Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

AMYLASE

-Amilasa
CNP-G3. Cinético. Líquido

PREPARACIÓN Y ESTABILIDAD
Ref. 41201 y 41202. Los reactivos se encuentran listos para su BX_400 / TYB_40 / RX_IMMOLA
uso.
Son estables hasta la fecha de caducidad indicada en el envase conservado a 2-8ºC.

CALIBRACIÓN
Utilizar un suero de calibración adecuado (CFAS Ref. 1002011).

CONDICIONES OPERATORIAS
Introducir en la memoria del ordenador los siguientes parámetros:

CHEMISTRY PARAMETER

Method * Name AMY Reagent Name Reagent (µL) Water (µL)


Print Name AMYLASE R1 AMYLASE 300
Sample SERUM R2 Enable
Unit U/L
Assay Type KINETIC Diluent Enable

Mesuring Points Decimal Points 1


Start End
1 17 37 Normal Range
Enable 2 Nº Range Name Min. Max.
1 * * *
Wave Lenght 2 * * *
Prim. 412 Sec x Disable 3 * * *
4 * * *
Normal
Sampling Sample (µL) Diluent (µL) LOW HIGH
Dilution 6 Technical (Conc.) * *
Rerun (High Prozone) Range (mAbs/10) -999999 999999
Dilution
Rerun (Low) SPT Wash Enable Reag. Name
Dilution Stirring Speed R1-Middle R2-Middle

Limit Checks Blank measurement


Duplicate Limit mAbs/10
Sensitivity Limit mAbs/10 Blank measurement Enable Reagent Bk for S1 (linear)
Linearity Limit % (mAbs/10)/min Meas. of reagent BK during run Next Run
Prozone Limit Reag. BK. meas. at calibration Reagent BK (system water)
SL1-S SL1-F Nº of measurement Duplicate
SL2-S SL2-F
SENS mAbs/10 Reagent Blank Limit Checks:
Duplicate Limit mAbs/10
x Absorbance limit
Reaction Increase Instrument Factor
Limit * mAbs/10 A: 1.00 B: 0.00

Calibration: 2 linear point.

Rev.01 Ed. 2023/07

SPINREACT, S.A.U. Ctra. Santa Coloma, 7 E-17176 SANT ESTEVE DE BAS (GI) SPAIN
Tel. +34 972 69 08 00 Fax +34 972 69 00 99 e-mail: spinreact@spinreact.com
AMYLASE

-Amylase
CNP-G3. Kinetic. Liquid

PREPARATION AND STABILITY


Ref. 41201 and 41202. Reagent ready to use.
BX_400 / TYB_40 / RXIMMOLA
Unopened reagent is stable until the expiration date printed on
the label when stored at 2-8ºC.

CALIBRATION
Use a human serum calibrator (CFAS Ref. 1002011).

OPERATORY CONDITIONS
Introduce the following data in the computer:

CHEMISTRY PARAMETER

Method * Name AMY Reagent Name Reagent (µL) Water (µL)


Print Name AMYLASE R1 AMYLASE 300
Sample SERUM R2 Enable
Unit U/L
Assay Type KINETIC Diluent Enable

Mesuring Points Decimal Points 1


Start End
1 17 37 Normal Range
Enable 2 Nº Range Name Min. Max.
1 * * *
Wave Lenght 2 * * *
Prim. 412 Sec x Disable 3 * * *
4 * * *
Normal
Sampling Sample (µL) Diluent (µL) LOW HIGH
Dilution 6 Technical (Conc.) * *
Rerun (High Prozone) Range (mAbs/10) -999999 999999
Dilution
Rerun (Low) SPT Wash Enable Reag. Name
Dilution Stirring Speed R1-Middle R2-Middle

Limit Checks Blank measurement


Duplicate Limit mAbs/10
Sensitivity Limit mAbs/10 Blank measurement Enable Reagent Bk for S1 (linear)
Linearity Limit % (mAbs/10)/min Meas. of reagent BK during run Next Run
Prozone Limit Reag. BK. meas. at calibration Reagent BK (system water)
SL1-S SL1-F Nº of measurement Duplicate
SL2-S SL2-F
SENS mAbs/10 Reagent Blank Limit Checks:
Duplicate Limit mAbs/10
x Absorbance limit
Reaction Increase Instrument Factor
Limit * mAbs/10 A: 1.00 B: 0.00

Calibration: 2 linear point.

REV.01 Ed. 2023/07

SPINREACT, S.A.U. Ctra.Santa Coloma, 7 E-17176 SANT ESTEVE DE BAS (GI) SPAIN
Tel. +34 972 69 08 00 Fax +34 972 69 00 99 e-mail: spinreact@spinreact.com