Professional Documents
Culture Documents
Forces - Medial Meniscal - 3D
Forces - Medial Meniscal - 3D
Forces - Medial Meniscal - 3D
PII: S0021-9290(22)00470-5
DOI: https://doi.org/10.1016/j.jbiomech.2022.111429
Reference: BM 111429
Please cite this article as: J.R. Brown, J.F. Hollenbeck, B.W. Fossum, H. Melugin, S. Tashman, A.F. Vidal, M.T.
Provencher, Direct Measurement of Three-Dimensional Forces at the Medial Meniscal Root: A Validation Study,
Journal of Biomechanics (2023), doi: https://doi.org/10.1016/j.jbiomech.2022.111429
This is a PDF file of an article that has undergone enhancements after acceptance, such as the addition of a cover
page and metadata, and formatting for readability, but it is not yet the definitive version of record. This version
will undergo additional copyediting, typesetting and review before it is published in its final form, but we are
providing this version to give early visibility of the article. Please note that, during the production process, errors
may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
1
Direct Measurement of Three-Dimensional Forces at the Posterior Medial Meniscal Root: A Validation Study
2
Abstract
The posterior medial meniscal root (PMMR) experiences variable and multiaxial forces during loading.
Current methods to measure these forces are limited and fail to adequately characterize the loads in all three
dimensions at the root. Our novel technique resolved these limitations with the installation of a 3-axis sensing
construct that we hypothesized would not affect contact mechanics, would not impart extraneous loads onto the
PMMR, would accurately measure forces, and would not deflect under joint loads. Six cadaveric specimens were
dissected to the joint capsule and a sagittal-plane, femoral condyle osteotomy was performed to gain access to the
root. The load sensor was placed below the PMMR and was validated across four tests. The contact mechanics test
demonstrated a contact area precision of 44 mm2 and a contact pressure precision of 5.0 MPa between the pre-
installation and post-installation states. The tibial displacement test indicated an average bone plug displacement of
< 1mm in all directions. The load validation test exhibited average precision values of 0.7 N in compression, 0.5 N
in tension, 0.3 N in anterior-posterior shear, and 0.3 N in medial-lateral shear load. The bone plug deflection test
confirmed <2 mm of displacement in any direction when placed under a load. This is the first study to successfully
validate a technique for measuring both magnitude and direction of forces experienced at the PMMR. This validated
method has applications for improving surgical repair techniques and developing safer rehabilitation and
3
INTRODUCTION
Meniscal tears are the most common indication for orthopedic surgery. As many as 1% of active people
suffer a meniscal injury, with the medial side being affected three times as often as the lateral (Gee et al., 2020;
Wilson et al., 2018). The outcomes of posterior medial meniscal root (PMMR) tears are well documented, although
the forces experienced at the root are only understood at a rudimentary level (Daney et al., 2019). The ability to
comprehensively characterize forces at the PMMR can guide the position of surgical techniques to withstand the
magnitude and direction of stress that results in injury. This also applies to making safer rehabilitation and
Meniscal root tears occur due to a variety of factors (Bahr et al., 2005). Tears are typically seen during
sudden movements at high flexion angles while weightbearing, such as descending stairs or pivoting during athletics
(Furumatsu et al., 2019; Song et al., 2017; Wang et al., 2021). Along with knee position, internal factors such as
advanced age, BMI, and varus alignment are all significant risk factors for PMMR tears (Hwang et al., 2012).
Higher flexion angles generate greater postero-medial pressure, leading many groups to hypothesize that PMMR
tears are caused by elevated compression and shear forces at the root (Habata et al., 2004; Vedi et al., 1999).
However, many of these hypotheses remain speculative due to the inability to measure PMMR loads in more than
one direction.
Previous methods have successfully estimated meniscal root force under various joint loading conditions.
However, the majority of these findings have focused on recording tension force leaving shear and compressive
forces to be discovered in future studies (Markolf et al., 2012; Stärke et al., 2013; Seitz 2012; Seitz et al., 2021).
Other methods have generated computational models to estimate three-dimensional (3D) root forces further
expanding the ways to measure root forces (Guess et al., 2017; Yao et al., 2006). Freutel et al. measured 3D forces
in a porcine meniscal root model making cadaveric specimens the next setting to record these forces (Freutel et al.,
2015). Fujie et al. used a cadaveric model that paired a force transformation scheme with an external sensor to take
the field another step closer to determining comprehensive forces in ligaments (Fujie et al., 1995). The current body
of literature surrounding root force measurements has laid the groundwork for an intact cadaveric model utilizing
the latest sensor technology to finally obtain comprehensive force measurements at the meniscal root.
Utilizing an established 3-axis force sensor (TR3D-B-250 Three Axis Load Cell, Michigan Scientific
Corporation, Charlevoix, MI), we have developed a novel method to measure the 3D magnitude and direction of
4
force transmitted to the PMMR in an intact knee joint. The objective of this study was to validate the accuracy of
this method for capturing PMMR loading conditions without influencing the native biomechanics of the knee joint.
We hypothesized that installation of the 3-axis sensing construct would not significantly affect the contact area of
the knee joint, would not significantly affect the average contact pressure of the knee joint, would not impart
significant extraneous loads onto the PMMR, would accurately measure forces applied to the PMMR, and would not
significantly deflect under joint loads. Successful validation has potential applications for tendon or ligament studies
that can guide clinicians towards surgical repairs better positioned to decrease forces and help develop safer
5
Methods
Specimen Preparation
Six male, fresh-frozen cadaveric knees (mid femur to mid tibia-fibula) with a mean age of 59.3 (Range: 53-
65 years) and no history of arthritis, previous surgery, or malignancy were used for this study. The skin,
subcutaneous tissue, and muscles down to the joint capsule were carefully dissected and removed, leaving all
ligaments intact (Figure 1). Each specimen was screened for meniscal pathology under direct visualization. The tibia
and fibula were potted in a cylindrical mold with polymethyl methacrylate (PMMA, Fricke Dental, Streamwood, IL)
15 cm distal to the joint line. Two holes were drilled into the femur in order to mount the specimen to a custom
fixture capable of changing flexion angles (fixture seen in Figure 2). Next, a medial condyle osteotomy was
performed with a micro-oscillating saw and k-wires for complete visualization of the medial meniscus (Figure 1B).
The osteotomy was secured with a removable steel plate and bicortical screws (Figure 1).
Sensor Installation
A high-grade stainless steel, environmentally sealed three-axis sensor (25.4 mm x 25.4 mm x 12.7 mm)
was acquired from Michigan Scientific Corporation that determines force measurements in three orthogonal
directions. The sensor consists of 3 four-arm strain gauge bridges, each with a full-scale output of 3.5 mV/V and
maximum load of 1.1 kN (250 lbs). Sensor non-linearity and hysteresis is reported to be less than 0.5% of full-scale
output. Sensor calibration was performed within one month of experimental testing. Manufacturer sensor calibration
was performed within one month of experimental testing. The sensor construct (Figure 1A) consisted of the 3-axis
load cell, a directional mounting bracket, and plastic potting cup. A section of bone was removed from the posterior
tibia via oscillating saw 2 cm inferior to the PMMR, maintaining 2 mm margins (Figure 1C). A bone plug was cut
out inferior to the PMMR via k-wire, carefully protecting and excluding the meniscus and footprint of the cruciate
ligaments protected by army navy retractors. This plug was made with 20⁰ of medial angulation in relation to the
tibial plateau to replicate the insertion angle of the meniscal root fibers (Haut Donahue et al., 2022). PMMA potting
material was used to cement the bone plug to the sensor and then secured by three converging screws. Once
preliminary testing for surface biomechanics concluded, final cuts were made to separate the bone plug from the
Force Transformation
6
Two coordinate systems were defined to standardize 3D forces to a tibial coordinate system across
specimens using a three-dimensional coordinate digitizer (Romer Absolute Arm, Hexagon Manufacturing
Intelligence, Cobham, Surrey, Great Britain). One system was positioned and oriented at the load cell, and the other
to the tibial plateau (Figure 3). A transformation matrix was calculated between the two coordinate systems such
that the forces measured in the sensor coordinate system could be transformed and reported with respect to the tibial
Four validation tests were performed: a contact mechanics test, a tibial plateau displacement test, a load validation
test, and a bone plug deflection test. Together, these tests ensured that the proposed method for measuring 3D forces
at the PMMR is accurate and does not affect the native biomechanics of the knee joint.
This test evaluated the hypothesis that the installation of the 3-axis sensing construct (independent variable) would
insignificantly affect the contact area and average contact pressure of the knee joint (dependent variables). Each
fully dissected specimen was measured in two states: before sensor installation and after. The specimen was
mounted to a materials testing system (Instron Electropuls 10,000, Norwood, MA, USA) via a custom fixture that
oriented the specimen so that each compartment was loaded evenly and applied load to the femoral epicondyles.
Equilibrated electronic pressure sensors (K-scan Model 4000, 1500 psi; Tekscan Inc, South Boston, MA) were
calibrated by performing Tekscan-defined calibrations at 20% of the expected maximum pressure (188.5 N) and
80% of the expected maximum pressure (754 N). At each loading value, cork (3 mm thickness) cylinder with a
contacting surface area of 314 mm2 and capped with a layer of silicone 30 was applied to the pressure sensor, and a
power calibration was performed per manufacturer’s specifications. The pressure-mapping sensors were installed
under the meniscus in the lateral and medial compartments, and the position of the sensors relative to the tibia was
marked. The specimen was loaded to 1000 N, and the total contact area, medial compartment contact pressure, and
lateral compartment pressure for the pre-installation state were recorded (Figure 2). The specimen was unloaded,
and the pressure-mapping sensor in the medial compartment was removed to install the 3-axis load cell via the steps
described above. Final cuts on the tibial bone plug were then performed. The pressure mapping sensors were then
reinstalled, and the loading protocol and pressure measurements were repeated for the post-installation state. Bias,
precision, and intra-class correlation coefficients (ICCs) were calculated from all outcome measures to compare the
7
pre-installation and post-installation testing states (Koo and Li 2015). Bias for contact area and average contact
pressure was calculated as the mean of the differences between states (Equation 1). Precision was calculated by first
squaring the difference between the pre-installation state with the bias subtracted and the post-installation state with
the bias subtracted. Then, the square root of the mean of these squared differences was calculated to obtain a
(∑ 𝑛
𝑥
𝑖 = 1 𝑝𝑟𝑒𝑖
― 𝑥𝑝𝑜𝑠𝑡𝑖 )
(1) 𝐵𝑖𝑎𝑠, 𝐵 = 𝑛, for 𝑖 : length(𝑛) where 𝑛 is the number of specimens and 𝑥 is the
(2) 𝑃𝑟𝑒𝑐𝑖𝑠𝑖𝑜𝑛 =
(∑ ((𝑥
𝑛
𝑖=1 𝑝𝑟𝑒𝑖 ― 𝑥𝑝𝑜𝑠𝑡𝑖) ― 𝐵)
2
) for 𝑖 : length(𝑛) where 𝑛 is the number of specimens and 𝑥
𝑛
is the outcome of the dependent variable (average contact pressure or contact area).
A custom-MATLAB script was used to calculate intraclass correlation coefficients based on a two-way mixed
effects model, using single measures, and absolute agreement as outlined by Koo and Li (2015). Agreement between
pre-installation state and post-installation state, based on the 95% confidence intervals of the ICC estimates, was
This test determined how the installation of the 3-axis sensing construct (independent variable) affected the position
Each specimen was recorded in two states: before sensor installation and after. Prior to sensor installation, the
locations of seven fiducial markers on the tibia and five fiducial markers on the bone plug were determined using a
3D digitizer (Romer Absolute Arm, Hexagon Manufacturing Intelligence, Cobham, Surrey, Great Britain) (Figure
4). These measurements were repeated following final installation of the sensor system. The displacement and
change in orientation of the bone plug from pre to post sensor installation was determined by calculating the
transformation matrix between the bone plugs in each state relative to the tibial location (Ellman et al., 2014).
This test served as a verification test of the 3-axis load cell and evaluated whether the applied load (independent
variable) was equivalent to the load output from the 3-axis load cell (dependent variable). The bone plug was
removed from the specimen by excising the medial meniscus from the tibia and unscrewing the bracket from the
tibia. The bone plug specimen was mounted to a materials testing system (Instron) via a custom fixture that oriented
8
the end effector of the tensile testing machine to the sensor’s coordinate system (Figure 5). Three-dimensional loads
from the sensor were recorded as the bone plug was loaded in compression (0 to 50 N in 10-N increments), in
medial-lateral (ML) shear (0 to 15 N in 5-N increments) (Figure 5). Loads were selected based on previous studies
in addition to what forces this group was seeing in piloting (Markolf et al., 2012; Stärke et al., 2013; Seitz 2012;
Seitz et al., 2021). The percent difference, bias, precision, and ICCs were calculated to compare the measured loads
to the applied loads and each applied loading state (e.g. 15 N of AP shear force) was calculated to compare the
measured loads to the applied loads. (Bias and precision were calculated with the same equations and methodology
Bias and precision were calculated for each load value (e.g. 0 N, 10 N, 20N, … 50 N) for each loading direction
(compression, tension, anterior-posterior shear, medial-lateral shear). A mean bias and mean precision were
calculated across all applied load values for each loading direction, and box-and-whisker plots were generated to
visualize the distribution of bias and precision values across load values for each loading direction (Figure 7). ICCs
were calculated in the same way as specified in the Contact Mechanics Test.
This test determined how shear load applied to the cortical layer of the bone plug (independent variable) affected
deflection in the bone plug (dependent variable). The specimen was mounted to a materials testing system via a
custom fixture that oriented the specimen in its neutral alignment with symmetric loads on the joint. A 1000 N
compression load was applied to the femoral epicondyles along the anatomical axis, and shear loads at the meniscal
root were recorded. Once removed, the bone plug was mounted to a materials testing system via a custom fixture
that oriented the end effector of the tensile testing machine to the sensor’s coordinate system. End effector
displacement was recorded as the shear loads, recorded earlier, were applied to the bone plug at the root attachment
site (Figure 5). As the gap distance between bone plug and tibia was standardized to 2 mm during installation, the
criterion for success was defined as an end effector displacement of less than or equal to 2 mm at maximal shear
load force.
9
Results
The average difference in contact area from pre to post sensor installation was 2 mm2 (-93 mm2 to 97 mm2 95%
confidence interval) for bias and 44 mm2 (13 mm2 to 63 mm2 95% confidence interval) for precision (Figure 6). An
ICC of 0.99 indicated strong agreement in contact area between the two testing states (Figure 6). In the lateral
compartment the average difference in contact pressure from pre to post sensor installation was -10.9 MPa (-40.0
MPa to 18.2 MPa 95% confidence interval) for bias and 13.6 MPa (2.7 MPa to 19.6 MPa 95% confidence interval)
for precision with an ICC of 0.97 (Figure 6). In the medial compartment the average difference in contact pressure
from pre to post sensor installation was -5.5 MPa for bias (-16.9 MPa to 27.9 MPa 95% confidence interval) and
10.4 MPa (0.9 MPa to 15.3 MPa 95% confidence interval) for precision with an ICC of 0.98 (Figure 6).
The average difference between the applied and the measured compression load was 0.5 N for bias (-0.3 N to 1.4 N
95% confidence interval) and 0.7 N (-0.2 N to 1.6 N 95% confidence interval) for precision (Table 1, Figure 7). The
average difference between the applied and the measured tensile load was -0.1 N for bias (-0.3 N to 0.1 N 95%
confidence interval) and 0.5 N (-0.4 N to 1.3 N 95% confidence interval) for precision (Figure 7). The average
difference between the applied and the measured AP shear load was -0.2 N for bias (-0.8 N to 0.3 N 95% confidence
interval) and 0.3 N (-0.8 N to 1.5 N 95% confidence interval) for precision (Figure 7). The average difference
between the applied and the measured ML shear load was -0.5 N for bias (-1.5 N to 0.5 N 95% confidence interval)
and 0.3 N (-1.1 N to 1.7 N 95% confidence interval) for precision (Figure 7). ICCs indicated very strong agreement
(r > 0.998) between the applied and measured loads in all loading directions.
Across all specimens, the bone plug displaced 0.4 ± 0.3 mm in the AP direction, 0.4 ± 0.3 mm in the ML direction
and 0.2 ± 0.1 mm in the SI direction (Table 1). The bone plug changed orientation 1.0° ± 0.7° about the AP axis,
1.3° ± 1.4° about the ML axis, and 2.7° ± 2.1° about the SI axis (Table 1).
Under AP shear, the bone plug deflected 0.90 mm ± 0.60 mm in the AP direction (Table 1). Under ML shear, the
10
Discussion
The most significant finding of this study was that a 3-axis load cell could be placed under the PMMR
without disrupting native joint mechanics and could accurately measure magnitude and direction of force. The
installation was validated by the contact mechanics test demonstrating a contact area precision of 44 mm2 and a
contact pressure precision of 5 MPa between the pre-installation and post-installation states, the tibial displacement
test indicating an average bone plug displacement of < 1mm in all directions, the load validation test exhibited
average precision values of 0.7 N in compression, 0.5 N in tension, 0.3 N in anterior-posterior shear, and 0.3 N in
medial-lateral shear load, and the bone plug deflection test that confirmed <2 mm of displacement in any direction
The method validated in this study offers a new approach towards understanding detailed biomechanical
injury mechanisms of the PMMR. Etiological mechanisms for injury to the root carry complex three-dimensional
forces that to this point have only been characterized in theory, but never quantified (Bahr et al., 2005; Kennedy et
al., 2020; Stärke et al., 2013). Flexion angle, increased age, BMI, activity level, posterior tibial slope, varus
alignment, and meniscal extrusion are known risk factors for meniscal injury that place strain and can alter the three-
dimensional forces on the meniscus (Krych et al., 2020). Previous studies have assessed uniaxial PMMR forces by
measuring tension along an assumed axis of force transmission and do not decompose this force into its orthogonal
components (Stärke et al., 2013). As PMMR strength varies along each dimension, these prior studies are non-
specific in drawing conclusions related to PMMR injury and injury prevention. Markolf et al. (2012) reported force
measurements at the PMMR recorded before and after ACL removal ranging from 11 N to 60 N in magnitude. The
study, however, measured only tensile force along an arbitrary orientation towards the PMMR, and was unable to
discern force components in shear or compression. It is well documented that the PMMR exhibits anisotropic load
tolerances (Ellman et al., 2014), so force measurements of the PMMR that fail to discern directional magnitudes
provide, at best, an incomplete assessment of injury risk. Measuring PMMR forces without significantly altering
mechanics of the joint has also been a challenge in previous studies. In their study of PMMR shear forces in a
porcine model, Freutel et al. (2015) failed to acknowledge that, in modifying the contact mechanics of the medial
compartment, meniscal function is altered and the relevance of their force measurements to PMMR injury risk
11
The ability to comprehensively characterize forces in three dimensions at the PMMR can guide the
development of surgical techniques. Knowing the magnitude and direction of force on the root will allow surgical
repairs to be positioned to best resist these strains. Specific procedures that may be guided by this biomechanical
study include PMMR repair, medial meniscus allograft transplantation, medial meniscus repair, and meniscal
centralization or meniscotibial ligament repair. Additionally, the measurement of forces at the PMMR may allow a
better understanding of how coronal and sagittal lower extremity alignment impact forces at the PMMR. For
example, it may be found that tibial osteotomies are required in some circumstances with concomitant meniscal
allograft transplantation or PMMR repair to decrease the forces experienced at the PMMR. Increased posterior tibial
slope has been found to increase forces on the ACL and has changed clinical practice as an anterior closing wedge
osteotomy is required in some circumstances to decrease the rate of ACL re-tear. This has yet to be studied in the
meniscus, but the development of this new force measurement technique at the PMMR will allow this.
An additional benefit of this study is its application to rehabilitation and postoperative protocols. Future
studies can take specimens through various motions and measure which movements impart the greatest loads at the
root. It may elucidate how much flexion is acceptable post-op day 1 or which maneuvers need to be avoided in
physical therapy. This will allow clinicians to modify protocols that can be both protective of surgically repaired
menisci while permitting earlier rehabilitation and potentially earlier return to play.
Limitations
The authors recognize that the proposed method of measuring 3D PMMR forces has limitations. Firstly,
there is a small amount of displacement that can occur within the 2 mm gap that separates the bone plug from the
tibial plateau. The gap is necessary to ensure that the force sensor only reads forces from the meniscal root, but
displacement within this space may have a small effect on force measurements. For specimen selection, all knees
were from male specimens. There were initially concerns about accommodating the size of the sensor, although the
method developed turned out to be adaptable to a wide range of tibial sizes. Furthermore, the reported forces are
based upon the anatomical reference points selected to create the coordinate frames. While the authors minimized
variability of point selection using anatomic landmarks, variation in point selection will influence variation of the
Notably, the applied loads were low for deflection testing (~20 N at the bone plug) and load validation (~15
N - 40 N at the bone plug). Despite the PMMR load tolerance exceeding 500 N (Ellman et al., 2014) and dynamic
12
joint loading yielding higher PMMR loads, the authors intended to mimic the quasi-static load magnitudes measured
in previous studies (Freutel et al., 2015; Markolf et al., 2012). Secondly, during the piloting process, the study team
recorded shear loads that did not exceed 15 N when the joint was loaded in 500 N of compression. These selections
are not indicative of validation failure at higher loads as the manufacturer reports nonlinearity specifications as less
than 0.56 N. As with all cadaveric studies, there will be bony deformation when high loads are applied that could
potentially affect results. In addition, the femoral and tibial osteotomies needed to install the sensor may alter bony
Conclusion
The current study successfully validated a method that can measure the three-dimensional forces seen at the
posterior medial meniscal root in a human cadaveric model that maintains native knee mechanics at low loads. This
force measurement methodology has potential to significantly advance repair techniques and alter protocols
13
References
1. Gee SM, Tennent DJ, Cameron KL, Posner MA, 2020. The Burden of Meniscus Injury in
Young and Physically Active Populations. Clin Sports Med. 13-27.
2. Wilson PL, Wyatt CW, Romero J, Sabatino MJ, Ellis HB, 2018. Incidence, Presentation,
and Treatment of Pediatric and Adolescent Meniscal Root Injuries. Orthop J Sports Med.
2325967118803888.
3. Daney BT, Aman ZS, Krob JJ, et al., 2019. Utilization of Transtibial Centralization Suture
Best Minimizes Extrusion and Restores Tibiofemoral Contact Mechanics for Anatomic
Medial Meniscal Root Repairs in a Cadaveric Model. Am J Sports Med. 1591-1600.
6. Song GY, Zhang H, Liu X, et al., 2017. Complete posterolateral meniscal root tear is
associated with high-grade pivot-shift phenomenon in noncontact anterior cruciate
ligament injuries. Knee Surg Sports Traumatol Arthrosc. 1030-1037.
7. Wang W, 2021. Artificial Intelligence in Repairing Meniscus Injury in Football Sports with
Perovskite Nanobiomaterials. J Healthc Eng. 4324138.
8. Hwang BY, Kim SJ, Lee SW, et al., 2012. Risk factors for medial meniscus posterior root
tear. Am J Sports Med. 1606-1610.
10. Vedi V, Williams A, Tennant SJ, Spouse E, Hunt DM, Gedroyc WM, 1999. Meniscal
movement. An in-vivo study using dynamic MRI. J Bone Joint Surg Br. 37-41.
11. Markolf KL, Jackson SR, McAllister DR, 2012. Force measurements in the medial
meniscus posterior horn attachment: effects of anterior cruciate ligament removal. Am J
Sports Med. 332-338.
12. Stärke C, Kopf S, Lippisch R, Lohmann CH, Becker R, 2013. Tensile forces on repaired
medial meniscal root tears. Arthroscopy. 205-212.
13. Seitz A, Kasisari R, Claes L, Ignatius A, Dürselen L, 2012. Forces acting on the anterior
meniscotibial ligaments. Knee Surg Sports Traumatol Arthrosc. 1488-1495.
14
14. Seitz AM, Schall F, Hacker SP, van Drongelen S, Wolf S, Dürselen L, 2021. Forces at the
Anterior Meniscus Attachments Strongly Increase Under Dynamic Knee Joint Loading.
Am J Sports Med. 994-1004.
15. Guess TM, Razu S, 2017. Loading of the medial meniscus in the ACL deficient knee: A
multibody computational study. Med Eng Phys. 26-34.
16. Yao J, Snibbe J, Maloney M, Lerner AL, 2006. Stresses and strains in the medial meniscus
of an ACL deficient knee under anterior loading: a finite element analysis with image-
based experimental validation. J Biomech Eng. 135-141.
17. Freutel M, Seitz AM, Ignatius A, Dürselen L, 2015. Influence of partial meniscectomy on
attachment forces, superficial strain and contact mechanics in porcine knee joints. Knee
Surg Sports Traumatol Arthrosc. 74-82.
18. Fujie H, Livesay GA, Woo SL, Kashiwaguchi S, Blomstrom G., 1995. The use of a universal
force-moment sensor to determine in-situ forces in ligaments: A new methodology. J
Biomech Eng. 1-7.
19. Haut Donahue TL, Pauly HM, 2022. Osteoarthritic meniscal entheses exhibit altered
collagen fiber orientation. Connect Tissue Res. 151-155.
20. Koo TK, Li MY, 2015. A guideline of selecting and reporting intraclass correlation
coefficients for reliability research. J Chiorop Med. 155-163.
21. Dahl KD, Dunford KM, Wilson SA, Turnbull TL, Tashman S., 2020. Wearable sensor
validation of sports-related movements for the lower extremity and trunk. Med Eng
Phys. 144-150.
22. ASTM Standard E177, 1996. Standard Practice for Use of the Terms Precision and Bias in
ASTM Test Methods. ASTM International, West Conshohocken, PA, 1996
23. Kennedy MI, Strauss M, LaPrade RF, 2020. Injury of the meniscus root. Clin Sports Med.
57-68.
24. Krych AJ, Bernard CD, Leland DP, et al., 2020. Isolated meniscus extrusion associated
with meniscotibial ligament abnormality. Knee Surg Sports Traumatol Arthrosc. 3599-
3605.
25. Ellman MB, LaPrade CM, Smith SD, et al., 2014. Structural Properties of the Meniscal
Roots. Am J Sports Med. 1881-1887.
15
Figure Legends
Figure 1. A) Sensor construct consisting of mounting bracket, 3-axis sensor, and potting cup. B)
Medial condyle osteotomy for direct visualization of the PMMR. C) Removed bone block 2 cm
below joint line. D) Sensor placement with 2 mm margins. E) Converging screw fixation and
Figure 2. Testing apparatus fixture for the contact mechanics test featuring the steel plate holding
the femoral condyle osteotomy together with 4 bicortical screws (left). Schematics of the pre-
installation (top right) and post-installation (bottom right) testing states, the TekScan sensors
were installed beneath the meniscus from anterior to posterior on the tibial plateau by making
Figure 3. Creation of sensor coordinate system, 𝑨, and tibial coordinate system, 𝑩, using a 3D
(center) that is utilized to transform the force vector in the sensor coordinate system, 𝑭, to the
Figure 4. Specimen with sensor installed before final cuts made. Dashed yellow lines show
direction of final cuts and bone plug (left). Schematics before final cuts made to create bone plug
16
(top right) and after bone plug is created from final cuts with the intention to prevent bone plug
Figure 5. Testing set up for the load validation test and bone plug deflection test (left). For both
tests, a dynamic tensile testing machine applied a known load to the cortical bone of the bone
plug (right). For the load validation test, forces were recorded from the sensor and compared to
the known applied load. For the bone plug deflection test, the displacement of the dynamic
BP – bone plug
Figure 6. Representative pressure maps in the lateral and medial condyles comparing the pre-
installation and post-installation testing states (top left). Chart of bias, upper 95% confidence
margin, lower 95% confidence margin, precision, and ICC of contact mechanics test (top right).
Confidence margins show 95% confidence margin of the difference between pre-installation and
installation contact area (bottom left), medial contact pressure (bottom center), and lateral
contact pressure (bottom right). Bland-Altman plots display the mean of the differences (SD), the
standard deviation of the differences (SD), and the limits of agreement (LOA).
17
Figure 7. Box and whisker plots of bias (left) and precision (right) calculated across all applied
loading values (e.g. 0 N, 10 N, 20 N, …, 50 N) for each loading direction. Within each plot, “X”
Table Legends
Table 1. Mean values, standard deviations, and 95% confidence interval for outcomes of the
tibial plateau displacement test (top left), the bone plug deflection test (top right), the contact
18
.
Brad Fossum – Conceptualization, Validation, Investigation, Writing – Review and Editing, Visualization.
Scott Tashman – Writing – Review and editing, Supervision, Project Administration, Funding Acquisition.
Armando Vidal – Writing – Review and editing, Supervision, Project Administration, Funding Acquisition.
Matthew Provencher – Conceptualization, Writing – Review and editing, Supervision, Project Administration,
Funding Acquisition.
Thank you,
Justin Brown, MD on behalf of the author list
19
Figures
F F
T
Fb T Fb
F
F
T Fb
Figure 1. T
20
Figure 2.
Figure 3.
21
Figure 4.
Figure 5.
22
Figure 6.
Figure 7.
23
Tables
Table 1.
24
25