Professional Documents
Culture Documents
Npu O&m Manual
Npu O&m Manual
Section Inserts
(this page does not go into the manual)
Section Material
2 nitrogen MSDS
activated carbon MSDS
3 Figure 3.1A
Figure 3.1B
4 Figure 4.2.A
Figure 4.2.B
6 table 6.6
Refill Procedure for Carbon Bed
9 chart 9-2
─────────────────────────────────────────────────────────────────
GENERON NPU Membrane Nitrogen System
Operating and Maintenance Manual OCT. 2007
─────────────────────────────────────────────────────────────────
SECTION 1
INTRODUCTION
1.1 SCOPE
This manual describes the general operating and maintenance instructions for a
GENERON* NPU Membrane Nitrogen System as designed by Generon IGS.. The purpose
of this manual is to enable a trained operator to start-up, shutdowns, make minor process
adjustments, and maintain a unit, which has been set-up initially by a trained Generon IGS.
Service representative. This manual is not intended to instruct the user on how to perform
detailed modifications or major set-up changes to the NPU membrane system.
Customers may direct specific questions related to the maintenance and operation of the
GENERON NPU Membrane System to the Generon IGS. Customer Service Department at
713-937-5200 or 800-871-0310.
1.2 CONFIDENTIALITY
This document contains confidential or proprietary information of Generon IGS.. Neither the
document nor the information herein is to be reproduced, distributed, used or disclosed,
either in whole or in part, except as specifically authorized by Generon IGS..
The membrane modules should never reach freezing temperatures. Keep the modules
between 40°F (4.4°C) and 150°F (65.6°C). Exposure to temperatures outside the specified
range could lead to membrane damage and performance reduction.
The system is equipped with a space heater, which is able to keep the temperature inside
the container at any time above 50°F. The space heater has a thermostat, which shall be
set to 50°F.
SECTION 2
SAFETY
The GENERON NPU Membrane System is capable of producing nitrogen with purities ranging from
95% to 99.9%. The nitrogen produced by the GENERON NPU Membrane System is colorless,
odorless; chemically inert gases defined as a simple asphyxiate. There are no definite warning
signs for exposure to nitrogen. Nitrogen is non-toxic, however, the liberation of large amounts of
nitrogen will displace the oxygen needed to support life.
A general indication of the symptoms and effects experienced when exposed to various
levels of oxygen deficient atmospheres is given below.
Avoid confined spaces until oxygen adequacy (20.9%) has been proven by analysis or other
positive means, or until breathing air has been provided by air pack, air hose or other
trusted sources. Confined areas/spaces include the following:
Do not allow the product nitrogen to be vented other than through proper piping. Warning
signs should be posted in any area and/or confined space which may be a potential
nitrogen deficient area. Appropriate warning signs are available from the GENERON IGS.
Service Department.
The enriched oxygen vent gas (permeate) consists of 30-45% oxygen as compared to
normal levels, which are 20.9% oxygen. While oxygen will not burn, oxygen concentrations
above 28% can support combustion of other materials much more readily than air. It is
imperative that the enriched oxygen vent gas be discharged only in well-ventilated and safe
areas. If the GENERON NPU Membrane System is located in an air tight or poorly
ventilated room, the vent gas should be piped outdoors for release.
The PVC permeate vent line must never be pressurized. This line should always be
allowed to flow un-restricted to atmosphere in a safe area.
The GENERON NPU Membrane System should not be subjected to excessive external
forces or impact. Excessive force or impact applied to a pressurized line may result in the
sudden release of pressurized air and possibly inflict injury to nearby personnel.
Leak checks should be performed on all piping components to ensure no compressed air or
product nitrogen is allowed to escape. When performing leak checks, use an accepted leak
detection solution.
─────────────────────────────────────────────────────────────────
GENERON NPU Membrane Nitrogen System
Operating and Maintenance Manual OCT. 2007
─────────────────────────────────────────────────────────────────
Personnel unfamiliar with the electrical equipment supplied with the GENERON NPU
Membrane System should not attempt to service this equipment. The electrical currents
supplied to the various electrical components on the GENERON NPU Membrane System
are at levels high enough to cause severe injury and/or death.
When installing a GENERON NPU Membrane System, ensure that all applicable earth
grounds are installed as required by the National Electric Code and/or local
requirements. Refer to the appropriate manufacturer's manuals for correct grounding
procedures for each piece of electrical equipment.
The Material Safety Data Sheets (MSDS) included in this section are provided to inform the
customer of the nature of the bulk substances contained in or produced by the GENERON
NPU Membrane System. It is the responsibility of the customer to provide access to this
information for personnel working around the equipment as appropriate. The MSDS sheets
included in this section are for Nitrogen gas and Activated Carbon.
─────────────────────────────────────────────────────────────────
GENERON NPU Membrane Nitrogen System
Operating and Maintenance Manual OCT. 2007
─────────────────────────────────────────────────────────────────
SECTION 3
ENVIRONMENTAL
Several components on the GENERON NPU Membrane System are equipped with
condensate drain lines. Refer to the GENERON NPU Membrane System Piping and
Instrument Diagram (P&ID) located in Section 10 of this manual for the specific equipment
and location of the drain lines. These drain lines collect oily condensate removed from the
compressed air supplied to the GENERON NPU Membrane system. It is the customer's
responsibility to ensure that the oily condensate released from these drain lines is disposed
of in accordance with local statutes. Anticipated quantities and oil concentrations of
condensate can be calculated utilizing Figures 3.1.A and 3.1.B.
Waste oil collected during compressor oil changes and oil soaked filter elements removed
from the coalescing and particulate filters are considered hazardous waste. It is the
customer's responsibility to ensure that all waste oil and oil soaked filter elements are
disposed of in accordance with governing statutes.
Oil spills or leaks from the compressor must be properly cleaned up. Commercial
absorbent materials are available for this purpose. Use only absorbent material that has
been approved by a responsible governing agency. In some cases, disposal of
contaminated soil may be necessary.
─────────────────────────────────────────────────────────────────
GENERON NPU Membrane Nitrogen System
Operating and Maintenance Manual OCT. 2007
─────────────────────────────────────────────────────────────────
SECTION 4
PROCESS DESCRIPTION
4.1 GENERAL
The GENERON NPU Membrane System is equipped with modules in which the nitrogen
separation takes place. The modules are filled with a cylindrical bundle of hollow fiber
membranes made of a polymer-based material. Each bundle contains several million
fibers, each about the size of a human hair. Compressed air enters the end of the fibers
and flows inside the fibers through the membrane bundle to the opposite end.
Gas separation takes place as the pressurized air contacts the membranes. "Fast" gases
such as oxygen, carbon dioxide, and water vapor quickly permeate through the fiber walls
and exit at atmospheric pressure through the vent port on the side of the module case.
Nitrogen, a slower gas, does not permeate through the fiber as quickly under flowing
conditions. It flows down bore of the fibers and exits at the product manifold on the end of
the module case. This enriched nitrogen stream exits the case at 2-10 psi (0.13-0.7 bar (g))
lower than the pressure at which air enters the case.
Membrane performance is generally described by the terms productivity (flow) and recovery
(efficiency). This performance is a function of the product purity, operating temperature, and
operating pressure.
In general, productivity increases with an increase in pressure and temperature (at constant
purity).
Recovery, on the other hand, increases slightly with pressure, but decreases as
temperature rises (at constant purity).
Both recovery and productivity increase as the oxygen concentration (product impurity)
increases (at constant pressure and temperature).
The NPU system operates most effectively at 335 psig 80o-110oF (23 bar(g), 27o- 43oC).
─────────────────────────────────────────────────────────────────
GENERON NPU Membrane Nitrogen System
Operating and Maintenance Manual OCT. 2007
─────────────────────────────────────────────────────────────────
SECTION 5
EQUIPMENT DESCRIPTION
5.1 GENERAL
For explanation purposes, the GENERON NPU Membrane System has been broken down
into four standard subsystems. These subsystems include the Feed Air subsystem, the Air
Pretreatment subsystem, the Product Gas Separation subsystem, and the HPLC Control
System. The descriptions included in this section are designed to provide the operator with
a basic understanding of the function of the components used on the GENERON NPU
Membrane System.
The Process and Instrument Diagram (P&ID) located in Section 10 of this manual reflects
these components and process system supplied to the customer.
The feed air source for the GENERON NPU Membrane System may consist of a dedicated
air compressor or customer-supplied compressed air from a plant or instrument air system.
A dedicated compressor, if supplied by Generon IGS., is usually a lubricated rotary screw
compressor sized to produce the required air flow rate at operating pressures with a minimal
horsepower requirement. The air compressors are typically equipped with the following
features: modulating controls, auto restart from Standby state, high dust filters, and
outdoor/freeze protection (if required), and are filled with synthetic lubricant. Refer to the
compressor manual itself for a description of the actual compressor.
Customers provided air sources of lower feed pressures will result in correspondingly lower
nitrogen production (see Section 4.2). The customer should ensure, however, that the feed
pressure of the air source remains stable and consistent to meet the intended membrane
system design performance.
The Air Pretreatment subsystem contains equipment required to properly condition the
process or feed air supplied to the membrane modules. Typical equipment includes an air
receiver, moisture separator and air dryer to dry the air prior to the container feed air inlet.
Within the container are a demister vessel (T-10), coarse coalescing filter (F-11), fine
coalescing filter (F-12), carbon filter (F-16) and a particulate filter (F-18). Refer to the
Process and Instrument Diagram (P&ID) located in Section 10 for specific equipment
supplied.
Proper operation and maintenance of this filtration system will prevent oily condensate,
airborne particles and pipe scale from contaminating and/or clogging the membrane fiber
openings.
─────────────────────────────────────────────────────────────────
GENERON NPU Membrane Nitrogen System
Operating and Maintenance Manual OCT. 2007
─────────────────────────────────────────────────────────────────
The demister vessel (T-10) is a coded pressure vessel with a design pressure of
375 psig (25.9 bar(g)). An internal demister pad conditions the air before leaving the
vessel. The primary function of the demister vessel is to remove large quantities of
oily water condensate from the feed air supply.
The demister vessel is equipped with a drain line and all components necessary to
continuously drain the collected condensate.
Coarse and fine coalescing filters (F-11 & F-12), which utilize a 1.0-micron and 0.01-
micron filter element respectively, are designed to remove virtually all of the
remaining oil aerosol still present in the feed air despite the moisture separator.
Regular replacement of the filter elements is essential for proper filter operation.
Refer to Section 6 for further information.
The coalescing filters are each also equipped with a condensate drain line.
Equipment on the drain line for the coalescing filters includes a block valve and an
electronic drain trap.
The process heater (EH-14) is an electric air circulation heater which optimizes the
membrane module performance by ensuring that the feed air stream is heated and
maintained at a constant temperature ranging between 86 oF and 113 oF (30oC and
45oC). In case of a higher inlet temperature than 113 oF (45oC), the heater will
increase the temperature further to superheat the feed air before entering the
membrane modules. The process heater receives its controlled power from the main
power panel mounted in the GENERON NPU.
The process heater (HE-14) is an heat exchanger which optimizes the membrane
module performance by ensuring that the feed air stream is heated and maintained
at a constant temperature ranging between 77 oF and 113 oF (25oC and 45oC).
The carbon filter (F-16) consists of a carbon steel ASME or locally coded vessel
containing a specially packed activated carbon bed. The activated carbon pellets
used in the carbon filter adsorb heavy hydrocarbon vapors that may be entrained in
the feed air supply. Refer to Section 6 for further information regarding the
maintenance intervals for replacing the activated carbon pellets.
The particulate filter (F-18), which utilizes a 0.01-micron filter element, is intended to
remove all remaining airborne particulate matter (rust, carbon dust, etc.) remaining
─────────────────────────────────────────────────────────────────
GENERON NPU Membrane Nitrogen System
Operating and Maintenance Manual OCT. 2007
─────────────────────────────────────────────────────────────────
in the feed air supply. The particulate filter is typically identical to the coalescing
filter, although the primary reason for this filter is solid particle removal.
The drain line used on the particulate filter contains a block valve for manual venting
to ensure that no liquid is present at this point.
Feed valve FSV-19-1, which is controlled by the PLC, provides on/off control of the
feed air supplied to the membrane modules. The feed valve is classified as a "fail
close" valve that enables the GENERON NPU Membrane System to isolate the
membrane modules from the feed air stream in case of a shutdown condition.
Air bypass valve FSV-19-2, which is controlled by the PLC, provides control of the
air bypass of the membrane modules. The air bypass valve is classified as a "fail
close" valve.
The Product Gas separation subsystem contains the membrane modules and the
components necessary for the proper operation and control of the GENERON NPU
Membrane System. The Process and Instrumentation Diagram (P&ID) located in Section
10 of this manual reflects the components and flow system specifically provided to the
customer.
Refer to Section 10 for information on the module cases used on the GENERON
NPU Membrane System.
The N2 Purity Control valve is used to control the N2 Purity from the membrane
modules. An electro-pneumatic positioner FY-23-1 located on the valve receives the
control signal from the PLC. The PLC System uses a PID control loop to maintain
purity at the preset set point. The O2 analyzer AT-20-1 monitors the actual O2
content downstream of the membrane modules and transmits the information to the
PLC.
The N2 control valve is classified as a Fail-Close type valve. When the control
signal from the PLC is less than 4 mA, the control valve will be fully closed.
Increasing flow corresponds to decreasing the purity; and reducing the flow will
correspond to increasing the N2 purity.
─────────────────────────────────────────────────────────────────
GENERON NPU Membrane Nitrogen System
Operating and Maintenance Manual OCT. 2007
─────────────────────────────────────────────────────────────────
The GENERON NPU Membrane System is equipped with purity assurance valves
(FSV-24-1A & FSV-24-1B). When actuated the product valve FSV-24-1A allows on-
spec product gas to flow into the customer's pipeline. The product valve FSV-24-1A
fails to the closed position and the product vent valve FSV-24-1B fails to the open
position, when product gas contains too high or too low oxygen impurity, or the
system shuts down. The off-spec gas vented from the product vent valve FSV-24-
1B is Oxygen deficient and can be hazardous to personnel safety. This gas shall be
vented outdoors.
The permeate gas is the waste gas stream from the nitrogen separation process.
Running the module in parallel, permeate gas typically contains 30-45% Oxygen
and should be piped outdoors to a safe area. Refer to Section 2.1.2 for proper
handling of the waste gas.
The control system on the GENERON NPU Membrane System consists of a PLC-based
control system, oxygen analyzer, pressure transmitters, front panel pushbuttons and
operator interface terminal, purity assurance system and purity control system.
The Programmable Logic Controller (PLC), located in the NPU Control Panel is pre-
programmed with the appropriate program for the individual system. It consists of an
assembly of modular components which is responsible for:
a) Taking input signals from front panel buttons and switches and displaying
status information on front panel indicators as well as remote status
indicators.
─────────────────────────────────────────────────────────────────
GENERON NPU Membrane Nitrogen System
Operating and Maintenance Manual OCT. 2007
─────────────────────────────────────────────────────────────────
b) All switching valve sequencing timers and counters. The PLC will energize
and de-energize piloting solenoids, as appropriate.
For specific instructions on programming the NPU control system, consult the
Generon IGS Service Center. In general, though, control system programming
changes are nothing more than changing the set points in the PLC memory.
The oxygen analyzer located in the GENERON NPU control cabinet continuously
monitors nitrogen product gas for oxygen content. During operation, sample gas
from the product header flows through the analyzer cell and generates a signal
proportional to oxygen concentration. The product sample purity is monitored by the
PLC, which adjusts the module outlet pressure setpoint to control nitrogen purity.
The front panel consists of Lighted Pushbuttons and an Operator Interface Terminal.
There are two lighted pushbuttons on the panel, ALARM & SHUTDOWN.
Two switches are POWER ON/OFF & AUTO/STOP/MANUAL
5.5.3.1.3 AUTO
5.5.3.1.4 MANUAL
5.5.4.1.5 STOP
Placing the AUTO/STOP/MANUAL switch in STOP will stop
the flow of air into the modules and shut off the heater.
There is an operator terminal built into the control panel front that allows the user to
easily interface with the PLC System within the panel. The OIT has several screens
that provide the user with menu-oriented functions to view operating, configuration,
and programming data. The terminal also has an Alarm Banner that can clear the
alarm from the Panelview screen only.
MAIN MENU
All Screens link to the Main Menu. From this menu
you may go to any screen in the system. To navigate
to any screen, push the corresponding “F” key. On all
other screens the “F1” key will take you back to the
MAIN MENU. O2 analyzer warm-up time as well as
current time and date can be found here also.
High oxygen impurity set points are entered into the PLC via the OIT Menu. The
menu items for HIGH OXYGEN ALARM/SHUTDOWN LEVEL are under the Analog
Setup Screen. If oxygen concentration of the product gas exceeds these set points,
the product gas will be vented by FSV-24-1B to prevent the contamination of the
customer's process with off-spec nitrogen. This venting will continue until purity is
re-establishment, change by user or the Vent Shutdown Time “F7” from the Analog
Setup Screen times out.
The Purity Control Valve position is controlled by the PLC to control a level of
expectable purity.
─────────────────────────────────────────────────────────────────
GENERON NPU Membrane Nitrogen System
Operating and Maintenance Manual OCT. 2007
─────────────────────────────────────────────────────────────────
SECTION 6
OPERATION OVERVIEW
6.1 GENERAL
The GENERON NPU Membrane System has been designed to run unattended except for
periodic operational and calibration checks.
This section describes the various operating modes, machine states, and alarm functions of
the HPLC control system, as well as the physical maintenance requirements with which the
on-site operator should be familiar.
The GENERON NPU Membrane System allows a choice of product pressure control
modes: Manual Control and Purity Control. The mode chosen determines how constantly
purity will be maintained over transitions of air supply, product demand, and ambient
conditions.
In the Manual control mode the Product PCV Opening is controlled with respect to
the value (0-100%) entered on the Analog Setup Screen. Because of ever
changing airflow this method should not be used. This is mainly for convenience of
troubleshooting the valve or stroking the valve to check movement.
Auto Control uses the O2 Analyzer to maintain the desired purity. The O2 purity
setpoint is entered on the Panelview Analog Setup screen and the PLC maintains
this setpoint reacting to flow, temperature and pressure changes.
The GENERON NPU Membrane System operates in one of six machine states. Of these
six states, five are considered Normal and one is considered Abnormal.
The five Normal states in which the GENERON NPU Membrane System operates
are STARTUP, WARMUP, MANUAL, VENT and AUTO CONTROL. When the
GENERON NPU Membrane System is operating in one of the Normal states, it is
operating correctly and within the parameters set up by the startup engineer.
The Startup state is a transitional state that must precede the Running state.
The Startup state is entered when the GENERON NPU Membrane System
energized for the first time and is designed to allow the O2 analyzers time to
warm up.
─────────────────────────────────────────────────────────────────
GENERON NPU Membrane Nitrogen System
Operating and Maintenance Manual OCT. 2007
─────────────────────────────────────────────────────────────────
Auto and Manual can not be obtained until startup timer is finished.
An alarm will energize if time has not expired.
Startup time remaining can be found on the Main Menu.
The duration of the Startup state is determined by the Startup Timer (60
seconds)
When the system is started from a cold state the system will be sent to vent
and the heater will fire to a hotter than normal set point in order to warm the
system up quickly. All these settings are settable in the Generon setup
screen.
In the Manual control mode the Product PCV Opening is controlled with
respect to the value (0-100%) entered on the Analog Setup Screen.
Because of ever changing airflow this method should not be used. This is
mainly for convenience of troubleshooting the valve or stroking the valve to
check movement.
The Auto Control State is nearly the same as the Running state, except that
the O2 set point controls the valve to maintain an oxygen concentration.
The Vent State is active when the system is first started up for 10 minutes or
when O2 purity has been above shutdown setpoint for over allotted time.
As the title suggests, when in the abnormal state, the GENERON NPU Membrane
System is not operating within the prescribed parameters as set up by the startup
engineer.
─────────────────────────────────────────────────────────────────
GENERON NPU Membrane Nitrogen System
Operating and Maintenance Manual OCT. 2007
─────────────────────────────────────────────────────────────────
The GENERON NPU Membrane System can enter this state from the
Running or Purity Control states when the oxygen level is above the High
Alarm/Shutdown level.
The purpose of this state is to notify an operator that a problem exists, while
allowing a set period of time to correct an off-spec condition without shutting
down. In this state:
If a heater alarm has occurred, the alarm must be corrected, and the system
must be put into STOP, then restarted.
Alarms can be acknowledged using the F2 key. This will clear the alarm from the screen but
will not stop the Alarm Light from flashing. If the alarm condition clears, the Alarm Light will
stop flashing on it's own. The ALARM HISTORY SCREEN will keep a history of the last 50
alarms. If a shutdown alarm occurs, the system AUTO/STOP/Manual switch must be place
in the STOP position and the alarm condition cleared before the system can be re-started.
E-STOP ACTIVE
Sytem emergency stop is pushed
AT-20-1 IS OFFLINE
The Oxygen Analyzer has dropped below 3.0 mA and has failed. Check for
analyzer failure, the system will become unstable and may be place into
manual to continue operation if necessary. The high/low O2 limit setting may
have to be adjusted if running the system is needed while the problem is
solved.
DT IN USE
Incoming air temperature (RTD 14-1) is within 7 degrees F of the heater set
point. Heater set point is being raised automatically to adjust for the hotter
incoming air.
HEATER TRIP
If electrical heater is used, the heater contactor has tripped and shutdown
the heater, check heating circuit for heater problems
The GENERON NPU Membrane System is designed to run unattended and requires only
minor attention to assure proper operation.
A listing of operational checks is given in Table 6.5. Some critical activities are
detailed below. For further information on these and other items please contact the
Generon IGS. Service Department.
Most particulate and coalescing filters can tolerate very high differential
pressures. In fact, manufacturers typically do not require element
replacement until the steady state differential exceeds 9 psi (0.6 bar(g)).
The Oxygen analyzer's fuel cell sensor used in the GENERON NPU control
have a service life of approximately one year. During that time, the
calibration may drift slightly, and should be checked monthly. See Section-9
and the Fujikora documentation in Section-10 for information of calibration
and service of this critical instrument.
─────────────────────────────────────────────────────────────────
GENERON NPU Membrane Nitrogen System
Operating and Maintenance Manual OCT. 2007
─────────────────────────────────────────────────────────────────
SECTION 7
STARTUP PROCEDURE
7.1 GENERAL
The following procedure should be used to bring the GENERON NPU Membrane System
on-line after a prolonged shutdown or major maintenance. Ensure that all piping
connections are properly fitted and electrical connections are secure before initiating
startup.
7.2 POWER UP
Switch the Main Power Panel CC-1 disconnect switch to the "ON" position.
Turn on the HPLC Control panel by turning the Power switch on. The HPLC Control panel
will activate the timer for the warm up period for the Oxygen analyzer.
The following equipment settings need to be established when starting up the unit.
Some must be done prior to startup, others can only be done when the system is
operating.
Set the instrument air regulators (PRV-19-1 & PRV-19-2) to 6-Bar 87-psi.
This can be done prior to startup, when the skid is pressurized.
Once there is pressure in the nitrogen product pipeline, also set the sample
regulator inside the CC-2 control panel (PRV-20-2) to 0.35-bar 5-psi. Adjust
the sample flow rate, using flow indicator FI-20-1 inside the panel, to 0.5 to
1.0 scfh.
Before turning the AUTO/STOP/MANUAL switch to AUTO, select the heater output signal to
either “Electrical” or “Heat Exchange” mode. This can be achieved by toggling “F9” on Main
screen of HMI. Read the instruction on the screen carefully to insure right selection. The
selection will only active before system starts. While system running, the selected mode will
not be active and the selection will stay remain the same.
Turn the AUTO/STOP/MANUAL switch to AUTO. The GENERON NPU Membrane System
will initiate a Startup state. During the Startup state, which lasts for 10 minutes until the
─────────────────────────────────────────────────────────────────
GENERON NPU Membrane Nitrogen System
Operating and Maintenance Manual OCT. 2007
─────────────────────────────────────────────────────────────────
temperature requirements are satisfied, the PLC will direct the N2 to vent off spec valve
FSV-24-1B. Once the temperature requirements are reached, the product valve FSV-24-1A
switches and directs the N2 to product.
Once there is pressure in the nitrogen product pipeline, set the sample PRV-20-1 to 0.35-
bar. Adjust the sample flow bleed off (flow indicator FI 20-1 in CC-2) to 1 SCFH (this value
may need revision to optimize for the on-site configuration).
Flow transmitter FT-15-1 will enable power to the process heater contactor when the inlet
air flow is high enough for heater control.
The GENERON NPU Membrane System will continue to vent product gas until the startup
timer times out. When the startup timer times out, the GENERON NPU Membrane System
will switch to purity control or manual control depending on the Control mode used and the
customers' product nitrogen demand.
Once the system has been pressurized, ensure that the drains are all operating properly.
Once the product pipeline O2 level falls within the preset vent limits, the PLC will
automatically direct the purity assurance valve FCV-23-1 to Open, thus placing the
GENERON NPU Membrane System On-Line.
The Operator Interface Terminal in the GENERON NPU Panel allows an operator to
change set points. The factory will set most of these, but some will occasionally
need to be set by the user. This is only the case when the unit is to operate
differently from its original set up at commissioning.
The software leads the operator through the programming procedure using menus
and sub-menus. The documents titled “Ladder Logic / Operator Interface Menu
Screens” in Section 10 shows the flow of these screens as they appear to the user.
The following procedure is used for all programming of the NPU Controller:
After power is applied to the PLC, there is a programmable warm-up period. At the
end of that period, use the “F” keys to navigate to all screens.
7.6.4 Screens
Purity Setpoint can be changed from .01 to 7%. Running Oxygen set points between 0.6
and 0.1 may require a change from series module configuration to parallel.
─────────────────────────────────────────────────────────────────
GENERON NPU Membrane Nitrogen System
Operating and Maintenance Manual OCT. 2007
─────────────────────────────────────────────────────────────────
SECTION 8
SHUTDOWNS
This section is intended to instruct an operator on procedures for shutting down the GENERON
NPU Membrane System as well as afford a means to identify a non-operator initiated shutdown.
. Isolate the GENERON NPU Membrane System from the feed air source
It is generally advisable to close the final product valve (if applicable) to keep the
product receiver (if supplied) at pressure.
Any time there has been a non-user system shutdown, the problem must be fixed
and the AUTO/STOP/MANUAL switch must me moved to STOP before the system
can be re-started.
If instrument air pressure drops, but is not lost completely, the valves will operate
sluggishly, not seal properly, etc. A high oxygen vent may eventually occur.
─────────────────────────────────────────────────────────────────
GENERON NPU Membrane Nitrogen System
Operating and Maintenance Manual OCT. 2007
─────────────────────────────────────────────────────────────────
SECTION 9
9.1 General
The GENERON nitrogen generator is provided with two Fujikora FCX-MVL-AC oxygen
percent analyzer transmitter, which is an integral part of the purity control system: One (1)
process nitrogen gas analyzer AT-20-1 and one (1) room oxygen analyzer AT-1-1. The
analyzers have been calibrated and tested by the factory prior to shipment. Although the
analyzers should not require any additional calibration, a certified oxygen content span gas
should be used to verify the calibration once every month.
Inside the control panel is a Span / Sample selection valve for AT-20-1. The pressure of the
span gas needs to be regulated with PRV-20-2 to 5 PSI (0.35 bar). The span gas used to
check (verify) the analyzer should have an oxygen content around the contract purity. The
nominal acceptable variation between oxygen span gas content and the analyzer reading is
+/- 0.20 (% O2).
The room analyzer AT-1-1 is also mounted inside the control cabinet CC-2 and detects low
oxygen concentrations inside the membrane system container. Therefore the analyzer is
mounted through the bottom of Control Panel CC-2 and the sensor is exposed to the
atmospheric air inside the container.
Calibration procedures are found on the following pages in this section. It is useful and
important to log the analyzer verifications and calibrations.