Professional Documents
Culture Documents
ImageDiagnost Installation Manual 4.6.0.en
ImageDiagnost Installation Manual 4.6.0.en
a GE Healthcare company
5391141-1-8EN
Revision 1
Copyright© 2010 General Electric Company.
All Rights Reserved.
GE Healthcare IDI Mammography Workflow Solutions
Table of Content
Table of Content
Table of Content . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Chapter 1: Safety
Definition of Warnings and Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Acronyms Glossary. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Revision History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Table of Content
Job Card CFG A12 - Desktop synchronisation from third party client to MammoWorkstation
(remote control) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 181
Job Card CFG A13 - Desktop synchronisation from MammoWorkstation to third party client
(remote control) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187
Job Card CFG A14 - Routing rules configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189
Job Card CFG A16 - Connection to a worklist server . . . . . . . . . . . . . . . . . . . . . . . . . . . 197
Job Card CFG A17 - DICOM Shuttle configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 201
Job Card CFG A18 - DICOM worklist SCP configuration . . . . . . . . . . . . . . . . . . . . . . . . 203
Job Card CFG A20 - WorklistBroker configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 205
Job Card CFG A25 - Storage Commit Broker (SC Broker) configuration . . . . . . . . . . . . 211
Job Card CFG A22 - DICOM printer configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 217
Job Card CFG A24 - Measurements calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 221
Job Card CFG A40 - Customer specific setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 223
Job Card CFG A28 - DICOM-Viewer languages configuration on DVD/CD . . . . . . . . . . 225
Job Card CFG A32 - HL7 export configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 227
Job Card CFG A34 - Unscheduled reporting configuration . . . . . . . . . . . . . . . . . . . . . . . 231
Job Card CFG A46 - Enable CESM feature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 237
Job Card IST A32 - Acceptance test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 241
Job Card IST A34 - Backup of the configuration folder . . . . . . . . . . . . . . . . . . . . . . . . . . 247
Chapter 5: Upgrade
Job Card UPG02 - Software upgrade preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 251
Job Card UPG03 - License Request . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 255
Job Card UPG04 - Basic software upgrade . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 259
Job Card UPG 07 - Backup of the configuration folder . . . . . . . . . . . . . . . . . . . . . . . . . . 261
Job Card UPG 08 - Post-Upgrade checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 263
Chapter 6: Uninstallation
Job Card UIST A02 - IDI applications uninstallation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 267
Safety
Chapter 1 Safety
This chapter contains information and warnings related to safety.
! Notice:
This Manual is primarily intended for GE Field Engineers or for Engineers trained by GE.
Safety
Language Warning
警告 本维修手册仅提供英文版本。
• 如果客户的维修服务人员需要非英文版本,则客户需自行提供翻
(ZH-CN) 译服务。
• 未详细阅读和完全理解本维修手册之前,不得进行维修。
• 忽略本警告可能对维修服务人员、操作人员或患者造成电击、机
械伤害或其他形式的伤害。
警告 本服務手冊僅提供英文版本。
• 倘若客戶的服務供應商需要英文以外之服務手冊,客戶有責任提
(ZH-HK) 供翻譯服務。
• 除非已參閱本服務手冊及明白其內容,否則切勿嘗試維修設備。
• 不遵從本警告或會令服務供應商、網絡供應商或病人受到觸電、
機械性或其他的危險。
警告 本維修手冊僅有英文版。
• 若客戶的維修廠商需要英文版以外的語言,應由客戶自行提供翻
(ZH-TW) 譯服務。
• 請勿試圖維修本設備,除非 您已查閱並瞭解本維修手冊。
• 若未留意本警告,可能導致維修廠商、操作員或病患因觸電、機
械或其他危險而受傷。
警告 このサービスマニュアルには英語版しかありません。
• サービスを担当される業者が英語以外の言語を要求される場合、
(JA) 翻訳作業はその業者の責任で行うものとさせていただきます。
• このサービスマニュアルを熟読し理解せずに、装置のサービス
を行わないでください。
• この警告に従わない場合、サービスを担当される方、操作員あ
るいは患者さんが、感電や機械的又はその他の危険により負傷
する可能性があります。
DANGER
Indicates an imminently hazardous situation that, if not avoided, will result in death or serious
injury.
WARNING
Indicates a potentially hazardous situation that, if not avoided, could result in death or serious
injury.
CAUTION
Indicates a potentially hazardous situation that, if not avoided, may result in minor or moderate
injury.
! Notice:
Used for instructions to the user to prevent damage to property.
Note:
Used to draw attention to information that is important for the user to know.
Acronyms Glossary
Acronyms Glossary
Acronyms Glossary
Revision History
Revision History
Revision History
Product Presentation
1 Introduction
IDI has developped a set of software solutions focused on improving the efficiency of the mammography
diagnostics capabilities in a digital network.
SIS
SIS
Server
Server
PACS
RIS
RIS Prefetch
Server priors
Server DICOM Worklist
(plus prior reports)
DICOM structured
report to PACS
Export to RIS
via HL7
Product Presentation
2 Physical description
2-1 Components Description
2-1-1 MammoWorkstation
MammoWorkstation is designed for diagnostic reading of mammograms.
The base software license is typically used in combination with a RIS client for reporting and does not
contain report creation.
In the base software license DICOM worklist provided by a RIS is available. The base software license
does not contain workflow management with user specific worklists provided by CollaborationServer.
2-1-1-1 Reporting Module (option)
The reporting module provides an integrated report creation mask for Mammography. It can be used to
document the report and access prior reports. The reports can be exported to an information system
using document formats such as DICOM-SR, XML, DICOM encapsulated PDF or PDF.
2-1-1-2 Workflow Module (option)
The workflow module provides support for CollaborationServer based workflow management including
double reading workflows and user specific worklists. If MammoWorkstation is used stand-alone, the
CollaborationServer functionality is built-in on the same PC. If multiple MammoWorkstations are used,
the CollaborationServer resides on a separate PC.
2-1-1-3 Backlight Module (option)
The backlight module provides image post-processing capabilites. It requires the "For Processing" raw
data from the mammography modality.
2-1-2 Mammo Documentation Station
The MammoDS is the MammoWorkstation software run in an environment not approved for diagnostic
reading. The technologist can use MammoDS to access patient data simultaneously to the examination.
MammoDS is installed close to the mammography system. MammoDS provides access to finding
annotations and reports created by the radiologist. In addition, the technologist can enter clinical patient
information obtained during the exam. This information is available to the radiologist when reading the
case.
2-1-3 DICOM Shuttle
The DICOM Shuttle provides routing services to distribute DICOM documents across devices and
locations.
Its WorklistBroker module consolidates worklists from different sources and presents them as one single
Worklist to one or more mammography modalities.
In addition to DICOM Shuttle node, some DICOM Shuttle satellites can be installed to connect to multiple
devices that are in remote locations.
2-1-4 CollaborationServer
The Collaboration server connects with RIS using DICOM worklists and manages the workflow for
MammoWorkstations. It provides workflow schemes for single reading and blind double reading.
It is proposed as an additional hardware in case there are multiple workstations on site.
Product Presentation
3 Functional description
3-1 Stand-Alone MammoWorkstation configuration
Illustration 1 Stand-alone MammoWorkstation
Product Presentation
Note:
The images are automatically pushed to the right PACS.
Product Presentation
Note:
- Collaboration Server enables advanced Mammo Workflow (multiple reading, rule based
worklists, user specific worklists).
- Structured report: homogenous vocabulary + accurate lesion localization.
Note:
Every hardware equipment shall be in compliance with IEC/EN 60950-1: Information technology
equipment - Safety.
Product Presentation
Hardware Requirements
Hardware Requirements
Images from other modalities are not interpreted as X-Ray Mammography. Instead they appear in the
"Additional Series" list.
*For Fuji-CR there is a "MG-Option", which is to be used instead of CR.
Hardware Requirements
Note:
For Agfa CR also the Series Description Description (0008,103E, General Series Module
Attributes C.7-5) is interpreted.
Translation table for standard orientations:
P L R CC
A R L CC
P FL R MLO
A FR L MLO
P F R ML
A F L ML
MammoWorkstation supports additional values for patient orientation with automatic flip and rotation of
images for proper display.
If Patient Orientation is missing, the "View Code Sequence" is used according to SNM3:
R-10224 medio-lateral ML ML
Hardware Requirements
R-10244 caudo-cranial FB CC
R-10228 latero-medial LM ML
If none of the above values are set, then the MammoWorkstation cannot sort the images according to
laterality and view position. The images will be displayed unsorted in the Patient Folder.
3. Measurements and Scales, Quadrant-Zoom, DICOM-Print
CR-Image Module Attributes (C.8-2) / X-Ray Image Module Attributes (C.8-26)
The following attribute overrides the Imager Pixel Spacing:Image Plane Module Attributes (C.7-10)
If none of these values are available, then the 1:1cm display function will be deactivated. Furthermore,
the scales in the Magnifying Glass and measurement function will be deactivated. DICOM print will not
be possible as well.
4. Windowing
VOI LUT Module Attributes (C.11-2)
Hardware Requirements
5. MQSA compliance
In order for MQSA compliant mammographic image identification, the modality must
fill the following DICOM tags:)
0010,0020 Patient ID
If the required information is encoded in other DICOM tags, these tags must be added to the display
preferences of MammoWorkstation (see Job Card CFG A06 - Language and display preference
configuration on page 169).
Hardware Requirements
For seamless PACS integration the following conditions on the PACS are to be validated:
• The MammoWorkstation prefers DICOM Q/R using the modern relational query model. The basic
hierarchic query model is also supported, but reduces the query performance of the workstation.
• For prefetch automation, the PACS should support the DICOM Tag Body Part Examined
(0018,0015) in queries.
• For storing of overlays, the PACS must support storing DICOM Grayscale Softcopy Presentation
State documents. (SOP class 1.2.840.10008.5.1.4.1.1.11.1).
• For the integrated reporting to work, the PACS must support storing DICOM Structured Report
documents. (Basic Text SR, SOP class 1.2.840.10008.5.1.4.1.1.88.11).
• For the CAD-display to be stored, the PACS must support storing DICOM CAD documents.
(Mammography CAD SR, SOP class 1.2.840.10008.5.1.4.1.1.88.50).
Additional useful features for decentralized tele-mammography networks:
• Rule-based auto-routing to archives and workstations
• Query-forwarding to secondary archives
• Extendible with web-server/web-client that supports display of reports in DICOM Structured Report
format.
Note:
For additional information, refer to the DICOM Conformance Statement.
2-3-2 Validated 3rd party PACS systems
Validated and in clinical use with Image Diagnost MammoWorkstation are the following PACS systems:
- Agfa IMPAX
- Agfa-GWI
- Aycan PACS
- Cerner Image Devices PACS
- ConVis PACS
Hardware Requirements
- Econmed PACS
- Fuji Axon PACS
- GE Centricity Archive
- GE Centricity RA 600
- GEMED PACS
- HIPAX PACS
- Image Devices PACS
- Kodak Carestream PACS
- Medavis PACS
- MedicalCommuniations Ashvins PACS
- Medigration PACS
- Medos PACS
- Philips PACS
- Rogan DelftPACS
- Sectra PACS
- Tiani PACS
- Unisolo PACS
- Vepro PACS
- Visus PACS
Hardware Requirements
- Exprivia RIS
- GE Centricity RIS
- iSoft/GAP RIS
- Image Devices RIS
- Kodak RIS 2010
- Medavis RIS
- Medos RIS
- Siemens Medico RIS
- Turbomed RIS
- Unisolo RIS
Software Installation
2 Installation of Windows (64-bit) Job Card IST A03 - Installation of Windows XP 45’
Professional (64-bit) on page 43
or Job Card IST A43 - Installation of Windows 7
(64-bit) on page 49
3 Setup of Network Connection Job Card IST A06 - Network connection setup 15’
on Windows XP on page 117
or Job Card IST A07 - Network connection setup
on Windows 7 on page 119
4 Basic Windows Configuration Job Card IST A05 - Windows XP configuration 15’
for IDI on page 69
or Job Card IST A45 - Windows 7 configuration
for IDI on page 83
5 IDI Publication Installation Job Card IST A11 - IDI Publication installation 15’
on page 99
6 IDI Software installation Job Card IST A12 - IDI Software installation on 20’
page 101
7 IDI Software licensing Job Card IST A13 - IDI application licensing on 10’
page 113
Note:
License delivery may take 24 hours; installation steps can still be completed, but license
code entry is mandatory to access the application and run functional tests.
8 VNC installation and configuration Job Card IST A22 - VNC Installation for 15’
Windows XP on page 149
9 Mandatory configuration Job Card IST A16 - Mandatory configuration on 15’
page 121
10 Desktop theme configuration Job Card IST A18 - Desktop theme 5’
configuration for Windows XP on page 125
11 Keypad installation Job Card IST A20 - MacroWorks II keypad 10’
installation on Windows XP on page 129
or Job Card IST A48 - MacroWorks 3 keypad
installation on Windows 7 on page 137
Software Installation
12 Local DICOM settings and database Job Card CFG A02 - Local DICOM settings and 10’
configuration database configuration on page 155
13 User accounts creation Job Card CFG A04 - User accounts creation on 10’
page 159
14 Language and display preference Job Card CFG A06 - Language and display 10’
configuration preference configuration on page 169
15 DICOM partners configuration Job Card CFG A08 - DICOM partners 20’
configuration on page 173
16 DICOM media import configuration Job Card CFG A10 - DICOM media import 5’
configuration on page 177
17 Remote control configuration Job Card CFG A12 - Desktop synchronisation 20’
from third party client to MammoWorkstation
(remote control) on page 181
18 RIS synchronisation Job Card CFG A13 - Desktop synchronisation 30’
from MammoWorkstation to third party client
(remote control) on page 187
19 Connection to a worklist server Job Card CFG A16 - Connection to a worklist 10’
server on page 197
20 DICOM printer configuration Job Card CFG A22 - DICOM printer 10’
configuration on page 217
21 Measurements calibration Job Card CFG A24 - Measurements calibration 10’
on page 221
22 Customer specific settings Job Card CFG A40 - Customer specific setting 5’
on page 223
23 Language configuration Job Card CFG A28 - DICOM-Viewer languages 5’
configuration on DVD/CD on page 225
24 Unscheduled Reporting Job Card CFG A34 - Unscheduled reporting
configuration configuration on page 231
25 HL7 export configuration Job Card CFG A32 - HL7 export configuration 20’
on page 227
Final sequence: required to complete the installation.
Software Installation
2 Installation of Windows (32-bit) Job Card IST A04 - Installation of Windows XP 35’
Professional (32-bit) on page 57
or Job Card IST A44 - Installation of Windows 7
(32-bit) on page 59
3 Setup of Network Connection Job Card IST A06 - Network connection setup on 15’
Windows XP on page 117
or Job Card IST A07 - Network connection setup
on Windows 7 on page 119
4 Basic Windows Configuration Job Card IST A05 - Windows XP configuration for 15’
IDI on page 69
or Job Card IST A45 - Windows 7 configuration
for IDI on page 83
5 IDI Publication Installation Job Card IST A11 - IDI Publication installation on 15’
page 99
6 IDI Software installation Job Card IST A12 - IDI Software installation on 10’
page 101
7 IDI Software licensing Job Card IST A13 - IDI application licensing on 10’
page 113
Note:
License delivery may take 24 hours; installation steps can still be completed, but
license code entry is mandatory to access the application and run functional tests.
8 VNC installation and Job Card IST A22 - VNC Installation for Windows 15’
configuration XP on page 149
9 Mandatory configuration Job Card IST A16 - Mandatory configuration on 15’
page 121
10 Desktop theme configuration Job Card IST A18 - Desktop theme configuration 5’
for Windows XP on page 125
11 Reporting mask configuration Job Card IST A17 - Reporting mask configuration 5’
for MammoDS on page 123
12 Keypad installation Job Card IST A20 - MacroWorks II keypad 10’
installation on Windows XP on page 129
or Job Card CFG A34 - Unscheduled reporting
configuration on page 231
13 Local DICOM settings and Job Card CFG A02 - Local DICOM settings and 10’
database configuration database configuration on page 155
Software Installation
14 User accounts creation Job Card CFG A04 - User accounts creation on 10’
page 159
15 Language and display Job Card CFG A06 - Language and display 10’
preference configuration preference configuration on page 169
16 DICOM partners configuration Job Card CFG A08 - DICOM partners 20’
configuration on page 173
17 DICOM media import Job Card CFG A10 - DICOM media import 5’
configuration configuration on page 177
18 Remote control configuration Job Card CFG A12 - Desktop synchronisation 20’
from third party client to MammoWorkstation
(remote control) on page 181
19 RIS synchronisation Job Card CFG A13 - Desktop synchronisation 30’
from MammoWorkstation to third party client
(remote control) on page 187
20 Connection to a worklist server Job Card CFG A16 - Connection to a worklist 10’
server on page 197
21 DICOM printer configuration Job Card CFG A22 - DICOM printer configuration 10’
on page 217
22 Measurements calibration Job Card CFG A24 - Measurements calibration 10’
on page 221
23 Customer specific settings Job Card CFG A40 - Customer specific setting on 5’
page 223
24 Language configuration Job Card CFG A28 - DICOM-Viewer languages 5’
configuration on DVD/CD on page 225
25 HL7 export configuration Job Card CFG A32 - HL7 export configuration on
page 227
Final sequence: required to complete the installation.
26 Acceptance test Job Card IST A32 - Acceptance test on page 241 20’
27 Backup of the configuration Job Card IST A34 - Backup of the configuration 15’
folder folder on page 247
Software Installation
2 Installation of Windows (32-bit) Job Card IST A04 - Installation of Windows XP 10’
Professional (32-bit) on page 57
or Job Card IST A44 - Installation of Windows 7
(32-bit) on page 59
3 Setup of Network Connection Job Card IST A06 - Network connection setup on 15’
Windows XP on page 117
or Job Card IST A07 - Network connection setup
on Windows 7 on page 119
4 Basic Windows Configuration Job Card IST A05 - Windows XP configuration for 15’
IDI on page 69
or Job Card IST A45 - Windows 7 configuration for
IDI on page 83
5 IDI Publication Installation Job Card IST A11 - IDI Publication installation on 15’
page 99
6 IDI Software installation Job Card IST A12 - IDI Software installation on 10’
page 101
7 IDI Software licensing Job Card IST A13 - IDI application licensing on 10’
page 113
Note:
License delivery may take 24 hours; installation steps can still be completed, but license
code entry is mandatory to access the application and run functional tests.
8 VNC installation and configuration Job Card IST A22 - VNC Installation for Windows 15’
XP on page 149
9 Mandatory configuration Job Card IST A16 - Mandatory configuration on 15’
page 121
10 Local DICOM settings and Job Card CFG A02 - Local DICOM settings and 10’
database configuration database configuration on page 155
11 Language and display preference Job Card CFG A06 - Language and display 10’
configuration preference configuration on page 169
12 DICOM partners configuration Job Card CFG A08 - DICOM partners 20’
configuration on page 173
13 Routing rules configuration Job Card CFG A14 - Routing rules configuration 20’
on page 189
Software Installation
14 Connection to a worklist server Job Card CFG A16 - Connection to a worklist 10’
server on page 197
15 DICOM Worklist SCP configuration Job Card CFG A18 - DICOM worklist SCP 15’
configuration on page 203
16 Worklist broker configuration Job Card CFG A20 - WorklistBroker configuration 15’
on page 205
Final sequence: required to complete the installation.
17 Acceptance test Job Card IST A32 - Acceptance test on page 241 20’
18 Backup of the configuration folder Job Card IST A34 - Backup of the configuration 15’
folder on page 247
1 Applicability
This job card applies to all PCs.
2 Personnel requirements
1 Field Engineer
Time: 15’
3 Reception
1. Examine the system on reception. Check any shock indicators present. If possible, make a visual
check of all items.
2. If problems occur (items missing, signs of shipping damage, excessive shock, etc.), note them down
on the delivery note before signing for reception.
1 Applicability
This job card applies to MammoWorkstation.
2 Personnel requirements
1 Field Engineer
Time:45’
3 Hardware requirements
MammoWorkstation software requires a PC that fulfills the following minimal requirements:
• Two Intel Quad-Core Xeon E5504 2.00GHz (or better)
• 8GB RAM (or better)
• 1x HDD SATA II 160GB 7.2k (or better) for operating system
• 2x HDD SATA II 500GB 7.2k (or better) for local database cache
• Display controller with standard DVI connector.
• network controller with 1GBit/s (or better)
• DVD+-RW double layer
• Localized professional keyboard, Mouse
• Windows XP x64 multilingual
Displays: MammoWorkstation software requires a three monitor setup, where there is:
• one display monitor with at least 1280×1024 pixel
• two 5MP diagnostic display monitors for primary diagnostic interpretation of mammographic images
! Notice:
Diagnostic displays used with MammoWorkstation must be cleared by local regulatory authorities
for use in digital mammography. Vendors evidence of regulatory approval for display systems shall
cover:
- displays,
- display controller,
- calibration equipment and procedures,
- display system QC procedures
! Notice:
For USA: Evidence of MQSA compliance is required.
Note:
Note that "Basic" turned to "Dynamic".
6. Right click on Disk 1(1) and create a New volume.
11. Choose "NTFS" as file system (default setting) and checkmark "Perform a quick format".
12. At the top of the disk management window you can now see that a striped volume (consisting of
Disk 1 and Disk 2) has been created.
1 Applicability
This job card applies to MammoWorkstation.
2 Personnel requirements
1 Field Engineer
Time:45’’
3 Hardware requirements
MammoWorkstation software requires a PC that fulfills the following minimal requirements:
• Two Intel Quad-Core Xeon E5504 2.00GHz (or better)
• 8GB RAM (or better)
• 1x HDD SATA II 160GB 7.2k (or better) for operating system
• 2x HDD SATA II 500GB 7.2k (or better) for local database cache
• Display controller with standard DVI connector.
• network controller with 1GBit/s (or better)
• DVD+-RW double layer
• Localized professional keyboard, Mouse
• Windows 7 x64 multilingual
Displays: MammoWorkstation software requires a three monitor setup, where there is:
• one display monitor with at least 1280×1024 pixel
• two 5MP diagnostic display monitors for primary diagnostic interpretation of mammographic images
! Notice:
Diagnostic displays used with MammoWorkstation must be cleared by local regulatory authorities
for use in digital mammography. Vendors evidence of regulatory approval for display systems shall
cover:
- displays,
- display controller,
- calibration equipment and procedures,
- display system QC procedures
! Notice:
For USA: Evidence of MQSA compliance is required.
9. Set a password for your account (e.g "idi"). Confirm by clicking [Next].
10. Read and accept the license terms. Confirm by clicking [Next].
12. Review your time and date settings. Confirm by clicking [Next].
4. Right click on Disk 1 (1)and convert the disk to a Dynamic Disk (2).
Note:
Note that "Basic" turned to "Dynamic".
10. Select "NTFS" as file system (default setting) and checkmark "Perform a quick format".
On the next page confirm the process with [Finish].
11. At the top of the disk management window you can now see that a striped volume has been created.
The final screen should look like this:
1 Applicability
This job card applies to:
• MammoDS / TechInsight
• CollaborationServer
• DICOMShuttle
2 Personnel requirements
1 Field Engineer
Time:35’
3 Hardware requirements
3-1 Hardware requirements for CollaborationServer and DicomShuttle Node
CollaborationServer and DICOMShuttle Node software option require a PC that fulfills the following
minimal requirements:
• One Intel Dual-Core Xeon 2.33GHz (or better)
• 2GB RAM (or better)
• 2x HDD SATA II 250GB 7.2k (or better)
• Display controller with standard DVI connector
• Network controller with 1GBit/s (or better)
• DVD+-RW double layer
• Localized professional keyboard, Mouse
• Windows XP x32 multilingual
Note:
For DICOM Shuttle PC and CollaborationServer PC, optionally one display monitor with at least
1280×1024 pixel can be attached.
8. Choose "skip" for "How will your computer connect to the Internet?".
The internet connection will be setup later.
9. Choose "No, not at this time" for "Ready to register with Microsoft?"
Setup is complete.
1 Applicability
This job card applies to:
• MammoDS /TechInsight
• CollaborationServer
• DICOMShuttle
2 Personnel requirements
1 Field Engineer
Time: 35’
3 Hardware requirements
3-1 Hardware requirements for CollaborationServer and DicomShuttle Node
CollaborationServer and DICOMShuttle Node software option require a PC that fulfills the following
minimal requirements:
• One Intel Dual-Core Xeon 2.33GHz (or better)
• 2GB RAM (or better)
• 2x HDD SATA II 250GB 7.2k (or better)
• Display controller with standard DVI connector
• Network controller with 1GBit/s (or better)
• DVD+-RW double layer
• Localized professional keyboard, Mouse
• Windows 7 x32 multilingual
Note:
For DICOM Shuttle PC and CollaborationServer PC, optionally one display monitor with at least
1280×1024 pixel can be attached.
3. The following screens will be displayed: Select the Boot Device. Change your setting via the arrow
keys.
7. Choose a user name for your account (e.g. "ImageDiagnost") and a name for your computer. Confirm
by clicking [Next].
8. Set a password for your account(e.g "idi"). Confirm by clicking [Next].
9. Read and accept the license terms. Confirm by clicking [Next].
11. Review your time and date settings. Confirm by clicking [Next].
1 Applicability
This job card applies to all PCs.
2 Personnel requirements
1 Field Engineer
Time: 15’’
3 Hardware requirements
• CAT6 Hospital Network cables must be used for Gigabit speed.
4 Physical connection
1. Connect the RJ45 connector to one network card ports.
5 Configuration of IP address
1. Click on Start ➜ Control Panel ➜ Network Connections and right click on "Local area connection"
properties.
Note:
In case the Control Panel window opens in Category View, click Switch to Classic View.
Note:
Since there are two network cards, there are also two "Local area connections". Choose the one
that corresponds to the port the network cable is connected to.
2. Scroll down to "Internet Protocol (TCP/IP)" and click on the button [Properties].
3. Enter the static IP address that was provided by the customer (A).
Note:
a) All IDI Workstations PCs have to be online and remotely accessible at any time.
b) There is no need to specify DNS servers (B). Doing so could affect the speed of the medical
application.
1 Applicability
This job card applies to all PCs.
2 Personnel requirements
1 Field Engineer
Time: 15’
3 Hardware requirements
• CAT6 Hospital Network cables must be used for Gigabit speed.
4 Configuration of IP address
1. Click on Start ➜ Control Panel ➜ Network and Internet ➜ Network and Sharing Center
2. On the left side click Change adapter settings.
3. Right click on Local Area Connection and select Properties.
Note:
In case the workstation has two or more interfaces, make sure to select the correct Local Area
Connection.
Note:
a) All IDI Workstations PCs have to be online and remotely accessible at any time.
b) There is no need to specify DNS servers (B). Doing so could affect the speed of the medical
application.
1 Applicability
This job card applies to all PCs.
2 Personnel requirements
1 Field Engineer
Time: 15’
3 Hardware requirements
NA
Verification: To ensure that all updates have been installed, return once more to the windows update and
check for updates.
Note:
If no internet connection is available you must install the Microsoft.net Framework. (See the
delivered CD or in the folder C:\install\windows_dotnet_1.1).
1. Click on Start ➜ right click on My Computer ➜ Manage ➜ "Local Users and Groups".
7. Assign admin rights: right click on the user that has just been created and select "Properties".
9. In the Select Groups window, type "Administrators" and click on the button [Check Names] to
make sure there is no typing error.
! Notice:
To avoid unauthorized access to any MammoWorkstation, one of the following security features
must be activated on all user accounts:
- Auto-logoff: the current user is logged-off when inactive for 30 min:
a. Select Start ➜ Control Panel (switch to classic view) ➜ Display.
b. On tab Screen Saver, choose a screen saver.
c. Checkmark the box On resume, display the welcome screen / On resume, password protect.
- Press the keys [Windows + L] when leaving the room.
Note:
For Administrator, choose password "operator".
11. Logout from the session: Select Start ➜ Log Off, click [Log Off].
The new user account should appear and the Administrator account should have disappeared.
12. Login to the new account. You are ready for the next installation steps.
7 Performance adjustment
1. Select Start ➜ Control Panel ➜ System - tab "Advanced", section "Performance" .
2. Click on button [Settings].
3. In tab "Visual effects" select "Adjust for best performance".
However, if the customer requires that Automatic Windows Updates is active, it is the responsibilty
of the customer to ensure that the system passes the acceptance test after any Windows Update.
See Job Card IST A32 - Acceptance test on page 241.
13 Firewall
1. It is recommended to turn off the firewall.
a. To turn off the firewall, select Start ➜ Control Panel ➜ Windows Firewall - tab "General".
b. Select "Off".
2. However, if the customer requires the firewall to remain active, make sure that the following ports are
open:
To open a port through the firewall, perform the following steps for each port:
a. Select Start ➜ Control Panel ➜ Windows Firewall - tab "Exceptions".
b. Click on [Add Port...].
c. Enter the name and port number of the port to keep open and select "TCP".
Note:
If the customer requires the firewall to remain active, make sure that the system passes the
acceptance test.
See Job Card IST A32 - Acceptance test on page 241.
14 Antivirus software
! Notice:
Antivirus software can have significant impact on the performance of the MammoWorkstation. It is
therefore recommended to turn off any antivirus software.
• To turn off the Antivirus software, select Start ➜ Control Panel ➜ Security Center.
However, if the customer requires that an Antivirus software is running observe the following
instructions:
1. Preparation:
• import and load several cases (note the number of cases)
• pay attention on how long it takes to import and load these cases
2. Procedure:
• Make sure the following folders and files are not scanned by the Antivirus software:
- the IDI Cache folder ("dbcache")
- the Image Diagnost Folder in the Windows Program Files
- 64bit systems (MammoWorkstation): C:\Program Files (x86)\ImageDiagnost
- 32bit systems (DICOM Shuttle, Collaboration Server, MDS): C:\Program Files\
ImageDiagnost
- all files with ".dcm" extension
- the MySQL Folder in C:/mysql
• Make sure the following services and applications are - depending on Antivirus software used-
not exclusively locked by the Antivirus software while the workstation is in use:
Folder dbcache
Services ApacheTomcat6forIDI
MySqI
IDDicomReceiveSCP
IDDicomWorklistSCP
Application MammoWorkstation.exe
! Notice:
Antivirus scan must not be performed while operators are using the MammoWorkstation.
Example: A full scan can be scheduled daily at 11pm.
3. Validation: After installing an Antivirus software, check the performance of the MammoWorkstation
by:
• importing and opening several cases (the same number of cases as before).
• comparing the loading time to the performance prior to the Antivirus software installation.
! Notice:
If the loading time is significantly longer than it was before, it is recommended to uninstall the
Antivirus software. Otherwise the performance of the MammoWorkstation will be impaired.
Note:
In the past, IDI customers have used the following Antivirus Software option:
- Antivirus option by Trend Micro.
So far, this option had no noticeable impact on the performance of the MammoWorkstation. This
Antivirus Software has not been validated and due to continuous change and update procedures
every Antivirus Software is subject to, an undisturbed performance can not be assured for the
future.
Note:
If the customer requires that an Antivirus software is running, make sure that the system passes
the acceptance test.
See Job Card IST A32 - Acceptance test on page 241.
3. In Computer Name field, enter the new computer name, then click [OK].
Note:
Type 15 characters max (e.g. "Helens-Shuttle1")
4. Reboot the computer.
16 Alert settings
1. Select Start ➜ Control Panel ➜ Security Center.
2. On the left side, click on the "change the way security center alerts me" and uncheck:
• Firewall
• Automatic Updates
• Virus Protection
It it possible to employ RDP (Remote Desktop Protocol), which is already a component of XP. In that
case, be sure that a user password is set. Otherwise RDP won’t work. The disadvantage though, is that
while being connected remotely to a system via RDP no interaction of the user on the other end is
possible. Thus connections for training and demonstration purposes are not viable.
Therefore we recommend a freeware tool such as VNC (see Job Card IST A22 - VNC Installation for
Windows XP on page 149).
Any other appropriate tool can be used as well.
Note:
After installing a Remote Software tool, make sure the system passes the acceptance test. See
Job Card IST A32 - Acceptance test on page 241.
18 Deactivate IPv6
To this date, IDI MammoWorkstation is not compatible with Internet Protocol version 6 (IPv6).
IPv6 must therefore be deactivated:
Note:
To deactivate IPv6 you must be a Member of the Administrator Group or at least have permission
to edit the Windows registry.
1 Applicability
This job card applies to all PCs.
2 Personnel requirements
1 Field Engineer
Time: 15’
3 Hardware requirements
NA
Verification: To ensure that all updates have been installed, return once more to the windows update and
check for updates.
Note:
If no internet connection is available you must install the Microsoft.net Framework. (See the
delivered CD or in the folder C:\install\windows_dotnet_1.1).
2. For new users, right click on "Users" and then on "New User".
Note:
User name is limited to 20 characters.
3. Set a password:
• Select "Users" and highlight the desired user.
• Right click the user and click "Set password"..
Note:
For Administrator, choose password "operator".
The new user account should appear as well as the Administrator account.
6. Login to the user account. You are ready for the next installation steps.
7 Performance adjustment
1. Select Start ➜ Control Panel ➜ System and Security ➜ System ➜ Advanced system settings - tab
"Advanced".
2. In the tab Advanced go to section Performance and click on [Settings].
3. In tab "Visual effects", select "Adjust for best performance".
4. Click [Apply] ➜ the windows appearance will change to windows standard theme.
However, if the customer requires that Automatic Windows Updates is active, it is the responsibilty
of the customer to ensure that the system passes the acceptance test after any Windows Update.
See Job Card IST A32 - Acceptance test on page 241.
3. Under section "Important updates" ➜ in the drop down menu select "Never check for updates (not
recommended)".
13 Firewall
1. It is recommended to turn off the firewall.
a. To turn off the firewall, select Start ➜ Control Panel ➜ System and Security ➜ Windows Firewall.
b. On the left side click "Turn windows firewall on or off".
- Set all location settings to "Turn off Windows Firewall (not recommended)".
2. If the customer requires the firewall to remain active, make sure the following TCP ports are open:
To open a port through the firewall, perform the following steps for each port:
a. Select Start ➜ Control Panel ➜ System and Security ➜ Windows Firewall.
b. On the left side click "Advanced settings".
c. On the left side select "Inbound Rules".
Note:
If the customer requires the firewall to remain active, make sure that the system passes the
acceptance test.
See Job Card IST A32 - Acceptance test on page 241.
14 Antivirus software
! Notice:
Antivirus software can have significant impact on the performance of the MammoWorkstation. It is
therefore recommended to turn off any antivirus software.
• To turn off the Antivirus software, select Start ➜ Control Panel ➜ System and Security.
However, if the customer requires that an Antivirus software is running observe the following
instructions:
1. Preparation:
• import and load several cases (note the number of cases)
• pay attention on how long it takes to import and load these cases
2. Procedure:
• Make sure the following folders and files are not scanned by the Antivirus software:
- the IDI Cache folder ("dbcache")
- the Image Diagnost Folder in the Windows Program Files
- 64bit systems (MammoWorkstation): C:\Program Files (x86)\ImageDiagnost
- 32bit systems (DICOM Shuttle, Collaboration Server, MDS): C:\Program Files\
ImageDiagnost
- all files with ".dcm" extension
- the MySQL Folder in C:/mysql
• Make sure the following services and applications are - depending on Antivirus software used-
not exclusively locked by the Antivirus software while the workstation is in use:
Folder dbcache
Services ApacheTomcat6forIDI
MySqI
IDDicomReceiveSCP
IDDicomWorklistSCP
Application MammoWorkstation.exe
! Notice:
Antivirus scan must not be performed while operators are using the MammoWorkstation.
Example: A full scan can be scheduled daily at 11pm.
3. Validation: After installing an Antivirus software, check the performance of the MammoWorkstation
by:
• importing and opening several cases (the same number of cases as before).
• comparing the loading time to the performance prior to the Antivirus software installation.
! Notice:
If the loading time is significantly longer than it was before, it is recommended to uninstall the
Antivirus software. Otherwise the performance of the MammoWorkstation will be impaired.
Note:
In the past, IDI customers have used the following Antivirus Software option:
- Antivirus option by Trend Micro
So far, this option had no noticeable impact on the performance of the MammoWorkstation. This
Antivirus Software has not been validated and due to continuous change and update procedures
every Antivirus Software is subject to, an undisturbed performance can not be assured for the
future.
Note:
If the customer requires that an Antivirus software is running, make sure that the system passes
the acceptance test.
See Job Card IST A32 - Acceptance test on page 241.
15 Action Center
1. Select Start ➜ Control Panel ➜ System and Security ➜ Action Center.
2. On the left side click on the "Change Action Center settings":
3. Uncheck:
• Windows Updates
• Network Firewall
• Virus Protection
It it possible to employ RDP (Remote Desktop Protocol), which is already a component of Windows 7. In
that case, be sure that a user password is set. Otherwise RDP won’t work. The disadvantage though, is
that while being connected remotely to a system via RDP no interaction of the user on the other end is
possible. Thus connections for training and demonstration purposes are not viable.
Therefore we recommend a freeware tool such as VNC. Any other appropriate tool can be used as well.
Note:
After installing a Remote Software tool, make sure the system passes the acceptance test. See
Job Card IST A32 - Acceptance test on page 241.
18 Deactivate IPv6
To this date, IDI MammoWorkstation is not compatible with the Internet Protocol Version 6 (IPv6). The
IPv6 must therefore be deactivated:
Note:
To deactivate IPv6 you must be a Member of the Administrator Group or at least have permission
to edit the Windows registry.
! Notice:
Since serious problems may occur if the registry was modified incorrectly, back up the registry
before modifying it. If any problem occurs after modifying the Windows registry, restore it with the
backup version.For detailed information on how to back up the Windows registry, refer to
http://support.microsoft.com/kb/322756/en-us.
Temporary solution:
1. Open the Local Area Network Status window and click on [Properties].
2. Uncheck the box next to The Internet Protocol Version 6 (TCP/IPv6).
Note:
The IPv6 will be enabled again whenever the system is restarted.
! Notice:
Since no procedure to deactivate IPv6 on Windows 7 has been validated so far, the following
procedure can only be a suggestion. Check every modification on Windows registry with your
Network Administrator.
Permanent solution:
1. Open the registry editor by clicking Start ➜ Run, type regedit ➜ click [OK].
2. Go to HKEY_LOCAL_MACHINE > SYSTEM ➜ CurrentControlSet ➜ services ➜ TCPIP6 ➜
Parameters key.
3. Right-click on [Parameters] on the left side. Select New ➜ DWORD (32-bit) Value.
4. Name the value as DisabledComponents.
5. Set the value to 0x20 to use IPv4 instead of IPv6 in prefix policies.
6. Click [OK].
7. Reboot the system.
Validation:
1. Open the Local Area Network Status window and click on [Properties]. The box next to Internet
Protocol Version 6 (TCP/IPv6) should be unchecked.
Note:
For further information on how to disable the IPv6 permanently, refer to
http://support.microsoft.com/kb/929852.
1. Select Start ➜ Control Panel ➜ Hardware and Sound . In the section Devices and Printers, select
Mouse.
2. A dialog window will open. Select the tab "Pointers".
3. In the drop down menu (1), select "Windows Standard (large) (system scheme)".
4. Click [Apply] to implement the configuration. Click [OK] to exit the menu.
1 Applicability
This job card applies to all PCs.
2 Personnel requirements
1 Field Engineer
Time: 15’
3 IDI Publication on CD
Two IDI Publication CDs are provided:
1. Mammography Workflow Solutions User Publication CD.
The Mammography Workflow Solutions User Publication CD contains all User Documentation.
2. Mammography Workflow Solutions Service Publication CD.
The Mammography Workflow Solutions Service Publication CD contains all Service Documentation.
! Notice:
At the end of the installation process, the IDI Publication CDs (User and Service) must be given to
the customer to make sure the customer can consult these CDs on others PCs as well.
Note:
The customer is responsible for regrouping and storing these publication CD-ROMs in a safe
place. These CDs are needed during the entire life cycle of the MammoWorkstation system.
1 Applicability
This job card applies to all PCs.
2 Personnel requirements
1 Field Engineer
Time: 20’
For MammoWorkstation you need to install the 64bit version of Image Mastering API v2.0.
For MammoDS you need to install the 32bit version of Image Mastering API v2.0.
Note:
Both packages, the 64bit version and the 32bit version of Image Mastering API, are included on
the Image Diagnost Software CD.
Note:
Make sure to choose the correct language for the Image Mastering API v.2.0 corresponding to your
Windows operating system.
1. Launch the Image Mastering API v2.0 installer. At the beginning a warning will be displayed. Click
[Run].
b. For MammoDS, click [Finish] to exit the Software Update Installation Wizard.
4. When you get the message box "Installing MySQL", click [OK].
5. An additional installation window will open to guide you through installing MySQL.Click [Next>]:
9. After MySQL setup is complete, confirm with [OK]. The software will be installed.
10. Click [Finish] to close the Image Diagnost 4.6.0 Setup Wizard.
! Notice:
This log file informs you about the installation process. If one of the lines starts with "ERROR"
instead of "INFO", please contact Image Diagnost for assistance.
12. Close this log file window. The license screen will appear.
Note:
To complete the installation of a DICOMShuttle, refer to Job Card CFG A17 - DICOM Shuttle
configuration on page 201.
7 Post-installation checks
7-1 Multiple items created
Due to the fact that MammoWorkstation is available as a 32-bit version and as a 64-bit version, several
multiple items might be created.
After the installation process, it is necessary to check that only the icons that correspond to your installed
system are available.
1
2. Right click on the icon you want to delete and select "Delete".
Note:
Make sure to delete the icon that does not correspond to your installed system.
2. Right click on the icon you want to delete and select "Remove from This List".
Note:
Make sure to delete the icon that does not correspond to your installed system.
Check that only one new program folder has been created.
Note:
If two program folders have been created, the program folder that does not correspond to your
installed system must be deleted.
Delete the program folder that does not correspond to your installed version:
1. Go to C:\Program Files\ImageDiagnost\MammoWorkstation.
2. Delete the x64 folder.
Is needed
for 64-bit
version
Is needed
for 32-bit
version
! Notice:
For Windows 7: Always start the MammoWorkstation as Administaror:
- Right click on "Setup_ImageDiagnost_[version]_[…].exe and click Run as administrator.
- Go to tab Compatibility and enable the Run this program as an administrator checkbox (1).
1 Applicability
This job card applies to all PCs.
2 Personnel requirements
1 Field Engineer
Time:10’’
3 Hardware requirements
• Internet connection
Note:
To operate the system, a valid license for Image Diagnost Software is required. The license key
depends on the PC hardware.
2. Click [Accept].
Installation Code
Note:
For a MWS standalone license request (single MammoWorkstation including a Collaboration
Server), use MammoWorkstation (standalone) as the Ordered system configuration:
Note:
If the Features of Version 4.x (not available in USA) and the Backlight creation (not available in
USA) boxes were ticked for the license request, the following warning will be displayed. Press
[Activate license]; the C-Echo test will be carried out.
Note:
For a MWS standalone license request, ensure that the CollaborationServer checkbox is checked.
Note:
If you confirm the dialog with [Finished], the C-Echo test will be carried out.
As the mandatory configuration has yet not been completed, a warning pop-up comes up (see
below). Confirm this warnings by clicking [OK]. Then click [OK] on the License tab.
If you confirm this warning and click [OK] on the License tab, you will receive the following message:
9. Confirm this message by clicking [OK]. Go to the tab Local (which will be marked by a red cross) and
continue the mandatory configuration, seeJob Card IST A16 - Mandatory configuration on page 121
! Notice:
For Windows 7: Always start the MammoWorkstation as Administaror:
- Right click on the MammoWorkstation icon and select Run as administrator.
or
- Right click on the MammoWorkstation icon.
- Select Properties
- Go to tab Compatibility and enable the Run this program as an administrator checkbox (1).
1 Applicability
This job card applies to all PCs.
2 Personnel requirements
1 Field Engineer
Time: 15’’
3 Configuration
1. Login to Settings. The default credentials are:
- Name: admin
- Password:admin
2. On the tab "Local", fill in the following fields:
1. Name of station: this name is used as the station name in DICOM evidence documents (such as
reports or presentation states) created with MammoWorkstation or MammoDS.
2. Name of institution: this name is used as the institution’s name in DICOM evidence documents
(such as reports or presentation states) created with MammoWorkstation or MammoDS.
1
2
3. Own AE Title (Other DICOM partners, e.g. a PACS archive, must know this AE Title.)
4. Insert the path to the folder "dbcache" (folder for the local image database) created earlier. The
file cache must be on a separate partition. For MammoWorkstation this is a dedicated RAID 0
partition.
5. Establish the cache size (minimum size is 150GB, maximum size is 80% of the harddisk size).
3. Click [OK] after completion. A warning pop-up comes up (see below). Confirm this warnings by
clicking [OK].
4 Verification (MammoWorkstation)
1. Login to MammoWorkstation using
- Login: demo
- Password: demo
This completes the mandatory configuration.
Job Card IST A17 - Reporting mask configuration for MammoDS Chapter 4
1 Applicability
This job card applies to MammoDS / TechInsight.
2 Personnel requirements
1 Field Engineer
Time:5’’
3 Hardware requirements
• HP Compaq dc7900 Base Model Convertible Minitower PC.
4 Introduction
If the Reporting Module is licensed, the Reporting mask in the IDI application will be available.
While setting up a MammoDS, the Reporting Mask Type needs to be adapted to meet the requirements
of the MammoDS.
Job Card IST A18 - Desktop theme configuration for Windows XP Chapter 4
1 Applicability
This job card applies to MammoWorkstation and MammoDS.
2 Personnel requirements
1 Field Engineer
Time: 5’
3 Procedure
Note:
It is not necessary to configure the desktop theme under Windows 7.
1. Navigate to:
• C:\Program Files(x86)\ImageDiagnost\MammoWorkstation\desktop-theme for
MammoWorkstation
• C:\Program Files\ImageDiagnost\MammoWorkstation\desktop-theme for MammoDS.
2. Double-click "imagediagnost_MWS_standard".
Job Card IST A19 - Additional language files for Windows XP Chapter 4
1 Applicability
This job card applies to all PCs.
Note:
This job card only applies to countries where either East Asian languages or complex script and
right to left languges are used.
2 Personnel requirements
1 Field Engineer
Time: 5’
3. Enable the check box next to Install files for East Asian languages.
➜ at the dialog box informing you about the required space for these language files, click [OK].
4. Enable the check box next to Install files for complex script and right-to-left languges (including Thai).
➜ at the dialog box informing you about the required space for these language files, click [OK].
5. Click [Apply].
6. After installation, you will be asked to restart your computer. Click [Yes].
1 Applicability
This job card applies to MammoWorkstation and MammoDS.
2 Personnel requirements
1 Field Engineer
Time: 10’
3 Hardware requirements
• IDI keypad, P/N 5314876
4 Installation
Note:
Make sure that the right file C:\Program Files (x86)\ImageDiagnost\MammoWorkstation\keypad\
[yyyymmdd]_keypad_v[version]CAD.mws (e.g. 20090519_keypad_v45CAD.mws) version is used.
Make sure that the correct keypad version 1.0.1.2 of MacroWorks II keypad is used. This version
is included in the MammoWorkstation Software version 4.6.0.
1. Attach the keypad device via USB. You can choose any available USB port.
Note:
Before you start the installation, make sure the keypad device is properly connected to the
MammoWorkstation.
2. Activate the keypad by turning the wheel on the right side of the keypad up: the control button will
turn red.
3. Execute the installer mwii.v[current version].exe from the install folder and follow the instructions.
5. After the restart, an icon named "X-keys MacroWorks II" will be visible in the system tray.
This icon indicates that the device driver has been correctly installed and loaded.
5 Verification of installation
1. Right click on the icon to display the context menu.
The displayed context menu must look like the following:
2. In case the system tray icon is not visible, reactivate it by going to Start ➜ All Programs ➜ P.I.
Engineering ➜ MacroWorks II ➜ MacroWorks II.
3. If this fails, right click on the taskbar and open the task Manager.
4. Highlight the process "MacroWorks.exe" on the tab "Processes" and click on [End Process].
6 Configuration
1. To start the configuration, open the keypad GUI (graphical user interface) and select "Xkeys
Desktop" from the X-keys’ task bar icon ’s context menu or scroll the sliding controller (1) anti-
clockwise.
Note:
The keys are blank as no
functions have been assigned yet.
7 Verification of configuration
1. Once the xkeys layout has been loaded, check that shortcuts are assigned to each key:
Note:
Correction for MammoWorkstation 4.3.0: Replace <F2> with "Ctrl-M" and <F3> with "Ctrl-C".
Note:
The following table shows an example of xkeys settings:
Table 1
2. Come back to this section for validation purpose as soon as images have been imported into the
"dbcache".
The shortcut configuration setting shown in the task before is equal to the following labels which can
be added to the keys on the keypad:
Spare keys
(if no reporting)
Note:
Available commands are listed at the end of the user manual.
The keypad layout can be customized using the MacroWorks II software. Labels for spare keys are
provided in /Program Files/ImageDiagnost/MammoWorkstation/keypad/.
3. Click the [Clear] button to unassign the current keyboard layout for this specific key.
b. Navigate to
C:\Program Files\ImageDiagnost\MammoWorkstation\keypad.
1 Applicability
This job card applies to MammoWorkstation and MammoDS.
2 Personnel requirements
1 Field Engineer
Time: 10’
3 Hardware requirements
• IDI keypad, P/N 5314876
4 Installation
Note:
Make sure that the right file C:\Program Files (x86) \ImageDiagnost\MammoWorkstation\ keypad\
[yyyymmdd]_keypad_default.mws (e.g. 20200617_keypad_default.mws) version is used.
Make sure that the correct keypad version 1.0.1.13 of MacroWorks 3 keypad is used. This version
is included in the MammoWorkstation Software version 4.6.0.
1. Right click on the installation file MW3InstallerWeb.exe from the install folder and click "Run as
administrator":
3. Click [Next].
4. Click [Install].
7. After the restart, attach the keypad via USB. You can choose any available USB port.
8. Windows will automatically install the device drivers.
9. An icon named "X-keys MacroWorks 3" will be visible in the system tray.
This icon indicates that the device driver has been correctly installed and loaded.
5 Verification of installation
1. Right click on the icon to display the context menu.
The displayed context menu must look like the following:
2. In case the system tray icon is not visible, reactivate it by going to Start ➜ All Programs ➜ P.I.
Engineering ➜ MacroWorks 3 ➜ MacroWorks 3.
3. If this fails, right click on the taskbar and open the Task Manager.
4. Highlight the process "MacroWorks3.exe" on the tab "Processes" and click on [End Process] (1).
6 Configuration
1. To start the configuration, press the programming switch to open the MacroWorks3 Graphical User
Interface.
4. Click on [Browse].
9. Click [Next].
10. A short overview of the buttons and the assigned macros appears. Click [Next].
11. Click [Finish]. The Script Conversion Wizard window will close and the main window of MacroWorks3
will become active. On the top, click on the dropdown menu and select your device (1).
12. The "Closing Script!" window appears. In the dropdown menu select your device and click [Assign].
7 Verification of configuration
1. Once the xkeys layout has been loaded, check that shortcuts are assigned to each key:
Note:
Correction for MammoWorkstation 4.3.0: Replace <F2> with "Ctrl-M" and <F3> with "Ctrl-C".
Note:
The following table shows an example of xkeys settings:
Table 2
2. Come back to this section for validation purpose as soon as images have been imported into the
"dbcache".
The shortcut configuration setting shown in the task before is equal to the following labels which can
be added to the keys on the keypad:
Spare keys
(if no reporting)
Note:
Available commands are listed at the end of the user manual.
The keypad layout can be customized using the MacroWorks II software. Labels for spare keys are
provided in /Program Files/ImageDiagnost/MammoWorkstation/keypad/.
3. Click the [Clear] button to unassign the current keyboard layout for this specific key.
4. There are two ways to configure this specific key:
a. Press the keys on the keyboard that are to be assigned.
b. Navigate to
C:\Program Files\ImageDiagnost\MammoWorkstation\keypad.
1 Applicability
This job card applies to all PCs.
2 Personnel requirements
1 Field Engineer
Time: 15’
3 Hardware requirements
• IDI software application CD.
Note:
A validated VNC version supporting Windows 7 can not be provided.
Contact the IDI Service team for further information.
4 Introduction
Virtual Network Computing (VNC) allows to view and fully interact with another computer.
VNC consists of two components: a server, which runs on the computer you want to remotely access,
and a viewer, which runs on the computer you are sitting in front of.
A remote desktop software such as VNC must be available to diagnose and fix problems.
5 Installation
1. Launch the VNC installer. Click [Run].
2. The VNC Setup Wizard will open. Click [Next].
3. Accept the license agreement by selecting the radio button "I accept the agreement".
6 Configuration
Note:
Make sure that port 5900 is open through the firewall (See section Firewall on page 76).
1. Click [Configure] to set a password for incoming connections. This is mandatory.
Click [Apply].
2. Click [Next].
3. Click [Finish]. The installation and configuration of VNC is complete.
7 Verification
7-1 IP address determination
1. Click on Start ➜ Run..., type in "cmd" and hit [Enter].
2. Next type "ipconfig".
3. Or else in the task bar, double click the system tray icon for your local area connection.
4. Go to tab Support.
3. Enter the just configured password when you are prompted for, and click [OK].
Job Card CFG A02 - Local DICOM settings and database configuration
Job Card CFG A02 - Local DICOM settings and database configuration Chapter 4
1 Applicability
This job card applies to all PCs.
2 Personnel requirements
1 Field Engineer
Time: 10’
1
2
3
7 8
4 5
6
11
10
1. Insert a descriptive name for the station. The lenght must not exceed 16 characters. This name will
appear on the login screen.
This name is used as the station name in DICOM evidence documents (such as reports or
presentation states) created with MammoWorkstation or MammoDS.
2. Insert the institution’s name.
This name is used as the institution’s name in DICOM evidence documents (such as reports or
presentation states) created with MammoWorkstation or MammoDS.
3. Displays the local computer’s name (on which the application is installed). Other DICOM partners,
e.g. a PACS, must know this name.
4. The AE Title of the local DICOM storage SCP. Other DICOM partners, e.g. a PACS archive, must
know this AE Title.
Job Card CFG A02 - Local DICOM settings and database configuration
5. Use C-ECHO to test the availability of the local DICOM Storage SCP. A green checkmark implies
that the SCP is active. If verification fails, a red cross appears. Typical cause for failure during
installation is an invalid license.
6. Port for the local DICOM Storage SCP. Port 104 is the standard port used for DICOM
communication. Keep this port unless the network administrator of the customer site tells you
otherwise. Other DICOM partners, e.g., a PACS, must know this AE Title.
7. The Image Diagnost authentication server provides user accounts for MammoWorkstation and
MammoDS. It also manages the administrator accounts for all other IDI products.
The location of the authentication server depends on the availability of a CollaborationServer:
a. There is no CollaborationServer: For a stand-alone installation, the authentication server is part of
the local installation, i.e., "localhost". It is already preconfigured. There is nothing to do.
Note:
When logging on to the MWS, MDS or DSH, the CollaborationServer’s user database will now be
contacted for authentication.
Job Card CFG A02 - Local DICOM settings and database configuration
8. Verify network connectivity to the authentication server. A green checkmark reflects a proper network
connection.
9. The path to the local file cache. This is the folder "dbcache" (created earlier).
The file cache must be on a separate partition. For MammoWorkstation this is a dedicated RAID 0
partition.
10. Set the size that will be reserved for local database, i.e., the space that the folder "dbcache" can
allocate.
At least 20% of the hard-disk must remain free.
11. For MammoWorkstation and MammoDS only:
If active, the user can search the local database (cache). This is recommended only for
demonstration purpose, if no PACS archive is available or the if PACS archive does not provide
appropriate DICOM Q/R services.
Normally, only worklist and PACS sources are available to the user. This has the advantage that the
user has access to the complete patient information and that evidence documents (reports,
presentation states) are automatically stored in the PACS archive. MammoWorkstation takes care of
prefetching images from the PACS and keeping a copy of these images on the local database
(cache).
Job Card CFG A02 - Local DICOM settings and database configuration
1 Applicability
This job card applies to MammoWorkstation and MammoDS.
2 Personnel requirements
1 Field Engineer
Time: 10’
3 Introduction
The Image Diagnost user manager allows identification of physicians working with the
MammoWorkstation and MammoDS. The accounts of administrators with access to settings are
managed.
The location of the user manager depends on the availability of a CollaborationServer:
a. There is no CollaborationServer: For a stand-alone installation, the user manager is part of the
local installation. User accounts are managed on the local PC. In this case, the User Manager
URL refers to "localhost".
b. There is a CollaborationServer: In this case, the CollaborationServer manages user accounts
centrally. In this case, the User Manager URL refers to the IP number of the CollaborationServer:
Note:
The tab "Local" in "Settings" displays the currently configured authentication server that provides
the User Manager.
Note:
Only users with the role "admin" can
configure the MammoWorkstation.
At least one user must have the
admin role. So do not delete the
admin account.
4. To validate the account creation, start MammoWorkstation and login with the just created account.
Note:
In the application login window, the Name will be the ID of the User account creation window.
• Once the role is created, it will be displayed in the User account creation window and can be
selected as a ROLE member:
2. Go to tab Routing.
- Choose the rule "Create worklist on import". The AE title of this rule is AE_MAMMO_WL.(1)
- Tick the box "Notify Collaboration Server" (2) and insert the right IP address (3) of the
Collaboration Server.
Note:
The IP address can either be "localhost" if the workstation includes a Collaboration Server, or the
IP address of a separate Collaboration Server.
2 3
The modality needs to be configured to send images to the AE title " AE_MAMMO_WL". The reception of
images at this AE title triggers a notification being sent to the Collaboration Server. The Collaboration
Server then creates a worklist task.
1
2
1. Go to http://localhost:8080/supportinterface/ .
2. Select Configuration Folder (1).
- Click on WfConf.xml (2) to open the file and search for the tag:
<worklist_definitions description= [...]>
Note:
The parameter of this tag states which file contains the active worklist definitions. The default file
is WorklistDefinitions.en.xml.
<item_creation>
<in_conjunction_with>
<local_aetitle>AE_MAMMO_WL</local_aetitle>
</in_conjunction_with>
<apply_preset_params>
<wl_order_task_type_code>SINGLE_READING</wl_order_task_type_code>
<wl_order_task_type_description>Reading</wl_order_task_type_description>
<sps_performing_actor_name>Demo User</sps_performing_actor_name>
<sps_class_uid>1.2.840.10008.5.1.4.31</sps_class_uid>
<sps_status>SCHEDULED</sps_status>
</apply_preset_params>
</item_creation>
This section is responsible for the creation of a SINGLE_READING task scheduled for Demo User when
an image is sent to the AE title AE_MAMMO_WL. This reading task will be available to any user that is
authorized to access the task READING and that is a member of the same group as the Demo User.
task_type_code task_type_description
KVSCREENING_TASK_2 Anamnesis
task_type_code task_type_description
KVSCREENING_TASK_5 Supervision
KVSCREENING_TASK_7 Assessment
CONSULTATION Consultation
SINGLE_READING Reading
DOUBLE_READING_1 Reading 1
DOUBLE_READING_2 Reading 2
When modifying the WorklistDefinition.en.xml file, it is mandatory to make use of these predefined task
types. Certain task types include a listener that will create a consecutive task after completion of the
previous one.
Example: After a "First Reading" task is completed, a "Second Reading" task will automatically be
created. The Second Reading will be followed by a "Consensus Reading" task if the findings of First
Reading and Second Reading were different.
<item_creation>
<in_conjunction_with>
<local_aetitle>AE_MAMMO_WL</local_aetitle>
</in_conjunction_with>
<apply_preset_params>
<wl_order_task_type_code>KVSCREENING_TASK_3</
wl_order_task_type_code>
<wl_order_task_type_description>First Reading<
wl_order_task_type_description>
<sps_performing_actor_name>Site XXX</sps_performing_actor_name>
<sps_class_uid>1.2.840.10008.5.1.4.31</sps_class_uid>
<sps_status>SCHEDULED</sps_status>
</apply_preset_params>
</item_creation>
5-1-3 Verification
1. Login to MammoWorkstation.
2. Send any patient to the rule "Create worklist on import".
Note:
It does not matter if the images received on the rule "Create worklist on import" were sent from
outside (modality) or from within the IDI workstation. The effect is the same.
3. Go to http://localhost:8080/supportinterface/
4. Select Findings Summary(1).
5. Select Worklist Status (2).
1
2
If a First Reading task for User "Site XXX" was created (see image above), the configuration is correct.
Job Card CFG A06 - Language and display preference configuration Chapter 4
1 Applicability
This job card applies to MWS and MDS.
2 Personnel requirements
1 Field Engineer
Time: 10’
1
2
3
4
6
1. Select the language for the user interface.
2. Adjust the brightness of the user interface.
3. Cancel or activate the display of the "Track viewed images warning".
Note:
This warning display may be mandatory in your country (e.g. US). Check the regulation before
canceling this warning. For further information refer to Job Card CFG A40 - Customer specific
setting on page 223 .
5. Edit information displayed as annotations on top of each image. The default setting follows the IHE
(Integrating the Healthcare Enterprise) recommendation. The default setting can be restored with the
button [Reset to IHE standard].
Note:
The first line, containing Patient ID and Patient’s Name is mandatory. Consequently, this line is
grayed out and cannot be edited.
Default DICOM attributes that fulfill IHE and MQSA requirements are:
Note:
The entry in bold script in the column "Dislayed image information in tab Preferences: Text string
+ DICOM tag ID" is the one corresponding to the attribute. Additional text was only added for easier
localization of the quoted attribute.
0010,0030 Patient's Birth Date "DOB %y, Age %s", 10-30, 10-1010
0008,0022 Acquisition Date "%s, Acq date %y %t", 8-60, 8-22, 8-32
0008,0032 Acquisition Time "%s, Acq date %y %t", 8-60, 8-22, 8-32
0018,1191 Anode Target Material "A/F %s/%s, D %fmm, Fcomp %fN", 18-
1191,18-7050, 18-11A0, 18-11A2
0018,11A0 Body Part Thickness "A/F %s/%s, D %fmm, Fcomp %fN", 18-
1191,18-7050, 18-11A0, 18-11A2
0018,700C Date of last Calibration "DetID %s, last calibration %y", 18-
700A, 18-700C
0008,1090 Manufacturer Model Name "%s, %s, SN %s" , 8-70, 8-1090, 18-
1000
0018,1000 Device Serial Number "%s, %s, SN %s" , 8-70, 8-1090, 18-
1000
0018,0010 Contrast Bolus Agent "Agent %s, Vol %s, Concentration %s",
18-10, 18-1040, 18-1049
0018,1040 Contrast Bolus Volume "Agent %s, Vol %s, Concentration %s",
18-10, 18-1040, 18-1049
0018,1049 Contrast Bolus Ingredient "Agent %s, Vol %s, Concentration %s",
Concentration 18-10, 18-1040, 18-1049
1 Applicability
This job card applies to all PCs.
2 Personnel requirements
1 Field Engineer
Time: 20’
3 Introduction
Partners are external DICOM devices the Image Diagnost applications communicate with. Partners can
provide a source and/or a destination. Typical partners are a local or remote PACS, other
MammoWorkstations, or a routing target such as DICOM Shuttle.
MammoWorkstation and MammoDS support as source: DICOM Storage SCP, DICOM Query-Retrieve
SCP, and DICOM-Media (DICOMDIR). Destinations can be DICOM Storage SCU or DICOM-Media.
CollaborationServer and Dicom-Shuttle support as source: DICOM Storage SCP. Destinations can be
DICOM Storage SCU.
3
4
5
6
7
8
9
10
1
11
12
13
5. TCP/IP address of the DICOM partner. Use [Ping] to verify that the IP is reachable.
6. Only for MammoWorkstation and MammoDS:
If the DICOM partner supports DICOM Query and Retrieve service, MammoWorkstation can query
patient information and retrieve images from this partner.
Note:
MWS does not support Q/R (it can only be destination).
7. AE title of the partner DICOM Q/R SCP. Use [C-ECHO] to verify the presence of a DICOM SCP.
8. Port of the partner DICOM Q/R SCP.
9. Select the type of Query/Retrieve model supported by the DICOM partner. Study-Root + Relational
provides the best performance. Select this if supported by the PACS.
10. The DICOM partner provides DICOM Storage SCP service. It can be target of send-actions.
AE Title and Port of the Storage SCP.
Use [C-ECHO] to verify the presence of a DICOM SCP.
11. Choose if images are to be compressed when sent.
Options are: None, JPEG 2000 lossless, JPEG 2000 1:10, JPEG 2000 1:15.
12. Define where the evidence documents created at MammoWorkstation and MammoDS (reports/
DICOM SR and presentation states/DICOM PR/DICOM encapsulated PDF (DOC)) should be sent.
Default is the partner itself. I.e., for cases from a PACS source reports would be sent back to the
same PACS.
Sometimes, an alternative target is required.
Example: The destination could be a rule on the DicomShuttle.
- This rule could contain preconditions e.g. only PR/SR (and/or DOC) depending on what the
archive accepts.
- This rule could contain a list of other members of a screening unit and the PACS.
The evidence documents would then be distributed amongst all recipients on that list.
Note:
For further information on Routing rules, refer to Job Card CFG A14 - Routing rules configuration
on page 189.
13. The DICOM partner provides DICOM Storage Commitment SCP. The DICOM Shuttle can pass
through storage commitment requests from the modality to the archive.
5 Verification
Verify your configuration using the [Ping] and [C-ECHO] functions.
Verification status:
• : Verification in progress
• : No service found.
1 Applicability
This job card applies to MammoWorkstation and MammoDS.
2 Personnel requirements
1 Field Engineer
Time: 5’
3 Prerequisite
Ensure that your data source is a DICOM Media. This requires that the data source (CD/DVD/folder) is
organized in conformance to the DICOM Media format standard. To verify this, check that this structure
can be seen in the root folder. The root folder must show a file "DICOMDIR":
Note:
To import DICOM images that are not in a DICOM structured media, use the tool "Import files from
disk" accessible from menu Start ➜ Program Files ➜ Image Diagnost ➜ Tools.
4 Procedure
1. Insert any CD in the CD driver of the MammoWorkstation.
2. Enter the settings of the MammoWorkstation and select the tab "Communication".
3. Configure a new DICOM Partner (Dicom – CD).
4. Insert the path to your CD/DVD-drive or to a folder holding the images for the import.
Note:
Ensure that the path directs to the "DICOMDIR" file. In case of a DICOM CD/DVD, use the path of
the drive, e.g., "E:\".
5. Click [OK] (the IDDicomReceive.scp service will be restarted and a C-Echo test will be performed).
6. Exit the IDI application (otherwise the new DICOM Partner won’t be visible).
5 Validation
1. Login as a user to MammoWorkstation and perform a seek mode on the DICOM Partner holding the
data as shown below (Interval of time "unlimited").
The images are still on the DICOM media and not yet in the local cache. The prefetch status is
therefore displayed red.
2. Click [Prefetch] to load the images from the DICOM media onto the local database (cache). To select
several cases at once, press [Ctrl] and contemporaneously select the desired cases with the mouse
cursor.
After the import is finished, the images will have a green prefetch status.
Job Card CFG A12 - Desktop synchronisation from third party client to MammoWorkstation (remote control)
Job Card CFG A12 - Desktop synchronisation from third party client to
MammoWorkstation (remote control) Chapter 4
1 Applicability
This job card applies to MammoWorkstation and MammoDS.
2 Personnel requirements
1 Field Engineer
Time: 20’
3 Introduction
MammoWorkstation supports context synchronization between a third party client and
MammoWorkstation. For example, a third party client may be a RIS, MIS or SIS application. This section
describes desktop synchronisation of the current context from the third party client to the
MammoWorkstation.
To verify proper configuration and perform an end to end test, access to the third party client is required.
1
2
Job Card CFG A12 - Desktop synchronisation from third party client to MammoWorkstation (remote control)
1. Allows other applications (RIS, MIS, SIS, etc.) to synchronize the context with the
MammoWorkstation. Images for the current context will be shown in MammoWorkstation.
2. Specify which source is queried for images, in case of a remote control event. This should be the
PACS associated with the RIS application that remote controls MammoWorkstation.
5 Validation
5-1 Allow remote control
1. To allow remote control by other applications, login to the settings, go to tab Remote Control and
enable this feature.
For test reasons, try one of the three URL invocation options:
1. Patient ID:
• The RIS Client must call
http://localhost:8080/ReceiverService/idimwscontrol.jsp?loadType=PATID&
loadCurrent=patient-id, with the Patient ID as loadCurrent parameter.
Job Card CFG A12 - Desktop synchronisation from third party client to MammoWorkstation (remote control)
•If the call was successful, the Internet Explorer displays the message:
IDI_REMOTE_CONTROL_EVENT_OK
• This call invokes the display of the patient with the Patient ID XXXXXXX on the
MammoWorkstation.
2. Study Instance UID:
• The RIS Client must call
http://localhost:8080/ReceiverService/idimwscontrol.jsp?loadType=
STUDYUID&loadCurrent=study-uid, with the StudyUID as loadCurrent parameter.
• To get the Study Instance UID:
- Go to http://localhost:8080/supportinterface/
default login admin/admin
- Select Import Processes (1)
3
1
- Select an arbitrary File path(4) ➜ the DICOM File Header Listing will be displayed.
Job Card CFG A12 - Desktop synchronisation from third party client to MammoWorkstation (remote control)
- Search for the Study Instance UID, with the Tag 0020,000D
•If the call was successful, the Internet Explorer displays the message:
IDI_REMOTE_CONTROL_EVENT_OK
• This call invokes the display of the patient with the StudyUID XXXXXXXXXXXX on the
MammoWorkstation.
3. Accession Number
• The RIS Client must call
http://localhost:8080/ReceiverService/idimwscontrol.jsp?
loadType=ACCNR&loadCurrent=accession, with the Accession Number as loadCurrent
parameter.
• To get the Accession Number:
- Go to http://localhost:8080/supportinterface/
- Select Import Processes (1)
3
1
Job Card CFG A12 - Desktop synchronisation from third party client to MammoWorkstation (remote control)
- Select an arbitrary File path (4) ➜ the DICOM File Header Listing will be displayed.
- Search for the Accession Number: Tag 0008,0050
• If the call was successful, the Internet Explorer displays the message:
IDI_REMOTE_CONTROL_EVENT_OK
• This call invokes the display of the patient with the Accession Number XXXXXXXXXXXX on the
MammoWorkstation.
Job Card CFG A12 - Desktop synchronisation from third party client to MammoWorkstation (remote control)
Job Card CFG A13 - Desktop synchronisation from MammoWorkstation to third party client (remote control)
1 Applicability
This job card applies to MammoWorkstation and MammoDS.
2 Personnel requirements
1 Field Engineer
Time: 30’
3 Introduction
MammoWorkstation supports context synchronization between a third party client and the
MammoWorkstation. For example, a third party client may be a RIS, MIS or SIS application.
This section describes desktop synchronisation of the current context from the MammoWorkstation to
the third party client.
To verify proper configuration and perform an end to end test, access to the third party client is required
1
2
3
1. Activate remote control: Allows the MammoWorkstation to synchronize the context with another
application (RIS, MIS, SIS, etc.).
Job Card CFG A13 - Desktop synchronisation from MammoWorkstation to third party client (remote control)
1 Applicability
This job card applies to DicomShuttle and CollaborationServer.
2 Personnel requirements
1 Field Engineer
Time: 20’
3 Introduction
Routing rules allow associating actions to storage AE titles. If a document is sent to the rule AE title, the
rule is executed.
2
3
4
1
6
7
1. List of all rules created. Create a new rule with the [plus] button.
Note:
Do NOT delete any of the default routing rules. Only modilications are allowed.
2. The display name of the rule. This should be descriptive.
3. AE title of the rule.
Note:
Always reference a rule by it’s AE title.
Rule display names are only relevant for the user interface.
4. Optionally, formulate a precondition under which this rule is executed. If the precondition is not met,
no action takes place.
b. Enter the DICOM tag and value and whether this condition is to be met (==) or not (!=).
This precondition means that the rule only processes documents where the DICOM tag 0008,0060
has the value "SR" or "PR". All other documents will be ignored.
5. Change DICOM Header.
a. It is possible to change DICOM tags of incoming documents.
WARNING
Use this option only if you are exactly aware of the consequences.
Modification of DICOM header information for incoming documents may destroy image and/or
patient data. A user must be exactly aware of the consequences when using this option
There are many DICOM tags that shall not be modified. In certain situations changing a tag requires
changing other tags as well, e.g., the instance UID.
b. Define one or more actions. Each action can have a precondition of its own.
Formulate the tags and values that need to be corrected. Refer to section How to correct patient
demographics on page 480 for details
7. Option to maintain a MPI (Master Patient Index) to anonymize patient’s identity towards external
targets (i.e. external diagnostic readings) and keeping track of the true identity internally. (Do not
use.)
8. Determine if images are decompressed when imported.
Must always be checked for all rules on MammoWorkstation and MammoDS.
On a DicomShuttle, images typically remain untouched in compressed format for further routing.
9. Configure send actions.
a. Click [New] to add targets to the send action.
The "Forward to" drop down menu lists all available destination DICOM partners (see Job Card CFG
A08 - DICOM partners configuration on page 173).
Note:
When adding a new DICOM partner the IDI application has to be terminated and restarted in order
for the new target to be visible on the "Forward to" drop down menu.
b. Choose the compression rate.
c. Choose whether the send action is to take place immediately or at a specific time.
The document does not contain the correct DICOM tags. Instead a private tag (0019,0010) is
provided. The rule corrects the DICOM header tags for view position and laterality.
Use Case:
After the reading for a client has been completed on a selected MWS/MDS, a report (SR) is being
created on this machine. This SR needs to be sent to a rule (listed on DSH) that distributes the SRs to all
other machines MWS/MDS (and especially to the Collaboration Server).
The following example shows the configurations needed to distribute the SR to different DICOM nodes.
2. Define a distribution rule on each MWS/MDS to send the created SR to the DICOM Shuttle (DSH).
a. Define a new partner on the MDS/
MWS
- Login on each MDS and MWS
that creates SR and go to tab
"Communication".
- Create a new partner (which will
be the DicomShuttle) and for the
target, choose the name of the
routing rule (configured in the
screenshot above) hosted on the
DicomShuttle "SR_FROM_B4".
- Test connectivity with ping and
[C-ECHO].
- Click [OK] and exit the
application to apply the changes.
- Exit out of the logon screens as
well.
This will result in all SR documents being automatically sent to the distribution rule on the DSH (which
then sends it to the CS and all other machines except the one where the SR was generated).
1 Applicability
This job card applies to MammoWorkstation, MammoDS, DicomShuttle with WorklistBroker,
CollaborationServer and DigitizingStation.
2 Personnel requirements
1 Field Engineer
Time: 10’
3
4
5
6
7
8
9
1 10
11
12
13
14
15
1. Overview of configured worklists. Create a new worklist with the plus button.
2. Display name of the worklist.
3. Description of the worklist.
4. AE title used to identify myself (Calling AE title).
If left blank, this is the same as the local AE title (tab Local).
To access a worklist scheduled for a specific modality, the AE title of the modality can be entered
here.
5. TCP/IP address of the worklist server. Use [Ping] to verify network connection.
6. AE title of the worklist server SCP. Use [C-ECHO] to verify DICOM connection.
7. Port on which the worklist server SCP listens.
8. Search Keys: Filter the worklist from the worklist server by the desired criteria.
Note:
The Image Diagnost WorkflowBroker provides client side filtering. In case the worklist server does
not support a search key, i.e., the server provides too many entries, the application locally filters
the answer and removes the surplus entries.
>Modality (0008,0060)
Requested Procedure
Patient Identification
Patient ID (0010,0020)
Patient Demographic
9. Enable "PPS" if the workflow server accepts DICOM Performed Procedure Steps (PPS). In that case
MammoWorkstation and MammoDS can commit worklist task status, i.e., the worklist server knows
when to close the task.
Note:
Must always be checked in case of a CollaborationServer.
10. Check "MammoWorkstation" if you configure a worklist for MammoWorkstation or MammoDS. The
worklist will then be available to the user in the Worklist tab of the Patient Manager.
11. Associated archive: location where the MWS will search for images corresponding to the worklist
task.
Note:
The local cache will always be searched no matter what this field contains.
Usually, this field contains the PACS archive that is used in combination with the RIS. Each worklist
can be associated to its own source.
Only under certain circumstances the local cache is configured (if the archive cannot perform Q/R
tasks or is exceptionally slow). It is then recommended to prefetch images from the PACS before
performing reading tasks.
12. Check "Image Diagnost CollaborationServer" if the worklist is provided by a CollaborationServer. In
that case, the worklist contains additional user and group information.
(only available with license module "Workflow").
Note:
Even though "Free search in worklist" is greyed out it has to be selected.
14. Check "WorklistBroker" to activate the worklist broker mode. In that case the worklist is stored in the
local WorklistBroker and is available from the WorklistBroker DICOM Worklist SCP.
Note:
For further information refer to Job Card CFG A20 - WorklistBroker configuration on page 205
15. Adjust port for HL7 communication.
Note:
For further information, see Job Card CFG A32 - HL7 export configuration on page 227.
1 Applicability
This job card applies to DICOM Shuttle.
2 Personnel requirements
1 Field Engineer
Time:
3 Introduction
To improve the performance of a DICOM Shuttle, two configurations can be modified.
4. Look for the following tag: <allow_roi_calculation description=[...]>, and set the value
of this tag to 0:
5. After any modification in the CcConf.xml, restart the Apache Tomcat service.
- Look for the following tag: <icon_width_pix description=[...]>, and set the value of this
tag to 0:
4. After any modification in the CcConf.xml, restart the Apache Tomcat service.
1 Applicability
This job card applies to DicomShuttle with WorklistBroker and CollaborationServer.
2 Personnel requirements
1 Field Engineer
Time: 15’
3 Introduction
The local DICOM Worklist SCP provides DICOM modality Worklist to external devices. The Dicom-
Shuttle WorklistBroker provides worklists for modalities. The CollaborationServer provides worklists for
MammoWorkstation and MammoDS.
4 Configuration
4-1 Worklist SCP
The default configuration of the local Worklist SCP is:
AE-Title: AE_WORK
TCP/IP port: 4005
1. Keep this AE-Title and port unless the network administrator of the customer site tells you otherwise.
Other DICOM partners, e.g. a modality, must know this AE Title.
2. To edit the local Worklist SCP parameters:
- Go to http://localhost:8080/supportinterface/
- Open the Configuration Folder.
- Edit IDWorklistSCP.xml
1 Applicability
This job card applies to DicomShuttle with WorklistBroker and CollaborationServer.
2 Personnel requirements
1 Field Engineer
Time: 15’
3 Introduction
The ImageDiagnost WorklistBroker has two functions:
1. Merge several worklists from exernal worklist servers into one single worklist and present this
worklist to a mammography station.
2. Duplicate an existing worklist in order to present it to more than one mammography station.
The WorklistBroker is part of the functional range of the DicomShuttle. It is a separate piece of software
though and has to be licensed separately.
The WorklistBroker acts as a DICOM Worklist SCU when querying Worklist from remote worklist server
such as a RIS. It uses the DICOM Worklist SCP to provide worklists to modalities or MammoWorkstation
and MammoDS.
3. Configuration example:
The screenshots below shows 2 worklists from 2 different sources (RIS_WORKLIST1 and
RIS_WORKLIST2).
Both reference the same worklist rule "EXTERNAL_TASK" , which means that they are implicitly
merged into one single worklist.
The EXTERNAL_TASK rule states the target (Mammography station) of the merged worklist. It will
be displayed at the mammography station "ANY_X_RAY_MASCHINE".
Click on [Edit] to enter the details:
Worklist
target
2. Configuration example:
The screenshots below shows one worklist (RIS_WORKLIST1), which references the worklist rule
item "DIST2MAMMO123". Two copies of this worklist rule item are created. Each item states a
different target.
a. Click [Edit rules] and create two copies of the existing rule so that there will be three instances.
target 1
Create
or delete
target 2
target 3
b. Combine the two functions of the worklist broker to merge several worklists into one single
worklist and present it to several mammography stations.
Job Card CFG A25 - Storage Commit Broker (SC Broker) configuration
Job Card CFG A25 - Storage Commit Broker (SC Broker) configuration Chapter 4
1 Applicability
This job card applies to DicomShuttle with Storage Commit Broker.
2 Personnel requirements
1 Field Engineer
Time: 45’
3 Introduction
The Storage Commit Broker is an intermediary between the PACS (document archive) and the modality.
It is responsible for scheduling the image routing as well as the DICOM Storage Commit requests and
responses. The Storage Commit Broker is needed in case a modality cannot directly send the DICOM
Storage Commit request to the PACS (document archive).
The DICOM Storage Commit service is needed to enable the modality to check whether image data was
successfully archived by the PACS (document archive).
The DICOM Storage Commit minimizes the risk of image data loss. However, the DICOM Storage
Commit cannot prevent the loss of image data since image data could be deleted or changed in the
PACS (document archive).
Note:
The IDI Storage Commit Broker schedules the image and storage commit routing.
Job Card CFG A25 - Storage Commit Broker (SC Broker) configuration
All necessary configurations can be done within the IDI MammoWorkstation GUI:
• Login to Settings. The default credentials are:
- Name: admin
- Password: admin
Job Card CFG A25 - Storage Commit Broker (SC Broker) configuration
2. Setup an additional DICOM Partner "Remote DICOM archive" via the Select archive type dialog:
Job Card CFG A25 - Storage Commit Broker (SC Broker) configuration
To ensure that the Storage Commit Broker forwards the DICOM Storage Commit response to the
modality that initiated the DICOM Storage Commit request in the first place, the Storage Commit Broker
needs the following information of the modality:
- DICOM AE Title
- IP-Address
- TCP Port
Job Card CFG A25 - Storage Commit Broker (SC Broker) configuration
Job Card CFG A25 - Storage Commit Broker (SC Broker) configuration
4. Look for the following tag: storage_commit_delay_time description= "" [...]>, and set
the value of this tag to the desired value:
Examples:
• for 60 seconds delay time, enter >60<.
• for 1 hour delay time, enter >1h<.
• for 30 minutes delay time, enter >30m<.
5. Click [Save] to save for applied modifications.
6. After any modification in the CcConf.xml, restart the Apache Tomcat service:
Note:
Generally it is possible to configure a delay time in the PACS (document archive) as well. This way
the PACS waits for the indicated delay time before processing the DICOM Storage Commit
requests.
For delay time configuration in the PACS, contact the respective PACS (document archive)
provider.
1 Applicability
This job card applies to MammoWorkstation.
2 Personnel requirements
1 Field Engineer
Time: 10’
3
4
4 Verification
1. Press [Check Printer] to perform basic functionality and connectivity test of the printer.
The output shall describe the printer name and supported formats.
2. In case the printer is not identified, MammoWorkstation provides default values and the printer name
is "default printer".
In this case:
a. Start the MammoWorkstation and open a patient.
b. Click on the [Print] button and press [Info].
A popup window will appear showing the DICOM information of the printer.
Illustration 1 Printer DICOM information
c. Make sure these information fit 100% to the printer settings in the PrinterDefaults.xml. To check
this:
- Go to http://localhost:8080/supportinterface/
- Open the Configuration Folder.
- Edit the PrinterDefaults.xml
For the example in Illustration 1, the printer configuration line should be:
<PrinterS0P PrinterName="DRYPIX1"
Manufacturer="FUJIFILM Corporation"
ManufacturerModelName="DRYPIX"/>
d. If the information in the DICOM information of the printer differ from the one in the
PrinterDefaults.xml, correct the PrinterDefaults.xml.
e. Restart the Apache Tomcat service.
3. If the printer model is not included in this file, contact the IDI service team.
4. Copy these three lines and set the values to the ones of your printer.
Note:
Make sure to use the correct DICOM information of the printer.
5. Restart the Apache Tomcat service.
! Notice:
If the printer was customized, send the printer information and printer configuration to
service@imagediagnost.de.
1. Go to http://localhost:8080/supportinterface/.
2. Open the Configuration Folder.
3. Edit the PrinterDefaults.xml
1 Applicability
This job card applies to MammoWorkstation and MammoDS.
2 Personnel requirements
1 Field Engineer
Time: 10’
1. Physical screen size must be configured for true scale (1:1) scaling to operate properly.
Size values must exactly match the dimensions of the two portrait monitors (the visible area). Either
refer to the manufacturer’s specification or take exact measurements with a ruler.
2. Correction of the gray scale display function. Change only after consultation with IDI.
With Gamma correction for digitized film, you can simulate a light-box appearance on screen (values
of 0.4 to 0.5)
3. The resolution an image has when transferred from the IDI image folder to a third party application
via drag-and-drop.
1 Applicability
This job card applies to MammoWorkstation and MammoDS.
2 Personnel requirements
1 Field Engineer
Time: 5’
! Notice:
This warning display may be mandatory in your country (e.g. US).
Check the regulation before applying the following:
To cancel the display of this warning:
1. Login to MammoWorkstation ➜ Settings.
- name: "admin"
- password: "admin"
2. Select the tab Preferences.
3. Untick the Checkbox next to "Track viewed images: Show warning upon closing a case in case some
images were not reviewed in full screen resolution ".
Note:
The time range is configurable by service.
! Notice:
This warning display may be mandatory in your country (e.g. US).
Check the regulation before applying the following:
To cancel the display of this warning:
1. Go to http://localhost:8080/supportinterface.
2. Open the Configuration Folder.
3. Edit MammoWorkstation.ini.
1 Applicability
This job card applies to MammoWorkstation and MammoDS.
2 Personnel requirements
1 Field Engineer
Time: 5’
3 Procedure
1. Open GlobalConfig.ini file in folder C:\Program\ImageDiagnost\MammoWorkstation\cdViewer\config.
2. Replace "Language = de " by the shortcut of your language (e.g. "en" for english)
3. Configure a new DICOM Partner (Dicom – CD).
Note:
Available languages and shortcuts are listed in C:\Program\ImageDiagnost\MammoWorkstation
\bin
1 Applicability
This job card applies to MammoWorkstation and MammoDS.
2 Personnel requirements
1 Field Engineer
Time:
6. Verify that HL7 export is enabled by going to Finding summary ➜ Worklist Status.
A new column (SRHL7 Export Status) should be visible now:
1 Applicability
This job card applies to MammoWorkstation and MammoDS.
2 Personnel requirements
1 Field Engineer
Time: 60’
3 Introduction
The configuration of unscheduled reporting is required for the following use case:
1. The physician selects a case in an external RIS (Patient Selection RIS). The RIS opens the selected
case in the MammoWorkstation via remote control.
2. The physician can read the case and enter the report in the MammoWorkstation. The report is then
going to be transferred to an IDI CollaborationServer.
3. The IDI CollaborationServer sends the report data to the external RIS (Report Storage RIS) via HL7
export.
Note:
To review the finished report, the physician needs to open the report in the Report Storage RIS.
4 Unscheduld reporting
The configuration of unscheduled reporting requires the following four configuration steps:
1. Configuring the MammoWorkstation.
2. Configuring remote control in the RIS (Patient Selection RIS).
3. Configuring the IDI CollaborationServer.
4. Configuring the RIS (Report Storage RIS) responsible for storing the sent report data .
! Notice:
For the RIS configuration it is important that the"MWSRemoteControl.exe" tool is always started
with the option "- -scheduled-task".
Note:
For further information on how to configure HL7 Export, refer to Job Card CFG A32 - HL7 export
configuration on page 227.
5. Restart the Apache Tomcat service.
1 Applicability
This job card applies to all PCs.
2 Personnel requirements
1 Field Engineer
Time: 15’
3 Introducation
Starting with software version 4.6, the MammoWorkstation will be able to identify and display CESM
images.
The configuration files required for the CESM feature are not automatically transferred during the
upgrade procedure.
To enable the CESM feature two configuration files need to be manually updated:
• HangingProtocolConfig.xml
• WindowLevelConfig.xml
4. Copy (select desired text and press "Ctrl + C") the content of the HangingProtocolConfig_CESM.xml
from the xml element <Protocol name="CESM"> to the xml element </Protocol>
[...]
7. Paste (press "Ctrl +V") the content of the HangingProtocolConfig_CESM.xml into the
HangingProtocolConfig.xml after the xml element <Protocol Config
Verion="preliminary"> (first line).
1
5. Insert the following section after the xml element </imagetype> (1).
<imagetype name="GE CESM" description="GE CESM" presetName="NORMAL"
presetIsVoiLut="no" defaultVoiFunction="SIGMOID">
<taggroup boolop="AND">
<tag id="0x00080008" vr="CS" valuemask="DERIVED*"/>
<tag id="0x00080060" vr="CS" value="MG"/>
<tag id="0x00080016" vr="CS" value="1.2.840.10008.5.1.4.1.1.1.2"/>
<tag id="0x00080070" vr="CS" value="GE MEDICAL SYSTEMS"/>
<tag id="0x00181030" vr="CS" valuemask="CESM"/>
</taggroup>
</imagetype>
7. Check the applied modifications. The result should look like the following:
1 Applicability
This job card applies to all PCs.
2 Personnel requirements
1 Field Engineer
Time: 20’
3 Introduction
The acceptance test is mandatory to complete the installation of the system.
The acceptance test must be repeated:
- Whenever third party software is installed on the same PC.
- Whenever a new interface to a third-party system is initially configured.
- Whenever a connected third-party system is changed (only the tests that apply to the specific
interface must be performed).
4 System configuration
Document the exact system configuration this acceptance test is applied to (refer to the Image Diagnost
Software CDAcceptanceTest_Report.pdf).
Note:
Product type refers to Mammography Modalities, PACS, RIS, CAD and Printers.
5 Acceptance test
5-1 File-path of local image cache
5-1-1 Purpose
• Validate the hard-disk resources for the local cache database are available.
• Validate that third party software does not block hard-disk resources.
5-1-2 Procedure
1. Go to Settings tab "Local" and read the cache file path.
2. Open the cache file path in Windows Explorer.
3. Verify that the cache file path directs to an empty hard-disk partition. The only folder on the partition
shall be the local image cache.
5-1-3 If the test fails
1. Correct the location of the local cache.
2. Remove all third-party files from the cache partition.
Note:
MammoWorkstation displays labeling according to the IHE Mammography Image Profile.
Vendors not always encode the required information in the appropriate DICOM tags.
5-3-2 Procedure
For each connected mammography modality:
1. Open a case in MammoWorkstation with at least one CC and one MLO view of each breast.
2. Verify that the following labeling appears on each image:
• Name of patient (top row)
• Patient identifier (top row, in brackets)
• Date of examination (bottom center)
• View and laterality (position near the axilla)
3. Activate "Annotation" in the toolbar.
4. Verify that the following labeling appear on each image:
• Facility name and location. At a minimum, the location shall include the city, State and zip code of
the facility.
• Technologist identification.
• Detector ID or Cassette ID (used for Cassette/screen identification).
• Device Serial Number (used for Mammography unit identification).
5-3-3 If the test fails
1. Login to MammoWorkstation Settings and go to the tab "Preferences". Select "Reset to IHE
Standard".
2. Verify in the DICOM header of the images that the required DICOM tags are properly filled in.
To display the DICOM header, open the Supportinterface (http://localhost:8080/supportinterface),
navigate to "import processes" and search to the patient.
The required DICOM tags are listed in Dicom Conformance Statements (http://www.imagediagnost.de/
conformance/).
Test Completed /
Passed
Check that the TCP/IP ports are available for Storage SCP, database and webserver
(Acceptance test 5.2)
Check that measurements in true size are displayed with correct size
(Acceptance test 5.4)
Check correct operation of DICOM worklist interface with worklist server (e.g. RIS)
(Acceptance test 5.6)
1 Applicability
This job card applies to all PCs.
2 Personnel requirements
1 Field Engineer
Time: 15’
3 Backup to HDD
1. Create a folder on the HDD D (or C), e.g. BackupV4.5.x.
2. In this backup folder, copy the directories:
• ~\ImageDiagnost\CadServer
• ~\ImageDiagnost\config
• ~\imageDiagnost\MammoWorkstation.
Upgrade
Chapter 5 Upgrade
This chapter contains all information necessary to upgrade all IDI products.
In order to minimize the downtime of the workflow, we recommend to use the following sequence for the
sequence:
1. Upgrade Collaboration Server (old DSH, MWS, MDS will work with the new Collaboration Server
version).
2. Upgrade each DICOM Shuttle (old MWS, MDS will work with the new DICOM Shuttle version).
3. Upgrade each MDS.
4. Upgrade each MWS.
Note:
For each IDI-node, a new license shall be requested before the upgrade is started.
New license is not needed to perform the upgrade but needed to hand over system to the
customer.
1 Backup and preparation Job Card UPG02 - Software upgrade 1FE 20’
preparation on page 251
2 Request license Job Card UPG03 - License Request on 1FE 30’
page 255
3 Upgrade IDI application Job Card UPG04 - Basic software upgrade on 1FE 20’
page 259
4 Backup of the configuration folder Job Card UPG 07 - Backup of the configuration 1FE 15’
folder on page 261
5 Post-Upgrade checks Job Card UPG 08 - Post-Upgrade checks on 1FE 10’
page 263
6 Enable CESM feature Job Card CFG A46 - Enable CESM feature on 1FE 15’
page 237
Final sequence: required to complete the upgrade
7 Acceptance Test Job Card IST A32 - Acceptance test on 1FE 20’
page 241
Upgrade
1 Applicability
This job card applies to all PCs.
2 Personnel requirements
1 Field Engineer
Time: 20’
3 Hardware requirements
• v4.6.0 software CD.
It is therefore advisable to create a backup of the database before performing any tasks that might pose
a risk.
1. Go to Start ➜ Run, type "cmd" and click [OK].
2. Type c: and hit [Enter] to change to the c drive.
3. Type cd c:\mysql\bin and hit [Enter] to change directory to the folder where all the mysql tools
are stored.
Example:
1. Type:
mysqldump –u root –h localhost –-opt test_cache > d:\backup_test_cache.sql
Note:
“D:\” stands for the data hard disk. If your data hard disk is named differently, change the “D:\”
element according to your data hard disk.
Note:
If you are using a MDS, only one hard disk might be available.
2. Hit [Enter].
3. Continue with:
mysqldump –u root –h localhost –-opt workflow > d:\backup_workflow.sql
WARNING
Be careful! The dataset can be destroyed if you use the wrong sign!
! Notice:
The less-than sign [<] or greater-than sign [>] indicates in which direction the information flow
takes place: [<] into the database; [>] out of the database.
4. Hit [Enter].
The above executed actions will extract the content of the two databases "workflow" and "test_cache"
and store it into a text file named "backup_workflow.sql" and "backup_test_cache.sql".
5. Validate backup:
To verify the result, navigate to the directory where the database backup files have been stored on
hard disk D:
Note:
If you are using a MDS, only one hard disk might be available.
6. Press the Control button [Ctrl] and select the files "backup_test_cache.sql" and
"backup_workflow.sql".
7. Select Copy.
8. Go to the created path D:\YYYYMMDD_IDI_Backup\IDI_v4.x.x.
9. Right click ➜ select Paste.
6 Stop services
1. Right click on My Computer.
2. Click on Manage ➜ Computer Management ➜ Services and Applications ➜ Services.
3. Stop the services IDReceiveSCP and IDWorklistSCP one after another!
If the MWS Tray icon is visible (which means the MWS Tray is still running), follow the below indicated
procedure:
1. Open the Task Manager:
- Press [Ctrl] + [Shift] + [Esc]
or
- Go to Start ➜ Run, type "taskmgr" and click [OK].
2. Go to the tab Processes.
3. Make sure the box next to Show processes from all users is ticked (1)
4. Look for all processes where the column "Image Name" contains "MWSTray.exe".
5. Right click on each MWSTray.exe and select End Process Tree (2).
1 Applicability
This job card applies to all PCs.
2 Personnel requirements
1 Field Engineer
Time: 30’
3 Hardware requirements
• v4.6.0 software CD.
4 License Request
For every upgrade a new license needs to be requested to ensure the tracking of installed base and to
check the license-configuration.
Note:
If the Installation code has been kept from the previous v4.5.x installation, the Preliminary steps
don't need to be performed. You can directly request a new licence with the kept installation code
before you go on site.
Use the instructions in Job Card IST A13 - IDI application licensing on page 113 for the request.
Note:
If the MammoWorkstation is connected to Insite, the Preliminary steps can be performed remotely
before you go on site.
5 Preliminary steps
5-1 Licensed modules check
1. Login to MammoWorkstation ➜ Settings.
- name: "admin"
- password: "admin"
2. Select the License tab.
Note:
If not exaclty the same modules are installed during the update, the installation will fail.
5. Click [Save].
5. Request the new license as described in Job Card IST A13 - IDI application licensing on page 113.
1 Applicability
This job card applies to all PCs.
2 Personnel requirements
1 Field Engineer
Time:30’
3 Hardware requirements
• v4.6.0 software CD.
4 Prerequisites
1. No customization has been applied beyond the Configuration User Interface.
Note:
If customized configuration has been applied beyond the Configuration User Interface, contact the
GE Service Team.
2. Document the current customized configuration in the Configuration User Interface.
Note:
If the documentation of the configuration has not already been done, manually look through the
different configuration dialogs.
Make sure to note down every applied configuration!
3. Click [Next>].
The software is being installed.
4. Click [OK] to avoid installing MySQL.
1 Applicability
This job card applies to all PCs.
2 Personnel requirements
1 Field Engineer
Time:15’
3 Backup to HDD
Note:
After every upgrade it is mandatory to save the current configuration to a backup folder.
1 Applicability
This job card applies to all PCs.
2 Personnel requirements
1 Field Engineer
Time:10’
Expected result:
No bin folder is available.
1
2. Right click on the icon you want to delete and select "Delete".
Note:
Make sure to delete the icon that does not correspond to your installed system.
2. Right click on the Icon you want to delete and select "Remove from This List".
Note:
Make sure to delete the icon that does not correspond to your installed system.
Check that only one new program folder has been created.
Note:
If two program folders have been created, the program folder that does not correspond to your
installed system must be deleted.
Delete the program folder that does not correspond to your installed version:
1. Go to C:\Program Files\ImageDiagnost\MammoWorkstation
2. Delete the x64 folder.
Is needed
for 64-bit
version
Is needed
for 32-bit
version
Chapter 6 Uninstallation
1 Applicability
This job card applies to all PCs.
.
2 Personnel requirements
1 Field Engineer
Time:
2. Stop (1):
- ApacheTomcat6forIDI
- IDDicomReceiveSCP
- IDDicomWorklistSCP
- MySql
- IDRMIRegistry
Note:
Stopping the IDRMIRegistry service should also stop the MWS Tray. If the MWS Tray does not
stop, continue with section 4 MWS Tray stop.
5 Configuration of IP address
1. Go to Start ➜ Control panel ➜ Add or Remove Programs.
2. Select "Image Diagnost Software" and click the button [Change/Remove].
7 Verification
Manually check that all MammoWorkstation services are removed:
1. Go to Control Panel ➜ Administrative Tools ➜ Services.
Note:
The services "ApacheTomcat6forIDI", "IDDicomReceiveSCP" and "IDDicomWorklistSCP" must
be removed.
Manufactured by:
GE Medical Systems SCS
283, rue de la Minière - BP 34
78533 Buc Cedex
France
Imagination at work