Download as pdf or txt
Download as pdf or txt
You are on page 1of 4

Time dependence of spacer charging in SF6

under DC stress
Michael Schueller, Ueli Straumann and Christian M. Franck
Power Systems and High Voltage Laboratories, ETH Zurich, 8092 Zurich, Switzerland

Abstract- As gas insulated DC transmission systems unite the potential on the spacer can be quite high, up to a few kV. So it
advantages of space saving installation and big transmission might happen that charges are lost because of discharges back
capacity, this technology will become essential soon. The main to the high voltage electrode. As SF6 has a dielectric strength
problem in operation with respect to insulation seems to be of 8.8 kV/mm·bar, only the upper end of the spacer can be
identified today, which is charge accumulation on the surface of
the insulators, so called spacers. These trapped charges can cause
affected. Thus this effect should be negligible also due to its
a reduction of dielectric strength and can lead to a breakdown, small spatial impact.
especially at polarity reversals and overvoltages. The measurement probe itself can also have pronounced
For experiments, a full encapsulated test setup was constructed. influences on the charging. If the probe is put too close to the
Test voltages up to 700 kV DC and a pressure up to 8 bar SF6 are spacer surface, discharges from the surface onto the sensor are
possible. The main component of this setup is a 3 axis fully possible. As a Kalvin Probe (explained later) was used, this
automated robotic handle system for moving the probe over the effect is minimized as the sensor head is put to the same
test samples inside the encapsulation. So the surface charge can potential as it is measuring. Thus there is no potential between
be measured at all positions at the sample surface. the surface and the probe except at very fast potential jumps
The time dependence of the charging mechanism till the
faster than 330 V/ms, which is the maximal velocity of the
saturated state is of big interest. Thus the experiments presented
in this paper deal with the influence of the measurement itself on probe.
the charging process. Charges can also decay via the bulk insulation material as
charges drift by the face of the electric field. As the time
I. INTRODUCTION constant in Teflon is huge, this effect can be neglected
Considering the great increase of new renewables and thus considering the measurements that take around 1 minute.
the increased need for high transmission capacities, HVDC Of course the gas phase itself has an influence on the charge
technology will become more and more important in the future decay. The main effects that have to be considered are
[4]. As gas insulated equipment can be installed in a very detachment and recombination of the charges in the gas. This
space saving way, gas insulated switchgears and gas insulated was investigated in detail in [5].
lines are becoming increasingly interesting due to their
advantages already known from AC applications [3]. III. EXPERIMENTAL SETUP AND MEASUREMENT
PROCEDURE
Due to DC, the electric fields are not capacitively but As it was shown in [1] it is not possible to conclude from
resistively determined. This can lead to considerable charging small scale models to full scale geometries taking the gas
and a change in the dielectric stress of the insulation [7]. The phase of the setup into account. That is because the larger the
neuralgic part of the insulation seems to be the surface of the geometry, the higher the ion current in the gas gets due to
insulators interfacing SF6 and solid insulation, as well as the natural ionization, as the possible way of an electrical field
triple junctions on the insulators [8]. Especially breakdown line also gets longer when the setup turns from small size
voltages in situations of polarity reversal may be reduced models into real applications [1].
under DC conditions due to charge accumulation [9]. Thus the constructed test setup should have dimensions like a
In order to investigate the unknown effects with respect to DC, real GIS application and should be operated at voltages close
a suitable full encapsulated test setup was constructed for to real applications. This was reached by using parts from a
accurate measurements of SF6 insulated GIS spacer charging real GIS for 550 kV AC as a test vessel.
under DC stress.
The main goal of this paper is to show the impact of the Test SetUp
measuring procedure itself on the accumulation of surface The used test vessel has an inner radius of 0.55 m and the
charges. So it was investigated if the measurements of the overall volume of the setup is 617 liters. The experimental
surface potential influence the charge accumulation and with it setup can be operated from vacuum up to 8 bar SF6 or any
especially the temporal evolution of the surface charges. other gas.
The electrodes between which the test sample is put have a
II. THEORY modular design, thus various forms of electrodes can be
Various factors can influence the charging process and thus installed.
have to be considered. The required DC voltage is generated by two Greinacher
If the voltage on the high voltage electrode is switched off for cascades that can be switched on the test setup alternately.
measuring, it has a potential of U = 0 V. But the surface Thus power reversal tests with voltages up to +/- 320 kV DC

978-1-4673-1252-3/12/$31.00 ©2012 IEEE 629


with a reversal time of less than four seconds can be
performed.
As the setup has to be gas-tight, a handling system for test
spacers and the measurement probe had to be constructed
inside the test vessel. This was realized with various systems
built by the company Festo. Three axes were used for that
purpose: One horizontal axis for moving the measuring probe
Bushing
to the sample and back again when high voltage is applied, a
High voltage electrode
vertical driving system is used for moving the earth side
Specimen
electrode up and down, and the third system is to rotate the
earth electrode with the test sample standing on it. Thus every Probe for charge measuring

point of the spacer models is accessible and the accumulated earth electrode automatic probe handling system
DC high voltage source earth connection
surface charges on the whole surface of the test sample can be
measured. voltage
measurement
As correct positions - especially the distance between the
surface of the spacer and the measuring probe - are of great
insulation
importance, all the moving axes have an accuracy of repetition automatic earth electrode
of 0.08 mm. All axes can be operated at a speed of 100 mm/s handling system,
= rotation and height
to keep measurement procedures short in order to minimize
the impact of the measurement procedure on the charging
Fig. 1 Schematic view of the constructed test setup with voltage supply and
process. measurement in the left part and the test setup itself on the right side.
All axes can be controlled via PC and thus fully automated
measurements are possible. After putting the specimen between the electrodes, a pressure
Measuring of 0.5 MPa SF6 is filled into the test vessel. Then the High DC
For measuring the accumulated surface charges, the DC- Voltage is applied. For measuring, the high voltage has to be
Stable Electrostatic Voltmeter 347 from Trek was used. This switched off and the probe has to be moved to the spacer
is a high precision electrostatic voltmeter with a vibrating surface with the horizontal axis of the handling system. For
Kalvin probe for doing non contacting surface charge or better measurement accuracy, the probe is then kept in its
voltage measurements. The sensor is placed on the horizontal position and the cylindrical shaped test sample is moved in
automatic handling system and has to be put in a distance of 2 front of the sensor.
mm to the surface of the test sample for measuring. The measured voltage signals are put together with the
The principle of the vibrating Kalvin probe was introduced by positions of the 3 axes handling system to map the measured
W.A. Zisman in 1932 [2]. surface charge on the right position on the spacer surface.
As electronics has improved massively over the last years, this
measurement method is very suitable for surface charge IV. RESULTS
To the Authors’ knowledge, the time dependency of the
measurements as it is also very stable and accurate.
charging itself has not yet been studied in detail. Only
A step of 1 kV can be followed within 3 ms. Thus fast
measurements of the final saturated state of the charge
measuring procedures are possible. As the drift with time is
accumulation have been presented in literature so far. As also
less than 100 ppm/h, long time decay measurements of the
the charging process itself is of big interest, any possible
accumulated surface charges can be done as well.
disturbances of the measurements in the charging process are
Due to minimization of the measuring probe, according to the
going to to be worked out in detail in this section.
manufacturer a resolution of 5 mm² can be achieved.
Charge decay
Test Samples
Charge decay during measurements has to be considered to
For the experiments presented in this paper, Teflon samples
investigate the maximal possible length of time a
were used to minimize the effect of bulk conduction due to the
measurement procedure can take without having a large
high bulk resistivity of 1018 Ω cm. In order to get a big normal
impact on the whole experiment.
component of the electric field on the surface of the spacer,
The decay of accumulated surface charges was studied in
inserts were designed and put into the test samples.
more detail in [5]. It was found that the charge decay time
Measurement Procedure constant strongly depends on the height of the surface
First it has to be ensured that the test spacer is absolutely potential. Also a slight difference in the decay time of positive
dry. The massive influence of the humidity on the electric and negative charges was observed [5].
conductivity κ and thus the charging process was worked out The main mechanism of decay is natural ionization. The
in [4] in detail. To ensure the dryness of the test samples, they natural ion production rate was measured in the lab 10 m
are only stored in vacuum. Further the samples have to be below ground level with 29 IP/cm3s. This value matches the
charge-free before the experiment. This is done by using data from literature, such as 26...55 IP/cm3s at 0.4 MPa [6],
Isopropanol for cleaning the sample and only touching it with quite well.
conductive grounded gloves.

630
represents six measurements at different spacer - sensor
separations. A separation of 3 mm is suggested by the
manufacturer of the sensor and measurements at a separation
of 2 and 1 mm were performed as well. In order to see if the
first measurement in certain spacing did corrupt the
accumulated charges, because of the reasons discussed before,
a second measurement was taken immediately after the first to
see any differences in the surface charge potential.

Fig. 2 Charge decay over the length of the spacer surface for 130 min with a
measurement every 10 minutes after 120h charging with 1 kV (dashed line)

Fig.2 and Fig.3 represent a decay measurement after stressing


the sample with 1 kV DC for 120 hours. In Fig.2 the charge
distribution over the full length of the spacer is shown every
10 minutes, while Fig.3 pictures the surface potential decay in
certain points over 130 minutes decay time.

Fig. 4 Two immediately in a row taken measurements at 3mm sensor spacing


(solid), 2mm spacing (dashed) and 1mm spacing (dotted)
Statistical spread, measurements with and without
interruption
Comparing all identical measurements of a certain electrode
arrangement and using the same charging voltage (more than
the ones shown here), a statistical spread of less than 2.9 %
could be achieved with the constructed setup.

Fig. 3 Measurement points and curve fit for charge decay for 130 min at point
of maximal negative charge (solid), middle of the spacer (dashed) and point of
maximal positive charge (dotted)

The decay of the negative charges follows an exponential law


with time constant neg=43.5 min, while the positive decay
constant is pos=100 min.
Measurements at different spacing
Using a Kalvin Probe [2], the housing of the sensor has to be
brought to the measured surface potential in order to nullify
Fig. 5 Charging of a Teflon spacer with + 1 kV for 24h with and without
the current and get a measurement signal [2]. Depending on interruptions for measurements during the 24h charging process
the accumulated surface potential, the probe can reach
potentials up to 3 kV. In Fig.5 two measurements lasting 24 hours at + 1 kV DC
As the spatial resolution of the surface potential increases with charging voltage are presented. The dotted line represents an
decreasing distance between sensor head and spacer surface, experiment where a measurement point was performed every
this had to be investigated in detail as well. It was considered 10 minutes within the first 100 minutes of the charging time.
that very close spacing would give the best local resolution of After that, some more measurement points were taken after 3,
the measured potential but would maybe lead to discharges on 5, 12 and finally 24 hours DC stress. The solid line represents
the sharp edged sensor or corruptions due to the potential of an experiment under the same conditions as the one just
the sensor. Large spacing, in contrast, would decrease the explained with the difference that only one measurement was
local resolution. So the impact of the sensor on the performed after 24 hours of DC stress.
accumulated charges had to be investigated as well. Fig.4

631
The slight differences in both curves are within the statistical the one without. This difference is within the measurement
spread. accuracy and thus not significant.
An interruption for performing a measurement did not take
more than 1 minute from switching off the high voltage VI. CONCLUSION
performing the measurement till stressing the sample again  justifiable small impact on the accumulation process if
with high DC voltage. measurements are kept below 1 minute
 no charge detachment through the Kalvin Probe also at
high surface potentials and small measuring distances
V. DISCUSSION  decay in the presented test setup only through natural gas
Charge decay due to natural neutralization as detachment and neutralization as detachment and recombination and not
recombination seems to be a major mechanism for charge through any equipment or the measuring procedure itself
decay on the spacer surface. In Fig.2 and Fig.3 the decay over
 negative surface charges decay faster than positive surface
130 minutes is shown. It was shown that positive charges
charges
decay slower than negative charges. This was already shown
in literature [12] before. This could be explained with the VII. OUTLOOK
smaller electron affinity of the SF6 compared to the surface of The effect of slower charge decay for positive surface
the Teflon sample. Still a widely accepted model has yet not potential has to be studied in more detail and the mechanism
been found [13]. behind that has to be found.
Decay time till only 50 % of the initial charges are present was As the presented paper has shown and insured that
observed with t = 31.5 minutes at the negatively charged part interruptions during the charging process only have a small
and t = 74.8 minutes at the positive surface potential. The impact on the charging, detailed studies of the time
negative decay time exactly matches the half-value period of dependence of the surface charging effects are going to be
charge decay on Teflon in SF6 worked out in [10]. The reason performed.
for the slower decay for positive charges has to be studied in
more detail. In [5] it was shown that charges decay slower VIII. REFERENCES
when the surface potential is not that high. This is not the case [1] U. Straumann, M. Schueller and C. M. Franck, “Theoretical
Investigation of HVDC Disc Spacer Charging in SF6 Gas Insulated
here as half-period time for the same low starting potential of Systems”, submitted to IEEE Trans on Dielectr. and Electr. Insul.,
+/- 63 V surface potential was still measured with 31 minutes 2012
for the negative case and 75 minutes for the positive. [2] W. A. Zisman. Revue of Science Instruments., 3:367–368, 1932
Keeping one measurement procedure at a measurement time [3] T.J. Hammons, V.F. Lescale, K. Uecker, M. Haeusler, D. Retzmann,
K. Staschus and S. Lepy, “State of the Art in Ultrahigh-Voltage
of 1 minute, only 2.3 % of the accumulated charges decay in Transmission”, Proc. IEEE, Vol. 100, No. 2, pp. 360-390, 2012
the faster negative case. This rather small failure can be [4] B. Lutz, “Einflussfaktoren auf die elektrische Feldverteilung in
accepted as it is beyond the statistical spread of measurement Isoliersystemen mit polymeren Isolierstoffen bei
repeat accuracy of 2.9 % achievable with the presented setup. Gleichspannungsbelastung”, PhD thesis, Technical University Munich
TUM, 2011
The measurements presented in Fig.4 show that no surface [5] S. Kumara, Y. Serdyuk and S. Gubanski, “Surface charge decay on
charges are ablated by the measuring sensor. One polymeric materials under different neutralization modes in air”, IEEE
measurement sequence consists of 2 measurements taken Trans on Dielectr. and Electr. Insul, Vol.18, No. 5; October 2011
directly in a row at a certain distance. As in every sequence [6] S. Sato, W. S. Zaengl and A. Knecht, “A Numerical Analysis of
Accumulated Surface Charge on DC Epoxy Resin Spacer”, IEEE
both measurements correspond well, it can be concluded that Trans. Electr. Insul., Vol. EI-22, No. 3, pp. 333-340, 1991
no charges were ablated by the probe at the first performed [7] C. W. Mangelsdorf and C. M. Cooke, “Static Charge Accumulated by
measurement. Further it can be observed that a close Epoxy Post Insulation Stressed at High DC Voltages”, IEEE Conf.
measuring distance improves the local measurement accuracy. Electr. Insul. Dielectr. Phenomena (CEIDP), pp. 220-227, 1978
[8] S.I. Bektas, O. Farish and M. Hizal, “Computation of the electric field
This can be seen quite well from the profile at position 52 mm at a solid/gas interface in the presence of surface and volume charges”,
and also at the upper part of the spacer where the maximum IEE Proc-A, Vol. 133, No. 9, pp. 577-586, 1986
charge is accumulated. As the difference of the measurements [9] K. Nakanishi, A. Yoshioka, Y. Arahata Y. Shibuya, “Surface Charging
between 1 mm and 2 mm distance between sensor and surface on Epoxy Spacer at DC Stress in Compressed SF6 Gas”, IEEE Trans.
Power App. Syst., Vol. 102, No. 12, pp. 3919-3927, 1983
is quite small, a measurement distance of 2 mm instead of [10] C.J. Dais, J.N. Marat-Mendes and J.A. Giacometti, “Isothermal charge
3 mm recommended by the manufacturer of the measurement decay of Teflon – FEP electrets under a corona discharge”, Electrets,
equipment will be used for further measurements as no 1988. Proceedings pp. 493-498, 1988
charges are ablated but the measurement accuracy is quite [12] Z. Ziari, S. Sahli, A. Bellel, Y. Segui and P. Raynaud, “Simulation of
Surface Potential Decay of Corona Char ged Polyimide”, IEEE Trans
improved. on Dielect. and Electr. Insul. Vol. 18, No. 5; October 2011
From what is illustrated in Fig.5 it can also be concluded that [13] Z. Xu, L. Zhang and G. Chen, “Measurement and analysis of electric
with the presented fast and accurate test setup, short potential decay in corona charged low-density polyethylene films”,
interruptions for doing measurements during the charging IEEE Conference on Solid Dielectrics, Winchester, UK, July 2007
process can be performed due to their small impact on the
charging process. In fact the measured surface potential is
even larger in the measurement with interruptions compared to

632

You might also like