportional to the current passing through the X-ray Insert. This
monitor voltage is connected to the MAX board as feedback (mA FEEDBACK).
Electrical tests
When the board is powered up, and x-rays are being generated, the following measurements
may be performed: '
1
With a VOM set on 200 VDC, you should measure about 18 to 21 VDC on (fP's 4, 13 and 5
referenced to TP3 (GND). - L. _ _ _ _ _ _ _ _ _ _ _ _
To check the current setting, measure the voltage at TP8. At this test point, a to 1 volt
corresponds to a to 1000 uA (1 rnA). If the current is set to 750 uA (0.75 mA), you should
measure 0.750 VDC +/- 0.010 VDC at TP8.
"
Appendix B.E
OINK Board Troubleshooting
Appendix 8.E
DPX-IQ Service Manual Rev. E (3/98)
Optical isolation is used to isolate the +26VDC supply from the +5VDC and +12VDC (logic)
supplies, and eliminate interference between the motor circuits and the event (CH1, CH2)
signals.
--- .
Appen ixB.F
OMI Board Troubleshooting
Appendix 8.F
DPX-IQ Service Manual Rev. E (3/98)
Electrical tests
The OMI has three wires connected to it:
• Red - +5 volts
• Green - Ground
• Black - Signal
The signal from the OMI board is nominally rectangular in shape. The signal is generated when
the Interrupter Disk alternately., passes and occludes a light beam between the two posts on the
photo-transducer. With the 'beam occluded by the disk, the output signal should be about 5
VDC (close counts). When the beam is allowed to pass through a slot in the disk, the Signal
output is near ground potential.
Appendix B.G
('
j
I
I
)
J)!J3 (
to ~~um flU! )
8R 9~22
:1
Appendix B.G
sse Board Troubleshooting
DPX-IQ Service Manual Rev. E (3/98)
Appendix 8. G
DPX-IQ Service Manual Rev. E (3/98)
• The SSC should have the proper EPROM chip installed. OPX-IQ scanners should have, '.' .r \ J
Appendix B.H
XORB Board Troubleshooting
.-~ "
Appendix 8. H
_L_U_N_A_R_®_ _ _..,.,--...,.......,___----..,....D-P-'X--I-O-S-e-rv-ic-e-M_a_n_ua_1_ _ _ _ _ _ _ _. . :. R.:. :;. e v _~: _(~,'~:: 8)
Electrica.l Tests
• Measure impedance between XORB ground plane and the electrical pan. It s~ c}, lk; bf'
less than an ohm.
• There should be no low impedance (less than 10 ohms) reading between any signal 0 1"
signal ground and chassis ground on the XORB when the associated signal connectors
are disconnected.
---,. - "\ • The following test points should have the following readings:
~" " ):
XORB TP 5
XORB TP 5
TP3 and TP7 are the kV programs from the SSC and should climb to 3.8VDC +/- O.04VDC.
J;8 )9
v4l /' )-, f1 ~J
/
('"(j t1 Ie
----=---,-,-~-,.,;-"""':"-~-:-----.:.....------------------___;:;_;_T--~"
XORB Board 8H-1
OPX-IQ Service Manual Rev. E (3/98)
;' - .
r}.: f .
. r·· ·..
,. j r> .
~ J" \" '. ,~~, - i'
, .
.:y
'~ ';:" . -
../ , ~,1\
k ' ...., ~. - • " .
... .... .... - ----------~- -' .. - ... .