Cutsheet - XRAY SIEMENS

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

SIEMENS

SOMATOM DEFINITION AS
TYPICAL ROOM PLAN

SIEME.ks

The intended use for this Cut Sheet is to communicate the spatial requirements as well as the basic architectural,
electrical, structural, and mechanical requirements for this piece of imaging equipment. The information provided in this
document is for reference only, during the pre-planning stage, and therefore does not contain any site specific detailed
requirements. This information is subject to change without notice. Federal, state and/or local requirements may impact
the final placement of the components. It is the customer's responsibility to ensure that the final layout and placement of
the equipment complies with all applicable requirements.
FOR REFERENCE ONLY,
SIEMENS NOT FOR CONSTRUCTION.

SOMATOM DEFINITION AS
TYPICAL ROOM PLAN
CT

18'-9 1/2"
M AN Al.'IERNA'llVE, A SEPARAlE SOIR:E
OF CHIL!.ED WAlER CNI BE PRINIDED BY
lHE CUIOMER/C0NIRACl'OR. SEE CHIL!.ED
WAlER REQUIREMENTS, SHEEr M-101A.
I
L
1r~:0:1
__
. __ _J
5'-1 1/2" 1HE OUIDOOR COOUNG UNIT IS
SUPPLIED BY SIEMENS, LOCAlED AND

~~fl=~T
EXCHANGER CABINEr IS SUPPLED BY
SEMENS, INSTAL!.ED BY 1HE
CUSTOMER/CONTIW:TOR. IF lHE SYSTEM
IS TO BE REILOCAlED, lHE OUTDOOR
COOUNG UNIT IS TO BE DEINSTALLED,
RELOCAlED AND REINSTAU.ED BY lHE
CUSTOMER/CONTIW:TOR, IF NECESSARY.
1HE INDOOR HEAT EXCHANGER CMINEr
IS RELOCAlED BY SIEMENS, INSTAL!.ED
BY lHE CUSlllMER/CONTRACrOR.

lHE MNMUM ROOM WWME


REQUIRED FOR lHE WAlER-AIR
SPUT HEAT EXCHANGER IS 424
CUBIC FEET.

IEOHT DIFFERENCE BEIWEEN


1£AT EXCHANGER AND
OUTDOOR UNIT IIJST NC1I"
EXCEED ±116'-7".

SURCE SUPPRESSOR TO BE SUPPLIED BY


SIEMENS. LOCAlED AND INSTAL!.ED BY
CUSTOMER/CONIRAClllR. F IT IS TO BE
RELOCAlED, IT MUST BE DEINSTAU.ED,
RELOCAlED AND REINSTAL!.ED BY lHE
CUSTOMER/CONIRAClllR. F NECESSARY. IT
MUST BE LOCAlED WITHIN 3'-0" OF lHE
aRCUIT BREAKER.

IEOHT OF WINDOW TO BE
COORDINAlED WITH
COUNlERTOP/DESI< HEIGHT.

IT IS 1HE RESPONSIBLRY OF lHE


CUSTOMER/CONl1WlroR 10 PRINIDE A
~=JF~ix:"1:l ~ ~~
AGAINST TlP-11\/ER, FLIIDS, IMPACI", ETC.
ADDITlONN. MOUNTING IS NC1I" NECESSARY IF
SIEMENS CONTAINER IS UTIUZED.

TYPICAL PLAN SCALE: 1 ;a· = 1'-o·

SIEMENS MEDICAL SOLUTIONS USA, INC. CUTSHEET FOR lYPICAL # 08006 REV 8 PIDE 1 OF 5
FOR REFERENCE ONLY,
SIEMENS NOT FOR CONSTRUCTION.

SOMATOM DEFINITION AS
SPECIFICATIONS
CT
EQUIPMENT LEGEND
NO DESCRIPTION SMS WEIGHT BTU/HR DIMENSIONS {INCHES) REMARKS
SYM {LBS) TO AIR w D H
(.1) OPERATING CONSOLE (OPTION) 0 132 43 47 1/4 31 1/2 29 1/4

~ SYNGO ACQUISITION WORKPLACE AND STANDARD @ <55 1,7D6 7 5/16 19 17 OFF FLOOR/IN CONTAINER
COMPONENTS
0 CONTAINER FOR ICS/IES (OPTION) @ 77 --- 31 1/2 31 1/2 29 1/4 HOUSING FOR ICS/IES

~ SOMATOM DEFINITION AS GANTRY @ 4,851 3,412* 93 11/16 36 5/8 78 *ADDITIONAL HEAT DISSIPATED
TO WATER
141 SOMATOM DEFINITION AS GANTRY @ 4,851 44,357* 93 11/16 36 5/8 78 * AIR COOLED GANTRY
15) PATIENT TABLE 0 1,103 1,024 28 95 11/16 33 7/16 2000mm TABLE

~ POWER DISTRIBUTION CABINET @ 1,373 6,824 35 7/16 26 15/16 76 3/4 UPS LOCATED INSIDE OF
PDC
7) IMAGE RECONSTRUCTION SYSTEM @ 62 5,122 10 26 3/16 20
181 HEAT EXCHANGER CABINET - WATER/AIR SPLIT (OPTION) ® 904 3,412 39 3/8 27 1/4 77
191 OUTDOOR UNIT - WATER/AIR SPLIT (OPTION) @ 410 102,364 100 3/4 42 1/8 45 5/16
110) CARE VISION DUAL MONITOR (OPTION) ® 122 --- --- --- --- CEILING MOUNTED

C!!) EATON SURGE PROTECTIVE DEVICE PANEL (OPTION) @ 13.5 --- 7 1/2 6 11/16 12 WALL MOUNTED

® MEDRAD DISPLAY CONTROL UNIT/BASE UNIT (OPTION) @ --- --- --- --- --- HEIGHT WITH SCREEN UP

® CEILING MOUNTED MEDRAD INJECTOR (OPTION) @ 106 --- --- --- --- SEE MFG SPECIFICATIONS

FINISHED ROOM HEIGHT FOR MORE INFORMATION


FOR CT GANTRY ONLY MINIMUM 7'-6 9/16" FOR MORE DETAILED PLANNING REQUIREMENTS FOR THIS SYSTEM, SEE
CAREVISION MONITOR/CEILING MOUNT SEE DETAIL ON S-102 SHEET THE TYPICAL FINAL DRAWING SET NUMBER: 08006

SIEMENS MEDICAL SOLUTlONS USA, INC. CUTSHEET FOR lYPICAL f 08006 REV 8 PIDE 2 OF 5
FOR REFERENCE ONLY,
SIEMENS NOT FOR CONSTRUCTION.

SOMATOM DEFINITION AS
SPECIFICATIONS
---------------CT
POWER REQUIREMENTS LINE IMPEDANCE RELATED TO
SYSTEM SUPPLY POWER MAIN MEASURED LINE VOLTAGE
VOLTAGE CONSUMPTION CIRCUIT IJ.I
BREAKER z 250
(VOLTS) (kVA)
(AMPS) ::J
0 200
SOMATOM 391 SEE IJ.I,......
DEFINITION AS 480/277Y BELOW 125 ~a
±10,r;
oE
...J.....,
150
~IJ.I
POWER CONSUMPTION (WITH STANDARD HOSPITAL CHILLED (.) 100
:::.z
WATER OR AIR COOLED SYSTEM) ::::,~
CT OPERATING FOR 3 SEC - 140 kVA ~IJ.I
50
CT OPERATING AT 35 SEC - 93 kVA za.
CT OPERATING AT 100 SEC - 43 kVA :E~ 0
CT SYSTEM ON (STAND-BY) - 4 kVA 360 380 400 420 440
CT SYSTEM ON {COMP ON) - 2.5 kVA MINIMUM MEASURED LINE VOLTAGE {VAC)
CT GANTRY OFF {EVA ON) - 1.7 kVA
POWER CONSUMPTION (WITH OPTIONAL WATER/AIR SPLIT THE MAXIMUM ALLOWED LINE IMPEDANCE DEPENDS ON THE
COOLING SYSTEM) ACTUAL MEASURED LOWEST LINE VOLTAGE.
CT OPERATING FOR 3 SEC - 140 kVA WE RECOMMEND MONITORING THE VOLTAGE OVER A PERIOD OF
CT OPERATING AT 35 SEC - 93 kVA 24 HOURS PREFERABLY ON WORK DAYS TO SPECIFY THE
CT OPERATING AT 100 SEC - 43 kVA MINUMUM.
CT SYSTEM ON {STAND-BY) - 4 kVA
CT SYSTEM ON {COMP ON) - 2.5 kVA
CT GANTRY OFF (EVA ON) - 1.7 kVA
COOLING SYSTEM - 16kVA
COOLING SYSTEM FLOW HEATER {OPTIONAL) - 12kVA CASEWORK & ACCESSORY NOTES
IF AN ON-SITE TRANSFORMER IS REQUIRED TO OBTAIN CT 1) ALL CASEWORK IS EITHER EXISTING OR IS TO BE DESIGNED,
OPERATING VOLTAGE, IT MUST BE OF SUFFICIENT CAPACITY AND DETAILED, FURNISHED AND INSTALLED BY THE CUSTOMER AND/OR
CHARACTERISTICS TO MAINTAIN SUPPLY VOLTAGE AND IMPEDENCE CONTRACTOR. FOLLOW DESIGN RECOMMENDATIONS INCLUDED
HEREWITH, AS THEY ARE ESSENTIAL FOR THE SUCCESSFUL
REQUIREMENTS {TRANSFORMER AND CONDUCTORS). INSTALLATION & OPERATION OF THE SIEMENS EQUIPMENT.
ALL STANDARD COMPONENTS AND ADD-ONS ARE SUPPLIED VIA
2) ALL FURNITURE (CHAIRS, ETC.) FOR THE CONTROL ROOM ARE TO
THE POWER DISTRIBUTION SYSTEM. BE PROVIDED BY THE CUSTOMER.
DO NOT CONNECT NON-SIEMENS COMPONENTS SUCH AS LASER
CAMERAS OR FILM PROCESSORS TO THE SIEMENS POWER
DISTRIBUTION SYSTEM {PDS).
THE EXAMINATION ROOM SHOULD BE EQUIPPED WITH AT LEAST
ONE EMERGENCY POWER OFF {PANIC) BUTTON.
TO ENSURE SATISFACTORY SYSTEM OPERATION THE PDS MUST
HAVE A DEDICATED PROTECTIVE GROUND CONDUCTOR.

SIEMENS MEDICAL SOLUTIONS USA, INC. CUISHEET FOR lYPICAI. # 08006 REV 8 PAGE J OF 5
FOR REFERENCE ONLY,
SIEMENS NOT FOR CONSTRUCTION.

SOMATOM DEFINITION AS

---------------
SPECIFICATIONS

INCHES~---------------------~
RADIATION SCATTER
CT

59.1 .0253 .0230 .0103 .0038 .0036 .0133 .0241 .0365 .0282 .0215 .0170 .0138 SOMATOM DEFINITION AS
VERTICAL LOCAL DOSE DISTRIBUTION
39.4 MEASUREMENT IN uGy/mAs SCALE 1/4"=1'-0"

19.7 .0424 .0666 .01309 293 .1316 .0649 .0390 .0258 .0183 .0143 SCANNING WAS PERFORMED USING A MAXIMUM SLICE
THICKNESS OF 64 x 0.6 mm (38.4 mm) AT 140 kV
0 THROUGH THE SYSTEM AXIS IN THE VERTICAL PLANE.
PHANTOM USED: CYLINDRICAL PMMA PHANTOM, 32 cm IN
DIAMETER, 16 CM LONG. THE PHANTOM WAS CENTERED
-19.7 .0400 .0620 .1112 IN THE TOMOGRAPHIC PLANE.

-39.4

-78.7-59.1-39.4-19.7 0 19.7 39.4 59.1 78.7 98.4 118.1 137.8


INCHES

DELIVERY RADIATION SCATTER


INCHES
TRANSPORTING INFORMATION:
TOTAL GANTRY TRANSPORT WEIGHT: 5,268 LBS.
78.7- .0254 .0374 .0418 .0432 .0425 .0363 .0250 -
GANTRY WITHOUT TRANSPORT DEVICE: 4,851 LBS. 59.1 >- .0237 .0491 .0639 .0673 .0616 .0495 .0200 -
TRANSPORT DEVICE: 417 LBS.
GANTRY TRANSPORTING WIDTH: 4' -8 5/8" MAXIMUM. 39.4>- .0098 .0496 .1189 .1578 .1243 .0427 .0079 -
3' -0 5/8" MINIMUM.
GANTRY TRANSPORTING LENGTH: 10'-5 9/16" MAXIMUM. .6021
19.7 >- .0034 .0029 -
8' -5 9/16" MINIMUM. '
NORMAL TRANSPORT REQUIREMENTS:
DURING THE MOVEMENT OF THE GANTRY THROUGH
CORRIDORS THE TRANSPORT CASTERS ARE SWIVELED OUT
O>-

-19.7>-
.0047
+ I
.0124 .0525 .JJJ2 .5814 .3525 .0485 .0091
.0028 -
-
FOR STABILITY. SEE MAXIMUM WIDTH AND MINIMUM LENGTH
ABOVE FOR TRANSPORT CASTERS SWIVELED OUT.
-39.4>- .0301 .0837 .122! .1514 1373 .0755 .0286 -
NARROW SPACE TRANSPORT REQUIREMENTS:
WHEN TRANSPORTING THE GANTRY THROUGH A NARROW -59.1 >- .0364 .0508 .062E .0676 0653 .0489 .0333 -
SPACE OR DOORWAY THE TRANSPORT CASTERS ARE
SWIVELED IN AS SHOWN IN THIS SKETCH.
-78.7 >- .0274 .0329 .035, .0386 0379 .0322 .0256 -
3'-0 5/B"
3'-2 3/16"
-98.4- .0206 .0226 .0231 .0210 0244 .0221 .0190 -
-118.1 >- .0155 .0166 .016, .0153 0176 .0161 .0142 -

-137.8- .0118 .0128 .0130 .0121 .0130 .0127 .0121 -


....I I I I I I I I I I
io -78.7-59.1-39.4-19.7 0 19.7 39.4 59.1 78.7

- - -- SOMATOM DEFINITION AS INCHES


HORIZONTAL LOCAL DOSE DISTRIBUTION
MEASUREMENT IN uGy/mAs SCALE 1/4"=1'-0"
SCANNING WAS PERFORMED USING A MAXIMUM SLICE THICKNESS
OF 64 x 0.6 mm (38.4 mm) AT 140 kV THROUGH THE SYSTEM
AXIS IN THE HORIZONTAL PLANE. PHANTOM USED: CYLINDRICAL
PMMA PHANTOM, 32 cm IN DIAMETER, 16 CM LONG. THE
PHANTOM WAS CENTERED IN THE TOMOGRAPHIC PLANE.

SIEMENS MEDICAL SOLUTIONS USA, INC. CUISHEET FOR lYPICAI. # 08006 REV 8 PAGE 4 OF 5
FOR REFERENCE ONLY,
SIEMENS NOT FOR CONSTRUCTION.

SOMATOM DEFINITION AS

---------------
SPECIFICATIONS

ENVIRONMENTAL REQUIREMENTS
TEMPERATURE ("F)
SYSTEM COMPONENT
NOISE LEVEL
CT

DECIBEL LEVEL
{AT 3'-3• DISTANCE)

95+----t--+--+---+--+--+---+--+--1-----1 GANTRY <70

PATIENT TABLE <60


H~:::i:-:=1=--=t::;:::;;t:;::::;:t:;::;::t:;::;::1;::;:::1;;::::;-,-r-----i
PDC CABINET ~55
77-+-----1----v
IRSmx2C TOWER (40/64 SLICE CONFIG.) 50 TO 55 (1)
g~:::i:-:=t=-4,.L..e.,~4-L.t.::,.LL,..L.t.~H-11 IRSmx2b TOWER {128 SLICE CONFIG.) <55
59+----t--+--+---+--+--+---+-+--+--1-----1
HEAT EXCHANGER - WATER/AIR SPLIT <60

1) NOISE DEPENDS ON THE ROOM TEMPERATURE AND THE


PROCESSOR LOAD.

32-+-----,t----+-+----l--+--+---+--+-+--+-----I
0 10 20 30 40 50 60 707580 90 100
HOSPITAL WAlER

TEMPERATURE, HUMIDITY, DUST, AIR


" RELATIVE HUMIDITY
CHILLED WATER
CONTAMINATION: THE GANTRY IS COOLED WITH CHILLED WATER IN A CLOSED
REFER TO THE CLIMATOGRAM ABOVE FOR THE PERMITTED LOOP CONNECTION FROM THE ON-SITE CHILLED WATER SUPPLY.
CLIMATE RANGE. AN ON-SITE CONNECTION TO THE CHILLED WATER SUPPLY MUST
BE AVAILABLE TO SUPPLY THE HEAT EXCHANGER LOCATED
THE MAXIMUM TEMPERATURE GRADIENT IS 6 K/HR.
INSIDE THE GANTRY. THE REQUIRED WATER TEMPERATURE IS
THE ENVIRONMENTAL REQUIREMENTS FOR THE OPERATOR AND 39.2 TO 53.6"F. THE NOMINAL OPERATING PRESSURE IS 29 TO
THE SYSTEM IS 64 TO 82 "F WITH A RELATIVE HUMIDITY OF 87 PSI, {MAX. 145 PSI). THE MINIMUM FLOW RATE DEPENDS
20-75" AND A BAROMETRIC PRESSURE OF 11.6 TO 15.4 PSI. ON THE WATER TEMPERATURE. DIFFERENTIAL PRESSURE AS
EXTERIOR AIR VENTS SHOULD BE EQUIPPED WITH A FILTRATION RELATES TO WATER CIRCULATION. HEAT DISSIPATION INTO THE
SYSTEM OF THE FILTER CLASS MERV 8 TO FILTER DUST WATER IS 40,946 BTU/HR.
PARTICLES >10 um.
THE ROOM AIR SHOULD BE PROTECTED AGAINST CONTAMINATION
BY HYDROGEN SULPHIDE, EVEN IN SMALL AMOUNTS. IF A
DANGER OF SUCH CONTAMINATION EXISTS, CORRECTIVE ACTIONS
WATER/AIR SPLIT
HAVE TO BE TAKEN. E.G., EXTRACTOR FANS, SIPHON,
MODIFICATION OF VENTILATION INTAKE, ETC •• GANTRY COOLING
THE GANTRY IS COOLED WITH CHILLED WATER IN A CLOSED
LOOP CONNECTION FROM THE HEAT EXCHANGER. THE HEAT
EXCHANGER CABINET IS COOLED WITH CHILLED WATER IN A
REMOTE SYSTEM DIAGNOSTICS CLOSED LOOP CONNECTION FROM AN OUIDOOR COOLING UNIT.
THE AMBIENT AIR TEMPERATURE RANGE REQUIRED FOR THE
SIEMENS REMOTE SERVICES (SRS} REQUIRES A CONNECTION BETWEEN OUTDOOR COOLING UNIT IS -22" TO 122" (-40" TO 122" WITH
THE SRS REMOTE SERVER AND SIEMENS SYSTEMS VIA REMOTE LOCAL FLOW HEATER OPTION). BTU/HR TO AIR (EXHAUST) IS 102,364.
AREA NETWORK ACCESS, TO ENSURE THE UPTIME OF YOUR SYSTEM.
A CUSTOMER VPN CAPABLE FlREWALL OR OTHER VPN APPLIANCE IS
PREFERRED.

AIR COOi.ED
AIR-COOLED GANTRY
THE AIR-COOLED GANTRY HAS INTEGRATED COOLING FANS FOR
AIR INTAKE AND AIR EXHAUST. ROOM AIR IS USED AS COOLING
AIR. THE REQUIRED AIR INTAKE TEMPERATURE IS 64.4 TO 82.4"F.
THE REQUIRED AIR FLOW RATE THROUGH THE GANTRY IS 81,224
CUBIC FEET/HOUR. HEAT DISSIPATION INTO THE AIR IS 44,357
BTU/HR. THE RATING CAPACITY OF THE ROOM AIR CONDmONER
HAS TO TAKE INTO ACCOUNT THE STRUCTURAL CONDITIONS (EX.
WINDOWS, BUILDING & ROOM THERMAL INSULATION, ROOM SIZE,
ROOM VOLUME, ETC.) OF THE SCAN ROOM TO ENSURE THAT THE
TEMPERATURE RANGE OF AIR NEEDED FOR THE SYSTEM IS
MAINTAINED.

SIEMENS MEDICAL SOLUTIONS USA. INC. CUISHEET FOR lYPICAI. # 08006 REV 8 PAGE 5 OF 5