Sodium Potassium Analyser
* The electrolyte analyser is a device for
measuring the electrolytes in the human body.
They are primarily used in the quantitative
measurement of sodium, potassium, and
chloride in whole blood, serum, or plasma. ...
Some electrolyte analysers employ
miniaturized ISE, which are fabricated in
planar type
Sodium Potassium Analyser
* Principle: The electrolyte analyzer has
different electrodes, specific for different ions
of interest. ... A difference in sodium ion
concentration between the sodium solution
inside the electrode and the sample causes an
electrochemical potential to form across the
membrane of the active electrode.ODIUM POTASSIUM ANALYSERS
The electrolyte analyzer is a device for measuring the electrolytes in the
human body. They are primarily used in the quantitative measurement of
sodium, potassium, and chloride in whole blood, serum, or plasma. The most
common methods of electrolyte analysis are the flame emission photometry
and ion-selective electrode (ISE). The ISE method of electrolyte
measurement is based on the principle of potentiometry in which the
valtage that develops between the inner and outer surfaces of an ISE
(membrane) is measured. Some electrolyte analysers employ miniaturized
ISE, which are fabricated in planar type. The optode technology does not use
electrodes or contact points, which eliminates the need for costly electrode
maintenance. Electrolyte analysers find applications for measurement of
electrolyte levels in the human body to detect metabolic imbalances and
measure renal and cardiac functions.iy
a
1-what analyte you want to measure ?
The external reference electrode and
measuring electrode are exposed to the
unknown sample solution, potential of
the sample solution is measured
relative to the external reference.
2- Buffering the voltage generated by
ion selective electrode by operational
amplifier.
3- Analong multiplexer selecting signal.
4- An (AID 1) converter for converting the
signal into digital domain .
5- Micro-processor based computer
control mechanical functions of the
analyzer.
6- Digital display for showing results
and displaying messages and questions
7- The motor driver turns drives D.C
motor.
8- D .C motor drives the Peristaltic
Pump9-Bubble detector verifies the presence
of air or fluid in the sample tubing.
10-An (AID 2) converts the bubble
detector reading into digital form to
the computer.
METHODS OF ELECTROLYTE
ANALYSIS
m Flame emission photometry
m lonselective electrodeFlame Photometer
* Aphotoelectric flame photometer is a device
used in inorganic chemical analysis to
determine the concentration of certain metal
ions, among them sodium, potassium, lithium,
and calcium. Group 1 and Group 2 metals are
quite sensitive to Flame Photometry due to
their low excitation energies
working
The principle of flame photometer is based on
the measurement of the emitted
light intensity when a metal is introduced into
the flame. The wavelength of the colour gives
information about the element and
the colour of the flame gives information
about the amount of the element present in
the sample.
applications
Flame photometer applications. Determining
the concentration of sodium and potassium
ions in infusion solutions, such as
NaCl solution, Ringer solution or others.
Product control and indirect quality testing of
various substances over sodium, potassium or
lithium. Concentration determination in
pharmaceutical reagents