ALBUMIN

Total proteins
Biuret

PREPARATION AND STABILITY OLYMPUS AU-480


Ref. 1001020. The reagent is ready to use.
This solution is stable at 2-8ºC up to the date of expiration specified.
Avoid direct sunlight.

CALIBRATION
Use a human serum calibrator (CFAS Ref. 1002011)

OPERATORY CONDITIONS
Introduce the following data in the computer:

SPECIFIC TEST PARAMETERS

Test Name TP Type Serum Operation Yes

Sample Volume 8.0 L Dilution 0 L

Reagent R1 Volume 300 L Dilution 0 L


R2 Volume L Dilution 0 L

Wavelength Pri 540 Sec


Method END
Reaction Slope +
Measuring point 1 First 0 Last 27
Measuring point 2 First Last
Linearity %
No-Lag-Time
Pre-dilution Rate

Min OD -0.100 Max OD 2.500

Reagent OD Limit
First L -0.100 First H 2.500
Last L -0.100 Last H 2.500
Dynamic Range
L 0.00 H 15.00
Correlation Factor
A 1 B 0

On-board stability period 30 days

Value/Flag * Level L * Level H *

Normal Ranges Age L Age H

Panic Value L H Unit g/dL

Cal. Type AB Formula Y = AX + B Counts *

Cal. No OL Conc. Factor OD-L Factor OD-H


Point 1-7 * **

LINEARITY

This method is linear up to 15 g/dL of Total proteins.


Ed. Febrero 2006

SPINREACT,S.A. Ctra. Santa Coloma, 7 E-17176 SANT ESTEVE DE BAS (GI) SPAIN
Tel. +34 972 69 08 00 Fax +34 972 69 00 99 e-mail: spinreact@spinreact.com