Simplified User Guide - DeVilbiss Ultraneb

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

Simplified User Guide

Equipment Nebuliser, DeVilbiss Ultraneb (EHOENEBU2--)


Standard Night stand Wheeled
Configuration kit. stand kit
(without medication (EHOENEBU210) (EHOEZFR0688)
cup)
Use with medication Also for use with
cups is described in medication cups.
section B)

A General Use
Install the antibacterial filter in the correct flow direction,
1 on the output of the device. Never change the direction of 1
antibacterial filter.

Put in place the nebulisation chamber on the coaxial jack


2 and fill to the indicated level, then fit the lid.
Connect the shortest tube between the lid of the
nebulisation chamber and the antibacterial filter.

Connecting the patient tube (the longest) to the lid of the


3 nebulisation chamber and, if necessary, connect a 2
mouthpiece to the other end.

Connect the device to the mains power and switch it on


4 with the switch below the power connector on the rear
(the green light on the front should light).

Fill the nebulisation chamber with medication as specified


by the treatment protocol.

Adjust the nebulisation power (d), the air flow (e), and the 3
5 timer (f) according to the requirements of the treatment
protocol used.

(a) Green power on light 4


(b) Red alarm or timer finished light
(c) Stop/start push-button
(d) Nebulisation power adjustment
(e) Air flow adjustment
(f) Timer adjustment

MSF Biomed – Simplified User Guide – DeVilbiss UltraNeb v1, Mar 2014, page 1/2
B Use with medication cup
For use with medication cup you will need the steel column
(EHOEZFR0689) and the retention bar (EHOEZFR0690) to
hold down the cup.
1

Cup Steel column Retention bar

1 Install the steel column in the slot at the rear of the


nebuliser and clamp it well with the screw. 2
Install the antibacterial filter in the correct flow direction, on
2 the output of the device. Never change the direction of
antibacterial filter.
Put in place the nebulisation chamber on the coaxial jack
3 and fill to the indicated level (marked on the side of the cup)
with clean water.
Add to the cup the prescribed drug (eg salbutamol) in small
3
quantities, and diluted according to the medical protocol
and fit the transparent lid of the cup.
Fix the lid in place with the retention bar by sliding it down
the steel column, and fix it with the screw. Raccord le tuyau
le plus court entre le filtre et la coupelle.
Connect the shortest tube between the lid of the
4 nebulisation chamber and the antibacterial filter. Then
connect the longer tube to the lid of the nebulisation
chamber, if necessary, connect a mouthpiece to the other
end.
Connect the device to the mains power and switch it on with 4
the switch below the power connector on the rear (the
green light on the front should light).

Adjust the nebulisation power (d), the air flow (e), and the
5 timer (f) according to the requirements of the treatment
protocol used.

6 5

(a) Green power on light


(b) Red alarm or timer finished light
(c) Stop/start push-button
(d) Nebulisation power adjustment
(e) Air flow adjustment
(f) Timer adjustment

MSF Biomed – Simplified User Guide – DeVilbiss UltraNeb v1, Mar 2014, page 2/2