A Clinical Guide

To
Aarhus Mini-implants and Skeletal Anchorage
“Conventional anchorage is built on the following principles: More teeth
Index Page

against fewer teeth, maximum rigidity, differentiated force systems, Historical Background 1
extra-oral anchorage, inter-maxillary elastics, reinforcement of occlusion,
and, sometimes in adult patients, the use of dental implants. Indications For Use 2

Skeletal anchorage can be used where conventional anchorage cannot Contra-Indications 2


be applied, but should not be a replacement of conventional anchorage.
In patients with missing teeth and having reduced periodiontium, Patient Profile 2
skeletal anchorage provides a framework for the regeneration of the
Selecting A Skeletal Anchorage System 3
alveolar process by the movement of teeth into the edentulous regions, thus improving the
rehabilitation possibilities for the patient. Patients presenting a therapeutic need for tooth
Mini-implants/Screws 3
configurations inconsistent with the normal movement, such as the need for the displacement
of multiple teeth in the same direction, asymmetric displacements, or borderline surgical cases, Thread Type – General Characteristics 3-4
are also candidates and can benefit greatly from skeletal anchorage. Mini-implants (TADs), when
used as either direct or indirect skeletal anchors, can provide absolute anchorage for a Intraosseous Components 4-6
multiplicity of tooth movements. They are easily inserted by the orthodontist and, in almost all Form
cases, loaded immediately. Their success rate is depended on the system used, the operator Diameter
(technique and experience), and the patient (multiple factors).” Length
Cut
- Professor Birte Melsen Tip

Extraosseous Components 6
Transmucosal Collar
Neck
Head Designs
6-7
Driver System
7
The Sterilization Tray

Screw Management and Organization 7

Placement and Insertion Protocols 7-8


Patient Information

Preparation for Insertion


Selecting the Proper Screw for the Case 8-9

Removal of the Mini-implants 9

Appendix A 10
Appendix B 11
References 12 - 13
Historical Background
The need for skeletal anchorage in orthodontics arose in response was born. Soon after, the possibility of replacing headgear with
to the increase in the numbers of adult patients presenting for Mini-implants was introduced by our Korean colleagues, who,
treatment. Case workups and analysis disclosed that a large in contrast to the general tendency in the profession, realized
number of these patients could not be treated by conventional a need for establishing maximum “anchorage in the treatment
anchorage means. Frequently, in a certain percentage of these of four premolar extraction cases for the purpose of flattening
cases, the number of teeth available was insufficient for the the profile”. Since then, a large and ever growing number of
establishment of the classic treatment strategy of using “fewer anchorage screws have been introduced and the use of skeletal
teeth against more teeth”. Additionally, the force systems nec- anchorage is, according to a recent survey in the Journal of
essary to obtain the desired tooth displacements would have Clinical Orthodontics (Buschang et al., 2008), a commonly used
generated undesirable movements of the reactive units. In other anchorage system.
cases, treatment required that all of the teeth be moved in the
same direction, a treatment goal that could not be fulfilled by an Development of the Aarhus Mini-implant system was initiated
orthodontic appliance in balance. in 1996 by Professor Birte Melsen at the University of Aarhus,
based on her prior studies of the use of skeletal anchorage for
As early as the nineteen seventies, maxillary anterior teeth had orthodontic treatment. The primary elements of the system as
been retracted against a surgical wire anchor inserted into a defined in her initial design requirements are described in the
hole drilled through the inferior part of the zygomatic arch. This following sections. The current state of the art represents the
method of skeletal anchorage, however, was limited to use in conclusions drawn from the information gathered over the past
those patients who only required retraction of maxillary teeth 12 years of many mechanical and clinical tests, and various
and thus it did not fulfill the anchorage requirements related to experiments related to the surgical screws; the result of a
a much broader range of tooth movements needed in the entire collaborative effort between two parties – Aarhus University and
adult patient population. An alternative to the zygoma wire was the Medicon eG.
use of palatal implant as introduced by Werhbein and Glatzmaier
(1996). Both the palatal implant as well as the “onplant” met the The company responsible for the manufacture of the system,
requirements for anchorage for retraction of maxillary anterior Medicon eG, is highly specialized in the production of surgical
teeth, but these were introduced primarily as an alternative to screws, instruments, and components for the field of craniofacial
extra-oral anchorage and other types of anchorage requiring surgery and has been a leading manufacturer in this field for many
patient compliance. years. The intraosseous component of the Aarhus system is
derived from Medicon’s knowledge, expertise in the arena of
Dental implants have been frequently used as anchorage in orthognathic surgery, and ongoing research of the basic param-
orthodontic patients when the implants are planned as an eters related to screw design and function. The extra-osseous
integral part of a full dental reconstruction (Odman et al. 1994). components of the Mini-implants have been designed by
The same type of implants, but with a slightly different design, Professor Melsen to meet specifically defined requirements related
were introduced as anchorage for the anterior displacement of to the goal-oriented orthodontics characterizing the Orthodontic
lower second and third molars following loss of Department at the Dental College of Aarhus University, Denmark.
the first molar due to a variety of reasons (Rob-
erts). The use of these small osseointegrated
implants allowed tooth movements that had
not previously been possible, i.e. movement of
teeth mesially without introducing any forces to
the reactive units.
The first use of a surgical screw as anchorage
was described by Creekmore in a case report
of a single patient but this did not immediately
attract a lot of attention (Creekmore and Eklund
1983). Much later, Kanomi demonstrated the
use of a small screw as anchorage for the
displacement of the lower incisors; Costa et
al., also used a surgical screw, primarily as an
easier alternative of the zygoma wire. Thus
the temporary orthodontic anchorage device
The HEAD is the terminal segment of the screw and is the
Indications For Use not be considered as either prudent or in the best interests of
the patient. There are, however, a number of conditions that
Selecting A Skeletal Anchorage System “working “end for the orthodontist. The Aarhus system has three
absolutely preclude their use. Chief among these contraindications distinct HEAD designs that are able to receive force modules of
Skeletal anchorage has, to a large degree, replaced conventional Screws are naturally at the very core of the (Mini-implant) skel-
are (Appendix B): varying types. These three HEAD designs provide the orthodon-
anchorage in situations where anchorage is considered either etal anchorage system. When evaluating a system it is natural,
• Patients who are suffering from metabolic bone diseases tist a complete range of options when planning and selecting the
critical, insufficient, or likely to result in undesirable side effects therefore, to focus closely on a detailed examination of the
• Patients receiving immune suppressive therapy biomechanics for each specific case. These general descriptions
such as vertical displacements generated by inter-maxillary force screw(s). One should be cautious, however, not to pay atten-
• Patients on chronic steroid or bispfosphonate medication for each segment: BODY, COLLAR, NECK, and HEAD will be
systems. Another frequently found indication for its use is in tion to the screws to the point of either ignoring completely or
• Patients who are incapable of following the instructions discussed below in greater detail.
cases of non-compliance. While many so called “compliance paying too little attention to the supporting elements (drivers,
for postoperative care. Causes for this are, for example, screw management and handling, delivery of instrumentation,
free” anchorage systems have been introduced for orthodontic
treatment, none of these have proven to deliver the absolute
patients with psychological/mental or neurological problems
• Patients with bone tissue either insufficient or of poor
etc.). It seems fair, then, that, in the decision making process, a Material Specifications
anchorage generated by skeletal anchorage systems. Finally but more prudent and rewarding approach is to examine a system
quality to provide primary stability in its entirety:
perhaps most importantly, skeletal anchorage has widened the Aarhus mini-implants are manufactured from a biocompatible
• Patients with circulatory disturbances or latent infections
spectrum of orthodontics allowing the orthodontist to perform titanium alloy composition, “Wrought Titanium-6 Aluminium-4
• Patients with hypersensitivity to specific materials, i.e. 1. Is it simple but complete?
treatments that could not, or, only with great difficulty, otherwise Vanadium ELI (Extra Low Interstitial)”, according to the ASTM
who react to foreign bodies. In these cases, appropriate 2. Does it offer options that apply to all clinical situations?
be done with conventional mechanics (Melsen 2004; Melsen F136-02 standards specification. In the medical field of Cranio-
tests are essential prior to implant (even if merely 3. What does the “package” look like?
2005a; Melsen 2005b; Melsen & Dalstra 2005; Melsen & Garbo Maxillo-Facial surgery, this titanium alloy is the material of choice
suspected) 4. Is it thoughtfully laid out?
2004). In general then, it can be stated that skeletal anchorage and has been used for several decades by all established
• Acute infections 5. Is the chair side delivery system user/patient friendly?
is indicated where the forces acting on the reactive units are manufacturers of titanium implants. Medicon eG, the manufac-
• Radiotherapy in the head region and patients with tissue 6. What about screw management-----packaging,
undesirable and cannot be neutralized by occlusal forces. turer of the Aarhus system, has been producing screw implants
damaged by radiation storage, delivery, tracking? according to the ASTM F 136-02 standard since the early eight-
• Recurring diseases of the oral mucosa and poor oral
The most common clinical indications are: ies. Furthermore, this specific titanium alloy composition covered
hygiene These questions and other details related to the physical
• Insufficient number of teeth and/or lack of occlusion in has been used for human implants, in contact with bone and
the anchorage unit, e.g. patients with agenesis or who makeup and practical use of the Aarhus System are described soft tissues, for more than 3 decades.
have lost teeth for various reasons. Patient Profile in the sections that follow.
The alloy exhibits a well documented level of biological inertness
• Extrusion or intrusion of single teeth or units of teeth
without antagonists (no opposing vertical forces acting
Patients as young as 10 years of age can benefit from treatment
utilizing skeletal anchorage. Great care, however, should be
Mini-implants/Screws being characterized as:
• Corrosion Free
on them). taken not to insert a Mini-implant in areas where teeth are either • Non Toxic
Using an analogy in anatomical
un-erupted or where root formation is still developing. Screws HEAD • Strong
terms, Aarhus screws can be
• Asymmetric tooth movements: the displacement of teeth inserted into areas of actively resorbing deciduous teeth have a NECK • Having a low module of elasticity
described as having four distinct
in a single direction, as in the case of A-anchorage, large risk of failure due to the active remodeling of the adjacent • Anti-magnetic
yet contiguous segments or zones:
C-anchorage, unilateral expansion; or the displacement bone. The upper age range, on the other hand, is set not by COLLAR
BODY, COLLAR, NECK, and HEAD
of all teeth, in either an arch or quadrant, in the same
direction.
chronological age but by the presence of cortical bone thickness
with a minimum of 1.0 mm (Fig 1). In cases of cortical bone
(Fig 2). Thread Type
thickness of less than 0.5 mm, the primary stability is dependent The screw BODY is comprised of a
General Characteristics Fig 3 Tip Detail
• Retraction and/or intrusion of anterior teeth with on either trabecular bone or bicortical bone. solid, cylindrical central shaft (core)
insufficient anchorage in the reactive unit. BODY Aarhus mini-implant screws have a self-drilling (self-cutting) thread
that terminates in a finely defined
sharp tip. Encircling and spiraling pattern by design. Self-tapping screws require a pre-drilling at full
• Mesial movement of molars in cases where the anterior length. The drilling is performed with a bur which has a slightly
around the full length of the shaft
teeth cannot afford to be retracted smaller diameter than the screw diameter and the primary stability
are the threads, formed to a specific
configuration and cut at a proscribed then depends on the pressure of the adjacent tissues against the
• Proclination of anterior teeth in cases where no posterior screw when it is manually inserted. The disadvantage is that the
depth and angle to the shaft. Fig 2 Screw Segments
anchoring element is available, or the reactive forces orthodontist/clinician does not have a clear tactile perception of
would have an adverse effect. the tissues when pre-drilling and damage to the roots can occur
The COLLAR is a smooth, polished, and thread less transitional
zone immediately extending from the screw BODY. It is the portion without the clinician realizing what has happened – It is therefore
• Space closure in maximum anchorage cases. recommended to choose a system that utilizes self-drilling screws
of the screw that resides in soft tissue. It is also cylindrical in
form with a slightly larger diameter than the BODY so as to that have a very fine sharp tip for engaging the host bone (Fig 3).
Contra-Indications provide a definite “stop” when placing screws in the host bone.
However, even the Aarhus anchorage system requires pre-drilling
It terminates in a broad, flat plate.
in cases of very thick and dense cortical bone, e.g. in the retro-
While there are many legitimate reasons to consider the use of molar area and into edentulous alveolar processes which have
mini-implants in orthodontic treatment, it may also be consid- The NECK area of the screw is another transitional area that joins
the COLLAR with the HEAD and extends into the oral cavity. It, been loaded by removable dentures over an extended period of
ered an imprudent use of them when the case can be just as time. However, the pre-drilling in these situations should not be
easily treated with conventional biomechanics. In addition, the too, is cylindrical and smooth with a diameter that is narrower
than both COLLAR and HEAD. Functionally, it serves as an “un- executed at full length; the clinician should drill a hole of only 1 to
best treatment plan should incorporate the minimum number of 2mm in depth, merely perforating the external particularly dense
mini-implants necessary to treat the case. Excessive use may Fig 1 Bone Strain and Cortical Thickness
der tie area” for the application of ligatures, elastomeric thread
and chain, or NiTi spring eyelets. layer of the cortex, before inserting the screw. When pre-drilling in

2 A Clinical Guide to Aarhus Mini-implants and Skeletal Anchorage A Clinical Guide to Aarhus Mini-implants and Skeletal Anchorage 3
an area of loose, unattached mucosa, it is of the utmost impor- Aarhus Mini-implant screws are produced in three different revealed that the center of rotation, when loaded perpendicular
tance that the mucosa is kept tight during drilling to avoid the thread diameters: 1.3 mm, 1.5mm, and 2.0mm. The 1.3mm to the long axis, corresponded to the endosteal side of the cor-
tissue being pulled around the bur. This also applies when the screw is recommended when the inter-radicular space is nar- tex, if the thickness of the cortex was 1.0mm (Fig 7.1 - 7.3).
screw is being inserted. row or in sites where the cortical bone is thin, e.g., between
the roots of teeth in the mandibular anterior region. The 1.5mm In the case of thicker cortical bone, the contribution of the
screws have application for placing in the alveolar process trabecular bone is minimal and a short screw will satisfy the
Intraosseous Components of both upper and lower jaws on the buccal side, and on the requirement for stability. If, on the other hand, the cortex is very
palatal side in the upper arch. The 2.0mm screws are used in thin, stability will be dependent on the trabecular bone and,
The intraosseous zone, the BODY, of the sites where the cortical bone is dense and thick, such as the therefore, a longer screw is required. This applies especially to a
screw has the following features: retro molar area of the lower jaw or in areas of the mandible that screw inserted inter-radicularly or in the retromolar region of the
• Form have been edentulous for a long time. Potential placement sites upper jaw. When the cortical thickness is less than 0.5 mm, the
• Diameter are shown in Fig 6.1 - 6.2. primary stability is frequently insufficient and skeletal anchorage
Body

• Length cannot be utilized with Mini-implants alone.


• Cut
Fig 9.1 Cellular adaptation to smooth,
• Tip polished surface of screw (Behrens).
Cut
Form
Aarhus mini-implant threads are cut at an angle of 11 degrees
The Aarhus Mini-implant is cylindrical in form. While in an asymmetrical pattern. This ensures a high resistance
both conical and cylindrical Mini-implants are against removal in a pull-out test (Fig 8). The thread depth
available, within the building industry it is well varies in relation to the overall screw diameter and the shaft
recognized that conical screws are reserved for diameter, e.g. for the 1.5mm diameter screws, the thread depth
use in concrete, not plastic materials, and cylin- is 0.025mm. These parameters were determined in order to op-
drical screws are used for plastic materials timize the screw to the material in which it will be cutting (bone)
Fig 6.1 - 6.2 Potential Placement Sites
such as wood and, by association, bone. and to the loading normally applied to it, either perpendicular
Consequently, screws produced for surgical or as a pull-out. The manner in which the screw penetrates into Fig 9.2 Close surface contact of bone to screw.
procedures are cylindrical in form (Fig 4, Length(s) the bone depends on the cut, the depth, and the angle of the
Appendix A). threading. With a smaller angle the surface area of the screw to the core, the higher the number of threads and the slower
Aarhus mini-implants are produced in three optional thread increases; this should, in theory, enhance stability, but it also
lengths. The production of different lengths and, hence, the the screw is inserted. The number of threads are, however, also
Diameter leads to a higher degree of trauma to the tissue. The angle cho- related to the impact on the bone when inserting the screw. The
selection of a specific length in treatment is based on the sen for Aarhus Mini-implants is the result of and optimization of
varying quality of bone found in different regions of the maxilla closer the threading, the flatter the resultant angle of the thread
Fig 4 the need for a larger surface and a minimum of trauma.
The diameter of Mini-implants (TADs) on the Cylindrical and mandible, and from patient to patient. The three different with more destruction to the surrounding bone. The sharper the
market vary anywhere from 1.1mm to 2.5mm. Shape lengths were chosen based on finite element analysis of the edges of the threads, the more precise the cut is in the bone,
The rationale for the production of the very thin stress/strain distribution related to loading of the screw in different whereas the blunter the edges, the more the pressure increases
screws is that there is a reduced risk of damaging the roots combinations of cortical bone thicknesses and trabecular bone with crushing of the surrounding bone.
when placing the screws inter-radicularly when the space be- densities. A finite element analysis of Aarhus mini-implants has
tween the roots is limited. The reduction of the diameter does, Angle 11˚ Finally, the treading can be formed either symmetrically or
on the other hand, increase the internal stress of the screw Sigma_X Depth asymmetrically. If the shape is asymmetrical, the angle is usually
during insertion and removal and thereby there is a greater risk 0.1 MPa .025mm sharper and if cut so that the oblique part is pointed towards the
of fracture as seen in Fig 5.1 - 5.2. 0.05 MPa apex of the screw and the horizontal part towards the head of
0.0 MPa the screw, it will have a higher resistance in relation to a pull-out
test in the lab or in the clinic to an applied extrusive force. An ini-
-0.05 MPa

Stress Factor for Constant Bending Movement -0.1 MPa


120
tial loading of 50cN is normally recommended. Higher loadings
can be accepted after the primary stability has been replaced by
100
Fig 7.1 Stress Distribution on cortical bone Fig 8 Cut Detail secondary stability, which will continue to increase as the density
Relative Stress Factor [%]

with a 50cN load perpendicular to the long axis. of the bone adjacent to the screw increases over time. Although
80 For a given diameter, the deeper the thread, the thinner the the screws may initially resist loosening when loaded with a mo-
60
Micro core diameter and, thereby, the more fragile it becomes. On the ment following a period of healing, it can be anticipated that the
Strain
other hand, the deeper the thread becomes, the greater the
Sigma_Y
_ 1.0000
10 shearing forces developed in relation to the type of loading will
increase in the intraosseous surface area of the Mini-implant.
_ 0.7500

40
most likely lead to loosening.
_ 0.5000
5
The surface of most Mini-implants are machine polished and
_ 0.2500
_ 0.0000
0
20 _ - 0.250
_ - 0.500
-5 the bone will adapt closely to the screw surface rendering an
Tip
_ - 0.750

0
_ - 1.000 -10 Fig 7.3 Stress osseointegration on a histological level (Fig 9.1 - 9.2).
0.75 1.00 1.25 1.50 1.75 2.00 2.25 concentration at
Fig 5.2 Stress gen- tip during insertion.
Diameter [mm] erated on removal. The angle of the cut in relation to the core determines how The tip is cut very fine. Therefore, it should not be forced as it
Note: If the neck is
Fig 5.1 Internal Stress vs. Diameter: increasing weakend, e.g., by a Fig 7.2 Strain (deformation) during the same many threads there are and also the rate with which the screw could slide or skip on the surface of the bone at the intended
the diameter from 1.0mm to 1.3mm results in a cross slot, it will be loading parameters. This strain will result in
50% reduction in stress prone to fracture. increased density of the trabecular bone. is inserted into the bone at each turn. The more perpendicular insertion site. This may cause a bending or fracture of the tip.

4 A Clinical Guide to Aarhus Mini-implants and Skeletal Anchorage A Clinical Guide to Aarhus Mini-implants and Skeletal Anchorage 5
It is therefore important that, when inserting the screw, the tip and elastic thread or chain. In the smooth or dome topped head systems. Equally important as screw pick-up, a simple reversal
should not be held at too oblique an angle to the surface (90-60 design, there is an .032in. (0.8mm) single round hole that passes of the sleeve smoothly and easily releases the blade tip from the Screw Head Thread Collar
degrees). Any necessary change in angulation can be made through the neck (Fig 11). This hole will accommodate round screw head. This feature helps to avoid unnecessary tugging Image Length Length LOT Number
after the screw has established a purchase in the host bone. wire hooks, ties, or auxiliaries as desired by the clinician. at the driver to release the screw, a step that could disturb the
immediate area of surrounding bone at the placement site and
Extraosseous Components Head Design(s) compromise primary stability (Fig 13).

Screw Part
The extraosseous zone of the screw is comprised of: Aarhus Mini-implants are produced with three different head de-
signs: 1. Button Head, 2. Round Head with a single round hole
The Sterilization Tray Number
Base
• Transmucosal Collar through the neck, 3. Bracket Head with two .022” x .025 “ slots
The Aarhus Sterilization Tray features
• The Neck oriented perpendicular to each other (Fig 12).
a compact design to minimize its
• Head Design(s)
footprint on the surgical cart or unit. It
is lightweight and roomy for efficient
Transmucosal Collar chairside delivery. The sliding cover Fig 16 Screw Cassette & Holding Base
plate locks securely to hold instru- Fig 14 Sterilization Tray
The Aarhus Mini-implants have smooth transmucosal collars in
Placement and Insertion Protocols
ments and screws in place. It is fully - Internal
order to reduce the risk of accumulation of plaque and sub- vented for superior steam circulation
sequent infection. This zone of the screw that passes through and drainage. Built-in screw cassette slots and multiple wells Patient Information
the soft tissue should be smooth and not threaded to mini- are thoughtfully laid out for all instrumentation needs. The
mize irritation and inflammation. The Aarhus mini-implants are ‘Handle Well’ has been designed to accept either screwdriver The anatomical site should be chosen so that the risk of root
produced with two different collar lengths 1.5mm and 2.5 mm version; the slim-line or the large grip handle. Solid one-piece contact is minimized and the clinician should avoid inserting
to accommodate varying tissue thicknesses in the upper and construction will provide years of dependable, trouble-free, screws in areas where nerves or vessels are located, e.g. the
lower jaws (Fig 10). Fig 12 Screw Heads: Bracket Head, Button Head, Through-Hole Head
service (Fig 14). posterior lateral part of the palate. Prior to considering the use of
The heads of Mini-implants on the market vary between those skeletal anchorage, it is important to assess its possibilities care-
The collar has been purposely with buttons or hooks allowing only a one point contact. Heads fully. Mini-screws, although generally well accepted, require that
1. Sliding
manufactured with a slightly larger penetrated by holes in one or two directions allow for the the unique biological environment of the individual patient has to
Cover
diameter than the body (threaded insertion of a wire thus providing two point dimensional control be thoroughly evaluated and understood. The bone turn-over of
section). This difference in diameter 1.5mm
while bracket designs allow a three point dimensional control. the patient may be influenced by factors that will interfere with
is important as it allows the clinician 2.5mm 2. Color Screw the normal tissue reaction surrounding the screw.
The latter also make it possible to utilize the screw as indirect Cassette
to feel when the collar has reached anchorage by consolidating the screw to a tooth with a full size
the periosteum during insertion. At stainless steel wire auxiliary. The tooth so consolidated to the A thorough exam and Medical History is required to rule out
this stage the insertion is complete 3. Base any contra-indications to treatment. If no contraindications are
screw can then be used as anchorage by conventional appli-
and further turning of the screw will ances and biomechanics. found, the patient must be informed about the advantages, oral
lead to loosening. In relation to Mini- Fig 10 Collar Lengths hygiene requirements, and the possible risk-factors before the
implants where the neck has the same Fig 15 Screw Magazine
start of treatment. This can be provided in either written form or
diameter as the threaded part, the clinician runs the risk of also Driver System verbally presented, e.g. by the hygienist or other qualified person
inserting the smooth neck into bone, resulting in a large pres-
sure zone and eventual ischemia of the bone surrounding the Unique to Aarhus, the driver blade incorporates a threaded Screw Management and Organization showing clinical images of patients with mini-implants. In either
case, a signed Consent Form for Treatment with Mini-screws
collar. In addition, the collar of the Aarhus mini-implant widens locking sleeve to assist in the engagement, pick-up, and delivery should be obtained from the patient/parent.
of screws to the host site. Simple advancement of the sleeve Aarhus screws are packaged and shipped in self-contained,
gradually towards the head and ends in a broad flat plate that
to the end of the driver blade securely captures screws and autoclavable units termed Screw Magazines. There are five (5)
suppresses tissue overgrowth so that the mucosa is not subject In order to get the first “take” in the bone, the point of the screw
allows them to be easily removed from either sterilization tray screws, of a single type, per magazine. Each magazine is actu-
to undue pressure by the plate carrying the head. This design is sharp and the point very fine, and the most “apical” part of
or cassette. Once so secured, the screws can be confidently ally a three part unit: sliding cover, base, and removable colored
facilitates the maintenance of healthy peri-implant tissues. the screw is therefore conical. [NOTE: It is important not to load
carried and driven without “mini-screw wobble” seen in other Screw Cassette (Fig 15). With the cover on, each screw is
maintained in its individual well and cannot fall out of place.
Neck Screw magazines provide safe, secure, hygienic handling, with
convenient storage and office to office transfer of screws. The
The neck of the screw is a short
colored screw cassettes are laser marked with complete and
NECK 0.032” cylindrical area that joins the
instant identification of contents (Fig 16). The ID markings
tissue collar with the head of
provide positive tracking and traceability, in addition to simplify-
the screw. This area is narrower
ing inventory control. Screw cassette colors are keyed to screw
in diameter than both the collar
diameter: WHITE for 1.3mm diameter, BLUE for 1.5mm diam-
and head; it provides an “under-
eter, and RED for 2.0mm diameter screws. In normal usage,
cut” area for the attachment of
the colored screw cassettes are removed from their magazine
various force modules selected
base and transferred into the sterilization tray. Alternatively, the
Fig 11 Round Hole by the clinician: the keyends of
magazine, taken together with its companion holding base, can
.032in Diameter NiTi coil springs, steel ligatures, Fig 13 Driver System, Blade and Sleeve function as a stand-alone “mini” delivery system. Fig 17.1 Sterilization Tray Fig 17.2 Tray Cover

6 A Clinical Guide to Aarhus Mini-implants and Skeletal Anchorage A Clinical Guide to Aarhus Mini-implants and Skeletal Anchorage 7
the screw hard against the bone. In such cases the point can be Thread length selection-In patients with a cortical thickness of the occlusal surface in the proximity with a light cured acrylic 4. A sterile surface has to be available and prepared to
bent or even fractured. When starting the insertion, the perora- 1mm or more, a 6mm screw may provide sufficient primary (Fig 20.1 – 20.3). While the PA radiograph with the template in receive all of the surgical armamentarium determined
tion of the periosteum and the initial introduction into the exter- stability. When the cortical bone is thin and the primary stabil- place provides an excellent guide for the insertion of the screw, necessary for placement of the screws.
nal layer of the cortex should be made perpendicular to or, as ity is primarily dependent on trabecular bone, a Mini-implant information about the cortical thickness cannot be obtained by 5. The doctor and assistants prep according to estab-
close to, perpendicular to the bone; once the screw feels stable, with longer thread length (8mm or more) may be indicated. In either OTP or PA radiographs; therefore, the doctor is referred lished protocols for oral or periodontal surgery (surgical
the direction can be changed to a more apically directed angle.] edentulous areas where the cortex may be very thin, bi-cortical to surveys provided in the literature (Costa). scrub, surgical head cover, mouth mask, and sterile
anchorage may be the only alternative and should be consid- gloves).
ered. It has been shown (Brettin et al. 2008, Buschang, Carillo, 6. The screw is picked up with the screwdriver, the grip
Ozenbaugh & Rossouw 2008) that bi-cortical anchorage is on the screw tightened, and the insertion performed
superior to uni-cortical anchorage. If planning for bi-cortical (Fig 22).
anchorage, both the total thickness of the alveolar processes
Fig 21.1 Endo file with stop
and the thickness of both labial and lingual mucosa should be When starting the insertion, the screw is kept perpendicular, or
measured and the insertion direction accurately determined close to perpendicular, to the bone surface. Once the screw is
(Fig 19.1 - 19.2). [NOTE: Perforation of the lingual cortex and felt to have engaged the bone, the direction can be altered to a
trauma to the periosteum will frequently lead to inflammation more apical direction. Keeping this direction steady, the screw
and loss of the screw, therefore one must be very precise when driver is slowly and continuously turned until the clinician feels
attempting bi-cortical placement.] an increase in resistance; a sign that the transmucosal collar
Fig 21.2 Perio Probe with 2mm markings has reached the periosteal surface. At this point, the turning
Site Selection - Once the most desirable insertion site has been of the screw is completed; continued turning would lead to a
determined based on the force system selected, the possibility loosening of the screw.
of actually being able to insert the screw in that site has to be A placement site with attached gingiva is much preferred. If this
verified based on a radiographic image. The optimal information is not possible due to the low border between attached gingival Following insertion, the peri-implant tissues are gently rinsed
1.3mm 1.5mm 2.0mm and the free mucosa, screws can be inserted in the mucosa
can be obtained from CBCT slices through the region of interest. with sterile saline solution before the screws are loaded. It is
Fig 18 Screw Diameters If CBCT imagery is not available, PA (Peri-apical) intraoral radio- and then covered so that only a coil spring or a wire passes recommended that the initial load not exceed a maximum force
graphs are preferable to panoramic radiographs as the latter can through the mucosa. [NOTE: It is important that the mucosa of 50cN and that the force be applied perpendicular, or as near
distort the mutual position of the roots, especially in the canine is held tight when the screw is inserted in order to avoid the to perpendicular, to the long axis of the screw as possible. An
Preparation for Insertion region. One recommendation is to bend a small wire template mucosa winding around the screw.] intrusive or extrusive pull-out force can also be accepted imme-
and to fixate this to either a bracket with a ligature wire or to diately following insertion; on the other hand, a moment in either
Selecting the Proper Screw for the Case Soft tissue consideration - Not only the bony cortex but also the the clock or counter clockwise direction will lead to a shearing
Fig 19.1 Uni-cortical Anchorage thickness of the mucosa has to be considered in the selection force resulting in a loosening of the screw with a loss of primary
Prior to the actual placement of any Mini-implant, it is mandatory of the proper screw – the lengths of both thread and transmu- stability. It is important to instruct the patient in proper care and
that the desired tooth movements be defined in all three planes cosal collar. Because it is desirable to avoid the presence of oral hygiene related to Mini-screws before dismissing them.
of space to insure a successful treatment result. Only with com- any threads in the soft tissues, the threadlength of the screw
prehensive and complete planning can sufficient information be has to be chosen giving due consideration to the mucosal
obtained of the specific anchorage requirements and force sys- thickness at the placement site.
tem necessary to treat the case. With this available information
at hand, the decision is then made as to whether the anchorage Selection of the proper transmucosal collar length is determined
screw is to be used directly or indirectly (through consolidation Fig 19.2 Bi-cortical Anchorage
by the mucosal thickness e.g., in areas with a thicker mucosa,
of screw and tooth which, as a unit, will provide the anchorage). such as in the palate or in the retromolar area, a longer trans-
mucosal collar type should be selected. The thickness of the
Head selection-The manner in which the skeletal anchorage mucosal can be assessed using either ran endodontic file with
is used, direct or indirect, will determine the type of head to rubber stop as a marker, or a perio probe with millimeter mark-
select. If indirect anchorage is indicated, (meaning that the point ings. In cases where the mucosa exceeds 3mm in thickness Fig 22 Insertion
of force application is not identical to the screw head) a Mini- e.g. in the palate, Mini-implants should not be considered for
implant with a bracket like head is indicated. skeletal anchorage as the point of force application will be too
far from the centre of resistance of the screw and the potential
Diameter selection-The diameter of the screw to use depends for failure is significantly increased. (Fig 21.1 – 21.2).
on the placement site. Between roots in the upper jaw a 1.3mm
screw of 11mm is recommended as the stability is also depen- Once the placement site has been determined:
dent on the trabecular bone. In the alveolar process the 1.5 mm Fig 20.1 Template on typodont 1. The patient is asked to rinse thoroughly (2 min.) with Removal of the Mini-implants
diameter with a shorter length is recommended. In areas with a.2% clorhexidine mouth wash.
very thick cortical bone, as in the retromolar area of the man- 2. Local anesthesia of the mucosa at the insertion site is The removal of the Mini-implants is easily accomplished with
dible, it is recommended that a 2.0mm screw be used, as the administered either by an injection of .5ml local anes- the same screw driver as used for insertion. The procedure can
torsion necessary for insertion in this region may lead to fracture thetic or by an alternative anesthesia of the mucosa. usually be done under application of a local anaesthetic gel.
of a thinner screw. Aarhus Mini-implants, like other screws from 3. Following the administration of anesthesia, the doctor The removal site should be gently swabbed with a 0.2%
the manufacturer, are color coded for positive visual identifica- has to prepare for the screw insertion procedure ac- Chlorhexidine. The wound present at time of screw removal is
tion. The 1.3mm diameter is YELLOW, the 1.5 mm is BLUE and Fig 20.2 Template positioned Fig 20.3 PA film of template cording to established sterile field standards estab- minimal and usually closes within a few days. In most cases,
the 2.0 mm is GREY (Fig. 18). in place with acrylic lished for intraoral surgery. healing will continue uneventfully.

8 A Clinical Guide to Aarhus Mini-implants and Skeletal Anchorage A Clinical Guide to Aarhus Mini-implants and Skeletal Anchorage 9
Bone Types file:///C:/Documents%20and%20Settings/bmachata/My%20Documents...

Appendix A Appendix B
Bone Types Contra-Indications: Complications and side-effects:
There are four types of bone in the human face and the length of treatment for placing and restoring implants with a “tooth” and crown depends on The Aarhus Mini-implant System must not be used when the • Complaints of pain, abnormal sensations or palpability
which type of bone the implant is placed in. Implants have to integrate with the surrounding bone before a tooth and crown is placed on it. following contraindications exist: of the implant.
There are four types of bone in the human face and the length of treatment for placing
and restoring implants with a "tooth" and crown depends on which type of bone the • Patients who are incapable of following the instructions • Material hypersensitivity of the patient due to the foreign
Type I bone is comparable to oak wood, which is very hard and dense. This type of bone has lessisblood
implant placedsupply thanhave
in. Implants all of
to the rest of
integrate the
with thetypes of
surrounding bone before a
bone. The blood supply is required for the bone to harden or calcify the bone next to the implant. Therefore, it takes approximately 5 months for this
tooth and crown is placed on it. for postoperative care; for example, patients with bodies in the form of allergic reactions.
type to integrate with an implant as opposed to 4 months for type II bone. psychological/mental or neurological problems. • The use of different materials may cause corrosion.
• Patients with insufficient or poor quality bone tissue. • Increased reaction of the connective tissue in the area of
Type II bone is comparable to pine wood, which isn’t as hard as type I. This type of bone usually takes 4 • Patients with circulatory disturbances or latent infections. the implant.
months to integrate with an implant. • Patients with hypersensitivity to specific materials, i.e. • Inadequate bone formation, osteolysis, osteomyelitis,
who react to foreign bodies. In these cases, appropriate osteoporosis, inhibited revascularization or infection
Type III bone is like balsa wood, which isn’t as dense as type II. Since the density isn’t as great as type II, tests are essential prior to implant (even if merely alignment that may result in loosening, bending, crack-
it takes more time to “fill in” and integrate with an implant. 6 months time is suggested before loading an
suspected). ing or breakage of the implants.
implant placed in this type of bone. Extended gradual loading of the implant can, however, improve the
bone density. • Patients with acute infections. • Breakage, bending, migration, or loosening of the implant.
• Patients having received Radiotherapy in the head • Inadequate osseointegration that could result in loosen-
Type IV bone is comparable to styrofoam, which is the least dense of all of the bone types. This type region and patients with tissue damage resulting from ing or breakage of the implant and failure of the
takes the longest length of time to integrate with the implant after placement, which is usually 8 months. radiation treatment. orthodontic treatment.
Additional implants should be placed to improve implant/bone loading distribution. Incremental loading of • Recurring diseases of the oral mucosa and poor oral • Decrease of bone density as a result of stress shielding.
the implants over time will improve bone density. Bone grafting or augmentation of bone are often required. hygiene.
Bone expansion and or bone manipulation can improve initial implant fixation. This surgical procedure may cause not only the above-mentioned
side-effects and complications, but also problems such as
Type I bone is comparable to oak wood, which is very hard and dense. This type of injuries to nerves, infections, pain etc., which are not necessarily
bone has less blood supply than all of the rest of the types of bone. The blood supply is
required for the bone to harden or calcify the bone next to the implant. Therefore, it
caused by the implant. If complications occur, they are often the
result of incorrect selection of the patient, inexperience, or lack
Bone Types and Locations
takes approximately 5 months
for this type to integrate with an implant as opposed to 4 months for type II bone.
of preoperative planning rather than caused by the implant itself.
Lekholm and Zarb explain the classification system of bone as follows4: Based on its radiographic
usually takes 4appearance and thewithresistance
an implant. at
Type II bone is comparable to pine wood, which isn't as hard as type I. This type of
bone months to integrate

drilling, bone quality has been classified in four categories: Type 1 bone in which almost
Typethe entire
III bone bone
is like is composed
balsa wood, which of homogenous
isn't as dense as type II. Since the density isn't
as great as type II, it takes more time to "fill in" and integrate with an implant. 6 months
compact bone; Type 2 bone in which a thick layer of compact bone surrounds a coregradual
of dense trabecular bone; Type 3 bone in which
time is suggested before loading an implant placed in this type of bone. Extended
loading of the implant can, however, improve the bone density.

a thin layer of cortical bone surrounds a core of dense trabecular bone; and Type 4 bone characterized
Type as a thin
IV bone is comparable layer of cortical
to styrofoam, bone
which is the least dense of all of the bone
types. This type takes the longest length of time to integrate with the implant after
surrounding a core of low density trabecular bone of poor strength. These differencesplacement,
in bone which
quality can be associated with different
is usually 8 months. Additional implants should be placed to improve
implant/bone loading distribution. Incremental loading of the implants over time will

areas of anatomy in the upper and lower jaw. Mandibles generally are more denselyexpansion
corticated than maxillas andcan
both jawsinitial
tendimplant
to fixation.
improve bone density. Bone grafting or augmentation of bone are often required. Bone
and or bone manipulation improve

decrease in their cortical thickness and increase in their trabecular porosity as they move posteriorly. It has been shown, although there
are have been some studies that argue the point that there is a decrease in success rates as the bone type increase. There have been a
range of statistics that have been reported from a 2% difference from type 1 (98% in 36 months) to type 4 (96% in 36 months) and a
14% difference in another group (90% type 1 vs. 76% type 4 in 36 months). These are important statistics as it indicates firstly that
1 of 2 2/29/2008 11:41 AM
the bone quality is significant when considering an implant placement site, and secondly that there appears to be other factors in the
success rates of implants as one considers the vast discrepancy between the results.
4
Lekholm U, Zarb GA, Albrektsson T. Patient selectino and preparation. Tissue integrated prostheses. Chicago: Quintessence Publishing Co. Inc., 1985;199-209.

10 A Clinical Guide to Aarhus Mini-implants and Skeletal Anchorage A Clinical Guide to Aarhus Mini-implants and Skeletal Anchorage 11
References 21. Roberts WE, Marshall KJ, and Mozsary PG (1990) Rigid Endosseous Implant Utilized as Anchorage to Protract Molars
and Close an Atrophic Extraction Site. The Angle Orthodontist 60 (2):135-152.
1. Block MS. and Hoffman DR. (1995) A New Device for Absolute Anchorage for Orthodontics. American Journal of
Orthodontics and Dentofacial Orthopedics. 107 (3):251-258. 22. Veltri M., Balleri B., Goracci C., Giorgetti R., Balleri P., Ferrari M. (2009) Soft Bone Primary Stability of 3 Different
Miniscrews for Orthodontic Anchorage: A Resonance Frequency Investigation. Am J Orthod Dentofacial Orthop.
2. Brettin BT., Grosland NM., Qian F., Southard KA., Stuntz TD., Morgan TA., Marhsall SD., Southard TE. (2008) May; 135(5):642-8.
Bicoritcal vs. Monocortical Orthodontic Skeletal Anchorage. American Journal of Orthodontics and Dentofacial
Orthopedics. Nov; 134(5):625-35. 23. Wawrzinek C., Sommer T., Fischer-Brandies H. (2008) Microdamage in Cortical Bone Due to the Overtightening of
Orthodontic Microscrews. J Orothofac Orthop. Mar; 69(2):121-34.
3. Costa A., Raffainl M., and Melsen B. (1998) Miniscrews as Orthodontic Anchorage: A Preliminary Report. The
International Journal of Adult Orthodontics and Orthognathic Surgery 13 (3):201-209. 24. Wehrbein H et al (1996) The Orthosystem--A New Implant System for Orthodontic Anchorage in the Palate.
J Orofac.Orthop. 57 (3):142-153.
4. Creekmore TD. and Eklund MK. (1983) The Possibility of Skeletal Anchorage. Journal of Clinical Orthodontics. 17 (4):266-269.
25. Wilmes B., Su YY., Sadigh L., Drescher D. (2008) Pre-drilling Force and Insertion Torques During Orthodontic Mini-implant
5. Dalstra M., Cattaneo PM., and Melsen B. (2004) Load transfer of Miniscrews For Orthodontic Anchorage. Orthodontics. Insertion in Relation to Root Contact. J Orthofac Orthop. Jan; 69(1):51-8.
1:53-62.
26. Melsen B, The Role of Skeletal Anchorage in Modern Orthodontics. Clinical Orthodontics. Chapter 7, Current Concepts,
6. Godtfredsen, Laursen M., Melsen, B. (2009) Multipurpose Use of a Single Mini-implant for Anchorage in an Adult Patient. Goals and Mechanics. Edited by Ashok Karad, 2010, Elsevier.
Journal of Clinical Orthodontics. Mar:43(3):193-9.

7. Kanomi R. (1997) Mini-implant for Orthodontic Anchorage. Journal of Clinical Orthodontics. 31 (11):763-767.

8. Kofod T., Wurtz V., Melsen B., (2005) Treatment of an Ankylosed Central Incisor by Single Tooth Dento-Osseous
Osteotomy and a Simple Distraction Device. American Journal of Orthodontics Dentofacial Orthopedics. Jan;127(1):72-80.

9. Kyung HM., et al. (2003) Development of orthodontic micro-implants for intraoral anchorage. Journal of Clinical
Orthodontics 37 (6):321-328.

10. Luzi C., Verna C., Melsen B. (2009) Guidelines for Success in Placement of Orthodontics Mini-implants. Journal of
Clinical Orthodontics. Jan;43(1):39-44.

11. Luzi C., Verna C., Melsen B. (2009) Immediate Loading of Orthodontic Mini-implants: A Histomorphometric Evaluation
of Tissue Reaction. European Journal of Orthodontics. Feb;31(1):21-9.

12. Luzi C., Melsen B., Verna C., (2007) A prospective clinical investigation of the failure rate of immediately loaded
mini-implants used for orthodontic anchorage. Prog Orthod., 8(1):192-210.

13. Melsen B, (2006) VIP interview: Birte Melsen. Interview by Samir E. Bishara. World J Orthod. Fall;7(3):313-6.

14. Melsen B., (2005) Mini-implants: Where are we? J Clin Orthod. Sep;39(9):539-47.

15. Melsen B., Costa A., (2000) Immediate Loading of Implants Used for Orthodontic Anchorage. Clin Orthod Res.
Feb;3(1)23-8.

16. Melsen B., Lang NP., (2001) Biological Reactions of Alveolar Bone to Orthodontic Loading of Oral Implants. Clin
Oral Implants Res. Apr;12(2):144-52.

17. Melsen B, Petersen JK, and Costa A (1998) Zygoma Ligatures: An Alternative Form of Maxillary Anchorage. Journal
of Clinical Orthodontics 32 (3):154-158.

18. Park HS et al (2001) Micro-implant Anchorage for Treatment of Skeletal Class I Bialveolar protrusion. Journal of
Clinical Orthodontics 35 (7):417-422.

19. Park HS and Kwon TG (2004) Sliding Mechanics with Microscrew Implant Anchorage. Angle Orthod 74 (5):703-710.

20. Roberts WE et al (1989) Rigid Endosseous Implants for Orthodontic and Orthopedic Anchorage. Angle Orthod 59 (4):247-256.

12 References References 13
NOTES
3524 Washington Avenue
Sheboygan, Wisconsin, 53081 USA

Phone: +1 920 457 5051


Email: cs@americanortho.com
Web: americanortho.com

©2018 American Orthodontics Corporation


All Rights Reserved

FOLLOW US ON

LITSCR-02 January 2018