Quick Guide
Unscrew locking nut. Adjust plunger to desired volume. Tighten locking nut. Adjust body / grip position using
locking screw on handle side.
1
2
1
2
3
4
5
1
2
3
4
5
Insert feed needle. Insert vent cannula. Attach tubing to syringe. Insert connecting tube and Pierce bottle rubber
Connect tubing. needle, assemble parts. stopper, lock bottle.
1
2
3
4
5
1
2
1
2
3
4
5
Activate handle until liquid Activate in vertical position to expell Introduce small pin and If insufficent, disassemble valves
reaches needle. liquid and eliminate any air bubble. push to loosen ball in valve. and clean.
2
3
4
1
2
3
4
5
Rinse with warm water and Unscrew body and remove. Clean residues in glass Detach plunger, clean and let dry.
purge completely. barrel with a brush.
Reassembling
1
2
3
4
2
3
4
Carefully reassemble glass barrel into Gently insert plunger into glass Activate one or two times
I.SYR.E - F0423 body. Attach plunger to handle rod. barrel then screw body. to check proper functioning.