Download as pdf or txt
Download as pdf or txt
You are on page 1of 7

Optimization analysis for plane orientation in 3-dimensional

cephalometric analysis of serial cone-beam computerized


tomography images
Manuel O. Lagravère, DDS, PhD, FRCD(C),a Marc Secanell, PhD,b
Paul W. Major, DDS, MSc, FRCD(C),c and Jason P. Carey, PhD, PEng,dAlberta, Canada
UNIVERSITY OF ALBERTA

Objectives. The purpose of this study was to evaluate and reduce potential errors associated with superimposition of
serial cone-beam computerized tomography (CBCT) images using planes based on cranial base landmarks.
Methods. CBCTs from 10 patients were analyzed. The potential impact of errors in cranial base landmark
identification on assessment of the relative position of distant landmarks is mitigated by means of a mathematical
algorithm that ensures that the distances and angles between landmark identification points are maintained for
different images by readjusting the landmark coordinates.
Results. Significant improvement was observed after optimization. The errors found in a previous study were
significantly reduced, some by more than 90%. Errors found in the standardization were viewed in both
infraorbitals and mentons ranging from 1 to 3 mm.
Conclusions. The mathematical transformation to readjust the coordinates of ELSA, left and right auditory external
meatus (AEM), and dorsum foramen magnum (DFM) points significantly improves their use for image superimposition.
(Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2011;111:771-777)

Superimposition of anatomical structures or landmarks be overcome. Many studies5,6 have reported accuracy
in lateral cephalometric analysis has traditionally been and reliability of location of landmarks in CBCT im-
done for determining orthodontic treatment results and ages but few4,7-9 have reported using these landmarks
growth. The accuracy of superimposition depends on as reference planes for pre- and posttreatment image
examiner experience and landmark location. Leonardi superimposition. Some results show that this is not a
et al.1 stated that landmarks automatically or manually simple task.10 The purpose of this study was to evaluate
located and traced are not sufficiently accurate to be and reduce potential errors associated with superimpo-
used in general practice. Of the available structures for sition of serial CBCT images using reference planes
superimposition, studies on implants and human skulls based on cranial base landmarks.
have shown that superimposition on highly stable areas
of the cranial base is preferable.2,3 MATERIAL AND METHODS
Two-dimensional (2D) imaging does not fully rep- CBCT’s volumetric data (NewTom 3G Volumetric
resent a 3D structure, because much of the information Scanner, Aperio, Italy) taken at 110 kV, 6.19 mA, and
is lost when 3D structures are converted to 2D images.4 8 mm aluminum filtration from 10 patients participat-
With the appearance of cone-beam computerized to- ing in a maxillary expansion clinical trial were used for
mography (CBCT), many of the problems that occur this study. Images were taken using a 12-inch detector
when locating landmarks and evaluating changes can field at baseline before treating the patients. Ethics
approval was obtained from the Health Research Ethics
Board at the University of Alberta.
a
Clinical Assistant Professor, Orthodontic Graduate Program, Faculty Images were obtained and converted to DICOM
of Medicine and Dentistry, University of Alberta, Alberta, Canada. (Digital Imaging and Communications in Medicine)
b
Assistant Professor, Mechanical Engineering, Faculty of Engineer- format using the NewTom software (Aperio, Italy) with
ing, University of Alberta, Alberta, Canada.
c
Professor, Chair of Dentistry, Faculty of Medicine and Dentistry,
a voxel size of 0.25 mm. Using AMIRA software
University of Alberta, Alberta, Canada. (AMIRA, Mercury Computer Systems, Inc., Berlin,
d
Associate Professor, Mechanical Engineering, Faculty of Engineer- Germany), the DICOM-format images were rendered
ing, University of Alberta, Alberta, Canada. into a volumetric image using 512 ⫻ 512 matrices
Received for publication Dec 17, 2010; returned for revision Feb 11, giving a range of 400 to 420 DICOM slices. Sagittal,
2011; accepted for publication Feb 11, 2011.
1079-2104/$ - see front matter
axial, and coronal volumetric slices, as well as the 3D
© 2011 Mosby, Inc. All rights reserved. image reconstruction were used to determine landmark
doi:10.1016/j.tripleo.2011.02.017 positions.

771
OOOOE
772 Lagravère et al. June 2011

Table I. Descriptive statistics of coordinates during both transformation procedures (mm)


First transformation Second transformation
Mean SD Minimum Maximum Mean SD Minimum Maximum
ELSA x 0.000 0.000 0.000 0.000 0.000 0.000 0.000 0.000
y 0.000 0.000 0.000 0.000 0.000 0.000 0.000 0.000
z 0.000 0.000 0.000 0.000 0.000 0.000 0.000 0.000
AEML x 0.997 1.323 –1.790 2.390 0.000 0.000 0.000 0.000
y –0.026 0.856 –1.190 0.940 0.000 0.000 0.000 0.000
z 0.000 0.000 0.000 0.000 0.000 0.000 0.000 0.000
AEMR x –0.098 1.254 –20.080 1.900 0.000 0.000 0.000 0.000
y 0.109 10.047 –1.260 2.140 0.000 0.000 0.000 0.000
z 0.000 0.000 0.000 0.000 0.000 0.000 0.000 0.000
DFM x 0.000 0.000 0.000 0.000 0.000 0.000 0.000 0.000
y 0.665 1.322 –0.530 3.850 0.000 0.000 0.000 0.000
z 0.681 1.731 –0.820 50.050 0.000 0.000 0.000 0.000
InfraOL x 0.061 10.071 –1.850 1.940 0.361 1.216 –1.240 2.480
y –0.033 0.766 –1.130 1.100 –0.170 0.860 –1.740 10.090
z –1.281 2.841 –8.330 1.650 –0.693 1.442 –3.360 10.010
InfraOR x –0.184 0.898 –20.000 1.490 0.128 10.010 –1.440 20.040
y 0.250 0.624 –0.570 1.490 0.343 0.613 –0.640 1.180
z –1.159 2.705 –7.940 1.610 –0.505 1.342 –2.530 0.890
MnL x 0.450 10.088 –1.170 1.790 0.868 1.575 –1.760 2.910
y 10.026 2.577 –1.380 7.580 0.291 1.381 –10.050 2.720
z –1.195 3.144 –8.980 20.080 –0.531 1.569 –3.170 1.460
MnR x 0.172 0.985 –1.250 1.550 0.595 1.387 –1.960 2.120
y 0.983 2.693 –10.050 80.060 0.483 1.174 –0.560 2.960
z –10.039 2.945 –8.410 1.610 –0.300 1.533 –2.460 1.770
AEML, auditory external meatus left; AEMR, auditory external meatus right; DFM, dorsum foramen magnum; ELSA, midpoint between both
foramen spinosums; InfraOL, infraorbitale left; InfraOR, infraorbitale right; MnL, menton left; MnR, menton right.

Four landmarks are required to define a 3D anatom- tablished that both AEM right (AEMR) and AEML lie
ical reference coordinate system. The left and right in the XY plane and thus have a zero Z coordinate,
auditory external meatus (AEM) and dorsum foramen whereas DFM lies in the YZ plane and thus has a zero
magnum (DFM) were selected based on position and X coordinate.
early formation in skeletal growth. It has been reported
that most of the cranial base growth (⬎85%) occurs in First coordinate transformation procedure
the first 5 years of life, presenting minor changes after A detailed description of the transformation process
this age. The fourth point, ELSA, is defined in a pre- is given in Lagravère et al.10 Briefly, to transform all
vious publication as the midpoint between the left and global landmark coordinates to an ELSA coordinate
right foramen spinosum.11 ELSA was selected as the system, the vector describing the position of ELSA in
origin of the new Cartesian coordinate system. From the AMIRA coordinate system was subtracted from all
the origin, 3D positional coordinates for AEM left, anatomical landmarks, thus making ELSA the (0,0,0)
AEM right, and DFM were determined. Intrareliability origin of the reference Cartesian coordinate system.
values were determined using intraclass correlation co- This is illustrated in Fig. 1.
efficient for all 4 landmarks, repeating the process 3 Coordinate system transformations were performed
times for each image. Intrareliability values for each in 2 steps. The coordinate system was constructed using
Cartesian coordinate for all landmarks were reported in 2 planes defined by anatomical landmarks. The first
Lagravère et al.,6 showing the lowest value of 0.965 in plane, which represents a new XY plane, was defined
the x-axis for AEM left (AEML). using both AEM and ELSA; the 2 vectors were taken
Landmarks used in the present study are defined in with respect to the ELSA coordinate system. The sec-
Table I. The principal investigator located the land- ond plane, defined as a new YZ plane, was formed by
marks on each image 3 times. Spherical markers of ELSA and DFM perpendicular to the new XY plane.
0.5-mm diameter were placed indicating the position of In this set of transformations, left and right AEM
the landmark, and the software used the center of these coordinates had zero z-component, as they lie in the
spherical markers as coordinates. For purposes of stan- X⬘⬘Y ⬘⬘ plane. The second transformation is a rotation of
dardization of the coordinate axial system, it was es- the X⬘⬘Y ⬘⬘plane defined in the above steps to set the
OOOOE
Volume 111, Number 6 Lagravère et al. 773

Fig. 1. Coordinate system transformation proposal. ELSA, AEMR, AEML, and DFM are lying outside of the coordinate system;
thus, a transformation is performed to make ELSA the (0,0,0) center of the coordinate system and the other 3 points lie on their
respective 0 axis.

Y ⬘⬘Z⬘⬘plane in which lies the DFM anatomical landmark. ELSA in images 1 and 2 have the same relative distance
After this transformation, the DFM had zero X-compo- between each other.
nent. To obtain the error correction for AEML, AEMR,
Using the initial coordinate transformation, values and DFM in image 2, we solved 9 equations:
were obtained for the new infraorbital (right and left)
㛳xAEML
1
㛳 ⫺ 㛳xAEML
2
㛳⫽0
and menton (right and left) landmarks. Images were
measured 3 times and thus it was expected that an error 㛳xAEMR
1
㛳 ⫺ 㛳xAEMR
2
㛳⫽0
close to zero would be obtained. After the analysis was 㛳xDFM
1
㛳 ⫺ 㛳xDFM
2
㛳⫽0
completed, a second transformation was performed on
㛳xAEML ⫺ xAEMR
1
㛳 ⫺ 㛳xAEML ⫺ xAEMR
2
㛳⫽0
the 4 initial points.
㛳xAEML ⫺ xDFM
1
㛳 ⫺ 㛳xAEML ⫺ xDFM
2
㛳⫽0 (1)
Second coordinate transformation procedure 㛳xAEMR ⫺ x 1
DFM 㛳 ⫺ 㛳xAEMR ⫺ x 2
DFM 㛳⫽0
involving the 4 initial landmarks
The relative position of the AEML, AEMR, and 共x · x 兲 ⫺ 共x · x 兲⫽0
1
AEML
1
AEMR
2
AEML
2
AEMR

DFM landmarks with respect to ELSA should be the 共xAEML


1 1
· xDFM 兲 ⫺ 共xAEML
2 2
· xDFM 兲⫽0
same between 2 images. However, because of operator 共xAEMR · xDFM兲 ⫺ 共xAEMR · xDFM兲⫽0
1 1 2 2

errors, small differences in the coordinates are observed


between images. To correct for the measurement errors, where㛳 㛳 represents the Euclidean distance, ⫻ repre-
1
an algorithm was developed to ensure that the distances sents the dot product, xAEML are the coordinates of
2
and angles between AEML, AEMR, DFM, and ELSA AEML in image 1 and xAEML are the corrected coordi-
are maintained. A system of equations is solved to find nates of AEML in image 2, i.e.,
the necessary perturbation so as to make all distances 2
xAEML ⬅ xAEML,original
2
⫹ ␦uAEML (2)
and angles the same. The system of equations solved is
as follows. where ␦uAEML is the unknown error correction vector.
The algorithm receives the coordinates of AEML, The system has 9 unknowns, corresponding to the
AEMR, DFM, and ELSA for 2 different images, i.e., measurement error corrections for AEML, DFM, and
image 1 and image 2. Given that operator error usually AEMR and 9 equations; therefore, the system can be
exists in recording these values, we would like to solved using any nonlinear solver. In this case, MAT-
guarantee that the relative angle and distances between LAB’s (MathWorks, Natick, MA, USA) fsolve routine
the 2 images are conserved. To achieve this, we ob- is used.
tained an error correction vector, ␦ui, for every point in Note that the first 3 equations guarantee that the
image 2 such that the points AEML, AEMR, DFM, and distance between ELSA and AEML, AEMR and DFM,
OOOOE
774 Lagravère et al. June 2011

Fig. 2. Illustration of the correction of the landmark points between images. The 3 points with different colors represent the
following: original landmarks in image 1 (blue), original landmarks in image 2 (red), and corrected landmarks in image 2 (green).
ELSA is not modified between images; therefore, superposition is observed at point (0,0,0) (mm).

respectively, are the same for the 2 images. The other 3 RESULTS
equations guarantee that the relative distances between Once the raw data were transformed to a standard-
AEML and AEMR, AEML and DFM, and AEMR and ized ELSA coordinate system using ELSA as refer-
DFM are the same between images 1 and 2, respec- ence (0,0,0), both AEM as (X, Y, 0) and DFM as (0,
tively. Finally, the last 3 equations guarantee that all Y, Z), the new values were used to determine the
angles are the same. mean measurement differences for the 3 trials for
To verify the effect of optimizing the distances and each axis of every landmark and then averaged. The
angles of AEML, AEMR, and DFM with respect to equation used is
ELSA, landmarks from 2 images were input to the
algorithm. Fig. 2 shows the landmarks on image 1, the 关ⱍ M2 ⫺ M1ⱍ ⫹ ⱍ M3 ⫺ M2ⱍ ⫹ ⱍ M3 ⫺ M1ⱍ兴
original landmarks on image 2, and the corrected land- M⫽
3
marks after both transformations.
Once the new landmark coordinates are established, (3)
the first coordinate transformation procedure is per- where M1, M2, and M3 are each trial’s measurement on
formed to have a standardization of the global axis. To an image and M is the mean measurement difference.
verify the changes, 4 points outside of the coordinate This calculation was done for each coordinate axis and
system formation were located. These points were in- each image.
fraorbital left and right (InfraOL and InfraOR) and Values obtained for each transformation can be
menton left and right (ML and MR). Descriptive sta- viewed in Table I. It should be noted that after the first
tistics on all new coordinates and pairwise comparisons transformation, mean differences are very large, in
among the 3 measurement trials were made. some cases approximately 1 cm. Following the second
OOOOE
Volume 111, Number 6 Lagravère et al. 775

transformation, these values have improved greatly giv- Table II. Pairwise comparisons of means from 3 trials
ing mean differences of less than 1 mm. In the worst of after both transformations
cases, the range for the second transformation was Pairwise comparisons
approximately 4 mm. The standard deviations have also Mean difference Standard error
been reduced with respect to the points obtained only ELSA x1 x2 0.000 0.000
after the first transformation. x3 0.000 0.000
A pairwise comparison of the 3 measurement differ- x2 x3 0.000 0.000
ences obtained from each trial measurement after both y1 y2 0.000 0.000
transformations was also done. Results are presented in y3 0.000 0.000
y2 y3 0.000 0.000
Table II. Table II shows there is no significant variation z1 z2 0.000 0.000
between measurement differences. Measurement dif- z3 0.000 0.000
ferences of approximately 0 mm are observed for the z2 z3 0.000 0.000
points used to generate the coordinate system. Mea- AEML x1 x2 0.000 0.000
surement differences of 1 mm are observed for points x3 0.000 0.000
x2 x3 0.000 0.000
that were away from the origin of the coordinate sys- y1 y2 0.000 0.000
tem. y3 0.000 0.000
y2 y3 0.000 0.000
z1 z2 0.000 0.000
DISCUSSION z3 0.000 0.000
Superimposition of images is a tool used by ortho- z2 z3 0.000 0.000
dontists to aid in diagnosis, treatment planning, treat- AEMR x1 x2 0.000 0.000
ment follow-up, and growth. Many types of superim- x3 0.000 0.000
x2 x3 0.000 0.000
positioning methods have been used for 2D images; for
y1 y2 0.000 0.000
example, lateral cephalograms. With the advent of 3D y3 0.000 0.000
imaging, however, further development in methods for y2 y3 0.000 0.000
superimposition is required. Some work has been done z1 z2 0.000 0.000
on the development of ways to superimpose 3D imag- z3 0.000 0.000
z2 z3 0.000 0.000
ing. One method is using gray scales to superimpose
DFM x1 x2 0.000 0.000
images. Cevidanes et al.12,13 used this style of CBCT x3 0.000 0.000
image superimposition to determine treatment changes x2 x3 0.000 0.000
associated with orthognathic surgery. Their method y1 y2 0.000 0.000
consisted of identifying the cranial base structures in y3 0.000 0.000
y2 y3 0.000 0.000
the images. These images were later subjected to dif-
z1 z2 0.000 0.000
ferent software to obtain mean surface distance and to z3 0.000 0.000
quantify how much, on average, the 2 surfaces differ z2 z3 0.000 0.000
from each other using color-coded graphical displays. InfraOL x1 x2 –0.229 0.167
Changes seen after using the software can only be x3 0.541 0.577
x2 x3 0.771 0.527
interpreted in the sense of surface distance changes in
y1 y2 0.144 0.357
tendencies (outward or inward) of large anatomical y3 –0.255 0.408
structures (maxilla and mandible). For this reason, a y2 y3 –0.399 0.327
different approach was investigated in this work to z1 z2 0.097 0.663
obtain changes in anatomical structures (whatever their z3 –1.039 0.684
z2 z3 –1.136 0.693
size) and reporting these changes in terms of each
InfraOR x1 x2 –0.777 0.218
coordinate axis individually. x3 0.191 0.479
Kawamata et al.14 also tried to superimpose 3D im- x2 x3 0.969 0.524
ages. The images used in their study were obtained y1 y2 0.432 0.299
from a CT machine and their method involved manu- y3 0.514 0.291
y2 y3 0.082 0.216
ally superimposing anatomical structures to view con-
z1 z2 0.213 0.752
dyle displacement after surgery. The steps involved in z3 –0.757 0.637
this process were to create lateral, axial, and frontal 3D z2 z3 –0.970 0.765
CT images of the pre- and postoperative temporoman-
dibular joint region. Then, both lateral images (pre and
post) were superimposed and rotated until anatomic
structures, such as zygomatic arch, mastoid process,
and infraorbital foramen overlapped. Once this was
OOOOE
776 Lagravère et al. June 2011

Table II. Continued errors in determining X, Y, and Z coordinates of dis-


Pairwise Comparisons tance landmarks. Errors present in landmarks chosen to
Mean Difference Standard Error standardize the reference plane can be magnified to
MnL x1 x2 0.002 0.411 points that are farther from these reference points. A
x3 1.302 0.747 minor error in setting the axis results in minor devia-
x2 x3 1.300 0.767 tions for points near the origin of the coordinate system;
y1 y2 0.020 0.576 however, this results in large deviations for points in
y3 0.437 0.655
y2 y3 0.417 0.736
regions farther away from the origin. There is a possi-
z1 z2 0.351 0.706 bility that errors in one axis could be canceled by an
z3 –0.796 0.744 error in another axis or errors in other reference points,
z2 z3 –1.147 0.723 but there is also a possibility that these could increase
MnR x1 x2 –0.475 0.505 the error present in the region of interest.
x3 0.892 0.658
x2 x3 1.367 0.779 The basis of this analysis and point selection has
y1 y2 –0.114 0.593 been described in a previous article.10 It has been
y3 0.725 0.557 reported that intrareliability of the landmarks forming
y2 y3 0.839 0.757 the coordinate system (ELSA, AEML, AEMR, and
z1 z2 0.388 0.755
DFM) was excellent in all axes.10 Four points (infraor-
z3 –0.450 0.727
z2 z3 –0.838 0.826 bital left and right and menton left and right) located at
the maximum distance from the cranial base reference
AEML, auditory external meatus left; AEMR, auditory external me-
atus right; DFM, dorsum foramen magnum; ELSA, midpoint be- system were located and used to determine the trans-
tween both foramen spinosums; InfraOL, infraorbitale left; InfraOR, formation potential to produce error when superimpo-
infraorbitale right; MnL, menton left; MnR, menton right. sitioning. When the potential envelope of error for the
No statistically significant difference was found (P ⬎ .05) on all reference plane system produced by the compounding
comparisons.
error of the landmarks defining the reference points was
applied using mathematical transformation, the error in
locating distant landmarks was high, as shown in Table
I in the first transformation column. By applying a
done, a joint image was created where colors demon- second mathematical transformation, it can be seen that
strated the amount of displacement of the condyle. This this envelope of error reduces significantly (Table I).
method, although practical, can present several draw- An alternative technique for CBCT image superim-
backs, as trying to determine coordinate displacements position is best-fit analysis of multiple cranial base
of structures cannot be done because no reference sys- landmarks and computer-aided superimposition based
tem has been determined. To apply this method to on best fit of object shapes in the cranial base.13 Opti-
growing patients is also difficult because anatomical mization analysis is another alternative to use when
structures are still changing and using this method in trying to determine a standard reference system based
the cranial base can be complicated, as visualization of on specific landmarks; this was the alternative verified
these structures is difficult when using only 3D images. in this article. Some variations of this technique are
Another way to approach superimposition is with the planned for future research; one such variation is to use
use of an optimization analysis. This analysis would 6 landmarks instead of 4 to optimize the coordinate
involve minimizing the total root mean square error system. Another would be to replace the landmarks
found over a series of fixed landmark positions. This is most prone to identification error and use other land-
a best-fit type of analysis. There are many ways of marks located in the vicinity of the original 4. Future
producing optimization analysis.15,16 Cevidanes et research is needed to critically evaluate the errors as-
al.’s12 approach involved, in the initial steps, a type of sociated with these alternative techniques.
optimization analysis. The method presented in the
present article has the goal to be a simple, precise, and CONCLUSIONS
reliable instrument to analyze changes within an indi- The objective of this article was to improve a 3D
vidual over time for assessment of growth and treat- method to superimpose CBCT images mentioned in a
ment outcomes. previous study.10 By using an optimization analysis to
Initial evaluation of the system presented in this reduce the mean square root error of landmark location
article found good reference plane reliability, but sec- of ELSA, AEML, AEMR, and DFM (landmarks used
ondary sensitivity analysis identified that small errors in to establish the reference coordinate system), the mean
locating cranial base landmarks had large potential errors were greatly reduced. This analysis gives good
OOOOE
Volume 111, Number 6 Lagravère et al. 777

prognosis on the establishment of an easy, accurate, and ing headgear treatment. Am J Orthod Dentofac Orthop
reliable method to superimpose 3D images. 2002;121:18-29; Discussion: 29-30.
10. Lagravère MO, Major PW, Carey J. Sensitivity analysis for plane
orientation in three-dimensional cephalometric analysis based on
REFERENCES superimposition of serial cone beam computed tomography im-
1. Leonardi R, Giordano D, Maiorana F, Spampinato C. Automatic ages. Dentomaxillofac Radiol 2010;39:400-8.
cephalometric analysis. Angle Orthod 2008;78:145-51. 11. Lagravère MO, Major PW. Proposed reference point for 3-di-
2. Bjork A. The use of metallic implants in the study of facial mensional cephalometric analysis with cone-beam computerized
growth in children: Method and application. Am J Phys Anthro- tomography. Am J Orthod Dentofac Orthop 2005;128:657-60.
pol 1968;29:243-54. 12. Cevidanes LH, Bailey LJ, Tucker GR Jr., Styner MA, Mol A,
3. Melsen B, Melsen F. The postnatal development of the palato- Phillips CL, et al. Superimposition of 3D cone-beam CT models
maxillary region studied on human autopsy material. Am J of orthognathic surgery patients. Dentomaxillofac Radiol
Orthod 1982;82:329-42. 2005;34:369-75.
4. Berkowitz S. A multicenter retrospective 3D study of serial 13. Cevidanes LH, Styner MA, Proffit WR. Image analysis and
complete unilateral cleft lip and palate and complete bilateral superimposition of 3-dimensional cone-beam computed tomog-
cleft lip and palate casts to evaluate treatment: part 1—the raphy models. Am J Orthod Dentofac Orthop 2006;129:611-8.
participating institutions and research aims. Cleft Palate Cranio- 14. Kawamata A, Fujishita M, Nagahara K, Kanematu N, Niwa K,
fac J 1999;36:413-24. Langlais RP. Three-dimensional computed tomography evalua-
tion of postsurgical condylar displacement after mandibular os-
5. Lagravère MO, Gordon JM, Guedes IH, Flores-Mir C, Carey JP,
teotomy. Oral Surg Oral Med Oral Pathol Oral Radiol Endod
Heo G, et al. Reliability of traditional cephalometric landmarks
1998;85:371-6.
as seen in three-dimensional analysis in maxillary expansion
15. Maes F, Collignon A, Vandermeulen D, Marchal G, Suetens P.
treatments. Angle Orthod 2009;79:1047-56.
Multimodality image registration by maximization of mutual
6. Lagravère MO, Low C, Flores-Mir C, Chung R, Carey JP, Heo
information. IEEE Trans Med Imaging 1997;16:187-98.
G, et al. Intraexaminer and interexaminer reliabilities of land-
16. Czaplicki A. Optimization solutions depend on the choice of
mark identification on digitized lateral cephalograms and format-
coordinate system. Acta Bioeng Biomech 2008;10:75-9.
ted 3-dimensional cone-beam computerized tomography images.
Am J Orthod Dentofac Orthop 2010;137:598-604.
7. Seckel NG, van der Tweel I, Elema GA, Specken TF. Landmark Reprint requests:
positioning on maxilla of cleft lip and palate infant—a reality? Manuel O. Lagravère, DDS, PhD
Cleft Palate Craniofac J 1995;32:434-41. Faculty of Medicine and Dentistry
8. Sachdeva RC. SureSmile technology in a patient— centered orth- Rm. 4051, Dentistry/Pharmacy Centre
odontic practice. J Clin Orthod 2001;35:245-53. University of Alberta
9. Ashmore JL, Kurland BF, King GJ, Wheeler TT, Ghafari J, Edmonton, Alberta, Canada T6G 2N8
Ramsay DS. A 3-dimensional analysis of molar movement dur- manuel@ualberta.ca

You might also like