PSLP Engstrom Carestation Rev 1.511
PSLP Engstrom Carestation Rev 1.511
Rev: 1.51
Date: October 2004
Doc: PSLP Engström Carestation Rev 1.51
By: Paul Renner / CARE P&L
Page 1
Table of Contents
1 Introduction
2 Product Information
Page 2
1 Introduction
This document contains the service support launch plans for the Datex-
Ohmeda Engström Carestation and is intended to be used as a tool by
Datex-Ohmeda Subsidiaries and Distributors as a guide in preparing local
product launch and service business plans.
The information contained within this plan is valid as of the date shown on
the front cover, but is subject to change without further notice.
2 Product Information
2.1 Product Launch Schedule
CARE Marketing, in conjunction with the channel organizations, will
determine the launch schedule for each country.
Page 3
2.3 Product Specifications
2.3.1 Technical Specifications
All specifications are approximate values and can change without notice.
Mechanical construction
Width 53 cm Width 38 cm
Depth 78 cm Depth 36 cm
Weight 66 kg Weight 29 kg
Environmental specifications
Pneumatic specifications
100 cmH2O
Maximum limited pressure
Range of inspiratory pressure 5 to 100 cmH2O ± 10%
Minimum limited pressure
-1.5 cmH2O
(subatmospheric)
Maximum working pressure range 12 to 100 cmH2O
Page 4
Electrical requirements
85 to 132 VAC, 47 / 63 Hz
Supply voltage (switchable)
187 to 264 VAC, 47 / 63 HZ
Internal batteries
The Engström Carestation is not designed for use during the transport of patients
between facilities. Two internal 12 VDC sealed lead-acid batteries supply battery
backup. Batteries are used as backup power in case of a power failure. The battery
is in a float charge state as long as the system is connected to a primary electrical
source.
• Capacity to operate for > 30 minutes minimum, 120 minutes typical, battery fully
charged
• Ampere-hour rating: 4 A-h
• Voltage requirement: 12 VDC
• Current requirement: 7.5 A
• Shelf life: 9 months following last charge.
• Recharge time: full discharge to full charge in 8 hours.
• Estimated life: four to six years; 250 100% discharge cycles.
• The System Status menu shows the status of the internal battery.
2.4 Warranty
2.4.1 Product Warranty
The warranty period for the Datex-Ohmeda Engström Carestation is 12
months from the date of original delivery to Buyer or to Buyer’s order, but in
no event for a period of more than two years from the date of original
delivery by Datex-Ohmeda to a Datex-Ohmeda Authorized Dealer.
(Reference warranty statement in the User Reference Manual)
Disposable accessories and supplies are intended for single use, therefore
the warranty is limited to ninety (90) days from the date of shipment or when
used, whichever occurs first.
Page 5
3 Service Support Policy
3.1 Technical Support
Technical support for the Engström Carestation will be provided by trained
personnel through the existing service channels currently utilized for support
of other CARE P&L systems.
Page 6
3.2 Technical Training
In order to ensure that local launches of Engström Carestation are
properly supported, at least one representative of each country service
organization must successfully complete technical training before
shipments will be authorized by the CARE P&L.
• The Technical Training Team will determine the best method and media
to use to deliver training to local country service organizations based on
the individual requirements of each territory.
• The development of the technical training materials required for class
presentations will be managed by the Technical Training Team.
Page 7
3.3 Installations and Configuration
The Datex-Ohmeda Service Representative should perform the following
system setup, calibration, and preoperative tests during the installation of
the Engström Carestation:
Unpack and setup the Engström Carestation per Section 3 “Setup and
Connections” of the User’s Reference Manual.
Turn system on per Section 4, Page 4-2 of the User’s Reference Manual.
Access the Installation and Service menus per Section 4.1 of the
Technical Reference Manual.
Access the Service menu per Section 4.4 of the Technical Reference
Manual and perform:
Configuration: Set language, altitude and units.
(Note: Copy Configuration Menu can be used to save a configuration to a
PCMCIA card and then copy the configuration to additional machines)
Service Log: Review and reset error and alarm log entries.
Perform the Pre-use Checkout per Section 4, Page 4-2 of the User’s
Reference Manual.
Page 8
3.4 Planned Maintenance
The recommended service level PM interval for the Engström Carestation
is every 12 months. Reference Section 6 of the Technical Reference
Manual for PM requirements.
3.4.1 PM Parts
Page 9
3.6 Factory Repair and Exchange Service
Does not apply. The CARE P&L does not offer a factory repair service for
the Engström Carestation.
3.7 Traceability
Country Sales Leader shall maintain serial and/or lot number and date of
shipment records for each Product so that, if necessary for corrective action
purposes, the Product can be identified and traced. In addition, each Party
shall maintain the same or similar records with respect to its components
that are incorporated within the Product so that, if necessary for corrective
action purposes, the manufacturing and delivery lot for such products can be
identified and traced.
Page 10
Description Stock Number
Windows based EV Service Application 1505-8001-000
RS-422 to USB Adapter w/PC cable 1505-8586-000
System Software / Service Diagnostic &
Troubleshooting Software (compact
flash) 1505-8000-000
Page 11
4.2 Documentation
User and Technical Reference Manuals for the Engström Carestation are
available through the normal Datex-Ohmeda distribution channels.
Page 12
4.3 Spare Parts
The Engström Carestation Technical Reference Manual identifies all sub-
assemblies and components that are available as repair or maintenance
parts. Following is a list of spare parts that the CARE P&L recommends be
stocked by local country service organizations to ensure proper support of
the Engström Carestation.
Page 13
Tools & Spare Parts X = Part Also Used
Minimum Stocking Recommendations QTY on Avance
Display Module
1009-8243-000 Backlight Kit 1 X
1009-8289-000 CPU Board, Display unit (with PCMCIA 1
Frame)
1503-3012-000 Encoder Assembly 1 X
898794 Knob, Com Wheel 2 X
1009-5505-000 Membrane Switches, Right Side 1 X
1009-5507-000 Membrane Switches, Lower 1 X
1009-5506-000 Membrane Switches, left 1
Language Specific (see TRM) Keypad Set 1
1009-5938-000 Display, LCD 12-inch Color (includes 1
backlights)
897010 Filter, Fan 5 X
1009-5679-000 Door, PCMCIA 1 X
Page 14
Tools & Spare Parts X = Part Also Used
Minimum Stocking Recommendations QTY on Avance
Vent Engine
1505-3256-000 Regulator w/screws 1
1505-3218-000 Restrictor, Pneumatic 2
1505-3220-000 Gasket, Pneumatic restrictor 1
1505-3208-000 Transducer Flow 0-160 LPM 2
1505-3209-000 Valve Flow Control 160 LPM 2
1505-3215-000 Transducer O2 Conc 2
1006-3616-000 O-ring, EPR, 2-016 5
1505-3262-000 Gasket Inspiratory Manifold 2
1503-3204-000 Valve Seat Free Breathing 1
0211-1454-100 Valve Flapper (Free Breathing) 5
1503-3208-000 O-ring, EPR, 2-126 5
1505-5706-000 Harness (vent engine to VCB/VMB) 1
1505-3267-000 Valve Umbrella Check 5
1505-8502-000 Manifold Inspiratory Valve Assy 1
1505-8503-000 Pressure Sensor Assembly 1
1505-3059-000 O-Ring, VITON, BS 4518 0717-16 2
Outlet manifold
1505-3227-000 Actuator Voice Coil 1
1505-8568-000 Exhalation Valve Housing Assy 2
1505-3224-000 Diaphragm 5
1505-3225-000 Seal Rolling Shaft 1
1505-3236-000 Pull Exh Housing Latch 1
1505-3251-000 PCA Ultrasonic Neb Control 1
1505-5508-000 PCA Exh Flow Interface Brd 1
1505-5507-000 PCA Exh Flow Snsr Brd 1
Inlet manifold
1006-3430-000 Check Valve 4
1504-3708-000 Filter Disc, Wire Mech 2 Micron 5
1504-3612-000 O-ring, EPR 2-122 70 duro 5
Page 15
Tools & Spare Parts X = Part Also Used
Minimum Stocking Recommendations QTY on Avance
AC Power Cords
1505-3041-000 Retainer, Outlet Power Cord Jumper 2
1505-3033-000 Retainer, Inlet Power Cord 2
0140-6219-130 Screw, M3X8 PAN HD Sems 10
Cart
21947-029 Caster Swivel 5” without Lock 2
21947-028 Caster Swivel 5” with Lock 2
Module Rack
1505-5601-000 Cable, Patient Module Rack 1
Air Compressor
6985795 Air inlet filter 5
5471001 Filter kit f/OF302-1.5SDA 1
1006-3614-000 O-Ring –019 70 EPDM 2
Page 16