Professional Documents
Culture Documents
186802-Model hc1
186802-Model hc1
| Service Manual
SYSTEM VERSION
COPYRIGHT
Copyright 2017, Human Gesellschaft für Biochemica und Diagnostica mbH, Wiesbaden,
Germany. All rights reserved.
No part of this documentation may be reproduced in any form, nor processed, copied or
distributed by means of electronic systems, without prior permission of HUMAN in
writing. Since all precautionary measures were taken into account in producing these operating
instructions, the manufacturer accepts no responsibility for any errors or omissions. This
includes any liability for damage that could arise from possible incorrect operation based on this
information. Subject to changes without notice as result of technical development.
TABLE OF CONTENTS
1 SAFETY INSTRUCTIONS 5
1.1 INTRODUCTION 5
1.2 USER WARRANTY 5
1.3 INTENDED USE OF THE INSTRUMENT 5
1.4 GENERAL SAFETY WARNINGS 6
1.5 DISPOSAL MANAGEMENT CONCEPT 6
1.6 BIOHAZARD WARNING 7
1.7 INSTRUMENT DISINFECTION 7
1.8 REPAIRS 7
2 INTRODUCTION 9
2.1 SERVICE MODE 9
2.1.1 Access to the service menu 9
2.1.2 Service functions 10
2.1.3 Quit service mode 23
2.2 USER MODE 24
2.2.1 UTILITIES menu 24
2.2.2 Access to the UTILITIES Menu 24
2.2.3 Functions of the UTILITIES Menu 25
2.2.4 Quit UTILITIES menu 33
2.2.5 PIN-Code forgotten 34
3 ADJUSTMENTS 37
3.1 REPAIR MATRIX 37
3.2 SERVICE CHECK 38
3.3 LCD CONTRAST 39
3.4 SPEED SENSOR DISTANCE 40
3.5 MIXER SPINNING 41
3.6 CHECK THERMAL BLOCK TEMPERATURE 41
3.7 CALIBRATE THERMAL BLOCK TEMPERATURE 43
3.8 SOFTWARE UPDATE 47
3.9 PARAMETER UPDATE 52
3.9.1 Parameter Import 52
3.9.2 Automatic Parameter Import 58
3.9.3 Parameter Export 64
3.9.4 Parameter Error 70
3.10 PARAMETER PROTOCOL 71
3.11 PRINTER CONNECTION 82
3.12 DPU 414 PRINTER SETUP 85
3.13 LOAD CUVETTECARD 89
3.14 PHOTOMETER CALIBRATION 93
3.15 VIRTUAL COMPORT DRIVER 96
4 INSTRUMENT ERRORS 97
7 APPENDIX 143
7.1 SOFTWARE STRUCTURE 143
CONTENTS
Safety Instructions 5
1 SAFETY INSTRUCTIONS
1.1 Introduction
This manual is considered part of the instrument and must be available to the
operator and the maintenance personnel. For accurate installation, use and
maintenance, please read the following instructions carefully.
In order to avoid damage to the instrument or personal injury, carefully read
the ”GENERAL SAFETY WARNINGS”, describing the appropriate operating pro-
cedures. Please contact your HUMAN authorised local Technical Service in the
event of instrument failure or other difficulties with the instrument.
Figure 1
Biological Hazard Symbol
1.8 Repairs
2 INTRODUCTION
-- While the line is scrolling then it is the only possible way to enter the service
menu by entering the service code 22926.
-- If the service code was entered correctly this message below will be dis-
played:
Warning
The Service Code is confidential and may not be accessed by unauthorized
person.
Test Functions
1. Photometer calibration
2. Optic-test
3. Channel-test
4. Mixer-test
5. Heating-test
6. Port A-test
7. Port B-test
8. Keyboard-test
9. Display-test
10. Chip-card-test
11. SD-card-test
12. Ext 5V-upply-test
13. Nebenlicht-test (ambient light test)
Calibration functions
1. Calibration temperature
2. Default parameters
Parameter functions
1. Parameter export
2. Parameter import
Photometer calibration
-- Press Enter key when < photometer calibr. > is displayed.
-- When this message below is displayed remove the cuvette from the meas-
uring channel, close the light protection cap and keep the cap closed while
processing then press Enter key.
-- When this message below is displayed press Enter key to complete the test
or you can exit the function by pressing the ESC key.
Optic-test
-- Press Enter key when <optic-test> is displayed.
-- The below message will be displayed after the optic test is completed (The
below values are an example only).
12
-- You can leave this function by pressing the ESC key and the LED will be
switched off.
Channel-test
-- Press Enter key when < channel-test > is displayed.
-- The below message will be displayed after the optic test is completed (The
below values are an example only).
Table 1
Minimum Maximum Unit
LED 0 1000 Digit 0 = off
Meas 0 65535 Digit
Ref 0 65535 Digit
Mixer-test
-- Press Enter key when < mixer-test > is displayed.
-- The below message will be displayed after the Mixer-test is completed (The
below values are an example only), to leave this function press ESC key and
the motor will keep spinning.
Motor 0 The actual target value for mixer motor speed [rpm]
rpm 0 The present value of mixer motor speed [rpm]
-- Enter the target motor revolution value and confirm with Enter.
-- The motor will turn accordingly.
-- The receiving signal of the hall-sensor on PCB will displayed as the revolu-
tions.
Table 3
Minimum Maximum Unit Tolerance
Motor 200 999 Rpm + / - 50
Heating-test
With this function you can check the heating and the temperature sensor in the
thermal block.
14
-- The below message will be displayed after the heating test is completed (The
below values are an example only), to leave this function press ESC key, but
before leaving the function put the heating back to 0% and it will switch it
off.
-- Enter the target heating power value [%] and confirm it with Enter.
-- The heater will start to heat up the thermal block according the selected
value.
-- The receiving signal of the NTC temperature sensor on PCB will be displayed
as digits.
-- Increasing values = thermal block is getting colder.
-- Decreasing values = thermal block is getting warmer.
-- When the test starts the heating will be turned off which means it will be 0%.
-- Enter a value between 1 and 100% and the heating will be switched on by
checking the pulse-pause relation.
-- By reading the NTC value you can check whether the temperature sensor
and the heating are working, if this value decreases while the thermal block
temperature rises then the heating is switched on.
Important:
In this mode it is possible to heat the thermal block up to 60°C.
Table 4
Minimum Maximum Unit
Heating 0 100 % 0 = off
Port A-test
With this function the serial port (Port A) can be checked, therefore special tests
plug which can be ordered as „service tool“.
-- The below message will be displayed after the Port A-test is completed (The
below values are an example only), to leave this function press ESC key.
Port B-test
With this function the USB Port (virtual ComPort) can be checked and the com-
munication between HumaClot Junior and the Computer can be tested, There-
fore an USB Cable, a computer, a virtual ComPort driver and terminal software
are required. If the USB port works a character wise communication between
terminal software and HumaClot Junior is possible.
-- The below message will be displayed after the Port B-test is completed(The
below values are an example only), to leave this function press ESC key.
-- With the Enter key you will start or stop repeat mode which will continuous-
ly transmit the code for the last pressed key to the serial port.
Table 6
Key Display Key Display
0 Number 0 Start Channel 1
1 Number 1
2 Number 2
3 Number 3
4 Number 4
5 Number 5
6 Number 6
7 Number 7
8 Number 8
9 Number 9
< Cursor left
> Cursor right
Mode Mode
Enter Enter
ESC ESC Test ends now
Keyboard- test
With this function the keyboard can be tested, the keys of the membrane key-
board can be pressed and will be confirmed by a beeping signal and shown in
the display.
-- Press any key when the message below is displayed, to leave this function
press ESC key.
Display-test
With this function the display can be tested.
-- Press Enter key when < display-test > is displayed:
-- Press any key when the message below is displayed, to leave this function
press ESC key:
-- The test will automatically fill the display screen with single characters.
-- After that a character set will be shown automatically.
ChipCARD-test
-- The readers unit will be tested by inserting and removing the chip card.
Table 8
Display shows Explanation / Reason
No card inserted
insert card
No card detected
Test failed
Reading or writing or card detect not possible
test failed
Check for correct direction/insertion of card
Check for correct card type
Card detected
please wait
Access to card in process
Test result OK
test ok
read / write access to card worked
Card detected
may remove card
Card can be removed
Ext 5V-supply-test
The HumaClot Junior provides (5VDC) voltage for an external Barcode Reader
plugged at (RS232) port, for this test a (RS232-test plug) is required to check the
outgoing signal.
-- This message below will be displayed (To leave this function press ESC key):
Calibration functions
Calibration temperature
This function is intended to calibrate the thermal block temperature to a target
value of 37.4 °C, for detailed calibration process please see chapter 3.7.
-- This message below will be displayed (to leave this function press ESC key):
Default parameters
This function is intended to load a default parameter setup from the memory of
the Analyzer, but this function will be processed without a 2nd safety query, so
any existing balanced cuvette will be deleted!
-- By pressing > key the displayed message will change as shown below:
-- When this message below is displayed, that’s mean the factory defaults
parameters have been loaded back to EEPROM:
Known Problem
Table 9
Problem Possible solution
Default parameter setup could not be loaded.
Low Level setup has been loaded.
Display shows: -- Check actual parameter setup
universal default -- Import parameters from SD-Card
parameters loaded -- Check / calibrate thermal block temperature
-- Calibrate photometer
-- Check application parameters
-- Check / re-enter conversion parameters
Parameter function
Parameter export
This function is intended to export the instruments setup parameter to a
PAREXP.TXT file on SD-Card.
-- Press Enter key when < param. im/export > is displayed.
Parameter import
This function is intended to import a parameter setup from PARIMP.TXT file on
SD-Card.
-- Press Enter key when < param. im/export > is displayed:
-- Press > key when < Import <- SD-Card > is displayed:
-- Press > key when < Import <- SD-Card > is displayed:
-- By pressing the > cursor key, the < UTILITIES > will be displayed:
-- The functions of the utilities menu will be displayed, by pressing the cursors
keys you can select between the functions and confirm it with the Enter key.
English Table 10
-- Off
-- On
-- AUTO
1 Printer -- Parameter-Protocol
• Global
• Actual Method
• All Parameter
-- Off
2 Port A
-- Ext. Printer
-- Off
3 Port B
-- Host
-- On
4 Beeper -- Off
-- Click
-- English
-- Deutsch
-- Espanol
5 Language
-- Francais
-- Portuges
-- Russian
6 Date/Time
-- on (= number)
7 PIN Code
-- off (= 0)
8 Cuvette-Test
-- Export -> SD-Card
9 Param. im/export
-- Import <- SD-Card
10 Photometer calibr.
26
Printer
-- When < Printer > is displayed, press the Enter key .
Table 11
Display shows Explanation
DEFAULT
Off Disables printout to external Printer con-
nected to Port A
Enables manually started printout to
external printer connected at Port A
On
Parameter Protocol
-- By pressing the right cursor the function will be changed , when the function
parameter protocol is displayed press the Enter key.
Port A
-- When the function Port A is displayed, press Enter key.
Port B
When the function Port B is displayed, press Enter key:
Beeper
When the function Beeper is displayed, press Enter key.
Language
In this function you will be able to change the language of the Analyzer expect
the language of the service menu, it will always be in English.
PIN Code
With this function the user will be able to change the PIN Code for the access to
the utilities menu and the setup menu for test applications.
-- When the function PIN Code is displayed, press Enter key:
Cuvette test
With this function it is possible to check the cuvette detection.
-- When the function Cuvette test is displayed, press Enter key:
-- By pressing the ESC key this function will be executed. ! Note: Make sure that are no
cuvettes in the measuring
channels while the instrument is
If any problem occurred with the detection of the cuvette during a measuring switched on and that are no cu-
run or during this test menu remove all cuvettes from the measuring channels vettes in the measuring channels
and switch the instrument off and on again. Then recheck this function. while Photometer check.
32
Parameter export
With this function you will be able to export the instruments setup parameter
to a PAREXP.TXT file on SD-Card.
-- When the function Parameter export is displayed, press Enter key.
Parameter import
With this function you will be able to import the instruments setup from PA-
RIMP.TXT file on SD-Card.
-- When the function Parameter import is displayed, press Enter key:
Photometer calibration
The HumaClot Junior is equipped with a 1 channel / 1 wavelength photome-
ter. This photometer is calibrated by factory. With this function it is possible to
re-calibrate the photometer.
-- When the function Photometer calibration is displayed, press Enter key:
-- Remove cuvette from measuring channel and Close Light Protection Cap.
Also Keep Light protection cap closed while processing.
-- Press Enter key to calibrate and the ESC key to leave this function.
OR
-- Press ESC key when this message below is displayed:
34
-- Key in „Superior Security Code“ (35001) and you will get an access to the
utilities menus.
-- Press > key until the function < PIN Code > is displayed and confirm it with
the Enter key.
Table 19
PIN Codes Explanation
DEFAULT
11111
Factory default PIN Code
Disables Requests for PIN Code for configurations in
00000
UTILITIES menu and test applications menu.
The range between 00001 and 59999 is allowed. The
00001 to 59999
code has always to be 5 digits.
ONLY FOR SERVICE
35001 Fixed „Superior Security Code“ in case PIN code was for-
gotten.
Important:
The „Superior Security Code“ must be kept confidential and should not be made
accessible to third parties.
3 ADJUSTMENTS
Check Softwareversion
Calibrate photometer
Check Mixer spinning
Check Parameter-ID
Parameter update
Software update
Keyboard test
Display test
Mixer test
Optic test
Replacement:
LCD-Module X
PCB Mainboard X X X X X X (X) X (X) X X (X) X X X X
Power Supply
Keyboard Overlay X
Display Cable X X
Thermal Block X X X
Thermal Switch
Mixer Motor X X X
Battery X
Housing X X X X X X (X) X (X) X X (X) X X X X
Light Protection Cap
Software / Parameter
Software Update (X) (X) X X (X) X (X)
Parameter Update X (X) X (X) X X (X)
Default Param. Loaded (X) (X) X X
38
Remarks:
Adjustment:
-- From the back side of the HumaClot Junior you will notice a potentiometer
inside a small hole.
-- Use a small screwdriver for adjusting the LCD-contrast.
Figure 2
Adjusting the LCD MODULE
Table 21
Space between magnet
and speed sensor
≤ 3mm Speed sensor malfunction
3-4mm Optimal distance
≥ 4mm Speed sensor malfunction
Figure 3
Adjusting the Hall sensor
-- Prepare a new cuvette with 200µl distilled water. ( It is required to use a new
cuvette with a new mixer to get an objective visual picture of the spinning
mixer).
-- Make sure that a new mixer is inside.
-- Make sure that there are no air bubbles inside.
-- Start Analyzer in service mode.
-- Select <mixer >.
-- Start motor with 200 rpm.
-- Put prepared cuvette into measuring channel.
-- Make a visual check:
1. Does the mixer spin in the centre of the cuvette.
2.
3. *
Does the spinning mixer seems to be a star like ( ).
In case it seems to be a triangle like () then check or replace the mag-
net / mixer motor unit.
Figure 4
Prefill postion(1) of inctubation
block with distilled water
-- Place external temperature sensor (probe) in the water filled position (1).
Figure 5
An External Probe is placed in
position (1)
-- Switch ON the Analyzer and wait until the Analyzer warms up.
-- When STANDBY is displayed that’s mean that the system has warmed up
successfully.
-- Read the value from the external Thermometer. If thermal block tempera-
ture is out of range refer to “thermal block calibration”.
Figure 6
-- When check process is completed, Remove the probe first then remove the
distilled water from incubation position.
Figure 7
Figure 8
44
Figure 9
-- Switch on the Analyzer and key in the service code while the text line is
scrolling.
-- Press the arrow keys until < calib. temperature > is displayed:
Calibration procedures:
1. Wait until temperature has stabilized to 37.4 and it will be displayed in
intern field.
4. Wait again until the temperature has stabilized to 37, 4 in the intern line.
5. If the values deviate repeat the procedure from step 2, till the both values
are the same then you will be able to finish the calibration and save it by
pressing the ESC key.
6. After pressing ESC key following display will appear:
10. Thermal block calibration has finished now, remove probe and remove the
distilled water from the incubation position.
46
Figure 10
Hardware Requirements:
-- SD-Card:
• Any SD-Card more than 32GB won’t work; also any SD-XD won’t work.
• The Acceptable Storage capacity for the SD-card is (128MB, 256MB,
512MB, 1 GB, 2 GB, 4 GB, 8 GB, 16 GB, and 32 GB).
-- CARD-Reading unit for PC
-- Desktop PC or Notebook if SD-Card is not prepared
-- Digital Photo Camera if SD-Card is not prepared
1. Formatting SD-Card:
For HumaClot Junior it is important to prepare the SD-Card at first.
Make sure that the file system is FAT or FAT32, Other file systems as NTFS
and others will not be accepted by the Analyzer. To format an SD-Card we
suggest using special software called SD Formatter which is provided by SD
Association www.sdcard.org or a digital photo camera. When the SD-cards
are formatted directly under windows or Linux operating systems will not be
accepted by the Analyzer.
For software update a special file- and folder structure is required, also it is
not allowed to have more than one pcs of *.MX file in CD12BOOT folder
at the same time.please follow the following procedures to update the soft-
ware:
Table 23
Step Action Folder-/ Filename Target
1 Create folder CD12BOOT SD-Card <ROOT>-Directory
2 File copy HC1_01v00.MX SD-Card CD12BOOT folder
Figure 11
-- For updating the 1-Channel, select and copy the folder (HC1_01v00.mx) into
the folder (CD12BOOT).
Figure 12
-- Pick the prepared SD-Card for software update; make sure that the SD-Card
is placed on the “lock” position. This prevents any information on the card
from being altered or deleted.
Figure 13
-- The write protection will enable the Boot-Loader routine in the Analyzer
software.
Figure 14
Figure 15
Figure 16
-- Insert SD-Card completely and a “click” tone will be heard when the SD-card
is inserted completely.
Figure 17
-- Turn the Analyzer back to original position, and turn on the Analyzer by plug-
ging the power supply.
Figure 18
Known Problems:
Hardware Requirements:
-- SD-Card:
• Any SD-Card more than 32GB won’t work; also any SD-XD won’t work.
• The Acceptable Storage capacity for the SD-card is (128MB, 256MB,
512MB, 1 GB, 2 GB, 4 GB, 8 GB, 16 GB, and 32 GB).
-- CARD-Reading unit for PC
-- Desktop PC or Notebook if SD-Card is not prepared
-- Digital Photo Camera if SD-Card is not prepared
1. Formatting SD-Card:
For HumaClot Junior it is important to prepare the SD-Card at first, make
sure that the file system is FAT or FAT32.Other file systems as NTFS and oth-
ers will not be accepted by the Analyzer. To format a SD-Card we suggest us-
ing special software called SD Formatter which is provided by SD Association
www.sdcard.org or a digital photo camera.
Figure 19
-- Copy the parameter file (PARIMP.TXT) and paste in the previous folder
(CD12DATA).
Figure 20
-- Pick the prepared SD-Card for software update; make sure that the SD-Card
is placed on the “lock” position. This prevents any information on the card
from being altered or deleted.
Figure 21
-- The write protection will enable the Boot-Loader routine in the Analyzer
software.
Figure 22
Figure 23
Figure 24
-- Insert SD-Card completely and a “click” tone will be heard when the SD-card
is inserted completely.
Figure 25
-- Turn the Analyzer back to original position, and turn on the Analyzer by plug-
ging the power supply.
56
Figure 26
-- Switch on the Analyzer and wait until the start up procedure has passed.
-- By pressing the cursor key, select < Import <- SD-Card > by pressing Enter key.
-- Parameters have been imported, press Esc key to save the new parameters.
-- Remove SD-Card.
-- Printout parameter-protocol “global”.
-- Check Parameter-ID.
-- Check thermalblock temperature.
-- Check calibration (conversion) parameters.
58
Known Problems:
Table 27
Problem Possible solution
Display message
BOOT-Loader V1.00
No application or firmware -- Reset write protection of SD-Card (switch)
update on the memory
card! Remove card.
-- Check folder structure on SD-Card
-- Check folder name
the folder name must written exactly as in-
Display message structed
Couldn´t open ´parimp.txt´ -- Check file structure on SD-Card
-- Check file system FAT on SD-Card
format SD-Card with different photo camera
model.
Display message
-- No valid parameter in line x
WARNING Line : ´xxxxxx´
-- Press Enter to over jump message and proceed
ignored
-- Check parameters after import procedure
ENTER= continue
Wrong parameter version
-- Check installation parameter file on SD-Card
installed
-- Refer to special instruction for parameter up-
date
All other cases
possibly special actions are requested.
Check of additional documentation.
Hardware Requirements:
-- SD-Card:
• Any SD-Card more than 32GB won’t work; also any SD-XD won’t work.
• The Acceptable Storage capacity for the SD-card is (128MB, 256MB,
512MB, 1 GB, 2 GB, 4 GB, 8 GB, 16 GB, and 32 GB).
-- CARD-Reading unit for PC
-- Desktop PC or Notebook if SD-Card is not prepared
-- Digital Photo Camera if SD-Card is not prepared
1. Formatting SD-Card:
For HumaClot Junior it is important to prepare the SD-Card at first, make
sure that the file system is FAT or FAT32.Other file systems as NTFS and oth-
ers will not be accepted by the Analyzer. To format a SD-Card we suggest us-
ing special software called SD Formatter which is provided by SD Association
www.sdcard.org or a digital photo camera.
Table 29
Step Action Folder-/ Filename Target
1 Create folder CD12DATA SD-Card <ROOT>-Directory
2 File copy PARIMP.AUT SD-Card CD12DATA folder
Figure 27
-- Copy the parameter file (PARIMP.TXT) and paste in the previous folder (CD-
12DATA).
Figure 28
-- Pick the prepared SD-Card for software update; make sure that the SD-Card
is placed on the “lock” position. This prevents any information on the card
from being altered or deleted.
Figure 29
-- The write protection will enable the Boot-Loader routine in the Analyzer
software.
Figure 30
Figure 31
Figure 32
-- Insert SD-Card completely and a “click” tone will be heard when the SD-card
is inserted completely.
62
Figure 33
-- Turn the Analyzer back to original position, and turn on the Analyzer by plug-
ging the power supply.
Figure 34
-- When this message below is displayed, that’s mean that the parameter have
be installed:
-- Remove SD-Card.
-- Remove PARIMP.AUT file from SD-Card.
-- Printout parameter-protocol “global”.
-- Check Parameter-ID.
-- Check thermal block temperature.
-- Check calibration (conversion) parameters.
Known Problems:
Automatic parameter
import every startup
After switching on Ana- -- Remove SD-Card
lyzer a parameter import -- Remove PARIMP.AUT file from SD-Card
starts automatically
Hardware Requirements:
-- SD-Card:
• Any SD-Card more than 32GB won’t work; also any SD-XD won’t work.
• The Acceptable Storage capacity for the SD-card is (128MB, 256MB,
512MB, 1 GB, 2 GB, 4 GB, 8 GB, 16 GB, and 32 GB).
-- CARD-Reading unit for PC
-- Desktop PC or Notebook if SD-Card is not prepared
-- Digital Photo Camera if SD-Card is not prepared
1. Formatting SD-Card:
For HumaClot Junior it is important to prepare the SD-Card at first, make
sure that the file system is FAT or FAT32.Other file systems as NTFS and oth-
ers will not be accepted by the Analyzer. To format a SD-Card we suggest us-
ing special software called SD Formatter which is provided by SD Association
www.sdcard.org or a digital photo camera.
Figure 35
66
-- Pick the prepared SD-Card for software update; make sure that the SD-Card
is placed on the “lock” position. This prevents any information on the card
from being altered or deleted.
Figure 36
-- The write protection will enable the Boot-Loader routine in the Analyzer
software.
Figure 37
Figure 38
Figure 39
-- Insert SD-Card completely and a “click” tone will be heard when the SD-card
is inserted completely.
Figure 40
-- Turn the Analyzer back to original position, and turn on the Analyzer by plug-
ging the power supply.
Figure 41
-- Switch on the Analyzer and wait unit the start up procedure has passed.
68
-- When < Export -> SD-Card > is displayed press Enter key:
Known Problems:
-- This below message will be displayed, press Enter key (“weiter”: means con-
tinue):
-- When this message below is displayed, press Enter key (“weiter” means con-
tinue):
Known Problems:
1. Global
2. Actual method
3. All Parameter
72
Printout global
-- Switch ON the Analyzer.
-- Wait until the start procedures ends.
-- By pressing the arrow keys, select < Parameter-Protocol > by pressing Enter
key.
-- Press Enter key when < Global > is displayed and printout will start right now
(See example 1 in the following pages):
Figure 42
-- By pressing the arrow keys, select < Parameter-Protocol > by pressing Enter
key.
-- Press Enter key when < Actual Method > and printout will start right now
(see example 2 in the following pages )
76
Figure 43
78
Mixer Setup:
Figure 44
Figure 45
-- adjust is fixed
pre-set by 2 sec.
All Parameters
Printout all parameters:
-- Switch ON the Analyzer.
-- Wait until the start procedures ends.
-- Select < Paramete-rProtocol > by pressing the arrow keys, and confirm it with
Enter key.
-- Press Enter key when < All Parameter > and printout will start right now (See
Example 3 in the following pages).
Figure 49
82
Figure 50
DPU-414 printer
Follow the following procedures to connect the HumaClot Junior with an exter-
nal printer:
-- Connect the printer cable to the serial port connector of the Analyzer.
Figure 51
Figure 52
Setup:
Port A: <ext. printer>
Port B: <Off
-- Select < Auto > by pressing the arrow keys, and confirm it with Enter key.
84
-- By pressing the arrow keys, select < ext. Printer > and confirm it with Enter
key.
-- For changing the setting of the Dip Switches, please refer to the printer’s
user manual.
Figure 53
86
Set the status of a Dip switch to “ON” by pressing ONLINE button or to “OFF” by
pressing FEED button as indicated in the tables below. After the ONLINE button
has been pressed the GREEN lamp indicates that the button operation has been
confirmed. When the FEED button has been pressed, the button operation is
confirmed by the RED lamp.
Alternately blinking of both lamps indicates that the new settings are written
to memory.
WARNING: Never turn the printer OFF while it is writing new settings to mem-
ory.
After each Dip switch setting via button, the status of the actual Dip switch is
printed out automatically. Wait until “DIP SW setting complete!!” is printed out
and the printer returns to the ONLINE mode before turn off the power.
Procedure
DIP SW -1
DIP SW -2
8 (OFF): = USA
Continue?:
23 Press FEED button GREEN lamp lights Push “On-line-SW”
Write?: Push
“Paper feed SW”
24 Press ONLINE button.
Move to chart Dip SW-3
25
Next page
DIP SW -3
Table 37
Step Action Lamps Printer prints:
With Dip Switches-3 show-
ing on the printer paper
1 (ON): Data Length
26 Press ONLINE button. RED lamp lights
= 8 bits
2 (ON): Parity Set-
27 Press ONLINE button. RED lamp lights
ting = No
3 (ON): Parity Con-
28 Press ONLINE button. RED lamp lights
dition = odd
4 (ON): Busy Con-
29 Press ONLINE button. RED lamp lights
trol = XON/XOFF
30 Press FEED button GREEN lamp lights 5 (OFF): Baud
31 Press ONLINE button. RED lamp lights 6 (ON): Rate
32 Press ONLINE button. RED lamp lights 7 (ON): Select
8 (ON): = 9600 bps
Continue ?: Push
“On-line-SW”
33 Press ONLINE button. RED lamp lights
Write ?: Push “Pa-
per feed SW”
Figure 54
Right view of HumaClot junior
-- The CuvetteCARD reading unit is placed on the right side of the Analyzer.
Figure 55
-- Insert the chip end of the CuvetteCARD into the Card reader Port.
Figure 56
90
-- When the CuvetteCARD is fully inserted into the Analyzer, it will identify a
valid cuvette card by:
• Showing a valid cuvette credit bookable : > 500 <
• Entering a quantity at debit: > 500 <
-- By pressing the Enter key the number of cuvettes will be loaded to the Hu-
maClot Junior, but we suggest loading fully 500 cuvettes in one step to the
Analyzer.
-- Wait until this below message is displayed (it may take up to 20 seconds).
-- When this message below is displayed, remove the CuvetteCARD from the
reading unit. (Keep CuvetteCARD and save it if there is a credit of cuvettes
left on the card.)
-- After loading the cuvettes to the HumaClot Junior an automatic printout will
appear when the printer is connected.
========================
========================
CuvCARD balance
loaded to
HumaClot Junior
V01.00
SerNo. M1360008
Date/Time:
17.10.2017, 08:22:27
-- CuvCARD Info --
Lot Number =’12345A67’
Balance = 500
CuvCARD devalued.
========================
========================
Known Problems:
-- When STANDBY is displayed that’s mean that the system has warmed up
successfully and ready to use.
-- Select < photometer calibr > by pressing the arrow keys, and confirm it with
Enter key.
-- When this message below is displayed remove the Cuvette from the meas-
uring channel,close the light protection cap and keep the cap closed while
processing then press Enter key:
------------------------
channel 1
------------------------
--- wavelength 405nm:
previous calibr. values
LED = 160
m-bright = 56518
r-bright = 42079
m-dark = 0
r-dark = 0
ext. gain = OFF
-- When this message below is displayed that’s mean that the new photome-
ter calibration have been saved.
Known Problems:
Table 39
Problem Possible solution
Display shows: Preparation procedure has not been followed
PHOTOMETER CALIBRAT. correctly.
WARM UP 10:27 -- Start process from beginning
Calibration procedure failed.
-- Keep printout saved
Display shows:
-- Check for cuvettes in the measuring channel
Bright calib failed!
-- Remove cuvette from meas. Channel
405nm LED ENERGY LOW
-- Retry calibration process by pressing Enter key
-- Contact Human s service
Calibration procedure failed.
-- Keep printout saved
Display shows:
-- Check for dust/dirt in the measuring channel
Bright calib failed!
-- Clean measuring channel
405nm LED ENERGY LOW
-- Retry calibration process by pressing Enter key
-- Contact Human s service
Known Problems:
Table 40
Problem Possible solution
Problems with STM Driver installation.
At Windows the USB is Unzip the file, run the executable, and then go to
detected as C:\Program Files (x86)\STMicroelectronics\Soft-
STM32 Virtual ComPort in ware\Virtua¬l comport driver in Windows Ex-
FS Mode plorer and double-click either dpinst_amd64.exe
for 64 bit systems, or dpinst_x86.exe for 32 bit.
4 INSTRUMENT ERRORS
Table 45
Problem Possible solution
-- Check if LIS system is ONLINE
-- Check cable connection
No LIS data output
-- Check Virtual ComPort driver on computer
Results should be send
-- Check Analyzer setup
when clotting is found
Port B = Host
-- Check correct setup of LIS System
-- Check Port B interface (service menu)
Problem /
Table 47 Possible solution
Display message
Parameter error!
Set to defaults -- See chapter 3.9.4
Parameter Error detected -- Contact Human service
in EEPROM memory
ERROR: xxx -- Contact Human service
-- Contact Human service
ERROR recursive
-- Change PCB mainboard
Problem /
Table 48 Possible solution
Display message
break time -- Check for correct pipetting sample and reagent
The measuring time -- Check for existing clot formation inside
reached cuvette
“max. measuring time” -- Check for correct test application in setup
-- Check for clear fluid inside cuvette
-- Check for direct ambient light
break top
from lamp or direct sunlight
Exceeded measuring
-- Close light protection cap
range is too low
-- Change Analyzers placement
-- Check for air bubbles inside cuvette
Table 49
Problem /
Possible solution
Display message
break
Measurement cancelled
n/a
by user with ESC + Start
Key
bright value failed
-- Check for cuvette inside measuring channel
Measuring signal too low
remove cuvette inside measuring channel
while Photometer check
-- Check for a clean measuring chamber
procedure.
-- Check for foreign parts in bottom of measuring
Measuring signal too low
channel
while Photometer check
-- Check for direct ambient light
procedure.
from lamp or direct sunlight
Calibrated photometer
-- Close light protection cap
bright value could not be
-- Change Analyzers placement
found.
-- Check of dirt in measuring channel
-- Check of dirt on magnet (inside Analyzer)
-- check if parts at magnet block the motor
break motor
example: mixing-bar
Mixer Motor or motor
-- Check if motor is spinning in service mode
speed sensor is not
-- Check if speed sensor is detecting correctly
working properly.
-- Check distance magnet vs. sensor
-- Change motor
-- Change PCB mainboard
Table 50
Problem /
Possible solution
Display message
Self starting -- Repeat startup process, make sure there is no
measurement cuvette inside measuring channel, keep light
Incubation timer starts protection cap closed while processing
without cuvette -- Calibrate Photometer, see chapter 3.14
-- Check for direct ambient light
from lamp or direct sunlight
-- Close light protection cap
Self starting measurement
-- Change Analyzers placement
Incubation timer starts
-- Repeat startup process, make sure there is no
without cuvette
cuvette inside measuring channel, keep light
protection cap closed while processing
-- Calibrate Photometer, see chapter 3.14
Table 52
Problem /
Possible solution
Display message
-- Check conversion parameters
ERR over -- Check calibration curve
Calculation overflow -- Check application setup
-- Contact Human service
-- Check conversion parameters
ERR div0 -- Check calibration curve
Calculation error -- Check application setup
-- Contact Human service
-- Remove cuvette and check if a clot is visible
ERR no clot inside cuvette
No clotting detected dur- -- Check application setup
ing measurement -- Repeat measurement
-- Contact Human service
-- Check conversion parameters
ERR log0 -- Check calibration curve
Calculation error -- Check application setup
-- Contact Human service
min-value reached- -- Check range of calibration
no more points -- Check conversion parameters
While input of calibration -- Check calibration curve
curve minimum value is -- Check application setup
reached. -- Contact Human service
max-value reached- -- Check range of calibration
no more points -- Check conversion parameters
While input of calibration -- Check calibration curve
curve maximum value is -- Check application setup
reached. -- Contact Human service
Problem /
Table 53
Possible solution
Display message
(Channel) locked
The measuring channel
is locked due to no re- -- Load new cuvette balance to the Analyzer
maining cuvette balance
available
CuvCARD read error -- Check for correct CuvetteCARD
Reading error while read- -- Try a different CuvetteCARD
ing CuvetteCARD -- Check Reading Unit for Chip- and CuvetteCards
Inserted card may be in service mode
defect. -- Replace PCB mainboard
Table 54
Problem /
Possible solution
Display message
Cuvette balance:
100 measurements
While start up procedure of -- Check for further cuvettes and Cuvette-
HumaClot Junior CARDs on stock.
the system indicates that -- Order new cuvettes
100 cuvettes or less
are available.
Warning -- Remove Card, press Enter key,
Wrong cuvette type -- Use matching cuvettes / CuvetteCARD
Wrong cuvette type -- Contact Human service
Warning
-- Remove Card
Invalid customer ID
-- Contact Human service
Wrong distributor ID code
Illegal card!
Please remove it -- Remove Card
Not certified Card or data on -- Contact Human service
Card
Empty/invalid card!
-- Remove Card
Empty or not certified Card or
-- Contact Human service
data on Card
System loads cuvettes Less than 100 cuvettes left on HumaClot
directly from card Junior. System loads new volume automat-
without showing ically from inserted CuvCARD without user
remaining balance communication
Balance fully used!
Only xx reserve measurm.
remain possible!
======================== -- Check for further cuvettes and Cuvette-
*** Please *** CARDs on stock.
*** order new *** -- Order new cuvettes
*** Cuvettes *** -- Proceed by loading new cuvette volume
========================
Cuvette balance on Analyzer
fully used.
No Card!
There is no ChipCARD inserted -- Insert ChipCARD into Card reading unit
in the Card reading unit.
-- Check for further cuvettes and Cuvette-
Balance and reserve fully used
There is no cuvette CARDs on stock.
balance on Analyzer. -- Load new cuvette balance from CuvCARD
-- Order new cuvettes
Problem / Table 55
Possible solution
Display message
-- Check for further cuvettes and Cuvette-
Balance fully used
CARDs on stock.
There is no cuvette balance
-- Load new cuvette balance from CuvCARD
on Analyzer.
-- Order new cuvettes
CuvCARD devaluated
-- Check for further cuvettes and Cuvette-
All cuvettes on CuvCARD
CARDs on stock.
are loaded to Analyzer. No
-- Order new cuvettes
cuvettes left on CuvCARD.
Only xx measurements remai-
ning.
-- Check for further cuvettes and Cuvette-
Warning:
CARDs on stock.
Cuvette balance on Analyzer
-- Order new cuvettes
is getting low. Only xx
measurements possible.
Please order new cuvettes
-- Check for further cuvettes and Cuvette-
Warning:
CARDs on stock.
Cuvette balance is getting
-- Order new cuvettes
low.
Only xx reserve measurements
-- Check for further cuvettes and Cuvette-
remain possible.
CARDs on stock.
Warning:
-- Load new cuvette balance from CuvCARD
Cuvette balance on Analyzer
-- Order new cuvettes
used up.
108
Table 56
Problem /
Possible solution
Display message
ChipCARD Checksum error
-- Check for correct ChipCARD
Checksum error while reading
-- Try a different ChipCARD
ChipCARD
-- Check for correct ChipCARD
ChipCARD read error
-- Try a different ChipCARD
Reading error while reading
-- Check Reading Unit for Chip- and Cuvette-
ChipCARD.
Cards in service mode
Inserted card may be defect.
-- Replace PCB mainboard
Warning: Please remove
ChipCard
-- Remove Card, press Enter key, retry as
Enter: continue
instructed
CuvetteCard inserted in
wrong direction
Warning: Please remove
ChipCard
-- Remove Card, press Enter key retry a
Enter: continue
different Card
CuvetteCard already
devaluated
Warning: Please remove
ChipCard
-- Remove Card, press Enter key, retry a
Enter: continue
CuvetteCARD
This card is not a
CuvetteCARD
Warning: incompatible me-
-- Remove Card, press Enter key
thod
-- Contact distributor
Enter: continue
-- Contact Human service
Incompatible Method CARD
No Card!
There is no CuvetteCARD
-- Insert CuvetteCARD into Card reading unit
inserted in the Card reading
unit.
Problem / Table 57
Possible solution
Display message
-- Reset write protection of SD-Card (switch)
SD-Card is write protected
-- Check for correct SD-Card type/size
Checksum error while reading
-- Check if SD-Card is formatted correctly
ChipCARD
-- Try a different SD-Card
BOOT-Loader V1.00
No application or firmware -- Reset write protection of SD-Card (switch)
update on the memory card! -- Remove SD-Card
Remove card.
-- Check folder structure on SD-Card
-- Check folder name
the folder name must written exactly as
instructed
Couldn´t open ´parimp.txt´
-- Check file structure on SD-Card
-- Check file system FAT on SD-Card
format SD-Card with different photo
camera model.
WARNING Line : ´xxxxxx´
ignored -- Press Enter to over jump message and
ENTER= continue proceed
No valid paramter in line xxx -- Check parameters after import procedure
of PARIMP.* file while param- -- Contact Human s service
eter import from SD-Card
-- Check for proper insertion of SD-Card in
reading unit.
-- Remove SD-Card from reading unit and
No SD-Card detected!
retry to insert SD-Card
SD-Card could not be detect-
-- Check file structure on SD-Card
ed in SD-Card reading unit.
-- Check file system FAT on SD-Card
format SD-Card with different photo
camera model.
-- Check for proper insertion of SD-Card in
No SD-Card found!
reading unit.
SD-Card could not be detect-
-- Remove SD-Card from reading unit and re-
ed in SD-Card reading unit.
try to insert SD-Card
110
Problem /
Table 58 Possible solution
Display message
-- Check for proper insertion of SD-Card in
reading unit
-- Remove SD-Card from reading unit and
retry to insert SD-Card
-- Check for correct SD-Card type/size
-- Check if SD-Card is formatted correctly
SD-Card: init error -- Try a different SD-Card
Initialisation Error while -- Check folder structure on SD-Card
writing data to SD-Card. -- Check folder name
the folder name must written exactly as
instructed
-- Check file structure on SD-Card
-- Check file system FAT on SD-Card
format SD-Card with different photo
camera model
Table 59
Problem /
Possible solution
Display message
-- Check folder structure on SD-Card
-- Check folder name
the folder name must written exactly as
Paramters are not imported instructed
from PARIMP.AUT file -- Check file structure on SD-Card
-- Check file system FAT on SD-Card
format SD-Card with different photo
camera model.
-- Check folder structure on SD-Card
-- Check folder name
the folder name must written exactly as
Paramters are not imported instructed
from PARIMP.DEL file -- Check file structure on SD-Card
-- Check file system FAT on SD-Card
format SD-Card with different photo cam-
era model.
-- Check installation parameter file on
Wrong parameter version
SD-Card
installed
-- See chapter 3.8
Table 60 Problem /
Possible solution
Display message
Invalid method param.
-- Check method parameters
Method parameters (Applica-
-- Contact Human s service
tion parameters) are incorrect
New name cancelled
-- Enter method name again
Method name was
-- Contact Human s service
entered incorrectly
Problem /
Possible solution
Table 61 Display message
-- Check setup in UTILITIES / Port B
set to OFF if no Host System available
-- Ignore message if no Host system availa-
ble BUT Patient ID is required.
HOST not connected
-- Check cable connection to Host System
Connection to HOST system
Port B, USB connection to PC
could not detected.
-- Check for correct virtual ComPort driver
on PC
-- Check if HOST/LIS application is properly
working on HOST computer
Problem / Table 62
Possible solution
Display message
-- Check for cuvette inside measuring
channel remove cuvette inside measuring
channel
-- Check for a clean measuring chamber
(Channel) locked
-- Check for foreign parts in bottom of meas-
Error during photometer
uring channel
check.
-- Check for direct ambient light from lamp
or direct sunlight
-- Close light protection cap
-- Change Analyzers placement
-- Check for cuvette inside measuring chan-
nel remove cuvette inside measuring chan-
nel
-- Check for a clean measuring chamber
ERR bright calib ch xx
-- Check for foreign parts in bottom of meas-
Error during photometer
uring channel
check.
-- Check for direct ambient light from lamp
or direct sunlight
-- Close light protection cap
-- Change Analyzers placement
-- Check for cuvette inside measuring
channel remove cuvette inside measuring
channel
-- Check for a clean measuring chamber
ERR dark calib ch xx
-- Check for foreign parts in bottom of
Error during photometer
measuring channel
check.
-- Check for direct ambient light from lamp
or direct sunlight
-- Close light protection cap
-- Change Analyzers placement
-- Wait until Analyzer has warmed up prop-
PHOTOMETER CALIBRAT.
erly.
WARM UP 10:27
-- Start process from beginning pay attention
Analyzer not ready for
to
calibration process.
-- See chapter 3.14
-- Keep printout saved
Display shows:
-- Check for cuvettes in the measuring channel
Bright calib failed!
-- Remove cuvette from meas. channel
405nm LED ENERGY LOW
Calibration procedure failed. -- Retry calibration process
-- Contact Human s service
114
Table 63
Problem /
Possible solution
transmitted message
E -- Check conversion parameters
Raw value [sec] [mE] -- Check calibration curve
[mE/Mn] faulty or range -- Check application setup
exceeded -- Contact Human service
-- Check if Analyzer is placed in direct sun-
light
-- Check if Analyzer is placed next to a heat-
ing
-- Check present thermal block temperature
with external thermometer
T
-- Recalibrate thermal block temperature
WARN: temp instable
-- Check thermal block connector to PCB
Mainboard
-- Check thermal block sensor (100KΩ at
25°C)
-- Change thermal block
-- Change PCB Mainboard
OK
Everything was OK, no errors n/a
or warnings appeared
Table 64 Problem /
Possible solution
transmitted message
< 1234 ** -- - Check conversion parameters
** example threshold value = -- - Check calibration curve
1234 -- - Check application setup
Calculated result < min value -- - Contact Human service
> 1234 ** -- Check conversion parameters
** example threshold value = -- Check calibration curve
1234 -- Check application setup
Calculated result > max value -- Contact Human service
-- Check conversion parameters
-- Check calibration curve
ERR_lin
-- Check application setup
-- Contact Human service
ERR_no_clot
channel 1 break time
-- Check conversion parameters
R_DIV0
-- Check calibration curve
Division by zero during calcu-
-- Check application setup
lation
-- Contact Human service
5 DISASSEMBLING / ASSEMBLING
-- Remove all cables to the Analyzer before opening the instrument.
-- Remove all vials from the thermal block.
-- Make sure that no cuvettes are in the measuring channels.
-- Make sure that no chip card is in the card reader.
-- Make sure that no SD-card is in the card reader.
-- Make sure to work on a clean and soft surface.
-- Collect all screws, clamps and clips in an appropriate container.
-- Make sure that no parts will disappear.
-- Immediately remove all parts which fell into the Analyzer by accident.
-- Leave no loose parts in the Analyzer.
-- During and upon completion of your work make sure that no loose parts are
in the Analyzer.
-- Make sure that no demounted parts will be left over after mounting the
instrument, except exchanged parts.
-- Make sure that exchanged modules have the proper program version. See
the information to the software version in chapter „Software status“.
Assembly
-- Assembling Light Protection Cap
-- Assembling Top- and Bottom
-- Assembling Top Housing w. Thermal Block & Mainboard
-- Assembling LCD-Module
-- Assembling Top Housing
-- Assembling Mainboard w. Thermal Block
118
Figure 57
Assembling Light Protection Cap
1
3
4
5 2
Figure 58
Assembling Top- and Bottom
1
Housing
7
6
Figure 59
1
Assembling Top Housing w.
2
Thermal Block & Mainboard
5
4
6
7
1 2 4
Figure 60
Assembling LCD-Module
6 HARDWARE COMPONENTS
HumaClot Junior consists of Printed circuit Boards, Keyboard Overlays, Housing,
Cables, Thermal Block and small parts, each component will be described in de-
tails in the following pages.
PCB Mainboard
Controls most of the hardware and the software functions
Figure 61
PCB Mainboard
Hardware functions:
-- Display control
-- Keyboard control
-- Control of measuring channel
1. Mixer-motor control
2. Light-source and –sensor control
Software functions:
-- Control of the analytical measurement range
-- Method selection
-- Calculation of the results
-- Output of the final results
124
LCD-Module
Figure 62
LCD-Module
Hardware function:
-- For a better communication with the HumaClot Junior everything will be
displayed on the LCD-module.
Power supply
Figure 63
Power Suppply
Hardware function:
-- Transform the mains voltage into small voltage
Figure 64
Hardware Function:
-- Responsible for the data Entries by pressing the keys to communicate with
the Analyzer.
Keyboard matrix
Figure 65
Figure 66
126
6.3 Housings
-- Housing unit
-- Display foil
-- Rubber feet
Housing Unit
Consist of top and bottom housing, left and right side housing
Figure 67
Housing Unit
Display foil
Figure 68
Dispaly foil
Specifications:
Rubber feet
Figure 69
6.4 CABLES
-- Cable for display
-- Data cable Mini-DIN / SUB-D
-- Data cable DPU-414
-- Data Cable for BC-Reader
-- USB Cable
-- Barcode Reader
-- RS232 TEST Plug
Figure 70
Cable for display
Figure 71
Data Cable Mini-DIN/SUB-D
Figure 72
Figure 73
Data Cable DPU-414
Figure 74
130
Figure 75
Data cable for BC-reader
Figure 76
USB CABLE
This cable is responsible for connecting the HumaClot Junior with the computer,
by connecting one of USB Ports with computer and the other with the Analyzer.
Figure 77
USB Cable
Barcode Reader
This cable is used to import the patient IDs into the HumaClot Junior.
Figure 78
Figure 79
RS232 Test plug
132
Figure 80
Y-CABLE
This cable is responsible for connecting the HumaClot Junior with the DPU-414
printer and the barcode reader at the same time.
Figure 81
Y- Cable
Figure 82
Specifications:
Comment:
Incl. heating foil
Incl. Temperature sensor
Incl. Temperature switch
Incl. Clamp
Without light protection caps and holders 115-93-150-00
Figure 84
Thermal Switch
Figure 86
Light Protection Cap
Specifications:
Contents:
Set of Fuses
These spare fuses are used for an electrical protection for the Analyzer and the
power supply.
Figure 87
Set of Fuses
136
Figure 88
Figure 89
Top view of PCB Mainboard
connector pin out
Figure 90
PCB MAIN BOARD FOR HumaClot
Junior - bottom view
Figure 91
Top view of the LCD-Module
138
Figure 92
Bottom view of the LCD-Module
Table 66
SD-Card connector (ST 1) Pin Assignment
1:CS
2:Data IN
3:GND
4:VCC
Figure 93 5:CLK
6:GND
7:Data Out
8:n.c.
9:n.c.
DET
WP
COM
1:+5VDC
2:GND
Figure 94 3:CTS
4:n.c.
5:TxD
6:RTS
7:n.c.
8:RXD
Pin Assignment
C1:VCC
C2:RES
C3:SCL
C4:n.c.
CARD reader (ST8) C5:GND
C6:VPP
C7:SDA
C8:n.c.
S1:SW
S2:SW
+ 12VDC
– GND
Figure 101
1: Heat +
2: Heat +
3: Heat -
4: Heat -
5: NTC +
6: NTC -
7 APPENDIX