Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

RF

FR-turbilatex
Turbidimetria Látex

PREPARACIÓN Y ESTABILIDAD TOYOBO_40 / FURUNO_CA400


Ref. 1107105. Ambos reactivos se encuentran listos para su uso. Son estables hasta la fecha de caducidad indicada en el envase.

CALIBRACION

Utilizar el calibrador de FR incluido en el kit.

CONDICIONES OPERATORIAS

Introducir en la memoria del ordenador los siguientes parámetros:

CHEMISTRY PARAMETER

Method * Name RF Reagent Name Reagent Water (µL)


(µL)
Print Name RF R1 RF 240
Sample SERUM R2 X Enable RF 60
Unit UI/mL
Assay Type End Point Diluent Enable

Mesuring Points Decimal Points 2


Start End
1 43 44 Normal Range
x Enable 2 55 56 Nº Range Name Min. Max.
1 * * *
Wave Lenght 2 * * *
Prim. 600 Sec x Disable 3 * * *
4 * * *
Normal
Sampling Sample (µL) Diluent (µL) LOW HIGH
Dilution 3 Technical (Conc.) * *
Rerun (High Prozone) Range (mAbs/10) -999999 999999
Dilution
Rerun (Low) SPT Wash Enable Reag.
Name
Dilution Stirring Speed R1-Middle R2-Middle

Limit Checks Blank measurement


Duplicate Limit mAbs/10
Sensitivity Limit mAbs/10 Blank measurement Enable Reagent Blank
Linearity Limit % (mAbs/10)/min Meas. of reagent BK during run Next Run
Prozone Limit Reag. BK. meas. at calibration Reagent BK (system water)
SL1-S SL1-F Nº of measurement Duplicate
SL2-S SL2-F
SENS mAbs/10 Reagent Blank Limit Checks:
Duplicate Limit mAbs/10
x Absorbance limit
Reaction increase Instrument Factor
Limit 25000 mAbs/10 A: 1.00 B: 0.00

• CALIBRATION: SPLINE

Ed. 2018

SPINREACT, S.A.U. Ctra. Santa Coloma, 7 E-17176 SANT ESTEVE DE BAS (GI) SPAIN
Tel. +34 972 69 08 00 Fax +34 972 69 00 99 e-mail: spinreact@spinreact.com
RF

RF-turbilatex
Latex Turbidimetry

PREPARATION AND STABILITY


TOYOBO_40 / FURUNO_CA400
Ref. 1107105. Both reagents are ready to use.
The reagents are stable until the expiration date stated on the label.

CALIBRATION

Use the calibrator included in the kit.

OPERATORY CONDITIONS

Introduce the following data in the computer:

CHEMISTRY PARAMETER

Method * Name RF Reagent Name Reagent Water (µL)


(µL)
Print Name RF R1 RF 240
Sample SERUM R2 X Enable RF 60
Unit UI/mL
Assay Type End Point Diluent Enable

Mesuring Points Decimal Points 2


Start End
1 43 44 Normal Range
x Enable 2 55 56 Nº Range Name Min. Max.
1 * * *
Wave Lenght 2 * * *
Prim. 600 Sec x Disable 3 * * *
4 * * *
Normal
Sampling Sample (µL) Diluent (µL) LOW HIGH
Dilution 3 Technical (Conc.) * *
Rerun (High Prozone) Range (mAbs/10) -999999 999999
Dilution
Rerun (Low) SPT Wash Enable Reag.
Name
Dilution Stirring Speed R1-Middle R2-Middle

Limit Checks Blank measurement


Duplicate Limit mAbs/10
Sensitivity Limit mAbs/10 Blank measurement Enable Reagent Blank
Linearity Limit % (mAbs/10)/min Meas. of reagent BK during run Next Run
Prozone Limit Reag. BK. meas. at calibration Reagent BK (system water)
SL1-S SL1-F Nº of measurement Duplicate
SL2-S SL2-F
SENS mAbs/10 Reagent Blank Limit Checks:
Duplicate Limit mAbs/10
x Absorbance limit
Reaction increase Instrument Factor
Limit 25000 mAbs/10 A: 1.00 B: 0.00

Ed. 2018

SPINREACT,S.A.U. Ctra.Santa Coloma, 7 E-17176 SANT ESTEVE DE BAS (GI) SPAIN


Tel. +34 972 69 08 00 Fax +34 972 69 00 99 e-mail: spinreact@spinreact.com