LIMITATIONS
GENERAL TECHNICAL PARAMETER
Samples that exceed the linearity limit (800 U/L) should be
diluted with an equal volume of saline and re-assayed and the Mode Kinetic
final results multiplied by two. Wavelength (Filter) 405 nm
Reaction Direction / Type Increasing
Reagent Blank No
QUALITY CONTROL Sample Vol. 50 µL
Reagent Vol. 1000 µL
To ensure adequate quality, control sera normal and abnormal
Delay Time 60 Sec
control should be used. These controls must be performed &
validated before the patient samples are assayed. The control Interval Time 60 Sec
frequency must be at least once a day, after each calibration No. of reading 2
and should be adapted to Quality Control procedures of each Measuring Time 120 Sec
laboratory and the regulatory requirements. Results should be Factor 2658
within the defined ranges. If values fall outside of the defined Reagent Blank Abs.(Max) NMT 1.0 Abs
ranges, each laboratory should take corrective measures. Calibration Method Fix factor
Quality control material should be used in accordance with Linearity 800.0 U/L
local guidelines.
Decimal Places 1
Temp. 37⁰C
EXPECTED VALUES
Unit U/L
Serum,plasma Ref. Low (Male / Female) 8/5 U/L
Ref. High (Male / Female) 61/36 U/L
Men Women
Serum, plasma: 8 - 61 U/L 5 - 36 U/L REFERENCES