PQ Test Protocol MPA Sample Compartment 20231211 222536

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 5

OVP - PQ Test Protocol

Company: Novartis Pharmaceuticals


Operator: Admin
Instrument Type: MPA Sample Compartment
Optics Configuration: Sample Compartment with: NIR, Quartz, TE-InGaAs [Internal]
Accessory: None
Instrument Serial Number: 2006068
Instrument Firmware Version: 1.470 Oct 5 2006
OPUS/DB Version: OPUS 8.5(SP1) Build: 8, 7, 10 / DB: 8,7,10,139
Overall Test Result PASSED
Test expires: 12/11/2024, 10:25:36 PM (GMT-8)
Test Date/Time: 12/11/2023, 10:25:36 PM (GMT-8)
Test Spectra Path: C:\Users\Public\Documents\Bruker\OPUS_8.7.10\Validation\Data
\20231211\222536
Date of last PQ Reference
Measurement: 12/11/2023
Comment:

Signal to Noise Test

Minimum S/N(area 1): 4000 Measured S/N: 11410

100% Line Test

Maximum 100% Line Deviation: 0.5 Measured 100% Line Deviation: 0.40

Interferogram Peak Test

Minimum Amplitude[%] : 70 Measured Amplitude[%]: 105.6

Energy Test

Maximum allowed Value: 30 Measured Value: 6.0

Wavenumber Accuracy Test - BRM 2065 (35.5 Deg. C, Peak is T. corr.)

Sample Material: BRM 2065


Specified Peak: 10245.50 cm-1 Maximum Deviation: 0.60 cm-1
Measured Peak: 10245.69 cm-1
Corrected Peak: 10245.46 cm-1 Measured Deviation: 0.04 cm-1

Photometric Reproducibility Test - Glass Filter A

Maximum Deviation[%]: 0.8 Measured Deviation[%]: 0.08

Overall Test Result = PASSED

Date and Signature Date and Signature

________________
Page 1 of 5 Page Signature
OVP - PQ Test Protocol
Signal to Noise Test, Spectra
100% Line Test, Spectra
Energy Test, Spectra
101.0
99.5 100.0 100.5
Transmittance [%]
99.0

10000 9000 8000 7000 6000 5000


Wavenumber cm-1
1 PQ_MPA_Sample_Compartment_Energy_1.0
2 PQ_MPA_Sample_Compartment_Energy_2.0
3 PQ_MPA_Sample_Compartment_Energy_3.0
4 PQ_MPA_Sample_Compartment_Energy_4.0
5 PQ_MPA_Sample_Compartment_Energy_5.0

1 df52d363-8c7e-4be9-8fb7-5712716ce180
2 44aa1424-3f09-4b62-b4fb-23d42a049f5c
3 5cc03752-284f-41eb-b466-a7124a43ea02
4 c44a47bb-c02d-4e8c-bce1-bfece1474f9b
5 b7c90f4f-dbe2-41af-8967-8e2aeeff9bfb

Energy Test, Measurement Parameters


Number of Scans: 10 Preamplifier Gain: Ref
Resolution [cm-1]: 8 Zero filling: 2
Aperture: Open Phase Correction: Power Spectrum
Acquisition Mode: Double Sided,Forward-Backward Velocity: 10 KHz
Apodization Function: Blackman-Harris 3-Term Phase Resolution: 128
Low Pass Filter: 10 KHz High Pass Filter: Open

________________
Page 2 of 5 Page Signature
OVP - PQ Test Protocol
Interferogram Peak Test, Spectrum
0.25
0.15
0.05
-0.05
-0.15

14000 12000 10000 8000 6000 4000 2000 0


Points
PQ_MPA_Sample_Compartment_Energy_1.0

df52d363-8c7e-4be9-8fb7-5712716ce180

Interferogram Peak Test, Measurement Parameters


Number of Scans: 10 Preamplifier Gain: Ref
Resolution [cm-1]: 8 Zero filling: 2
Aperture: Open Phase Correction: Power Spectrum
Acquisition Mode: Double Sided,Forward-Backward Velocity: 10 KHz
Apodization Function: Blackman-Harris 3-Term Phase Resolution: 128
Low Pass Filter: 10 KHz High Pass Filter: Open

________________
Page 3 of 5 Page Signature
OVP - PQ Test Protocol
Wavenumber Accuracy Test BRM 2065 (35.5 Deg. C, Peak is T. corr.), Spectrum
1.0

10245.69
0.8
Absorbance Units
0.4 0.6
0.2

10300 10250 10200 10150


Wavenumber cm-1
PQ_MPA_Sample_Compartment_Frequency.1

18f2f2d7-256b-43ec-a1d9-e1485b5ab244

Wavenumber Accuracy Test BRM 2065 , Measurement Parameters


Number of Scans: 32 Preamplifier Gain: Ref
Resolution [cm-1]: 4 Zero filling: 8
Aperture: Filter BRM2065 Phase Correction: Power Spectrum
Acquisition Mode: Double Sided,Forward-Backward Velocity: 10 KHz
Apodization Function: Blackman-Harris 3-Term Phase Resolution: 64
Low Pass Filter: 10 KHz High Pass Filter: Open

________________
Page 4 of 5 Page Signature
OVP - PQ Test Protocol
Photometric Reproducibility Test - Glass Filter A, Spectrum
25
Transmittance [%]
10 15 20
5

10000 9000 8000 7000 6000 5000


Wavenumber cm-1
PQ_MPA_Sample_Compartment_GFA.0

8370ebe3-9015-4422-9767-dd7e67bf1000

Photometric Reproducibility Test - Glass Filter A, Measurement Parameters


Number of Scans: 32 Preamplifier Gain: Ref
Resolution [cm-1]: 8 Zero filling: 2
Aperture: Filter NG9 Phase Correction: Mertz
Acquisition Mode: Double Sided,Forward-Backward Velocity: 10 KHz
Apodization Function: Blackman-Harris 3-Term Phase Resolution: 64
Low Pass Filter: 10 KHz High Pass Filter: Open

________________
Page 5 of 5 Page Signature